8-monitoring during anesthesia

Upload: sanjivdas

Post on 30-May-2018

222 views

Category:

Documents


0 download

TRANSCRIPT

  • 8/14/2019 8-Monitoring During Anesthesia

    1/108

    1

    MonitoringMonitoring

    duringduring AnesthesiaAnesthesia

  • 8/14/2019 8-Monitoring During Anesthesia

    2/108

    2

    Perioperative periodPerioperative period

    operationoperation postoperationpostoperationpreoperationpreoperation

    operative evaluationoperative evaluation

    and preparationand preparation anesthesiaanesthesia

    Induction ofInduction of

    anesthesiaanesthesia

    Maintenance ofMaintenance of

    anesthesiaanesthesia

    Recovery fromRecovery from

    anesthesiaanesthesia

    Monitoring during anesthesiaMonitoring during anesthesia

  • 8/14/2019 8-Monitoring During Anesthesia

    3/108

    3

    The word monitor is derived from the LatinThe word monitor is derived from the Latin

    verbverb moneremonere to warn. The purpose of a to warn. The purpose of amonitoring device is to measure amonitoring device is to measure aphysiological variable and to indicatephysiological variable and to indicatetrends of change, thus enablingtrends of change, thus enabling

    appropriate therapeutic action to be taken.appropriate therapeutic action to be taken. It is essential to ensure that all monitoringIt is essential to ensure that all monitoring

    equipment is maintained correctly andequipment is maintained correctly andthat it functions accurately, so that thethat it functions accurately, so that the

    information which it provides is reliableinformation which it provides is reliable The user should understand the basicThe user should understand the basic

    principles on which monitoring equipmentprinciples on which monitoring equipmentis based and be able to interpret theis based and be able to interpret the

    information provided.information provided.

  • 8/14/2019 8-Monitoring During Anesthesia

    4/108

    4

    MonitoringMonitoring duringduring

    AnesthesiaAnesthesia Cardiovascular SystemCardiovascular System

    Respiratory systemRespiratory system

    Body Temperature: geriatric, pediatricBody Temperature: geriatric, pediatric

    Neuromuscular Junction: muscularNeuromuscular Junction: muscular

    relaxantrelaxant

    Nervous System: depth of anesthesiaNervous System: depth of anesthesia

    Coagulation System: cardiothoracicCoagulation System: cardiothoracic

    surgerysurgery

  • 8/14/2019 8-Monitoring During Anesthesia

    5/108

    5

    PARTPART

    THE CARDIOVASCULARTHE CARDIOVASCULAR

    SYSTEMSYSTEM

  • 8/14/2019 8-Monitoring During Anesthesia

    6/108

  • 8/14/2019 8-Monitoring During Anesthesia

    7/108

    7

    Electrocardiogram (EElectrocardiogram (ECCG)G)

    ECG monitors have become increasinglyECG monitors have become increasingly

    reliable and less subject to interference.reliable and less subject to interference.

    As the technique is non-invasive, simpleAs the technique is non-invasive, simple

    and accurate, it is now regarded asand accurate, it is now regarded as

    mandatory in the UK that the ECG shouldmandatory in the UK that the ECG should

    be monitored in all patients undergoingbe monitored in all patients undergoinganaesthesia, no matter how minor theanaesthesia, no matter how minor the

    surgery procedure.surgery procedure.

  • 8/14/2019 8-Monitoring During Anesthesia

    8/108

    8

    Electrocardiogram (EElectrocardiogram (ECCG)G)

    One of the standardized monitors duringOne of the standardized monitors during

    any form of anesthesia.any form of anesthesia.

    For detection and diagnosis ofFor detection and diagnosis of

    DysrhythmiasDysrhythmias: Tachycardia, Bradycardia,: Tachycardia, Bradycardia,VFVF

    CConduction defectsonduction defects: AVB: AVB auriculo-ventricular blockCCardiac ischemiaardiac ischemia: CAD: CAD

    EElectrolyte disturbancelectrolyte disturbance: Potassium,: Potassium,

    MagnesiumMagnesiumhyperkalemia, hypokalemia

  • 8/14/2019 8-Monitoring During Anesthesia

    9/108

    9

    Types of MonitoringTypes of Monitoring

    Five-electrode system: one onFive-electrode system: one on

    each leach limbimb and one precordial leadand one precordial lead

    (V(V5,5, along the anterior axillary linealong the anterior axillary line

    in the fifth intercostal spacein the fifth intercostal space forfor

    detection of anterior ischemia).detection of anterior ischemia).

  • 8/14/2019 8-Monitoring During Anesthesia

    10/108

    10

    PPrecordial leadrecordial lead

    Midclavicular line

    Anterior axillary line Midaxillary line

    Posterior axillary line

  • 8/14/2019 8-Monitoring During Anesthesia

    11/108

    11

    Standard limb lead

  • 8/14/2019 8-Monitoring During Anesthesia

    12/108

    12

    Types of MonitoringTypes of Monitoring

    Standard lead II monitoring is usedStandard lead II monitoring is used

    widely.widely.

    However, the CMHowever, the CM55 lead configuration haslead configuration has

    been advocated for routine intraoperativebeen advocated for routine intraoperative

    monitoring because it reveals moremonitoring because it reveals more

    readily ST segment changes produced byreadily ST segment changes produced by

    left ventricular ischaemia.left ventricular ischaemia.

  • 8/14/2019 8-Monitoring During Anesthesia

    13/108

    13

    Types of MonitoringTypes of Monitoring

  • 8/14/2019 8-Monitoring During Anesthesia

    14/108

    14

    Types of MonitoringTypes of Monitoring

    Right arm lead overmanubrium sterni

  • 8/14/2019 8-Monitoring During Anesthesia

    15/108

    15

    Electrocardiogram (EElectrocardiogram (ECCG)G)

    KaplanKaplan et al.et al. (Anesthesiology, 1976):(Anesthesiology, 1976):

    90% of intraoperative cardiac ischemia will90% of intraoperative cardiac ischemia will

    be detected by multiple Ebe detected by multiple ECCG, especially VG, especially V55..

