alarm monitor – stationary dose rate monitoring system ... · pdf filealmo alarm monitor...

4
ALMO ALarm MOnitor – stationary dose rate monitoring system with external detectors Fields of application workplace and room monitoring, e.g. in hot cells and laboratories system monitoring, e.g. in isotope production ward and/or patient monitoring in nuclear medicine/radiotherapy, e.g. in radio iodine therapy or afterloading monitoring and selection in sorting boxes for radioactive waste exhaust air monitoring monitoring of test facilities in nondestructive material testing warehouse monitoring, e.g. collection sites for radioactive waste Use and function The ALMO alarm monitors are used to control permanently the dose rate measurement level in the laboratory and in production facilities. The systems consist of a basic device with LCD display and electronics, combined with one or more probes and external alarm units. Depending on the version, it is possible to connect 1, 3 or 6 detectors to the ALMO basic device. The measurement value display of the multi-channel systems ALMO-3 and ALMO-6 is displayed individually for each channel. The stand-alone ALMO-systems can be linked to a superior PC-system for central monitoring and documentation

Upload: ngotuong

Post on 06-Mar-2018

233 views

Category:

Documents


6 download

TRANSCRIPT

Page 1: ALarm MOnitor – stationary dose rate monitoring system ... · PDF fileALMO ALarm MOnitor – stationary dose rate monitoring system with external detectors Fields of application

ALMO ALarm MOnitor – stationary dose rate monitoring system with external detectors

Fields of application

• workplaceandroommonitoring,e.g.inhotcellsandlaboratories• systemmonitoring,e.g.inisotopeproduction• wardand/orpatientmonitoringinnuclearmedicine/radiotherapy, e.g.inradioiodinetherapyorafterloading• monitoringandselectioninsortingboxesforradioactivewaste• exhaustairmonitoring• monitoringoftestfacilitiesinnondestructivematerialtesting• warehousemonitoring,e.g.collectionsitesforradioactivewaste

Use and function

TheALMOalarmmonitorsareusedtocontrolpermanentlythedoseratemeasurementlevelinthelaboratoryandinproductionfacilities.ThesystemsconsistofabasicdevicewithLCDdisplayandelectronics,combinedwithoneormoreprobesandexternalalarmunits.Dependingontheversion,itispossibletoconnect1,3or6detectorstotheALMObasicdevice.Themeasurementvaluedisplayofthemulti-channelsystemsALMO-3andALMO-6isdisplayedindividuallyforeachchannel.Thestand-aloneALMO-systemscanbelinkedtoasuperiorPC-systemforcentralmonitoringanddocumentation

Page 2: ALarm MOnitor – stationary dose rate monitoring system ... · PDF fileALMO ALarm MOnitor – stationary dose rate monitoring system with external detectors Fields of application

Performance features

• µ-controller-basedmeasurementelectronics• digitalmeasurementvalueinformationonlarge-area,illuminatedLCD• measurementvaluedisplayofdoserateinn/µ/mSv/hwithautorangingfunction• externallyconnectabledetectors(GMcountertube,NaI-detectors)withintegratedhightension andelectronics• automaticdetectoridentification,calibrationdataarereadoutbythemeasurementelectronics, allowingsimplereplacementofthedetector• 2freelydefinablealarmthresholdsperprobe• easy-to-operatemeasurementsystemwithuserguidance• ergonomicallyshapedhousing,desktoporwallversion• cableexitofthewallversionupward(totheceiling)ordownward(tothefloor)• variousvisual/acousticalarmunitscanbeconnected• serialdatainterfaceformeasurementdatatransferandstorageonexternalPC-system• softwareforcontinuousdoseratemeasurement,includingdatastorage• alsoavailablewithintegratedbatterybackup• maximalcablelength: o GMdetector:100m o NaI-scintillationdetector:20m

Versions and detectors

TheALMO-systemisavailableinthreeversions:• ALMO-1one-channelsystem,1detectorconnectable• ALMO-3multi-channelsystem,upto3detectorsconnectable• ALMO-6multi-channelsystem,upto6detectorsconnectable

Special versions

AfterloadingTheALMO-1systemisusedbydifferentmanufacturersofafterloadingsystems.Inthiscase,themaintaskofthealarmmonitoristoindicateifthetherapysourceisopenedorclosed.ForthatpurposetheALMO-1ismostlycombinedwithGM-probetype18550.Generallytheversionwithintegratedbatterybackupisusedinordertokeepthesystemrunningincaseofapowerbreakdown.AnexternalLED-lampwith3levels(red/yellow/green)andahornelementshowstheoperatingstatusofthesystemvisuallyandacoustically:green=closedsource,yellow=error,red=opensource,red+acousticalarm=opensourceandopendoor.AdditionallytotheLED-lamp,somemanufacturersuseasecondLED-lamporamatrixtableauwithalarmsignalfortheareaoutsidethetherapyroom(corridororcontrolroom).TheseunitsarecontrolledinasimilarwayviatheALMO-system.Anelectronicinterlock- ifexistent-canalsobeoperatedviatheALMO-system.Suchafunctionisoftenusedassafetyfeature,particularlyinhotcells.

