agilent 7890b- 5977a msd introduction · · 2016-09-03faster and easier method development 7...
TRANSCRIPT
Agilent 7890B- 5977A
MSD Introduction
ASTS
March 15, 2013
Graham Robinett
GC-MS Product Specialist
1
240 IT 7890 GC
7000 TQ 7890 GC
5975E SQ 7820 GC
5975T LTM SQ
5977A SQ 7890B GC
7200 Q-TOF 7890GC
More Choices – Better Solutions
Agilent GC/MS/MS
Agilent GC/MS
Today’s Challenges Economic Backdrop: Tension and Uncertainty
March 18, 2013 3
• Operating expenses increasing
• Capital Budgets decreasing
• Asset management and protection a priority Operations
• Regulations, compliance and technology driving need for lower level detection
• Sample complexity increasing
• Many analytes are labile
Technology
•Skill reduction at local level; expertise often sequestered at COE •Lean operation drives cross-training •Higher staff turnover
Human Resources
Agilent Response Solutions to Assure Business Continuity & Reduce Uncertainty
March 18, 2013 4
Reduced operating
expenses: gas use
Improved asset
utilization & protection:
maintenance feedback &
improved GC:MSD
communication
Increased productivity:
fast MSD vent mode &
Rapid consumables
Parts Finder tool
Economic
A more facilitated user
experience
Powerfully Intuitive
Software Solutions:
New OpenLAB CDS
MassHunter for MSD
Integrated Intelligence
Analyzer Solutions
Execution
Better analyte detection:
Improved performance
specifications
Better protection of
labile compounds:
Agilent Inert Flow Path
Expanded lab
capabilities:
improved FPD, ECD as
3rd detector, dual MMI
Technical
Innovations in Hardware, Software, Chemistries and Workflows
Improve Asset Utilization Leverage and Protect Your Investment Better
5
Economic
5977A 7890B
During MSD vent… Increases GC carrier gas flows,
Cools MSD faster
40% Faster Vent Times
(280C 100C)
In event of MSD Failure… Stops GC Flows
Don’t waste expensive He
Don’t build-up H2
5977A 7890B
GC monitors GC/MSD
connection.
If communication lost… GC shuts down various
(GC) thermal zones
Integrated GC-MSD Communication
Faster Maintenance, better asset protection, safer H2 use
Reducing Operating Expenses Reducing Dependence on He
6
Economic
Can save as much as 15% in gas usage – up to $4,000 per year per instrument for many labs
Sleep-Wake Function
Manage He Use Optimally
• Set time to sleep
• Set time to wake
• Choose to condition GC before first run
• Use only the gas you need
Increase Productivity Faster and easier method development
7
Economic
Bar Code Scanning Option
AutoInput column, liner and septum information into method. Integrated with Parts Finder for seamless consumables management.
Improved Asset Utilization Avoid Unplanned Downtime
March 18, 2013
Confidentiality Label
8
Economic
Optimize your maintenance plan
• You determine your maintenance schedule,
not your instrument
• Use only the consumables you need, when
you need them
• A maintenance SOP built into your GC
When production or shipment decisions are on the line,
unplanned downtime can cost $1000’s
Early Maintenance Feedback
Saving Energy and Time
Agilent Confidential Tonga Project Definition Check Point
9
0
50
100
150
200
250
300
350
400
NormalOp
Standby
Power Consumption - 5977
0
50
100
150
200
250
0 4 6 7 15 20 23 30 35 40
Vent Fast
Vent
Power
saved
Time saved
Po
we
r( V
A)
So
urc
e T
em
p
Time(min)
Time saved using Fast Vent
Economic
Reducing Operating Expenses Reducing Dependence on He
March 18, 2013 10
Economic
Switching to N2 gas overnight and on weekend downtime can save up to 180 days of He per
GC instrument
He Conservation Module
Manage He Use Optimally
• Use He only when needed
• Extend He tank usage
• Reduce operating costs
• Reduce gas supply uncertainty
Improved Performance Deliver better results
11
UltiMetal – TCD,
FPD, NPD/FID Jets
UltiMetal Capillary Flow Technology
Devices, Ultimate Union
UltiMetal Inlet Weldment, Shell
and Transfer Lines
Ultra Inert GC Column
Ultra Inert Gold Seal
Ultra Inert Inlet
Liner
New UltiMetal
FlexiMetal Ferrules
Technical …now from a single supplier
Agilent Inert Flowpath SSL Inlet Option Proprietary UltiMetal-Plus Treatment
12
Enhanced performance for trace GCMS and
GC-ECD analysis of pesticides and drugs of
abuse
Protects labile analytes contacting injector
metal structure during high volume injections
Available as 7890B Inert Flowpath option #114
or upgrade G3453B for 7890A/B
Technical
Improved Performance Deliver better results
March 18, 2013 13
Technical
Detector Performance
MSD Scan 1500:1 S/N; SIM 10 fg IDL
FID <1.4 pg C/sec
ECD <4.4 fg/sec
FPD <2.5 pg S/sec; <45 fg P/sec
SCD <0.5 pg S/sec
NPD <0.01 pg P/sec, 0.08 pg N/sec (Blos bead)
TCD 400 pg/mL
New Extractor Ion Source for 5977
Page 14
Why the Extractor EI Source?
