agreement and repeatability of central corneal thickness measurements made by ultrasound pachymetry...

5
Original article Agreement and repeatability of central corneal thickness measurements made by ultrasound pachymetry and anterior segment optical coherence tomography Chao-Wen Lin, Tsing-Hong Wang, Yu-Hsuan Huang, Jehn-Yu Huang * Department of Ophthalmology, National Taiwan University Hospital, Taipei, Taiwan article info Article history: Received 12 December 2012 Received in revised form 29 March 2013 Accepted 30 April 2013 Available online 10 June 2013 Keywords: agreement anterior segment optical coherence tomography central corneal thickness repeatability ultrasound pachymetry abstract Purpose: To evaluate repeatability within and between examiners and to assess agreement in corneal thickness measurements obtained by anterior segment optical coherence tomography (ASOCT) and ul- trasonic pachymetry (USP). Methods: This was a prospective, observational study. Fifty-one patients who visited a glaucoma service with suspected glaucoma, glaucoma, or cataract were enrolled. Patients with a history of corneal surgery or abnormalities and those who were uncooperative for examinations were excluded. Each patient underwent ASOCT followed by USP, with each test performed by two different examiners. Measurement repeatability was evaluated using intraclass correlation coefcient (ICC) values. Agreement in corneal thickness measurements was evaluated by the Bland and Altman plot method. Linear regression analysis was used to assess the relationship between ASOCT and USP measurements. Results: USP and ASOCT corneal thickness measurements revealed high intraexaminer and interexaminer repeatability (ICCs 0.978). ASOCT corneal thickness measurements of a central 2-mm zone showed higher intraexaminer (ICC ¼ 0.999) and interexaminer (ICC ¼ 0.999) repeatability than USP measure- ments or ASOCT measurements of the vertex. The 95% limit of agreement (LoA) between the vertex- centered ASOCT and the USP scan was between 3.68 and 24.76 mm. The 95% LoA between the central 2-mm zone ASOCT and the USP scan was between 3.75 and 23.39 mm. Average ASOCT corneal thickness was less than USP measurements, but ASOCT thickness could be converted to USP thickness through linear regression equations. Conclusion: Both devices have good intraexaminer and interexaminer repeatability, though ASOCT has slightly better interexaminer repeatability. ASOCT accurately and reliably measures corneal thickness in a noninvasive manner. Copyright Ó 2013, The Ophthalmologic Society of Taiwan. Published by Elsevier Taiwan LLC. All rights reserved. 1. Introduction Central corneal thickness (CCT) is an important measurement in the clinical management of glaucoma. Many studies have reported that high or low corneal thickness can lead to overestimation or underestimation of intraocular pressure (IOP). 1e3 Although adjusting IOP for CCT is not necessary, CCT can help ophthalmolo- gists properly perform risk assessments for the development of glaucoma. 4 The results of the Ocular Hypertension Treatment Study (OHTS) suggest that patients with thinner corneas may be at a greater risk of developing glaucoma. 4 Central corneal thickness is an important evaluation before laser refractive surgery and is a predictive factor of postoperative kera- toectasia. 5 Overestimating corneal thickness may lead to excessive ablation and potentially result in iatrogenic keratoectasia. Knowl- edge of the corneal thickness is also critical in avoiding endothelial injury during collagen cross-linking treatment for progressive keratoconus. 6 All CCT data in the above-mentioned studies were obtained with ultrasound pachymetry (USP). 4e6 This is currently the most frequently used clinical technique and the gold standard to evaluate CCT. Ultrasound pachymetry has some advantages, including its ease of use, portability, and low cost. However, the accuracy of this method is heavily examiner dependent. Any misalignment or variation in probe placement can introduce measurement error. Furthermore, USP probes must be placed directly onto the cornea, which introduces risk of corneal abrasion and infection transmission. 7,8 Because of these * Corresponding author. Department of Ophthalmology, National Taiwan Uni- versity Hospital, Number 7, Zhongshan South Road, Taipei 100, Taiwan. E-mail address: [email protected] (J.-Y. Huang). Contents lists available at SciVerse ScienceDirect Taiwan Journal of Ophthalmology journal homepage: www.e-tjo.com 2211-5056/$ e see front matter Copyright Ó 2013, The Ophthalmologic Society of Taiwan. Published by Elsevier Taiwan LLC. All rights reserved. http://dx.doi.org/10.1016/j.tjo.2013.04.007 Taiwan Journal of Ophthalmology 3 (2013) 98e102

