GBM5320GBM5320Dispositifs Mdicaux IntelligentsDispositifs Mdicaux Intelligents
Sensors : IntroductionSensors : IntroductionMohamad Sawan, Professor
Benoit Gosselin, and Louis-Francois Tanguay, Ph.D. CandidatesLaboratoire de neurotechnologies Polystim
http://www.cours.polymtl.ca/gbm5320/mohamad.sawan@polymtl.ca
M5418
9 February 2008
GBM5320 - GBM5320 - Dispositifs Mdicaux IntelligentsDispositifs Mdicaux IntelligentsGBM5320 - Dispositifs Mdicaux Intelligents 2
Biomedical microsensors : Course outline
Microsensors- Overview- Definitions
Microsensors types:- Strain- Pressure- Displacement- Temperature- Gas (Electrode-based)- Chemical sensors (ISFET, CHEMFET)
Biosensors Lab-on-chip technology
GBM5320 - GBM5320 - Dispositifs Mdicaux IntelligentsDispositifs Mdicaux IntelligentsGBM5320 - Dispositifs Mdicaux Intelligents 3
Microsensors?
Microsensors are small devices that convert physicalor chemical signals to electrical signals. They enableobjects to interface to the real world;
Implantable microsensors enables monitoringbiological parameters. They could allow real-timemeasurement of temperature, pressure, pH, oxygenand nitric oxide concentrations in vivo;
They allow to help the medical research community inlearning about the progression of diseases and assessdegree of response to treatment;
More & better access to measurement sites- Do not perturb the system under test- More accurate measurements and less invasive- Less psychological trauma & feedback
More functionality, better portability, and lower cost.
Pressure sensor
Gas sensorhttp://www.nist.gov/public_affairs/techbeat/tb2003_0910.htm
-
GBM5320 - GBM5320 - Dispositifs Mdicaux IntelligentsDispositifs Mdicaux IntelligentsGBM5320 - Dispositifs Mdicaux Intelligents 4
Definitions
MicrosensorA microdevice that transforms a signal in measured/analyte format inan electrical signal.
Direct sensorSignal to be measured is directly transformed to electrical signal.Example: photo-conductor converts light to change of resistance.
Indirect sensorSignal to be measured is first converted to some other variable that isthen converted to an electrical signal Example: acceleration sensor converts acceleration to strain
which is then sensed. Biosensor
A microsensor dedicated for medical implantable and cellular devices.
GBM5320 - GBM5320 - Dispositifs Mdicaux IntelligentsDispositifs Mdicaux IntelligentsGBM5320 - Dispositifs Mdicaux Intelligents 5
Sensor Performance Characteristics Transfer Function: The functional relationship between physical
input signal and electrical output signal.
Sensitivity: The sensitivity is the ratio between a small change inelectrical signal resulting from a small change in the physical signal tobe measured.
Dynamic Range: The range of input physical signals which may beconverted to electrical signals by the sensor. Signals outside of thisrange are expected to cause unacceptably large inaccuracy.
Linearity: The maximum deviation from a linear transfer function overthe specified dynamic range.
Accuracy: Generally defined as the largest expected error betweenactual and ideal output signals.
Resolution: The resolution of a sensor is defined as the minimumdetectable signal fluctuation.
GBM5320 - GBM5320 - Dispositifs Mdicaux IntelligentsDispositifs Mdicaux IntelligentsGBM5320 - Dispositifs Mdicaux Intelligents 6
Sensor Performance Characteristics
Hysteresis: Some sensors do not return to the same output valuewhen the input stimulus is cycled up or down. The width of theexpected error in terms of the measured quantity is defined as thehysteresis.
Noise: All sensors produce some output noise in addition to theoutput signal. The noise of the sensor limits the performance of thesystem based on the sensor. Noise is generally distributed across thefrequency spectrum.
Bandwidth: All sensors have finite response times to aninstantaneous change in physical signal. In addition, many sensorshave decay times, which would represent the time after a step changein physical signal for the sensor output to decay to its original value.The reciprocal of these times correspond to the upper and lowercutoff frequencies, respectively. The bandwidth of a sensor is thefrequency range between these two frequencies.
-
GBM5320 - GBM5320 - Dispositifs Mdicaux IntelligentsDispositifs Mdicaux IntelligentsGBM5320 - Dispositifs Mdicaux Intelligents 7
Microsensors: General architecture
A generalized architecture of a microsensor system:
Sensor/ActuatorArray
Signal Conditioners(Analog + Digital)
Embedded Controller(Calibrate-measure, process & compress, store & forward)
Drivers Comm.Interface
Inputs
GBM5320 - GBM5320 - Dispositifs Mdicaux IntelligentsDispositifs Mdicaux IntelligentsGBM5320 - Dispositifs Mdicaux Intelligents 8
Strain sensors - ResistiveResistance is related to length and area of cross-section of theresistor and resistivity of the material as
By differentiating both sides, the equation becomes
DimensionalPiezoresistance
Strain gage component can be related by poissons ratio (v) as
LengthTransfer Function : Input is strain, output is dR.
