This laboratory maintains A2LA accreditation to ISOIIEC 17025 for the specificcalibrations listed in Certificate # 2166.01 and meets the relevant requirementsof ISO 9001:2008
IACCREDITEDlCALlBRATION CERT #2166.01
FL.UKE @ Everett Service Center1420 75th St. SW
Everett, Washington 98203 Calibration CertificateUSAe
NQA ISO 9001:2008 (10100/2)
Description:Manufacturer:
Model:
Serial Number:
Customer Name:METRICA SA de CV
City, State: MONTERREY N L, MXCustomer Item ID: C-801PO Number: CIARMA Number: 4698843
Result Summary: MV<Spec = VESReceived Date: 29 March 2011
CALlBRATORFLUKE5500A 120
6265013
Certificate Number: 945162-6265013:1301616947Date of Calibration: 01 April 2011
Date of Certificate: 01 April2011Recommended Due Date:Procedure Name:
FLUKE 5500A:(1 YEAR) ACAL VER COMBINEProcedure Revision: 4.1
Data Type: FOUND-LEFTTemperature: 22.9 °Celsius
Relative Humidity: 47 %
This calibration certificate may contain data that is not covered by the A2LA Scope of Accreditation. The unaccredited material,where applicable, is indicated by an asterisk (*), or confined to clearly marked sections. PASS I FAIL tests are not accredited.If there is non accredited material, the non accredited data is compliant with ANSI/NCSLI Z540.1-1994(R2002) requirements.
Measurement uncertainties at the time of test are given in the following pages, where applicable. They are calculated in accordancewith the method described in NIST TN1297, for a confidence level of approximately 95% using a coverage factor of approximately 2(K=2).
In the attached measurement results, deviation may be expressed with units, Measured Value (MV) - Nominal Value (NV) or as aproportion ofthe nominal value «MV-NV)/NV), expressed without units with a scalar multiplier such as % (0.01), or as a ratio oftheunits (mAlA, INN, etc.) Descriptions such as IlAlA, IlVN, and others, where used to annotate results or column headings are thepreferred replacements for what was historically labeled as "ppm" or parts-per-million and described the results in that column,unless otherwise noted by units symbols.
The Data type that could be found in this certificate must be interpreted as:As-Found - Calibration data collected before the unit is adjusted and/or repaired.
. As-Left - Calibration data collected after the unit is adjusted and/or repaired.Found-Left - Calibration data collected without any adjustment and/or repair performed.
No statement of compliance with specifications is made or implied on this certificate. However, results are reviewed, if applicable, toestablish where any measurement results exceeded the manufacturer's specifications and to communicate results by means of thiscertificate. Measured values areater than the Manufacturer's scecification (Scec) are indicated bv "!".
Comments:
~~Authorized Slgnature
Sr. Spec;alty Tech
Metrology Technician
Fluke Corporation
1420 75th 5treet 5W, Everett WA 98203 USA
Telephone Facsirnile Internet Page 1 of 15
Rev 7.4, 02123/11888.993.5853 425.446.6390 www.f1uke.com
This certificate applies to only the item identified and shall not be reproduced other than in tull, without the specific written approval byFluke Corporation. Calibration certificates without signatures are not valid. This certificate shall not be used to claim product .endorsement bv A2LA.
FL.UKE@ Certificate Number:945162-6265013:1301616947
Calibration Date:1-Apr-ll
Traceability InformationFor each parameter listed below the calibration was conducted using an unbroken chain of standards to:
DCVoltageThe Voltage Reference standard group, traceable to the Fluke PrimaryElectricalStandards Laboratory,which is traceable to the U.S. representationof the volt, through the internationallyaccepted value of the Josephson constant Kj=483597.9GHzIVand a 10 VoltJosephson ArrayVoltageStandard.
Frequency and PeriodA GPS Disciplined Rubidiumoscillatorfrequency standard which is traceable to the Nationallnstitute of Standards and Technology(NIST).
ACVoltage, Resistance, DCCurrent, ACCurrent, Capacitance, Inductance, PhaseThe Fluke Primary Electrical Standards Laboratory, which is traceable to the NIST.
ACVoltage FlatnessThe Fluke Primary Electrical Standards Laboratory or Agilent Technologies Standards Laboratory which are traceable to NIST.
Humidity
The Vaisala Measurement Standards Laboratory Primary Salt calibration bath, with traceability based on the physical phenomena in which theequilibrium relative humidity values associated with certain saturated salt solutions are known.
Rise Time
The Tektronix GmbH Calibration Laboratory which is traceable to the Physikalisch-Technische Bundensanstalt (PTB), and the National PhysicalLaboratory (NPL).
Radiation TemperatureTraceableto NIST,PTB,andthe FlukePrimaryTemperatureStandardsLaboratory.
Contact TemperatureTraceable to the Fluke Primary Temperature Standards Laboratory which is traceable to the NIST.
Gas FlowThe Fluke Primary Gas Flow Laboratory which is traceable to the NIST.
Pressure
The Fluke Primary Pressure Laboratory (Phoenix), which is traceable to the Laboratoire National de Metrologie et D'Essais (LNE) and PTB.