Upload: jehn-yu

Post on 01-Jan-2017

213 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Agreement and repeatability of central corneal thickness measurements made by ultrasound pachymetry and anterior segment optical coherence tomography

at SciVerse ScienceDirect

Taiwan Journal of Ophthalmology 3 (2013) 98e102

Contents lists available

Taiwan Journal of Ophthalmology

journal homepage: www.e-t jo.com

Original article

Agreement and repeatability of central corneal thicknessmeasurements made by ultrasound pachymetry and anteriorsegment optical coherence tomography

Chao-Wen Lin, Tsing-Hong Wang, Yu-Hsuan Huang, Jehn-Yu Huang*

Department of Ophthalmology, National Taiwan University Hospital, Taipei, Taiwan

a r t i c l e i n f o

Article history:Received 12 December 2012Received in revised form29 March 2013Accepted 30 April 2013Available online 10 June 2013

Keywords:agreementanterior segment optical coherencetomography

central corneal thicknessrepeatabilityultrasound pachymetry

* Corresponding author. Department of Ophthalmversity Hospital, Number 7, Zhongshan South Road, T

E-mail address: [email protected] (J.-Y. Hu

2211-5056/$ e see front matter Copyright � 2013, Thhttp://dx.doi.org/10.1016/j.tjo.2013.04.007

a b s t r a c t

Purpose: To evaluate repeatability within and between examiners and to assess agreement in cornealthickness measurements obtained by anterior segment optical coherence tomography (ASOCT) and ul-trasonic pachymetry (USP).Methods: This was a prospective, observational study. Fifty-one patients who visited a glaucoma servicewith suspected glaucoma, glaucoma, or cataract were enrolled. Patients with a history of corneal surgeryor abnormalities and those who were uncooperative for examinations were excluded. Each patientunderwent ASOCT followed by USP, with each test performed by two different examiners. Measurementrepeatability was evaluated using intraclass correlation coefficient (ICC) values. Agreement in cornealthickness measurements was evaluated by the Bland and Altman plot method. Linear regression analysiswas used to assess the relationship between ASOCT and USP measurements.Results: USP and ASOCT corneal thickness measurements revealed high intraexaminer and interexaminerrepeatability (ICCs � 0.978). ASOCT corneal thickness measurements of a central 2-mm zone showedhigher intraexaminer (ICC ¼ 0.999) and interexaminer (ICC ¼ 0.999) repeatability than USP measure-ments or ASOCT measurements of the vertex. The 95% limit of agreement (LoA) between the vertex-centered ASOCT and the USP scan was between 3.68 and 24.76 mm. The 95% LoA between the central2-mm zone ASOCT and the USP scan was between 3.75 and 23.39 mm. Average ASOCT corneal thicknesswas less than USP measurements, but ASOCT thickness could be converted to USP thickness throughlinear regression equations.Conclusion: Both devices have good intraexaminer and interexaminer repeatability, though ASOCT hasslightly better interexaminer repeatability. ASOCT accurately and reliably measures corneal thickness in anoninvasive manner.Copyright � 2013, The Ophthalmologic Society of Taiwan. Published by Elsevier Taiwan LLC. All rights

reserved.

