an easy-to-use precision power analyser. - mhz … · an easy-to-use precision power analyser. ......

12

Upload: lynga

Post on 11-Aug-2018

219 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: An easy-to-use Precision Power Analyser. - MHz … · An easy-to-use Precision Power Analyser. ... Many loads, such as a rectifier with capacitor filter produce a distorted current
Page 2: An easy-to-use Precision Power Analyser. - MHz … · An easy-to-use Precision Power Analyser. ... Many loads, such as a rectifier with capacitor filter produce a distorted current

2

An easy-to-use Precision Power Analyser .The touch of a button gives direct readout oftrue power and a host of other usefulmeasurements .

Easy to use-fully autoranging forvoltage, current and frequency

Accurate measurements of realpower (W) and apparent power (VA)even with distorted waveforms

Displays real power factor

True RMS measurement of voltage(2V to 700V AC) and current(20mA to 20A AC)

Computes the crest factor (Pk/RMS)of input voltage and load current

• Harmonic analysis of current andvoltage waveforms for checkingconformance to specifications

Wide frequency range - from DC to 20KHz

Determines peak inrush current ofmotors, transformers, power suppliesand similar loads

Page 3: An easy-to-use Precision Power Analyser. - MHz … · An easy-to-use Precision Power Analyser. ... Many loads, such as a rectifier with capacitor filter produce a distorted current

Applications

wMeasures accurately the power absorbed by a load ordelivered from an AC power source, even with distortedwaveforms . The instrument is (AC + DC) coupled andtherefore measures the total real power .

Connect the PM 1000 as shown in the appropriateconnection diagram and select'W' . The instrument isfullyautoranging and the displaywill read the actualpower in watts .

VA

Real Power

Apparent PowerDetermines the apparent power absorbed by a load ordelivered from a source . Connect the instrument as forpower measurements and select VA' .

The apparent power is computed as VA = Vrms x Armsand displayed in volt-amperes .

PF Power FactorDisplays the true power factor of an AC load . Powerfactor is not cos 4 except for undistorted waveforms ;cos $ is merely the displacement factor of the load, ameasure of the phase shift between the voltage andcurrent waveforms .

If the current waveform is distorted, the fundamentalcomponents of voltage and current may well be in phase,with cos 4 = 1, but the true power factor will be less thanunity . The PM 1000 computes the ratio W/VA, which isthe true power factor even for distorted waveforms .

Vrms RMS VoltageDetermines the true rms value of the input voltage, evenwith distorted waveforms . The input is (AC + DC)coupled, and the instrument will record the rms value ofthe combined AC and DC components . Only the voltageinput leads need be connected for Vrms measurements .

3

Page 4: An easy-to-use Precision Power Analyser. - MHz … · An easy-to-use Precision Power Analyser. ... Many loads, such as a rectifier with capacitor filter produce a distorted current

4

ArmsDetermines the true rms value of the load currentirrespective of waveform . As for Vrms, the input is(AC + DC) coupled, and displays the total RMS value .

Only the current input need be connected when takingmeasurements of Arms .

Vcf

RMS Current

Voltage CrestFactor

Display the ratio VpkNrms, where Vpk is the magnitudeof the repetitive peak voltage, irrespective of polarity . Thevalue of Vcf for a pure sine-wave is 1 .41 .

A non-linear load will cause distortion of the input voltagewaveform, resulting in a reduction in the value of voltagecrest factor .

A measurement of voltage crest factor thereby provides auseful indication of the effect of a load on the AC supply .

Acf Current CrestFactor

Displays the ratio Apk/Arms, where Apk is the magnitudeof the repetitive peak current, irrespective of polarity .

If the load takes a sinusoidal current, Acf = 1 .41 (i)Many loads, such as a rectifier with capacitor filterproduce a distorted current waveform (ii) and Acf willbe much greater than 1 .41 .

Page 5: An easy-to-use Precision Power Analyser. - MHz … · An easy-to-use Precision Power Analyser. ... Many loads, such as a rectifier with capacitor filter produce a distorted current

This measurement is useful in determining the repetitivepeak current requirement of a power source . For examplein a UPS installation the crest factor must be determinedto ensure that the system can handle the repetitive peakcurrent without being overloaded .