Standards Used
Asset#10066
10242
10383
10384
13821
14386
6619
7602
7678
9220
9532
9567
9866
9911
9911S
J277
Fluke Corporation
Instrument ModelHEWLETI PACKARD 3458A DIGITAL MULTIMETER
HEWLETI PACKARD 3458A DIGITAL MULTIMETER
FLUKE Y5020 CURRENT SHUNT/SL013
FLUKE Y5020 CURRENT SHUNT/SL013
KROHN-HITE CORP 6900B DISTORTION ANAL YZER
OMEGA TJ36ICINI8UI2SMPWM J PROBE
FLUKE 5450A RESISTANCE CALIBRATOR
FLUKE 8842A MUL TIMETER
FLUKE 5450A RESISTANCE CALIBRATOR
FLUKE 8520AIPRT DIGITAL MULTIMETER W/PRT
FLUKE PM6680B TIMER/COUNTER
FLUKE 5500A CALIBRATOR
FLUKE PM6304C PROGRAMMABLE RCL METER
FLUKE 5790A AC MEASUREMENT STANDARD
FLUKE 5790A/A40 SET ACI MEASUREMENT SYSTEM
CLARKE-HESS 6000 PHASE METER
Cal Date08 March 2011
II February 2011
27 October 2010
07 December 2010
08 September 2010
20 December 2010
14 March 2011
25 September 2010
25 January 2011
19 October 2010
31 August 2010
08 October 2010
26 January 2011
21 January 2011
09 March 2011
04 August 2010
Telephone Facsimile Internet
1420 75th Street SW, Everett WA 98203 USA www.fluke.com888.993.5853 425.446.6390
CalDue08 September 2011
11 November 2011
27 October 2011
07 June 2011
08 September 2011
20 December 2011
14 July 2011
25 September 2011
25 April 2011
19 April 2011
31 August 2011
08 October 2011
26 July 2011
21 July 2011
09 April2011
04 August 2011
Page 2 of 15
Rev 7.4, 02/23/11
FLUKE@ Certificate Number: Calibration Date:945162-6265013:1301616947 l-Apr-ll
Calibration ResultsManufacturer's
Nominal Measured Measurement Specifications
Function/Range Value Value Uncertainty Lower Limit Upper Limit
1 Year Specification Limits shown
DC VOLTAGE ACCURACY TEST (NORMAL)
330 mV Range0.0000 mV 0.0000 0.00028 2.2e-007 V -0.00300 0.00300
300.000 mV 300.000 299.9994 1.2e-006 V 299.9790 300.0210
-300.000 mV -300.000 -299.9953 1.2e-006 V -300.0210 -299.9790
3.3 V Range0.000000 V 0.000000 0.0000012 8.4e-007 V -0.0000050 0.0000050
0.300000 V 0.300000 0.3000006 1.2e-006 V 0.2999800 0.3000200
-0.300000 V -0.300000 -0.2999990 1.2e-006 V -0.3000200 -0.2999800
3.000000 V 3.000000 2.9999951 9.0e-006 V 2.9998450 3.0001550
-3.000000 V -3.000000 -3.0000256 9.0e-006 V -3.0001550 -2.9998450
33 V Range0.00000 V 0.00000 0.000007 1.0e-005 V -0.000050 0.000050
30.00000 V 30.00000 29.999971 1.7e-004 V 29.998450 30.001550
-30.00000 V -30.00000 -30.000263 1.7e-004 V -30.001550 -29.998450
330 V Range30.0000 V 30.0000 30.00007 1.7e-004 V 29.99785 30.00215
-30.0000 V -30.0000 -29.99998 1.7e-004 V -30.00215 -29.99785
300.0000 V 300.0000 300.00001 1.8e-003 V 299.98300 300.01700
-300.0000 V -300.0000 -300.00304 1.8e-003 V -300.01700 -299.98300
1000 V Range100.0000 V 100.0000 100.00006 5.6e-004 V 99.99300 100.00700
-100.0000 V -100.0000 -99.99994 5.5e-004 V -100.00700 -99.99300
1000.000 V 1000.000 999.9990 1.7e-002 V 999.9435 1000.0565
-1000.000 V -1000.000 -1000.0106 1.7e-002 V -1000.0565 -999.9435
DC VOLTAGE ACCURACY TEST (AUX)
330 mV Range0.000 mV 0.000 -0.0189 3.ge-006 V -0.3500 0.3500
300.000 mV 300.000 299.9883 9.8e-006 V 299.5300 300.4700
-300.000 mV -300.000 -300.0234 9.8e-006 V -300.4700 -299.5300
3.3 V Range0.330000 V 0.330000 0.3299883 5.5e-006 V 0.3295180 0.3304820
-0.330000 V -0.330000 -0.3300170 5.5e-006 V -0.3304820 -O.3295180
3.00000 V 3.00000 3.000018 3.3e-005 V 2.998450 3.001550
-3.00000 V -3.00000 -3.000059 3.3e-005 V -3.001550 -2.998450
DC OFFSET ACCURACYTEST
50 mV Range0.000 mV 0.000 0.0021 4.1e-006 V -0.0330 0.0330
50.0000 mV 50.0000 50.01025 1.1e-006 V 49.91700 50.08300
-50.0000 mV -50.0000 -50.01090 7.8e-007 V -50.08300 -49.91700
500 mV Range
Fluke Corporation Telephone Facsimile Internet Page 3 of 15
1420 75th Street SW, Everett WA 98203 USA 888.993.5853 425.446.6390 www.fIuke.com Rev 7.4, 02/23/11
FL.UKE@ Certificate Number: Calibration Date:
945162-6265013:1301616947 l-Apr-ll
Calibration ResultsManufacturer's
Nominal Measured Measurement Specifications
Function/Range Value Value Uncertainty Lower Limit Upper Limit
0.00 mV 0.00 0.025 6.6e-005 V -0.330 0.330
500.00 mV 500.00 500.150 3.2e-005 V 499.170 500.830
-500.00 mV -500.00 -500.206 2.ge-005 V -500.830 -499.170
5 V Range0.0000 V 0.0000 0.00050 7.6e-004 V -0.00330 0.00330
5.0000 V 5.0000 5.00233 5.3e-004 V 4.99170 5.00830
-5.0000 V -5.0000 -5.00158 2.1e-004 V -5.00830 -4.99170
50 V Range0.000 V 0.000 -0.0010 1.1e-003 V -0.0330 0.0330
45.0000 V . 45.0000 45.02202 3.0e-004 V 44.92202 45.07798
-45.0000 V -45.0000 -45.01758 3.ge-004 V -45.07798 -44.92202