1. Introduction

Central corneal thickness (CCT) is an important measurement inthe clinical management of glaucoma. Many studies have reportedthat high or low corneal thickness can lead to overestimation orunderestimation of intraocular pressure (IOP).1e3 Althoughadjusting IOP for CCT is not necessary, CCT can help ophthalmolo-gists properly perform risk assessments for the development ofglaucoma.4 The results of the Ocular Hypertension Treatment Study(OHTS) suggest that patients with thinner corneas may be at agreater risk of developing glaucoma.4

ology, National Taiwan Uni-aipei 100, Taiwan.ang).

e Ophthalmologic Society of Taiw

Central corneal thickness is an important evaluation before laserrefractive surgery and is a predictive factor of postoperative kera-toectasia.5 Overestimating corneal thickness may lead to excessiveablation and potentially result in iatrogenic keratoectasia. Knowl-edge of the corneal thickness is also critical in avoiding endothelialinjury during collagen cross-linking treatment for progressivekeratoconus.6

All CCT data in the above-mentioned studies were obtained withultrasound pachymetry (USP).4e6 This is currently the mostfrequently used clinical technique and the gold standard to evaluateCCT. Ultrasound pachymetry has some advantages, including its easeof use, portability, and lowcost. However, the accuracy of thismethodis heavily examiner dependent. Any misalignment or variation inprobe placement can introduce measurement error. Furthermore,USPprobesmustbeplaceddirectlyonto the cornea,which introducesriskof cornealabrasionand infection transmission.7,8Becauseof these

an. Published by Elsevier Taiwan LLC. All rights reserved.

Page 2: Agreement and repeatability of central corneal thickness measurements made by ultrasound pachymetry and anterior segment optical coherence tomography

Fig. 1. (A) Pachymetry map obtained with anterior segment optical coherence to-mography (ASOCT). Average corneal thickness of each region is shown. (B) The ASOCTreport also provides information on corneal thickness at the corneal vertex.

C.-W. Lin et al. / Taiwan Journal of Ophthalmology 3 (2013) 98e102 99

drawbacks, devices and methods for noninvasive measurement ofCCT, suchas theOrbscan (Bausch&Lomb Inc., Rochester,NY,USA), thePentacam (Oculus Optikgerate GmbH, Wetzler, Germany), and ante-rior segment optical coherence tomography (ASOCT), were invented.Anterior segment OCT evaluates anterior segment structure andmeasures corneal thickness in a noncontact manner. According toprevious studies, CCT readings obtained by ASOCT were significantlylower than those obtained by USP.9e14

In this study, we used Fourier-domain OCT and USP to measurecorneal thickness. The agreement both between and within mea-surement modalities was examined and the possibility of con-verting results from one imaging modality to the other throughlinear regression analysis was explored.

2. Materials and methods

2.1. Patients

This prospective, cross-sectional, observational study adhered tothe tenets of the Declaration of Helsinki and was reviewed andapproved by the institutional review board of the National TaiwanUniversityHospital. Informed consentwas obtained fromall patientsbefore any study procedure or examination was performed. Thestudy population was composed of patients who were seen at theglaucoma service of National Taiwan University Hospital due tosuspected glaucoma (n¼ 23), glaucoma (n¼ 16), or cataract (n¼ 12).Patients were recruited randomly between August 2009 and June2010. Patientswhohadahistoryof corneal surface surgeryor cornealabnormalities and those who were uncooperative were excludedfrom participation. Study measurements were made in a total of 51eyes from 51 volunteers (20 men, 31 women) who enrolled in thisstudy. Mean patient age was 56.5 � 14.3 years (range: 21e89 years).

2.2. Sample size estimation

Based on the results of a previous study,15 we calculated thenecessary sample size to be 34 eyes with nQuery advisor (version4.0; Statistical Solutions, Boston, MA, USA). A sample size of 34 had80% power to detect a difference in mean corneal thickness of 5 mm.We assumed a standard deviation of 10 mm and that comparisonsbetweenmeans would bemadewith a two-tailed paired t-test witha 0.05 significance level.

