Quick-Install Guide
Serial No.
Date Installed
Installed By If you have any questions regarding this installation, call your
local Waters office. In the U.S. call 1-800-252-HPLC.
The quality management system of Waters' chromatography
applications software design and manufacturing facility,
Milford, Massachusetts, complies with the International
Standard ISO 9001 Quality Management and Quality
Assurance Standards. Waters' quality management system is
periodically audited by the registering body to ensure
compliance.
Waters® 2465
Electrochemical Detector
Waters and Connections are registered
trademarks of Waters Corporation.
Printed in U.S.A. May, 2004
715000483 Rev. B
Waters Corporation
34 Maple Street
Milford, MA 01757
508 478-2000
http://www.waters.com
Waters
3 Power Up
After completing the electrical
and fluidic connections, power up
the unit using the switch on the
rear panel. The screens at the left
appear.
CAUTION
To prevent substantial damage to
the working electrode or the
electronics, never power-on the
flow cell when:
• the cell cable is not properly
connected,
• the flow cell is not filled with
buffer, or,
• the outside of the flow cell is
wet.
Refer to the 2465 Electrochemical Detector Operator's Guide for more
information on startup
procedures.
Flow Cell Type
Reference Electrode
Working Electrode
FIRST DISPLAYON POWER UP
CHECKING MEMORY OKCHECKSUM OK
SECOND DISPLAYON POWER UP
WATERS 2465 EC DETECTOR MAIN
DIAGSCANPULSEDCCONFIGfirmware version 2.00
FINAL DISPLAYON POWER UP
INSTRUMENT READY
Electrochemical Detector2465Waters
CHECKING MEMORY . . . . .CHECKSUM . . . .
NOTE: The following are the basic steps to install the 2465 detector. Check the after completing each section. For more detailed information and cautionary statements regarding this product, refer to the 2465 Electrochmechanical Detector Operator's Manual.
NOTE: The following are the basic steps to install the 2465 detector. Check the after completing each section. For more detailed information and cautionary statements regarding this product, refer to the 2465 Electrochemical Detector Operator's Guide.
Electrical Connections1 Fluid Connections2NOTE: The following is an example of typical
fluid connections. Your system may require a different setup. All tubing is PEEK.
Continued on the back page.
Connectors A and B
Signals on these connectors let the
2465 control (or be controlled by)
external equipment. Together with
time files, these signals provide a
powerful tool for the development of
automated methods.
Refer to the 2465 Electrochemical Detector Operator's Guide for more
information on I/O signal
connections.
RS-232
Power
Connect an RS-232 cable
between the detector and
an Empower™ personal
computer, workgroup or
enterprise client.
Connect the power cord to
a suitable receptacle.
CONNECTOR
WORKINGELECTRODE
BLOCK
INLETBLOCK
LC IN(from column)
LC OUT(to waste)
AUX(BLUE)
WORK(RED)
REF(BLACK)
FLOW CELL ASSEMBLY
Unpack the unit and lift it from the
shipping box.
NOTE: Never lift the unit by its front door, only by its sides.
Unpacking
1. If using a pulse dampener, run
tubing from the pulse dampener
and through the hole at the
bottom of the side panel. Connect
this tubing to the inlet (bottom) of
the column.
2. Connect the oulet (top) of the
column to the flow cell's inlet
block. Rotate the flow cell holder
to 45º, as shown in the
illustration, to avoid bubble
formation.
Also, keep the tubing connecting
the column to the flow cell as
short as possible.
3. Feed a piece of tubing from the
outlet of the working electrode
block and through the hole at the
top of the unit's side panel. This
tubing goes to waste.
NOTE: Place the waste container above the level of the flow cell to supply back pressure.
Refer to the 2465 Electrochemical Detector Operator's Guide for more
information on fluidic connections.
To install your 2465 Detector:
A
B RS-232
ANALOGOUTPUT
1 8
159
1 8
159
5
9
1
6
For instrument control, data acquisition
For chart marker, inject marker.
Event inputs and outputs(See chart above)
Voltage output via SAT/IN to Breeze system, Millenium32 client or workstation, or recording device.
GROUND STUD
123456789101112131415
123456789101112131415
GNDGNDN/CN/C
MARKN/CN/CN/CN/CN/CN/CN/C
INJECT MARKERN/CN/C
RELAY 1
RELAY 2
CELL ONRESET
OVERLOADAUX 1AUX 2
CELL OFFSTART
AUTO ZEROGND
Connector A
BLACKBROWN
REDORANGEYELLOWGREENBLUE
PURPLEGREY
WHITEPINK
LIGHT GREENBLACK with STRIPE
BROWN with STRIPERED with STRIPE
Wire Color Connector B
IN
OUT
110V WAT097194220V WAT097410