DC CURRENT ACCURACY TEST
3.3 mA Range0.00000 mA 0.00000 0.000006 1.8e-008 A -0.000050 0;000050
3.00000 mA 3.00000 2.999988 2.ge-008 A 2.999559 3.000441
-3.00000 mA -3.00000 -2.999951 2.ge-008 A -3.000441 -2.999559
33 mA Range0.00000 mA 0.00000 0.000041 7.ge-008 A -0.000250 0.000250
30.0000 mA 30.0000 29.99982 3.2e-007 A 29.99676 30.00324
-30.0000 mA -30.0000 -29.99950 3.2e-007 A -30.00324 -29.99676
330 mA Range0.0000 mA 0.0000 0.00049 1.6e-007 A -0.00330 0.00330
300.000 mA 300.000 299.9977 7.ge-006 A 299.9667 300.0333
-300.000 mA -300.000 -299.9924 7.ge-006 A -300.0333 -299.9667
2.2 A Range0.0000000 A 0.0000000 0.00000p21 2.6e-007 A -0.00004400 0.00004400
2.000000 A 2.000000 2.0000095 1.3e-004 A 1.9993560 2.0006440
-2.000000 A -2.000000 -2.0000249 1.3e-004 A -2.0006440 -1.9993560
11 A Range0.0000000 A 0.0000000 0.00002258 1.5e-006 A -0.00033000 0.00033000
10.0000 A 10.0000 9.99957 6.ge-004 A 9.99367 10.00633
-10.0000 A -10.0000 -10.00000 6.6e-004 A -10.00633 -9.99367
DC POWER ACCURACY TEST (NORMAL)
20 mV @ 2.9 A20.0000 mV 20.0000 20.00061 3.7e-007 V 19.99580 20.00420
20mV@11 A20.0000 mV 20.0000 20.00050 3.7e-007 V 19.99580 20.00420
DC POWER ACCURACY TEST (AUX)
0.1 mA @ 1000 V0.100000 mA 0.100000 0.1000035 2.3e-009 A 0.0999370 0.1000630
1 mA@ 1000 V1.00000 mA 1.00000 0.999993 2.0e-008 A 0.999820 1.000180
FlukeCorporation Telephone Facsimile Internet Page4 of 15
1420 75th Street SW, Everett WA 98203 USA 888.993.5853 425.446.6390 www.fluke.com Rev 7.4, 02/23/11
FL.UKE@ Certificate Number: Calibration Date:945162-6265013:1301616947 l-Apr-ll
Calibration ResultsManufacturer's
Nominal Measured Measurement Specifications
Function/Range Value Value Uncertainty Lower Limit Upper Limit
2A @ 300 V2.00000 A 2.00000 2.000009 1.3e-004 A 1.999356 2.000644
10 A @ 1000 V10.0000 A 10.0000 9.99969 6.6e-004 A 9.99367 10.00633
PHASE ACCURACY TEST (Voltage ys. Voltage)
O degrees
0.00 o @ 60 Hz 0.00 0.054 5.2e-002 o -0.150 0.150
0.00 o @ 400 Hz 0.00 -0.040 5.2e-002 o -0.900 0.900
0.00 o @ 1 kHz 0.00 -0.060 5.2e-002 o -2.000 2.000
0.00 o @ 5 kHz 0.00 -0.370 5.2e-002o -6.000 6.000
0.00 o @ 10 kHz 0.00 -0.030 5.2e-002 o -10.000 10.000
60 degrees
60.00 o @ 60 Hz 60.00 59.982 6.4e-002 o 59.850 60.150
60.00 o @ 400 Hz 60.00 59.970 5.2e-002 o 59.100 60.900
60.00 o @ 1 kHz 60.00 59.930 5.2e-002 o 58.000 62.000
60.00 o @ 5 kHz 60.00 59.618 5.2e-002 o 54.000 66.000
60.00o @ 10kHz 60.00 60.070 5.2e-002o 50.000 70.000
90 degrees
90.00 o @ 60 Hz 90.00 90.030 5.2e-002 o 89.850 90.150
90.00o @400 Hz 90.00 89.960 5.2e-002o 89.100 90.900
90.00o @ 1 kHz 90.00 89.940 5.2e-002o 88.000 92.000
90.00o @5 kHz 90.00 89.630 5.2e-002o 84.000 96.000
90.00o @ 10kHz 90.00 90.080 5.2e-002o 80.000 100.000
PHASE ACCURACY TEST (Voltage ys. Current)
0.00 degrees (33 V, 300 mA @ 65 Hz)
0.00 o @ 65 Hz 0.00 0.030 5.2e-002o -0.150 0.150
0.00 degrees (33 V, 2 A @ 65 Hz)
0.00 degrees (33 V, 2 A @ 65 Hz)
0.00 o @65 Hz 0.00 0.018 5.2e-002o -0.150 0.150
0.00 degrees (33 V, 5 A @ 65 Hz)
0.00 o @ 65 Hz 0.00 -0.004 5.3e-002o -0.150 0.150
0.00 degrees (33 V, 5 A @ 400 Hz)
0.00 o @400 Hz 0.00 -0.070 6.8e-002o -0.900 0.900
FREQUENCY ACCURACY TEST
119.00000 Hz @ 3 V 119.00000 119.000301 4.ge-005 Hz 118.996025 119.003975
120.00000 Hz @ 3 V 120.00000 119.999394 1.8e-005 Hz 119.996000 120.004000
1.00000000 kHz @ 3 V 1.00000000 1.000000484 2.0e-005 Hz 0.999974000 1.000026000
100.0000000 kHz @ 3 V 100.0000000 99.99998690 6.0e-005 Hz 99.99748500 100.00251500
DISTORTION AND NOISE TEST (NORMAL)
300 mV Range
0.000 % @ 10 Hz 0.000 0.0650 0.0000 0.1800 *
0.000 % @ 45 Hz 0.000 0.0205 0.0000 0.0650 *
Fluke Corporation Telephone Facsimile Internet Page 5 of 15
1420 75th Street SW. Everett WA98203 USA 888.993.5853 425.446.6390 www.fIuke.com Rev 7.4,02/23/11
...
FLUKE@ Certifica te Number: Calibration Date:
945162-6265013:1301616947 l-Apr-ll
Calibration ResultsManufacturer's
Nominal Measured Measurement Specifications
Function/Range Value Value Uncertainty Lower Limit Upper Limit
0.000 % @ 10 kHz 0.000 0.0204 0.0000 0.0650 *
0.000 % @ 20 kHz 0.000 0.0350 0.0000 0.0900 *0.000 % @ 50 kHz 0.000 0.0854 0.0000 0.1800 *0.000 % @ 100 kHz 0.000 0.1177 0.0000 0.2300 *
0.000 % @ 500 kHz 0.000 0.1143 0.0000 0.9300 *300 V Range0.000 % @ 45 Hz 0.000 0.0215 0.0000 0.1583 *0.000 % @ 1 kHz 0.000 0.0238 0.0000 0.0583 *0.000 % @ 10 kHz 0.000 0.0275 0.0000 0.0583 *
0.000 % @ 20 kHz 0.000 0.0514 0.0000 0.6083 *
DISTORTION AND NOISE TEST (AUX)
