1 _______________________________________________________________________________________________ NOTICE: this is the author’s version of a work that was accepted for publication in BioChip Journal. Changes resulting from the publishing process, such as peer review, editing, corrections, structural formatting, and other quality control mechanisms may not be reflected in this document. Changes may have been made to this work since it was submitted for publication. A definitive version will be subsequently published in BioChip Journal, [Vol.7(4), pp.367374, (2013)].” _______________________________________________________________________________________ A Microfluidic Device for Partial Cell Separation and Deformability Assessment Diana Pinho 1, 2 , Tomoko Yaginuma 1 & Rui Lima *1, 2 1 Polytechnic Institute of Bragança, ESTiG/IPB, C. Sta. Apolonia, 5301-857 Bragança, Portugal. 2 CEFT, Faculdade de Engenharia da Universidade do Porto (FEUP), R. Dr. Roberto Frias, 4200-465 Porto, Portugal. * Correspondence and requests for materials should be addressed to R. Lima (ruimec@ipb.pt) Abstract Blood flow in microcirculation shows several interesting phenomena that can be used to develop microfluidic devices for blood separation and analysis in continuous flow. In this study we present a novel continuous microfluidic device for partial extraction of red blood cells (RBCs) and subsequent measurement of RBC deformability. For this purpose, we use polydimethylsiloxane (PDMS) microchannels having different constrictions (25%, 50% and 75%) to investigate their effect on the cell-free layer (CFL) thickness and separation efficiency. By using a combination of image analysis techniques we are able to automatically measure the CFL width before and after an artificial constriction. The results suggest that the CFL width increases with enhancement of the constriction and contributes to partial cell separation. The subsequent measurements of RBCs deformation index reveal that the degree of deformation depends on the constriction geometries and hematocrit after the cell separation module. The proposed microfluidic device can be easily transformed into a simple, inexpensive and convenient clinical tool able to perform both RBC separation and deformability analysis in one single device. This would eliminate the need for external sample handling and thus reducing associated labor costs and potential human errors. Keywords: Biomicrofluidics, Microfluidic devices, Microcirculation, Blood on chips, Red blood cells, Cell separation, Cell deformability, Deformation index. Introduction Cell separation and identification are essential in a variety of biomedical applications including cell biology, diagnostic and therapeutic methods. Blood is a non-Newtonian fluid containing extremely rich amount of information about the physiological and pathological state of the human body. However, due to its complexity there are few accurate analysis methods. Most of the standard techniques used cell separation and sorting are often labor intensive or require additional external labels to identify cells. Blood flow in microcirculation shows several interesting phenomena that can be used to develop microfluidic devices for blood separation and analysis in continuous flow. These phenomena include plasma skimming [1], cell-free layer (CFL) [2, 3, 4], leukocyte margination [5] and bifurcation law [6]. Recently, several researchers have used these effects and replicated in microfluidic systems. In microchannels RBCs, due to their deformability and lift forces, tend to be concentrated around the center of the microchannels while white blood cells (WBCs) and rigid RBCs (such as Malaria RBCs) tend to migrate to the CFL originated at regions close to the walls [1-5]. Bifurcation law [6] states that RBCs behavior, in microchannels with bifurcations, tends to be oriented to the wide microchannel. A number of microfluidic devices have been developed to take advantage of these natural hemodynamics phenomena. Shevkoplyas et al. [7] developed a microdevice to isolate WBCs from a blood sample by using the margination effect, whereas Hou et al. [8] have very recently proposed a biomimetic separation device to separate normal RBCs from malaria infected RBCs. Other researchers have found several advantages to control and manipulate blood flow in microfluidic devices. Fujiwara et al. [9] have found evidence that it is possible to create an artificial CFL under appropriate hemodynamic and geometrical conditions, and also the CFL thickness is strongly influenced by the RBC deformability.

Upload: others

Post on 07-Dec-2021

6 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: A Microfluidic Device for Partial Cell Separation and

1_______________________________________________________________________________________________“NOTICE: this is the author’s version of a work that was accepted for publication in BioChip Journal. Changes resulting from the publishing process, such as peer review, editing, corrections, structural formatting, and other quality control mechanisms may not be reflected in this document. Changes may have been made to this work  since  it was  submitted  for  publication.  A  definitive  version will  be  subsequently  published  in BioChip Journal, [Vol.7(4), pp.367‐374, (2013)].” _______________________________________________________________________________________              

A Microfluidic Device for Partial Cell Separation and Deformability Assessment

Diana Pinho1, 2, Tomoko Yaginuma1 & Rui Lima*1, 2

1 Polytechnic Institute of Bragança, ESTiG/IPB, C. Sta. Apolonia, 5301-857 Bragança, Portugal. 2 CEFT, Faculdade de Engenharia da Universidade do Porto (FEUP), R. Dr. Roberto Frias, 4200-465 Porto, Portugal. * Correspondence and requests for materials should be addressed to R. Lima ([email protected])

