air master fume hood product catalog (2009)
DESCRIPTION
Air Master System is a U.S. manfacturer or laboratory fume hoods. Located in Michigan, they produce a variety of bench top, walk-in, demonstration, perchloric and radio isotope fume hoods.TRANSCRIPT
telephone 231-798-1111 1
EH-HP High Performance Low Velocity Series Hood ___________________________ 2
Student Work Station ________________________________________________ 4
Eliminator 100 Series Standard Aerodynamic Hood ___________________________ 6
Eliminator 200 Series Double Faced 100 Series Hood _________________________ 8
Eliminator 300 Series Slim Line Hood ____________________________________ 10
Eliminator 400 Series Walk-In Hood _____________________________________ 12
Eliminator 500 Series ADA Hood (Barrier Free Design) ________________________ 14
Eliminator 600 Series Radio Isotope Hood ________________________________ 16
Eliminator 700 Series Perchloric Acid Hood _______________________________ 18
Eliminator 800 Series Flat Front Hood ___________________________________ 20
Eliminator 900 Series Demonstration Hood ________________________________ 22
Paint Spray Booth __________________________________________________ 24
Self-Contained and Portable Hoods _____________________________________ 26
Canopy Hoods ____________________________________________________ 27
Air Monitor Alarms _________________________________________________ 28
Ceiling Enclosures and Finished Backs ___________________________________ 29
Plumbing Accessories _______________________________________________ 30
Electrical Accessories and Specifications _________________________________ 31
Sash Options _____________________________________________________ 32
Base Cabinets ____________________________________________________ 33
Epoxy Resin and Stainless Steel Surfaces _________________________________ 34
Air Chamber Option ________________________________________________ 35
QUICK SHIP Fume Hoods, Epoxy Tops and Base Cabinets __________________ 36
QUICK SHIP Stainless Steel Products _______________________________ 37
Stainless Steel Countertops ___________________________________________ 38
Blowers, Lattice Assemblies and Peg Boards ______________________________ 39
Knock Down Hoods _________________________________________________ 40
Fume Hood General Design Requirements ________________________________ 41
Glossary ________________________________________________________ 44
Color Chart ___________________________________________ inside back cover
Ordering Guide ______________________________________________ back cover
Contents
www.airmastersystems.com2
U.L. 1805Classified
60 FPM • FITS THROUGH STANDARD DOOR • 30” COUNTERTOPThe HP Eliminator is THE hood for remodeling, new construction or retrofitting.
The Eliminator HP SERIES: High Performance/Energy Efficient
A Green AlternativeThe Eliminator HP hood is designed to
operate economically while creating a safe work environment without compromising containment. Buy operating at 60 FPM, air flow across the sash opening results in a 40% savings in fume hood operation costs.
The HP model comes standard with all the same features as other Eliminator models, along with some additional innovations to set this hood apart from its competitors. Features of the HP Series include:• 40-60%energysavings,accomplished
without compromising performance or safety.
• High-efficiencyL.E.D.lightingwithlowmaintenance costs.
• Standard30"deepworksurfaceforzerowork space loss.The HP hood was designed to do all this
andstillfitonastandard-depthcountertopand cabinet. This means no disassembly and reassembly during the installation process. TheHPhoodfitsthroughastandarddooropening, making this hood the choice for thatretrofitjob.
EH-HPStandard Features
• PowderCoatFinish
• FullFrameConstruction
• ShaftDriveSash
• U.L.ListedPolyResinLiner
• LEDLights
• FlushMountAirFoil
• KnockDownCapability
• InteriorAccesstoLight
• Full-LengthFingerLift
• SashSweep
• AirflowMonitor/Alarm
• SashAlarm
KNOCK DOWN—
SEE PAGE 40
telephone 231-798-1111 3
APPLICATIONS: Higher Education; Industrial; Private Research
The Eliminator HP SERIES: High Performance/Energy Efficient
EH-HPOptional Accessories
ITEM AMS PT# PAGE #
CeilingEnclosures CE 29
PlumbingAccessories 30
ElectricalAccessories 31
EpoxyTops ET 34
Alarms ALARM 28
BaseCabinets BC 33
Blowers BL 39
Pre-Plumbing PREP
Pre-Wiring PREW
ITEM AMS PT# PAGE #
SashStops SASHSTOP 32
Chain&SprocketDrive CHNSPK Call
StainlessSteelSash SSSASH 32
StainlessSteelAirfoil SSAF
PushButtonSash PUSHBUTTON 32
CombinationSash H&V 32
LatticeAssembly LATTICE 39
AutoSashReturn 32
Part NumberSuperstructure Worktop Hood Opening CFM
@ 28”*CFM
@ 18”* SPBlowerModel #
ShippingWeightW H D W H D W H
EH-HP-48 48” 56” 33-1/4” 48” 1-1/4” 30” 38” 28” 469 313 .25 BL-HP-48 200
EH-HP-60 60” 56” 33-1/4” 60” 1-1/4” 30” 50” 28” 619 413 .34 BL-HP-60 225
EH-HP-72 72” 56” 33-1/4” 72” 1-1/4” 30” 62” 28” 769 513 .34 BL-HP-72 270
EH-HP-96 96” 56” 33-1/4” 96” 1-1/4” 30” 86” 28” 1069 713 .34 BL-HP-96 300
*Height of open sash; CFM calculated at 60 FPM
EH-HP-48 EH-HP-60
EH-HP-72 EH-HP-96
www.airmastersystems.com4
U.L. 1805Classified
The Eliminator SWS SERIES: Student Work Station
Viewing Everything
TheStudentWorkStationisasafe,highly-visibleworkingenvironmentwithfull360ºviewing.Thishoodhasmany of the standard features that are incorporatedintoeveryAMShood.Inaddition, this hood has the flexibility necessary to allow you to easily design yourworkspacebyputtinghoodsback-to-back,side-by-sideorinjustaboutanyotherconfigurationneeded.Theunique design of this product allows for the plumbing and electrical devices to be easily accessed for installation and maintenance personnel.
Student Work StationStandard Features
• Two-tube,rapidstart,ULlistedfluorescentlightfixturewithsound rated ballast installed on exterior of roof. Provide safety glass panel sealed to the hood roof.
• Fullviewlaminatedsafetyglass.
• Endviewpanels.
• Hoodsmeetorexceed thespecificationssetby ASHRAE110-1995and SAMAstandardLF-10.
• Sashcables:steel,coated, 1/8”diameter.
• Sashcablepulleyassemblyis 2”diameter,ballbearingtype,with cable retaining device.
APPLICATIONS: Higher Education; Industrial; Private Research
telephone 231-798-1111 5
The Eliminator SWS SERIES: Student Work Station
Student Work StationOptional Accessories
ITEM AMS PT# PAGE #
PlumbingAccessories 30
BaseCabinets 33
Electrical 31
Part NumberSuperstructure Worktop Hood Opening CFM
@ 100LFM SP at Hood** Duct SizeW H D W H D W H
EH-SWS-48 48" 60" 32-3/8" 48" 1-1/4" 29" 45-3/8" 29" 770 .25” 4” x 24”
EH-DWS-48 48" 60" 64-3/4" 48" 1-1/4" 58" 45-3/8" 29" 770 .25” 2 – 4” x 24”
EH-SWS-60 60" 60" 32-3/8" 60" 1-1/4" 29" 57-3/8" 29" 1000 .33” 4” x 30”
EH-DWS-60 60" 60" 64-3/4" 60" 1-1/4" 58" 57-3/8" 29" 1000 .33” 2 – 4” x 30”
EH-SWS-72 72" 60" 32-3/8" 72" 1-1/4" 29" 69-3/8" 29" 1250 .5” 4” x 36”
EH-DWS-72 72" 60" 64-3/4" 72" 1-1/4" 58" 69-3/8" 29" 1250 .5” 2 – 4” x 36”
SWS = Single Sided Work StationDWS = Double Sided Work Station (order two single sided hoods and face them back to back)
www.airmastersystems.com6
Minimized Turbulence, Increased Performance
TheEH100seriesairfoil-type
fume hood features a 45-degree angle
around the fascia and a flush-mount
radiused airfoil across the bottom,
whichresultsinminimizedturbulence
and increased performance.
Liketheexteriorofthehood,
theairfoilismadefromchemical-
resistant,powder-coated,16-gauge
cold-rolledsteel.Astainlesssteel
airfoil is available as an option.
Postsare5”(127mm)wide.
Fivestandardsizesofferdesign
flexibility,orwe’llcustomizetoyour
specificneeds.
