charging contamination
Post on 14-Nov-2021
2 Views
Preview:
TRANSCRIPT
Charging Contamination
Charge balance in the sample
I bI b
I b
sc
Electrons cannot becreated or destroyedso currents at a pointmust sum to zero. Thecurrent flow to earth Iscis the difference betweenthe in and out currents
E2 values
Material E2(keV) Material E2 (keV)
Resist 0.55 Kapton 0.4Resist on Si 1.10 Polysulfone 1.1PMMA 1.6 Nylon 1.2Pyrex glass 1.9 Polystyrene 1.3Cr on glass 2.0 Polyethylene 1.5GaAs 2.6 PVC 1.65Sapphire 2.9 PTFE 1.8Quartz 3.0 Teflon 1.8
Determining E2 in the Low Voltage SEM
1. Set the magnification to 100x and scan at TV rate
2. Increase the magnification to 1000x as quickly as possible
3. Count to five
4. Drop back to 100x magnification
5. Look at the scan square that is visible in the center of the screen .....
Negative charging
» If the scan square is brighter than the background then the sample is charging negative and the beam energy is greater than E2 (or - just possibly - less than E1)
Paper pulp at 2.5keV
Positive Charging
» If the scan square is dark compared to the background then the sample is charging positive and the beam energy is less than E2 (and greater than E1)
Paper pulp at 1.2keV
Imaging non-conductors
» On a new SEM this will be the lowest available energy
» On older machines you must decide how low to go before the performance becomes too poor to be useful for the purpose intended
» The goal is to avoid implanting charge deep beneath the surface. If this is allowed to occur then stable imaging may never be achieved.
» Step #1 - Set the SEM to the lowest operating energy
Next……...
» If the sample is charging positive (i.e. a dark scan square) then E1< E<E2. Increase the beam energy and proceed to image
» If sample is charging negatively (i.e. bright scan square) then E>E2.
» Since we cannot reduce E
any further go on to step 3.
» Step #2 - Determine the charging state of the sample using the scan square test
Step 3
» Tilt the sample to 45 degrees and repeat the usual scan square test
» Can E2 be reached now?
» E2() = E2(0)/cos2
so tilting by 45 degrees raises E2 by a factor of 2x
» But ..because E2 varies with the angle of incidence the ‘no charge’ condition can never be satisfied everywhere on the surface at the same time and charging will always occur
Note that tilting the sample reduces charging at all energies
Living with charging - #1
» Reduce the beam current since the charging varies directly with IB
» Use a smaller aperture, or reduce the gun emission current
» Reduces the S/N ratio so longer scan times may be required
Living with charging - #2
» Reduce the magnification
» This minimizes Dynamic Charging (internal charge production from electron-hole pairs). The magnitude of this depends on the dose and hence on the magnification
» Dynamic charging is worst when E0 is close to the E2 value
» Limits resolution by limiting magnification
Living with charging - #3
Charging is time dependent because the sample acts like a leaky capacitor
Samples charge up more quickly than they discharge
Depending on the scan rate and the relative charge up and decay rates samples can
float at a steady potential or
gradually acquire a charge
Generally at TV scan rates sample potentials float in a stable manner so focussing and stigmation are possible
Beam On
Cha
rge
Time
One frame Fast rise
Fast decay
Fast rise
Slow decay
Choosing a detector
» The choice of detector can have a significant effect on the apparent severity of charging
» The conventional ET (Everhart - Thornley) detector is much less sensitive to charging than...
Individual polymer macro- molecules on Si at 1.5keV -
Lower (ET) detector
Upper detector
» …a through the lens detector. This is because TTL systems act as simple SE spectrometers and preferentially select low energy electrons
» Note however that charging can be a useful form of contrast mechanism when properly employed
Same area as before, TTL detector
BSE imaging to avoid charging
» Backscattered electrons are less affected by charging and offer the same resolution at LV
» Newer technologies such as conversion plates, and ExB filters, for BSE actually improve in efficiency as the beam energy is reduced, so using this mode to avoid charging problems becomes a good choice
Uncoated Teflon S-4700 ExB BSE image
If all else fails…..coat the sample
» Coatings do not make the sample a conductor
» They form a ground plane - i.e. the free electrons in the metal move so as to eliminate the external field
» The charge is not eliminated but the disruptive field is removed
» Successful coating means paying attention to the details...
-----------------
-----------------
Charge in sample
Field deflectsincident and exit electrons
++++++++
coating
'image charge'+++++
metal is equipotentia ground plane
NO EXTERNAL FIELDS
Charge insample
Field deflects electrons
ground planeField lines do not leak away from the surface
To ensure good coatings (for high resolution or low voltage)
» Keep it clean - wipe the glass vessel clean after every run, and clean the anodes weekly
» Keep it slow - reduce the gas pressure and/or the anode voltage till plasma just stays on
» Keep it thin - thicker coatings do not work better and they obscure surface detail. Aim for no more than 5nm of Au/Pd, 2nm of Cr
» Keep it dry - the argon gas (never air!) must be perfectly dry. Check the color of the plasma.
and furthermore…..
» All that glitters - do not use pure gold because of its high surface energy. Use Au-Pd, Ir, or Pt to ensure good, thin, particulate, films
» Evaporated Carbon is a contaminant not a coating. Carbon is a poor conductor, produces only a small number of secondary electrons, evaporated carbon is usually mixed with a filler compound, and the thickness cannot be well controlled. Either use a ion sputter coater for C or avoid it altogether
Unwanted Beam Interactions
Radiation Damage
IonizationDisplacementHeating
ContaminationEtching
Intrinsic to electronbeam irradiation
Results fromvacuum problems
Both are usually important
Radiolysis
» Ionization damage is most important threat to organic, and some inorganic, materials.
» Electrons are the most intense source of ionizing radiation available - the typical dose in an SEM is equivalent to standing 6 foot from a 10 megaton H-bomb
Compare SEM to Sun and SPEAR
Effects of radiolysis
» Destroys the crystalline structure of polymers, and other organic crystals, leaving them amorphous
» The probability of radiolysis is 10x to 100x bigger than the chance of generating an X-ray
» Damage competes with signal generation - damage usually wins
Damage to Protein Protoxein crystals from imaging
Contamination - Etching
» Contamination is beam induced polymerization of the hydrocarbons present on the sample surface
» Etching is the removal of surface layer by impact of ions (C + H2 O 2- --> CO + H2 )
» Both phenomena are affected by surface charging and often occur together
» Both are temperature dependent
Contamination and Etching
Electrons break down the hydrocarbon film. The residue charges +ve and the field pulls in fresh material for radiolysis. If water
vapor is present then H2 02- ions go to the + ve charged region and etch that area away
Low magnification
» At low magnification the hydrocarbon film is polymerized into a thin sheet.
» This will charge positive (and so look black in the SE image) but is not a serious problem
Schematic of contamination build-up at low magnification
scans
High magnification
» At high magnification the contamination grows a cone which prevents the beam reaching the surface
» So avoid spot mode - always keep the beam scanning the sample
» Try and pre-pump samples before use
» Keep your hands off the sample
~ 0.03
Virtue of necessity..
» Contamination cones can grow to a height of hundreds of angstroms and are very tough - used for high resolution AFM tips
» Prevent this growth by irradiating the area at low magnification before going up to a high magnification
300A high cone grown on a silicon wafer in 5 minutes
exposure
The Cold Finger
» Standard fitting on S-4800 & S- 4700, available as an option for the S-4500
» The finger is held at LN2 temperatures, a few mm from the specimen surface
» After filling the cold finger allow the sample enough time to reach thermal equilibrium before starting to image
top related