Ainst Peak InrushCurrent

The PM1000 can determine the peak inrush current thatflows into a load when the load is first energised . Forexample an ordinary tungsten lamp passes a significantlyhigher current when first switched on than the currentthat flows when it is running .

It is useful to determine the magnitude of this current inorder to correctly specify fuses and circuit breakers andto determine the effect on the AC supply feeding the load .

To measure Ainst :-a) Connect the PM 1000 as shown .

b) Select Ainst . The PM1000 is now 'armed' ; it iscontinuously scanning the signal on the currentchannel and is ready to display the peak value .

c) Energise AC supply by closing SW1, which may be anelectro-mechanical or a solid state switch . The peakinrush current will be captured and displayed .

d) To take another reading of peak current reset thePM 1000 by pressing Ainst again .

FREQ FrequencyThis displays the frequency in Hz of the voltage or currentinput. If both the voltage and current inputs are connectedand they are not the same frequency (e.g . a PWMsystem) then the PM1000 will choose and display thelower frequency . This feature allows the PM1000 to workin applications where one input is 'chopped' such asvariable speed motor drives etc .

5

Page 6: An easy-to-use Precision Power Analyser. - MHz … · An easy-to-use Precision Power Analyser. ... Many loads, such as a rectifier with capacitor filter produce a distorted current

6

HARM HarmonicAnalysis

This function is used to determine the magnitude of the fundamentalcomponent and harmonics of distorted current or voltage waveforms . ManyAC loads, such as DC power supplies, florescent lamps and thyristorconvertors draw a distorted current waveform from the AC supply .

There is an increasing awareness of the need to limit the level of harmonicsdrawn from the supply, and a number of regulatory authorities already specifylimits for the harmonic content of certain loads . For example VDE 0712 Part 2provides a limit for ballast lamps of :

The PM1000 will measure the harmonics of the voltage or currentwaveforms .

To determine harmonic content of the current waveform :

a) Connect the PM1000 so that it measures the correct fundamentalfrequency when 'FREQ' is selected (see page 5) .

b) Select 'HARM' . The red L.E.D. will turn on to indicate that the PM 1000 isin the harmonic analysis mode . Now press the button marked 'Arms' toselect current analysis . All L .E.D.'s above the buttons will light and thedisplay will read the Total Harmonic Distortion (THD) of the currentwaveform as a percentage .

c) To determine the value of a specific harmonic simply press the pushbuttonappropriate to the harmonic required . Harmonic '0' (DC) will be displayedin amps DC, while harmonics 2, 3, 5, 7, 9, 11, 13 will be displayed as apercentage of the fundamental . The fundamental is displayed in ampsrms .

d) To exit this function press 'HARM' once again . The red L .E .D. will go out,and the instrument will reset to Vrms' .

To determine harmonic content of the voltage waveform :

a) Connect the PM1000 so that it measures the correct fundamentalfrequency when 'FREQ' is selected (see page 5) .

b) Select 'HARM' . The red L.E.D. will turn on to indicate that the PM1000 isin the harmonic analysis mode . Now press the button marked Vrms' toselect voltage analysis . All L .E.D .'s above the buttons will light and thedisplay will read the Total Harmonic Distortion (THD) of the voltagewaveform as a percentage .

c) To determine the value of a specific harmonic simply press the pushbuttonappropriate to the harmonic required . Harmonic '0' (DC) will be displayedin volts DC, while harmonics 2, 3, 5, 7, 9, 11, 13, will be displayed as apercentage of the fundamental . The fundamental is displayed in voltsrms .

d) To exit this function press 'HARM' once again . The red L .E.D . will go out,and the instrument will reset to Vrms' .

NoteIn the harmonic analysis mode the frequency of the fundamental is taken asthe frequency which is displayed when 'FREQ' is selected.

IHarmonic Limit (% of Fundamental)2 53 25 x Power Factor/0 .95 77 49 3

11 213 1

Page 7: An easy-to-use Precision Power Analyser. - MHz … · An easy-to-use Precision Power Analyser. ... Many loads, such as a rectifier with capacitor filter produce a distorted current

Connection DiagramsSingle PhaseMeasurements

Three PhaseMeasurements

Page 8: An easy-to-use Precision Power Analyser. - MHz … · An easy-to-use Precision Power Analyser. ... Many loads, such as a rectifier with capacitor filter produce a distorted current

8

FunctionalDescriptionThe PM 1000 is a micro-processor based instrument thateffects measurements by sampling and digitizing thevoltage and current waveforms and computing the valueof the required measurement in the processor .

