OrbiterOrbiter** IM512P/E IM512P/E UpgradeUpgrade

OrbiterOrbiter** IM512P/E Upgrade IM512P/E Upgrade

DOT Terminal Removed. ALL DOT BIN Boards Removed. All DOT Terminal Tasks now done in the IM512P/E. Detector Energy & Linearity Correction Boards not used

anymore. New Motion Controller. New Detector Controller. New IM512P/E Acquisition & Processing Software.

DOT BIN with the New OrbiterDOT BIN with the New Orbiter** Controller Bd.Controller Bd.

New Detector Interface Bd.New Detector Interface Bd.

DOT BIN Removed Boards (ALL 9)DOT BIN Removed Boards (ALL 9)

Digitrac Detector Removed Boards (4)Digitrac Detector Removed Boards (4)

Removed DOT TerminalRemoved DOT Terminal

AdvantagesAdvantages

On the fly Energy, Linearity, Uniformity corrections done by Acquisition Software.

Real Time Isotope Spectrum Display.

Better Linearity & Uniformity Specs.

Intrinsic and User Uniformity Correction Tables for any Isotope-Collimator pair.

Last generation USB V2 Acquisition Interface.

Less boards, more reliability.

Less power consumption, less generated heat. Last generation processor + Windows XP/Vista/Windows 7**

PC. Easy to Maintain and Upgrade.

Quality Control Quality Control AcquisitionsAcquisitions

Quality Control Quality Control AcquisitionsAcquisitions

Quality Control AcquisitionsQuality Control Acquisitions

UFOV Uniformity: 1.6 % Int. / 1.4 % Diff.UFOV Uniformity: 1.6 % Int. / 1.4 % Diff.

Absolute Linearity [mm]: 0.28 CFOV 0.38 Absolute Linearity [mm]: 0.28 CFOV 0.38 UFOVUFOV

Resolution [mm]: FWHM=3.92 Resolution [mm]: FWHM=3.92 FWTM=7.39FWTM=7.39

Specifications Comparison TableSpecifications Comparison Table

Original IM512P/E Upgrade

Resolution: FWHM UFOV [mm] 3.94 3.92

Resolution: FWTM UFOV [mm] 7.52 7.39

Absolute LIN UFOV [mm] 0.9 < 0.5 (0.35 typ.)

Absolute LIN CFOV [mm] 0.5 < 0.4 (0.30 typ.)

Integral UNIF UFOV [%] 4.3 < 2.5 (1.6 typ.)

Differential UNIF UFOV [%] 3.0 < 2.0 (1.4 typ.)

Max. Raw Count Rate [kcps] > 200 > 200

Multiple Window Registration [mm] < 2.0 < 2.0

ConnectivityConnectivity

To and from most of the current systems via DICOM.

InterFile.

Exported graphic formats: PCX, JPG, AVI.

Can convert local and imported Cine Displays into standard AVI files to be seen in any Windows** based computer.

Remote Help Desk & Service via Integrated VNC® with file transfer.

IM512P/E Lin. Phantom exported to IM512P/E Lin. Phantom exported to e.softe.soft

IM512PE screen exported to syngoIM512PE screen exported to syngo

EndEnd

Contact usContact us

Address:Address: Elpidio González 4717. Elpidio González 4717. (C1407GBC) Ciudad de Buenos Aires (C1407GBC) Ciudad de Buenos Aires República Argentina República Argentina

Phone:Phone: (54 11) 4566-1202 / 5119 (54 11) 4566-1202 / 5119

Fax:Fax: (54 11) 4566-5822 (54 11) 4566-5822

e-mail:e-mail: admin@alfanuclear.com admin@alfanuclear.com

* Orbiter is a trademark of Siemens Medical Systems, Inc.** Windows, Windows XP, Vista and Windows 7 are registered trademarks of Microsoft Corporation. All other trademarks mentioned are property of their respective owners.

top related