300 mV Range
0.000 % @ 10 Hz 0.000 0.0597 0.0000 0.2667 *
0.000 % @ 20 Hz 0.000 0.0154 0.0000 0.1267 *0.000 % @ 45 Hz 0.000 0.0182 0.0000 0.1267 *0.000 % @ 1 kHz 0.000 0.0256 0.0000 0.1467 *
0.000 % @ 5 kHz 0.000 0.1630 0.0000 0.3667 *
0.000 % @ 10 kHz 0.000 0.4313 0.0000 0.6667 *
3 V Range0.000 % @ 10 Hz 0.000 0.0649 0.0000 0.2067 *0.000 % @ 20 Hz 0.000 0.0148 0.0000 0.0667 *
0.000 % @ 45 Hz 0.000 0.0173 0.0000 0.0667 *0.000 % @ 1 kHz 0.000 0.0271 0.0000 0.0867 *0.000 % @ 5 kHz 0.000 0.0927 0.0000 0.3067 *0.000 % @ 10 kHz 0.000 0.3302 0.0000 0.6067 *
1 Year Specification Limits shown
AC VOLTAGE ACCURACY TEST (NORMAL)
33mV Range
30.00 mV @ 9.5 Hz 30.00 30.153 3.0e-005 V 28.335 31.665
30.000 mV @ 10 Hz 30.000 30.0017 6.8e-006 V 29.8750 30.1250
30.000 mV @ 45 Hz 30.000 29.9993 2.7e-006 V 29.9350 30.0650
30.000 mV @ 1 kHz 30.000 29.9991 2.7e-006 V 29.9350 30.0650
30.000 mV@ 10 kHz 30.000 29.9992 2.7e-006 V 29.9350 30.0650
30.000 mV@20 kHz 30.000 30.0012 2.7e-006 V 29.9200 30.0800
30.000 mV @ 50 kHz 30.000 30.0077 4.8e-006 V 29.9050 30.0950
30.000 mV @ 100 kHz 30.000 30.0139 8.1e-006 V 29.8620 30.1380
30.00 mV@ 500 kHz 30.00 29.988 2.3e-005 V 29.640 30.360
330mV Range300.0 mV @ 9.5 Hz 300.0 300.39 3.0e-004 V 283.35 316.65
300.00 mV @ 10 Hz 300.00 300.001 5.0e-005 V 299.200 300.800
300.000 mV @ 45 Hz 300.000 300.0014 9.1e-006 V 299.8300 300.1700
300.000 mV @ 1 kHz 300.000 300.0006 9.1e-006 V 299.8300 300.1700
Fluke Corporation Telephone Facsimile Internet Page 6 of 15
1420 75th Street SW, Everett WA98203 USA 888.993.5853 425.446.6390 www.fluke.com Rev 7.4,02/23/11
FL-UKE@ Certifica te Number: Calibration Date:945162-6265013:1301616947 l-Apr-ll
Calibration ResultsManufacturer's
Nominal Measured Measurement Specifications
Function/Range Value Value Uncertainty Lower Limit Upper Limit
300.000 mV @ 10 kHz 300.000 299.9930 9.0e-006 V 299.8300 300.1700
300.000 mV @ 20 kHz 300.000 299.9810 9.0e-006 V 299.6800 300.3200
300.00 mV @ 50 kHz 300.00 299.953 1.4e-005 V 299.480 300.520
300.00 mV @ 100 kHz 300.00 299.907 2.1 e-005 V 299.110 300.890
300.00 mV @ 500 kHz 300.00 299.372 7.8e-D05 V 297.570 302.430
3.3V Range
3.000 V @ 9.5 Hz 3.000 3.0035 3.0e-D03 V 2.8335 3.1665
3.0000 V@ 10 Hz 3.0000 2.99995 4.7e-004 V 2.99525 3.00475
3.00000 V @ 45 Hz 3.00000 2.999988 6.0e-005 V 2.999040 3.000960
3.00000 V @ 1 kHz 3.00000 3.000000 6.0e-D05V 2.999040 3.000960
3.00000 V @ 10kHz 3.00000 2.999932 6.0e-005 V 2.999040 3.000960
3.00000 V @ 20 kHz 3.00000 2.999846 6.0e-005 V 2.997540 3.002460
3.0000 V @ 50 kHz 3.0000 2.99964 1.4e-004 V 2.99550 3.00450
3.0000 V @ 100 kHz 3.0000 2.99942 2.1e-004 V 2.99110 3.00890
3.0000 V @ 500 kHz 3.0000 2.99913 9.6e-004 V 2.98170 3.01830
33V Range
30.00 V @ 9.5 Hz 30.00 30.008 3.0e-002 V 28.335 31.665
30.000 V @ 10 Hz 30.000 29.9994 4.7e-003 V 29.9525 30.0475
30.0000 V @ 45 Hz 30.0000 29.99998 7.5e-D04V 29.98740 30.01260
30.0000 V @ 1 kHz 30.0000 29.99844 7.6e-004 V 29.98740 30.01260
30.0000 V @ 10 kHz 30.0000 29.99840 7.5e-004 V 29.98740 30.01260
30.0000 V @ 20 kHz 30.0000 29.99836 7.6e-004 V 29.97340 30.02660
30.000 V @ 50 kHz 30.000 29.9971 1.4e-003 V 29.9380 30.0620
30.000 V @ 100 kHz 30.000 29.9987 2.3e-003 V 29.9110 30.0890
330V Range300.000 V @ 45 Hz 300.000 300.0094 9.7e-003 V 299.8434 300.1566
300.000 V @ 1 kHz 300.000 299.9848 9.7e-003 V 299.8434 300.1566
300.000 V @ 10 kHz 300.000 299.9776 9.7e-003 V 299.7450 300.2550
300.000 V @ 20 kHz 300.000 299.9882 9.7e-003 V 299.6970 300.3030
1000V Range
1000.00 V @ 45 Hz 1000.00 1000.050 3.2e-002 V 999.420 1000.580
1000.00 V @ 1 kHz 1000.00 1000.021 3.2e-002 V 999.420 1000.580
1000.00 V @ 5 kHz 1000.00 999.988 3.2e-002 V 997.900 1002.100
1000.00 V @ 10 kHz 1000.00 999.965 3.2e-002 V 997.500 1002.500
AC VOLTAGE ACCURACY TEST, SQUAREWAVE (NORMAL)
66mV Range60.000 mVpp @ 10Hz 60.000 60.0274 59.5200 60.4800 *60.000 mVpp @ 45 Hz 60.000 60.0036 59.6850 60.3150 *60.000 mVpp @ 1 kHz 60.000 60.0136 59.6850 60.3150 *60.000 mVpp @ 20 kHz 60.000 60.2316 59.5350 60.4650 *60.000 mVpp @ 100 kHz 60.000 61.0792 56.6700 63.3300 *660mV Range
Fluke Corporation Telephone Facsimile Internet Page 7 of 15
1420 75th Street SW, Everett WA 98203 USA 888.993.5853 425.446.6390 www.fluke.com Rev 7.4, 02/23/11
FL.UKE@ Certifica te Number: Calibration Date:945162-6265013:1301616947 l-Apr-ll
Calibration ResultsManufacturer's
Nominal Measured Measurement Specifications