2.3. Central corneal thickness measurements

The central corneal thickness (CCT) for each patient’s right eyewas measured using the ASOCT (RTVueModel RT100; Optovue Inc.,Fremont, CA, USA) first and then USP (PacScan 330P; Sonomed Inc.,Lake Success, NY, USA). For intraexaminer and interexaminerrepeatability testing, the first examiner performed two consecutivemeasurements. Then, the second examiner performed twoconsecutive measurements with the same instrument. Both ex-aminers completed the ASOCT measurement before the USP mea-surement. Between the two ASOCT measurements, patients weregiven the opportunity to sit back and leave the chin rest. Betweenthe two USP measurements, the probe was moved away from thecorneal surface so that two independent readings could be made.

TheRTVueOCT is a Fourier-domainOCT thatuses an840-nmsuperilluminated diode as its scanning source. The transverse resolution is15mm,and theaxial resolution is5mm.Whencorneal-anteriormodule(CAM) adapters are installed on the RTVue OCT, images of the corneaand anterior chamber can be obtained. The two types of CAMadaptersare the CAM-L and the CAM-S. The CAM-L was selected for use in thisstudy because it allows a corneal pachymetrymap to be obtained. Forthis study, an OCTcorneal pachymetrymapwas obtained by selecting

the “Pachymetry þ CPwr (corneal power)” and aiming the measure-ment beam at the pupil center. The built-in software (version 3.5)automatically generated a pachymetry map (Fig. 1A) and dividedit into the central (0e2 mm), pericentral (2e5 mm), and transitional(5e6 mm) zones. The two outer zones were further subdivided intoeight sectors each by radial lines. Average corneal thickness for thecentral zoneandeachsubzone isprovidedontheASOCTreport, aswellas the thickness of the corneal vertex (Fig. 1B). Both the central zonecorneal thickness (CCTc) and the corneal vertex thickness (CCTv)werecompared to USP central corneal thickness (CCTu) measurements.

After ASOCT measurements were completed, the cornea wasanesthetized with one drop of 0.5% proparacaine (Alcaine, AlconInc., Fort Worth, TX, USA), followed by 10-second eyelid closure. Toavoid a possible increase in corneal thickness secondary to topicalanesthesia, USP measurements were started after 90 seconds ofnormal blinking. The ultrasound probe was placed as close to thecenter of the cornea as possible, perpendicular to the surface, aspatients gazed at a distant target. The measurements were usedonly if the probe was within 10� of perpendicular. Using the defaultsettings of the PacScan 330P USP, five consecutive measurementswere averaged to obtain a single value for CCT.

2.4. Statistical analysis

For intraexaminer repeatability, we compared two measure-ments of the samemethod by the same examiner. For interexaminer

Page 3: Agreement and repeatability of central corneal thickness measurements made by ultrasound pachymetry and anterior segment optical coherence tomography

Table 2Interexaminer repeatability of optical coherence tomography and ultrasoundpachymetry measurements of corneal thickness.

Examiner 1 mean (SD) Examiner 2 mean (SD) ICC

CCTv (mm) 540.22 (30.00) 541.24 (29.70) 0.998a

CCTc (mm) 540.86 (29.81) 540.04 (29.54) 0.999a

CCTu (mm) 554.43 (30.42) 551.39 (29.85) 0.980a

ap < 0.001.CCTc ¼ central 2-mm zone corneal thickness measured by ASOCT; CCTu ¼ centralcorneal thickness measured by ultrasound pachymetry; CCTv ¼ vertex cornealthickness measured by anterior segment optical coherence tomography (ASOCT);ICC ¼ intraclass correlation coefficient; SD ¼ standard deviation.