Abstract Blood flow in microcirculation shows several interesting phenomena that can be used to develop microfluidic devices for blood separation and analysis in continuous flow. In this study we present a novel continuous microfluidic device for partial extraction of red blood cells (RBCs) and subsequent measurement of RBC deformability. For this purpose, we use polydimethylsiloxane (PDMS) microchannels having different constrictions (25%, 50% and 75%) to investigate their effect on the cell-free layer (CFL) thickness and separation efficiency. By using a combination of image analysis techniques we are able to automatically measure the CFL width before and after an artificial constriction. The results suggest that the CFL width increases with enhancement of the constriction and contributes to partial cell separation. The subsequent measurements of RBCs deformation index reveal that the degree of deformation depends on the constriction geometries and hematocrit after the cell separation module. The proposed microfluidic device can be easily transformed into a simple, inexpensive and convenient clinical tool able to perform both RBC separation and deformability analysis in one single device. This would eliminate the need for external sample handling and thus reducing associated labor costs and potential human errors. Keywords: Biomicrofluidics, Microfluidic devices, Microcirculation, Blood on chips, Red blood cells, Cell separation, Cell deformability, Deformation index.

Introduction

Cell separation and identification are essential in a variety of biomedical applications including cell biology, diagnostic and therapeutic methods. Blood is a non-Newtonian fluid containing extremely rich amount of information about the physiological and pathological state of the human body. However, due to its complexity there are few accurate analysis methods. Most of the standard techniques used cell separation and sorting are often labor intensive or require additional external labels to identify cells.

Blood flow in microcirculation shows several interesting phenomena that can be used to develop microfluidic devices for blood separation and analysis in continuous flow. These phenomena include plasma skimming [1], cell-free layer (CFL) [2, 3, 4], leukocyte margination [5] and bifurcation law [6]. Recently, several researchers have used these effects and replicated in microfluidic systems. In microchannels RBCs, due to their deformability and lift forces, tend to be concentrated around the center of the microchannels while white blood cells (WBCs) and rigid RBCs (such as Malaria RBCs) tend to migrate to the CFL originated at regions close to the walls [1-5]. Bifurcation law [6] states that RBCs behavior, in microchannels with bifurcations, tends to be oriented to the wide microchannel. A number of microfluidic devices have been developed to take advantage of these natural hemodynamics phenomena. Shevkoplyas et al. [7] developed a microdevice to isolate WBCs from a blood sample by using the margination effect, whereas Hou et al. [8] have very recently proposed a biomimetic separation device to separate normal RBCs from malaria infected RBCs. Other researchers have found several advantages to control and manipulate blood flow in microfluidic devices. Fujiwara et al. [9] have found evidence that it is possible to create an artificial CFL under appropriate hemodynamic and geometrical conditions, and also the CFL thickness is strongly influenced by the RBC deformability.

Page 2: A Microfluidic Device for Partial Cell Separation and

2

Recently, Ishikawa et al. [10] and Leble et al. [11] have shown the existence of thin CFL in the centre of the microchannel, just downstream of a confluence. This phenomenon is due to the existence of a CFL in both inner walls and a consequent formation of a triangular CFL in the region of the confluence apex [10, 11]. Faivre et al. [12] and Sollier et al. [13] have demonstrated that the CFL could be enhanced by using a microchannel containing a constriction followed by sudden expansion to separate plasma from the whole in vitro blood. However, most of these studies aim at the complete extraction of cells from plasma, which is not the case of the present study.

The main objective of the proposed microfluidic device is to separate certain amount of RBCs from plasma and then measure the deformability of individual RBCs downstream in one single step. RBC deformability is important in a clinical sense as it is related to several diseases such as diabetes, malaria, as well as cardiovascular disorders [14]. These diseases are at times fatal and early detection is crucially preferable. In this sense, RBC deformability can be a new biomarker and the fast and easy measurement methods are required. Former studies on RBC deformability have introduced some methodologies such as micropipette aspiration and optical tweezers where basically RBCs are stretched and the pressure or force for their extension is measured [15, 16]. Whilst these studies have revealed useful RBCs mechanical properties, these methods reqire large amount of preparation and are very time consuming. For fast analysis of vast amount of blood samples, these traditional measurement methods are unlikely to be appropriate. Moreover, since the preceded cell separation process is needed, the microfluidic devices which are easy to manipulate their microchannel geometries are more suitable for our purposes. A typical microdevice for RBC deformation studies uses a microchannel having a shape of sudden constriction in order to elongate cells. Zhao et al. [17] used a straight microchannel with a sudden narrowing and expanding constriction. Lee et al. [18], Yaginuma et al. [19-20] and Faustino et al. [21], on the other hand, used microchannels with a hyperbolic shape contraction followed by a sudden expansion region. These microfluidic experiments showed high deformability of RBCs when travelling through a contraction region. However, those studies did not include the preceded cell separation process as they have solely performed deformability measurements after an additional sample preparation [19-21].

The proposed microfluidic device aims to obtain a CFL with a low enough RBC concentration to perform cell deformability measurements downstream the separation constriction. This study is divided in two main parts. Firstly, a simple microfluidic device with different constrictions (25%, 50% and 75%) is used to test the RBCs separation. Secondly, a more complex device, able to perform in a single step both RBC separation and deformability analysis, was tested. This first tentative to integrate in one single device both tasks of separation and deformation have shown not only the viability of this new clinical strategy but also new findings that will be crucial to optimize the design for this kind of microfluidic device.

