ac procedure

Upload: r-wah-larounette

Post on 03-Apr-2018

214 views

Category:

Documents


0 download

TRANSCRIPT

  • 7/28/2019 AC Procedure

    1/5

    Procedure 1: Creating and Measuring AC Voltages

    In this procedure the Function Generator was connected to the oscilloscope to provide AC

    voltage signals, each time with different shape: sinusoidal wave or square wave or triangle

    wave, also each with different settings for the frequency, the offset voltage and for the peak-to-peak amplitude. Then, the mathematical equation describing the signal produced was

    recorded. Afterwards, the display of the signal's wave on the oscilloscope's screen was

    adjusted to view a specified number of cycles of the signal. Measurements regarding the

    signal produced were recorded using two ways. First, the values of the horizontal setting,

    vertical setting, period, frequency, amplitude and offset were taken from the oscilloscope's

    screen and were recorded. Second, measurements of the period, the frequency, the amplitude

    and the offset voltage were taken from the "Quick Measure" option provided by the

    oscilloscope and were also recorded. Eventually, the percentage error of the frequency and the

    amplitude for each signal was calculated.

    Results and Discussions:

    The 1st

    signal produced was for a sinusoidal wave with a frequency of 200Hz, no offset

    voltage, and peak-to-peak amplitude of 4Vpp. The mathematical equation describing the

    voltage signal was written as the following:

    V(t)= 2sin((400)t)

    The measurements taken from the oscilloscope's screen were recorded in table (1.1). Whereas,

    the measurements received from the "Quick Measure" option were recorded in table (1.2).

  • 7/28/2019 AC Procedure

    2/5

    It's noticeable that the values in both tables are close to each other and almost equivalent to

    the specified settings. However, the values in table (1.2) show more precise measurements

    compared to the values in table (1.1). For example, the period's value in both tables equals 5

    but the resolution/error (least count) is better when using quick measure with a value of 0.001while using the screen gives a resolution of 0.2. Also, it's obvious in the case of the amplitude

    (peak to peak) which shows in table (1.1) a resolution of 0.25 while a resolution of 0.01 is

    shown in table (1.2). On the other hand, the offset's value in table (1.1) is more accurate.

    Signal of Procedure 1.1

    ( Screen)Measurement

    error

    (Resolution) Units

    Wave shape sin wave -

    Oscilloscope Horizontal

    Setting 1 0.2 MsOscilloscope Vertical Setting 1 0.25 V

    Period 5 0.2 Ms

    Frequency 0.2 - KHz

    Amplitude (peak to peak) 4 0.25 V

    Offset 0 0.25 V

    Signal of Procedure 1.1 ( Quick

    measure)Measurement

    error

    (Resolution) Units

    Wave shape sin wave -Period 5 0.001 ms

    Frequency 200 0.1 Hz

    Amplitude (peak to peak) 4.06 0.01 v

    Offset (avg) 156 1 mv

  • 7/28/2019 AC Procedure

    3/5

    The 2nd signal produced was for a square wave with a frequency of 2000Hz, no offset voltage,and peak-to-peak amplitude of 4Vpp. It was described by the following mathematicalequation:

    In table (1.3) the measurements shown were taken from the screen. While, the measurementsprovided by the "Quick Measure" option were recorded in table (1.4).

    Signal of Procedure 1.4 ( Screen) Measurement error (Resolution)

    Units

    Wave shape Square wave -

    Oscilloscope Horizontal Setting 200 40 s

    Oscilloscope Vertical Setting 1 0.25 v

    Period 500 40 s

    Frequency 0.002 - Hz

    Amplitude (peak to peak) 4 0.25 v

    Offset 0.25 0.25 v

    As seen in table (1.3) the period's resolution is the same as the horizontal setting's resolution

    and equals 40 s. This is due to the fact that the period's value is taken from the horizontal

    axis, whereas, when using the quick measure the settings for the horizontal and the vertical

    aren't needed.

