Advantages of Ultrasonic Thickness Gages over Flaw Detectors for Corrosion Thickness Measurements

Greg Bauer | Product Specialist, Ultrasonic Thickness Gages January 11, 2017

Advantages of Flaw Detectors § More versatile

–  Straight beam, angle beam, scanning

§  Includes specialized software for sizing defects: –  Distance Amplitude Correction (DAC) –  Time Varied Gain (TVG) –  Distance/Gain/Size (DGS)

OLYMPUS Scientific Solutions

Instrument Priority §  Flaw detectors

–  Accuracy of signal amplitude

§  Thickness gages –  Consistency and accuracy of the time of flight measurement

§ Use the right tool for the job

OLYMPUS Scientific Solutions

Why were flaw detectors used for measuring thickness? § History of ultrasonic (UT) equipment

–  Ultrasonic instruments were released in the late 1940s –  Digital thickness gages released in the 1960s

–  No waveform = less confidence –  Thickness gage with on-board waveform released in 1980s

OLYMPUS Scientific Solutions

Additional Information from Live A-Scan

OLYMPUS Scientific Solutions

Waveform 1 Waveform 2

Detection of backwall Detection of noise

Additional Information from Live A-Scan

OLYMPUS Scientific Solutions

Waveform 1 Waveform 2

Heavily corroded backwall Cleaner backwall

Advantages of Thickness Gages over Flaw Detectors § Size and cost

§ Ease of calibration and operation

§ More consistent and accurate time of flight measurement

§ Specialized features or software for special applications

OLYMPUS Scientific Solutions

Thickness Gage Measurement Advantages §  Auto zero compensation

§  V-path correction

§  Auto probe recognition

§  Detection algorithms and digital signal processing (DSP)

Auto Zero Compensation

OLYMPUS Scientific Solutions

Why are dual element transducers used for corrosion applications?

OLYMPUS Scientific Solutions

V-Path Correction When Using Thickness Gages

OLYMPUS Scientific Solutions

Sample

TX RX

Angular Sound Path

True Thickness

Thickness Measurement Error When Using a Flaw Detector

OLYMPUS Scientific Solutions

Auto Probe Recognition § Optimizes setup and receiver gain

–  Easier and faster setup for the user

§ Recalls transducer V-path correction –  Accurate measurements over large thickness ranges

OLYMPUS Scientific Solutions

V-Path Builder on Olympus 38DL PLUS® Thickness Gage

OLYMPUS Scientific Solutions

Step 3 Step 4

Step 1 Step 2

Flaw Detector Detection Mode — Flanking/Edge Gate

OLYMPUS Scientific Solutions

Detection point at 60 dB Detection point at 54 dB

§  Thickness reading changes 0.011 in. (0.280 mm) from a 6 dB change

Flaw Detector Detection Mode — Peak Gate

OLYMPUS Scientific Solutions

Signal from the pit Majority of signal not from thinnest area

Corrosion Thickness Gages Use Algorithms and Digital Signal Processing (DSP)

OLYMPUS Scientific Solutions

Detection Threshold

Thickness Gage Application Advantages §  Temperature compensation

§  Electromagnetic acoustic transducer (EMAT)

§  Internal oxide

§  Encoded B-scan

§  Echo-to-echo measurement

§  THRU-COAT® technology

High-Temperature Applications

OLYMPUS Scientific Solutions

§ Setup adjustments –  Increase gain –  Highest update rate –  Auto zero –  Temperature compensation

Olympus 38DL PLUS Thickness Gage Temperature Compensation § Sound velocity changes with temperature

–  Approximately 1% for every 100 °F (55 °C) change in temperature for carbon steel

–  V0: Sound velocity at calibration –  T0: Temperature at calibration (user inputted) –  T1: Temperature at measurement (user inputted) –  k: Temperature coefficient (user inputted)

OLYMPUS Scientific Solutions

Olympus 38DL PLUS Thickness Gage Temperature Compensation

OLYMPUS Scientific Solutions

Boiler Tube Applications §  Two solutions

–  Measuring tube wall thickness using EMAT –  Measuring tube wall and internal scale thickness using 38DLP-OXIDE

OLYMPUS Scientific Solutions

Olympus 38DL PLUS Thickness Gage — Measurements with the E110-SB EMAT § Electromagnetic acoustic transducer (EMAT)

–  Requires external oxide scale to measure tube wall thickness –  No couplant required –  Fast and easy

OLYMPUS Scientific Solutions

§ Measure boiler tube wall thickness and internal scale thickness –  Minimum scale with M2017: 0.010 in. (0.250 mm) –  Minimum scale with M2091: 0.006 in. (0.150 mm)

–  Shear wave couplant (SWC-2) required

OLYMPUS Scientific Solutions

Olympus 38DL PLUS Thickness Gage — Optional 38DLP-OXIDE Software

§ Enables the 38DL PLUS thickness gage to perform encoded B-scans –  Data is transferrable –  Less expensive

OLYMPUS Scientific Solutions

Olympus 38DL PLUS Thickness Gage — Optional 38DLP-EBSCAN Software

Distance traveled

§ Echo-to-echo –  Used to ignore coatings and measure only the true metal thickness

OLYMPUS Scientific Solutions

Measuring Thickness on Painted Metal Applications

Auto echo-to-echo Manual echo-to-echo

§ Measuring thin materials –  About 0.080 to 0.100 in. (2.0 to 2.5 mm) minimum

OLYMPUS Scientific Solutions

Echo-to-Echo Limitations

0.080 in. (2.0 mm) sample 0.040 in. (1.0 mm) sample Valid echo-to-echo measurement No separation, false reading

§ Samples with a heavily corroded back surface –  No multiple backwalls

OLYMPUS Scientific Solutions

Echo-to-Echo Limitations

§  THRU-COAT technology –  Used to measure paint thickness and steel thickness at the same time

OLYMPUS Scientific Solutions

Measuring Thickness on Painted Metal Applications

§ Measure thinner samples –  Minimum of about 0.040 in. (1.0 mm)

§ Measure thicker samples § Measuring painted samples with corroded back surfaces

OLYMPUS Scientific Solutions

Advantages of THRU-COAT Technology

Echo-to-echo THRU-COAT technology

§ Used in standard, THRU-COAT, and echo-to-echo mode §  Limitations

–  Rough or textured surfaces –  Coating has to be at least 0.005 in. (0.127 mm) –  Cannot be used for high temperatures

OLYMPUS Scientific Solutions

THRU-COAT Transducers (D7906-SM, D7906-RM, D7908)

Thickness Gage Feature Advantages §  Doubling warning

§  Gain optimization

Thickness Gage Feature — Doubling Warning § Definition of doubling

–  Gage detects the second backwall reflection and misses the first backwall signal

OLYMPUS Scientific Solutions

Dual Element Transducers — Measuring Thick and Thin Samples

OLYMPUS Scientific Solutions

Thick sample Thin sample

Calibration Doubling Warning

OLYMPUS Scientific Solutions

Potentially wrong echo detected!

CAL VEL on thick CAL ZERO on thin

Doubling Issue with Dual Element Transducers § How to correct the doubling issue

–  Increase gain so the first backwall is above the 20% detection threshold

Waveform 1 Waveform 2

OLYMPUS Scientific Solutions

Gain Optimization Calibration § Advantages

–  Increase useful life of transducer –  Help prevent calibration and measurement errors

OLYMPUS Scientific Solutions

Default gain of 66 dB for this D790 Default gain of 59 dB for this D790