Collision or Reaction Mode Collision/Reaction Cells (CRCs) in ICP-MS can operate either as reaction cells (when a reactive gas is used) or as collision cells (when an inert gas is used). Since the first commercial CRC ICP-MS systems were introduced 10 years ago, the relative performance of these two approaches has been the subject of intense debate. With a reactive cell gas, interference removal is based on the relative rates at which the selected reaction gas reacts with the analyte and interfering ion pair. In order to select which reactive gas should be used, the interference must therefore be identified in advance, which means the matrix must be known, relatively simple, and consistent. In some applications (e.g. semiconductor process chemicals) these criteria are met, and reaction mode can be used successfully. However, in most applications the sample matrix is unknown, variable, and often complex. Under these circumstances, reactive cell gases will lead to serious analytical problems: Each reaction gas will only react with certain interferences; unreactive interferences will remain, leading to errors. All reaction gases will create some new polyatomic ions (cell-formed product ions) causing new, matrix-dependent interferences. Reaction gases react with some analyte ions, leading to severe signal loss for some analytes, degrading detection limits. With an inert cell gas, interference removal is mainly by kinetic energy discrimination (KED). This process works because polyatomic interferences are by definition molecular ions, and have a larger ionic cross section than analyte (single atom) ions at the same mass. Polyatomics collide more often with the cell gas and lose more energy, and so can be prevented from passing into the analyzer by applying a fixed KED bias voltage. The key benefits of collision mode are that it can remove multiple interferences on multiple analytes, no new polyatomic interferences are produced regardless of the sample matrix, and selective analyte signal loss by reaction does not occur. The 3 rd generation Octopole Reaction System (ORS 3 ) featured in the Agilent 7700 Series ICP-MS has been developed to give the highest performance in helium (He) collision mode. This requires an extremely narrow initial ion energy spread (delivered by the ShieldTorch System), and the capability to maintain high ion transmission at the high cell gas pressure needed for effective KED (provided by the small internal volume, octopole-based ORS 3 cell). Helium Mode in the ORS 3 Figure 1 illustrates the relative performance of different cell modes for the removal of multiple interferences on the 1 st row transition elements. No gas mode (top), H 2 reaction mode (middle) and He mode (bottom) spectra are shown for the same matrices (5% HNO 3 , 5% HCl, 1% H 2 SO 4 and 1% IPA), on the same intensity scale. The complex polyatomic interferences created in 7700 ORS 3 and Helium Mode More effective interference removal in complex samples this matrix mix are evident in the no gas mode spectra shown in Figure 1 (top). Figure 1 (middle) shows that H 2 reaction mode leaves some residual overlaps, and creates several new ones. By contrast, Figure 1 (bottom) confirms that He mode effectively removes all the polyatomic interferences in this complex matrix. High sensitivity is maintained in He mode, as shown by the inset spectrum of a 10 ppb spike measured under the same conditions. Figure 1. Mixed matrix in no gas, H 2 and He mode (inset: with 10 ppb spike) For more information on the 7700 Series ICP-MS visit the Agilent Technologies web site at: www.agilent.com/chem/icpms © Agilent Technologies, Inc., 2011, February 21, 2011, 5990-7574EN

Upload: others

Post on 12-Feb-2022

5 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: 7700 ORS3 and Helium Mode - United States Home | Agilent

Collision or Reaction Mode Collision/Reaction Cells (CRCs) in ICP-MS can operate either asreaction cells (when a reactive gas is used) or as collision cells(when an inert gas is used). Since the first commercial CRC ICP-MSsystems were introduced 10 years ago, the relative performanceof these two approaches has been the subject of intense debate.

With a reactive cell gas, interference removal is based on therelative rates at which the selected reaction gas reacts with theanalyte and interfering ion pair. In order to select which reactivegas should be used, the interference must therefore be identifiedin advance, which means the matrix must be known, relativelysimple, and consistent.

In some applications (e.g. semiconductor process chemicals)these criteria are met, and reaction mode can be used successfully.However, in most applications the sample matrix is unknown,variable, and often complex. Under these circumstances, reactivecell gases will lead to serious analytical problems:• Each reaction gas will only react with certain interferences;

unreactive interferences will remain, leading to errors.• All reaction gases will create some new polyatomic ions

(cell-formed product ions) causing new, matrix-dependent interferences.

• Reaction gases react with some analyte ions, leading to severesignal loss for some analytes, degrading detection limits.

With an inert cell gas, interference removal is mainly by kineticenergy discrimination (KED). This process works because polyatomicinterferences are by definition molecular ions, and have a largerionic cross section than analyte (single atom) ions at the samemass. Polyatomics collide more often with the cell gas and losemore energy, and so can be prevented from passing into the analyzer by applying a fixed KED bias voltage.

The key benefits of collision mode are that it can remove multipleinterferences on multiple analytes, no new polyatomic interferencesare produced regardless of the sample matrix, and selective analytesignal loss by reaction does not occur.

The 3rd generation Octopole Reaction System (ORS3) featured in theAgilent 7700 Series ICP-MS has been developed to give the highestperformance in helium (He) collision mode. This requires anextremely narrow initial ion energy spread (delivered by theShieldTorch System), and the capability to maintain high iontransmission at the high cell gas pressure needed for effective KED(provided by the small internal volume, octopole-based ORS3 cell).

Helium Mode in the ORS3

Figure 1 illustrates the relative performance of different cellmodes for the removal of multiple interferences on the 1st rowtransition elements. No gas mode (top), H2 reaction mode(middle) and He mode (bottom) spectra are shown for the samematrices (5% HNO3, 5% HCl, 1% H2SO4 and 1% IPA), on the sameintensity scale. The complex polyatomic interferences created in

7700 ORS3 and Helium ModeMore effective interference removal in complex samples

this matrix mix are evident in the no gas mode spectra shown inFigure 1 (top).

Figure 1 (middle) shows that H2 reaction mode leaves some residualoverlaps, and creates several new ones. By contrast, Figure 1(bottom) confirms that He mode effectively removes all thepolyatomic interferences in this complex matrix. High sensitivityis maintained in He mode, as shown by the inset spectrum of a10 ppb spike measured under the same conditions.

Figure 1. Mixed matrix in no gas, H2 and He mode (inset: with 10 ppb spike)

For more information on the 7700 Series ICP-MS visit the AgilentTechnologies web site at: www.agilent.com/chem/icpms

© Agilent Technologies, Inc., 2011, February 21, 2011, 5990-7574EN