APPLICATIONS: K-12; Higher Education; Industrial; Private Research
U.L. 1805 Classified
The Eliminator 100 SERIES: Airfoil Fume Hood
EH-100Standard Features
• PowderCoatFinish
• FullFrameConstruction
• ShaftDriveSash
• U.L.ListedPolyResinLiner
• T-8FluorescentLightwithbulbs
• FlushMountAirFoil
• KnockDownCapability
• InteriorAccesstoLight
• Full-LengthFingerLift
• LouveredFrontforBy-Pass
• 10”RoundS/SDuctCollar
KNOCK DOWN—
SEE PAGE 40
telephone 231-798-1111 7
The Eliminator 100 SERIES
Part NumberSuperstructure Worktop Hood Opening CFM
@ 29”*CFM
@ 18”* SPBlowerModel #
ShippingWeightW H D W H D W H
EH-111-36 36” 56” 33-1/4” 36” 1-1/4” 30” 26” 29” 567 354 .25 BL-100-36 190
EH-111-48 48” 56” 33-1/4” 48” 1-1/4” 30” 38” 29” 833 521 .25 BL-100-48 200
EH-111-60 60” 56” 33-1/4” 60” 1-1/4” 30” 50” 29” 1100 688 .34 BL-100-60 225
EH-111-72 72” 56” 33-1/4” 72” 1-1/4” 30” 62” 29” 1367 854 .34 BL-100-72 270
EH-111-96 96” 56” 33-1/4” 96” 1-1/4” 30” 86” 29” 1900 1188 .40 BL-100-96 300
EH-100Optional Accessories
ITEM AMS PT# PAGE #
CeilingEnclosures CE 29
PlumbingAccessories 30
ElectricalAccessories 31
EpoxyTops ET 34
Alarms ALARM 28
BaseCabinets BC 33
Blowers BL 39
Pre-Plumbing PREP
Pre-Wiring PREW
ITEM AMS PT# PAGE #
SashStops SASHSTOP 32
Chain&SprocketDrive CHNSPK Call
StainlessSteelSash SSSASH 32
StainlessSteelAirfoil SSAF
PushButtonSash PUSHBUTTON 32
CombinationSash H&V 32
AutomaticSashReturn AUTORETURN 32
LatticeAssembly LATTICE 39
LEDLights LED
EH-111-36 EH-111-48 EH-111-60
EH-111-72 EH-111-96
*Height of open sash; CFM calculated at 100 FPM
www.airmastersystems.com8
APPLICATIONS: K-12 classrooms and other demonstration settings
The Eliminator 200 SERIES: Double-Faced 100 Series
Dual Access
TheEliminator200seriesprovidesthesame aerodynamic design features as the EH-100AirfoilHoodseries,withentrancefrom two sides. This makes it ideal for use in classroom or prep room settings or whenever demonstration of experiments is desired.
A45-degreeanglearoundthefasciaandaflush-mountradiusedairfoilacrossthebottomprovideminimizedturbulenceandincreased performance.
TheEH-200featuresstandard5”(127mm)poststoallowforelectricalandremotecontrol plumbing services on both sides, andthesecanbeconfiguredtoyourexactspecifications.
Fivestandardsizesandanarrayofoptionsmake this one of the most flexible fume hoods available anywhere.
EH-200Standard Features
• PowderCoatFinish
• FullFrameConstruction
• U.L.ListedPolyResinLiner
• T-8FluorescentLightwithbulbs
• FlushMountAirfoil
• KnockDownCapability
• InteriorAccesstoLight
• Full-LengthFingerLift
• LouveredFrontforBy-Pass
U.L. 1805Classified
telephone 231-798-1111 9
The Eliminator 200 SERIES: Double-Faced 100 Series
Part NumberSuperstructure Worktop Hood Opening CFM
@ 29”*CFM
@ 18”* SPBlowerModel #
ShippingWeightW H D W H D W H
EH-211-36 36” 56” 35-1/2” 36” 1-1/4” 29” 26” 29” 567 354 .25 BL-100-36 190
EH-211-48 48” 56” 35-1/2” 48” 1-1/4” 29” 38” 29” 833 521 .33 BL-100-48 200
EH-211-60 60” 56” 35-1/2” 60” 1-1/4” 29” 50” 29” 1100 688 .34 BL-100-60 225
EH-211-72 72” 56” 35-1/2” 72” 1-1/4” 29” 62” 29” 1367 854 .34 BL-100-72 270
EH-211-96 96” 56” 35-1/2” 96” 1-1/4” 29” 86” 29” 1900 1188 .40 BL-100-96 300
EH-200Optional Accessories
ITEM AMS PT# PAGE #CeilingEnclosures CE 29
PlumbingAccessories 30
ElectricalAccessories 31
EpoxyTops ET 34
Alarms ALARM 28
BaseCabinets BC 33
Blowers BL 39
Pre-Plumbing PREP
Pre-Wiring PREW
ITEM AMS PT# PAGE #SashStops SASHSTOP 32
StainlessSteelSash SSSASH 32
StainlessSteelAirfoil SSAF
PushButtonSash PUSHBUTTON 32
CombinationSash H&V 32
AutomaticSashReturn AUTORETURN 32
LatticeAssembly LATTICE 39
LEDLights LED
SashInterlock 32
*Height of open sash; CFM calculated at 100 FPM with one sash open
EH-211-36 EH-211-48 EH-211-60
EH-211-72 EH-211-96
www.airmastersystems.com10
Expanded Workspace
TheEliminator300seriesisathin-wall fume hood with slimmer front posts to expand the workspace. Posts are 2-1/2”(64mm)wide.
Because the slimmer posts remove the option for wiring and plumbing in the posts, plumbing services must be deck mounted”onepoxytop,andelectricalservices must be mounted on the face ofthecabinet.(Seecabinetoptionsonpage33.)
Fivestandardsizesandanarrayof options make this one of the most flexible fume hoods available anywhere.
APPLICATIONS: K-Middle School
The Eliminator 300 SERIES: Thin Wall
EH-300Standard Features
• PowderCoatFinish
• FullFrameConstruction
• ShaftDriveSash
• U.L.ListedPolyResinLiner
• T-8FluorescentLightwithbulbs
• KnockDownCapability
• InteriorAccesstoLight
• Full-LengthFingerLift
• LouveredFrontforBy-Pass
KNOCK DOWN—
SEE PAGE 40
U.L. 1805Classified
telephone 231-798-1111 11
The Eliminator 300 SERIES: Thin Wall
Part NumberSuperstructure Worktop Hood Opening CFM
@ 28”*CFM
@ 18”* SPBlowerModel #
ShippingWeightW H D W H D W H
EH-311-36 36” 56” 29-1/2” 36” 1-1/4” 30” 26” 28” 646 431 .25 BL-300-36 190
EH-311-48 48” 56” 29-1/2” 48” 1-1/4” 30” 43” 28” 896 597 .25 BL-300-48 175
EH-311-60 60” 56” 29-1/2” 60” 1-1/4” 30” 55” 28” 1146 764 .39 BL-300-60 200
EH-311-72 72” 56” 29-1/2” 72” 1-1/4” 30” 67” 28” 1417 944 .37 BL-300-72 250
EH-311-96 96” 56” 29-1/2” 96” 1-1/4” 30” 91” 28” 1896 1264 .43 BL-300-96 275
EH-300Optional Accessories
ITEM AMS PT# PAGE #
CeilingEnclosures CE 29
PlumbingAccessories 30
ElectricalAccessories 31
EpoxyTops ET 34
Alarms ALARM 28
BaseCabinets BC 33
Blowers BL 39
ITEM AMS PT# PAGE #
SashStops SASHSTOP 32
Chain&SprocketDrive CHNSPK Call
StainlessSteelSash SSSASH 32
CombinationSash H&V 32
AutomaticSashReturn AUTORETURN 32
LatticeAssembly LATTICE 39
LEDLights LED
EH-311-36 EH-311-48 EH-311-60
EH-311-72 EH-311-96
*Height of open sash; CFM calculated at 100 FPM
www.airmastersystems.com12
For Those Tall Experiments
The Eliminator 400 series, with a double-hungsashconfigurationand72-inchhighopening,isdesignedtoaccommodatealarge-apparatusapplication,upto85”inheight.Postsare5”(127mm)wide.
AndwiththeAMSexclusiveknock-downability,evenahoodthisbigisabreezetoinstall,maintain,repair and move.
APPLICATIONS: University, Research, Private Industry
The Eliminator 400 SERIES: Walk-In Hood
EH-400Standard Features
• PowderCoatFinish
• FullFrameConstruction
• ShaftDriveSash
• U.L.ListedPolyResinLiner
• T-8FluorescentLightwithbulbs
• FlushMountAirfoil
• KnockDownCapability
• InteriorAccesstoLight
• Full-LengthFingerLift
• LouveredFrontforBy-Pass
KNOCK DOWN—
SEE PAGE 40
U.L. 1805Classified
telephone 231-798-1111 13
The Eliminator 400 SERIES: Walk-In Hood
EH-400Optional Accessories
ITEM AMS PT# PAGE #
CeilingEnclosures CE 29
PlumbingAccessories 30
ElectricalAccessories 31
Alarms ALARM 28
Blowers BL 39
Pre-Plumbing PREP
Pre-Wiring PREW
SashStops SASHSTOP 32
ITEM AMS PT# PAGE #
Chain&SprocketDrive CHNSPK Call
StainlessSteelSash SSSASH 32
PushButtonSash PUSHBUTTON 32
CombinationSash H&V 32
AutomaticSashReturn AUTORETURN 32
LatticeAssembly LATTICE 39
LEDLights LED
EH-411-72 EH-411-96
Part NumberSuperstructure Hood Opening
CFM @ 36”* CFM @ 72”* SP Blower Model # Approx. Shipping WeightW H D W H
EH-411-48 48” 93” 35” 38” 72” 990 1979 .25 BL-400-48 200
EH-411-60 60” 93” 35” 50” 72” 1306 2613 .34 BL-400-60 225
EH-411-72 72” 93” 35” 62” 72” 1623 3246 .34 BL-400-72 270
EH-411-96 96” 93” 35” 86” 72” 2256 4513 .34 BL-400-96 300
EH-411-48 EH-411-60
*Height of open sash; CFM calculated at 100 FPM
www.airmastersystems.com14
Meets ADA Dimensional Guidelines
TheEliminator500seriesisanairfoil-typefumehooddesignedtomeetADAdimensionalguidelines.