Materials and methods Working fluids The working fluid used in this study was dextran 40 (Dx40) containing about 9% (i.e. Hematocrit, Hct = 9) by volume of human RBCs. More details can be found in Supplementary Materials. Microdevice geometries The microfluidic devices tested in this study were fabricated using a soft lithography technique and consist of two main parts: a cell separation region and a cell deformation region. The microchannel height was measured by a profilometer to be 51 m.

Fabrication of the microfluidic devices The polydimethylsiloxane (PDMS) rectangular microchannels were fabricated using a soft lithographic technique. A detailed description of the fabrication process can be found elsewhere [22, 23]. Briefly, the microchannel geometry was drawn using Autocad, and a high resolution photomask was manufactured. The solid master was then fabricated on a silicon wafer with an ultrathick photoresist (SU-8 50; Kayaku MicroChem, Japan). The PDMS prepolymer was prepared by mixing a comercial prepolymer and catalyzer (Silpot 184; Dow Corning, USA) at a weight ratio of 10:1. After the mixture was degassed under vacuum, the PDMS was poured into the SU-8 photo-resist master mold and cured by baking for about 2h at 70°C. Both master and PDMS were cooled to room

Page 3: A Microfluidic Device for Partial Cell Separation and

temperatmicro-pislide gla Cell sepaTo studydifferentof a straiand 25 test the R Cell sepaTo invesmicroflutested. Ndevice hdifferentdimensio

Figure ExperimeIn this ssystem a ContractiIn this st( ) of tdifferentwas defi

Note tharespectiv

To analyconstrictconstrictupstream

ture and the ipette tips. Fass, where a r

aration(CS) my the effect ot constrictionight microchm wide (RBCs separa

aration and destigate the abuidic device hNote that thihad feed mict kind of coons can be se

e 1. Schemati

ental set-up study we havand the imag

tion ratio, CFLtudy we usedthe constritiot artificial coined as follow

at, the ofvely.

yze the CFLtions, as shotion, and the

m and downs

PDMS was Finally, the Preversible sea

microfluidic deof a single cn sizes was dhannel with 1

). Figure 6aation.

eformation (Cbility to perfohaving a cels latter regiocrochannel onstrictions teen in Figure

ic view of th

ve used a hige analysis pr

FL thickness ad microchannons varying fonstrictions (ws:

f the constric

L thickness, ow in FigureCFL to thetream of the

peeled from PDMS was wal formed sp

evice constriction odeveloped. T100 m wide shows the m

CSD) microfluorm in a singl separation on has a conof 100 m (o perform Re 1 in supplem

e contraction

gh-speed vidrocedure are

nd deformationels with thefrom 25 m, 25%, 50% a

ctions 25 m

measuremere 7. CFLe thickness dconstriction

m the master. washed with pontaneously

on the cell sThis first deve ( ) with dmicrofluidic

uidic device gle step both region followncentration o

), a constRBC deformmentary mat

n-expansion

deo microscoexplained in

on index (DI)e width ( ) 50 m, up t

and 75%) on

m, 50 m an

ents were takcorrespondsdownstream

n was given b

The input/ouethanol and

y.

separation, avice had onlydifferent consdevice with

RBC separawed by an oof cells lowetriction of 50

mability. Morterials.

geometry, id

opy system. n supplement

) of the feed

to 75 m. Tothe CFL thi

(1)

nd 75 m co

ken upstreams to the thic

the constricby:

(2)

utput ports abrought into

a simple micy a separationstriction regithe different

ation and defoutlet cell defer than the f0 % ( re detailed in

dentifying the

The detailstary material

channel of 1o examine theckness, the c

orresponds to

m and downckness of thection. The di

are made by o contact wit

crofluidic devn region thations of 75 mt contraction

formability anformation refeed Hct. Th50m) foll

nformation a

he relevant va

s of the expels.

100 m and te effect of thcontraction r

o 0.25, 0.5 a

nstream the he CFL upstdifference of

3

means of th a clean

vice with t consists m, 50 m ns used to

nalysis, a egion was his tested lowed by about the

ariables.

erimental

the width hese three ratio ( )

and 0.75,

artificial tream the

the CFL

Page 4: A Microfluidic Device for Partial Cell Separation and

For char

Note thalength er(≪ 1µm

Results CS microIn the Cconsequeand 0.75

QualiFurthermTo obtaithicknesrates (se

Figure 2flow rater = 0.75

Overathat all tclear tha0.25 wheshow a tresults otendency

racterizing de

at Lmajor and rror which m

m) throughout

s

ofluidic deviceCS device wently on the

5), at a Hct ofitative flow

more, it is clein more dets were takene Figure 2).