Operation is as follows :An input attenuator (8MS2 impedance) generates a lowlevel voltage signal and a 25mu1 current shunt generatesa low level current signal . These signals are buffered bydifferential amplifiers (1), which permit the inputs to floatwith respect to each other and with respect to ground,greatly simplifying accurate measurements .

The signals from the differential amplifiers are amplifiedby programmable-gain amplifiers (2), the gain beingselected by the processor according to the magnitude ofthe input voltage and current signals .

The outputs from the gain amplifiers are converted todigital information by fast A-D converters (3), whicheffect simultaneous scanning of the signals on thevoltage and current channels . The rate of scanning iscontrolled by the processor (4) depending on thefrequency of the voltage signals .

The digital data captured by the processor is used tocompute the value of the measurement selected by thekeyboard (5), and the result is presented on the 4 1/2 digitLED display (6) .

Page 9: An easy-to-use Precision Power Analyser. - MHz … · An easy-to-use Precision Power Analyser. ... Many loads, such as a rectifier with capacitor filter produce a distorted current

Specifications

9

RMSVoltage(Vrms)

Range

Display

Frequency range

Crest Factor

Accuracy 23

Input Impedance

PeakInputVoltage

± 5°C Sine wave

Continuous;

1 second

2V to 700V rms (autoranging) in 4 ranges :8V, 35V, 140V and 700V rms

4'/ digits

DC and 5Hz to 20kHz

Up to 19 .9 (limit 1000V Pk)

+ 0 .20% ± 0 .02%/kHz ± 0 .10% of range

8M11

1000V

2500V

RMSCurrent(Arms)

Range

Display

Frequency range.

Crest Factor

Accuracy 23

Input Resistance

Standard

Option 001

± 5"C Sinewave

20mA to 20A rms (autoranging) in 4 ranges :-0 .2A, 1A, 5A, and 20A rms

5OmA to 20A rms (autoranging) in 4 ranges :-0 .5A, 2 .5A, 12 .5A and 20A rms

4'/2 digits

DC and 5Hz to 20kHz

Up to 19.9 (Limit 40A Pk)(Limit 11 OA Pk on option 001)

± 0.20% ± 0 .10%/kHz ± 0 .10% of range

0.050 max inc . fuse circuit

Power(W)

Range

Display

Frequency rang( -:

Accuracy 23

Standard

Option 001

± 5 -'C Sinewave

40mW to 14kW (autoranging) in 16 rangescorresponding to V and A

100mW to 14kW

4'/z digits with polarity accordingto direction of power flow

DC and 5Hz to 20kHz

+ 0 .40% ± (0 .25 - PF)%'kHz± (0.1 x PF)% of range

ApparentPower(VA)

Range

Display

Frequency range

Accuracy 23

Standard

Option 001

± 5 C Sinewave

40mVA to 14kVA (autoranging) in 16 rangescorresponding to V and A

100mVA to 14kVA

4'/:--, digits

DC and 5Hz to 20kHz

± 0,40% ± 0.25°'o kHz ± 0 .10% of range

PowerFactor(PF)

Range

Display

Accuracy 23 ± 5°C S newave

0.000 ± 1 .000

4 1/z digits (+ = leading PF~\- = lagging PF I

± 0.002 ± 0 .0025/kHz

Page 10: An easy-to-use Precision Power Analyser. - MHz … · An easy-to-use Precision Power Analyser. ... Many loads, such as a rectifier with capacitor filter produce a distorted current

1 0

VoltageCrest Factor(Vcf)

Range

Display

Accuracy 23 + 5"C Sinewave

1 .00 to 19 .9

3 digits

± 0.05

CurrentCrest Factor(Acf)

Range

D splay

4ccurac, 23 + , C newave

1 .00 to 19 .9

3 digits

± 0.05

InstantaneousPeak Current(Ainst)