C.-W. Lin et al. / Taiwan Journal of Ophthalmology 3 (2013) 98e102100

repeatability, we compared the first measurement made by eachexaminer. The repeatability was evaluated by the intraclass corre-lation coefficient (ICC) value, which was interpreted as follows: 0e0.2 indicated poor agreement, 0.3e0.4 indicated fair agreement,0.5e0.6 indicated moderate agreement, 0.7e0.8 indicated strongagreement, and >0.8 indicated almost perfect agreement.

Agreement between ASOCT and USP measurements was eval-uated by the Bland-Altman method. Ninety-five percent limits ofagreement (LoA) were defined as the mean � 1.96 standard de-viations (SDs). Difference between ASOCT and USP measurementswas evaluated by theWilcoxon signed ranked test. The relationshipbetween ASOCT and USP corneal thickness measurements wasdetermined using linear regression analyses, where USP measure-ments were compared individually to ASOCT vertex scans andcentral 2-mm zone results.

3. Results

3.1. Intraexaminer repeatability of corneal thickness measurements

The ICCs for the measurement of CCTv, CCTc, and CCTu by thefirst examiner were 0.998, 0.999, and 0.995, respectively. The ICCsfor the measurement of CCTv, CCTc, and CCTu by the secondexaminer were 0.996, 0.999, and 0.978, respectively (Table 1). BothASOCT and USP showed good intraexaminer repeatability. Inaddition, ICC for corneal thickness measurement by ASOCT wassimilar between the first and second examiners. However, USP ICCsfor corneal thickness measurements were lower for both the first(0.995) and the second examiner (0.978) than for ASOCTmeasurements.

3.2. Interexaminer repeatability of corneal thickness measurements

We also analyzed interexaminer repeatability of corneal thick-ness measurements made by the ASOCT and USP by comparing thefirst measurement made by each examiner. Inter-examinerrepeatability of both devices was acceptable for clinical practiceand the ICCs for CCTv, CCTc, and CCTu were 0.998, 0.999, and 0.980,respectively. The CCTc measurement showed the highest inter-examiner repeatability and the CCTu measurement showed thelowest interexaminer repeatability (Table 2).

3.3. Agreement between optical coherence tomography andultrasound measurements

Mean corneal thickness and the 95% LoA for both ASOCT andUSP measurements are summarized in Table 3. Average ASOCTcorneal thickness was statistically less than USP measurements forboth vertex and central OCTmeasurements (Table 3). Differences inindividual patient ASOCT and USP corneal thickness measurementsare shown in the Bland-Altman plots of Figs. 2 and 3. Vertex CCTwas thinner than CCTu in all but one case and CCTc was always

Table 1Intraexaminer repeatability of optical coherence tomography and ultrasound pachymetr

Examiner 1

1st measurement(mean [SD])

2nd measurement(mean [SD])

ICC

CCTv (mm) 540.22 (30.00) 541.35 (29.53) 0.99CCTc (mm) 540.86 (29.81) 540.90 (30.09) 0.99CCTu (mm) 554.43 (30.42) 555.18 (30.49) 0.99

ap < 0.001.CCTc ¼ central 2-mm zone corneal thickness measured by ASOCT; CCTu ¼ central cornemeasured by the anterior segment optical coherence tomography (ASOCT); ICC ¼ intrac

thinner than CCTu. The mean difference between CCTu and CCTvmeasurements was 14.22 � 5.38 mm (95% LoA ¼ 3.68e24.76 mm,ICC¼ 0.984, p < 0.001) and the mean difference between CCTu andCCTc measurements was 13.57 � 5.01 mm (95% LoA ¼ 3.75e23.39 mm, ICC ¼ 0.986, p < 0.001).