2. CFL thickes (1 μl/min,5).

all the quantthe constrict

at the CFL ere CFL is tendency of

obtained usiny is more pro

eformability

Lminor are thmight be caust the measure

e: constrictionwe have ev separation ef 9% and wit

w visualizatiear that the etail results an both upstre

kness for both, 5 μl/min an

titative meastions tested increases wabout three the CFL thi

ng a microchaonounced for

of RBCs, we

he major andsed by the caed regions so

n effect in thevaluated the effectivenessth three diffeon results

enhancementabout the conam ( CFL ) a

h upstream (nd 10 μl/min

surements ofenhance the

with decreasinorders of maickness to deannel with sir r = 0.5 an

e use deform

d minor axisamera’s expoo that these p

e CFL thickneeffect of t

s. This studyerent flow ratshow that

t is more pronstriction efand downstre

( CFL ) and n) and differe

f the CFL the CFL thicknng r and thagnitude greaecrease with imilar constrnd 0.75.

mation index

(3)

lengths of tosure time bupossible devi

ess the constricty was perfortes (1 μl/minall constric

onounced for ffect, quantiteam CFL

downstream ent constricti

hickness, preness at downhis enhancemater than theincreasing f

rictions and a

(DI) which i

the RBC. Wut the values ation was ign

tion on the rmed with din, 5 μl/min anctions enhanthe microch

tative measuthe constrict

CFL conion ratios (r

esented in Finstream regiment is moreCFL . Addi

flow rate, wha Hct of 2.6%

is defined as

We also calcuwere sufficinored.

CFL thickdifferent (nd 10 μl/minnce CFL t

hannel with rurements of tions at diffe

nstrictions at = 0.25, r

igure 2, showion CFL . e pronounceditionally, thehich corrobo

% [12]. Note

4

follows.

ulated the ently low

ness and (0.25, 0.5 n). thickness. r = 0.25. the CFL

erent flow

different =0.5 and

w clearly It is also d for r = se results orates the that, this

Page 5: A Microfluidic Device for Partial Cell Separation and

CSD micDownstrmicrochadifferentwere tesupstreamused.

The Hthem stamicrochato be smaxial DI contractisudden nparachutFigure 3shear strvariable meaningideal forhas to be

Figurfor defortype of cmore dogood imthis cond

Figure 3contracti

crofluidic deviream the Cannel as shot constrictionsted in orde

m constrictio

Hct at this reart becomingannel with a

maller of the values of on

ion entrance narrowing ofte, reaching 3a.) The sudress. Howevand the def

g for the ellipr deformatione carefully do

re 3b shows rmability meconstriction,

ominant overmage analysis

dition is idea

3. DI valuesion.

ice: constrictiCS region, Rown in Figun regions, oner to examinn (CS region

egion shouldg like a parac

width of 7 normal size ne individua(x/Lc=0), th

f the flow fia low DI vaden contract

ver, proceedefinition of mpsoid shape. n measuremeone in the up

the DI valueeasurement. in the middl

r shear force,s results it isal.

s of a single

ion effect andRBC deformure 1 in supne with a sune efficient n) with =

d be around 2chute or umbm. This RBof human R

al RBC flowihe highest Deld. Then thalue at the etion is effected DI valueajor and minBesides this

ents especialpper stream.

es of one sinIt is clear th

le of the mic, where RBCs important t

e RBC flow

d deformationmation meapplementary

udden contracdeformabilit0.5, an inlet

2.4% and Rbrella shape w

BC deformatiRBC (8 m) a

ing through DI was obtainhe RBC startexit of the cotive to obser

es are slightlnor axis of ts, overlaps olly when an

ngle RBC flohat the highcrochannel (yCs tend to floto measure R

wing through

measuremensurement rematerials. I

ction and anty measuremt Hct of 9%,

RBCs tend towhen they pon behaviouand to the shthe sudden c

ned due to a ts changing tontraction (xrve a strong ly unreliablethe cell in thof the cells wautomatic m

owing throughest DI is oby=0 and z=0ow stably wiRBCs flowin

h (a) a sudde

nts egion is folIn this study

nother with aments. For th, and a flow

o flow in lineass through ar is mainly d

hear flow. Ficonstriction strong shear

their shape fx/Lc=1, see

deformatione as parachuhis shape doewere often ob

method is app

gh the smooserved aroun

0) extensionaith no rotatio

ng in the sam

en contractio

llowed in ty, two geoma smooth conthis examina

rate of 1 μl

e and the maa sudden con

due to the conigure 3a showmicrochanner stress causefrom an ellipalso the iman of RBCs wute shapes aes not have bserved, whi

plied. The Hc

oth contractiond x/Lc= 0.

al force is likon or orientame orientatio

on and (b) a

5

the same metrically ntraction, ation, the /min was

ajority of nstriction nstriction ws the x-el. At the ed by the

psoid to a age IV in with high are rather the same ich is not ct control

on region 5. In this

kely to be ation. For on so that

a smooth

Page 6: A Microfluidic Device for Partial Cell Separation and

Figure 4flow rate

Figure 5measureconfiden

DiscussTo analythe ∆increasesis about worth mFor r =higher flCFL thic

Figurfor r = remaininimage anconsider

4. The differes and differe

5. Separationd values are

nce interval.