Webster, Medical Instrumentation
GBM5320 - GBM5320 - Dispositifs Mdicaux IntelligentsDispositifs Mdicaux IntelligentsGBM5320 - Dispositifs Mdicaux Intelligents 9
Strain sensors - Resistive
Gage Factor of a strain gage
G is a measure of sensitivity
Put mercury strain gage around an arm orchest to measure force of musclecontraction or respiration, respectively
Used in prosthesis or neonatal apneadetection, respectively.
= dL/L
Webster, Medical Instrumentation www.microstrain.com/
-
GBM5320 - GBM5320 - Dispositifs Mdicaux IntelligentsDispositifs Mdicaux IntelligentsGBM5320 - Dispositifs Mdicaux Intelligents 10
Piezoelectric Sensors
What is piezoelectricity ?
Strain causes a redistri-bution of charges andresults in a net electricdipole
where q = charge, f = force
k = 2.3 pC/N for quartz = 140 pC/N for Barium
Different transducerapplications:- Accelerometer,- Microphone.
q = k f & V = q / C
www.ipodlinux.org/group27imaging.com/RespiratorySensor.aspx
GBM5320 - GBM5320 - Dispositifs Mdicaux IntelligentsDispositifs Mdicaux IntelligentsGBM5320 - Dispositifs Mdicaux Intelligents 11
Displacement Sensor - LVDT
LVDT
www.pages.drexel.edu/~pyo22/mem351-2004/lecture04/pp062-073lvdt.pdf
An LVDT (Linear Variable Differential Transformer) is used as asensitive displacement sensor: for example, in a cardiac assist deviceor a basic research project to study displacement produced by acontracting muscle.
Signal ConditioningElectronics Muscle
Inductive displacement sensors:- Self inductance;- Mutual inductance;- Differential transformer.
GBM5320 - GBM5320 - Dispositifs Mdicaux IntelligentsDispositifs Mdicaux IntelligentsGBM5320 - Dispositifs Mdicaux Intelligents 12
Capacitance-based Sensors
DifferentialMode
Variable Dielectric ModeVariable Area Mode
-
GBM5320 - GBM5320 - Dispositifs Mdicaux IntelligentsDispositifs Mdicaux IntelligentsGBM5320 - Dispositifs Mdicaux Intelligents 13
Acceleration sensor
Accelerometer fordisplacement monitoring- Surface micromachined,
capacitive sensor- Sensor + Electronics on same
substrate= smart
Analog Devices ADXL-50
C1 C2
g
GBM5320 - GBM5320 - Dispositifs Mdicaux IntelligentsDispositifs Mdicaux IntelligentsGBM5320 - Dispositifs Mdicaux Intelligents 14
Pressure sensors
Collins Miniature Passive Pressure Transensor for Implanting, 1967.
Miniature Passive Pressure Transensor for Implanting in the Eye Measurement of intraocular and other physiological pressures. Displacement transducer contained in a small distensible pillbox. This passive
resonant transensor absorbs energy from an oscillating detector coil outsideof the animal at a frequency dependent upon the pressure in the eye.
GBM5320 - GBM5320 - Dispositifs Mdicaux IntelligentsDispositifs Mdicaux IntelligentsGBM5320 - Dispositifs Mdicaux Intelligents 15
15
The value of capacitorchange with pressuredue to the deflectablediaphragm. Thisvariation change theresonant frequency ofthe LC circuit and ismeasured wirelessly.
Pressure sensors
Allen, GA Tech, 1999-2002.
PTFE = PolytetrafluoroethyleneFEP = Fluorinated Ethylene Propylene
Ceramic chamber
Flexible Wireless Passive PressureSensors for Biomedical Applications.
The sensor consists of a cavity,bounded on 2 sides by capacitorplates interconnected with inductance.
-
GBM5320 - GBM5320 - Dispositifs Mdicaux IntelligentsDispositifs Mdicaux IntelligentsGBM5320 - Dispositifs Mdicaux Intelligents 16
Micromachined pressure sensors
Pressure Sensor-Resistive / capacitive based measurements-Thin Silicon Membrane deforms with
pressure-Piezoresistors change with strain induced
by bending membrane-Packaging requires sealing to maintain
pressure differential.
www.dolphin.fr/flip/mems/mems_cps.htmlwww.memstouch.net/
High sensitivity capacitivestrain sensor.