    At least two leads should be simultaneouslyAt least two leads should be simultaneously

    showed on the monitor.showed on the monitor.

  • 8/14/2019 8-Monitoring During Anesthesia

    16/108

    16

    Position of the LeadsPosition of the Leads

    The four limb leads should be placed onThe four limb leads should be placed on

    the back of shoulders and hips, wherethe back of shoulders and hips, where

    they will disturb the operative field thethey will disturb the operative field theleast.least.

    Every lead should be fixed and protectedEvery lead should be fixed and protected

    with tape to prevent dislodgement ofwith tape to prevent dislodgement ofleads during operation.leads during operation.

  • 8/14/2019 8-Monitoring During Anesthesia

    17/108

    17

    Electrocardiogram (EElectrocardiogram (ECCG)G)

    It is important to appreciate that the ECGIt is important to appreciate that the ECG

    is an index only of electrical activity. It isis an index only of electrical activity. It is

    possible for a normal electrical waveformpossible for a normal electrical waveformto exist in the presence of a negligibleto exist in the presence of a negligible

    cardiac output.cardiac output.

    Consequently, information from the ECGConsequently, information from the ECGshould be used in conjunction with datashould be used in conjunction with data

    acquired from monitoring of perfusionacquired from monitoring of perfusion

  • 8/14/2019 8-Monitoring During Anesthesia

    18/108

    18

    Monitoring the circulationMonitoring the circulation Maintenance of perfusion of vital organs is oneMaintenance of perfusion of vital organs is one

    of the principal tasks of the anaesthetist duringof the principal tasks of the anaesthetist during

    surgery. Adequate perfusion is dependent onsurgery. Adequate perfusion is dependent on

    adequate venous return to the heart, cardiacadequate venous return to the heart, cardiac

    performance and arterial pressure.performance and arterial pressure. Direct measurements of cardiac output andDirect measurements of cardiac output and

    blood volume are difficult during anaesthesiablood volume are difficult during anaesthesia

    and require invasive procedures which areand require invasive procedures which are

    inappropriate in many situations.inappropriate in many situations.

  • 8/14/2019 8-Monitoring During Anesthesia

    19/108

    19

    Monitoring the circulationMonitoring the circulation However, adequacy of cardiac output andHowever, adequacy of cardiac output and

    circulating blood volume may be inferredcirculating blood volume may be inferred

    indirectly from observation of the followingindirectly from observation of the following

    variables:variables:

    1. Peripheral pulse.1. Peripheral pulse.

    2. Arterial oxygen saturation.2. Arterial oxygen saturation.

    3. Peripheral perfusion.3. Peripheral perfusion.

    4. Urine production.4. Urine production.

    5.Arterial pressure.5.Arterial pressure.

  • 8/14/2019 8-Monitoring During Anesthesia

    20/108

    20

    1.Regular palpation of the peripheralpulse is one of the simplest and most

    useful methods of monitoring during

    anaesthesia and is mandatory for even

    the most minor surgery. (Radial artery)

    2.Information may be obtained by

    observation of the rate, volume and

    rhythm.

    The peripheral pulseThe peripheral pulse

  • 8/14/2019 8-Monitoring During Anesthesia

    21/108

    21

    1.Pulse oximeters measure the arterial

    oxygen saturation and the pulse rate

    non-invasively and accurately to within

    2%.

    2.A simple probe is attached to a finger

    an ear lobe, flexed across the nasalbridge, or wrapped around a childs digit

    and connected to the oximeter.

    3.The probe contains two light-emittingdiodes, one for red and one for infrared

    light, and a single detector positioned

    on the opposite side of the digit or ear

    lobe.

    Pulse oximetryPulse oximetry

  • 8/14/2019 8-Monitoring During Anesthesia

    22/108

    22

    Pulse oximetry----ProbePulse oximetry----Probe

    Finger probe forpulse oximeterattached to afinger

  • 8/14/2019 8-Monitoring During Anesthesia

    23/108

    23

    Pulse oximetry----Pulse oximetry----

    ProbeProbe

    Disposable fingerprobe for pulseoximeterwrapped around a

    childs digit

  • 8/14/2019 8-Monitoring During Anesthesia

    24/108

    24

    Pulse oximetry----Pulse oximetry----

    OximeterOximeter

    Pulseoximeter

    control anddisplaymodule,probe is

    connected to

    Waveform S

    pO2

    Pulse rate

  • 8/14/2019 8-Monitoring During Anesthesia

    25/108

    25

    Pulse oximetryPulse oximetry

    Pathophysiology:

    Oxygen is exchanged by diffusion fromOxygen is exchanged by diffusion from

    higher concentrations to lowerhigher concentrations to lower

    concentrationsconcentrations

    Most of the oxygen in the arterial blood isMost of the oxygen in the arterial blood is

    carried bound to hemoglobincarried bound to hemoglobin

    -97% of total oxygen is normally bound to-97% of total oxygen is normally bound tohemoglobin (hemoglobin (SpOSpO22))

    -3% of total oxygen is dissolved in the-3% of total oxygen is dissolved in the

    plasma (Paplasma (PaOO22))

  • 8/14/2019 8-Monitoring During Anesthesia

    26/108

    26

    Pulse oximetryPulse oximetry

    Oxygen Saturation: Percentage of hemoglobin saturated withPercentage of hemoglobin saturated with

    oxygenoxygen

    Normal SpONormal SpO

    22 is 95-98%is 95-98%

    Suspect cellular perfusion compromise ifSuspect cellular perfusion compromise if

    SpOSpO22 is less than 95%is less than 95%

    -Insure adequate airway-Insure adequate airway

    -Provide supplemental oxygen-Provide supplemental oxygen

    -Monitor carefully for further changes and-Monitor carefully for further changes and

    intervene appropriatelyintervene appropriately

  • 8/14/2019 8-Monitoring During Anesthesia

    27/108

    27

    Pulse oximetryPulse oximetry

    Oxygen Saturation:

    Suspect severe cellular perfusionSuspect severe cellular perfusion

    compromise when SpOcompromise when SpO22 is less than 90%is less than 90% Insure airway and provide positiveInsure airway and provide positive

    ventilations if necessary.ventilations if necessary.

