ENVIRONMENTAL SCANNING ELECTRON MICROSCOPY (ESEM)
Jean-Luc BIAGI
Topography
Chemical contrast
EDX spectrum
Electron Interaction with sample
Auger Electrons
� Versatile all in one SEM
- High vacuum (down to 6e-6 mbar)� Common SEM mode
- Low vacuum (around 1 mbar)� Low pressure gas inside the analysis chamber � Eliminate sample charging
- Extended Low Vacuum mode ESEM (0.1 to 30 mbar)
� Control of humidity conditions
� Quanta FEG 200 (FEI) coupled with an EDS GENESIS XM 4i (EDAX)
� Electron gun : Field Emission Gun (FEG) � Better beam stability (compared to a hot cathode)� High lateral resolution (a few nanometers)
� Samples � All types with minimum preparation� Max. dimensions / weight: 8cm×8cm×5cm / 1kg
Features
� Porosity in brass
Topographical / Morphological characterization
Topographical contrast Chemical contrast
Colour scale = f (Z atomic)Depth of field
Secondary e- Backscattered e-
High vacuum mode
Low vacuum mode
� No sample preparation
� Quick results
� Spatial resolution (tens of nm)
� Magnification of×50 000
easily obtained
� Tick on carbon tape
Characterization of non-conductive samples× 70
× 800
× 200
× 1600
Qualitative analysis: Ti, Zn, O
100 µm
C
FeO
Ca
Elemental chemical analysis and mapping
Elemental chemical analysis
Si Mo Cr Mn Fe Co Ni Cu
Wt% 0.7 2.8 18.9 1.5 64.1 0.2 11.8 0.1
Si Mo Cr Mn Fe Co Ni Cu
Wt% 0.5 2.4 18.5 1.6 64.3 0.1 12.3 0.2
Standard ANSI 316 Stainless Steel
Experimental result of EDS analysis
Semi-quantitative analysis
Ag nanoparticles
X 450.000X 300.000 X 100.000
Nanotubes in a composite
Transmission imaging (STEM detector)8 TEM sample grid positions
� Ultrathin sections placed on copper-grid
� Better resolution than in classic mode
Micro-tensile machine
Increase of stretching
In-situ materials testing: micro-tensile machine
Visualization of sample damage as a function of stretching
DYNAMIC IN-SITU ANALYSES
For more information, see Yves Fleming / Frederic Addiego’spresentation this afternoon
Temperature control: heating stage up to 1500oC
Heating stage
Increase of temperature
Appearance of surface defects
473oC 610oC 625oC 646oC
Material change under temperature effect
DYNAMIC IN-SITU ANALYSES
� Temperature control� Peltier cooling stage down to –25oC
� Humidity (RH) control� Variation of humidity up to 100%
Peltier cooling stage Temperature and/or RH control
Pressure & temperature conditions
to maintain water in a liquid state in ESEM.
Analysis under humid conditions
DYNAMIC IN-SITU ANALYSES
View of the chamber in ESEM configuration
Gaseous Secondary Electron Detector GSED
Cooling stage
Maximum sample size is limited to 6×6 mm2
Analysis under humid conditions
DYNAMIC IN-SITU ANALYSES
SEM column
Peltier cooling stage
Analytical parameters
• Pressure: 6 to 7 Torr • Temperature: 5oC• Humidity: 0 to 100 %
Analysis under humid conditions
DYNAMIC IN-SITU ANALYSES
� Example 1: Hydration / Dehydration
Magnification× 2000
Pressure 6 Torr
Temperature 3oC
Humidity 100 %.
Due to coatings defects (cracks, uncoated area, thickness variation, …), droplets are not of perfect circular shape.
Increase of water drops size
� Example 2: silicon wafer coated with plasma-polymer
Drops appear and increase in diameter randomly.
Analysis under humid conditions
DYNAMIC IN-SITU ANALYSES
20 µm 20 µm
Carbon fibre treated with a water-based solution
containing a catecholamine at different concentrations
30 µm 30 µm
50 µm 10 µm
As received, no surface treated
Carbon fibre treated with a water-based polyelectrolyte solution then with an aged
catecholamine solution during different time of
treatment.µm
Hydrophobic
Hydrophilic
HV 10Kv
Pressure 6.5 Torr
Temperature 4oC
Humidity 100 %
Analysis under humid conditions
DYNAMIC IN-SITU ANALYSES
� Example 3: carbon fibers
With courtesy to Arnaud Martin
� Example 4: Liquid drops on the insect body
20 µm20 µm
20 µm20 µm
50 µm
10 µm
Pressure 5.2 Torr - Temperature 1oC - Humidity 100 %
Analysis under humid conditions
DYNAMIC IN-SITU ANALYSES
� Topographical / Morphological characterization
� Phase distribution / Elemental chemical analyses (EDS)
� Transmission imaging (STEM detector)
� Dynamic in-situ analyses
� In situ materials testing (micro-tensile machine)
� Temperature control (heating stage up to 1500oC)
� Under humid conditions (ESEM)
� Temperature control (Peltier cooling stage down to –25oC)� Humidity (RH) control (variation of humidity up to 100%)
Special features
Type of analyses
SEM HITACHI SU-70 CHARACTERISTICS
Electron gun (FEG)
EDXWDX
Specimen chamber
Source: FEG gun
Equipment:
- Only High Vacuum
- Accelerating voltage range: 0.1-30kV
- SEM detectors (upper, lower)
Multiple (SE/BSE) in-Lens
- Resolution at 15kV = 1nm, 1kV = 1.6nm
- Magnification: 30x - 800.000x
- EDS and WDS spectrometers
(Oxford instruments)
- Samples: polymers (low energy), metals,
composites, ceramics
- Size of specimen: up to 5cm x 5cm x3 cm
- Rotation of sample 360o and tilt 70o
Advantage of Wavelength Dispersive Spectrometry (WDS)
Overlap of Silicon K and Tantalum M lines are easily resolved using WDS (red spectrum)
instead of EDS (green spectrum).
� Better sensitivity, better accuracy
� Requires (time-consuming) element per element calibration
SEM HITACHI SU-70
EDS spectrumWDS spectrum
SEM HITACHI SU-70
E-SEM FEI Quanta FEG 200
Ultimate imaging performances 1 nm Tens of nm
Modes High vacuum only High vacuum, low vacuum (better for the analysis of insulatingsamples)
Sample size 5cm x 5cm x 3 cm 8cm × 8cm × 5cm / 1kg
Detectors EDS, WDS EDS only
Accessories Mechanical strain TemperatureRelative humidity (Wet SEM)
� High performance SEM
� Highly
flexible SEM
Science is sometimes art