agilent making a distance-to-fault measurement using the agilent …€¦ ·  · 2007-04-14making...

8
Agilent Making a Distance-to-Fault Measurement Using the Agilent CSA White Paper Introduction In wireless communication systems, cables, connectors, and antennas and other components cause problems. The Agilent CSA spectrum analyzer with the optional stimulus/response measurement suite can help you quickly find these problems. Within the stimulus/ response suite you can use the return loss and the distance- to-fault (DTF) measurements with built-in signal source and bridge to help you determine the severity of the problems in your communication system. The built-in signal source sends signals through your system and looking for reflections back to help pin point fault locations. The bridge sepa- rates the input signal and the reflected signal so they can be measured and compared. It redirects the returning signal internally back to the RF input for analysis. When your system has minor problems, your return loss could be 30 dB or higher while a system with severe faults will show a return loss close to 0 dB. The DTF measurement will allow you to determine how far away from the Agilent CSA signal source input a fault is located, up to 300 m away. This paper will step you through making a DTF measurement with the Agilent CSA. Figure 1 shows a screen capture of a DTF measurement. Figure 1. DTF screen capture Return loss Measurement result table Calibration menu Largest fault 6.76 m away from input Start/Stop distance of your cable

Upload: vananh

Post on 15-Apr-2018

220 views

Category:

Documents


4 download

TRANSCRIPT

Page 1: Agilent Making a Distance-to-Fault Measurement Using the Agilent …€¦ ·  · 2007-04-14Making a Distance-to-Fault Measurement Using the Agilent CSA White Paper ... To return

AgilentMaking a Distance-to-Fault MeasurementUsing the Agilent CSA

White Paper

Introduction

In wireless communication systems,cables, connectors, and antennas andother components cause problems.The Agilent CSA spectrum analyzerwith the optional stimulus/responsemeasurement suite can help youquickly find these problems. Withinthe stimulus/ response suite you canuse the return loss and the distance-to-fault (DTF) measurements withbuilt-in signal source and bridge tohelp you determine the severity of the problems in your communicationsystem.

The built-in signal source sends signals through your system and looking for reflections back to help pinpoint fault locations. The bridge sepa-rates the input signal and the reflectedsignal so they can be measured andcompared. It redirects the returningsignal internally back to the RF inputfor analysis. When your system hasminor problems, your return loss couldbe 30 dB or higher while a systemwith severe faults will show a returnloss close to 0 dB.

The DTF measurement will allow youto determine how far away from theAgilent CSA signal source input a faultis located, up to 300 m away. Thispaper will step you through making aDTF measurement with the Agilent CSA.Figure 1 shows a screen capture of aDTF measurement.

Figure 1. DTF screen capture

Return loss

Measurement result table

Calibration menu

Largest fault 6.76 m away from input

Start/Stop distanceof your cable

Page 2: Agilent Making a Distance-to-Fault Measurement Using the Agilent …€¦ ·  · 2007-04-14Making a Distance-to-Fault Measurement Using the Agilent CSA White Paper ... To return

2

Figure 3 shows the hard key frontpanel of the Agilent CSA. Hard keyswill be noted in bold text while softkeys will be noted in italics in thisdocument.

Familiarize Yourself with the Agilent CSA

Figure 2. Agilent CSA front panel overview

Signal source/RF output

On/Off Help RF input

Soft keys Hard keys

Figure 3. Hard key front panel

Page 3: Agilent Making a Distance-to-Fault Measurement Using the Agilent …€¦ ·  · 2007-04-14Making a Distance-to-Fault Measurement Using the Agilent CSA White Paper ... To return

3

Demonstration

To start using the DTF measurement,push the following keys:

Mode: Stimulus/response: Distance-to-fault

There are two hard keys that you willuse the most when using the DTFmeasurement. They are the FreqChannel and Mode Preset keys,which are circled in red in Figure 3.Mode Preset will reset the instrumentand start you in two-port insertionloss. To return to the DTF measure-ment window press Meas and selectDistance-to-fault.

Freq channel menuIn DTF you will be concerned with theStart and Stop distance of your cable.Push the Freq Range button once toselect Auto as in Figure 5 if you do notknow the frequency range that youwant to use.

Next enter your Start/Stop distance ofthe cable you wish to measure. Themaximum distance is 304.8 m withAuto frequency range selected. If yourstart distance is 0 m you will alwaysshow a fault at 0 m. This is caused bythe reflection from the connection atthe RF output. This DC component issometimes referred to as the “deadzone.”

To calibrate the instrument, you willpress the Calibrate button and walkthrough the step-by-step instructionsshown next.