Increase in ion count
True increase in sensitivity
Better trace level precision
Lower IDL and MDL
Proven in the 7000 Series GC/TQ!
Technical
Comparison of Inert and Extractor Ion Source
Agilent Confidential
Page 15
Repeller
(Push)
Drawout Plate
Ion Focus Lens
Entrance Lens
Ionization area
Ions
Inert Ion Source
Extractor Lens
(Pull)
Extractor Source
More ions to
Mass Filter
Mass Filter
January 8, 2013
Extractor Source “Etune” Sensitivity Increase
16
Blue – Extractor 40ppb solution (SVOC)
Red – Standard 40ppb solution (SVOC)
5977 Extractor Source: Improved Thermal Profile
Page 17
Better performance of high boilers
i.e. PAH even at lower concentration!
5ppb PAH
Top of the peak shows
expected imprecision, but no
response trending with new
5977 Extractor Source
GC Conditions:
80 minute separation
Longer than normal
to enhance visibility
of trending
Technical
Better Specs for Working Methods:
Instrument Detection Limit (IDL)
Page 19
More ions = Better precision = Better IDL = Better MDL
10 fg Extractor EI (24 fg Inert EI/TP)
Method conditions
TESTED at INSTALLATION (factory spec even lower)
Technical
European Pharmacopoeia 6.2 Council of Europe, Strasbourg, France, 2008
20x pk width half
height
LCGC Europe, Vol 22 (2), Feb 2009; R Boqué, Y Vander Heyden
S/N = 2H
h
H is measured from
mid-point of baseline
immediately before
and after peak (not
mid-point of h)
Noise time segment
carefully defined
Three Improved Specifications
Page 22
Market’s ONLY SQ IDL < 10 fg
Market’s Best S/N (SNR) > 1500:1
Technical
Better HW Features: Agilent Turbo Pump (V304)
Page 23
Agilent Quality Standard
TwisTorr 304 FS Innovation for:
•Higher pumping performance
• especially for He and H2
•High reliability
• lower cost of operation
Technical
Economic
24
TwisTorr 304 FS: Vacuum Design
8 EDM turbo stages
(rotor) designed to
optimize pumping
speed for light gases
8 Turbo stators
made by pressed
SST sheet
technology
3 TwisTorr drag stages (spiral
channels) designed for high
compression ratio performances
Agilent’s new patent-
pending Floating
Suspension for robust,
low noise operation
255 L/s helium
220 L/s hydrogen
Technical
Better HW Features: Agilent Rough Pumps
Page 25
New DS42i New IDP-3c
High performance
Energy saving design “Inverter” not compatible with diff pump
High performance
Oil-free scroll pump
NH3 CI compatible
with Agilent Quality Standards
MassHunter GC/MS Acquisition for GC/MSD
Unified GC/MS Acquisition and
MassHunter Data Analysis; with
MSD ChemStation workflow for GC/MSD
26
Economic
The Value of G1701FA Software
All Inclusive G1701FA Software Package
• Acquisition, Quant, Qual, MSD ChemStation
• 21CFR11 Compliance tools (Acquisition and
Quant)
• GC/MS and LC/MS Translators
Collated User Information
• Workflow guides, Video Tutorials, all in one
location
Support
• 1yr Upgrade, Updates, and Phone Support
• Installation and Familiarization options
• New Familiarization material for GC/MSD
users
MSD ChemStation DA
MassHunter Qual
MassHunter Quant
27
GC/MSD Software MassHunter Qual
