D R I L L I N GS E Q U E N C EG U I D E
IS-II active
D R I L L I N GS E Q U E N C EG U I D E
IS-II activeØ 4.5 10mm fixture will be placed
2
D R I L L I N GS E Q U E N C EG U I D E
IS-II activeIncision
3
D R I L L I N GS E Q U E N C EG U I D E
IS-II active
1,200 RPM, 30 ~ 45Ncm
Point Lindemann drilling withirrigation
4
D R I L L I N GS E Q U E N C EG U I D E
IS-II active
1,200 RPM, 30 ~ 45Ncm
Point Lindemann drilling withirrigation
5
D1 D1 : Ø1.2D2 : Ø2.5D2
7
1.3mm0
33.5mm
8.5 10 11.5 13 16.5
D R I L L I N GS E Q U E N C EG U I D E
IS-II activePoint Lindemann drill
Diameter Product nameØ2.5 LDS25
LDSØ
25
6
D R I L L I N GS E Q U E N C EG U I D E
IS-II activePositioning for the initial drilling
7
D R I L L I N GS E Q U E N C EG U I D E
IS-II active
1,200 RPM, 30~45Ncm,10mm stopper
Initial drilling(Ø2.2)with irrigation
8
D R I L L I N GS E Q U E N C EG U I D E
IS-II active
1,200 RPM, 30~45Ncm,10mm stopper
Initial drilling(Ø2.2)with irrigation
9
0
33.5mm
D R I L L I N GS E Q U E N C EG U I D E
IS-II activeInitial drill
Diameter Product nameØ2.2
Ø2.2
TSD22FTS
D22
F
7
1.3mm
8.5 10 11.5 13 16.5
10
D R I L L I N GS E Q U E N C EG U I D E
IS-II activeØ 2.2 depth 10mm hole has appeared
11
D R I L L I N GS E Q U E N C EG U I D E
IS-II activeChecking the depthwith depth gauge
12
D R I L L I N GS E Q U E N C EG U I D E
IS-II activeChecking the path with a parallel pin
13
D R I L L I N GS E Q U E N C EG U I D E
IS-II activeTaking radiography (X-ray)
14
D R I L L I N GS E Q U E N C EG U I D E
IS-II activeRemoval of the parallel pin
15
D R I L L I N GS E Q U E N C EG U I D E
IS-II active
1,200 RPM, 30~45Ncm,10mm stopper
Twist drilling(Ø2.9)with irrigation
16
D R I L L I N GS E Q U E N C EG U I D E
IS-II activeØ 2.9 depth 10mm hole has appeared
17
D R I L L I N GS E Q U E N C EG U I D E
IS-II active
1,200 RPM, 30~45Ncm,10mm stopper
Twist drilling(Ø3.4)with irrigation
18
D R I L L I N GS E Q U E N C EG U I D E
IS-II activeØ 3.4 depth 10mm hole has appeared
19
D R I L L I N GS E Q U E N C EG U I D E
IS-II active
1,200 RPM, 30~45Ncm,10mm stopper
Twist drilling(Ø3.9)with irrigation
20
D R I L L I N GS E Q U E N C EG U I D E
IS-II activeØ 3.9 depth 10mm hole has appeared
21
D R I L L I N GS E Q U E N C EG U I D E
IS-II active
50 RPM, 50Ncm
Profile tapping(Ø4.5) with irrigation
22
D R I L L I N GS E Q U E N C EG U I D E
IS-II activeØ 4.5 Tapping has appeared
23
D R I L L I N GS E Q U E N C EG U I D E
IS-II active
Ø 4.5 X 10mm, 25~30 RPM,35Ncm
Fixture placementwith contra angle
24
D R I L L I N GS E Q U E N C EG U I D E
IS-II active
25
D R I L L I N GS E Q U E N C EG U I D E
IS-II active
Initial fixation at the apex isexcellent to both immediate placement and immediateloading
Apex
26
2525
Tighten slowly until it reaches 35Ncm
When Implant is being placed w i t h t o r q u e wrench, 35Ncm is suggested with irrigation
27
D R I L L I N GS E Q U E N C EG U I D E
IS-II activeConnecting a coverscrew with Ø1.2 hexdriver (by hand)
28
D R I L L I N GS E Q U E N C EG U I D E
IS-II activeSuture
29
D R I L L I N GS E Q U E N C EG U I D E
IS-II active
1
2
1. Bioseal2. Powerful thread
2 Main Factors for primary fixation duringosseointegration
30
D R I L L I N GS E Q U E N C EG U I D E
IS-II active
Bioseal increases marginal sealing of soft tissue therefore minimizes bone loss.
1.
Powerful thread also calledMagic thread, is specially designed to withstand occlusal and lateral forces.
2.
31
D R I L L I N GS E Q U E N C EG U I D E
IS-II activeAfter osseointegration,remove the cover screwwith Ø1.2 hex driver(by hand)
32
D R I L L I N GS E Q U E N C EG U I D E
IS-II activeConnectinga healing abutment
33
D R I L L I N GS E Q U E N C EG U I D E
IS-II activeAfter soft tissue healing,remove the healingabutment withØ1.2 hex driver(by hand)
34
D R I L L I N GS E Q U E N C EG U I D E
IS-II activeConnecting solidabutment withØ1.2 hex driver(by hand)
35
Connectingtorque ratchet
36
Tightening withtorque ratchet(Max. 35Ncm)
Tighten slowly until it reaches 35Ncm
37
D R I L L I N GS E Q U E N C EG U I D E
IS-II activeSeating impression cap over the abutment
38
Taking impression
39
Impression capcomes out withimpression
40
Insert abutmentlevel lab analoginto impression
41
D R I L L I N GS E Q U E N C EG U I D E
IS-II activeSeating protectivecap over abutmentand fabricatingprovisional restoration
42
D R I L L I N GS E Q U E N C EG U I D E
IS-II activeWhen finalprosthesis is ready,remove theprotective cap
43
D R I L L I N GS E Q U E N C EG U I D E
IS-II activeCementing thefinal restoration
44
D R I L L I N GS E Q U E N C EG U I D E
IS-II active
1. Bioseal2. Coronal Macro Thread3. S.L.A. Surface treatment
Characteristics of IS-II active Implant that has long-term stability
1
2
3
45
D R I L L I N GS E Q U E N C EG U I D E
IS-II active
IS-II active implant is optimized for maintenance of primary stability with its powerful thread.Soft tissue healing process is predictable due toS.L.A surface treatment composed of macro andmicro pores which allow excellent biosealing.
S.L.A. Surface treatmentBioSeal (0.5mm) Coronal Macro Thread
2.1. 3.
46