GBM5320 - GBM5320 - Dispositifs Mdicaux IntelligentsDispositifs Mdicaux IntelligentsGBM5320 - Dispositifs Mdicaux Intelligents 17
Temperature sensors
Temperature sensors have become common elements in wide range ofmodern integrated circuits
The main parameters of temperature sensors are: temperature range,sensitivity, output range, linearity, accuracy
Types of integrated temperature sensors:- Resistance based : Thermistors, RTDs- Thermocouples & CMOS PTAT references.
www.singleiteration.com/
GBM5320 - GBM5320 - Dispositifs Mdicaux IntelligentsDispositifs Mdicaux IntelligentsGBM5320 - Dispositifs Mdicaux Intelligents 18
Thermistors
Thermistors are made fromsemiconductor material
Generally, they have a negativetemperature coefficient (NTC), that isNTC thermistors are most commonlyused
Ro is the resistance at a reference point(in the limit, absolute 0), B is materialconstant, and T and T0 are absolute andreference temperatures.
Webster, Medical instrumentation
-
GBM5320 - GBM5320 - Dispositifs Mdicaux IntelligentsDispositifs Mdicaux IntelligentsGBM5320 - Dispositifs Mdicaux Intelligents 19
Thermocouples
A conductor generates a voltage when subjected to a temperature gradient. Tomeasure this voltage, one must use a second conductor material whichgenerates a different voltage under the same temperature gradient. So,Thermocouples measure temperature differences and need a known referencetemperature to yield the absolute readings.
When a pair of dissimilar metals are joined at one end, and there is atemperature difference between the joined ends and the open ends, thermalelectromotive force (emf) is generated, which can be measured in the openends. There are three major effects involved : the Seebeck, Peltier, andThomson.
Webster, Medical Instrumentationwww.efunda.com/.../images/thermocouple_A.gif
GBM5320 - GBM5320 - Dispositifs Mdicaux IntelligentsDispositifs Mdicaux IntelligentsGBM5320 - Dispositifs Mdicaux Intelligents 20
CMOS temperature sensor
VBE 2 !VBE1 = "VBE (T)
=kT
qln
pIE
IS2
#
$ %
&
' ( !
kT
qln
IE
IS1 /r
#
$ %
&
' ( =
kT
qln p ) r[ ]
The voltage difference between the two diodes, operated at a differentcurrent density, is used to generate a Proportional To AbsoluteTemperature (PTAT) current.
This voltage difference is PTAT with a temperature coefficient of+0.085 mV/C at room temperature.
Pertijs et al, Precision Temperature Measurement using ,IEEE Sensors, v4, 2004.
GBM5320 - GBM5320 - Dispositifs Mdicaux IntelligentsDispositifs Mdicaux IntelligentsGBM5320 - Dispositifs Mdicaux Intelligents 21
IPTAT = I5 =W5/W4*I2
CMOS temperature sensor : Complete PTAT circuit
The current mirrored at theoutput is PTAT:
VX!V
Y
VR1 =VY !VZ "VX !VZ
VR1 =VEB1 !VEB 2 =kT
qln
A1
A2
#
$ %
&
' (
IR1
= I2
=VR
1
R1
=1
R1
!kT
qln
A1
A2
"
# $
%
& '
-
GBM5320 - GBM5320 - Dispositifs Mdicaux IntelligentsDispositifs Mdicaux IntelligentsGBM5320 - Dispositifs Mdicaux Intelligents 22
pH Electrodes
Glass electrodes develop a gel layerwith mobile hydrogen ions when dippedinto an aqueous solution;
pH changes cause ion diffusionprocesses generating an electrodepotential. Lithium-rich glasses are wellsuited for this purpose;
The potential is measured in comparisonto a reference electrode which is usuallyan Ag/AgCl system;
The electric circuit is closed via adiaphragm separating the referenceelectrolyte from the solution.
Sonnleitner, Bioanalysis and Biosensors for Bioprocess Monitoring,Springer, 1999.
GBM5320 - GBM5320 - Dispositifs Mdicaux IntelligentsDispositifs Mdicaux IntelligentsGBM5320 - Dispositifs Mdicaux Intelligents 23
Oxygen Partial Pressure (pO2) electrode
A membrane through which oxygenmust diffuse separates the measuringsolution from the electrolyte
Oxygen is reduced by electronscoming from the central platinumcathode which is surrounded by aglass insulator.
This design, a so-called polaro-graphicelectrode, needs an external powersupply.
For oxygen, the polarization voltage isin the order of 700 mV and the typicalcurrent for atmospheric pO2 is in theorder of 107 A.
Sonnleitner, Bioanalysis and Biosensors for Bioprocess Monitoring,Springer, 1999.