Range

Display

Accuracy 23

Standard

Option 001

± 5°C Sinewave

0.1A to 44A Pk (autoranging in 3 ranges1 .5, 7 .5 and 44 A Pk)

0.2A to 11 OA Pk (autoranging in 3 ranges3.8, 19 and 11 OA Pk)

3 digits with polarity

± 2.0% of range

100 psAperture time

AC supplyFrequency(FREQ)

Range

Display

Accuracy 23 ± 5"C Sinewave

5Hz to 20 kHz

4 1/2 digits

± 0 .2% of reading

HarmonicAnalysis(HARM)

RMScurrent

RMS voltage

Frequency range

Standard

Option 001

20mA to 20A rms in 4 ranges0 .2, 1, 5 and 20A rms)

50mA to 20A rms in 4 ranges0.5, 2 .5, 12 .5 and 20A rms

2 to 700V rms (autoranging) in 4 ranges :8, 35, 140 and 700V rms

DC, 5Hz to 20kHz

4 1/2 digits

Fundamental : ± 0 .5% ± 0.2%/kHz

± 0 .1 % of range

Harmonics : (+ 0 .25 ± 0.01/kHz)% of

fundamental

Display

Accuracy 23Triangular Wave

± 5°C

Input Terminal 4mm Safety Sockets A max. 1000V to

Operating Temperature and Humidity 0 50"C, 10-90% RH

Overcurrent Protection Fuse HRC I Vi" (32mm) 20AT

DielectricStrength

Input - CaseInput - Power SupplyPower Supply - Case

4kV AC 50/60 Hz 1 minute

PowerSupply

AC Input Voltage

FrequencyProtectionConsumption

11OV ± 20% (11OV position)220V ± 20% (220V position)48 to 65 HzFuse 20mm 315mAT15W, 20VA max .

Page 11: An easy-to-use Precision Power Analyser. - MHz … · An easy-to-use Precision Power Analyser. ... Many loads, such as a rectifier with capacitor filter produce a distorted current

Dimensions

1 1

Page 12: An easy-to-use Precision Power Analyser. - MHz … · An easy-to-use Precision Power Analyser. ... Many loads, such as a rectifier with capacitor filter produce a distorted current

Safety ConsiderationsThis test equipment is constructed and tested to the requirements of IEC 348'Safety requirements for electronic measuring apparatus' . The instruction manualand the notes below should be read and followed to ensure the safety of theinstrument and the user . Their safety is not assured if the instrument is misused orcarelessly handled .

1 . This equipment is constructed to IEC348 Safety class 1 and the A .C . line plugshould be inserted in a socket with a protective ground contact .

2. The line plug should be inserted before connections are made to measuring orcontrol circuits .

3. Do not attempt to remove outer cover without first disconnecting auxiliary andtest power supplies .

WarrantyThis Voltech Instruments product is warranted against defects in materials andworkmanship for a period of one year from date of shipment . During the warrantyperiod, Voltech Instruments will, at its option, either repair or replace products whichprove to be defective .

For repair or service under warranty, this instrument must be returned to a servicecenter designated by Voltech . Purchaser shall pre-pay shipment charges to servicecenter and Voltech shall pay shipment charges to return the instrument topurchaser .

Limitation of WarrantyThe foregoing warranty shall not apply to defects resulting from unauthorizedmodifications or misuse, or operation outside the specification of the instrument. Noother warranty is expressed or implied .

VoltechWhilst every care has been taken in compiling the information in this publication VoltechInstruments Ltd, cannot accept legal liability for any inaccuracies . Voltech Instruments Ltdhas an intensive programme of design and development which may alter productspecification.Voltech Instruments Ltd. reserves the right to alter specification without notice andwhenever necessary to ensure optimum performance from its product range.

All rights reserved.

No part of this publication may be produced, stored in a retrieval system . or transmitted inany form . or by any means, electronic, mechanical, photocopying, recording or otherwise .without prior permission of Voltech Instruments Ltd .

Distributed by:-

Voltech Instruments Ltd .,65 Milton Park,Milton,Abingdon,Oxford OX14 4RX,England .

Tel: (0235) 861173Fax: (0235) 861174

Voltech, Inc.200 Butterfield Drive,Ashland,Massachusetts, 01721 .U.S.A .

Tel: (508) 881-7329Fax: (508) 879-8669