Lens Identification to Prevent Radiation Induced Cataracts Using Convolutional Neural Networks

Ross Filice, M.D.Associate Professor

Chief of Imaging InformaticsMedStar Georgetown University Hospital

MedStar Medical Group Radiology

#CMIMI18

Disclosures

Nvidia Corporation; Academic GPU grant

#CMIMI18

The Problem

“…repeated exposure to head and neck CT is significantly

associated with increased risk of cataract incidence.”

Yuan MK, Tsai DC, Chang SC, Yuan MC, et al. The Risk of Cataract Associated With Repeated Head and Neck CT Studies: A Nationwide Population-Based Study. AJR Sep 2013; 201(3):626-30.

#CMIMI18

The Problem

In computed tomographic (CT) imaging of the brain, the lens of the eye is of special concern, as cataract formation is a well-documented result of radiation damage.

Particular attention is paid to…cataracts…hence, threshold doses appear to be lower than previously considered.

ICRP, 2012 ICRP Statement on Tissue Reactions / Early and Late Effects of Radiation in Normal Tissues and Organs – Threshold Doses for Tissue Reactions in a Radiation Protection Context. ICRP Publication 118. Ann. ICRP 41(1/2).

Reimann et al. Organ-Based Computed Tomographic (CT) Radiation Dose Reduction to the Lenses: Impact on Quality for CT of the Head. JCAT May/June 2012; 36(3):334-8.

#CMIMI18

The Problem

In computed tomographic (CT) imaging of the brain, the lens of the eye is of special concern, as cataract formation is a well-documented result of radiation damage.

Particular attention is paid to…cataracts…hence, threshold doses appear to be lower than previously considered.

ICRP, 2012 ICRP Statement on Tissue Reactions / Early and Late Effects of Radiation in Normal Tissues and Organs – Threshold Doses for Tissue Reactions in a Radiation Protection Context. ICRP Publication 118. Ann. ICRP 41(1/2).

Reimann et al. Organ-Based Computed Tomographic (CT) Radiation Dose Reduction to the Lenses: Impact on Quality for CT of the Head. JCAT May/June 2012; 36(3):334-8.

#CMIMI18

AAPM – Routine Adult Head CT

To reduce or avoid ocular lens exposure, the scan angle should be parallel to a line created by the supraorbitalridge and the inner table of the posterior margin of the foramen magnum. This may be accomplished by either tilting the patient’s chin toward the chest (“tucked” position) or tilting the gantry. While there may be some situations where this is not possible due to scanner or patient positioning limitations, it is considered good practice to perform one or both of these maneuvers whenever possible.

https://www.aapm.org/pubs/CTProtocols/documents/AdultRoutineHeadCT.pdf

#CMIMI18

AAPM – Routine Adult Head CT

To reduce or avoid ocular lens exposure, the scan angle should be parallel to a line created by the supraorbital ridge and the inner table of the posterior margin of the foramen magnum. This may be

accomplished by either tilting the patient’s chin toward

the chest (“tucked” position) or tilting the gantry. While there may be some situations where this is not possible due to scanner or patient positioning limitations, it is considered good practice to perform one or both of these maneuvers whenever possible.

https://www.aapm.org/pubs/CTProtocols/documents/AdultRoutineHeadCT.pdf

#CMIMI18 Futacsi B, Kovacs B, Karlinger, K. Dose alertness: how to lower lens dose in standard brain CT scans? Our clinical experience. ECR 2014.

#CMIMI18

Previous Work at MGUH

Manual paper-based technologist feedbackResulted in improvement….

#CMIMI18

Previous Work at MGUH

Manual paper-based technologist feedbackResulted in improvement….

…when project stopped returned to baseline.

#CMIMI18

Continuous Feedback Works

Lumish et al. Reporting Scan Time Reduces Cardiac MR Examination Duration. JACR Apr 2014; 11(4):425-8.

#CMIMI18

Continuous Feedback Works

Lumish et al. Reporting Scan Time Reduces Cardiac MR Examination Duration. JACR Apr 2014; 11(4):425-8.

#CMIMI18

Methods

- 100 routine head CTs- 4128 512x512 3mm axial soft tissue kernel images- Manual bounding boxes drawn on eyes and lenses- CNN based on GoogLeNet with modification to allow simultaneous detection of two objects (eyes; lenses)- 10 studies consisting of 404 images used for testing- 138 routine outpatient exams over 2 months used to establish baseline technologist compliance

#CMIMI18

Modified GoogLeNet

#CMIMI18

Example Bounding Boxes

#CMIMI18

Example Bounding Boxes

#CMIMI18

Saliency Features

#CMIMI18

Image and Study Level ResultsModel

Lens None

Trut

h Lens 45 2

None 1 356

Model

Eye None

Trut

h Eye 66 2

None 1 335

Sensitivity: 95.7%Specificity: 99.7%Accuracy: 99.3%

Sensitivity: 97.0%Specificity: 99.7%Accuracy: 99.3%

* Both FP had relatively low confidence scores* 1 lens FP was an implant; 1 eye FP was a lens

Study level accuracy: 100%

#CMIMI18

Baseline Technologist Performance

#CMIMI18

Baseline Technologist Performance

#CMIMI18

Future Directions

- Automatically screen exams on ingestion- Notifications and/or reports for technologist leads- Evaluate for improvement

- Limitations: Sensitivity perhaps could be improved- Some patients cannot comply with positioning; some feedback mechanism to override or exempt “noncompliance”