3.4. Relationship between optical coherence tomography andultrasound measurements

There was a strong linear relationship between OCT and USPmeasurements of corneal thickness (Figs. 4 and 5). Using linearregression analyses, we found the equations of the best-fit lines toconvert OCT results to USP results. These equations are as follows(all CCT measurements in mm):

CCTu ¼ 7:360þ 1:011� CCTv�adjusted r2 ¼ 0:974

�(1)

CCTu ¼ 10:124þ 1:006� CCTv�adjusted r2 ¼ 0:973

�(2)

4. Discussion

Central corneal thickness is an important parameter in deter-mining whether or not patients are candidates for corneal refrac-tive surgery and in assessing a patient’s risk when glaucoma, or thesuspicion of glaucoma, is present. Adjustment of IOPmeasurementsfor CCT may lead to decision changes regarding medical therapy,laser treatment, and surgical intervention in 8.5%, 2.1%, and 3.2% ofpatients, respectively.16 Therefore, an accurate and precise CCTmeasurement is helpful for physicians in the clinical managementof their patients.

Currently, USP is the standard device for CCT measurement, butwith the introduction of ASOCT, ophthalmologists have an alter-native, noncontact method tomeasure CCT. Measuring CCTwithouttouching the cornea reduces the risk of iatrogenic corneal injuryand infection transmission. Additionally, we have shown thatASOCT is less examiner-dependent than USP.

In agreement with previous studies,10e12 our ASOCT CCT mea-surements were lower than USP measurements, with differences

y measurements of corneal thickness.

Examiner 2

1st measurement(mean [SD])

2nd measurement(mean [SD])

ICC

8a 541.24 (29.70) 540.76 (29.78) 0.996a

9a 540.04 (29.54) 540.39 (29.55) 0.999a

5a 551.39 (29.85) 555.55 (30.06) 0.978a

al thickness measured by ultrasound pachymetry; CCTv ¼ corneal vertex thicknesslass correlation coefficient; SD ¼ standard deviation.

Page 4: Agreement and repeatability of central corneal thickness measurements made by ultrasound pachymetry and anterior segment optical coherence tomography

Table 3Interdevice agreement of optical coherence tomography and ultrasound pachymetry measurements of corneal thickness.

CCTu (mm)(mean � SD)

CCTv (mm)(mean � SD)

pb CCTu e CCTv (mm)(mean � SD)

95% of LoA ICC

554.43 � 30.42 540.22 � 30.00 <0.001 14.22 � 5.38 3.68 w 24.76 0.984a

CCTu (mm)(mean � SD)

CCTc (mm)(mean � SD)

pb CCTu e CCTc (mm)(mean � SD)

95% of LoA ICC

554.43 � 30.42 540.86 � 29.81 <0.001 13.57 � 5.01 3.75 w 23.39 0.986a

CCTc ¼ central 2-mm zone corneal thickness measured by ASOCT; CCTu ¼ central corneal thickness measured by ultrasound pachymetry; CCTv ¼ vertex corneal thicknessmeasured by anterior segment optical coherence tomography (ASOCT); ICC ¼ intraclass correlation coefficient; LoA ¼ limit of agreement; SD ¼ standard deviation.

a Indicates p < 0.001.b Differences between CCTu and CCTv, CCTu and CCTc were evaluated by the Wilcoxon signed-rank test.

C.-W. Lin et al. / Taiwan Journal of Ophthalmology 3 (2013) 98e102 101

ranging between 6.4 and 49.4 mm. Possible explanations for thesedifferences include topical anesthesia-induced corneal edema anduncertain patient-to-patient variations in the speed of soundthrough the cornea, for USP measurements, and tissue index ofrefraction, for ASOCT measurements.10

Good measurement repeatability depends on short data acqui-sition time, consistent positioning between scans, and minimalcorneal thickness variation along neighboring points.

