Journal of the Optical Society of America and Review of Scientific Instruments Vol. 7 MAY, 1923 Number 5 A CHART OF THE FLICKER PHOTOMETER BY HERBERT E. IVES INTRODUCTION In various investigations of the phenomena of critical frequency and of the flicker photometer the writer has obtained a considerable body of experimental data. Part of this was obtained primarily in order to have information as to the behavior of the flicker photometer, irrespective of any theory as to its action, and part was obtained in order to test theoretical speculations. Taken together the data con- stitute a picture of the behavior of the flicker photometer probably more complete than any purely experimental study would be likely to afford. It is the purpose of this paper to assemble the data from the writer's various papers, presenting them in the form of a single chart. This chart may be considered either as an exposition of the character- istics of the flicker photometer sufficient for practical purposes, or as the body of facts which any theory of the instrument must explain. NATURE OF THE CHART The chart of .the flicker photometer consists of the extension to all working intensities of the diagram shown in Fig. 1, which is reproduced from an earlier paper,' which should be consulted for a full description of its significance. It shows the critical frequencies of disappearance of flicker for all ratios of illumination of the two alternating elements of the flicker field, starting with illumination of one element alone (e.g. by red light) and extending through all intermediate ratios to the illumination of the other element alone (e.g. by green light) under the condition that the sum of the two illuminations shall be constant. The IA Polarization Flicker Photometer. Ives. Phil. Mag. April 1917, p. 360. 363

Upload: herbert-e

Post on 02-Oct-2016

215 views

Category:

Documents


1 download

TRANSCRIPT

Page 1: A CHART OF THE FLICKER PHOTOMETER

Journalof the

Optical Society of Americaand

Review of Scientific InstrumentsVol. 7 MAY, 1923 Number 5

A CHART OF THE FLICKER PHOTOMETER

BY HERBERT E. IVES

INTRODUCTION

In various investigations of the phenomena of critical frequencyand of the flicker photometer the writer has obtained a considerablebody of experimental data. Part of this was obtained primarily inorder to have information as to the behavior of the flicker photometer,irrespective of any theory as to its action, and part was obtained inorder to test theoretical speculations. Taken together the data con-stitute a picture of the behavior of the flicker photometer probablymore complete than any purely experimental study would be likely toafford. It is the purpose of this paper to assemble the data from thewriter's various papers, presenting them in the form of a single chart.This chart may be considered either as an exposition of the character-istics of the flicker photometer sufficient for practical purposes, or asthe body of facts which any theory of the instrument must explain.

NATURE OF THE CHART

The chart of .the flicker photometer consists of the extension to allworking intensities of the diagram shown in Fig. 1, which is reproducedfrom an earlier paper,' which should be consulted for a full descriptionof its significance. It shows the critical frequencies of disappearance offlicker for all ratios of illumination of the two alternating elementsof the flicker field, starting with illumination of one element alone(e.g. by red light) and extending through all intermediate ratios to theillumination of the other element alone (e.g. by green light) under thecondition that the sum of the two illuminations shall be constant. The

IA Polarization Flicker Photometer. Ives. Phil. Mag. April 1917, p. 360.

363

Page 2: A CHART OF THE FLICKER PHOTOMETER

[J.O.S.A. & R.S.I., 7

amounts of the two illuminations corresponding to any critical speedare shown in the lower diagram. Practically such a variation of rela-tive illumination could be achieved by using two extended bright sur-faces as the light sources, and arranging a variable diaphragm overeach, the diaphragms being so connected that as one is opened theother is closed.