Before you start the calibrationprocess, you may want to select thetype of cable you are testing as thiswill help make your calibration moreaccurate. Under Meas Setup you willfind Cable Type, where there is a longlist of radio guide (RG) and base transceiver station (BTS) cable types.If you are not using an RG or BTScable you can select “custom.” A sample of the BTS cables in the menuis shown in Figure 4.

Calibration – the first stepAfter selecting your Start/Stopdistance and Cable type select theCalibrate key under the Freq Channelbutton.

Calibrating the Agilent CSA willensure your measurements are accu-rate and will save you time in makingduplicate measurements to check yourresults. The easy-to-use, step-by-stepguide minimizes the need for trainingand helps technicians master theinstruments and get their work doneefficiently.

Figure 5. Frequency channel menu

Figure 4. BTS cables in the menu

Page 4: Agilent Making a Distance-to-Fault Measurement Using the Agilent …€¦ ·  · 2007-04-14Making a Distance-to-Fault Measurement Using the Agilent CSA White Paper ... To return

4

Step-by-Step Menu

The step-by-step menu is as follows:

Specify frequency rangeYou should have already specified the Start/Stop distance under the FreqChannel menu or selected Auto after entering your Start/Stop distance.

Once confirmed select Continue.

Connect the openYou may connect the open directly to the RF output or use a Type-N cable connected to an open.

Once confirmed select Continue.

Do not change or remove the connectionMake sure you do not remove the connection until the next screen appears.The larger the distance you select the longer each step of the calibration willtake. In addition, the more averages you have set the longer this will take. Toturn Averaging off: Meas Setup: Avg Mode: Off.

Connect the shortYou may connect the open directly to the RF output or use a Type-N cable connected to the short.

Once confirmed select Continue.

Connect the loadYou may connect the open directly to the RF output or use a Type-N cable connected to the load.

Once confirmed select Continue.

Now that you have calibrated your instrument, the calibrated frequency rangewill appear in the upper left hand corner of the screen.

Figures 6a-6d. Step-by-step menu

a

b

c

d

Page 5: Agilent Making a Distance-to-Fault Measurement Using the Agilent …€¦ ·  · 2007-04-14Making a Distance-to-Fault Measurement Using the Agilent CSA White Paper ... To return

5

Making a Distance-to-FaultMeasurementAfter calibrating your instrument, con-nect the cable you want to test to theRF output on the Agilent CSA. You willnotice that the top four faults over thedistance that you selected are shownin severity order with yellow markers.These are fault indicators and can beturned On/Off under the View/Display menu. Even when these areturned off, you will see the return loss,distance of the fault from the RF out-put, and voltage standing wave ratio(VSWR) of those top four faults asshown in Figure 7. VSWR measuresthe impedance mismatch between thetransmission line and a load; the higherthe VSWR, the greater the mismatch.

Again, if your start distance is 0 m youwill see a fault at 0 m that correspondsto your connection at the RF output.This is shown as fault two in Figure 1.

In order to determine the severity ofthe any other faults on screen you canalso use markers to see the distanceand return loss of up-to-four otherfaults at one time as shown in Figure 7.Press Marker to use up-to-four mark-ers and their associated delta marker.

How to Pick the Best CableMatches and ConnectionsIt is important that your connectionsare solid so that you have little lossbetween cables. When choosingbetween different connectors andmatching cables the Agilent CSA willhelp you make the best connectionpossible.

Figure 8 shows the DTF measurementwith two 6-feet RG-214 Type-N cablesconnected with a Type-N female-to-female connector. You can see thefault at 6-feet is 32.3 dB. The higherthe return loss, the better the connec-tion/cable match.

Figure 7. Distance-to-fault measurement, fault indicators

Figure 8. Good connector and matching of cables

Page 6: Agilent Making a Distance-to-Fault Measurement Using the Agilent …€¦ ·  · 2007-04-14Making a Distance-to-Fault Measurement Using the Agilent CSA White Paper ... To return

6

Figure 9 shows the DTF measurementwith the same two cables as shownabove with a less than perfect con-nection. The return loss at 6-feet, fault#2, is now 20 dB. This measurementcan be used to pick the best cablematches as well. If you have deter-mined a good cable and connector youwill be able to find the best matchingcable using this same technique ofcomparing the return loss at a knownconnection. You can compare thesedifferent cables/matches by using thefollowing technique of comparing twotraces.

Comparing Two TracesFaults in cables and connections canoccur due to weather, erosion overtime, construction damage, and manyother reasons. The Agilent CSA allowsyou to save and name traces so thatyou can recall them at a later date. Itis a good idea to make a DTF mea-surement and save that trace on newor newly repaired systems so that youcan compare them later when doingmaintenance check ups.