Function/Range Value Value Uncertainty Lower Limit Upper Limit
600.000 mVpp @ 10 Hz 600.000 600.2320 595.2000 604.8000 *600.000 mVpp @ 45 Hz 600.000 600.0240 596.8500 603.1500 *600.000 mVpp @ 1 kHz 600.000 600.1180 596.8500 603.1500 *600.000 mVpp @ 20 kHz 600.000 602.1240 595.3500 604.6500 *600.000 mVpp @ 100 kHz 600.000 610.0640 566.7000 633.3000 *
6.6V Range
6.00000 Vpp @ 10 Hz 6.00000 6.002180 5.952000 6.048000 *6.00000 Vpp @ 45 Hz 6.00000 6.000220 5.968500 6.031500 *6.00000 Vpp @ 1 kHz 6.00000 6.001060 5.968500 6.031500 *6.00000 Vpp @ 20 kHz 6.00000 6.019520 5.953500 6.046500 *6.00000 Vpp @ 100 kHz 6.00000 6.092640 5.667000 6.333000 *
66V Range
60.0000 Vpp @ 10 Hz 60.0000 60.02220 59.52000 60.48000 *
60.0000 Vpp @ 45 Hz 60.0000 60.00220 59.68500 60.31500 *60.0000 Vpp @ 1 kHz 60.0000 60.00940 59.68500 60.31500 *60.0000 Vpp @ 20 kHz 60.0000 60.23540 59.53500 60.46500 *60.0000 Vpp @ 100 kHz 60.0000 61.29180 56.67000 63.33000 *AC VOLTAGE ACCURACY TEST, TRIANGLEWAVE (NORMAL)
93mV Range90.000 mVpp @ 10 Hz 90.000 90.0022 89.3100 90.6900 *90.000 mVpp @ 45 Hz 90.000 90.0040 89.5425 90.4575 *90.000 mVpp @ 1 kHz 90.000 90.0064 89.5425 90.4575 *90.000 mVpp @ 20 kHz 90.000 90.0123 89.3175 90.6825 *90.000 mVpp @ 100 kHz 90.000 90.0691 85.0350 94.9650 *930mV Range900.000 mVpp @ 10 Hz 900.000 899.9287 893.1000 906.9000 *900.000 mVpp @ 45 Hz 900.000 899.9945 895.4250 904.5750 *900.000 mVpp @ 1 kHz 900.000 900.0257 895.4250 904.5750 *900.000 mVpp @ 20 kHz 900.000 899.9599 893.1750 906.8250 *900.000 mVpp @ 100 kHz 900.000 899.9148 850.3500 949.6500 *9.3V Range9.00000 Vpp @ 10 Hz 9.00000 8.999079 8.931000 9.069000 *9.00000 Vpp @ 45 Hz 9.00000 8.999806 8.954250 9.045750 *9.00000 Vpp @ 1 kHz 9.00000 9.000187 8.954250 9.045750 *9.00000 Vpp @ 20 kHz 9.00000 8.999737 8.931750 9.068250 *9.00000 Vpp @ 100 kHz 9.00000 9.00167T 8.503500 9.496500 *93V Range90.0000 Vpp @ 10 Hz 90.0000 89.98940 89.31000 90.69000 *90.0000 Vpp @ 45 Hz 90.0000 89.99876 89.54250 . 90.45750 *90.0000 Vpp @ 1 kHz 90.0000 89.99772 89.54250 90.45750 *90.0000 Vpp @ 20 kHz 90.0000 90.02162 89.31750 90.68250 *90.0000 Vpp @ 100 kHz 90.0000 90.01538 85.03500 94.96500 *
Fluke Corporation Telephone Facsimile Internet Page 8 of 15
1420 75th Street SW, Everett WA 98203 USA 888.993.5853 425.446.6390 www.fluke.com Rev 7.4, 02/23/11
FLUKE@ Certifica te Number: Calibration Date:945162-6265013:1301616947 1-Apr-ll
Calibration ResultsManufacturer's
Nominal Measured Measurement SpecificationsFunction/Range Value Value Uncertainty Lower Limit Upper Limit
AC VOLTAGE ACCURACYTEST, SINEWAVE(AUX)
330mVRange10.000 mV@ 10 Hz 10.000 10.0022 5.1e-006 V 9.6100 10.3900
10.000 mV@ 20 Hz 10.000 10.0037 5.2e-006 V 9.6200 10.3800
10.000 mV@ 45 Hz 10.000 10.0047 5.3e-006 V 9.6200 10.3800
10.000 mV @ 1 kHz 10.000 10.0075 5.3e-006 V 9.6200 10.3800
10.000 mV@ 5 kHz 10.000 10.0203 5.3e-006 V 9.5300 10.4700
10.000 mV @ 10 kHz 10.000 10.0362 5.3e-006 V 9.5100 10.4900
300.0 mV @ 9.5 Hz 300.0 299.71 3.0e-004 V 283.35 316.65
300.00 mV@ 10Hz 300.00 299.945 5.1e-005 V 299.030 300.970
300.00 mV@ 20 Hz 300.00 299.990 2.1e-005 V 299.330 300.670
300.00 mV@ 45 Hz 300.00 299.995 1.2e-005 V 299.330 300.670
300.00 mV@ 1 kHz 300.00 300.036 1.2e-005 V 299.330 300.670
300.00 mV@ 5 kHz 300.00 299.973 1.2e-005 V 298.950 301.050
300.00 mV @ 10 kHz 300.00 299.927 1.4e-005 V 298.350 301.650
3.3V Range3.000 V @ 9.5 Hz 3.000 3.0042 3.0e-003 V 2.8335 3.1665
3.0000 V @ 10 Hz 3.0000 2.99947 5.1e-004 V 2.99355 3.00645
3.0000 V @ 20 Hz 3.0000 2.99994 2.1e-004 V 2.99655 3.00345
3.0000 V @ 45 Hz 3.0000 3.00000 1.2e-004 V 2.99685 3.00315
3.0000 V @ 1 kHz 3.0000 3.00023 1.2e-004 V 2.99685 3.00315
3.0000 V @ 5 kHz 3.0000 2.99994 1.2e-004 V 2.99260 3.00740
3.0000 V @ 10 kHz 3.0000 2.99991 1.2e-004 V 2.98660 3.01340
AC VOLTAGEACCURACYTEST, SQUAREWAVE(AUX)
660mV Range600.000 mVpp@ 10 Hz 600.000 600.0640 595.2000 604.8000 *
600.000 mVpp @ 45 Hz 600.000 600.0960 596.8500 603.1500 *
600.000mVpp@ 1 kHz 600.000 602.2120 596.8500 603.1500 *
600.000 mVpp @ 10 kHz 600.000 621.4640 566.7000 633.3000 *
6.6V Range6.00000 Vpp @ 10Hz 6.00000 6.000600 5.952000 6.048000 *
6.00000 Vpp @ 45 Hz 6.00000 6.001080 5.968500 6.031500 *
6.00000 Vpp @ 1 kHz 6.00000 6.024500 5.968500 6.031500*6.00000Vpp@ 10kHz 6.00000 6.236860 5.667000 6.333000*AC VOLTAGEACCURACYTEST,TRIANGLEWAVE (AUX)930mV Range
900.000 mVpp@ 10 Hz 900.000 899.8005 893.1000 906.9000 *
900.000 mVpp@ 45 Hz 900.000 899.9980 895.4250 904.5750 *
