Also Available: AgraStrip ® Allergens AgraQuant ® Plus Allergens AgraQuant ® Allergens AgraStrip ® Gluten G12 Quick Guide AgraStrip ® Gluten G12 Test Kit includes: 10 test strips 1 dropper bottle: extraction buffer 1 dropper bottle: dilution buffer 10 extraction tubes & caps 10 extraction tube dropper tips 10 dilution tubes & caps 10 swabs 1 stand for incubation Shelf life: 12 months This test kit’s performance was reviewed by AOAC Research Institute and was found to perform to the manufacturer’s specifications. The following matrices have been validated according to AOAC Performance Tested Methods (PTM) protocols: rice flour, bread, cookies, ice cream and dark chocolate. Stainless steel has been validated per AOAC PTM protocol for environmental surface swab testing. * The rinse water test method has been validated internally by Romer Labs ® Test procedures for: • Raw materials/processed food • Rinse waters* • Swab samples

Upload: others

Post on 19-Jul-2020

10 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: AgraStrip Gluten G12 Quick Guide - keydiagnostics.com.au · QG_AS_G12_EN_1215_JKR AgraStrip® Allergen Quick Guide Fill level Homogenise the sample and weigh 0.2 g of sam-ple and

Also Available:• AgraStrip® Allergens• AgraQuant® Plus Allergens• AgraQuant® Allergens

AgraStrip® Gluten G12 Quick Guide

AgraStrip® Gluten G12 Test Kit includes:

• 10 test strips• 1 dropper bottle: extraction buffer• 1 dropper bottle: dilution buffer• 10 extraction tubes & caps• 10 extraction tube dropper tips• 10 dilution tubes & caps• 10 swabs• 1 stand for incubation

Shelf life: 12 months

This test kit’s performance was reviewed by AOAC Research Institute and was found to perform to the manufacturer’s specifications. The following matrices have been validated according to AOAC Performance Tested Methods (PTM) protocols: rice flour, bread, cookies, ice cream and dark chocolate. Stainless steel has been validated per AOAC PTM protocol for environmental surface swab testing.

* The rinse water test method has been validated internally by Romer Labs®

Test procedures for:• Raw materials/processed food• Rinse waters*• Swab samples

User
Key 2017
Page 2: AgraStrip Gluten G12 Quick Guide - keydiagnostics.com.au · QG_AS_G12_EN_1215_JKR AgraStrip® Allergen Quick Guide Fill level Homogenise the sample and weigh 0.2 g of sam-ple and

20

105

Glu

ten

Ag

ra S

trip

® G

lute

n A

gra

Str

ip®

Glu

ten

Ag

ra S

trip

® G

lute

n A

gra

Str

ip®

Glu

ten

Ag

ra

20

105

Transfer Dilution Buffer (red label) into the dilution tube already filled with extract to one of the lines demarcating the test cut off levels (5/10/20 ppm Gluten). Close the dilution tube with the cap and shake vigorously by hand for 15 seconds.

Transfer 3 drops (100 μl) of the sample supernatant into the dilution tube by squeezing the extraction tube.

After the liquid has soaked up to the “flow” level remove the test strip from the dilution tube and place it upright (arrows pointing down) into a slot of the Tube Holder and allow to develop for 10 minutes and then read off the result immediately.

Take off cap and place a test strip vertically (arrows pointing down) into the dilution tube and allow liquid to flow up the strip to the „flow“ level line. (this takes about 45 seconds).

Glu

ten

Ag

ra S

trip

® G

lute

n A

gra

Str

ip®

Glu

ten

Ag

ra S

trip

® G

lute

n A

gra

Str

ip®

Glu

ten

Ag

ra

Result Interpretation

Negative Result (N):

One single blue line in the control zone (c) of the strip

Positive Result (P):

One red line in the test zone (t) and one blue line in the control zone (c). Even a faint line in the test zone implies a positive result(see strip at bottom). The sample contains the specific allergen higher than the cutoff level (which needs to be validated for individual matrices).

Nc t

P

PFaint test line:

QG

_AS_G

12_E

N_1

215_

JKR

AgraStrip® Allergen Quick Guide

Filllevel

Homogenise the sample and weigh 0.2 g of sam-ple and add to the ex-traction tube. (If balance is unavailable 0.2 g can be estimated by filling the extraction tube cap). Fill tube with extraction solution to level shown in picture.

Close tube using the ex-traction tube cap, and shake vigorously by hand for 1 minute. Remove extraction tube cap and replace with dropper tip.

1 2

Fill extraction tube with Extraction Buffer up to where the tube gets thicker, wet the swab by dipping into the buffer. Wipe an area of 5 cm x 5 cm using side to side movements, rotating the swab tip as you go.

Place the swab into the extraction tube and care-fully break off the end at the pre-scored point. Close the tube with a cap and shake vigorously for 1 minute. Remove cap and replace with dropper tip.

Procedure for finished products Procedure for environmental swabs

1 2

© C

opyr

ight

Rom

er L

abs

Div

isio

n H

oldi

ng G

mbH

, 20

17

3 4 5 6

Romer Labs Inc. - America · Newark, DE, USA · T +1 302 781 6400 · F +1 302 781 6378 · E [email protected] Labs Diagnostic GmbH - Europe · Tulln, Austria · T +43 2272 615 33 · F +43 2272 615 33 13177 · E [email protected] Labs Singapore Pte. Ltd. - Asia · Singapore · T +65 6631 8018 · F +65 6275 5584 · E [email protected] www.romerlabs.com