    Administer high flow oxygen.Administer high flow oxygen. Head injured patients, SpOHead injured patients, SpO22shouldshould

    never drop below 90%.never drop below 90%.

  • 8/14/2019 8-Monitoring During Anesthesia

    28/108

    28

    Pulse oximetryPulse oximetry

    SpO2 and PaO2:

    SpOSpO22 indicates the oxygen bound toindicates the oxygen bound to

    hemoglobinhemoglobin-Closely corresponds to SaO-Closely corresponds to SaO22 measured inmeasured in

    laboratory testslaboratory tests

    -SpO-SpO22 indicates the saturation wasindicates the saturation was

    obtained with non-invasive oximetryobtained with non-invasive oximetry

    PaOPaO22 indicates the oxygen dissolved in theindicates the oxygen dissolved in the

    plasmaplasma

    -Measured in ABGs (artery blood gases)-Measured in ABGs (artery blood gases)

  • 8/14/2019 8-Monitoring During Anesthesia

    29/108

    29

    Pulse oximetryPulse oximetry

    SpO2 and PaO2:

    Normal PaONormal PaO22 is 80-100 mmHgis 80-100 mmHg

    NormallyNormally

    80-100 mm Hg corresponds to 95-100%80-100 mm Hg corresponds to 95-100%

    SpOSpO22

    60 mm Hg corresponds to 90% SpO60 mm Hg corresponds to 90% SpO22

    40 mm Hg corresponds to 75% SpO40 mm Hg corresponds to 75% SpO22

  • 8/14/2019 8-Monitoring During Anesthesia

    30/108

    30

    1.Pulse oximeters are simple to use, non-

    invasive and require no warm-up time.

    2.Provide an overall assessment of the

    integrity of all the systems involved in

    delivering oxygen to the tissues

    Oxygen supply, intake and delivery.

    3.unaffected by pigmented skin

    Pulse oximetryPulse oximetry

    Advantages:

  • 8/14/2019 8-Monitoring During Anesthesia

    31/108

  • 8/14/2019 8-Monitoring During Anesthesia

    32/108

    32

    1.It is assessed most usefully by

    observation of the patient's extremities.

    2.Warm, dry, pink skin indicates adequate

    peripheral perfusion; cold, white

    peripheries the converse

    3.This is particularly true in children, inwhom cool peripheries usually indicate

    hypovolaemia.

    Peripheral perfusionPeripheral perfusion

  • 8/14/2019 8-Monitoring During Anesthesia

    33/108

    33

    1.The core-peripheral temperature

    gradient is a useful index of adequacy

    of peripheral perfusion.

    2.One temperature probe is placed

    centrally (e.g. in the nasopharynx) and

    the other peripherally (e.g. on the greattoe)

    3.The temperature gradient increases

    with vasoconstriction and low cardiacoutput, and decreases gradually as

    vasodilation occurs with increasing limb

    blood flow consequent upon increasing

    cardiac output.

    Peripheral perfusionPeripheral perfusion

  • 8/14/2019 8-Monitoring During Anesthesia

    34/108

    34

    1.Adequacy of renal perfusion may be

    inferred from the volume of urine

    produced

    2.The kidney is the only organ whose

    function may be monitored directly in

    this way.

    3.Adequate production of urine implies

    that perfusion of other vital organs is

    Urine outputUrine output

  • 8/14/2019 8-Monitoring During Anesthesia

    35/108

    35

    1.Accurate measurement of urine

    volumes with a urimeter is particularly

    indicated in the following situations:

    Urine outputUrine output

    6. Surgery in the jaundiced

    patient.

    3. Major trauma.

    5. Massive fluid or blood

    loss

    2. Cardiac surgery.

    4. Critically ill/shocked

    patients

    1. Major vascular

    surgery.

    aim is to achieve a urine output of 0.5~1ml.kg-1.h-1

  • 8/14/2019 8-Monitoring During Anesthesia

    36/108

    3 It is an indirect method of estimating

  • 8/14/2019 8-Monitoring During Anesthesia

    37/108

    37

    3. It is an indirect method of estimating

    adequacy of

    cardiac output, because:

    Blood pressure CO peripheral

    resistance

    4. In conjunction with estimation ofperipheral

    perfusion, it is an invaluable

    measurement.

    6. Indirect, non-invasive methods of

    measurement

    Systemic arterialSystemic arterial

    pressurepressure

  • 8/14/2019 8-Monitoring During Anesthesia

    38/108

    38

    Palpation of the radial pulse as the

    sphygmoma-nometer cuff is deflated is

    a simple method of measuring systolic

    pressure, but is inaccurate at low

    pressures or when vasoconstriction is

    present.

    Systemic arterialSystemic arterial

    pressurepressurePalpation :

  • 8/14/2019 8-Monitoring During Anesthesia

    39/108

    39

    1.Auscultation of the Korotkoff sounds is

    too cumbersome for routine use during

    anaesthesia.

    Systemic arterialSystemic arterial

    pressurepressureAuscultation :

  • 8/14/2019 8-Monitoring During Anesthesia

    40/108

  • 8/14/2019 8-Monitoring During Anesthesia

    41/108

    41

    Systemic arterialSystemic arterial

    pressurepressureOscillometry (2):

    an automated oscillometer

  • 8/14/2019 8-Monitoring During Anesthesia

    42/108

    42

    Systemic arterialSystemic arterial

    pressurepressureOscillometry (2):

    Multifunctional Monitor

  • 8/14/2019 8-Monitoring During Anesthesia

    43/108

  • 8/14/2019 8-Monitoring During Anesthesia

    44/108

    44

    Cuff SizeCuff Size

    Too small cuff will result in highToo small cuff will result in high

    blood pressure reading.blood pressure reading. A loosely applied cuff will alsoA loosely applied cuff will also

    produce a reading higher than itproduce a reading higher than it

    should be.should be.

