Page 1: Analysis of IMS Data

Analysis of IMS Data

By Ed Sittler

CAPS Team Meeting #28

Marseille, France

July 26-28, 2004

Page 2: Analysis of IMS Data

Analysis of TOF Data• Purpose:

– Determine ion composition

– Revise calibration parameters for FM

• Peak shifts

• Peak Amplitudes and widths

• Algorithm

– Read routine which sums bcycle data in 2D arrays for ST & LEF.

• Use highest data rate modes (I.e., 4, 8, 16 kbps).

• Set start and stop times.

• Store data in file.

– Read file from previous step.

• Make 2D color spectrograms.

• Make line plots of TOF data for specific estep.

– Perform non-linear least-squares fit to TOF data.• Can do up to 10 ions. Use line profile parameters as given. Only adjust ion counts.

• Presently we can shift the ion peaks manually. We also fit simultaneously to the background.

• First do manual fit.

• Then do non-linear fit. Compute chisqr.

• Plot Results.

• Further generalizations being considered.

Page 3: Analysis of IMS Data
Page 4: Analysis of IMS Data
Page 5: Analysis of IMS Data
Page 6: Analysis of IMS Data
Page 7: Analysis of IMS Data
Page 8: Analysis of IMS Data
Page 9: Analysis of IMS Data
Page 10: Analysis of IMS Data
Page 11: Analysis of IMS Data
Page 12: Analysis of IMS Data
Page 13: Analysis of IMS Data

ST TOF Spectrum 182(23:24) No Shift

Page 14: Analysis of IMS Data

ST TOF Spectrum 182(23:24) Shifted

Page 15: Analysis of IMS Data

LEF TOF Spectrum 182(23:24)

Page 16: Analysis of IMS Data

ST TOF Spectrum 182(18:24) No Shift

Page 17: Analysis of IMS Data

ST TOF Spectrum 182(18:24) Shifted

Page 18: Analysis of IMS Data

LEF TOF Spectrum 182(18:24)

Page 19: Analysis of IMS Data

ST TOF Spectrum 182(18:24) E = 30 eV

Page 20: Analysis of IMS Data

LEF TOF Spectrum 182(18:24) E = 30 eV

Page 21: Analysis of IMS Data

Summary of TOF Analysis Results

• Composition of H+, N+, Water Group Ions & O2+. Detection of H2+ ?

• Shift ~ 16 channels (2048) seems to be required.• With no shift H3O+ tends to dominate, C+ more important and N+ less

important.• With shift H2O+ tends to dominate, N+ more important and C+ less

important.• LEF data shows H+, N+ and Water Group Ions. No C+ seen. O2

+ can produce peak near N+ peak, but to shorter TOF. Therefore, less favored by data. Although O2

+ is present it is not sufficiently intense.• LEF data set ideal for detection of N+ since O+ has low sensitivity in

LEF data (I.e., O+ tends to want electrons).• ST O- peak shoulder to lower TOF not understood.• Electronic peaks? Remove bad data (I.e., revise script pgm).

Page 22: Analysis of IMS Data

Analysis of Ion Data

• Purpose:– Determine ion composition.– Determine distribution functions of individual ion species.– Compute ion moments.– Study performance of SAM algorithm and Group Tables.– Incorporate revised FM cal parameters into ion analysis. Reconstruct ion counts

from data. Required high rate data (4, 8 and 16 kbps)

• Procedure– Read level0 data products for period 182(18:24). Sum data for various ion species

as function of estep and angular sector.– Study statistical properties and search for systematic effects as function of estep and

angular sector.– Plot data in appropriate format.– Apply non-linear fitting algorithm which uses revised FM cal data derived from

analysis of TOF data. Reconstruct true ion counts for all ion species. Manually, we have already been doing this.

Page 23: Analysis of IMS Data
Page 24: Analysis of IMS Data
Page 25: Analysis of IMS Data
Page 26: Analysis of IMS Data
Page 27: Analysis of IMS Data
Page 28: Analysis of IMS Data
Page 29: Analysis of IMS Data
Page 30: Analysis of IMS Data
Page 31: Analysis of IMS Data
Page 32: Analysis of IMS Data

Summary of Ion Data Analysis• Overall ion data looks good.

– But, RTIU problem, especially within inner magnetosphere caused us to throw away a lot of the data. Script program helped but then bad data leaked into results. Localized spikes in data. Needs further work.

• Definite anti-correlations with estep between OH+ & H2O+

– Interpolation between esteps may cause systematic effects.– If there were no systematic effects than anti-correlations would go away

by summing over many spectra.– Adding nearby esteps may remove some of the problem.– Going to hardware binning should definitely fix the problem.

• Summing water group ions together removes anti-correlations. This was expected.

• Signal peaks in sector 2 as expected. Enhancement in sector 8 indicated pressure anisotropy effect?

• We will next need to apply non-linear fitting algorithm to reconstruct ion counts. Here, we would use revised FM cal parameters from analysis of TOF data.