Based on our results, both intra- and inter-examiner repeat-ability were higher for OCT than for USP measurements. This in-dicates that OCT may be more suitable for repeated CCTmeasurements over time. The procedure of measuring CCT withASOCT is less examiner-dependent than the process of obtainingmeasurements using USP. While taking OCT measurements, a real-time corneal cross-sectional image is visible during the examina-tion, and examiners can easily adjust the scanning position toensure that the measurement is being made perpendicular to thecorneal surface. If the location of the corneal vertex does not changeover time, this method could serve as a location registration tomake serial measurements at the same location possible. Theability to quickly acquire corneal thickness data is another advan-tage of ASOCT over USP. The RTVue OCT has an acquisition speed of0.32 seconds, which is much faster than the minutes it takes toobtain USP measurements. Rapid data acquisition is essential inminimizing both patient and machine motion artifacts, and thehigher axial resolution of the RTVue OCT may contribute to betterrepeatability.

The intraexaminer and interexaminer repeatability with theASOCT central 2-mm zone method (CCTc) was slightly better than

Fig. 2. BlandeAltman plots comparing ultrasound pachymetry and anterior segmentoptical coherence tomography (vertex-scan) corneal thickness measurements.

the vertex method (CCTv). The central 2-mm zone measurement isthe result of averaging corneal thickness within the central 2-mmzone. Because it is not a single point measurement, errors due tomisalignment areminimized, but CCTc does not provide the cornealthickness of a single central point.

The range of 95% LoA between CCTc and CCTu was narrowerthan the one between CCTv and CCTu. According to a previousstudy,17 the 95% LoA between RTVue OCT and USPmeasurements isbroad, and the corneal thickness measurements cannot be usedinterchangeably in either the clinical or research setting. However,our study demonstrated that reliable USP and ASOCT measure-ments are linearly related to each other and that it is possible toconvert between the two types of CCT measurements. Linearregression analyses showed that adjusted R-squared values of bothregression equations (CCTv vs. USP, CCTc vs. USP) were greater than0.97. Therefore, it is possible to use ASOCT, instead of USP, tomeasure corneal thickness in healthy individuals. This may even bebetter because of the higher intraexaminer and interexaminerrepeatability, and the noncontact nature, of ASOCT.

There were some limitations to our study. First, the sample sizewas relatively small and more corneal thickness measurements areneeded in a larger group of patients to validate our conversionequations. Second, all patients enrolled in this study had normalcorneas, so it is unknown how ocular and/or corneal alterations/pathologies would have affected our results. Further studiesincluding patients who have undergone refractive surgery (i.e.,LASIK) or who have corneal abnormalities (i.e., keratoconus,corneal opacity) are needed because changes in corneal curvatureand/or structure may lead to different results.

Fig. 3. BlandeAltman plots comparing ultrasound pachymetry and anterior segmentoptical coherence tomography (central 2-mm zone) corneal thickness measurements.

Page 5: Agreement and repeatability of central corneal thickness measurements made by ultrasound pachymetry and anterior segment optical coherence tomography

Fig. 4. Relationship between ultrasound pachymetry and anterior segment opticalcoherence tomography (vertex-scan) corneal thickness measurements. Equation:CCTu ¼ 7.360 þ 1.011 � CCTv (adjusted r2 ¼ 0.974).

Fig. 5. Relationship between ultrasound pachymetry and anterior segment opticalcoherence tomography (central 2-mm zone) corneal thickness measurements. Equa-tion: CCTu ¼ 10.124 þ 1.006 � CCTc (adjusted r2 ¼ 0.973).

C.-W. Lin et al. / Taiwan Journal of Ophthalmology 3 (2013) 98e102102

Corneal thickness as measured by ASOCT was consistentlythinner than corneal thickness as measured by USP. Both mea-surement modalities had good intraexaminer and interexaminerrepeatability, but ASOCT was slightly better. Using linear regressionanalysis to find the equation of the best-fit line, it was possible toconvert between ASOCT and USP measurements. Therefore, ASOCTis a useful and precise technology that can reliably measure cornealthickness in a noncontact manner.

Acknowledgments

The authors indicated nofinancial support or conflicts of interest.

