active stabilization - crossroads extremity€¦ · lapidus plate surgical technique step 01 plate...
TRANSCRIPT
AC TIVESTABILIZ ATION®
- - - - - - - - - -
L A P I D U S I M P L A N T S Y S T E M
O V E R V I E W
A C T I V E S T A B I L I Z A T I O N®
The Active Stabilization® System is the only hybrid plating system to replace static compression with dynamic
continuous compression. It is designed to address the most challenging foot & ankle procedures.
Active Stabilization®Bending & Rotational Stability Continuous Compression
P R O D U C T F E A T U R E S
Two distinct plate options
Plate and clip designed with 10˚ contour to match
patient anatomy
3x more effective at maintaining compression than a
traditional plate and cross screw*
Sterile recoverable instrumentation with EcoSMART®
*data on file
A NEW ERA IN BONE FIXATION
I N S T R U M E N T A T I O NActive Stabilization® comes with sterile packed stainless steel instrumentation. Inserters are pre-loaded with implants
for each surgery.
Inserter Drill Guide Tamp
R O T A T I O N A L S T A B I L I T YThe Active Stabilization® system has
higher rotational stability compared
to a plate and cross screw.
COMPETITIVE PLATE + CROSS SCREW
Z-PLATE + CLIP 3.3
8.4
ROTATIONAL STABILIT Y AT 15°
VS.
TO
RQ
UE
(N
-M)
9
8
7
6
5
4
3
2
1 Cup Reamer
G A P R E C O V E R Y The Active Stabilization® system
presented full gap recovery, while
the plate with no clip did not
recover fully.
GAP AFTER LOADING WITH NO CLIP
GAP AFTER LOADING WITH CLIP
Active Stabilization® 4-Hole Plate + Clip
PL
AN
TA
R G
AP
(M
M)
4-Hole Plate + No Clip
1
2
3
4
5
6
7
1.56
0.00
VS.
T E S T I N G D A T A
S P E C I F I C A T I O N S
NON-LOCKING LOCKING
SIZE RANGE 8-30mm* 8-30mm*
SIZE OPTION 3.0 & 3.5mm 3.0 & 3.5mm
*2mm increments
L Z - P L A T EAlpha, 18 and Beta, 18
L C - P L A T EAlpha, 18 and Beta, 18
S C R E W SNon-locking
18 x 18 18 x 20
INTERAXIS LENGTH 18mm 18mm
LEG LENGTH 18mm 20mm
REAMER SIZE 3.2mm 3.2mm
MAX CLIP THICKNESS 2.7mm 2.7mm
LEG WIDTH 1.8mm 1.8mm
CATALOG 7118-1818KT-C 7118-2020KT-C
H I M A X™ - C C L I P S
31.80
10º CONTOUR 10º CONTOUR
44.16
ALPHA, 18 BETA, 18
ACCEPTABLE CLIP SIZE 18 18
SLOT LENGTH / CLIP SIZE 18 18
OVERALL LENGTH 31.8 31.8
THICKNESS 1.7 1.7
CATALOG 7100-LZ18-A 7100-LZ18-B
ALPHA, 18 BETA, 18
ACCEPTABLE CLIP SIZE 18 18
SLOT LENGTH / CLIP SIZE 18 18
OVERALL LENGTH 44.16 44.16
THICKNESS 1.7 1.7
CATALOG 7100-LC18-A 7100-LC18-B
A
B
CD E
F
G
HI
I N S T R U M E N T S I N C L U D E D I N E A C H K I TInstrumentation is packaged in sterile kits for efficiency
and convenience. After surgery, the instrumentation is
returned to CrossRoads® via the EcoSMART® Instrument
Recovery Service providing a significant cost savings
over typical disposable reusable instrument kits.
A Driver Handle F Driver Bits (2)
B Plate Bender (2) G Fixation Pins (2)
C 2.5mm Reamer H Locking Drill Guide (2)
D 2.0mm Reamer I Non-Locking Drill Guide
E Depth Gauge Probe
I N S T R U M E N T A T I O N
L O A N E R K I T S
C L I P S• 1 each, (18x18 Kit)
• 1 each, (18x20 Kit)
• 2 each Single Pack, (18x18,18x20)
P L A T E S• 2 each, Lapidus Z Alpha
• 2 each, Lapidus Z Beta
• 2 each, Lapidus C Alpha
• 2 each, Lapidus C Beta
• 1 each, Plate Instrument Kits
L A P I D U S P L A T E S U R G I C A L T E C H N I Q U E
S T E P 0 1Plate SelectionAfter temporary fixation is applied,
trial to determine best anatomic fit.
Note: Trial with sterile plates.
S T E P 0 2Temporary FixationOnce plate is selected, use provided
fixation pins to secure plate and
confirm proper location.
S T E P 0 3Clip PreparationUse provided drill guide to prepare
holes for the HiMAX™-C Clip.
LC-PLATE
LC-PLATE
LC-PLATE LZ-PLATE
LZ-PLATE
LZ-PLATE
S T E P 0 4Clip PlacementInsert HiMAX™-C Clip through the
plate and tamp flush.
S T E P 0 5Screw PlacementStarting with the most distal hole,
prepare screw holes and fill with
appropriate length non-locking
screws using non-locking drill guide.
Repeat until all holes are filled.
Note: Non-locking screws are recommended
LC-PLATE
LC-PLATE
LZ-PLATE
LZ-PLATE
*Patents Issued
6055 Primacy Parkway, Suite 140, Memphis, Tennessee 38119
(901) 221-8406 | [email protected] | crextremity.com LA0319 REV A