aerosol particle counter models 310b, 310c, 350l, 5100€¦ · comprehensive validation manual...
Post on 09-Jun-2018
226 Views
Preview:
TRANSCRIPT
BENEFITS
•
•
•
•
•
•
•
•
•
•
APPLICATIONS
•
•
•
•
•
•
•
ISO 21501 compliant with built-in Pulse Height Analyzer (PHA)
– 100 LPM flow samples 1m³ in 10 minutes
Ultra quiet pump improves work environment
Calculates and creates reports for ISO 14644-1 and EC GMP (PIC/S GMP) Annex #1
Comprehensive validation manual makes it easy to meet pharmaceutical regulatory requirements
Operates in cleanrooms via a web browser: Setup, sample, display, print, download data
Choose from nine languages for display and printout
Removable batteries with optional external charger for continuous mobile use
Stores sampling recipes to reduce operator errors
Cleanroom monitoring
Facility certification for ISO or EC GMP
Trend analysis
Statistical process control
Troubleshooting particle excursions
Manifold compatible
Portable or dedicated use
100 LPM, 50 LPM or 1 CFM flow rates available
Large IR touch screen for simplified use
Without measurement, there is no control.
The Lasair III particle counter sets the new standard for aerosol particle counters. The on-board Pulse Height Analyzer (PHA) ensures ISO 21501 compliance. The large IR touch screen controls and intuitive user interface make the system very easy to use. The particle counter’s rugged enclosure is chemically resistant and lightweight for portability. The enclosure also has a streamlined design that minimizes particle traps and makes the unit easy to clean.
Aerosol Particle Counter Models 310B, 310C, 350L, 5100
Exclusive Distributor and Service Centers:
www.enviroterm.com
Aerosol Particle Counter
Specificationq/Modelu 310BChannels
Flow rate 1.0 CFM (28.3 LPM) ± 5%
aMaximum Concentration > 1,380,270/ft.³
Sample Tubing ID
Zero count
U.S. and foreign patents pending.
www.pmeasuring.com
0.3, 0.5, 1.0, 3.0, 5.0, 10µm
310C
1.0 CFM (28.3 LPM) ± 5%
> 1,380,270/ft.³
ID: 3/8"; Max. Length: 8 m
Data storage 3000 complete data sets. Data cannot be edited or altered meeting 21 CFR Part 11 compliance.
Communication Modes Ethernet or RS-232; USB data downloading; optional wireless Ethernet
Controlling Software DataAnalyst®, Facility Net, Pharmaceutical Net, FacilityPro, Microsoft® Internet Explorer 5.0+, Firefox®
Remote web browser operation; USB download, downloadable to Facilty Net/PharmaNet.Remote Operation
Environmental Sensors Four 4-20 mA inputs
Languages English, French, German, Italian, Japanese, Korean, Chinese (Traditional or Simplified), Spanish, Russian
8.4” color VGA display (640 x 480); IR touch screen (IP65 rated); Display Printer Built-in thermal printer
External Surface Polycarbonate (PC)
Enclosure Cleaning Materials Bleach, formaldehyde, ethyl/isopropyl alcohol, peroxide/quaternary ammonium solutions
Sample Output Internally filtered to HEPA standards (> 99.97% at 0.3 µm) filtering
Power and Battery 85-264 V, 50-60 Hz; Lithium battery: operation 3 hrs. (single) 6 hrs. (dual) for 1 CFM unit
12 x 13 x 10 inch (30x 33 x 26 cm) Dimensions (h, w, d)Weight 13 lbs. (5.9 kg) without battery, 16 lbs. (7.3 kg) with two optional batteries
Operating Environment Temperature: 0-30°C; Humidity:5-95% RH non-condensing
Lasair® is a registered trademark of Particle Measuring Systems, Inc.Microsoft® is a registered trademark of Microsoft Corporation.
