Improved Sensitivity in LA-ICP-MS Applications through the use of Mixed Gas Plasmas and Collisional Focusing Julian D. Wills, Christoph Wehe and Shona McSheehy Ducos Thermo Fisher Scientific, Bremen, Germany Application Note 43261 Key Words Laser Ablation, ICP-MS, iCAP Q, Collision Cell technology Goal To demonstrate the sensitivity improvement possible in LA-ICP-Q-MS through the use of mixed plasma gases and advanced collision cell techniques. Introduction Laser ablation (LA) sampling has continually taken advantage of instrument based advancements in inductively coupled plasma mass spectrometry (ICP-MS) to improve analytical performance. For example: 1. Collision and reaction cells for the reduction of spectral interferences. 2. High mass resolution (ICP-SF-MS) for the ultimate in performance. 3. Improved interface designs leading to increased robustness and reduced contamination in dry plasma applications. For many LA-ICP-MS applications however, it’s the pursuit of higher instrumental sensitivity that’s the most interesting. When performing discrete LA-ICP-MS analyses improved instrumental sensitivity enables the use of increasingly smaller laser spots. Smaller spots are advantageous in many applications, for example in detrital zircon studies in order to maximize the number of analyzable zircons in provenance studies or to improve spatial resolution in elemental mapping analyses in both geological and biomedical studies. This paper demonstrates the sensitivity improvement possible in LA-ICP-Q-MS through the use of mixed plasma gases and advanced collision cell techniques. The Thermo Scientific iCAP Q ICP-MS system was used to perform laser ablation ICP-MS (LA-ICP-MS) analyses using the standard Ar/He plasma, a N 2 doped Ar/He plasma and a N 2 doped Ar/He plasma in combination with the collision cell pressurized with He. Instrument Settings All analyses were made using a combination of the Teledyne Photon Machines Analyte G2 Excimer Laser Ablation with the iCAP Q ICP-MS. Both the iCAP Q ICP-MS and the Analyte G2 were controlled by the Thermo Scientific Qtegra Intelligent Scientific Data Solution (ISDS) Software. A dedicated laser ablation dual mode sample introduction system was used to introduce the combination of He from the laser and additional argon or argon/nitrogen mixtures to the plasma. Figure 2 summarizes the analytical parameters used for the laser ablation system and the iCAP Q ICP-MS. Figure 1: The Thermo Scientific iCAP Q ICP-MS Dual Mode Sample Introduction System for Laser Ablation Applications

Upload: vantuyen

Post on 29-Apr-2018

216 views

Category:

Documents


3 download

TRANSCRIPT

Page 1: AN43261 - Improved Sensitivity in LA-ICP-MS Applications ... · Qtegra ISDS Software provides plug-in based support of ... Norway +46 8 556 468 00 Russia/CIS +43 1 333 50 34 0 Singapore

Improved Sensitivity in LA-ICP-MS Applications through the use of Mixed Gas Plasmas and Collisional Focusing Julian D. Wills, Christoph Wehe and Shona McSheehy Ducos Thermo Fisher Scientific, Bremen, Germany

Ap

plica

tion

No

te 4

32

61

Key WordsLaser Ablation, ICP-MS, iCAP Q, Collision Cell technology

Goal To demonstrate the sensitivity improvement possible in LA-ICP-Q-MS through the use of mixed plasma gases and advanced collision cell techniques.

IntroductionLaser ablation (LA) sampling has continually taken advantage of instrument based advancements in inductively coupled plasma mass spectrometry (ICP-MS) to improve analytical performance. For example:

1. Collision and reaction cells for the reduction of spectral interferences.

2. High mass resolution (ICP-SF-MS) for the ultimate in performance.

3. Improved interface designs leading to increased robustness and reduced contamination in dry plasma applications.

For many LA-ICP-MS applications however, it’s the pursuit of higher instrumental sensitivity that’s the most interesting. When performing discrete LA-ICP-MS analyses improved instrumental sensitivity enables the use of increasingly smaller laser spots. Smaller spots are advantageous in many applications, for example in detrital zircon studies in order to maximize the number of analyzable zircons in provenance studies or to improve spatial resolution in elemental mapping analyses in both geological and biomedical studies. This paper demonstrates the sensitivity improvement possible in LA-ICP-Q-MS through the use of mixed plasma gases and advanced collision cell techniques.

