analysis of electrolyte solution in lithium ion …...no.m264a n results of electrolyte solution...

LAAN-A-MS-E030A No. M264A Gas Chromatography Mass Spectrometry Analysis of Electrolyte Solution in Lithium Ion Rechargeable Battery (LIRB) and Evolved Gas from LIRB The electrolyte solution in a lithium ion rechargeable battery consists of an organic solvent (consisting chiefly of carbonates), the electrolyte, and additives. Evaluation of the electrolyte solution and analysis of its deterioration due to repetitive charge/discharge cycles is important for the further development of the lithium ion rechargeable battery. GC-MS is applicable for composition analysis of the electrolyte solution, and for analysis of components in the electrolyte solution that have been generated due to repeated charging and discharging. This Application News introduces examples of analysis of the electrolyte solution of a lithium ion rechargeable battery and of the gas evolved from a cell that has been maintained at a high temperature. n Analytical Conditions Analysis of the electrolyte solution was conducted using the analytical conditions of Table 1, and analysis of gas evolved from the cell was conducted using the conditions of Table 2. The evolved gas to be measured was collected from an aluminum-laminated lithium ion battery that had been stored for 5 days at 80 ºC. The needle tip of a gas-tight syringe was inserted directly into the cell, and the gas phase was aspirated into the syringe. Model : GCMS-QP2010 Ultra Column : Rtx-200MS (RESTEK Corporation) [30 m × 0.25 mm I.D. df = 1 μm] -GC- -MS- Injector Temp. : 250 °C Interface Temp. : 250 °C Injection Method : Split Ion Source Temp. : 200 °C Split Ratio : 1: 100 Ionization Mode : EI Carrier Gas : He (Constant Linear Velocity Mode) Acq. Mode : Scan Linear Velocity : 40 cm/sec Scan Range : m/z 35-500 Column Temp. : 40 °C (3 min) - 8 °C/min - 280 °C (5 min) Event Time : 0.3 sec Injection Volume : 1 μL Model : GCMS-QP2010 Ultra Column : Rt-Q-BOND (RESTEK Corporation) [30 m × 0.32 mm I.D. df = 10 μm] + Guard column (RESTEK Corporation) [3 m × 0.32 mm I.D.] -GC- -MS- Injector Temp. : 200 °C Interface Temp. : 200 °C Injection Method : Split Ion Source Temp. : 200 °C Split Ratio : 1: 30 Ionization Mode : EI Carrier Gas : He (Constant Linear Velocity Mode) Acq. Mode : Scan Linear Velocity : 61.6 cm/sec Scan Range : m/z 10-300 Column Temp. : 35 °C (3 min) - 10 °C/min - 260 °C (5 min) Event Time : 0.3 sec Injection Volume : 500 μL Table 1 Analytical Conditions for Electrolyte Solution Analysis Table 2 Analytical Conditions for Evolved Gas Analysis

Upload: others

Post on 27-Feb-2020

5 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Analysis of Electrolyte Solution in Lithium Ion …...No.M264A n Results of Electrolyte Solution Analysis Fig. 1 shows the results of analysis of the lithium ion rechargeable battery

LAAN-A-MS-E030A

No.M264AGas Chromatography Mass Spectrometry

Analysis of Electrolyte Solution in Lithium Ion Rechargeable Battery (LIRB) andEvolved Gas from LIRB

The electrolyte solution in a lithium ion rechargeable battery consists of an organic solvent (consisting chiefly of carbonates), the electrolyte, and additives. Evaluation of the electrolyte solution and analysis of its deterioration due to repetitive charge/discharge cycles is important for the further development of the lithium ion rechargeable battery. GC-MS is applicable for composition analysis of the electrolyte solution,

and for analysis of components in the electrolyte solution that have been generated due to repeated charging and discharging.This Application News introduces examples of analysis of the electrolyte solution of a lithium ion rechargeable battery and of the gas evolved from a cell that has been maintained at a high temperature.

n Analytical ConditionsAnalysis of the electrolyte solution was conducted using the analytical conditions of Table 1, and analysis of gas evolved from the cell was conducted using the condi t ions of Table 2. The evolved gas to be measured was collected from an aluminum-laminated

lithium ion battery that had been stored for 5 days at 80 ºC. The needle tip of a gas-tight syringe was inserted directly into the cell, and the gas phase was aspirated into the syringe.