Clark-type oxygenpartial pressure (pO2)
electrode
GBM5320 - GBM5320 - Dispositifs Mdicaux IntelligentsDispositifs Mdicaux IntelligentsGBM5320 - Dispositifs Mdicaux Intelligents 24
Carbon Dioxide Partial Pressure (pCO2) electrode
CO2 diffuses through the membrane into or out of the electrolyte whereit equilibrates with HCO3 thus generating or consuming protons.
The respective pH change of the electrolyte is sensed with a pHelectrode and is logarithmically proportional to the pCO2 in themeasuring solution.
Sonnleitner, Bioanalysis and Biosensors for Bioprocess Monitoring,Springer, 1999.
-
GBM5320 - GBM5320 - Dispositifs Mdicaux IntelligentsDispositifs Mdicaux IntelligentsGBM5320 - Dispositifs Mdicaux Intelligents 25
Ion-Sensitive Field Effect Transistors (ISFETS and CHEMFETs) arebasically metal oxide semiconductor field-effect devices.
The construction of an ISFET differs from the conventional MOSFETdevices, in that the gate metal is omitted and replaced by amembrane sensitive to the ions of interest.
ISFET/CHEMFET sensors
www.sentron.nl/nieuw/index.php?id=4Shepherd, Weak Inversion ISFETs Sensing , S&A B , v107, 2005.
GBM5320 - GBM5320 - Dispositifs Mdicaux IntelligentsDispositifs Mdicaux IntelligentsGBM5320 - Dispositifs Mdicaux Intelligents 26
pH ISFET equivalent model
The drain current for the weak inversion ISFET in saturation is given by:
Shepherd & Toumazou, Weak Inversion ISFETs for Ultra-Low PowerBiochemical Sensing , Sensors and Actuators B (Chemical), v107, 2005.
pH-ISFET Macromodel
GBM5320 - GBM5320 - Dispositifs Mdicaux IntelligentsDispositifs Mdicaux IntelligentsGBM5320 - Dispositifs Mdicaux Intelligents 27
Classification of biosensors
Ferrari et al, BioMEMS and Biomedical Nanotechnology: Vol IV:Biomolecular Sensing, Processing and Analysis, Springer, 2006.
-
GBM5320 - GBM5320 - Dispositifs Mdicaux IntelligentsDispositifs Mdicaux IntelligentsGBM5320 - Dispositifs Mdicaux Intelligents 28
Biosensors
A bioreceptor is a biological molecular species (e.g., an antibody, anenzyme, a protein, or a nucleic acid) or a living biological system (e.g.,cells, tissue, or whole organisms) that utilizes a biochemicalmechanism for recognition
The sampling component of a biosensor contains a bio-sensitive layer.The layer can either contain bioreceptors or be made of bioreceptorscovalently attached to the transducer.
The most common forms of bioreceptors used in biosensing are basedon:- Antibody/antigen interactions- Nucleic acid interactions- Enzymatic interactions- Cellular interactions (i.e. microorganisms, proteins)- Interactions using biomimetic materials (i.e., synthetic bioreceptors).
GBM5320 - GBM5320 - Dispositifs Mdicaux IntelligentsDispositifs Mdicaux IntelligentsGBM5320 - Dispositifs Mdicaux Intelligents 29
Example : Glucose Sensors
Enzymatic Approach
Glu e O GluconicAcid H OGlu eOxidase
coscos
+ ! "!!!!!! +2 2 2
Makes use of catalytic (enzymatic)oxidation of glucose
The setup contains an enzymeelectrode and an oxygen electrode andthe difference in the readings indicatesthe glucose level.
The enzyme electrode has glucoseoxidase immobilized on a membrane ora gel matrix*.
Platinumelectrode
Plasticmembrane
Glucose
O2
Gluconicacid
Silveranode
O2
H2O2O2
*In the enzyme electrode, when glucose is present it combines with O2, so less O2 arrives to the cathode.
Webster, Medical Instrumentation
GBM5320 - GBM5320 - Dispositifs Mdicaux IntelligentsDispositifs Mdicaux IntelligentsGBM5320 - Dispositifs Mdicaux Intelligents 30
Example : Glucose SensorsAffinity Approach (Optical)
This approach is based onthe immobilized competitivebinding of a particularmetabolite (glucose) and itsassociated fluorescent labelwith receptor sites specific tothe metabolite and thelabeled ligand. This changein light intensity is thenpicked up. 3 mm
0.3 mm
Hollow dialysis fiber
Excitation
Emission
OpticalFiber
Glucose
Schultz et al, Affinity sensor : A new technique, Diabetes Care, 1982.
Measure of glucose concentration by detecting changes in fluorescent lightintensity caused by competitive binding of a fluorescein-labeled indicator.