    Too large cuff will severelyToo large cuff will severely

  • 8/14/2019 8-Monitoring During Anesthesia

    45/108

    45

    Cuff SizeCuff Size

    from Barbara Bates: A Guide to Physical

    Examination

  • 8/14/2019 8-Monitoring During Anesthesia

    46/108

    46

    Systemic arterialSystemic arterial

    pressurepressureOscillometry (3):

    a.Inaccurate when SBP less than 60mmHg

    .

    b. Unable to follow rapid swings in arterial

    pressure .

    c. Under-reading occurred at high systolic

    pressures

    Disadvantages:

  • 8/14/2019 8-Monitoring During Anesthesia

    47/108

    47

    1. It is achieved by attaching a

    transducer to an intra-arterial cannula

    inserted into a peripheral artery.

    2. It is an invasive procedure which

    carries potential morbidity.

    3. The method is only justified whenrapid changes in arterial pressure are

    anticipated during anaesthesia.

    Direct measurement ofDirect measurement of

    ABPABP

  • 8/14/2019 8-Monitoring During Anesthesia

    48/108

    48

    IABP system

    CannulaManometertubeTransducerMonitor

  • 8/14/2019 8-Monitoring During Anesthesia

    49/108

    49

    1. The transducer should be zeroed, and

    the system should be calibrated.

    2. The transducer is connected to the

    cannula via a piece of stiff-walled,saline-filled manometer tube.

    3. The pressure signal is displayed as a

    waveform on an oscilloscope screen and

    systolic, diastolic and mean arterial

    pressures displayed digitally.

    Direct measurement ofDirect measurement of

    ABPABP

  • 8/14/2019 8-Monitoring During Anesthesia

    50/108

    50

    Common indications for arterialcannulation:

    Direct measurement ofDirect measurement of

    ABPABP

    Major vascular surgery

    Cardiothoracic surgery

    Induced hypotension

    Critically ill and shocked patients

    Surgery for pheochromocytoma

    Neurosurgery

    Necessity for frequent blood gas analysis

  • 8/14/2019 8-Monitoring During Anesthesia

    51/108

    51

    Morbidity associated with arterialcannulation :

    Direct measurement ofDirect measurement of

    ABPABP

    1. Arterial wall damage and

    thrombosis

    2. Embolization

    3. Disconnection and

    haemorrhage4. Sepsis

    5. Tissue necrosis

  • 8/14/2019 8-Monitoring During Anesthesia

    52/108

    52

    Direct measurement ofDirect measurement of

    ABPABP

    Site for Arterial Cannulation:

    1. Radial artery:

    2. Femoral artery:

    3. Brachial artery:

    4. Dorsalis pedis artery:

  • 8/14/2019 8-Monitoring During Anesthesia

    53/108

    53

    1. A central venous catheter positioned

    with its tip in the superior (inferior) vena

    cava.

    2. It provides valuable information

    concerning the volume status of thecirculation during anaesthesia.

    3. A direct measurement of RV filling

    pressure, CVP is good measurement of

    LV filling pressure only in the absence of

    pulmonary hypertension or mitral

    Central venousCentral venous

    pressure (CVP)pressure (CVP)

  • 8/14/2019 8-Monitoring During Anesthesia

    54/108

    54

    Central venousCentral venous

    pressure (CVP)pressure (CVP)

    tip in the superior

    Internal jugular

    vein

    External jugular

    vein

  • 8/14/2019 8-Monitoring During Anesthesia

    55/108

    55

    Central venousCentral venous

    pressure (CVP)pressure (CVP)

    Femoral Vein

  • 8/14/2019 8-Monitoring During Anesthesia

    56/108

    56

    Central venousCentral venous

    pressurepressure

    Site for Vein Cannulation:

    1. Internal jugular vein

    2. External jugular vein3. Subclavian vein

    4. Femoral vein

    5.Peripheral arm vein

  • 8/14/2019 8-Monitoring During Anesthesia

    57/108

    57

    Peripheral arm vein :

    Central venousCentral venous

    pressurepressure1.This route is the least likely to provide

    correct placement of the catheter

    (approximately 40%)

    2.It avoids most of the serious

    complications of other routes of

    insertion.

    3.Thrombophlebitis and sepsis are

    common when a peripheral arm vein is

    used, particularly if the catheter is left

  • 8/14/2019 8-Monitoring During Anesthesia

    58/108

    58

    Internal jugular cannulation:

    Central venousCentral venous

    pressurepressure

    This route is associated with the

    highest incidence of correct

    catheter placement (approximately

    90%).

  • 8/14/2019 8-Monitoring During Anesthesia

    59/108

    59

    Technique for right internal jugularTechnique for right internal jugular

    vein central venous cannulation.vein central venous cannulation.

    P t

  • 8/14/2019 8-Monitoring During Anesthesia

    60/108

    60

    (A) Important surface landmarks are identified.(A) Important surface landmarks are identified.

    (B) The course of the internal carotid artery is palpated.(B) The course of the internal carotid artery is palpated. (C) The internal jugular vein is punctured at the apex of(C) The internal jugular vein is punctured at the apex of

    the triangle formed by the two heads of thethe triangle formed by the two heads of thesternocleidomastoid muscle with the needle tip directedsternocleidomastoid muscle with the needle tip directedtoward the ipsilateral nipple.toward the ipsilateral nipple.

    (D) A guide wire is introduced through the thin-wall(D) A guide wire is introduced through the thin-wallneedle into the vein.needle into the vein.

