A new suite of tools for Radioactivity Measurement Laboratories João Abrantes , Gonçalo Carvalhal, Mário Reis Instituto Superior Técnico, Universidade Técnica de Lisboa, Portugal Estrada Nacional 10, 2686-953 Sacavém (PORTUGAL) Advances in Liquid Scintillation Spectrometry 18-22 March 2013, Barcelona

Upload: trinhkien

Post on 10-Nov-2018

217 views

Category:

Documents


0 download

TRANSCRIPT

A new suite of tools for Radioactivity Measurement Laboratories

João Abrantes, Gonçalo Carvalhal, Mário Reis

Instituto Superior Técnico, Universidade Técnica de Lisboa, Portugal Estrada Nacional 10, 2686-953 Sacavém (PORTUGAL)

Advances in Liquid Scintillation Spectrometry 18-22 March 2013, Barcelona

Laboratory workload

Advances in Liquid Scintillation Spectrometry

18-22 March 2013, Barcelona

• ~2500 samples/year • 10-12 assay methods (gross alpha/beta, H-3, Sr-90, Cs-137, Rn-

222) • Different sample types (water, milk, food) • Different detection systems (GPC, LSC) • Different measurement purposes (National Surveillance,

external clients, R&D, internal studies)

ISO 17025 accreditation

Advances in Liquid Scintillation Spectrometry

18-22 March 2013, Barcelona

First steps

Advances in Liquid Scintillation Spectrometry

18-22 March 2013, Barcelona

• Identify commom processes • Define workflow • Classify items

• Samples by class (cal. standard, background, blank, cert. ref. material, internal QC, …) and type (water, milk, food, …)

• Batches (ordered sequence of samples) • Analyses (ordered sequence of batches) • Measurements (associated to each analysis and detector

system) • Methods (assay/routine, reference/QC)

Requirements (development)

Advances in Liquid Scintillation Spectrometry

18-22 March 2013, Barcelona

• Major tasks defined as modules • Easy to use • Easy to check/debug • Parameterizable • Rapid implementation

R: free software environment for statistical computing and graphics (http://www.r-project.org/) Data stored in ASCII files

Requirements (outcome)

Advances in Liquid Scintillation Spectrometry

18-22 March 2013, Barcelona

• Get data from measurements and samples • Perform activity calculations including uncertainties and

detection limits • Generate reports • Determinate ROIs for LSC • Calculate calibration parameters • Fit calibration curves • Estimate MDA from calibration curves • Check detector performance • Check method QC • …

Detector performance (I)

Advances in Liquid Scintillation Spectrometry

18-22 March 2013, Barcelona

Reference methods defined for detector system • alpha/beta standards (GPC, LSC) • beta standards (LL LSC)

Standard sources and backgrounds assigned to detectors Am-241 (and BKG) on Tricarb 3170

Detector performance (II)

Advances in Liquid Scintillation Spectrometry

18-22 March 2013, Barcelona

Ability to monitor tSIE for standard set kit (Am-241, Cl-36, BKG)

Method QC

Advances in Liquid Scintillation Spectrometry

18-22 March 2013, Barcelona

For each method: • check background and

blank count rates • monitor tSIE

Sr-90 in food and milk

ROI determination in LSC

Advances in Liquid Scintillation Spectrometry

18-22 March 2013, Barcelona

Loads standard and background files to determinate ROIs (from FOM)

Calibration

Advances in Liquid Scintillation Spectrometry

18-22 March 2013, Barcelona

Uses standard (working) solution and standards data, measurement data to calculate activity, efficiency and interference (spillover/crosstalk).

Handles repeatability on measurements to calculate uncertainties (type A and B)

Fitting data (I)

Advances in Liquid Scintillation Spectrometry

18-22 March 2013, Barcelona

Loads output from calibration, performs fit and plots calibration curves.

H-3 in water (LSC) Gross alpha (Am-241) in water (LSC)

Fitting data (II)

Advances in Liquid Scintillation Spectrometry

18-22 March 2013, Barcelona

Also returns goodness of fit parameters for evaluation

Gross beta (K-40) in water (GPC)

MDA estimated after calibration

Advances in Liquid Scintillation Spectrometry

18-22 March 2013, Barcelona

Gross alpha (Am-241) in water (LSC) (10x conc. @ top, no concentration @ bottom) H-3 in water (LSC)

Activity calculation

Advances in Liquid Scintillation Spectrometry

18-22 March 2013, Barcelona

Implementation for: • Gross alpha/beta activities on water and suspended matter

(GPC, LSC) • Sr-90 in water (GPC) • Cs-137 in water (GPC) • Sr-90 in food and milk (LSC) • Rn-222 in water (LSC, 2 methods) • H-3 in water (LSC, 2 methods) • H-3 in RBA (LSC)

Returns all calculations to be used in uncertainty budget

Other tools

Advances in Liquid Scintillation Spectrometry

18-22 March 2013, Barcelona

• Get data from detector systems (only Tricarb) • Send worklist to detectors • Create and fill batches and analysis • Generate report (ASCII) • Statistics on data (managment)

Conclusions

Advances in Liquid Scintillation Spectrometry

18-22 March 2013, Barcelona

• Main laboratory procedures can easily be streamlined in a systematic, time-saving way

• Data and documentation traceability

• Quality of results assured

• Less error prone

Future work

Advances in Liquid Scintillation Spectrometry

18-22 March 2013, Barcelona

• Database development • Graphical User Interface (integrating modules) • Sample data module • Extend capabilities to other detector systems • Spectra view • Statistical and database tools (managment) • …

Thank you for your attention.