The Thermo Scientific™ iCAP™ Q ICP-MS system was used to perform laser ablation ICP-MS (LA-ICP-MS) analyses using the standard Ar/He plasma, a N2 doped Ar/He plasma and a N2 doped Ar/He plasma in combination with the collision cell pressurized with He.

Instrument SettingsAll analyses were made using a combination of the Teledyne Photon Machines Analyte G2™ Excimer Laser Ablation with the iCAP Q ICP-MS. Both the iCAP Q ICP-MS and the Analyte G2 were controlled by the Thermo Scientific™ Qtegra™ Intelligent Scientific Data Solution™ (ISDS) Software. A dedicated laser ablation dual mode sample introduction system was used to introduce the combination of He from the laser and additional argon or argon/nitrogen mixtures to the plasma. Figure 2 summarizes the analytical parameters used for the laser ablation system and the iCAP Q ICP-MS.

Figure 1: The Thermo Scientific iCAP Q ICP-MS Dual Mode Sample Introduction System for Laser Ablation Applications

Page 2: AN43261 - Improved Sensitivity in LA-ICP-MS Applications ... · Qtegra ISDS Software provides plug-in based support of ... Norway +46 8 556 468 00 Russia/CIS +43 1 333 50 34 0 Singapore

2

ResultsIn order to compare instrumental sensitivities in each of the three analysis modes, five replicate analyses of NIST 610 were performed using the laser parameters shown in Figure 2.

An automated tuning procedure (using NIST 610) was performed in each mode in order to maximize sensitivity while minimizing oxide formation (ThO/Th). Figures 3-5 show the gas blank subtracted intensities for NIST 610 LA-ICP-MS analyses in each of the three analysis modes tested.

Figure 3 presents the data for an Ar plasma with He from the laser cell. With a standard analysis mode that combines the particle containing He stream from the laser cell with Ar make-up gas1 from the iCAP Q ICP-MS, sensitivities of > 1 Mcps are achieved for elements in NIST 610 (~450 µg·g-1) at a spot size of 40 µm and a fluence of 3.2 J·cm-2.

Figure 4 presents the data for an Ar plasma with He from the laser cell and an additional flow of N2 at 3 mL·min-1 to the Ar/He stream. By adding a small flow of nitrogen to the Ar/He mixture2, sensitivities across the entire mass range are improved by approximately 30%, with the largest gains made for the lighter masses.

Figure 3. Ar/He plasma.

Figure 4. N2 doped Ar/He plasma.

Figure 5. N2 doped Ar/He plasma, He collisional focusing.

Figure 5 presents the data for an Ar plasma with He from the laser cell, the additional flow of N2 to the Ar/He stream and the use of collisional focusing in the iCAP Q ICP-MS collision/reaction cell (QCell). With a low He flow at 4 mL·min-1 added to the iCAP Q ICP-MS QCell, sensitivities for the LA-ICP-MS analysis of NIST 610 are further improved with > 3 Mcps recorded for a spot size of 40 µm at a fluence of 3.2 J·cm-2.

Collisional FocusingCollisional focusing is theoretically possible with any collision cell, but works very effectively with the iCAP Q ICP-MS QCell due to the inner dimensions and volume of the cell. With collisional focusing any ions that would otherwise be lost by beam expansion in an empty (no gas) collision cell are better focused / transmitted in a (He) pressurized cell, resulting in higher instrumental sensitivities.