Type Energy range Measurement range Dimensions, weight

GM-detector 18545 C 40keV-1.3MeV 150nSv/h-200µSv/h length345mm,Ø25/40mm,380g

GM-detector 18550 C 40keV-1.3MeV 1µSv/h-20mSv/h length10mm,Ø40mm,150g

GM-detector 18509 C 55keV-1.3MeV 10µSv/h-1Sv/h length110mm,Ø40mm,150g

GM-detector 18529 C 70keV-3MeV 100µSv/h-10Sv/h length110mm,Ø40mm,150g

NaI-scintillation detector 25B38 25keV-1.3MeV 40nSv/h-200µSv/h length38mm(activerange),Ø25mm,200g

AllALMO-systemscanbecombinedwiththefollowingdetectortypes

Page 3: ALarm MOnitor – stationary dose rate monitoring system ... · PDF fileALMO ALarm MOnitor – stationary dose rate monitoring system with external detectors Fields of application

ALMO build-in version

Especiallyforthemanufacturersofhotcellsandisotopeproductionfacilities,wehavedevelopedabuild-inversionforthealarmmonitorsALMO-3andALMO-6.Insteadofthestandardhousingfordesktoporwallmounting,theALMOisintegratedinametaldie-castcasewithmountingboreholes.Thecaseisinstalledbehindthehotcellcoverorinthedoorofaswitchcabinet.ThefrontpaneloftheALMO-systemwithdisplay,operatorkeyboardandLEDsisembeddedintotheintendedcut-outsofthecoveringorthedoortoformaflatsurfacewiththeoutside.Thissolutionhasessentialadvantagesparticularlywithregardtothepermanentlyin-creasing clean room conditions, because the contamination risks and the extensivecleaningprocessforhousingpartsandcablingareomitted.

ALMO-1Electronics: μ-controller-basedmeasuringelectronicswith1measuringchannel (1detectorconnectable)Display: LCDwithLEDilluminationduringcontinuoususeKeyboard: foilkeyboardHousing: 200x150x75mm(LxWxH),availableaswallordesktophousingWeight: approx.700gPowersupply: 100-240V~,47-63HzPowerconsumption: 15WAlarm: opticalandacoustic,optionallyexternalalarmunits,quitfunctionTemperaturerange: 0°Ctill+50°C,0–95%relativehumidity(nocondensation)Interfaces: 3relayoutputs(24V,500mAorpotential-free,max.24V,1A) e.g.forlampwith3levels,siren,interlock RS232/422/485orUSB-interface(selectableviamenu

ALMO-3LikeALMO-1,butwiththefollowingdifferences:Electronics: μ-controller-basedmeasuringelectronicswith3measuring channels(upto3detectorsconnectable)Housing: 280x300x120mm(LxWxH) availableaswallordesktophousingWeight: approx.2.2kgPowerconsumption: 60WInterfaces: 3relayoutputsperchannel(24V,500mAorpotential- free,max.24V,1A)e.g.forlampwith3levels,siren,interlock 2serialinterfaces(selectableviamenu): A:selectableRS232/422/485orUSB B:selectableRS232/422/485

Technical data

ALMO-6WLikeALMO-3,butwiththefollowingdifferences:Electronics: μ-controller-basedmeasuringelectronics with6measuringchannels(upto6detectorsconnectable)

Page 4: ALarm MOnitor – stationary dose rate monitoring system ... · PDF fileALMO ALarm MOnitor – stationary dose rate monitoring system with external detectors Fields of application

Options and accessories

External alarm unitsAlarmunits,e.g.theLED-lampwith3levels,canbeconnectedtoallALMO-systemstovisualizeanopticalalarm.TheLED-lampsareavailable indifferentversions (1/2/3 levels,withorwithoutacoustics,flashlight).Theycanbeordered forwallmountingaswellas fordesktoporbuild-upinstallation.

Additional external displayDependingonthelocalcircumstancesitmaybenecessarytodisplaythemeasuringvalue of one or more detectors at the measurement or monitoring place, apartfrom the central measurement display of the ALMO. The external measurementdisplayoffersahighcomfortwithtask-specificparametersetting.

SoftwareAll measurement data can be displayedgraphically by the software for measure-mentanddatastorage.Thereforethedoseratemeasurementvaluescanbedisplayedinacurve,asafunctionoftime.Duringtheactualmonitoringprocess,thelamp symbols (green/yellow/red) can beshown for a quick interpretation of themeasurementvalues.

Examples of use

UniversityofManchester,UK,WolfsonMolecularImagingCenter(ALMOwallversion,cableexitupward)

PETResearchCenterRossendorf(ALMOwallversion,cableexitdownward)

Subjecttotechnicalchanges09/10

Headquarters:|D-48249Dülmen

+49-2594-9424-0|[email protected]

+49-2594-942414www.sea-duelmen.de