MassHunter Quant MH
CS
MassHunter GC/MS Acquisition
And Use Either
MassHunter Data Analysis
OR
MSD ChemStation Data Analysis MSD ChemStation DA
G1701FA
28
GC/MSD Acquisition Methods
MSD Productivity ChemStation MassHunter GC/MS Acquisition
29
GC/MS Acquisition Methods are Transparent
from MSD ChemStation to MassHunter GC/MS (Except PAL)
Moving from MSD Productivity ChemStation EA to
MassHunter with MSD ChemStation Data Analysis (FA)
GC & GC/MS Acquisition data is compatible with both
MSD Productivity ChemStation EA
MSD ChemStation Data Analysis FA (MassHunter)
MSD Productivity ChemStation (EA)
MSD ChemStation DA (FA)
30
Migrate Data to MassHunter Format
Data
Data
Data
MassHunter Quant
MassHunter Unknowns Analysis
MassHunter Qual
ChemStation (.D)
WorkStation (.SMS)
Generic (.CDF)
31
Migrate Quantitative Methods to
MassHunter Format
MSD ChemStation (QDB.MTH)
WorkStation (.mth)
MassHunter Quant
Method
Method
32
MSD Productivity ChemStation OR MassHunter?
MSD ChemStation
Established Agilent GC/MSD User
• Adding instrument to routine production
• Established workflows
• Macros, Macros, Macros
DRS or Screener DB
EnviroQuant User
• Uses the US EPA workflow, especially
reports and QA/QC criteria.
• MassHunter EnviroQuant targeted for
late 2013
MassHunter
Look and feel of modern software
Superior Quantitation
21CFR11 & Other Regulated Labs
Multi disciplinary MS lab
• Have Agilent Triple Quad or (Q-)TOF
• LC/MS, ICP-MS, & GC/MS
• MS and MS/MS
33
1-Year Software Maintenance Agreement
Supported Add-on Software
DRS and Analyzers
DRS is for MSD Productivity
ChemStation mode ONLY
Headspace PAL
Fully supported in GC/MS Acquisition
Mass Profiler Professional GC/MS Libraries & RTL Database
34
Flexibility: Performance and Budget
Page 35
Lowest Price
Performance Meets
Method Requirement
5977E Turbo MS
5977E Diff MS
7820A
Stainless source
Highest
GC Performance
& MS Performance
5977A EI
5977A EI/CI
7890B GC
Extractor source
DS42i
Balance of Price,
GC performance,
& MS performance
5977A Turbo MS
5977A Diff MS
7890B GC
Inert source
Introducing the 5977A Series GC/MSD Re-defining Premium Single Quad GC/MS
Improved Performance Specifications Best-in-class Single Quadrupole Performance
New & Expanded Accessories & Options Agilent Turbomolecular and Roughing pumps
New Inert Flowpath option
Dual MMI inlets
Improved non-MS detectors
ChemStation AND MassHunter Data Handling
Integrated Intelligence Functions GC-MSD communication
Sleep-wake function
Method development calculators
3D Interactive Graphical Parts Finder
Early Maintenance Feedback
Barcode Scanning with Auto-Input
36
37
7890B/5977A GC/MS Solution Portfolio Driving Business Value and Success
Proven 7890 and 5977 Reliability
MassHunter & ChemStation SW
Integrated Intelligence
Improved Performance
Greater Operating Efficiency
Expanded Accessories and Options
MassHunter Superior Quantitative Analysis
• Batch Design for Comprehensive Data Review
• Automated Quality Checking and Outlier
display
• Interactive Calibration Results
Unified Agilent MS solution
38