A45-degreeanglearoundthefasciaandaflush-mountradiusedairfoilacrossthebottomresultinminimizedturbulenceandincreasedperformance.
TheADAhood’sairfoilflipsbackoutofthewayforcleanup.Undertheairfoilisacontainment trough to capture accidental spills that may flow over the dished worksurface. The trough assembly is fabricatedfromchemical-resistant16-gaugecold-rolledsteel.
AcombinationverticalandhorizontalsashisstandardontheEH-500.
ADAkneespace-configured”basecabinets,picturedatright,areavailable.Seepage33for cabinet options.See
page 33 for cabinet
options
APPLICATIONS: K-12; University, Private Industry
The Eliminator 500 SERIES: ADA Design
EH-500Standard Features
• PowderCoatFinish
• FullFrameConstruction
• ShaftDriveSash
• U.L.ListedPolyResinLiner
• T-8FluorescentLightwithbulbs
• FlushMountAirfoil
• KnockDownCapability
• InteriorAccesstoLight
• Full-LengthFingerLift
• LouveredFrontforBy-Pass
• PaintedSteelTroughAssembly
• AirflowAlarm
• SashStop
• CombinationSash
• 1/4TurnBallValves
KNOCK DOWN—
SEE PAGE 40
U.L. 1805Classified
telephone 231-798-1111 15
The Eliminator 500 SERIES: ADA Design
EH-500Optional Accessories
ITEM AMS PT# PAGE #
CeilingEnclosures CE 29
PlumbingAccessories 30
ElectricalAccessories 31
EpoxyTops ET 34
Alarms ALARM 28
BaseCabinets BC 33
RollOutBase ROB 33
Blowers BL 39
Pre-Plumbing PREP
ITEM AMS PT# PAGE #
Pre-Wiring PREW
Chain&SprocketDrive CHNSPK Call
StainlessSteelSash SSSASH 32
StainlessSteelAirfoil SSAF
PushButtonSash PUSHBUTTON 32
AutomaticSashReturn AUTORETURN 32
LatticeAssembly LATTICE 39
LEDLights LED
BaseFramesandCabinets 33
Part NumberSuperstructure Worktop Hood Opening CFM
@ 30”*CFM
@ 18”* SPBlowerModel #
ShippingWeightW H D W H D W H
EH-521-48 48” 56” 33-1/4” 48” 1-1/4” 30” 38” 29” 833 521 .25 BL-500-48 200
EH-521-60 60” 56” 33-1/4” 60” 1-1/4” 30” 50” 29” 1100 688 .34 BL-500-60 225
EH-521-72 72” 56” 33-1/4” 72” 1-1/4” 30” 62” 29” 1367 854 .34 BL-500-72 270
EH-521-48 EH-521-60 EH-521-72
Base frames and cabinets sold separately. See page 33.
*Height of open sash; CFM calculated at 100 FPM
www.airmastersystems.com16
For Safe Handling of Radiochemicals
TheEliminator600RadioIsotopeHoodisspecificallyengineeredandbuilttohandleapplications involving radiochemicals.
Ithasthesameaerodynamicdesignasthe standard airfoil hood, with the added featureofradiusconstructiononaone-piecestainless-steellinerandcountertopto prevent absorption of radioactive and corrosive chemicals.
The interior of the hood and integral countertoparefabricatedof304stainlesssteel, buffed and polished to a smooth #4 finish.Thisreducesthechanceofchemicalbuildupandsimplifiesdecontamination.
Theflush-mountradiusedairfoilisalsotype304stainlesssteel,asisthestandardverticalrisingsash.Acombinationsashisavailable as an option.
Fiveavailablesizesandmanyoptionalaccessories meet all your design needs and the end user’s application requirements.
APPLICATIONS: University, Private Industry
The Eliminator 600 SERIES: Radio Isotope Hood
EH-600Standard Features
• PowderCoatFinish
• FullFrameConstruction
• ShaftDriveSash
• T-8FluorescentLightwithbulbs
• FlushMountAirfoil
• InteriorAccesstoLight
• Full-LengthFingerLift
• LouveredFrontforBy-Pass
• 304StainlessSteelOne-PieceIntegralWorkSurfaceandLiner
• StainlessSteelSash
• StainlessSteelAirfoil
U.L. 1805Classified
telephone 231-798-1111 17
The Eliminator 600 SERIES: Radio Isotope Hood
Part NumberSuperstructure Integral Worktop Hood Opening CFM
@ 29”*CFM
@ 18”* SPBlowerModel #
ShippingWeightW H D W H D W H
EH-613-36 36” 56” 33-1/4” 36” 1-1/4” 30” 26” 29” 567 354 .25 BL-100-36 190
EH-613-48 48” 56” 33-1/4” 48” 1-1/4” 30” 38” 29” 833 521 .25 BL-100-48 200
EH-613-60 60” 56” 33-1/4” 60” 1-1/4” 30” 50” 29” 1100 688 .34 BL-100-60 225
EH-613-72 72” 56” 33-1/4” 72” 1-1/4” 30” 62” 29” 1367 854 .34 BL-100-72 270
EH-613-96 96” 56” 33-1/4” 96” 1-1/4” 30” 86” 29” 1900 1188 .40 BL-100-96 300
EH-600Optional Accessories
ITEM AMS PT# PAGE #
CeilingEnclosures CE 29
PlumbingAccessories 30
ElectricalAccessories 31
Alarms ALARM 28
BaseCabinets BC 33
Blowers BL 39
Pre-Plumbing PREP
Pre-Wiring PREW
ITEM AMS PT# PAGE #
SashStops SASHSTOP 32
Chain&SprocketDrive CHNSPK Call
PushButtonSash PUSHBUTTON 32
CombinationSash H&V 32
AutomaticSashReturn AUTORETURN 32
LatticeAssembly LATTICE 39
LEDLights LED
EH-613-36 EH-613-48 EH-613-60
EH-613-72 EH-613-96
*Height of open sash; CFM calculated at 100 FPM
www.airmastersystems.com18
Safe Handling of Perchloric Acid
TheEliminator700PerchloricAcidHoodhas the same aerodynamic design as the 100series,witha45-degreeanglearoundthefasciaandaflush-mountradiusedairfoilacrossthebottomtominimizeturbulenceandprovide increased performance.
Thishard-workinghoodalsofeaturesabuilt-introughtocapturerunoffinwashdownprocedureswhereperchloricacidisinuse.Inaddition, there is a manual washdown feature for the areas behind the baffles as well as the duct-collarareas.
Theinteriorisfabricatedwithtype316stainlesssteel,integrallyweldedtoa316stainless-steelworksurface,thenbuffedandpolishedtoa#4finish.Thetopis1-1/4’witha1/4-inchdishtoretainspills.Theairfoilandsasharealsomadeof316stainlesssteel.
TheEH-700isbuilttoughtowithstandthedaily use of perchloric acid, while keeping the workplace safe.
APPLICATIONS: University, Private Industry
The Eliminator 700 SERIES: Perchloric Acid Hood
EH-700Standard Features
• PowderCoatFinish
• FullFrameConstruction
• ShaftDriveSash
• T-8FluorescentLightwithbulbs
• FlushMountAirfoil
• InteriorAccesstoLight
• Full-LengthFingerLift
• LouveredFrontforBy-Pass
• 316StainlessSteelOne-PieceIntegralWorkSurfaceandLiner
• Built-InTrough
• ManualWashDown
• StainlessSteelSash
• StainlessSteelAirfoil
U.L. 1805Classified
telephone 231-798-1111 19
The Eliminator 700 SERIES: Perchloric Acid Hood
Part NumberSuperstructure Integral Worktop Hood Opening CFM
@ 29”*CFM
@ 18”* SPBlowerModel #
ShippingWeightW H D W H D W H
EH-716-36 36” 56” 33-1/4” 36” 1-1/4” 30” 26” 29” 567 354 .25 BL-100-36 190
EH-716-48 48” 56” 33-1/4” 48” 1-1/4” 30” 38” 29” 833 521 .33 BL-100-48 200
EH-716-60 60” 56” 33-1/4” 60” 1-1/4” 30” 50” 29” 1100 688 .34 BL-100-60 225
EH-716-72 72” 56” 33-1/4” 72” 1-1/4” 30” 62” 29” 1367 854 .34 BL-100-72 270
EH-716-96 96” 56” 33-1/4” 96” 1-1/4” 30” 86” 29” 1900 1188 .40 BL-100-96 300
EH-700Optional Accessories
ITEM AMS PT# PAGE #
CeilingEnclosures CE 29
PlumbingAccessories 30
ElectricalAccessories 31
Alarms ALARM 28
BaseCabinets BC 33
Blowers BL 39
Pre-Plumbing PREP
Pre-Wiring PREW
ITEM AMS PT# PAGE #
SashStops SASHSTOP 32
Chain&SprocketDrive CHNSPK Call
PushButtonSash PUSHBUTTON 32
CombinationSash H&V 32
AutomaticSashReturn AUTORETURN 32
LatticeAssembly LATTICE 39
LEDLights LED
EH-716-36 EH-716-48 EH-716-60
EH-716-72 EH-716-96
*Height of open sash; CFM calculated at 100 FPM
www.airmastersystems.com20
Economical School Hood
AbasicfumehoodforK-12educational applications, the EH 800seriesissimilarindesignandconstructiontotheEH-300(ThinWall)hood,butwith5”(127mm)poststoprovideplumbingandelectricalfixtureson the face of the hood. This hood has many of the same features found in otherAMShoods,butitiseconomicallypriced for secondary school budgets.