sion yse in more d for each cos with decreatwo to four o

mentioning th 0.5 and r =low rates. Hockness (see Fre 5 shows th 0.5 this co

ng 27% of cnalysis approred that all th

rence of the ent constricti

n rate of the expressed as

detail the effeonstriction aasing the rorders of maat the highes= 0.25 for loowever theseFigure 5). he separationonstriction ccells to the doach used to he white par

CFL upstreaion ratios (r

RBCs flowis the mean ±

fect of the arts a function and this en

agnitude greast ∆ corow flow ratee results show

n rate of cellscan efficientldeformation calculate the

rt, upstream

am and down = 0.25, r

ing to the ou

± standard de

tificial constrn of the flownhancement iater than ∆rresponds toe there is a tw evidence t

s flowing to tly separate region. This

e CFL thicknthe constrict

nstream of th0.5 and r 0

utlet region neviation acco

rictions on thw rate. Theseis more pron

for r = 0.r = 0.25 antendency to that the const

the outlet regabout 73%

s separation ness. Briefly,tion, corresp

he constrictio0.75).

number 2 (upording to a t-t

he CFL thicke results shoounced for r5 and r = 0.

nd is independecrease butriction has a

gion numberof blood ceestimation i, we have usonds to a un

on (∆ ) at

pper side outest analysis

kness, Figureow clearly thr = 0.25 wh.75, respectiv

ndent of the fut tend to flaa strong imp

r 2 (see Figurells to regiois based on sed binary imniform distri

6

different

tlet). The at a 95%

e 4 shows hat ∆

here ∆ vely. It is flow rate. attened to act in the

re 6), i. e., on 2 and the same

mages and bution of

Page 7: A Microfluidic Device for Partial Cell Separation and

the RBCwere ablconstrictshown aHence, wr = 0.5decided single stfeed Hctrecently,contractirate shouthe geomlocal Hc

Figure 6separatiom, respof deformThe heig

Figure 7contracti

Cs core to bele to estimattion with higa significant we believe th5 is by decreto use the coep both sepat to be abou, by using ions [20]. Inuld be arounmetrical paracts less than 1

6. Schematicon part. The pectively. Thmation meas

ght of the mic

7. Mean DI ions.

e 100%. Hente the separagher efficiendifference b

hat the most easing the geonstriction waration and dut 9%, the lo1% Hct we n order to a

nd 90% and ameter L2, to1%.

c drawing of dimensions

he different csurement parcrochannel i

values of tw

nce, by measation rate depncy was the between r =effective wa

eometrical pawith r = 0.5 tdeformation oocal Hct in t

have succechieve 1% Hconsequentl

o a dimensio

f the microfluof L , L , w

constrictions rt. The dimes 51 µm.

wo deformatio

suring the CFpending on one with r

= 0.25 and ay to get a coarameter L2

to test the abof blood cellthe outlet 3

essfully meaHct with thely r must beon less than

uidic device.w , w , w ,

were used (ensions of w

on measurem

FL thicknessthe size of tr = 0.25. Ar = 0.75, tonsiderable d(see Figure

bility of the ps. Note that (deformatio

asured the Re proposed me less than 0100 µm, it w

. (a) the entirare 300 m,

= 75 mw and w are

ment module

s downstreamhe constricti

Although statthis is not thdifference be6). In the pr

proposed devfor r = 0.5 n region) is

RBC deformamicrofluidic d0.25. Alternawill be possi

re view and 100 m, 10

, 50 m and e 7 m and

es: (a) a sudd

m the constriion. As expetistical analy

he case for retween r = resent study

vice to perfor and by consaround 2.4%

mability in hydevice the s

atively, by deible to achie

close-up vie00 m, 80 m

25 m). (b)10 m, resp

den and (b)

7

iction we ected, the ysis have r = 0.5. 0.25 and we have

rm in one sidering a %. Very yperbolic eparation ecreasing

eve easily

ew of cell m and 20 the view

pectively.

a smooth

Page 8: A Microfluidic Device for Partial Cell Separation and

8

In terms of the deformation measurements, Figure 7 shows average DI values of ten RBCs flowing through the two different constriction regions. The maximum DI value is in both cases relatively the same, however, for the sudden contraction case (a) the maximum DI is located close to the constriction entrance whereas for the smooth case (b) the maximum DI is located around the middle of the constriction. In addition, in the constricted region in (a) RBCs tend to flow as parachute like shape and the DI values are not truly comparative in the sequence of cell behaviour throughout the deformation analysis region of the device. At least, the region for measurements has to be carefully selected for a meaningful comparison. On the other hand, the smooth contraction provides less deviation of DI results and indicates clearly an appropriate measurement region (x/Lc=0.5), i. e., middle of constriction. The majority of the cells are flowing with the same orientation at the centreline of the channel where the cells are under an extensional flow dominated regime and as a result they tend to elongate in y direction induced mainly by this extensional force. This stable state of the RBCs makes this geometry suitable to measure small changes of RBC deformability. More detailed studies on RBCs extensional flow effects can be found elsewhere [19-21].