    The two tables show similar values for the period and the frequency. The slightest difference

    is noticed in the values of the amplitude and the offset where in table (1.3) the amplitude is

    equivalent to the specified settings provided by the function generator which is 4Vpp.

    However, since there's no offset for this signal table (1.4) gives a close value to zero

    compared to table (1.3).

    Signal of Procedure 1.4

    ( Quick measure)Measurement

    error

    (Resolution) Units

    Wave shape square wave -

    Period 500 1 s

    Frequency 2 0.01 KHz

    Amplitude (peak to peak) 4.1 0.01 v

    Offset 156 1 mv

    500

    2

    500:2

    2

    5000:2

    )(t

    t

    tV

  • 7/28/2019 AC Procedure

    4/5

    The 3rd signal produced was for a triangle wave with a frequency of 100Hz, 2V offset, andpeak-to-peak amplitude of 4Vpp. The mathematical equation describing the produced signalwas recorded as the following:

    Measurements using the screen were recorded in table (1.5), while, measurements using quickmeasure were recorded in table (1.6).

    Signal of Procedure 1.7

    ( Screen)

    Measurement

    error

    (Resolution)/least

    count

    Units

    Wave shape Triangle wave -

    Oscilloscope Horizontal Setting 2 0.4 ms

    Oscilloscope Vertical Setting 1 0.25 v

    Period 10 0.4 ms

    Frequency 0.1 - KHz

    Amplitude (peak to peak) 4 0.25 V

    Offset 2 0.25 v

    Once again the values in both tables are close to each other. But, it can be noted that using

    the screen as in tables (1.5) provides values that are much closer to the specified setting

    produced by the function generator and this is mostly noticeable in the value of the offset

    which is exactly 2, whereas, it's 2.163 in table (1.6).

    Signal of Procedure 1.7

    ( Quick measure)Measurement

    error

    (Resolution) Units

    Wave shape Triangle wave -

    Period 10.01 0.01 ms

    Frequency 99.9 0.1 Hz

    Amplitude (peak to peak) 4.06 0.01 v

    Offset 2.163 0.001 v

    104

    30:2)10(

    4/10

    2

    4

    30

    4

    10:2)

    2

    10(

    4/10

    2

    4

    100:2

    4/10

    2

    )(

    tt

    tt

    tt

    tV

  • 7/28/2019 AC Procedure

    5/5

    The 4th and the last signal produced was also for a square wave and had a frequency of450Hz, an offset voltage of 2.5V, and peak-to-peak amplitude of 5.5Vpp. It was described bythe following mathematical equation:

    The measurements taken from the screen were recorded in table (1.7). While, the

    measurements provided by quick measure were recorded in table (1.8).

    Signal of Procedure 1.10

    ( Quick measure)Measurement

    error

    (Resolution) Units

    Wave shape square wave -

    Period 2.22 0.01 ms

    Frequency 450.5 0.1 Hz

    Amplitude (peak to peak) 5.5 0.1 V

    Offset 2.62 0.01 V

    The most noticeable values when comparing the two tables are the values of the frequencyand the offset. In table (1.7) the frequency equals approximately 0.0004545 Hz which equals

    454.5 Hz, while, the frequency in table (1.8) equals 450.5 Hz and is closer to the actual value.However, the value of the offset in table (1.7) is 2.5 V, whereas, it's 2.62 V in table (1.8). In

    this case the value of the offset in table (1.7) is equivalent to the actual value.

    Signal of Procedure 1.10 ( Screen) Measurement

    error

    (Resolution) Units

    Wave shape Square wave -

    Oscilloscope Horizontal Setting 500 100 s

    Oscilloscope Vertical Setting 1 0.25 v

    Period 2200 100 s

    Frequency 0.0004545 - Hz

    Amplitude (peak to peak) 5.5 0.25 V

    Offset 2.5 0.25 V

    22002

    2200:75.2

    2

    22000:75.2

    )(

    t

    t

    tV