Examining Fig. 1-in which we shall confine our attention to thecurved lines marked by circles-we see that at either end a high speed

Id 10~~~~~~~~~~~~~~~~~~~~~I

R zo 18 16 1.4 12 10 8 5 4 *2 0G 0 2 .4 *6 * 1.0 I 14 1 6 1.8 20

FIG. 1. Critical speeds for disappearance of flicker for various relative amounts of alter-

nated red and green light.

of alternation is called for to abolish flicker; as the central portion of

the diagram is approached the critical speed decreases, the two branchesof the critical frequency curve descending toward zero. The criticalspeed does not however actually fall to zero (unless the two illumina-

tions are of the same color) because of the occurrence of color flicker,

or flicker still present when the two illuminations are equal in intensitybut different in color. This lower limit to the critical frequency curveis shown by the approximately horizontal line joining the two descend-ing branches of the critical frequency curve already discussed. The

HERBERT E. IVES364

Page 3: A CHART OF THE FLICKER PHOTOMETER

May, 1923] A CHART OF THE FLICKER PHOTOMETER 365

region above the two descending branches and their connecting arm isa region in which no flicker occurs, while the region below is one inwhich flicker is present everywhere. The correct speed for operatingthe flicker photometer is the lowest possible which will not cause colorflicker. The ratio of intensities which is evaluated as equality by theflicker photometer is that corresponding to the mid-point of theno-flicker region, which is found by oscillating the ratio of intensitiesbetween the two points of appearance of flicker. If the two branchesof the critical frequency curve of Figure 1 are symmetrically placedabout the equal illumination axis the flicker photometer setting isidentical with the direct comparison or equality of brightness setting.If the branches are unsymmetrically placed, the ratio of intensitiescorresponding to the middle of the no-flicker region is no longer unity,i.e. the flicker photometer setting differs from the equality of brightness.The lower the color flicker critical speed and hence the lower the prac-tical operating speed of the flicker photometer the narrower is thepossible region of setting and hence the greater the precision.

The data incorporated in this diagram were obtained with thewriter's polarization flicker photometer,' in which the transition fromone color to the other is absolutely uniform, so that no question arisesof flicker due to edges of discs or other mechanical imperfections. Thisis the only design of flicker photometer suitable for constructing acomplete chart of the kind under discussion.

DATA AVAILABLE FOR CONSTRUCTION OF THE COMPLETE CHART

The experimental difficulties in the way of securing curves of thetype of Fig. 1 with the polarization flicker photometer, extending overa wide range of intensities, were rather formidable, largely because sogreat a fraction of the available light was lost in the filtering andpolarizing apparatus. With light sources now available higher bright-nesses of the photometer field could be obtained so that it should bepossible to complete the diagram experimentally, which would be highlydesirable. However, by the use of data available from other studies,the gaps can be filled in well enough to make the result worth while.

Three sets of data will be used to complete the chart. The first setof data is shown in Fig. 2. It consists of critical frequency versusillumination data for various monochromatic radiations2 for a largephotometric field (50X80). The critical frequencies are in cycles per

2 Studies in the Photometry of Lightswof Different Colors. Ives, Phil. Mag. 1912, p. 149,352, 744, 845, 853.

Page 4: A CHART OF THE FLICKER PHOTOMETER

second; the illuminations are in arbitrary units, and are plotted on a

logarithmic scale. The observations for red light lie on a single straight

line, those for the other colors on broken lines, the slopes varying from

color to color. The experimental data may be represented by equations

of the formX:= a log I+b

a and b each having one value for red light, and two, over different

ranges of illumination, for each of the other colors.

00

e1 .

40

ILo9 . .1 ArAtayUntj)

FIG. 2. Critical frequency versus log illumination data for mnonochromatic radiations.

The second set of data, also drawn from the paper on the polariza-

tion flicker photometer, consists of observations on the critical speeds

of disappearance of color flicker, as compared with the critical speeds of

the compared colors separately. The experimental points are shown

in Fig. 3, where the upper line is the mean of the speeds for red and

green, the colors studied; while the lower line, which is approximately

straight, but of different slope, is drawn through the color flicker

critical frequency observations.The third set of data, taken from a recent paper on a theory of inter-

mittent vision,3 are embodied in the mathematical expression for the

critical speeds for various amplitudes of the fluctuating stimulus.

Calling a the amplitude, (excursion from the mean position, e.g.

a =12 at the ends of the diagram, Fig. 1) it is shown in the paper quoted

3 A Theory of Intermittent Vision. Ives, J. 0. S. A. & R. S. I., June, 1922, p. 343.

[J.O.S.A. & R.S.I. 7HERBERT E. IES366

Page 5: A CHART OF THE FLICKER PHOTOMETER

May, 1923] A CHART OF THE FLICKER PHOTOMETER 367

that the critical speed a, as the amplitude is changed, is represented bythe equation:

c =a log I a+bwhere a is the same constant as in the previously quoted equation of thesame form, and [a log a+ b] corresponds to the b of the previous equation,that is, the equation for alternation of light with complete darkness isa special case of the equation just given. Now in the lower diagram ofFig. 1, representing the illuminations of the two alternated elementsof the flicker photometer, the dovetailing of the two stimuli correspondsto a variation of the total (summed) amplitude from 12 at the ends, tozero at the center, so that the two constituent curves of Fig. 1 can be

30.

EJ20

0Ioq brIgIhtess

FIG. 3. Critical frequencies for the disappearance of color flicker as compared with themean critical frequencies for the alternated colors separately. Upper points, mean critical fre-quencies for red and green light. Lower points, critical frequencies for red and green alternated.against each other.

calculated from the experimentally established end points (Fig. 2).This expression is strictly applicable only where the dovetailed illumi-nations are of the same color, a point which will be discussed in thefollowing section. Assuming that the use of this expression can bedefended, as will be established, it is clear that we have available allthe data necessary to build up the curves of Fig. 1 for all illuminations.

UTILIZATION OF THE DATA

In order to utilize the data above assembled, several steps arenecessary. First of all the critical frequency data of Fig. 2 must bereferred to a common scale of illumination, which is at least approxi-

Page 6: A CHART OF THE FLICKER PHOTOMETER

[J.O.S.A. & R.S.I., 7

mately correct, instead of the arbitrary scales used in obtaining thedata. A rigidly correct scale of illumination, applicable to all thecolors, could be assigned only upon agreement as to how we are toevaluate different colors photometrically. Avoiding a discussion ofthat problem we can secure the approximate scale sufficient for ourpresent problem by noting that two straight lines of different slope,representing the critical frequencies for two colors, must cross some-where, and that it is well established experimentally that criticalfrequencies for different colors are at least approximately the same forequal luminous values anywhere in the ordinary working range ofilluminations. It has therefore been assumed that the red and greencritical frequency-log I lines cross at 25 meter candles illumination ofthe flicker disc, and that the critical speed at this illumination is givenby the constants in Porter's equation.' The choice of this point at 25meter candles, which represents a practical working illumination, is inaccord with considerable experimental evidence, but the exact pointpicked is not vital to the establishment of the main characteristics ofthe chart. On the basis of these assumptions, the equations of the linesof red and green light, which will be used for the chart are:

For redco=9.9 log I a+39.1

For green, upper portion,co= 11.05 log I a+37.9

lower portionX 4.45 log I a+ 24.9

The color flicker critical frequency line is for simplicity taken as thestraight line most nearly fitting the experimental points establishedby Fig. 3. This stands in a proportional relation to the red and greencritical frequency lines, which leads to its replotting on the assumedillumination scale according to the equation

c=7.25 log +22.5These various lines are shown plotted to scale in Fig. 4.There remains now the question of the applicability of the equation= a log I a+ b where the amplitude is that resulting from the dove-

tailing of illuminations of unlike color. In general we note that theilluminations dovetailed call separately for different critical speeds. Alower critical speed pretty obviously means that the reaction producedin the eye makes smaller excursions from its mean position, that is,that it is of smaller amplitude, and vice versa. Consequently when twoilluminations calling for different critical speeds are dovetailed the

368 HERBERT E. IVES

Page 7: A CHART OF THE FLICKER PHOTOMETER

A CHART OF THE FLICKER PHOTOMETER

hollows of the resultant fluctuating reaction will be lower, the peakshigher than if the illumination needing higher critical speed had beendovetailed with its own color.4 Therefore, we may expect that thecritical speeds called for by the above equation will in general be toolow for the higher branch and too high for the lower branch of criticalfrequency curves such as that of Fig. 1, the error being progressivelygreater toward the center. Such differences in speed as the calculatedbranches indicate on either side of the center will therefore be some-what less than they should be to correspond with the experimentallydetermined end points. It may be pointed out that this simply meansthat the center of the diagram corresponds to a somewhat smaller sizeof observing field than the ends. It does however correspond to somefield size, which is sufficient for our present purpose.

_ L ~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~~10

FIG. 4. Critical frequency log illumination data for red and green light replotted to chosenlluminatio n scale.

THE CHART

Fig. 5 shows the complete chart, as constructed according to theabove description, for red and green light. At the bottom are shownthe relative amounts of the two stimuli which are dovetailed at thecorresponding abscissae of the chart. At the sides of the chart areplotted red and green critical frequencies, the separations betweensuccessive points corresponding to equal distances along the log I axis

4 This conception of the action of the flicker photometer as a producer of fluctuatingreactions which differ in amplitude for the same intensity of stimulus according to the colorand illumination level, is discussed at length in "The Theory of the Flicker Photometer"(Phil. Mag., November, 1914, p. 708).

May, 19�31 369

Page 8: A CHART OF THE FLICKER PHOTOMETER

[J.O.S.A. & R.S.I., 7

of Fig. 4. Extending inward and downward from these points are thecurves calculated by introducing the successive amplitudes (a = 1 2 at

GRdEEN .R :0

7 0 .9 .).760 4 . 2 .

6 0 7m 3 =f =

h~~~~~~~~~~~

FIG. 5. Thte chlart of thze flicker phlotom)eter. Heavy line close to vertical axis shosE, byreference to thte diagramz of relative proportions at thte bottom, thte settinzg made wit/i thte flickerphotometer at thte illurminlationz corresponding to thte critical frequentcy curve cut.

ends, and dropping to 0 at the center of the diagram) into the criticalfrequency-log I expressions for red and green respectively. These

370 HERBERT E. IVES

Page 9: A CHART OF THE FLICKER PHOTOMETER

A CHART OF THE FLICKER PHOTOMETER

curves are continued until they meet the approximately horizontallines which are drawn through the critical frequencies for color flicker,also taken from Fig. 4. Each complete curve corresponding to oneillumination therefore consists of two branches descending according tothe logarithm of the amplitude, limited and connected by a shortapproximately horizontal branch. The various curves are letteredfrom a, corresponding to the condition of equal critical speeds for bothcolors, down to , and upward by primed letters to b', c', d'. Theexperimentally obtained points of Fig. 1 are reproduced on the chart(circles); it will be seen that they fall very satisfactorily among thecomputed values. They show however much less dissymmetry, whichis to be expected since the observing field was of much smaller diameterin the polarization flicker photometer than was the field used in ob-taining the critical frequency data of Fig. 2.

Short dashed lines are drawn which bisect the no-flicker regionsjust above the limiting speeds for color flicker. These lines are the lociof the flicker photometer setting-positions for the speeds given by thecorresponding ordinates. The heavy curved line closely hugging theaxis of the figure connects the intersection of these dashed lines with thecolor-flicker critical-speed lines, that is, the points of most sensitiveflicker setting.

Examining the chart in detail we note that it indicates a zig-zagcourse of the flicker setting about the axis of the figure as the illumina-tion is progressively changed. At very high illuminations, at the topof the chart, the median of the flicker region is on the red side, that is,red will be underrated. At an illumination near but not exactly co-incident with the equal critical speed point the flicker setting lies onthe axis. On decreasing the illumination the setting point movestoward the green side, that is red will be overrated, constituting thereversed Purkinje effect, discovered by the writer.2 At still lower illumi-nations, due to the change of slope of the green critical frequency-logillumination line the flicker setting point moves again toward the redside.

It is to be noted however that the most violent shift in position ofthe flicker photometer setting with change in illumination is indicatedfor the frequency region below about 12 cycles per second. No experi-mental observations have been made this low, and it is in fact doubtfulwhether the critical frequency relations here utilized still hold for thosevery low intensities. Moreover, the curves are computed on the basisof equal physical steps in intensity, neglecting the relatively greater

May, 1923] 371

Page 10: A CHART OF THE FLICKER PHOTOMETER

[J.O.S.A. & R.S.I., 7

subjective judgment of short wave energy (the Purkinje effect) whichwould influence a photometric setting by direct comparison. Thelack of complete consistency of the central portions of the

computed curves with the end points, already alluded to, causes

the lowest shift in direction of the flicker setting line to be dis-

placed vertically, so that it does not correspond as nearly as it shouldto the locus of the second (lower) symmetrical critical-frequency curve.For these various reasons the lower end of the flicker setting curvehas been drawn in as a dashed line only; it is useful as indicating the

7W-

FIG. 6. Experimental "slit-widthz" verses log illumoination data for thze flicker photometer.

direction the curve is taking, but must not be taken too seriously as a

quantitative guide.A fact of importance to notice is that these shifts to either side of

the axis are not of the same magnitude for the same (logarithmic)change of illumination. Because of the fact that the color-flickercritical-frequency drops below the homogeneous critical frequency byprogressively greater amounts as the illumination increases, the two

branches of the critical frequency-amplitude curves continue to rela-tively low values for the higher illuminations. Conversely at the lower

372 HERBERT E. IVES

Page 11: A CHART OF THE FLICKER PHOTOMETER

May, 1923] A CHART OF THE FLICKER PHOTOMETER 373

illuminations the greater relative values of the color-flicker criticalfrequencies prevent these curves from closely approaching the axis.As a net result the greatest variations of the flicker photometer settingfrom the axis occur at low illuminations; at high illuminations they areto a large extent offset by the narrowing of the flicker region near thesetting point.

All of these findings are very beautifully checked by the experimentalcurves reproduced in Fig. 6 from an earlier paper of the writer.2 In thisfigure the ordinates are illuminations (meter candles with field viewedthrough small artificial pupil) on a logarithmic scale, and so correspondto the critical frequency ordinate scale of Fig. 5, bearing in mind thelogarithmic relation between illumination and critical frequency. Ineach case the colors alternated are a monochromatic spectrum patchand the "white" of a carbon lamp. This latter color, according to thefindings of the study from which Fig. 3 is reproduced, yields criticalfrequency-illumination data practically coincident with green light.Consequently the "slit width" versus log illumination curves of Fig. 6for red light should correspond to the case studied in the present paper.The qualitative corresponcence between Fig. 5 and 6 is in fact complete,showing the validity of the methods used in constructing the chart.

RESEARCH LABORATORIES OF THEAMERICAN TELEPHONE AND TELEGRAPH CO.,

AND THE- WESTERN ELECTRIC COMPANY, INC.NOVEMBER 24, 1922.

On the Emission of Light by Canal Rays.-This article f Wien'sconsists mainly of equations developed from his conception of canalrays, which may be summarized as follows: (a) in a canal ray beamthere are three kinds of particles; ionized atoms, neutral "unexcited"atoms and neutral "excited" atoms; (b) the relative proportions ofthese tend towards a statistical equilibrium, depending on the pressureof the gas which the beam is traversing; (c) the transitions of particlesfrom one kind to another are similar to collisions between gas molecules,the "free time" or "free path" of a particle between transitions havingvalues distributed exponentially about mean values corresponding tothe mean free paths of gas molecules; (d) emission of light takes placewhen a particle passes from the neutral excited to the neutral unexcitedstate, and the mean value of the path preceding this transition isindependent of the pressure of the gas; (e) but if a collision between aparticle in the neutral excited state and a gas molecule occurs, theparticle reverts to the neutral unexcited state without emission oflight; (f) the mean free path preceding such a collision, and all theother kinds of mean free paths, are inversely as the pressure of the gas.