Carotid stenting-in stent restenosis and

treatment modalities

Jagiellonian University Institute of Cardiology,

John Paul II Hospital Krakow, Poland

Prof. Piotr Pieniazek MD, PhD

Disclosure

Speaker name: PIOTR PIENIAZEK.

I have the following potential conflicts of interest to report:

X Consulting

Employment in industry

Stockholder of a healthcare company

Owner of a healthcare company

Other(s)

I do not have any potential conflict of interest

• comprehensive patient evaluation including non-invasive imaging:

• extra- and intracranial Doppler

• extra- and intracranial CT angio

• brain perfusion (MRI)

• independent neurological consultation (NIHSS, MMSE, Rankin Scale)

• patient-and-lesion-tailored neuroprotection before meshstent era

(Roadsaver & CGuard) - 2015!!!!!!

• independent histologic evaluation of captured material

• coronary angiography prior to CAS (except pts. after CABG or recent PCI)

rigorous follow-up is mandatory !!!!!

The Krakow Program of Stroke Prevention

by Carotid Artery Stenting (from Jan 2001)

4055 CAS procedures

in a single Krakow Center

The largest meta-analysis comparing outcome in different stent

design before meshstents era.

In-stent restenosis after carotid stenting is not a trivial problem as

it still exists.

• Occurs less frequently than in subclavian, vertebral, renal or peripheral arteries.

• Is less frequent than restenosis after carotid endarterectomy.

• 10% according guidelines is not acceptable.

• Treatment modality in-stent restenosis has still growing progress.

• Re- PTA for in-stent restenosis needs always temporary NPD.

• In recurrent in-stent restenosis of label devices need to be use.

• Rigorous US investigation in such patients is mandatory.

Never compare incomparable groups.

Then it's easy to get compromising conclusions!!!

12m. follow-up

Treatment modality of ISR

Patient 78 y. with bilateral ICA stenosis after syncope two months earlier with recurrent -

bilateral RESTENOSIS

LICA ReRe in-stent restenosis: 80% IVUS stenosis; MLA 4.2mm . CAS - 20 Feb. 2018

Viatrac 4.5/20mm STENTYS X position 3.0-3.5/27mm, MAVERIC 5.0/20mm & FilterWire

Diffuse in-stent & edge in-stent restenosis. Standard balloon & DCB5.0/40mm 3 inflation 8 atm for 45 min!! Our standard of treatment for in-stent restenosis

Standard

Balloon

4,5/20mm

DCB

5.0/40mm

Filter

Our current approach with Carotid In-Stent Restenosis. DCB is used after

excellent artery preparation with Stantard, Cutting or NC Balloon. Temporary

NPD use is mandatory.

DEB

Standard

Balloon

NC Balloon

24 Atm

COMPARE : Does Ranger DCB with dose density of

2μg/mm2 have similar efficacy compared to and In.

Pact DCB with dose density of 3.5μg/mm2?

Important knowledge after publication from Dec.2018 which should be considered

Ranger

5mm

In.Pact

5mm

40mm 1301 2553

Nominal dose (ug/device) to treat a typical

lesion.

Lenght

Ranger5mm

1301

All balloons have CE registrations

Treatment modality for in-stent restenosis.

Balloon angioplasty always with bigger diameter.

Cutting Balloon (only whey the standard balloon slides away from

lesion).

Second self expanding stent (stent in stent technique, do not use

Carotid Wallstent).

Drug eluting balloon with Paclitaxel (problem with 45 sec.

inflation).

Self expanding sirolimus coronary stent – STENTYS X position

(withdrawn from the market in 2019 ) but the new DCB is coming.

Surgery should be avoided!!!!

Take home message :

• Restenosis after carotid stenting is not a frequent complication

and should not exceed 5%

• Treatment strategy for in-stent restenosis is changing

• New technological solutions should be taken for consideration

• We use DCB with Paclitaxel as a standard choice for in-stent

restenosis treatment

• For restenosis after CAS or CAE do not use Carotid Wallstent

Thank you

John Paul II Hospital Krakow PL.