The sensitivity improvement possible with collisional focusing is mass dependent. Heavier masses, due to their increased numbers of collisions with He, have a lower energy in a pressurized cell. This prolongs their residence time in the cell allowing them to be better focused leading to improved transmission.

Figure 2. LA-ICP-MS instrument settings.

Analyte G2 Laser Ablation System:

• 40 µm spot • 30% of 7 mJ • 3.2 J.cm2 • 4 Hz • 70 s per site • 20 s gas blank • 30 s ablation • 0.95 L.min-1 He • 3 mL.min-1 N

2

iCAP Q ICP-MS:

• High sensitivity insert • Ni cones • 41 isotopes • 25 ms persotope

Page 3: AN43261 - Improved Sensitivity in LA-ICP-MS Applications ... · Qtegra ISDS Software provides plug-in based support of ... Norway +46 8 556 468 00 Russia/CIS +43 1 333 50 34 0 Singapore

Ap

plica

tion

No

te 4

32

61

AN43261-EN 0616S

Driven by Qtegra ISDS SoftwareQtegra ISDS Software provides plug-in based support of all major laser ablation systems from ESI, TELEDYNE CETAC, and TELEDYNE PHOTON MACHINES. A dedicated data evaluation for laser ablation analyses (trQuant) with a range of integration and quantification features comes as standard in Qtegra ISDS software. Additionally, multiple export formats, including a format for export into the Iolite program are a standard part of any iCAP Q ICP-MS system running on Qtegra ISDS software.

ConclusionThrough the use of N2 mixed gas plasma and He collisional focusing, sensitivity improvements of over a factor of three are possible using the Thermo Scientific iCAP Q ICP-MS system. With such a significant sensitivity improvement, smaller laser spot sizes can be routinely used opening up new LA-ICP-MS applications in geological and biomedical studies.

References1. D. Günther and C. Heinrich, J. Anal. At. Spectrom., 14,

1999, 1363–1368.

2. Z. Hu, S. Gao, Y. Liu, S. Hu, H. Chen, and H. Yuan, J. Anal. At. Spectrom., 2008, 23, 1093.

Figure 6. Relative sensitivities improvements for mixed gas plasma and collisional focusing.

The sensitivity changes for N2 doped Ar/He and N2 doped Ar/He with He collisional focusing relative to a standard Ar/He plasma are shown in Figure 6.

Through a combination of mixed gas plasma and QCell flatapole collision cell technology, 3 fold sensitivity improvements can be achieved with the Thermo Scientific iCAP Q ICP-MS system.

Africa +43 1 333 50 34 0Australia +61 3 9757 4300Austria +43 810 282 206Belgium +32 53 73 42 41Canada +1 800 530 8447China 800 810 5118 (free call domestic)

400 650 5118

Denmark +45 70 23 62 60Europe-Other +43 1 333 50 34 0Finland +358 10 3292 200France +33 1 60 92 48 00Germany +49 6103 408 1014India +91 22 6742 9494Italy +39 02 950 591

Japan +81 45 453 9100Korea +82 2 3420 8600Latin America +1 561 688 8700Middle East +43 1 333 50 34 0Netherlands +31 76 579 55 55New Zealand +64 9 980 6700Norway +46 8 556 468 00

Russia/CIS +43 1 333 50 34 0Singapore +65 6289 1190Spain +34 914 845 965Sweden +46 8 556 468 00Switzerland +41 61 716 77 00UK +44 1442 233555USA +1 800 532 4752

www.thermofisher.com©2016 Thermo Fisher Scientific Inc. All rights reserved. Analyte G2 is a trademark of Teledyne Photon Machines, CETAC is a trademark of Teledyne CETAC Technologies, ESI is a trademark of ELECTRO SCIENTIFIC INDUSTRIES, INC. CORPORATION. NIST is a Service Mark of National Institute of Standards and Technology AGENCY OF THE GOVERNMENT UNITED STATES. All other trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is presented as an example of the capabilities of Thermo Fisher Scientific products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details.