An improved treatment of the linearity correction of IR detectors Massimo Robberto JWST/NIRCam STScI TIPS – Sep. 16, 2010

Upload: hesper

Post on 23-Feb-2016

33 views

Category:

Documents


0 download

DESCRIPTION

An improved treatment of the linearity correction of IR detectors. Massimo Robberto JWST/ NIRCam STScI TIPS – Sep. 16, 2010. Ouverture. IR detectors are non linear. Linearity is assumed at the beginning of the ramp. linear fit to the first 20 samples. - PowerPoint PPT Presentation

TRANSCRIPT

Page 1: An improved treatment of the linearity correction of IR detectors

An improved treatment of the linearity correction of IR detectors

Massimo RobbertoJWST/NIRCam

STScI TIPS – Sep. 16, 2010

Page 2: An improved treatment of the linearity correction of IR detectors

OUVERTURE

Page 3: An improved treatment of the linearity correction of IR detectors

IR detectors are non linear

Page 4: An improved treatment of the linearity correction of IR detectors

Linearity is assumed at the beginning of the ramp

linear fit to the first 20 samples

Page 5: An improved treatment of the linearity correction of IR detectors

The “true” slope depends on the range of the assumed linear regime

Page 6: An improved treatment of the linearity correction of IR detectors

In fact, the angular coefficient of the true slope is hard to find…

Page 7: An improved treatment of the linearity correction of IR detectors

ACT 1CURRENT STATUS

Page 8: An improved treatment of the linearity correction of IR detectors

How we do it now

In the case of NICMOS and WFC3, we apply the following correction

F are the measured counts Fc are the true counts. The calibration process assumes that they are known (fit to the first part of the ramp). Known both F’s, we derive the correction coefficients c2, c3 and c4 used for general linearity correction.

Page 9: An improved treatment of the linearity correction of IR detectors

Problems with this approach

1) We do not really know what is the real slope of the calibration frame, and our estimate depends on the samples we use.

2) Physically, one has a linear true flux which is converted in a non-linear measured count rate by the detector. This is not what we model!

We modulate the observed data to get the real flux; instead, we should modulate the real flux to get the observed data.

Page 10: An improved treatment of the linearity correction of IR detectors

A controlled experiment using simulated data

THIS IS THE WEIRD (NON POLYNOMIAL)NON-LINEARITY TERM

Page 11: An improved treatment of the linearity correction of IR detectors

Let’s plot our baseline…

Page 12: An improved treatment of the linearity correction of IR detectors

… and derive the correction “a’la HST”

I will assume that we know perfectly the true slope, i.e. problem 1 has been solved. I therefore get the best possible c coefficients.

THIS IS THE POLYNOMIALCORRECTION TERM

Page 13: An improved treatment of the linearity correction of IR detectors

The result is:

Page 14: An improved treatment of the linearity correction of IR detectors

Residuals

Page 15: An improved treatment of the linearity correction of IR detectors

ACT 2A DIFFERENT APPROACH

Page 16: An improved treatment of the linearity correction of IR detectors

Let’s look at the equationInstead of

We can try with the physically more correct expression:

i.e. we modulate the real flux Fc to get F, not viceversa

Fc × 1+ c2 × Fc + c3 × Fc

2 + c4 × Fc3( )=F

Fc =F× 1+ c2 × F+ c3 × F2 + c4 × F3( )

Page 17: An improved treatment of the linearity correction of IR detectors

MethodIn Equation

the Fc and c2,c3,c4 values are unknown. I use IDL/curvefit.pro to derive them from the set of known ti and measured Fi:

having defined the function:

Fc × 1+ c2 × Fc + c3 × Fc

2 + c4 × Fc3( )=F

0.3% error on the slope!

Page 18: An improved treatment of the linearity correction of IR detectors

Linearity correctionFrom the values of c2, c3, an c3 one can derive Fc by solving the equation:

Need to use an iterative method:

Fc =

F1+ c2 × Fc + c3 × Fc

2 + c4 × Fc3( )

Page 19: An improved treatment of the linearity correction of IR detectors

Results

i=0

1

2

4

Page 20: An improved treatment of the linearity correction of IR detectors

Check: different flux rate

Same “detector”, i.e. exponential non-linearity term

Page 21: An improved treatment of the linearity correction of IR detectors

Correction: old vs. new method

Old

New

Page 22: An improved treatment of the linearity correction of IR detectors

ConclusionThe current strategy we implement to correct for non-linearity seems less than ideal.1) Problems with the estimate of the coefficients, which depend

on the assumed “linearity” region of the detector2) Problems with the equation, which does not correctly

describes the non-linearity effect

The new method has two advantages3) Coefficients are estimated without any assumption on the

true, linear flux4) The correct equation, with an iterative solve, seems to

provide a much better estimate of the true linear flux.

Check on real data is in progress