    (E) The central venous cannula is inserted over the(E) The central venous cannula is inserted over theguide wire, making sure that the proximal end of theguide wire, making sure that the proximal end of theguide wire protrudes beyond the catheter and isguide wire protrudes beyond the catheter and is

    controlled by the operator. See text for greater detail.controlled by the operator. See text for greater detail.

    PunctureProcedure

    C l

  • 8/14/2019 8-Monitoring During Anesthesia

    61/108

    61

    Complications of internal jugularcannulation:

    Central venousCentral venous

    pressurepressure

    1. Air embolism

    2. Carotid artery puncture

    3. Brachial plexus/phrenic nerve

    damage

    4. Ectopic placement (numerous

    sites)5. Sepsis

    6. Pneumothorax

    lC t l

  • 8/14/2019 8-Monitoring During Anesthesia

    62/108

    62

    Subclavian vein:

    Central venousCentral venous

    pressurepressure

    1.This approach is more hazardous

    than the internal jugular, and less

    likely to provide correct catheter

    placement.

    2.It is the most suitable route if long-

    term parenteral feeding.

    C lC t l

  • 8/14/2019 8-Monitoring During Anesthesia

    63/108

    63

    Complications of subclavian veincannulation :

    Central venousCentral venous

    pressurepressure

    1. Pneumothorax

    2. Subclavian artery puncture

    3. Air embolism

    4. Damage to thoracic duct (left

    side)

  • 8/14/2019 8-Monitoring During Anesthesia

    64/108

    64

    Measurement should be done atMeasurement should be done at end-end-

    expirationexpiration phase.phase.

  • 8/14/2019 8-Monitoring During Anesthesia

    65/108

    65

    Right atriumCVP

  • 8/14/2019 8-Monitoring During Anesthesia

    66/108

    C t lC t l

  • 8/14/2019 8-Monitoring During Anesthesia

    67/108

    67

    Measurement of CVP:

    Central venousCentral venous

    pressurepressure1.The normal range of values is

    0~6cmH2O in the spontaneously

    breathing patient.

    2.In patients receiving IPPV values of

    CVP are approximately 5 cmH2O higher

    because of the increased mean

    intrathoracic pressure

    3.Trends in measured observations are

    C t lC t l

  • 8/14/2019 8-Monitoring During Anesthesia

    68/108

    68

    Increase of CVP :

    Central venousCentral venous

    pressurepressure

    1. Over hydration

    2. Right-sided heart failure3. Cardiac tamponade

    4. Constrictive pericarditis

    5. Pulmonary hypertension6. Tricuspid stenosis and regurgitation

    7. stroke volume is high

    Pulmonary artery pressurePulmonary artery pressure

  • 8/14/2019 8-Monitoring During Anesthesia

    69/108

    69

    Pulmonary artery pressurePulmonary artery pressure

    monitoringmonitoring

    It can measure:It can measure:

    2.2. PA pressurePA pressure

    3.3. Pulmonary capillary wedged pressurePulmonary capillary wedged pressure(PCWP): a balloon at catheter tip(PCWP): a balloon at catheter tip

    (volume 1.5 ml), when the balloon is(volume 1.5 ml), when the balloon is

    inflated and the vessel is wedged, ainflated and the vessel is wedged, a

    valveless hydrostatic column existsvalveless hydrostatic column existsbetween the distal port and LA.between the distal port and LA.

    PAPPAC (Swan-PAPPAC (Swan-Ganz)Ganz)

    Pulmonary artery pressurePulmonary artery pressure

  • 8/14/2019 8-Monitoring During Anesthesia

    70/108

    70

    Pulmonary artery pressurePulmonary artery pressure

    monitoringmonitoring

    1.1. CVP: a port for CVP measurement isCVP: a port for CVP measurement is

    located at 30 cm from the tiplocated at 30 cm from the tip

    2.2. Cardiac output: measurement of RVCardiac output: measurement of RV

    outputoutput

    3.3. Blood temperatureBlood temperature

    4.4. Derived hemodynamic dataDerived hemodynamic data

    n Mixed Venous OMixed Venous O22 saturation (SvOsaturation (SvO22))

    The Pulmonary ArteryThe Pulmonary Artery

  • 8/14/2019 8-Monitoring During Anesthesia

    71/108

    71

    The Pulmonary Arteryy y

    CatheterCatheter

    (PA catheter, Swan-Ganz)(PA catheter, Swan-Ganz)

    PAP, PCWP, SvO2

    CVP

    Blood temperature CO

    Administration

  • 8/14/2019 8-Monitoring During Anesthesia

    72/108

    72

    Pulmonary Artery CatheterPulmonary Artery Catheter1. The proximal lumen. This is situatedapproximately 25 cm from the tip and should lie in

    the right atrium after final placement of the catheter.

    CVP may be measured using this lumen.

    2. The distal lumen Situated at the tip of thecatheter, this lumen lies in a major branch of the

    pulmonary artery when the catheter is placed

    correctly and is used to measure pulmonary artery

    pressure by connecting it to a suitable transducer.

    3. The balloon lumen. This lumen permits the

    introduction of approximately 1.5 ml of air into the

    ballon which surrounds the distal tip of the lumen.

    4. Thermistor lumen. A bead thermistor is

    situated 4cm from the tip of the catheter and

    measures the tem erature of blood at this site.

