USER’S MANUAL V1 - Rev. 01/2002
CD Automation S.r.l. Via F.lli Cervi, 42/44 20020 Cantalupo di Cerro Maggiore (M) - Italy Tel +39 0331 533512 – Fax +39 0331 533516 e-mail: [email protected] web: www.cdautomation.com
APPLICATION NOTES
KANTHAL SUPER MOLIBDENUM
DI SILICIDE MoSi2
C.D. Automation srl Application Note-Kantal Super Molibdenum V 1
www.cdautomation.com 1
KANTHAL SUPER MOLIBDENUM DI SILICIDE MOSI2
It’s an electric heating element material for use in industrial and
laboratory furnaces. It can be used in air to heat furnaces to a
temperature as high as 1850 °C.
These resistance has a particular feature
• The resistivity of Kanthal super element increases sharply with
temperature and when is cold can have a current that is up to 17
times, the nominal current if is not limited. It has a strong positive
coefficient of resistance 17:1 from ambient to working temperature
• Thus a current limit thyristor unit with phase angle must be used like:
• CD 2200 – CD 3200: for one phase loads
• CD 3000E – MULTIDRIVE: for three phase loads
• As Kanthal Super elements
do not increase in resistance
with use, there is no need
for a voltage reserve, and
should one fail, it may be
replaced without affecting
the others in circuit.
C.D. Automation srl Application Note-Kantal Super Molibdenum V 1
www.cdautomation.com 2
• These Kanthal Super elements are fragile at ambient temperature
and soft when are working at high temperature.
• MoSi2 elements are manufactured in the form of giant hair pins. The
force between the arms due to current flowing is proportional to the
square of current. When cold, if current is not limited, the force can
reach 100 times the normal one damaging the elements.
• It’s also necessary to use Phase Angle plus current limit to reduce
the thermal shock that can damage the furnace building materials.
• Due to this reasons also if we size a thyristor unit or a contactor to
supply cold elements we can damage them and the furnace building
materials. In the past when where not used thyristor units was
necessary to use multi-tapped or continuously variable transformers
and, in this case, human errors in forgetting part of the start
procedure was possible. With thyristor every think is done
automatically.
C.D. Automation srl Application Note-Kantal Super Molibdenum V 1
www.cdautomation.com 3
IMPORTANT
With MoSi2 elements due to the very low cold resistance the inrush current
is very quick so we suggest:
• Soft start (setted ≅ 2 sec)
• Current limit
• Phase Angle Firing
CD Automation Thyristor unit can accept an external current limit profiling
signal. This is an important feature because is possible to have a density
of circulating current up to 1200°C and after a linear decreasing up
1700°C.
For example it’s possible to use:
20W/cm2 from O : 1200°C
10W/cm from 1200 : 1700°C
This external signal can be given by a temperature controller with
retransmission of Process Variable.
The controller can be configured by front to have
From 0 : 1200°C → 10V Retransmission
From 1200 : 1700°C → 5V Retransmission
This output from controller can be connected on the external current
profiling of CD Automation Thyristor Unit.
This feature is also known as Dallas profiling.
C.D. Automation srl Application Note-Kantal Super Molibdenum V 1
www.cdautomation.com 4
L2 or N
L1
FiringCircuit
Main Supply
CurrentLinitProfiling
1200
CurrentLimit
V
1400
C.D. Automation srl Application Note-Kantal Super Molibdenum V 1
www.cdautomation.com 5
Sometimes the elements MoSi2 must be supplied at low voltage via a
transformer that can be one or three phase, depending on load
configuration.
There are no particular problems for thyristor units and when a
transformer is coupled with MoSi2 and the thyristors are connected on
primary, use the same precaution mentioned above for the MoSi2 powered
directly.
If a three phase transformer is used the voltage on the three phase must
be keepad balanced and this will lead certainly to an imbalance of
currents.
Power factor is low when the voltage is low with cold resistance and
increase with resistance value of element.