900.000mVpp@ 1 kHz 900.000 899.9564 895.4250 904.5750 *
900.000 mVpp@ 10 kHz 900.000 890.5375 850.3500 949.6500*9.3VRange9.00000Vpp@ 10Hz 9.00000 8.997797 8.931000 9.069000*
Fluke Corporation Telephone Facsirnile Internet Page 9 of 15-1420 75th Street SW, Everett WA 98203 USA 888.993.5853 425.446.6390 www.fIuke.com Rev 7.4,02/23/11
FLUKE@ Certifica te Number: Calibration Date:945162-6265013:1301616947 l-Apr-ll
Calibration ResultsManufacturer's
Nominal Measured Measurement Specifications
Function/Range Value Value Uncertainty Lower Limit Upper Limit
9.00000 Vpp @ 45 Hz 9.00000 8.999945 8.954250 9.045750 *
9.00000 Vpp @ 1 kHz 9.00000 8.998629 8.954250 9.045750 *
9.00000 Vpp @ 10kHz 9.00000 8.902430 8.503500 9.496500 *
HARMONIC AMPLlTUDE ACCURACY TEST (NORMAL)
33mV Range
30.000 mV @ 1 kHz 30.000 30.0003 29.9150 30.0850 *
30.000 mV @ 5 kHz 30.000 30.0004 29.9150 30.0850 *
30.000 mV @ 10 kHz 30.000 30.0004 29.9150 30.0850 *
330mV Range
300.000 mV @ 1 kHz 300.000 300.0030 299.8101 300.1899 *
300.000 mV @ 5 kHz 300.000 300.0000 299.8101 300.1899 *
300.000 mV @ 10 kHz 300.000 299.9920 299.8101 300.1899 *
3.3V Range
3.00000 V @ 1 kHz 3.00000 3.000010 2.998980 3.001020 *
3.00000 V @ 5 kHz 3.00000 2.999980 2.998980 3.001020 *
3.00000 V @ 10kHz 3.00000 2.999940 2.998980 3.001020 *
33V Range
30.0000 V @ 1 kHz 30.0000 29.99860 29.98680 30.01320 *
30.0000 V @ 5 kHz 30.0000 29.99850 29.98680 30.01320 *
30.0000 V @ 10 kHz 30.0000 29.99830 29.98680 30.01320 *
330V Range
300.000 V @ 1 kHz 300.000 299.9860 299.8368 300.1632 *300.000 V @ 5 kHz 300.000 299.9770 299.7300 300.2700 *300.000 V @ 10 kHz 300.000 299.9770 299.7300 300.2700 *
1000V Range
1000.00 V @ 1 kHz 1000.00 1000.022 999.340 1000.660 *
1000.00 V @ 5 kHz 1000.00 999.982 997.800 1002.200 *1000.00 V @ 10 kHz 1000.00 999.958 997.000 1003.000 *HARMONIC AMPLlTUDE ACCURACY TEST (AUX)
330mV Range300.000 mV @ 1 kHz 300.000 300.0460 298.9600 301.0400 *300.000 mV @ 5 kHz 300.000 299.9830 298.5000 301.5000 *300.000 mV @ 10kHz 300.000 299.9340 297.9000 302.1000 *3.3V Range
3.00000 V @ 1 kHz 3.00000 3.000280 2.996400 3.003600 *3.00000 V @ 5 kHz 3.00000 3.000020 2.991200 3.008800 *3.00000 V @ 10 kHz 3.00000 2.999990 2.985200 3.014800 *AC CURRENT ACCURACY TEST
330pA Range300.00 IJA@ 10 Hz 300.00 299.981 3.0e-008 A 299.100 300.900
300.00 IJA@ 20 Hz 300.00 299.993 3.0e-008 A 299.475 300.525
300.00 IJA@ 45 Hz 300.00 299.995 2.3e-008 A 299.475 300.525
FlukeCorporation Telephone Facsimile Internet Page 10 of 15
1420 75th Street SW, Everett WA 98203 USA 888.993.5853 425.446.6390 www.fluke.com Rev 7.4, 02/23/H
FL.UKE@ Certificate Number: Calibration Date:
945162-6265013:1301616947 l-Apr-ll
Calibration ResultsManufacturer's
Nominal Measured Measurement Specifications
Function/Range Value Value Uncertainty Lower Limit Upper Limit
300.00 IJA@ 1 kHz 300.00 300.032 2.3e-008 A 299.375 300.625
300.00 IJA@ 5 kHz 300.00 300.013 6.1 e-008 A 298.650 301.350
300.00 IJA@ 10 kHz 300.00 300.092 7.7e-008 A 296.100 303.900
3.3mA Range
3.0000 mA @ 10 Hz 3.0000 2.99984 3.0e-007 A 2.99370 3.00630
3.0000 mA @ 20 Hz 3.0000 2.99999 3.0e-007 A 2.99670 3.00330
3.0000 mA @ 45 Hz 3.0000 2.99999 2.0e-007 A 2.99670 3.00330
3.0000 mA @ 1 kHz 3.0000 3.00043 2.0e-007 A 2.99670 3.00330
3.0000 mA @ 5 kHz 3.0000 3.00001 2.6e-007 A 2.99370 3.00630
3.0000 mA @ 10kHz 3.0000 3.00009 3.0e-007 A 2.98170 3.01830
33mA Range30.000 mA @ 10 Hz 30.000 30.0001 4.5e-006 A 29.9370 30.0630
30.000 mA @ 20 Hz 30.000 30.0009 4.5e-006 A 29.9670 30.0330
30.000 mA @ 45 Hz 30.000 30.0013 2.ge-006 A 29.9700 30.0300
30.000 mA@ 1 kHz 30.000 30.0039 3.0e-006 A 29.9700 30.0300
30.000 mA @ 5 kHz 30.000 30.0037 , 3.0e-006 A 29.9370 30.0630
30.000 mA @ 10 kHz 30.000 30.0014 3.0e-006 A 29.8170 30.1830
330mA Range
300.00 mA @ 10 Hz 300.00 299.875 3.ge-005 A 299.370 300.630
300.00 mA @ 20 Hz 300.00 299.893 4.0e-005 A 299.670 300.330
300.00 mA @ 45 Hz 300.00 299.895 2.0e-005 A 299.700 300.300
300.00 mA @ 1 kHz 300.00 299.927 2.0e-005 A 299.700 300.300
300.00 mA @ 5 kHz 300.00 299.920 2.0e-005 A 299.370 300.630
300.00 mA @ 10 kHz 300.00 299.919 2.0e-005 A 298.170 301.830
2.2A Range
2.0000 A @ 10 Hz 2.0000 1.99994 2.6e-004 A 1.99570 2.00430
2.0000 A @ 45 Hz 2.0000 2.00006 1.6e-004 A 1.99770 2.00230
2.0000 A @ 1 kHz 2.0000 2.00005 1.6e-004 A 1.99770 2.00230
2.0000 A @ 5 kHz 2.0000 1.99989 1.6e-004 A 1.98470 2.01530
11A Range
10.000 A @ 45 Hz 10.000 10.0003 1.6e-003 A 9.9920 10.0080