When saving a trace you can eithername it yourself or the Agilent CSAwill name it automatically. To nameyour trace, make sure you select Askunder the following menu: Save:Name: Filename: Ask.

To save your measurement pressSave: Type: Trace: Device: Internal (or USB): Save Now. See menu inFigure 10a.

When recalling a saved file you canchoose whether to view it as trace 1or trace 2. The fault indicators andtable will only show on trace 1; how-ever, you can use markers on eithertrace 1 or trace 2.

To Recall a saved trace push the fol-lowing: Recall: Type: Trace: Device:Internal (or USB): Destination: Trace 1or 2: Recall Now. On your Agilent CSA,trace 1 will be in yellow and trace 2will be in blue as in Figure 11.

Figure 9. Less than perfect connection between two similar cables

Figure 10a - 10b. Save and recall menus

Page 7: Agilent Making a Distance-to-Fault Measurement Using the Agilent …€¦ ·  · 2007-04-14Making a Distance-to-Fault Measurement Using the Agilent CSA White Paper ... To return

7

To see the change in severity from thesaved trace to the new trace pressMarker and push Marker Trace toselect trace 2. Select the Normalmarker and then use the knob to scrollto the fault you want to evaluate. InFigure 11 you can see the markerreadout for the old #3 fault is 29.8 dBwhile the current reading in the FaultIndicator Table for fault #3 is 21.4 dB.This difference of 8.4 dB may indicatea connection that has degraded, andmay be loose, dirty or broken.

Conclusion

Your communication system needs tobe reliable, so you need to be sureyour cables and connections areclean, tight, and in good condition.The Agilent CSA spectrum analyzerwith built-in bridge and signal sourcemakes it easy to measure distance-to-fault and return loss, so you know thecondition of the cables when youinstall them and during routine main-tenance, and you can quickly find theproblem when a repair is needed. Testresults can be stored and recalled soyou can monitor your system andidentify potential problems before theybecome big problems. The AgilentCSA has outstanding RF performancein an easy-to-use, portable package soyou can be confident in the measure-ments. And, the built in step-by-stepcalibration and test set-up graphicsminimize the learning curve on a newinstrument.

Figure 11. Comparing two traces

Page 8: Agilent Making a Distance-to-Fault Measurement Using the Agilent …€¦ ·  · 2007-04-14Making a Distance-to-Fault Measurement Using the Agilent CSA White Paper ... To return

Remove all doubt

Our repair and calibration services will get

your equipment back to you, performing

like new, when promised. You will get

full value out of your Agilent equipment

throughout its lifetime. Your equipment

will be serviced by Agilent-trained techni-

cians using the latest factory calibration

procedures, automated repair diagnostics

and genuine parts. You will always have the

utmost confidence in your measurements.

Agilent offers a wide range of additional

expert test and measurement services for

your equipment, including initial start-up

assistance onsite education and training,

as well as design, system integration, and

project management.

For more information on repair and

calibration services, go to

www.agilent.com/find/removealldoubt

www.agilent.com

For more information on Agilent

Technologies’ products, applications

or services, please contact your local

Agilent office. The complete list is

available at:

www.agilent.com/find/contactus

Phone or Fax

United States:(tel) 800 829 4444(fax) 800 829 4433

Canada:(tel) 877 894 4414(fax) 800 746 4866

China:(tel) 800 810 0189(fax) 800 820 2816

Europe:(tel) 31 20 547 2111

Japan:(tel) (81) 426 56 7832(fax) (81) 426 56 7840

Korea:(tel) (080) 769 0800(fax) (080) 769 0900

Latin America:(tel) (305) 269 7500

Taiwan:(tel) 0800 047 866 (fax) 0800 286 331

Other Asia Pacific Countries:(tel) (65) 6375 8100 (fax) (65) 6755 0042Email: [email protected]: 11/08/06

Product specifications and descriptions

in this document subject to change

without notice.

© Agilent Technologies, Inc. 2006, 2007

Printed in USA, March 1, 2007

5989-5209EN

www.agilent.com/find/emailupdates

Get the latest information on the products

and applications you select.

www.agilent.com/find/agilentdirect

Quickly choose and use your test

equipment solutions with confidence.

www.agilent.com/find/open

Agilent Open simplifies the process of

connecting and programming test systems

to help engineers design, validate and

manufacture electronic products. Agilent

offers open connectivity for a broad range

of system-ready instruments, open industry

software, PC-standard I/O and global sup-

port, which are combined to more easily

integrate test system development.

Agilent Email Updates

Agilent Direct

AgilentOpen