Particle Measuring Systems, Inc. reserves the right to change specifications without notice.a. Less than 10% coincidence loss at maximum recommended concentration
Compliance to ISO-21501-4 requirementsCalibration
Reports USB reports; sample printouts; cleanroom certification reports for ISO and EC GMP (PIC/S GMP); averaging
ID: 3/8"; Max. Length: 8 m
5100
100 LPM ± 5%
> 688,495/ft.³
2 counts/m (100 LPM)³
ID: ¾"; Max. Length: 8 m
0.3, 0.5, 1.0, 5.0,10.0, 25µm
7.07 counts/m (1 CFM) ³4 counts/m³
7.07 counts/m (1 CFM) ³4 counts/m³
350L
50 LPM ± 5%
> 832,599/ft.³
ID: ½"; Max. Length: 8 m
2 counts/m (50 LPM)³
Counting Efficiency50% ± 20% for most sensitive threshold (100%±10% at 1.5 times the most sensitive sensitivity), fully complies with ISO 21501-4
Exclusive Distributor and Service Centers:
ENVIROTERM SDN. BHD. 369410-P
Particle Measuring Systems Headquarters
5475 Airport Blvd., Boulder, CO 80301
1-303-443-7100 1-800-238-1801 Fax: 1-303-546-7331
URL www.enviroterm.com Email : sales@enviroterm.com
37-1, Jalan Setia Tropika 1/25,Taman Setia Tropika, 81200 Johor Baru,Johor, Malaysia.Tel : 07-2362768 Fax: 07-2362769
SOUTHERN OFFICE
0.3, 0.5, 1.0, 5.0,10.0, 25µm 0.3, 0.5, 1.0, 5.0,10.0, 25µm
H igh Pressure
• Monitor pressurized gases with air particle counter• Pressura range: 25 - 100 psi (60 - 100 psi for 100 LPM) • Compatible with CDA, Argon, N , or other non-toxic inert gases. 2
• Flow rate selection for 0.1, 1.0 CFM, 50, & 100 LPM
APPLICATION
Temperature/Relative Humidity handheld probe TRH-9008TR
Temperature/Relative Humidity Sensor TRH-100AND
Temperature Range:-40 to+80°C, ±0.3°C RH Range: 0-100%,±2% RH Cable Length: 2 meters, with handheld.
Temperature Range: 0 to+50°C, ±0.5°C RH Range: 0-100%,±3% RH Fixed Module on the rear panel of Lasair III
Differential Air Pressure Sensor DAP-100-AO Pressure Range:-50 to+50 Pascal, ±1.5% Tubing Length: 2 meters x 2 pcs
Filter Scanning Probe, For 1.0 CFM Flow rate FSP-ET-28
Many more options available to cater for your specific applications, includes, trace moisture, VOC, and etc. please contact one of our sales to discuss further on your application.
Instrument Service & Support: 1-800-557-6363
Customer Response Center: 1-877-475-3317
© 2012 Enviroterm Sdn. Bhd. All rights reserved.
OPTION: CAM
20, Lorong PS1, Bandar Perda,14000 Bukit Mertajam,Pulau Pinang, Malaysia.Tel : 04-5371371 Fax: 04-537 1373
NORTHERN OFFICELot 5035, Jalan 18/62, Taman Perindustrian Sri Serdang, 43300 Seri Kembangan, Selangor D.E., Malaysia.Tel : 03-89434558 Fax: 03-89436012
HEAD OFFICE
D iffuser
, Model : CAM
This CAM can be used in applications where users want to measurethe noted parameters at pressure (25 to 100 psi range)
Compressed Air ModuleAPPLICATION
Wireless ModuleThis
wireless data communication from LASAIR III to PCAPPLICATION wireless unit allow
OPTION: HPD-III
OPTION: WM
Dew Point: -60ºC to 40ºC,Temperature: -20ºC to 75ºCPressure: 0 to 200 psia
OPTION: ENVIRO-SENSORS
top related