As mentioned before, by using a constriction with r = 0.5, we have observed collisions and the overlap of neighbouring cells flowing within the measurement region. In the CS region, the Hct of the flowing fluid is reduced from 9% to ≈2.4% but ideally it needs to be less than that. Nevertheless, it is worth mentioning that by reducing Hct to less than 2% the number of RBCs to measure may not be large enough to obtain a significant statistical picture of the results. Therefore, an optimization of channel, i.e. change in depth, needs to be done in order to obtain the best representative results of the RBCs DI.

Conclusions The conventional microfluidic methods for measuring RBC deformability are often labor intensive and require additional sample modification and preparation. In this paper, we present a new, simple microfluidic device able to perform both RBCs separation and deformability assessment in one single step. In general, our results indicate that the proposed device can perform both operations (separation and deformation) successfully. The reported results show evidence that the constriction has a strong impact on the CFL thickness and consequently on separation rate. Moreover, the deformability results show clearly that most appropriate geometry to measure RBC deformability is the microchannel containing a smooth constriction region. In this kind of geometry due to the existence of a dominant extensional force the majority of the RBCs tend to flow with the same orientation. This stable performance of the RBCs may prove to be enough sensitive to detect small changes of RBC deformability and thus it may have the ability to diagnose early stage RBC related diseases such as diabetes, malaria and sickle cell anemia. Therefore, the integrated and simple continuous system operations make the proposed microfluidic device a potential diagnostic technique to be applied to both healthy cells and blood cell diseases. Acknowledgement The authors acknowledge the financial support provided by 2007 Global COE Program “Global Nano-Biomedical Engineering Education and Research Network”, Japan and grant-in-Aid for Science and Technology PTDC/SAU-BEB/108728/2008, PTDC/SAU-BEB/105650/2008, PTDC/EME-MFE/099109/2008, PTDC/SAU-ENB/116929/2010 and scholarship SFRH/BD/89077 /2012 from FCT (Science and Technology Foundation), COMPETE, QREN and European Union (FEDER). The authors are also very grateful to Dr. Mónica Oliveira (Strathclyde University), Professor Takuji Ishikawa and Professor Takami Yamaguchi (Tohoku University) for their suggestions and support to this research work.

References

[1] Caro, C., Pedley, T., Schroter, R. & Seed, W. The Mechanics of the Circulation. Oxford University Press, 1978.

[2] Lima, R. et al. Measurement of individual red blood cell motions under high hematocrit conditions using a confocal micro-PTV system. Ann Biomed Eng. 37, 1546-59, (2009).

[3] Fujiwara, H., et al. Red blood cell motions in high-hematocrit blood flowing through a stenosed microchannel. J Biomech 42, 838-843, (2009).

Page 9: A Microfluidic Device for Partial Cell Separation and

9

[4] Lima, R., et al. Axisymmetric polydimethysiloxane microchannels for in vitro hemodynamic studies. Biofabrication, 1: 1-7, (2009).

[5] Goldsmith, H., & Spain, S. Margination of leukocytes in blood flow through small tubes. Microvasc Res, 27, 2, 204-222, (1984).

[6] Yang, S., Undar, A., & Zahn, J.D. A microfluidic device for continuous, real time blood plasma separation. Lab Chip 6:871, (2006).

[7] Shevkoplyas, S.S., Yoshida, T., Munn, L.L., & Bitensky, M.W. Biomimetic Autoseparation of Leukocytes from Whole Blood in a Microfluidic Device. Anal Chem, 77, 933-937, (2005).

[8] Hou, H.W., et al. Deformability based cell margination - A simple microfluidic design for malaria-infected erythrocyte separation. Lab Chip, 10, 2605–2613, (2010).

[9] Fujiwara, H., et al. Red blood cell motions in high-hematocrit blood flowing through a stenosed microchannel. J Biomech 42, 838-843, (2009).

[10] Ishikawa, T., et al. Asymmetry of blood flow and cancer cell adhesion in a microchannel with symmetric bifurcation and confluence. Biomed Microdevices, 13, 159, (2011).

[11] Leble, V., et al. Asymmetry of red blood cell motions in a microchannel with a diverging and converging bifurcation. Biomicrofluidics, 5, 044120, (2011).

[12] Faivre, M., Abkarian, M., Bickraj, K. & Stone, H. Geometrical focusing of cells in a microfluidic device: an approach to separate blood plasma. Biorheology, 43, 147-159, (2006).

[13] Sollier, E., Cubizolles, M., Fouillet, Y. & Achard J. Fast and continuous plasma extraction from whole human blood based on expanding cell-free layer devices. Biomed Microdevices, 12, 485-97, (2010).

[14] Mokken, F.C., Kedaria, M., Henny, C.P., Hardeman, M.R., & Gelb, A.W. The clinical importance of erythtrocyte deformability, a hemorrheological parameter. Ann Hematol, 64, 113-122, (1992).

[15] Musielak, M. Red blood cell deformability measurement: Review of techniques. Clin Hemorheol Micro, 42, 47–64, (2009).

[16] Shiga, T., Maeda, N., & Kon, K. Erythrocyte rheology. Crit Rev Oncol Hemat, 10: 9-48, (1990).

