a study of time over threshold (tot) technique for plastic scintillator counter

12
A Study of Time over Threshold (TOT) Technique for Plastic Scintillator Counter 高高高高高高高 高高高

Upload: jonah-contreras

Post on 01-Jan-2016

21 views

Category:

Documents


0 download

DESCRIPTION

A Study of Time over Threshold (TOT) Technique for Plastic Scintillator Counter. 高能物理研究所 吴金杰. Outline. Introduction Research method-wave analysis Signal characteristics Relative charge resolution by TOT Summary. Introduction to TOT (1). TOT -­ T ime O ver T hreshold Pulse width - PowerPoint PPT Presentation

TRANSCRIPT

Page 1: A Study of Time over Threshold (TOT) Technique for Plastic Scintillator Counter

A Study of Time over Threshold (TOT) Technique

for Plastic Scintillator Counter

高能物理研究所 吴金杰

Page 2: A Study of Time over Threshold (TOT) Technique for Plastic Scintillator Counter

Outline

Introduction Research method-wave analysis Signal characteristics Relative charge resolution by TOT Summary

Page 3: A Study of Time over Threshold (TOT) Technique for Plastic Scintillator Counter

Introduction to TOT (1)

TOT- Time Over Threshold

Pulse width

TOT is somewhat dependent the charge

Different threshold

Advantage:

One of alternatives to charge measurement

Simplify the electronics

Reduce the cost

Time over threshold

Page 4: A Study of Time over Threshold (TOT) Technique for Plastic Scintillator Counter

Introduction to TOT (2)

Apply in gas detectors Multi-gap Resistive Plate Chamber of

ALICE and STAR:

NINO-Correction of front edge time walk

Drift Chamber of CLEO-III: Relative charge resolution: 10%

TOT for Plastic scintillator counter?The correlation between TOT and Charge?

Relative charge resolution?Relative Charge resolution of CLEO

TOT versus measured time

Page 5: A Study of Time over Threshold (TOT) Technique for Plastic Scintillator Counter

Research method-wave analysis

Scintillator

PMT1

PMT3

Scintillator EJ200

PMT2

Sintillator 1

PMT R5924

PMT R5924

Signal from the scintillator bar

Delay cable

Oscilloscope

PMT1PMT2PMT3

DiscriminatorLogic unit 1 ∩ 2 ∩ 3

Trigger

CH1

CH2

To be measure TOT, Amplitude, Charge

How?

Chip?

Based on CAMAC system?

Wave analysisUsing high quality oscilloscope

sample rate 20 GS/s

TOF module

Scintillator EJ200 +R5924 Experimental setup

Page 6: A Study of Time over Threshold (TOT) Technique for Plastic Scintillator Counter

Pedestal, Noise, Rise time

The data within the first 250 ns are noise without any particle passing through the scintillator counter

Noise: 2.0 mV

Pedestal: 2.2 mV

Rise time:4.2ns A wave shape of pulse

The noise spectrum Spectrum of rise time

Reference to the design BESIII TOF electronics

Page 7: A Study of Time over Threshold (TOT) Technique for Plastic Scintillator Counter

Amplitude, Charge

Spectrum of amplitude

Spectrum of charge

Charge versus amplitude

2000

0

50

50

i

ii

ps VQ

The charge is obtained from the time integral of the current pulse

Page 8: A Study of Time over Threshold (TOT) Technique for Plastic Scintillator Counter

Distribution of TOT

The distribution of TOT when the threshold is set to 65 mV

15% of the most probable value of amplitude

The distribution of TOT when the threshold is set to 195 mV

45% of the most probable value of amplitude

The distribution of TOT with different threshold

Page 9: A Study of Time over Threshold (TOT) Technique for Plastic Scintillator Counter

TOT VS Charge--threshold 65mV

The relative charge resolution versus charge

TOT versus charge

Linear correlation between TOT and charge

Relative charge resolution can reach 7%-18% when the threshold is set to 65mV

Q TOT actual

actual

Q Q

Q Q

The relative error of Q by TOT can be obtained by formula

Page 10: A Study of Time over Threshold (TOT) Technique for Plastic Scintillator Counter

TOT VS Charge--threshold 195mV

The relative charge resolution versus charge

TOT versus charge

The correlation between TOT and charge

Relative charge resolution can reach 4%-7% when the threshold is set to 195mV

/TOT A B Q

Page 11: A Study of Time over Threshold (TOT) Technique for Plastic Scintillator Counter

Summary

A Study of TOT technique for Scintillator counter firstly

Wave analysis The Characteristics of TOF module

Pedestal, Noise, Amplitude, Charge, Rise time

Relative charge resolution From 7% to 18% (Threshold is set to 65mv)

From 4% to 7% (Threshold is set to 195mv)

Page 12: A Study of Time over Threshold (TOT) Technique for Plastic Scintillator Counter

Thanks