    P l A tPulmonary Artery

  • 8/14/2019 8-Monitoring During Anesthesia

    73/108

    73

    Pulmonary ArteryPulmonary Artery

    CatheterCatheter

    PA Port:PA Port:YELLOWYELLOW

    CVP Port:CVP Port:BLUEBLUE

    PA balloon Port:PA balloon Port:REDRED

    1 Distal port opening is at1 Distal port opening is at

  • 8/14/2019 8-Monitoring During Anesthesia

    74/108

    74

    1. Distal port opening is at1. Distal port opening is at

    the tip (end) of thethe tip (end) of the

    cathetercatheter.. pulmonary artery pressures (PAP)pulmonary artery pressures (PAP) systolic (PAS)systolic (PAS)

    diastolic (PAD)diastolic (PAD) pulmonary capillary wedge pressure (PCWP)pulmonary capillary wedge pressure (PCWP)

    when balloon is inflated PA pressures shouldwhen balloon is inflated PA pressures shouldalways be monitored continuouslyalways be monitored continuously

    NEVER USENEVER USE for medication infusionfor medication infusion Can be used for drawing "mixed venous" bloodCan be used for drawing "mixed venous" blood

    gas samplegas sample

  • 8/14/2019 8-Monitoring During Anesthesia

    75/108

    75

    2.2.Balloon portBalloon port

    located about < 1 cm from tip of the catheterlocated about < 1 cm from tip of the catheter

    the balloon is inflated with proximately 0.8 to 1.5 ccthe balloon is inflated with proximately 0.8 to 1.5 cc

    of airof air do not inflate with liquid---- always inflate with airdo not inflate with liquid---- always inflate with air

    when deflated, turn stopcock to off position andwhen deflated, turn stopcock to off position and

    leave syringe connect to portleave syringe connect to port

    3 Th i t d3 Th i t d

  • 8/14/2019 8-Monitoring During Anesthesia

    76/108

    76

    3. Thermistor and3. Thermistor and

    connector portconnector port the thermistor connector connects thethe thermistor connector connects thepulmonary catheter to the cardiac outputpulmonary catheter to the cardiac outputcomputercomputer

    thermistor wire within the lumen transmitsthermistor wire within the lumen transmitsblood temperature (core temperature is mostblood temperature (core temperature is mostaccurate reflection of the body temperature)accurate reflection of the body temperature)used in determiningused in determining cardiac outputcardiac output

    i l4 P i l t

  • 8/14/2019 8-Monitoring During Anesthesia

    77/108

    77

    4.Proximal port 4.Proximal port

    approximately 30 cm fromapproximately 30 cm fromtip of catheter.tip of catheter. also known as CVP port (central venousalso known as CVP port (central venous

    pressure) lies in the right atrium andpressure) lies in the right atrium and

    measures CVP can be used for infusion of IVmeasures CVP can be used for infusion of IV

    solutions or medicationssolutions or medications

  • 8/14/2019 8-Monitoring During Anesthesia

    78/108

    78

  • 8/14/2019 8-Monitoring During Anesthesia

    79/108

    79

    i f h hL i f h h

  • 8/14/2019 8-Monitoring During Anesthesia

    80/108

    80

    Location of the catheterLocation of the catheter

  • 8/14/2019 8-Monitoring During Anesthesia

    81/108

  • 8/14/2019 8-Monitoring During Anesthesia

    82/108

    82

    W f d i I i

  • 8/14/2019 8-Monitoring During Anesthesia

    83/108

    83

    Waveform during InsertionWaveform during Insertion

  • 8/14/2019 8-Monitoring During Anesthesia

    84/108

    84

    Length of InsertionLength of Insertion

    Usual conditions (just for ease toUsual conditions (just for ease to

    memorize)memorize)

    35 cm: RV35 cm: RV 4545 cm: PAcm: PA

    55 cm: wedge55 cm: wedge

    But the actual length may varyBut the actual length may vary

    greatly between patients!greatly between patients!

    HemodynamicHemodynamic

  • 8/14/2019 8-Monitoring During Anesthesia

    85/108

    85

    yMeasurements--Measurements--NormalNormal

    RangeRange RA pressure: 2~6 mmHgRA pressure: 2~6 mmHg

    RV pressure:RV pressure: systolic 15~25 mmHg;systolic 15~25 mmHg;

    diastolic 0~4 mmHgdiastolic 0~4 mmHg PA pressure: systolic 15~25 mmHg;PA pressure: systolic 15~25 mmHg;

    diastolic 8~16 mmHgdiastolic 8~16 mmHg

    Mean PAP: 10~20 mmHgMean PAP: 10~20 mmHg

    Pulmonary Capillary Wedge PressurePulmonary Capillary Wedge Pressure

    (PCWP): 6~12 mmHg(PCWP): 6~12 mmHg

    Derived HemodynamicDerived Hemodynamic

  • 8/14/2019 8-Monitoring During Anesthesia

    86/108

    86

    Derived HemodynamicDerived Hemodynamic

    ProfilesProfiles

    Systemic Vascular ResistanceSystemic Vascular Resistance

    (SVR): 80 x (MAP-CVP)/CO;(SVR): 80 x (MAP-CVP)/CO;

    800~1200 dyne-sec-cm800~1200 dyne-sec-cm-5-5

    Pulmonary Vascular ResistancePulmonary Vascular Resistance

    (PVR): 80 x (PAP-PCWP)/CO;(PVR): 80 x (PAP-PCWP)/CO;20~130 dyne-sec-cm20~130 dyne-sec-cm-5-5

    Derived HemodynamicDerived Hemodynamic

  • 8/14/2019 8-Monitoring During Anesthesia

    87/108

    87

    Derived HemodynamicDerived Hemodynamic

    ProfilesProfiles

    Cardiac Output: thermodilutionCardiac Output: thermodilution

    method; 4~8 L/minmethod; 4~8 L/min

    Cardiac Index: CO/BSA; 2.5~4.2Cardiac Index: CO/BSA; 2.5~4.2L/min/mL/min/m22

    Continuous CardiacContinuous Cardiac

  • 8/14/2019 8-Monitoring During Anesthesia

    88/108

    88

    Continuous CardiacContinuous Cardiac

    OutputOutput

    Continuous Cardiac Output (CCO)Continuous Cardiac Output (CCO)

    measurement can be achieved bymeasurement can be achieved by

    a electric coil attached on the tipa electric coil attached on the tipof PA catheter. It automaticallyof PA catheter. It automatically

    measures CO every 3 min.measures CO every 3 min.