10.000 A @ 65 Hz 10.000 10.0002 1.6e-003 A 9.9920 10.0080
10.000 A@ 500 Hz 10.000 10.0001 1.1e-003 A 9.9880 10.0120
10.000 A @ 1 kHz 10.000 9.9997 1.0e-003 A 9.9650 10.0350
AC POWER ACCURACY TEST (High Voltage)
3mA, 1000V (O° Phase)
3.0000 mA @ 65 Hz 3.0000 2.99997 2.0e-007 A 2.99670 3.00330
3mA, 1000V (90° Phase)
3.0000 mA @ 65 Hz 3.0000 3.00039 2.0e-007 A 2.99670 3.00330
30mA, 1000V (O° Phase)
30.000 mA @ 500 Hz: 30.000 29.9994 3.0e-006 A 29.9700 30.0300
30mA, 1000V (90° Phase)
FlukeCorporation Telephone Facsimile Internet Page 11 of 15
1420 75th Street SW, Everett WA 98203 USA 888.993.5853 425.446.6390 www.fluke.com Rev 7.4,02/23/11
FL.UKE@ Certificate Number: Calibration Date:945162-6265013:1301616947 1-Apr-ll
Calibration ResultsManufacturer's
Nominal Measured Measurement SpecificationsFunction/Range Value Value Uncertainty Lower Limit Upper Limit
30.000 mA @ 500 Hz 30.000 30.0017 3.0e-006 A 29.9700 30.0300
30mA, 1000V (00 Phase)
30mA, 1000V (00 Phase)
30.000 mA @ 1 kHz 30.000 30.0041 3.0e-006 A 29.9700 30.0300
30.000 mA @ 5 kHz 30.000 30.0016 3.0e-006 A 29.9370 30.0630
30.000 mA @ 10 kHz 30.000 30.0034 3.0e-006 A 29.8170 30.1830
AC POWER ACCURACY TEST (High Current)
33mV, 10A (00 Phase)
33.000 mV @ 65 Hz 33.000 33.0033 3.0e-006 V 32.9635 33.0365
33mV, 10A (900 Phase)
33.000 mV @ 65 Hz 33.000 33.0225 3.0e-006 V 32.9635 33.0365
33mV, 10A (00 Phase)33.000 mV @ 500 Hz 33.000 33.0117 3.0e-006 V 32.9635 33.0365
330mV, 10A (00 Phase)
0.33000 V @ 1 kHz 0.33000 0.329984 1.0e-005 V 0.329841 0.330159
3.3V, 2A (00 Phase)
3.30000 V @ 5 kHz 3.30000 3.299714 6.7e-005 V 3.298079 3.301921
3;3V, 300mA (00 Phase)
3.30000 V @ 10kHz 3.30000 3.299704 6.8e-005 V 3.298079 3.301921
AC POWER ACCURACYTEST (High Power)300V, 2A (00 Phase)
300.000 V @ 5 kHz 300.000 299.9736 9.7e-003 V 299.7450 300.2550
300V, 2A (900 Phase)300.000 V @ 5 kHz 300.000 299.9802 9.7e-003 V 299.7450 300.2550
1000V, 10A (00 Phase)
1000.00 V @ 1 kHz 1000.00 1000.012 3.2e-002 V 999.420 1000.580
1000V, 10A (900 Phase)
1000.00 V @ 1 kHz 1000.00 1000.028 3.2e-002 V 999.420 1000.580
1 Year Specification Limits shown
RESISTANCE 2W COMP TEST
2W Comp Test PASS*
RESISTANCE ACCURACY TEST
11 Ohm RangeO.OOOOOOhm 0.00000 0.000093 4.1e-0050hm -0.006000 0.006000
1.00000 Ohm 1.00000 1.000168 3.1e-005 Ohm 0.993880 1.006120
1.90000 Ohm 1.90000 1.900203 5.6e-005 Ohm 1.893772 1.906228
10.00000hm 10.0000 10.00057 1.2e-004 Ohm 9.99280 10.00720
33 Ohm Range11.90000 Ohm 11.90000 11.901017 1.4e-004 Ohm 11.888572 11.911428
19.00000 Ohm 19.00000 19.001655 2.2e-004 Ohm 18.987720 19.012280
Fluke Corporation Telephone Facsimile Internet Page 12 of 15-1420 75th Street SW, Everett WA 98203 USA 888.993.5853 425.446.6390 www.fluke.com Rev 7.4,02/23/11
FLUKE@ Certifica te Number: Calibration Date:945162-6265013:1301616947 l-Apr-ll
Calibration ResultsManufacturer's
Nominal Measured Measurement Specifications
Function/Range Value Value Uncertainty Lower Limit Upper Limit
30.0000 Ohm 30.0000 30.00237 3.5e-004 Ohm 29.98640 30.01360
110 Ohm Range33.0000 Ohm 33.0000 33.00197 3.8e-0040hm 32.98703 33.01297
100.00000hm 100.0000 100.00606 5.ge-0040hm 99.98100 100.01900
330 Ohm Range119.00000hm 119.0000 119.00427 7.0e-004 Ohm 118.97929 119.02071
190.0000hm 190.000 190.0075 1.1e-003 Ohm 189.9729 190.0271
300.000 Ohm 300.000 300.0098 2.1e-003 Ohm 299.9630 300.0370
1.1 kOhm Range0.330000 kOhm 0.330000 0.3300059 2.4e-003 Ohm 0.3299103 0.3300897
1.000000 kOhm 1.000000 1.0000191 5.8e-003 Ohm 0.9998500 1.0001500
3.3 kOhm Range1.190000 kOhm 1.190000 1.1900204 1.0e-002 Ohm 1.1898329 1.1901671
1.90000 kOhm 1.90000 1.900061 1.6e-002 Ohm 1.899769 1.900231
3.00000 kOhm 3.00000 3.000141 3.4e-002 Ohm 2.999670 3.000330
11 kOhm Range3.30000 kOhm 3.30000 3.300092 3.5e-002 Ohm 3.299103 3.300897
10.00000 kOhm 10.00000 10.000246 9.6e-002 Ohm 9.998500 10.001500
33 kOhm Range11.9000 kOhm 11.9000 11.90035 1.2e-001 Ohm 11.89833 11.90167
19.0000 kOhm 19.0000 19.00065 2.2e-001 Ohm 18.99769 19.00231
30.0000 kOhm 30.0000 30.00083 3.3e-001 Ohm 29.99670 30.00330
110 kOhm Range33.0000 kOhm 33.0000 33.00027 3.5e-001 Ohm 32.99037 33.00963
100.0000 kOhm 100.0000 100.00085 9.5e-001 Ohm 99.98300 100.01700
330 kOhm Range119.0000 kOhm 119.0000 119.00090 1.0e+000 Ohm 118.97972 119.02028
190.0000 kOhm 190.0000 190.00223 1.6e+000 Ohm 189.97120 190.02880
300.000 kOhm 300.000 300.0019 2.5e+000 Ohm 299.9580 300.0420
1.1 MOhm Range0.330000 MOhm 0.330000 0.3300014 2.8e+000 Ohm 0.3298955 0.3301045
1.000000 MOhm 1.000000 1.0000034 9.3e+000 Ohm 0.9997950 1.0002050
3.3 MOhm Range1.19000 MOhm 1.19000 1.190039 1.5e+001 Ohm 1.189767 1.190233
1.90000 MOhm 1.90000 1.900040 2.4e+001 Ohm 1.899660 1.900340
3.00000 MOhm 3.00000 3.000034 3.5e+001 Ohm 2.999495 3.000505
11 MOhm Range3.30000 MOhm 3.30000 3.300068 9.2e+001 Ohm 3.297470 3.302530
10.00000 MOhm 10.00000 10.000031 6.3e+002 Ohm 9.993450 10.006550
33 MOhm Range11.90000 MOhm 11.90000 11.899956 7.3e+0020hm 11.887550 11.912450
19.00000 MOhm 19.00000 18.999556 1.5e+003 Ohm 18.980450 19.019550
Fluke Corporation Telephone Facsimile Internet Page 13 of 15
1420 75th Street SW, EverettWA 98203 USA 888.993.5853 425.446.6390 www.fluke.com Rev 7.4, 02123/11
FLUKE@ Certifica te Number: Calibration Date:945162-6265013:1301616947 l-Apr-ll
Calibration ResultsManufacturer's
Nominal Measured Measurement Specifications
Function/Range Value Value Uncertainty Lower Limit Upper Limit
30.00000 MOhm 30.00000 29.999078 2.2e+003 Ohm 29.969450 30.030550
110MOhm Range33.00000 MOhm 33.00000 32.998928 2.8e+003 Ohm 32.829500 33.170500
100.00000 MOhm 100.00000 100.021295 2.7e+004 Ohm 99.494500 100.505500
330MOhm Range119.00000 MOhm 119.00000 118.965854 8.7e+0040hm 118.388500 119.611500
300.00000 MOhm 300.00000 300.047426 1.0e+005 Ohm 298.483500 301.516500
CAPACITANCE ACCURACY TEST
500pF Range
350 pF @ 1 kHz 350 350.2 2.ge-012 F 338.3 361.7
480 pF @ 1 kHz 480 480.6 2.2e-012 F 467.6 492.4
1.1nF Range
0.600 nF @ 1 kHz 0.600 0.6005 2.0e-012 F 0.5870 0.6130
1.000 nF @ 1 kHz 1.000 1.0008 1.8e-012 F 0.9850 1.0150
3.3nF Range1.200 nF @ 1 kHz 1.200 1.2008 1.7e-012 F 1.1840 1.2160
3.000 nF @ 1 kHz 3.000 3.0009 3.3e-012 F 2.9750 3.0250
11nF Range
3.300 nF @ 1 kHz 3.300 3.3007 4.3e-012 F 3.2735 3.3265
10.900 nF @ 1 kHz 10.900 10.9013 6.6e-012 F 10.8355 10.9645
33nF Range
12.000 nF @ 1 kHz 12.000 12.0070 7.2e-012 F 11.8700 12.1300
30.00 nF @ 1 kHz 30.00 30.007 1.8e-011 F 29.825 30.175
110nF Range
33.00 nF @ 1 kHz 33.00 33.007 2.7e-011 F 32.818 33.182
109.00 nF @ 1 kHz 109.00 109.020 6.6e-011 F 108.627 109.373
330nF Range
120.00 nF @ 1 kHz 120.00 120.017 7.3e-011 F 119.400 120.600
300.0 nF @ 1 kHz 300.0 300.03 1.8e-010 F 298.95 301.05
1.1uF Range
0.3300 F @ 100 Hz 0.3300 0.33003 2.7e-010 F 0.32818 0.33182
1.0900 F @ 100 Hz 1.0900 1.09007 6.6e-010 F 1.08627 1.09373
3.3uF Range1.2000 F @ 100 Hz 1.2000 1.20037 7.3e-010 F 1.19280 1.20720
3.000 F @ 100 Hz 3.000 3.0009 1.ge-009 F 2.9865 3.0135
11uF Range
3.300 F @ 100 Hz 3.300 3.3009 2.7e-009 F 3.2785 3.3215
10.9000 F @ 100 Hz 10.9000 10.90233 8.ge-009 F 10.85185 10.94815
33uF Range12.000 F @ 100 Hz 12.000 12.0033 9.7e-009 F 11.9220 12.0780
30.00 F @ 100 Hz 30.00 30.005 2.4e-008 F 29.850 30.150
110uF Range
Fluke Corporation Telephone Facsimile Internet Page 14 of 15
1420 75th Street SW, Everett WA 98203 USA 888.993.5853 425.446.6390 www.fluke.com Rev 7.4, 02/23/11
FLUKE@ Certifica te Number:945162-6265013:1301616947
Calibration Date:
l-Apr-ll
Calibration Results
Function/RangeNominal
ValueMeasured
Value
33.00 IJF@ 100 Hz
109.00 IJF@ 100 Hz
330uF Range120.0 IJF@ 100 Hz
300.0 IJF@ 50 Hz
1100uF Range
330.0 IJF@ 50 Hz 330.0 330.05
800.0 IJF@ 50 Hz 800.0 800.04
THERMOCOUPLE ("J" Type) MEASUREMENT ACCURACY TEST25.030 °C 25.030 24.96
THERMOCOUPLE SOURCING ACCURACY TEST (10uVlOC)0.00000 mV 0.00000 -0.000053
1.0000 mV 1.0000 0.99992
-1.0000 mV -1.0000 -1.00004
10.0000 mV 10.0000 9.99984
-10.0000 mV -10.0000 -9.99992
100.0000 mV 100.0000 99.99925
-100.0000 mV -100.0000 -99.99827
THERMOCOUPLE MEASUREMENT ACCURACY TEST (10uV/°C)0.000 mV 0.000 0.0005
100.000 mV 100.000 100.0026
-100.000 mV -100.000 -100.0003
33.00
109.00
33.001
109.007
120.0
300.0
120.13
300.00
MeasurementUncertainty
2.7e-008F1.2e-007F
1.3e-007 F
3.3e-007 F
4.3e-007 F
2.6e-006 F
4.0e-002 °C
1.Oe-007V
1.6e-006 V
1.6e-006 V
3.1e-007 V
3.0e-007 V
8.0e-007 V
8.1e-007 V
2.5e-007 V
6.8e-007 V
6.8e-007 V
Manufacturer'sSpecifications
Lower Limit Upper Limit
End of Report
Fluke Corporation Telephone
1420 75th Street SW, Everett WA 98203 USA 888.993.5853
Facsimile
425.446.6390
Internet Page 15 of 15
Rev 7.4,02/23/11www.fluke.com
32.735 33.265
108.355 109.645
118.86 121.14
297.60 302.40
326.40 333.60
791.70 808.30
24.89 25.17
-0.003000 0.003000
0.99694 1.00306
-1.00306 -0.99694
9.99640 10.00360
-10.00360 -9.99640
99.99100 100.00900
-100.00900 -99.99100
-0.0030 0.0030
99.9910 100.0090
-100.0090 -99.9910