[17] Zhao, R., et al. Microscopic investigation of erythrocyte deformation dynamics. Biorheology, 43(6), 747-765, (2006).

[18] Lee, S.S., Yim, Y., Ahn, K.H., & Lee, S.J. Extensional flow-based assessment of red blood cell deformability using hyperbolic converging microchannel. Biomed Microdevices, 11, 1021-1027, (2009).

[19] Yaginuma, T., Oliveira, M.S.N., Lima, R., Ishikawa, T., & Yamaguchi, T. Red blood cell deformation in flows through a PDMS hyperbolic microchannel, Proceedings of NSTI-Nanotech, 2 505-507, (2011).

[20] Yaginuma, T., Oliveira, M.S.N., Lima, R., Ishikawa, T., & Yamaguchi, T. Human red blood cell behavior under homogeneous extensional flow in a hyperbolic-shaped microchannel. Biomicrofluidics, 7:054110, (2013).

[21] Faustino V., Pinho D., Yaginuma T., R. Calhelha, M. Oliveira, I. Ferreira, Lima R., Flow of red blood cells suspensions through hyperbolic microcontractions. Visualization and simulation of complex flows in biomedical engineering. In: R.Lima, T. Ishikawa, Y. Imai & M. S. N. Oliveira (Eds), Springer, 2013 (in press).

[22] Lima, R., Wada, S., Tanaka, S., Takeda, M., Ishikawa, T., Tsubota, K., Imai, Y., Yamaguchi, T., In vitro blood flow in a rectangular PDMS microchannel: experimental observations using a confocal micro-PIV system. Biomedical Microdevices, 10: 153-167, 2008.

[23] Lima, R., et al. Microscale flow dynamics of red blood cells in microchannels: an experimental and numerical analysis. In: Tavares and Jorge (Eds) Computational Vision and Medical Image Processing: Recent Trends, Springer 19, 297-309, (2011).

[24] Abramoff, M., Magelhaes, P. & Ram, S. Image Processing with ImageJ. Biophotonics International, 7, 11, 36-42, (2004).

Page 10: A Microfluidic Device for Partial Cell Separation and

10

Supplementary Materials

1. MATERIALS AND METHODS

1.1. Working fluids

The working fluid used in this study was dextran 40 (Dx40) containing about 9% (i.e. Hematocrit

(Hct = 9%). Briefly, blood was collected from a healthy adult volunteer and heparin was added in

order to prevent coagulation. The red blood cells (RBCs) were separated from bulk blood by

centrifugation and aspiration of the plasma and buffy coat. The RBCs were then washed twice with a

physiological saline solution and diluted with Dx40 to make up the required RBC concentration. All

blood samples were stored hermetically at 4ºC until the experiments were performed at controlled

temperature of approximately 37ºC.

1.2. Fabrication of the microfluidic devices

The polydimethylsiloxane (PDMS) rectangular microchannels were fabricated using a soft

lithographic technique. A detailed description of the fabrication process can be found elsewhere [22,

23]. Briefly, the microchannel geometry was drawn using Autocad, and a high resolution photomask

was manufactured. The solid master was then fabricated on a silicon wafer with an ultrathick

photoresist (SU-8 50; Kayaku MicroChem, Japan). The PDMS prepolymer was prepared by mixing a

comercial prepolymer and catalyzer (Silpot 184; Dow Corning, USA) at a weight ratio of 10:1. After

the mixture was degassed under vacuum, the PDMS was poured into the SU-8 photo-resist master

mold and cured by baking for about 2h at 70°C. Both master and PDMS were cooled to room

temperature and the PDMS was peeled from the master. The input/output ports are made by means of

micro-pipette tips. Finally, the PDMS was washed with ethanol and brought into contact with a clean

slide glass, where a reversible seal formed spontaneously.

1.3. Experimental set-up

The high-speed video microscopy system used in this study is shown in Figure 1. This system

consisted mainly by an inverted microscope (IX71; Olympus, Japan) combined with a high-speed

camera (Phantom v7.1; Vision Research, USA). All the microfluidic devices were placed on the stage

of the inverted microscope and by using a syringe pump (KD Scientific, USA), a constant pressure-

driven was maintained. Additionally, a thermo plate controller (Tokai Hit) was set to 37ºC in order to

have an environment close to in vivo conditions.

Page 11: A Microfluidic Device for Partial Cell Separation and

Figure devices.

1.4. Ima

Cell sepa

All imag

pixels, a

with the

L/min.

[24].