    Mixed Venous OxygenMixed Venous Oxygen

  • 8/14/2019 8-Monitoring During Anesthesia

    89/108

    89

    Mixed Venous OxygenMixed Venous Oxygen

    Saturation (SvOSaturation (SvO22))

    Mixed by blood from both SVCMixed by blood from both SVC

    and IVC, sampled at PAand IVC, sampled at PA

    OO22 consumption= SaOconsumption= SaO22-SvO-SvO22SvOSvO22=SaO=SaO22 - (VO- (VO22/Q x Hb x 13)/Q x Hb x 13)

  • 8/14/2019 8-Monitoring During Anesthesia

    90/108

  • 8/14/2019 8-Monitoring During Anesthesia

    91/108

    Clinical monitoring ofClinical monitoring of

  • 8/14/2019 8-Monitoring During Anesthesia

    92/108

    92

    Clinical monitoring ofClinical monitoring of

    ventilationventilation

    Continuous observation should beContinuous observation should bemade of the following :made of the following :

    1. patient's colour .1. patient's colour .

    2. respiratory rate .2. respiratory rate .3. adequacy of chest movement .3. adequacy of chest movement .

    4. movement of the reservoir bag or4. movement of the reservoir bag or

    ventilator bellows .ventilator bellows .

    Clinical monitoring ofClinical monitoring of

  • 8/14/2019 8-Monitoring During Anesthesia

    93/108

    93

    Clinical monitoring ofClinical monitoring of

    ventilationventilation Auscultation of both lung fields shouldAuscultation of both lung fields should

    also be performed frequently in orderalso be performed frequently in order

    to detect :to detect :

    1. equality of air entry .1. equality of air entry .

    2. intubation of a bronchus .2. intubation of a bronchus .

    3. presence of secretions .3. presence of secretions .

    4. occurrence of a pneumothorax .4. occurrence of a pneumothorax .

    Clinical monitoring ofClinical monitoring of

  • 8/14/2019 8-Monitoring During Anesthesia

    94/108

    94

    Clinical monitoring ofClinical monitoring of

    ventilationventilation

    Anaesthetist must check regularly for signs ofAnaesthetist must check regularly for signs of

    respiratory obstruction as evidenced by:respiratory obstruction as evidenced by:

    1. tracheal tug .1. tracheal tug .

    2. paradoxical abdominal movement .2. paradoxical abdominal movement .

    3. absence of bag deflation .3. absence of bag deflation .

    Some ventilators make a regular noise duringSome ventilators make a regular noise during

    part of the ventilating cycle and this is apart of the ventilating cycle and this is avaluable audible monitor.valuable audible monitor.

    Measurement of airwayMeasurement of airway

  • 8/14/2019 8-Monitoring During Anesthesia

    95/108

    95

    Measurement of airwayMeasurement of airway

    pressurepressure

    Airway pressure may reflect changes in lung andAirway pressure may reflect changes in lung and

    chest wall compliance :chest wall compliance :

    Chest wall compliance may be influenced by :Chest wall compliance may be influenced by :

    1. degree of muscle paralysis .1. degree of muscle paralysis .

    2. surgical manipulation .2. surgical manipulation .

    3. position of the patient .3. position of the patient .

    Measurement of airwayMeasurement of airway

  • 8/14/2019 8-Monitoring During Anesthesia

    96/108

    96

    Measurement of airwayMeasurement of airway

    pressurepressure

    Lung compliance may be influenced by :Lung compliance may be influenced by :

    1. accumulation of secretions .1. accumulation of secretions .

    2. development of a pneumothorax .2. development of a pneumothorax .

    3. position of the patient .3. position of the patient .

    Increased resistance to air flow caused byIncreased resistance to air flow caused by

    bronchospasm or obstruction of the trachealbronchospasm or obstruction of the trachealtube is reflected by an increased peak airwaytube is reflected by an increased peak airway

    pressure.pressure.

    Measurement of airwayMeasurement of airway

  • 8/14/2019 8-Monitoring During Anesthesia

    97/108

    97

    Measurement of airwayMeasurement of airway

    pressurepressure

    1.1. Kinking of ventilator tubing or tracheal tube.Kinking of ventilator tubing or tracheal tube.

    2.2. Overinflation of the tracheal tube cuff withOverinflation of the tracheal tube cuff with

    consequent obstruction of the lumen of the tube.consequent obstruction of the lumen of the tube.3.3. Increased secretions.Increased secretions.

    4.4. Pneumothorax.Pneumothorax.

    5.5. Bronchospasm.Bronchospasm.

    6.6. Inadequate muscle relaxation.Inadequate muscle relaxation.

    ses of elevation of airway pressure

  • 8/14/2019 8-Monitoring During Anesthesia

    98/108

    98

    Disconnection alarmDisconnection alarm

    When the lungs are ventilated mechanically, theWhen the lungs are ventilated mechanically, the

    continuity of the anaesthetic breathing system,continuity of the anaesthetic breathing system,

    and thus of gas delivery to the patient, should beand thus of gas delivery to the patient, should be

    monitored using a disconnection alarmmonitored using a disconnection alarm The alarm is activated if the airway pressureThe alarm is activated if the airway pressure

    decreases below a preset minimum for a presetdecreases below a preset minimum for a preset

    time interval.time interval.

  • 8/14/2019 8-Monitoring During Anesthesia

    99/108

    99

    Disconnection alarmDisconnection alarm

    A large leak, or total disconnection, is indicated ifA large leak, or total disconnection, is indicated ifthe alarm is triggered.the alarm is triggered.

    In addition, most of these devices sound an alarmIn addition, most of these devices sound an alarm

    if excessive airway pressures are generated.if excessive airway pressures are generated.

    A disconnection alarm does not obviate the needA disconnection alarm does not obviate the need

    for visual surveillance of the continuity of thefor visual surveillance of the continuity of the

    breathing system.breathing system.

    End-tidal carbon dioxideEnd-tidal carbon dioxide

  • 8/14/2019 8-Monitoring During Anesthesia

    100/108

    100

    End tidal carbon dioxideEnd tidal carbon dioxide

    tension (PE'cotension (PE'co22)) PE'coPE'co

    22

    , correlates well with Paco, correlates well with Paco22

    , in patients who, in patients who

    have no significant pulmonary disease.have no significant pulmonary disease.