APPLICATIONS: K-12
The Eliminator 800 SERIES: Flat Front
EH-800Standard Features
• PowderCoatFinish
• FullFrameConstruction
• ShaftDriveSash
• U.L.ListedPolyResinLiner
• T-8FluorescentLightwithbulbs
• KnockDownCapability
• InteriorAccesstoLight
• Full-LengthFingerLift
• LouveredFrontforBy-Pass
• 10"RoundDuct
KNOCK DOWN—
SEE PAGE 40
U.L. 1805Classified
telephone 231-798-1111 21
The Eliminator 800 SERIES: Flat Front
Part NumberSuperstructure Worktop Hood Opening CFM
@ 29”*CFM
@ 18”* SPBlowerModel #
ShippingWeightW H D W H D W H
EH-811-36 36” 56” 29-1/2” 36” 1-1/4” 30” 26” 29” 567 354 .25 BL-100-36 190
EH-811-48 48” 56” 29-1/2” 48” 1-1/4” 30” 38” 29” 833 521 .33 BL-100-48 200
EH-811-60 60” 56” 29-1/2” 60” 1-1/4” 30” 50” 29” 1100 688 .34 BL-100-60 225
EH-811-72 72” 56” 29-1/2” 72” 1-1/4” 30” 62” 29” 1367 854 .34 BL-100-72 270
EH-811-96 96” 56” 29-1/2” 96” 1-1/4” 30” 86” 29” 1900 1188 .40 BL-100-96 300
EH-800Optional Accessories
ITEM AMS PT# PAGE #
CeilingEnclosures CE 29
PlumbingAccessories 30
ElectricalAccessories 31
EpoxyTops ET 34
Alarms ALARM 28
BaseCabinets BC 33
Blowers BL 39
Pre-Plumbing PREP
Pre-Wiring PREW
ITEM AMS PT# PAGE #
SashStops SASHSTOP 32
Chain&SprocketDrive CHNSPK Call
StainlessSteelSash SSSASH 32
PushButtonSash PUSHBUTTON 32
CombinationSash H&V 32
AutomaticSashReturn AUTORETURN 32
LatticeAssembly LATTICE 39
LEDLights LED
EH-811-36 EH-811-48 EH-811-60
EH-811-72 EH-811-96
*Height of open sash; CFM calculated at 100 FPM
www.airmastersystems.com22
Three Windows for Optimum Viewing
TheEH-900seriesisaneducationalviewing fume hood with one slimmed down(2”or51mm)frontpostandonestandard(5”or127mm)posttoallow for more workspace and electrical services. Plumbing service must be deck mounted.
TheEH-900isavailableineitherright-handorleft-handorientation.Thishoodfeaturesatwo-sashconfigurationandathird(end)viewingwindow, making it ideal for use in high schools or any environment where demonstration of experiments is desired. The end window may beeliminatedforthrough-wallinstallations.
(Left-handshown)
APPLICATIONS: K-12
The Eliminator 900 SERIES: Demonstration Hood
EH-900Standard Features
• PowderCoatFinish
• FullFrameConstruction
• U.L.ListedPolyResinLiner
• T-8FluorescentLightwithbulbs
• KnockDownCapability
• InteriorAccesstoLight
• Full-LengthFingerLift
• LouveredFrontforBy-Pass
• 10"RoundDuct
U.L. 1805Classified
telephone 231-798-1111 23
The Eliminator 900 SERIES: Demonstration Hood
EH-900Optional Accessories
ITEM AMS PT# PAGE #CeilingEnclosures CE 29
PlumbingAccessories 30
ElectricalAccessories 31
EpoxyTops ET 34
Alarms ALARM 28
BaseCabinets BC 33
Blowers BL 39
Pre-Plumbing PREP
Pre-Wiring PREW
ITEM AMS PT# PAGE #SashStops SASHSTOP 32
StainlessSteelSash SSSASH 32
PushButtonSash PUSHBUTTON 32
CombinationSash H&V 32
AutomaticSashReturn AUTORETURN 32
LatticeAssembly LATTICE 39
LEDLights LED
SashInterlock 32
Part NumberSuperstructure Worktop Hood Opening CFM
@ 26.5”*CFM
@ 18”* SPBlowerModel #
ShippingWeightW H D W H D W H
EH-911-60 60” 48” 30” 60” 1-1/4” 32” 53” 26.5”** 1030 722 .75 BL-900-60 185
Please specify Left or Right (Left shown) *Height of open sash; CFM calculated at 100 FPM with one sash open**Each side
EH-911-60LEH-911-60R
www.airmastersystems.com24
The Eliminator Paint (Spray) Booth
TheAMSPaintBoothisfabricatedof18-gauge steel with a corrosion resistant, powder coat baked epoxy finish. The upper booth enclosure includes a standard vapor-proof light as well as two filters set in an easy-to-change flange bracket system. The globed interior light is mounted and wired toacommonjunctionbox.AswithallAMSproducts,thelightfixtureisshippedwithbulbs, standard.
The optional blower motor operates at anRPMdesignedtoprovide100LFMandapproximately3⁄4”ofstaticpressure.Iftheblower is ordered, it will be mounted on top andwiredtothecommonjunctionbox.Thetransition piece from the booth to blower will be provided and connected. Both the light and theblowercanbepre-wiredtoswitchesinthefront post area of the booth when required.
Sold separatelyPart No. TT12CR
Dimensions:Paintboothsareavailableinwidthsof24”,30”,36”,48”and60” Allare48”Hx24”D.Specialsizesavailableuponrequest.
telephone 231-798-1111 25
Part Number Description CFM @ 100 FPM
AMS-24-PB 24” Paint (Spray) Booth 472
AMS-30-PB 30” Paint (Spray) Booth 613
AMS-36-PB 36” Paint (Spray) Booth 755
AMS-48-PB 48” Paint (Spray) Booth 1040
AMS-60-PB 60” Paint (Spray) Booth 1322
Paint (Spray) Booth
Paint (Spray) Booth with Blower
The Eliminator Paint (Spray) Booth
Explosion-Proof options
NOTE: Explosion-proof (X-P) switches must mount on the exterior of booth or be shipped loose for remote operation.
Paint BoothOptional Accessories
ITEM AMS PT# PAGE #LightSwitch LTSW 31
BlowerSwitch BLSW 31
StainlessSteelTop 34
ITEM AMS PT# PAGE #GalvanizedTop Call
Pre-WireService PREWIRE
AMS-30-PB AMS-48-PB
AMS-24-PB AMS-36-PB
Part Number Description CFM @ 100 FPM
AMS-24-XPB 24” Booth w/X-P Light 472
AMS-30-XPB 30” Booth w/X-P Light 613
AMS-36-XPB 36” Booth w/X-P Light 755
AMS-48-XPB 48” Booth w/X-P Light 1040
AMS-60-XPB 60” Booth w/X-P Light 1322
Part Number Description CFM @ 100 FPM
AMS-24-PBW 24” Booth and Blower 472
AMS-30-PBW 30” Booth and Blower 613
AMS-36-PBW 36” Booth and Blower 755
AMS-48-PBW 48” Booth and Blower 1040
AMS-60-PBW 60” Booth and Blower 1322
Part Number Description CFM @ 100 FPM
AMS-24-XPBW 24” Booth w/X-P Light and Blower 472
AMS-30-XPBW 30” Booth w/X-P Light and Blower 613
AMS-36-XPBW 36” Booth w/X-P Light and Blower 755
AMS-48-XPBW 48” Booth w/X-P Light and Blower 1040
AMS-60-XPBW 60” Booth w/X-P Light and Blower 1322
www.airmastersystems.com26
Optional Accessories
• 208,230or277VoltMotor
• Pre-filterTypesandEfficiencies
• PTEFandGelSealFilters
• ULPAUpgrades
• 1”EpoxyTop
• MobileCartw/LockingCasters
• PhotohelicGauge
• StainlessSteelConstruction
• Totally-ClosedFront
Standard Features
• PaintedSteelContruction
• Energy-SavingLEDLightFixture
• CombinationLight/Blower Switch
• GFIDuplexOutlet
• Pre-wired
• 8’PowerCord
• MagnahelicGauge
TheAMSself-containedhoodisacompletelyself-containedunit.Thebackward-curved120voltblowerhasavariablespeedcontroltomaximizeefficiencyandminimizemotornoise.The53mm99.99%efficientHEPAfilterisreplaceablefromthetopoftheunit.Italsocontainsaone-inch30%pre-filterthatisreplaceablefromtheinterioroftheunit.Thissitsatopafully-weldedtubechamberwithplexiglas windows for viewing from all four sides. The front hasafold-updoortoallowforloadingandunloadingofapparatus and experiments.
AMS Self-Contained HoodAPPLICATIONS: K-12, University, Private Industry
Model numbers: AMS-36-SC hoodonlyAMS-36-SCT hoodwithepoxytopAMS-36-SCTC hoodwithtopandcartForstainlesssteelconstructionusetheprefixSS
AMS Portable HoodThisportablehoodhasthesameconstructionfeaturesastheself-
containedunit,butmustbeconnectedtoanHVACsystem.
Standard Features
• PaintedSteelContruction
• Energy-SavingLEDLightFixture
• CombinationLight/Blower Switch
• GFIDuplexOutlet
• Pre-wired
• 8’PowerCord
Model numbers: AMS-36-PH hoodonlyAMS-36-PHT hoodwithepoxytopAMS-36-PHTC hoodwithtopandcartForstainlesssteelconstructionusetheprefixSS
telephone 231-798-1111 27
AMScanopyhoodsaredesignedtocollectandexhaust
heat, steam and odors when mounted above hot plates,
water baths or portable equipment. They are available in
three different models as pictured here and in six standard
widths:24”,30”,36”,48”,60”,and72”.Customsizes
are also available.
Standardcanopyhoodsareavailablein18-gaugecold
rolledsteel,finishedwithacidwashandpowdercoatepoxy
paint.(Seeinsidebackcoverforavailablestandardcolors.)
Stainlesscanopyhoodsarefabricatedof18-gauge304
stainlesssteelwitha#4smoothgrainfinish
Two stainless steel threaded rods come with FS and SE
models;fouraresuppliedwithILmodels.
To order canopy hoods:
CH-FS forflatbackandflatsides
CH-IL forcenterstyle
CH-SE forflatbackandslopedsides
For stainless steel option, add -SS”
For example:
CH-FS-4824 isapaintedsteel,48”wide,24”deep
canopy hood with flat back and flat sides.
CH-IL-6030-SS isa60”wide,30”deepislandcanopywith
stainless steel construction.
Canopy Hoods
www.airmastersystems.com28
AFA 1000/1Mk3
Fume Hood Airflow MonitorSpecifications
Part #AMS-AL1D
Display range 0-999 fpm (0-5.0 m/s)Alarm range 0-999 fpm (0-5.0 m/s)Field set-up 2-point velocity calibration ( with on-screen instructions )Accuracy Sensor / Display resolution 1 fpm Face velocity accuracy +/-10%Alarm delays User configurable – 0 to 60secsRelay output 4 (1 on board – 3 on optional plug in relay interface unit)Analog output 0-10V output, directly proportional to velocity (Optional)Relay input 3Comm. Port RS232 – Can be connected via Serial interface to LAN network (Full software available)Sash high indication Yes, with separate plug-in connectionNight setback YesExternal alarm indication YesPower Requirement Input–120VAC, 60Hz Output–15VDC, 500maUnits English and Metric (user selectable)Display–visual Analog bar graph or Fault Timeline LEDs: red, alarm; yellow, caution; green, normal Digital display of velocity reading (can be turned off)Alarm indication Red LED & audible alarm (to a certain sound spec)Horn silence Yes (temporary / permanent/automatic depending on type of alarm or external input)Mounting Semi FlushOperating temperature 55-86 F (13-30 C)Storage temperature -40-150 F (-40-65 C)Instrument dimensions Instrument Case: 5.2” H x 3.19” W x 1.14” D (132mm H x 81mm W x 29mm D)Agency listings UL & CE
Alarm range 30-400 fpm (.15 -2.0 m/s)Accuracy Face velocity accuracy +/- 10%Airflow sensor On-board or remote sensorCalibration Single or 2-point ( Installer selectable )Low Air Alarm delay Fixed 5 secsRelay output 1 -–( Low Air alarm )Analog output Not availableRelay input 2 – Night setback & sash highComm. Port RS232 – Can be connected via serial interface to LAN network (Full hood performance software reporting available)Sash high indication Yes – using a Micro switch or Proximity switch input With repeat alarm feature factory set to 5 min. (Can be adjusted via laptop up to 30 mins)Night setback Yes – using a relay inputExternal alarm indication Not availablePower Requirement Input—120VAC, 60Hz Output—15VDC, 500maDisplay—visual LEDs: red, alarm; green, normalAlarm indication Red LED & audible alarmHorn silence Yes – temporaryMounting Semi FlushOperating temperature 55-86 F (13-30 C)Storage temperature -40-150 F (-40-65 C)Instrument dimensions Instrument Case: 5.2” H x 3.19” W x 1.97” D (132mm H x 81mm W x 50mm D)Agency listings UL & CE
AFA 500Mk3
Fume Hood Airflow MonitorSpecifications
Part #AMS-ALRG
Air Monitor Alarms
telephone 231-798-1111 29
Ceiling Enclosures & Finished BacksAirMasterSystemsCorporationoffersceilingenclosuresandfinishedbackstoprovideaprofessional,finished
designinanysetting.Ceilingenclosuresandfinishedbacksarefabricatedofcoldrolledsteelandfinishedto
match the fume hood superstructure.
The ceiling enclosures are available in two types: for standard fume hoods and also for those fume hoods
utilizingtheoptionalAirChamber.Bothenclosuresconcealandprotectthehoodductwork,electricalconduit
and other supply lines.
Ceiling Enclosures & Finished Backs
Finished Backs: Order FINBACK-W where
W equals hood width.
Ceiling Enclosures: Order CE-W-H where W is hood width and H is height of enclosure required. Example: CE-48-18 is a set of enclosures for a 48” wide hood which are 18” high.
Finished backs are used to close off the back of the fume hood,
givingitafinishedlookwhenlocatedinthecenterofaroom.
www.airmastersystems.com30
Remote Control Fixtures
90º Water FaucetPart # FXTDS
Deck Mount WaterPart # FXTDMW
AllAirMasterSystemsCorporationremoteoperatedfixturescomestandardwithfour-pronghandles,bakedcolorcodedepoxyupperassembliesandcontrolrodscuttolength.Chromeandotherstylesareavailableuponrequest.Pre-plumbedservicesare3/8”blackpipeforgasesand1/2”rigidcopperwithsolderedjointsforwaterandotherlabservices.
Deck Mount GasSingle Outlet
Part# FXTDMG1
Deck Mount GasDouble Outlet
Part # FXTDMG2
Hot & Cold Water MixingPart # FXTCWM
Gooseneck Faucet with Vacuum Breaker
Part # FXTGNWVBSpecify Hot or Cold
Gooseneck FaucetPart # FXTGN
Specify Hot or Cold
Description Part #
Gas FXTGAS
Air FXTAIR
Vacuum FXTVAC
Nitrogen FXTNIT
Cold water FXTCW
Plumbing Accessories
Others available upon request
Panel Mounted Mixing Faucetwith Swing Spout, Lever Handles
Part # FXTPMSGN-6Part # FXTPMSGN-8Part # FXTPMSGN-10Part # FXTPMSGN-12Part # FXTPMSGN-14Part # FXTPMSGN-16
telephone 231-798-1111 31
Pre-Wire service is completed in accordance with NEC and UL-1805 standards.Note: If the fume hood is not pre-wired, wiring in the field must be accomplished using UL listed electrical fixtures while observing NEC standards and local electrical codes.
Electrical Accessories and Specifications
Single Pole 15 Amp Light Switch
Standard Black Color & Cover PlatePart# LTSW
Red Illuminated Toggle Switch with Black Cover Plate
Part# BLSW
Explosion Proof OutletClass 1, Group 1
Part# XPO
Explosion Proof SwitchClass 1, Group 1
Part# XSW
208 Volt, 20 Amp Single Pole Receptacle
230, 277 or other available upon request
Part# 208
120 volt 20 Amp Ground Fault Interrupter (GFI)
Duplex Outlet Standard Black color with cover
Part# GFI
Explosion Proof LightClass 1, Group 1
Available in incandescent or fluorescent
Part# XPL
www.airmastersystems.com32
Sash Options
Combo Horizontal/Vertical Sash
EH-121-72
Horizontal Sash Only
EH-131-72
Interlocking Sash
EH-141-72
Hanging Horizontal Doors
EH-451-72
Additional Sash Options
•SASHALARM: a red light alarm activated when sash is openedaboveapre-setheight
•SASHSTOP: a device to limit sash opening, with manual override
•AUTOSASHRETURN:returnssashto18”workingheight when opened beyond set height
•SASHINTERLOCK: allows only one sash to be opened at anytimeonadouble-sidedhood
•DOUBLEHUNGSASH:for larger openings with ceiling height limitations
•SASHLOCK: a keyed lock that keeps the sash closed if needed in a classroom setting
•PUSHBUTTONSASH*: opens and closes sash with the push of a button
•AUTO-SENSINGSASH*: automatically closes sash when operator walks away
Toordertheseoptionalsashes,pleaserefertotheOrderingGuideonthebackcoverofthiscatalog.Thelarge,underlined number in each of the part numbers below is the Sash Style number to use when ordering.
*These two types of sashes have a wide range of capabilities. Please call an Air Master Systems representative or the factory to determine which application is right for your needs.
telephone 231-798-1111 33
AlthoughAMSisyour fume hood expert, we do offer a limited selection of metal base cabinets manufactured by our approved cabinet suppliers to help you complete your fume hoodproject.Whena large quantity of standard laboratory cabinets is required, however, we suggest you work directly with oneoftheAMScabinetpartners.Callusforan approved list of suppliers.
AMSfumehoodbasecabinets conform to the method of testing and performance requirements set forth in theScientificApparatusMakersAssociation(SAMA).AllADAandhood base cabinets can be manufactured in stainless steel.
Base Cabinets
HBF-3048HBF-3348
ADA-3012ADA-3018ADA-3024
HB-3612*HB-3630HB-3636HB-3648
ASC-30ASC-36ASC-48
FSC-30FSC-36FSC-48
ROB-35
ADA Base Frames ADA Hood Bases
Must be specified left- or right-hand; right-hand shown
ADA-3312ADA-3318ADA-3324
Must be specified left- or right-hand
Hood Bases
EXAMPLE: HB-3630 = 36”h x 30”w
NOTE: These cabinets also come in 33” high and 30” high
*Single door; specify left- or right-hand
NOTE: ADA cabinet configurations come in both 30” and 33” high.
Acid Storage Cabinet
Acid cabinet can be painted
any standard color or safety blue.
Flammable Storage Cabinet Roll Out Flammable Storage Cabinet**
**Designed for kneespace application.
NOTE: All flammable storage cabinets come standard in safety yellow.
LEFT-HAND
RIGHT-HAND
www.airmastersystems.com34
Work SurfacesAnotherkeycomponentofeffective
andefficientfumehoodutilizationisthetypeofworksurfacetouse.Inappropriateworksurfaces can interfere with lab processes and beapotentialdangertolabpersonnel.AirMasterSystemsCorporationprovidesdurable,high-performanceepoxyresinandstainlesssteel surfaces that feature top quality materials and workmanship.
Epoxy Resin
Stainless Steel
Chemical Resistance S G
Heat Resistance S G
Stain Resistance G G
Moisture Resistance E S
Impact Resistance E S
Abrasion Resistance G E
Cleanability G E
Sanitation P E
Spill Containment G G
Maintenance Cost Low Low
Smoke Toxicity Low Low
Disposal Impact Low Low
S=Superior E=Excellent G=Good F=Fair P=Poor
The above chart of chemical resistance test results represents a sampling of data. Complete test results for additional solvents and chemicals can be obtained by contacting us.
Epoxy ResinPrimarily used for all general purpose fume hoodsandbasecabinets,AMSFumeHoodepoxyresintopsare1-1⁄4”(3.2cm)thickwithablack,lowsheenfinishandfeatureadeepdished3/8”(0.97cm)workarea.Oval3”x6”(7.65x15.3cm)cupsinksarestandard;locationsmustbespecifiedwhenordering.
Stainless SteelStainless steel work surfaces are fabricated of304stainlesssteel*,are1-1/4”(3.2cm)thickanddished3/8”witha#4smoothsatinfinish.Square3”x6”(7.65x15.3cm)weldedcupsinksarestandard;locationsmustbespecifiedwhenordering.
*In perchloric acid applications, 316 stainless steel is used for lab safety.
200 Series Epoxy Work Surfaces
Epoxy Resin & Stainless Steel Surfaces
300 Series Epoxy Work Surfaces
100 & 500 Series Epoxy Work Surfaces
800 Series Epoxy Work Surfaces
telephone 231-798-1111 35
TheAirChambersystemreducesenergycostsforheatingandcoolingtoanyfumehoodbyasmuchas70percent.This is achieved by capturing air from outside the laboratory and supplying it to the hood through a supply duct to the AddAirchamber.Anadditional,separateblowerandductare required for the air supply.
Increased FlexibilityTheAddAirchambercanbeusedonanyEliminatorhood
and coordinates with all Eliminator standard features in design,manufacturing,construction,finishing,andfixtures.
ToordertheAddAiroption,includethelettersAA”afterthewidthdimension.SeeOrderingGuide”onbackcover.
Optional feature reduces energy costs by up to 70 percent
Blower information is provided for general purposes only based on a full open sash and static pressure for the hood only. Corners and additional duct runs will increase the static pressure and blower requirements. Please consult with your HVAC professional for your specific requirements.
Supply Blower InformationModel 70% of Hood CFM SP @ Hood Duct Size
EH-111-36AC 400 .15 Inches 4” x 17”
EH-111-48AC 580 .18 Inches 4” x 17”
EH-111-60AC 765 .22 Inches 4” x 22”
EH-111-72AC 950 .28 Inches 4” x 22”
EH-111-96AC 1315 .42 Inches (2) 4” x 17”
Supply Blower InformationModel 70% of Hood CFM SP @ Hood Duct Size
EH-411-48AC 580 .18 Inches 4” x 17”
EH-411-60AC 765 .22 Inches 4” x 22”
EH-411-72AC 950 .28 Inches 4” x 22”
EH-411-96AC 1315 .42 Inches (2) 4” x 17”
Supply Blower InformationModel 70% of Hood CFM SP @ Hood Duct Size
EH-311-36AC 470 .15 Inches 4” x 17”
EH-311-48AC 645 .18 Inches 4” x 17”
EH-311-60AC 820 .22 Inches 4” x 22”
EH-311-72AC 995 .28 Inches 4” x 22”
EH-311-96AC 1345 .42 Inches (2) 4” x 17”
The Eliminator Air Chamber–an option for all hood styles
www.airmastersystems.com36
QUICK SHIP Fume Hoods, Epoxy Tops and Base CabinetsAllAMSQUICKSHIPproductsshipwithin10days.Someitemsareavailablewithin24hours.Fordetailedinformation and available quantities, go to www.airmastersystems.com, contact a factory representative or callAMSdirectly.
HOOD ONLY100 Series Hoods
QS-148QS-160QS-172
HOOD, TOP AND SINK100 Series Hoods
QS-148-TSQS-160-TSQS-172-TS
NOTE: Allhoodscomewithlightswitch,blowerswitch,twoGFIoutlets,onegasspout,oneCWdownspout,prewiredandpreplumbed on the right.
QUICK SHIPFume Hoods and Epoxy Tops
HOOD BASESQS-12-HBQS-30-HBQS-36-HBQS-48-HB
QUICK SHIPBase Cabinets
All QUICK SHIP products come in standard Champagne color.
ADA CONFIGURATIONSQS-12-ADA BASE CABINETQS-48-ADA BASE FRAME
FLAMMABLE STORAGE CABINETSQS-30-FSCQS-36-FSCQS-48-FSC
COLOR: Safety Yellow
ACID STORAGE CABINETSQS-30-ASCQS-36-ASCQS-48-ASC
COLOR: Safety Blue
For additional detailed information, see page 33.
telephone 231-798-1111 37
QUICK SHIP Stainless Steel Products
Stainless Steel Wall Shelving Model Dimensions B x A
WS-1224 12”x24”
WS-1236 12”x36”
WS-1248 12”x48”
WS-1260 12”x60”
WS-1272 12”x72”
WS-1296 12”x96”
WS-1624 16”x24”
WS-1636 16”x36”
WS-1648 16”x48”
WS-1660 16”x60”
WS-1672 16”x72”
WS-1696 16”x96”
Most stainless steel items ship within 72 hours— please check www.airmastersystems.com for availability.
Stainless Steel Drop-in Sinks Model Dimensions
DIS-15155 15”x15”x5”
DIS-14101 14”x10”x10”
DIS-14105 14”x10”x5”
DIS-16145 16”x14”x5”
Stainless Steel Work Tables and Sinks
Model Length = A
WT-3-1818-12 90” w/3 sinks
WT-3-1620-12 84” w/3 sinks
WT-3-1014-10 60” w/3 sinks
Other sizes and configurations available upon request—please call for a quote.For more information and available quantities, go to www.airmastersystems.com
Specialty Hand Steel Sinks/Wall Mount
Model Sink Size
AMHS-W-SS 9” x 9” x 4-1/4”
AMHS-ADA-S 14”x 16”x 5”
Drip Troughs with TubingModel Size
AMHDT-18-2 18”
AMHDT-24-2 24”
AMHDT-30-2 30”
AMHDT-36-2 36”
Stainless Steel Pegboards with Pegs
Model Size
AMV2430 24”w x 30”tall 32 pegs
AMV3030 30”w x 30”tall 50 pegs
AMV3024 30”w x 24”tall 25 pegs
AMV3630 36”w x 30”tall 60 pegs
Stainless Steel Double-Locking Narcotics Cabinet
Model Size
AMNL-2K 12”x 13”x 10”deep
www.airmastersystems.com38
Primarily used for all general purpose fume hoods and base cabinets, standard stainless steel countertops are fabricated of16-gauge304*stainlesssteelwitha#4smoothgrainfinishandare11⁄4”(3.2cm)thickanddished3/8”(0.97cm).A4”integral back splash comes standard and, when sinks or end splashes are required, they are continuously welded to the top and professionallybuffedtoa#4finish.Standarddepthis30”andtheycanbemadeuptoanylengthrequiredtoincludeafieldjointorcontinualcounterdependingonthespecificjobrequirements.Thetopsalsohavearubberizedsounddeadeningmaterialappliedto soften the sound of the top and make it more acoustically appealing.
Square3”x6”(7.65x15.3cm)weldedcupsinks are standard; locations must be specifiedwhenordering.
*Use of perchlorics requires use of 316 stainless steel for lab safety.
SECTION THRUSECTION THRU
SPECIAL COUNTERTOP w/DOUBLE SINK SPECIAL COUNTERTOP w/EXTRA DEEP SINK
MARINE EDGE COUNTERTOPS ALSO AVAILABLE
SECTION THRU
Custom Stainless Steel Countertops
telephone 231-798-1111 39
Callforpricing.
Lattice Assembly
Callforpricing.
Peg Boards
Blowers, Lattice Assemblies and Peg Boards
AirMasterSystemscansupplytheblower units for all your fume hood needs. OurstandardblowersareallAMCA(AirMovementandControlAssociation)certifiedrated. The blowers are available in direct drive,beltdriveandFRP(fiberglassreinforcedplastic)tomeetstheneedsofyour laboratory environment. Please contact our customer service department to help chose the right fan for your application.
Latticeassemblies,commonlyreferredtoas
monkeybars,”areneededforcomplexdistillation
setups.Standarddesignsaremadewith1/2”solid
aluminumrod.Theyarealsoavailablein304
stainlesssteel.Manydifferentsizesareavailable;
please call a factory representative for details.
Blower
AMSstainlesssteelpegboardsaredesignedto
accommodate every application. These boards are
available in stainless steel or acrylic, and can be
wall mounted, freestanding, or may be used with
wallstandardstoallowforheightadjustments.
Thereisafinishedbackoptionforuseinisland
endapplications.Contactafactoryrepresentative
forsizesandavailability.
www.airmastersystems.com40
Knock Down Hoods
MostAMShoodscanbeknockeddowninto manageable modules. This can be very usefulwhenretrofittingoldlabs,orforeasierinstallation of large hoods. This feature also allows for container shipment packaging. ContactyourcustomerservicerepresentativeforacopyoftheHow-Tovideo.
Easy disassembly and reassembly for a multitude of applications
telephone 231-798-1111 41
1.01 FUMEHOODGENERALDESIGNREQUIREMENTS A. Fumehoodsshallfunctionasventilated,enclosed
workspaces,designedtocapture,confineandexhaust fumes, vapors and particulate matter produced or generated within the enclosure.
B. Designfumehoodsforconsistentandsafeairflowthroughthehoodface.Negativevariationsoffacevelocityshallnotexceed20%oftheaverageface velocity at any designated measuring point as definedinthissection.
C. Averageilluminationofworkarea:Minimum80foot-candles.Workareashallbedefinedastheareainside the superstructure from side to side and from face of baffle to the inside face of the sash, and fromtheworkingsurfacetoaheightof28inches.
D. Fumehoodshallbedesignedtominimizestaticpressure loss with adequate slot area and stainless stealexhaustcollarconfiguration.Maximumaveragestatic pressure loss readings taken three diameters abovethehoodoutletfromfourpoints,90degreesapart, shall not exceed the following maximums.
FaceVelocity MeasuredS.P.L.(W.G) F.P.M. .18inches 100F.P.M. .30inches 125F.P.M. .45inches 150F.P.M. .60inches
E. Fume hood shall maintain essentially constant exhaust volume at any baffle position for safety. MaximumvariationinexhaustCFM,staticpressureand average face velocity as a result of baffle adjustmentshallnotexceed10%foranybafflepositionatthespecifiedfacevelocity.
1.02 SUBMITTALS
A. ShopDrawings:Indicateequipmentlocations,large-scale plans, elevations, and cross sections, rough in and anchor placement dimensions and tolerances and all required clearances.
B. ProductData:Submitmanufacturer’sdataforeach component and item of laboratory equipment specified.Includecomponentdimensions,configurations,constructiondetails,jointdetails,and attachments, utility and service requirements and locations.
C. Samples:Submit3”x6”inchsamplesoffinishfor fume hood, work surfaces and for other pre finishedequipmentandaccessoriesforselectionbyArchitect.
D. TestReports:Submittestreportsverifyingconformance to test performances specified.Submitindependenttestsasspecified.
1.03 QUALITYASSURANCE
A. Singlesourceresponsibility:Fumehoodcasework,work surfaces, and other laboratory equipment and accessories shall be manufactured or furnished by a single laboratory furniture company.
B. Manufacturer’squalifications:Modernplantwithpropertools,dies,fixturesandskilledworkertoproduce high quality laboratory casework and equipment, and shall meet the following minimum requirements:
1. Fiveyearsormoreexperienceinmanufactureoflaboratory casework and equipment of type specified.
2. Teninstallationsofequalorlargersizeandrequirements.
C. Installer’squalifications:Factorycertifiedbythemanufacturer.
1.04 DELIVERY,STORAGEANDHANDLING
A. Scheduledeliveryofequipmentsothatspacesaresufficientlycompletethatequipmentcanbeinstalled immediately following delivery.
B. Protectfinishedsurfacesfromsoilingordamageduringhandlingandinstallation.Keepcoveredwithpolyethylenefilmorotherprotectivecoating.
C. Protectallworksurfacesthroughoutconstructionperiodwith1/4”corrugatedcardboardcompletelycovering the top and securely taped to edges. Mark cardboardinlargeletteringNoStanding.”
1.05 PROJECTCONDITIONS
A. Donotdeliverorinstallequipmentuntilthefollowing conditions have been met:
1. Windowsanddoorsareinstalledandthebuildingissecure and weather tight.
2. Plumbing,overheadductworkandlightingareinstalled.
3. Allpaintingiscompletedandfloortilelocatedbelowcasework is installed.
Fume Hood General Design Requirements
www.airmastersystems.com42
Fume Hood General Design Requirements (continued)
2.01 FUMEHOODMATERIALS
A. Steel:Highquality,coldrolled,mildsteelmeetingrequirementsofASTMA366;gaugesU.S.Standard.
B. StainlessSteel:Type304;gaugesU.S.Standard.
C. Ceilingclosurepanels:Minimum18gauge;finishtomatch hood exterior.
D. Bypassgrilles:Lowresistanttype,18gaugesteel,upward directional louvers.
E. Safetyglass:7/32”thicklaminatedsafetyglass.
F. Sashcables:7x7steel,coated,1/16”diametercoatedto3/32”.(Militaryspec.quality.)
G. Sashguides:Afulllengthextrudedcorrosionresistant polyvinyl chloride or powder coated steel withPVCguidestoprotectagainstmetaltometalcontact.
H.Pulleyassemblyforsashcable:2”diameter,nylontired, ball bearing type, with cable retaining device.
I. Sashpull:Fullwidth16gaugesteeltomatchhoodcolor.
J. Interioraccesspanels:Tobemadeofthesamematerial as the fume hood liner. Panels are to be heldinplacebymeansofanon-mechanicaltypefastening systems.
K. Fastenings: 1. Exteriorstructuralmembersattachments:Sheet
metalscrews,zincplated. 2. Interiorfasteningdevicesconcealed.Exposed
screws not acceptable. 3. Exteriorpanelmemberfasteningdevicestobe
corrosionresistantnon-metallicmaterial.Exposed screws not acceptable.
2.02 FUMEHOODCONSTRUCTION
A. Superstructure:Rigid,self-supportingassemblyofdoublewallconstruction,maximum5-1/4”thick.
1. Wallconsistsofasheetsteeloutershellandacorrosion resistant inner liner, and houses remoteoperatingservicefixturemechanismsand electrical services.
2. Accesstofixturevalvesconcealedinwallprovidedby exterior removable access panels, access
panels on the inside liner walls, or through removable front posts.
3. Hoodsmustbeoffullframeconstruction.Hoods that use metal brackets and spacers to hold interior and exterior panel in place are unacceptable.
B. Exhaustoutlet:10”round,20gaugestainlesssteelexhaust collars.
C. Accessopeningperimeter:Topandsidesoffaceopening to be radiuses or angled.
1. Bottomhorizontal:foilshallbeaflushmounttypeandprovidea1”bypasstoinsureacleansweepandtominimizeeddiesalongtheworksurfacewhen sash is in the closed position.
2. Bottomsashrail:1-1/2”framesection,16-gaugesteelorPVC.Provideflushpull,fullwidthofbottom rail.
3. Setsafetyglassintorailsindeepform,extrudedpolyvinylchlorideorneopreneglazingchannelsif a steel sash frame is being used.
4. Counterbalancesystem:Singleweight,pulley,cable, counter balance system which prevents sashtiltingbymeansofashaftdriven”systemandpermitsonefingeroperationatanypointalong full width pull. Sash not using this type of counter balance systems are unacceptable. Maximum9poundspullrequiredtoraiseorlower sash throughout its full length of travel. Designsystemtoholdsashatanypositionwithout creep and to prevent sash drop in the event of cable failure.
5. Openandclosesashagainstrubberbumperstops.
D. Fumehoodliner:1/4”Polyresin:Reinforcedpolyesterpanelsmoothfinishandwhitecolorinfinalappearance.Flexuralstrength:14,000psi.Flamespread:15orlessperU.L.723andASTME84-80.
E. Baffles:Fabricatefixed”bafflesprovidingcontrolledair vectors into and through the fume hood of the samematerialastheliner.Hoodswithadjustablebafflesareunacceptable.Allbafflesupportbracketstobenon-metallic.
F. Servicefixturesandfittings:Colorcodedhosenozzleoutlets and valves mounted inside the fume hood and controlled from the exterior with color coded indexhandles(whenspecified).
1. Valves:Needlepointtypewithself-centeringconetip and seat of hardened stainless steel.
2. Providepre-pipingforallservicefixturesfromvalvetocommonpointforfinalconnectionbyrespectivetrades.Copperforwater,airand
telephone 231-798-1111 43
Fume Hood General Design Requirements (continued)
vacuum and black iron for gas services. 3. Fixturesexposedtohoodinterior:Brasswith
chemically resistant powder coating. 4. Remotecontrolhandles:Prongtype,easytograsp. 5. Services:Asfollows:Gas,water,air,andvacuum.
G. Hoodlightfixtures:Twolamp,rapidstart,T-8ULlistedfluorescentlightfixturewithsoundratedballast installed on top panel.
1. Interioroffixture:White,highreflectingplasticenamel.
2. Provide:Removablesafetyglasspaneloninteriorofhoodforre-lampingoflights.
3. Sizeoffixture:Largestpossibleupto48”forhoods with superstructures up to six feet. Providetwo24”fixturesforhoodswitheightfoot superstructures.
4. Includelampswithfixtures. 5. Illumination:Perperformancevalues,part1ofthis
section. 6. Provide switch with black acid resistant
thermoplastic(whenspecified). 7. 3-wayswitchoneachsideofdoublesidedhoods
(whenspecified).
H. Electrical services: Provide on each front post of hoods. Three wire grounding type receptacles rated at120vGFI,20ampereswherespecified.FlushPlates: Black acid resistant thermoplastic.
I Worksurfaces:1-1/4”thickdishedanominal1/4”to contain spills.
1. MoldedresinworksurfacesforhoodswithwhiteResistoRocorPoly-resinliners.Frontraisededgenomorethan1/2”wide.
J. SafetyMonitor/AlarmSystem:ProvidesafetyMonitor/Alarmsystemthatmonitorsfacevelocityand provides audible and visual alarm if face velocity drops below safe levels. The technology usedintheTEL1000/1willbebasedonthermallycompensated thermistor based in the alarm module. Astheinternalfumehoodpressurechangesasthesash opening is closed and opened, the flow passing over the thermistor is calibrated to a face velocity that is displayed on the front of the monitor.
1. Safetymonitor:ULlisted,tamperproof,withallalarm circuits, electric components, external tubing, and manifolds furnished complete and factory installed. Monitor shall have light emitting diode display that provides clear indication of airflow conditions.
2. Calibrationistheresponsibilityoftheownerandis required once the hood is stationed and the hood exhausts and room supply systems are
balanced.Asecondarycalibrationhasbeenfactory set into the alarm’s memory only to determine that the alarm is functional and ready for shipment. The primary calibration must be completedinthefield.
3. Airflowsensor:Thermallycompensatedglassbeaded thermistor, factory connected to a sidewall port on the interior of the fume hood.
4. AlarmSignal:Audiblesignalandvisual,redlightemitting diode:
a. Silence pushbutton, which disables the audible alarm, shall be accessible on the front of the safety monitor.
b. Provide alternate mode in which visible alarm issilencedindefinitelybutvisualalarmremains activated until the alarm condition is corrected.
c. Whenalarmconditioniscorrectedandfacevelocityandvolumereturntospecifiedlevels,the Safety Monitor will automatically reset and begin routine monitoring.
d. Provide test circuit to verify proper Safety Monitor operation.
e. Electricalrating:Maximum12VDC,andmaximumcurrentratingof20OMA.
f. Provideaoptionforasashalarm/sensorifrequired.
2.03 CEILINGENCLOSURE:Provideceilingenclosurefromtopofhoodtoaccommodateaceilingheight(verify).Fabricateenclosurefrom18gaugesteeltomatchthehoodmaterialandfinish.
A. Preparation:Spraycleanmetalwithaheatedcleaner/phosphatesolution.
B. Application:Electrostaticallyapplypowdercoatof selected color and baked in controlled high temperature oven to assure a smooth, hard satin finish.Surfacesshallhaveachemicalresistant,highgradelaboratoryfurniturequalityfinishofthefollowing thickness:
1. Exteriorandinteriorsurfacesexposedtoview:1.5milaverageand1.2milminimum.
2.04 SOURCEQUALITYCONTROL
A. Demonstratefumehoodperformancebymeansofdocumentation of a third party testing company to theASHRAE110-1995methodsoftesting.
www.airmastersystems.com44
Airfoil: Shaped or streamlined member at hood entrance designed to enhance movement of air into the hood
Air Volume:Rateofairflow,normallyexpressedincubicfeetperminute(CFM)
ASHRAE:AmericanSocietyofHeating,Refrigerating,andAirConditioningEngineers,aprofessionalorganizationthat sets industry standards for fume hood testing procedures
Auxiliary Air: Supply or makeup air delivered external to the chamber of a fume hood to reduce air consumption
Baffle: Panels located across back of hood interior, which control pattern of air moving through the hood
Blower:Airmovingdevice(orfan)consistingofmotor,impeller, and scroll
Bypass:Compensatingopeningthathelpsmaintainconstant volume exhaust from fume hood, regardless of sash position
Canopy Hood:Ceilingorwallsuspendedventilatingdevice for noncritical use with heat, water vapor, odors, etc.
CFM:CubicFeetPerMinute,aunitofmeasurementofair volume
Combination Sash:Horizontalpanelsinaverticallyrisingframe; see sash
Constant Volume: Type of fume hood exhaust system that exhausts the same volume of air, regardless of sash position
Containment:Extenttowhichfumesareconfinedwithinthe hood compartment
Damper:Deviceinstalledinducttocontrolairvolume
Demonstration Hood: Fume hood with glass panels on two or three sides to improve visibility for demonstrating experiments in a classroom setting
Exhaust Volume/Parameters:Quantityofairexhaustedbythe fume hood; quantity of air required to maintain desired face velocity, expressed in cubic feet per minute(CFM)
Face Velocity: Speed of air moving into the fume hood throughthefaceopening(throughthesash),measuredinfeetperminute(FPM)
FPM: Feet per minute; measurement of air velocity
Liner: Fume hood interior sides, back, and top, including baffle
Lintel: Portion of fume hood front located above access opening
Louvers:Slit-likeopeningsinthelintelthatallowbypassair to enter the hood when the sash is closed
NFPA:NationalFireProtectionAssociation
Negative Pressure: Pressures lower than one atmosphere
Positive Pressure: Pressures higher than one atmosphere
Restricted Bypass Fume Hood: Fume hood operating type, designed with limited bypass area; commonly used in conjunctionwithVariableAirVolume(VAV)exhaustsystems and restricted sash opening designs
Sash: Sliding glass panel set in the fume hood face that provides access to the hood interior
Service Fitting/Plumbing:Waterfaucetsandgasvalvesmounted on or fastened to the fume hood
Static Pressure:Airpressure,orresistance,infumehoodor duct, expressed in inches of water
U.L. 1805:UnderwritersLaboratoriescertificationthatverifiesconformancetoelectrical,mechanical,andairflow standards
Variable Air Volume (VAV): Type of fume hood exhaust system that typically maintains constant fume hood facevelocitybyadjustingblowermotorspeedora balance damper in response to changes in sash position
Velocity: Speed of air, measured in feet per minute (FPM)
Velocity Pressure: Force per square inch applied by moving air
Volume:Quantityofair,usuallymeasuredincubicfeetperminute(CFM)
Work Surface: Top material; area in fume hood where apparatus rests and where work takes place
Fume Hood Technology — Glossary of Terms