References

1. Johnson M, Kass MA, Moses RA, Grodzki WJ. Increased corneal thicknesssimulating elevated intraocular pressure. Arch Ophthalmol 1978;96:664e5.

2. Whitacre MM, Stein RA, Hassanein K. The effect of corneal thickness onapplanation tonometry. Am J Ophthalmol 1993;115:592e6.

3. Stodtmeister R. Applanation tonometry and correction according to cornealthickness. Acta Ophthalmol Scand 1998;76:319e24.

4. Gordon MO, Beiser JA, Brandt JD, Heuer DK, Higginbotham EJ, Johnson CA, et al.The ocular hypertension treatment study: baseline factors that predict theonset of primary open-angle glaucoma. Arch Ophthalmol 2002;120:714e20.

5. Randleman JB, Woodward M, Lynn MJ, Stulting RD. Risk assessment for ectasiaafter corneal refractive surgery. Ophthalmology 2008;115:37e50.

6. Ashwin PT, McDonnell PJ. Collagen cross-linkage: a comprehensive review anddirections for future research. Br J Ophthalmol 2010;94:965e70.

7. Hashemi H, Roshani M, Mehravaran S, Parsafar H, Yazdani K. Effect of cornealthickness on the agreement between ultrasound and Orbscan II pachymetry.J Cataract Refract Surg 2007;33:1694e700.

8. Drexler W, Baumgartner A, Findl O, Hitzenberger CK, Sattmann H, Fercher AF.Submicrometer precision biometry of the anterior segment of the human eye.Invest Ophthalmol Vis Sci 1997;38:1304e13.

9. BechmannM,ThielMJ,NeubauerAS,Ullrich S, LudwigK,KenyonKR, et al. Centralcorneal thickness measurement with a retinal optical coherence tomographydevice versus standard ultrasonic pachymetry. Cornea 2001;20:50e4.

10. Kim HY, Budenz DL, Lee PS, Feuer WJ, Barton K. Comparison of central cornealthickness using anterior segment optical coherence tomography vs ultrasoundpachymetry. Am J Ophthalmol 2008;145:228e32.

11. Li H, Leung CK, Wong L, Cheung CY, Pang CP, Weinreb RN, et al. Comparativestudy of central corneal thickness measurement with slit-lamp optical coher-ence tomography and visante optical coherence tomography. Ophthalmology2008;115:796e801.

12. Li Y, Shekhar R, Huang D. Corneal pachymetry mapping with high-speed op-tical coherence tomography. Ophthalmology 2006;113:792e9.

13. Wong AC, Wong CC, Yuen NS, Hui SP. Correlational study of central cornealthickness measurements on Hong Kong Chinese using optical coherence to-mography, Orbscan and ultrasound pachymetry. Eye 2002;16:715e21.

14. Prospero Ponce CM, Rocha KM, Smith SD, Krueger RR. Central and peripheralcorneal thickness measured with optical coherence tomography, Scheimpflugimaging, and ultrasound pachymetry in normal, keratoconus-suspect, andpost-laser in situ keratomileusis eyes. J Cataract Refract Surg 2009;35:1055e62.

15. Huang JY, Pekmezci M, Yaplee S, Lin S. Intra-examiner repeatability andagreement of corneal pachymetry map measurement by time-domain andFourier-domain optical coherence tomography. Graefes Arch Clin Exp Oph-thalmol 2010;248:1647e56.

16. Shih CY, Graff Zivin JS, Trokel SL, Tsai JC. Clinical significance of central cornealthickness in the management of glaucoma. Arch Ophthalmol 2004;122:1270e5.

17. Chen S, Huang J, Wen D, Chen W, Huang D, Wang Q. Measurement of centralcorneal thickness by high-resolution Scheimpflug imaging, Fourier domainoptical coherence tomography and ultrasound pachymetry. Acta Ophthalmol2012;90:449e55.