    Normal PacoNormal Paco22-PE'co-PE'co22 gradient is approximatelygradient is approximately

    5mmHg. (Paco5mmHg. (Paco22 35-45mmHg, PE'co35-45mmHg, PE'co22 30-40mmHg.)30-40mmHg.)

    End-tidal carbon dioxide concentration may beEnd-tidal carbon dioxide concentration may be

    measured using the principle of infrared absorptionmeasured using the principle of infrared absorption

    spectrophotometry.spectrophotometry.

    PPETETCOCO

    22

    End-tidal carbon dioxideEnd-tidal carbon dioxide

  • 8/14/2019 8-Monitoring During Anesthesia

    101/108

    101

    End tidal carbon dioxideEnd tidal carbon dioxide

    tension (PE'cotension (PE'co22))

    PE'coPE'co22 is useful particularly in the followingis useful particularly in the followingcircumstances.circumstances.

    b.b. To provide evidence of correct placement of theTo provide evidence of correct placement of the

    tracheal tube. Capnography is the only methodtracheal tube. Capnography is the only method

    available which provides rapid and reliableavailable which provides rapid and reliablediagnosis of intubation of the oesophagus.diagnosis of intubation of the oesophagus.

    c.c. For routine monitoring of the adequacy ofFor routine monitoring of the adequacy of

    ventilation and the effects of IPPV.ventilation and the effects of IPPV.

    d.d. To detect rebreathing.To detect rebreathing.

    PPETETCOCO

    22

    End-tidal carbon dioxideEnd-tidal carbon dioxide

  • 8/14/2019 8-Monitoring During Anesthesia

    102/108

    102

    End tidal carbon dioxideEnd tidal carbon dioxide

    tension (PE'cotension (PE'co22))

    To detect air, fat or pulmonary embolism; a suddenTo detect air, fat or pulmonary embolism; a suddendecrease in PE'codecrease in PE'co22 occurs as a result of increasedoccurs as a result of increased

    dead spacedead space

    n To detect malignant hyperthermia; a progressiveTo detect malignant hyperthermia; a progressive

    increase in PE'coincrease in PE'co22, results from increased muscle, results from increased musclemetabolism.metabolism.

    n To ensure normocapnia in elderly patients in anTo ensure normocapnia in elderly patients in an

    attempt to maintain adequate cerebral perfusion.attempt to maintain adequate cerebral perfusion.

    n To maintain normal PE'coTo maintain normal PE'co22 during carotid arteryduring carotid artery

    surgery in order to maintain cerebral perfusion.surgery in order to maintain cerebral perfusion.

    PPETETCOCO

    22

  • 8/14/2019 8-Monitoring During Anesthesia

    103/108

    103

    7-5 a Normal CO2 waveform

    b CO2 drop to zero Disconnection

    c CO2 decrease gradually

    hyperventilation

    d CO increase raduall h oventilation

  • 8/14/2019 8-Monitoring During Anesthesia

    104/108

    104

    PARTPART

    OTHER SYSTEMOTHER SYSTEM

    Measurement ofMeasurement of

  • 8/14/2019 8-Monitoring During Anesthesia

    105/108

    105

    easu e e o

    TemperatureTemperature General anaesthesia inhibits the patientsGeneral anaesthesia inhibits the patients

    ability to maintain body temperature byability to maintain body temperature by

    depressing the thermoregulatory centre indepressing the thermoregulatory centre in

    the hypothalamusthe hypothalamus

    Heat loss during anaesthesia is potentiatedHeat loss during anaesthesia is potentiated

    by surgery of long duration and exposure oby surgery of long duration and exposure of

    large surface areas of tissue, e.g. thelarge surface areas of tissue, e.g. the

    abdominal contents during gastrointestinalabdominal contents during gastrointestinal

    operations.operations.

    Measurement ofMeasurement of

  • 8/14/2019 8-Monitoring During Anesthesia

    106/108

    106

    TemperatureTemperature The use of wet packs and dry inspiredThe use of wet packs and dry inspired

    gases compounds the problem.gases compounds the problem.

    These sources of heat loss assume These sources of heat loss assume

    even more importance in children,even more importance in children,

    especially small babies, whose surfaceespecially small babies, whose surface

    area is much larger in proportion toarea is much larger in proportion to

    body weight than in the adult.body weight than in the adult.

    Measurement ofMeasurement of

  • 8/14/2019 8-Monitoring During Anesthesia

    107/108

    107

    TemperatureTemperature

    1. The operating room temperature should be as high1. The operating room temperature should be as high

    as is comfortable for the theater staff.as is comfortable for the theater staff.

    2. A warming mattress should be placed beneath the2. A warming mattress should be placed beneath the

    patient.patient.3. Exposed surfaces should be swaddled with warm3. Exposed surfaces should be swaddled with warm

    gauze or foil, especially in neonates.gauze or foil, especially in neonates.

    4. All i.v. infusion fluids should be warmed.4. All i.v. infusion fluids should be warmed.

    5. Inspired gases should be warmed and humidified.5. Inspired gases should be warmed and humidified.

    easurement to minimize heat loss

    Measurement ofMeasurement of

  • 8/14/2019 8-Monitoring During Anesthesia

    108/108

    TemperatureTemperature

    1. The nasopharynx (approximates to brain1. The nasopharynx (approximates to brain

    temperature)temperature)

    2. The oesophagus (approximates to cardiac2. The oesophagus (approximates to cardiac

    temperature).temperature).

    3. The tympanic membrane (best for core3. The tympanic membrane (best for core

    temperature, but the membrane is delicate andtemperature, but the membrane is delicate and

    easily damaged)easily damaged)

    4. The rectum.4. The rectum.

    The probe may be placed in thefollowing positions in order to measurecore temperature: