toc analyser - minitoc-arrowlabs.net/resources/brochures/minitoc short presentation.pdf · toc...

Post on 17-Oct-2020

12 Views

Category:

Documents

1 Downloads

Preview:

Click to see full reader

TRANSCRIPT

TOC Analyser - MiniTOC-

membraPure GmbH Wolfgang-Küntscher Straße 14

16761 Hennigsdorf

membraPure GmbH founded 1993

office building production hall Product range:

2 22.11.2018

TOC Analyser

measurement range 0.1 - 1000 ppb

cell constant accuracy 1.5%

temperature sensor : PT 1000

temperature accuracy: +/- 0.25 K

ports 1 for sample, 1 for SST (Basic)

1 for sample, 3 for SST (Professional)

max. conductivity of sample:

< 2 μS/cm (online)

3 22.11.2018

TOC Analyser

sample volume flow 14 ml/min

analysis time 2 sec

inlet pressure < 0.5 bar otherwise

pressure reducer

sample temperature 10 - 50 °C if

more than heat exchanger

particles < 100 micron

4 22.11.2018

TOC Analyser

Housing

dimensions (w x d x h)

300 mm x 200 mm x 500 mm

weight 14 kg

protection: IP 54 (option)

Operating Conditions

power supply 110 - 230 VAC, 50 Hz

power consumption < 50 W

ambient relative humidity 10 - 95 %

altitude : up to 3500 m 5 22.11.2018

TOC Analyser

Configuration

control internal PC, 8“ TFT touch screen

monitor

operating system Windows 7 embedded

software miniTOC control software for

measurement, qualification, diagnosis

interfaces USB 2.0, analogue out (option)

6 22.11.2018

Design

UV-Reactor

Pump

Conductivity sensors

Valves

7 22.11.2018

Scheme fluidic

8 22.11.2018

Fluid connections

9 22.11.2018

Menu

10 22.11.2018

Online measurement

11 22.11.2018

Calibration - start

12 22.11.2018

Calibration- finished

Factor & R²

Measured data

Calibration curve

13 22.11.2018

SST - start

14 22.11.2018

Viewer

15 22.11.2018

Interval Mode MiniTOC pro

16 22.11.2018

Service Change UV-Reactor: Plug out power supply Remove plug connection Unscrew fluidic tubes Take reactor out of holder

Change pump tube: Open cable ties Remove fluidic lines from

pump tube Open pump clamps, remove

tube by turning pump manually

Reset hourmeters for pump tube and UV Lamp!!

17 22.11.2018

Installation

waste

Cooler (if T > 50 °C)

Option:

Pressure reducer

p ≈ 0.25 bar

connector ¼“ -28 UNF

connector ¼“ BSP x 6 mm

Water Connection

connector

Gate valve with

Adapter 6 mm

Tube 6 mm AD

Tube 1/8“ 3.2 mm AD

18 22.11.2018

Installation

19 22.11.2018

top related