Figure 21 L/min

First, the

with ma

function

region c

converte

1. High-spe

age analysis

aration (CS)

ges were cap

at a rate of 80

flow of RBC

The record

2. Original imn.

e captured v

aximum inten

n, which resu

corresponds t

ed to binary i

ed video m

) microfluidi

ptured around

000 frames/s

Cs (halogen

ded images w

mage with th

videos were

nsity of all

ults in a sing

to RBCs cor

images with

icroscopy sy

ic device

d the middle

s and an expo

illumination

were transferr

he flow of RB

converted to

the images

le image hav

re. To obtain

thresholding

ystem used

e plane of the

osure time o

n) through a 2

rred to the co

BCs through

o a sequence

in the stack

ving a region

n quantitativ

g (see Figure

to test the

e microchann

f 0.125 ms. F

25% constric

omputer and

a 75% const

e of static im

k was select

n brighter th

ve measurem

3).

performance

nels with a re

Figure 2 sho

ction for a co

then evaluat

triction for a

mages (stack

ed by using

han the backg

ments the gre

e of the mic

esolution of

ows an origin

onstant flow

ated in Image

a constant flo

k). Then, eac

g ImageJ “Z

ground. This

ey scale ima

11

crofluidic

800304

nal image

w rate of 1

e J (NIH)

ow rate of

ch pixel’s

Z project”

s brighter

ages were

Page 12: A Microfluidic Device for Partial Cell Separation and

Figure 3The chan

Cell sepa

RBC de

contracti

the regio

flowing

Figure 4 Deforma

Deforma

where Lm

like shap

3. Binary imnnel dimensi

aration and

eformation a

ion region o

on. Ten cell

cells were m

4. Schematic

ation Index

ation Index (

major and Lmin

pe RBCs, we

mage obtaineions are: w1 =

deformation

analysis has

f the microc

s were selec

measured man

c diagram of

DI) is define

nor are the ma

e measured th

ed after imag= 100 m an

n (CSD) micr

been achiev

hannel (W5

cted for each

nually frame

RBC deform

ed as:

ajor and mino

he major and

ge processingnd w2 = 25 m

rofluidic dev

ved by mea

and W6 in Fi

h contraction

e by frame.

mation measu

or axis lengt

d minor axis

g and binarim.

vice

asuring RBC

igure 4), inc

ns, and majo

urement regio

hs of the RB

lengths as sh

zation using

C deformatio

luding upstre

or and mino

ons with deta

BC (see Figur

hown in Figu

g “Image J”

on index (D

ream and dow

or axis length

ailed dimens

re 5a). For p

ure 5b.

12

software.

DI) at the

wnstream

hs of the

sions.

arachute-

Page 13: A Microfluidic Device for Partial Cell Separation and

Figure 5measurethe majo The mea

(NIH) [2

The 2D

opened a

flowing

by the R

RBC’s a

it by the

found to

deviation

2. RESUFigure 6

measure

Figure 6side) for

Figure 7

device u

5. Schematicment of para

or and minor

asurements w

24]. We also

image cont

and this shad

direction (y)

RBC’s actual

actual velocit

e frame inter

o be suffici

ns were igno

ULTS 6 shows repr

ments obtain

6. Original br different art

7 shows a re

upstream the

c view of DIachute shapeaxes are def

were done m

calculated th

ains the flow

dow should

). To avoid t

velocity tim

ty, we measu

rval time. Th

iently low (

ored.

resentative o

ned at differe

blood flow vtificial constr

epresentative

separation re

I measuremee cells. The bfined.

manually wit

he length err

wing cell’s s

not be consi

this error the

mes the expos

ured the trav

he cell’s trav

(≪ 1µm) th

original imag

ent artificial c

isualization rictions (r =

e original im

egion with r

ents. (a) the box surround

th the assista

ror which mi

shadow whi

idered as the

e cell’s trave

sure time of

veling length

veling length

hroughout th

ges and corre

constrictions

images (left = 0.25, r = 0

mage of the

= 0.5.

measuremending the cell

ance of an i

ight be cause

ich is captur

e cell’s actua

ling length (

f the high spe

h of the cells

h values obt

he measured

espondent bi

s (r = 0.25, r

side) and co0.5 and r = 0

deformation

nt of ellipsoidefines the c

mage handli

ed by the cam

red when the

al length esp

(shadow) in y

eed camera. I

frame by fra

tained by the

d regions so

inary images

r = 0.5 and

orrespondent0.75).

region of th

id shape cellcenter of the

ing software

mera’s expos

e camera’s s

pecially in th

y-axis was c

In order to o

ame and then

ese calculati

o that these

s from in vit

r = 0.75).

t binary imag

he CSD mic

13

s. (b) the e cell and

e, ImageJ

sure time.

shutter is

he axis of

calculated

obtain the

n divided

ons were

possible

tro blood

ges (right

crofluidic

Page 14: A Microfluidic Device for Partial Cell Separation and

Figure 7containin

Figure 8

The Hct

the cell c

Figure 8containin

7. Blood flowng a sudden

8 shows a rep

at this regio

can be observ

8. Blood flowng a smooth

w visualizaticonstriction

presentative

on should be

ved at the na

w visualizaticonstriction

ion images omicrochanne

original ima

e also around

arrowest part

ion images omicrochann

of the deformel with a wid

age of the de

d 2.4%. In th

t of the contr

of the deformnel with a wid

mation regiondth of 7 m.

eformation r

his contractio

raction.

mation regiondth of 10 m

n of the CSD

egion with a

on, the maxi

n of the CSDm.

D microfluid

a smooth con

imum deform

D microfluid

14

dic device

ntraction.

mation of

dic device