Model : GCMS-QP2010 UltraColumn : Rtx-200MS (RESTEK Corporation) [30 m × 0.25 mm I.D. df = 1 μm]-GC- -MS-

Injector Temp. : 250 °C Interface Temp. : 250 °CInjection Method : Split Ion Source Temp. : 200 °CSplit Ratio : 1 : 100 Ionization Mode : EICarrier Gas : He (Constant Linear Velocity Mode) Acq. Mode : ScanLinear Velocity : 40 cm/sec Scan Range : m/z 35-500Column Temp. : 40 °C (3 min) - 8 °C/min - 280 °C (5 min) Event Time : 0.3 secInjection Volume : 1 μL

Model : GCMS-QP2010 UltraColumn : Rt-Q-BOND (RESTEK Corporation) [30 m × 0.32 mm I.D. df = 10 μm]

+ Guard column (RESTEK Corporation) [3 m × 0.32 mm I.D.]-GC- -MS-

Injector Temp. : 200 °C Interface Temp. : 200 °CInjection Method : Split Ion Source Temp. : 200 °CSplit Ratio : 1 : 30 Ionization Mode : EICarrier Gas : He (Constant Linear Velocity Mode) Acq. Mode : ScanLinear Velocity : 61.6 cm/sec Scan Range : m/z 10-300Column Temp. : 35 °C (3 min) - 10 °C/min - 260 °C (5 min) Event Time : 0.3 secInjection Volume : 500 μL

Table 1 Analytical Conditions for Electrolyte Solution Analysis

Table 2 Analytical Conditions for Evolved Gas Analysis

Page 2: Analysis of Electrolyte Solution in Lithium Ion …...No.M264A n Results of Electrolyte Solution Analysis Fig. 1 shows the results of analysis of the lithium ion rechargeable battery

SHIMADZU CORPORATION. International Marketing Division3. Kanda-Nishikicho 1-chome, Chiyoda-ku, Tokyo 101-8448, Japan Phone: 81(3)3219-5641 Fax. 81(3)3219-5710

No.M264A

n Results of Electrolyte Solution AnalysisFig. 1 shows the results of analysis of the lithium ion rechargeable battery electrolyte solution. The dimethyl carbonate, ethyl methyl carbonate and ethylene

carbonate used as the solvent were identified from a mass spectral library search. In addition, vinylene carbonate used for the additive was also identified.

Dimethyl carbonate (solvent)

Ethyl methyl carbonate (solvent)

Vinylene carbonate (additive)

Ethylene carbonate (solvent)

Total Ion Current Chromatogram15,000,000

4.0 10.0 19.0min

Dim

ethy

l car

bona

te

Ethy

l met

hyl c

arbo

nate

Viny

lene

car

bona

te

Ethy

lene

car

bona

te

50,000,000

1.0 10.0 20.0 25.0 1.0 2.0 3.0min

Wat

er

Prop

ane

Buta

ne

Ethy

l met

hyl c

arbo

nate

(Sol

vent

)

Die

thyl

car

bona

te(S

olve

nt )

Air +

CO

Met

hane

Ethy

lene

Car

bon

diox

ide

Etha

ne

Ethy

l flu

orid

e

Total Ion Current Chromatogram Total Ion Current Chromatogram (Enlarged)

Fig. 1 Total Ion Current Chromatogram (TICC) of Electrolyte Solution and Mass Spectra of Target Compounds

n Analytical Results for Evolved Gas from LIRBThe results obtained from analysis of the gas evolved from the cell are shown in Fig. 2. Many substances that originated from the degenerated solvent and addit ive, as well as from the electrolyte, were

identified. Moreover, a fluorinated compound of electrolyte origin was also identified in the library search.

Fig. 2 Total Ion Current Chromatogram (TICC) of Evolved Gas from Lithium Ion Rechargeable Battery

n Analytical NoteA PLOT column sol id phase consis t ing of an adsorbent or porous polymer is suitable for analysis of gases and low-boi l ing compounds. However, compared to a capillary column that uses a liquid phase, there is a greater risk of particle scattering,

resulting in clogging or ion source contamination. It is therefore necessary to connect a 2 to 3 m guard column at the interface end of the PLOT column to lessen the risks associated with particle scattering.