5975 inert msd - agilent

72
5975 inert MSD Chris Sandy European GC-MS Product Specialist Agilent Technologies

Upload: others

Post on 29-Dec-2021

25 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: 5975 inert MSD - Agilent

5975 inert MSD

Chris SandyEuropean GC-MS Product SpecialistAgilent Technologies

Page 2: 5975 inert MSD - Agilent

Presenter: Chris Sandy

Welcome to our e-Seminar:

Introduction of the new 5975 inert MSD for expanding your application capabilities

Page 3: 5975 inert MSD - Agilent

5975

Che

mSta

tion

eMet

hod

Sum

mar

y

Slide 3Chairperson: John Vis, Audio: +44 207 162 0125

What Has Changed?

What’s the same?

• Reliability

• Quality

• Inertness

But NOT Performance – it’s better!

Page 4: 5975 inert MSD - Agilent

5975

Che

mSta

tion

eMet

hod

Sum

mar

y

Slide 4Chairperson: John Vis, Audio: +44 207 162 0125

What’s Different?

• Functional industrial design

• More efficient vacuum system

• Extended mass range

• AutoCI

• New sensitivity specifications

• Synchronous SIM/Scan

• EI with CI ion source

• 6850 GC control

• QuickSwap MSD interface

• Review of differentiators

Page 5: 5975 inert MSD - Agilent

5975

Che

mSta

tion

eMet

hod

Sum

mar

y

Slide 5Chairperson: John Vis, Audio: +44 207 162 0125

Quick Remove Top Cover

Easy analyzer access

Page 6: 5975 inert MSD - Agilent

5975

Che

mSta

tion

eMet

hod

Sum

mar

y

Slide 6Chairperson: John Vis, Audio: +44 207 162 0125

Functional Front Window

• Wire connections

• Filament operation

• Column insertion

• CI Source identification

Page 7: 5975 inert MSD - Agilent

5975

Che

mSta

tion

eMet

hod

Sum

mar

y

Slide 7Chairperson: John Vis, Audio: +44 207 162 0125

Modified MSD mainframe

for quick/easy turbo pump

maintenance

Modified Mainframe

New turbo and mechanical vacuum pumps are more efficient for Helium

and Hydrogen

• New 70 L/sec standard pump

• New 262 L/sec performance pump

• New 2.5 m3/hour mechanical pump

(67% more capacity)

Page 8: 5975 inert MSD - Agilent

5975

Che

mSta

tion

eMet

hod

Sum

mar

y

Slide 8Chairperson: John Vis, Audio: +44 207 162 0125

G3397A – optional gauge/controller

Integrated Ion Gauge/Controller

• Greater vacuumprecision

• Excellent for trouble shooting small leaks –every system should have one!

Software displayLocal display

Page 9: 5975 inert MSD - Agilent

5975

Che

mSta

tion

eMet

hod

Sum

mar

y

Slide 9Chairperson: John Vis, Audio: +44 207 162 0125

New Sensitivity Specifications –EI Scan

182:1 s/n100:160:1EI1 pg OFN

59755973

Page 10: 5975 inert MSD - Agilent

5975

Che

mSta

tion

eMet

hod

Sum

mar

y

Slide 10Chairperson: John Vis, Audio: +44 207 162 0125

New Sensitivity Specifications –CI Scan

125:175:1PCI100 pg BZP

300:1 (100 fg)

500:1(1 pg)

NCI1 pg/100 fg

OFN

59755973

704:1 s/n (100 fg OFN under NCI)

Page 11: 5975 inert MSD - Agilent

5975

Che

mSta

tion

eMet

hod

Sum

mar

y

Slide 11Chairperson: John Vis, Audio: +44 207 162 0125

Extended Mass Range

• 1050 m/z for 5975 inert MSD

• Quadrupole modified

• New electronics

• 800 m/z for all 5973 MSDs

(Software upgrade does not enable 1050 m/z)

Page 12: 5975 inert MSD - Agilent

5975

Che

mSta

tion

eMet

hod

Sum

mar

y

Slide 12Chairperson: John Vis, Audio: +44 207 162 0125

Extended Mass Range – PBDE 209

Page 13: 5975 inert MSD - Agilent

5975

Che

mSta

tion

eMet

hod

Sum

mar

y

Slide 13Chairperson: John Vis, Audio: +44 207 162 0125

959

957

961

955

963

953

965

960 962

958956

953

954

955

956

957

958960

961

962

963

964

965

959

953 954 955 956 957 958 959 960 961 962 963 964 965

5975 inert MSD Theoretical

Extended Mass Range –PBDE 209

Page 14: 5975 inert MSD - Agilent

5975

Che

mSta

tion

eMet

hod

Sum

mar

y

Slide 14Chairperson: John Vis, Audio: +44 207 162 0125

AutoCI

• AutoCI – Fully automated setup/tuning

• No external manual control

• CI made as easy as EI!

The new 5975

CI interface

The old 5973 manual CI interface

Page 15: 5975 inert MSD - Agilent

5975

Che

mSta

tion

eMet

hod

Sum

mar

y

Slide 15Chairperson: John Vis, Audio: +44 207 162 0125

Automated EI with CI Source

• March 2004: 5973 MSD

• Application note on EI with CI ion source

5989-0595EN

• Now: 5975 MSD

• Automated switching between EI and CI in a

sequence

Page 16: 5975 inert MSD - Agilent

5975

Che

mSta

tion

eMet

hod

Sum

mar

y

Slide 16Chairperson: John Vis, Audio: +44 207 162 0125

Automated EI with CI Source

Setup steps:

1. Install CI Source

2. Tune instrument in PCI – create PCICH4.u

3. Auto create EI tune file – create ei-cisource.u

4. Tune instrument in NCI – NCICH4.u

5. Assign tune file with method and run in sequence

Page 17: 5975 inert MSD - Agilent

5975

Che

mSta

tion

eMet

hod

Sum

mar

y

Slide 17Chairperson: John Vis, Audio: +44 207 162 0125

6850 GC Control via ChemStation

• 5973N MSD (G2570A)

• 5973 inert MSDs

• 5975 inert MSDs

Page 18: 5975 inert MSD - Agilent

5975

Che

mSta

tion

eMet

hod

Sum

mar

y

Slide 18Chairperson: John Vis, Audio: +44 207 162 0125

Agilent MSD Unique Features

• Inert Ion source

• Hyperbolic quad

• Quad temp

• CI system

• Analyzer modularity

Page 19: 5975 inert MSD - Agilent

Synchronous SIM/Scan

Page 20: 5975 inert MSD - Agilent

5975

Che

mSta

tion

eMet

hod

Sum

mar

y

Slide 20Chairperson: John Vis, Audio: +44 207 162 0125

Set Up Synchronous SIM/Scan

• SIM/Scan data

analysis requiresmixed mode dataview.

• SIM/Scan data issaved in the samedata file directory.

Page 21: 5975 inert MSD - Agilent

5975

Che

mSta

tion

eMet

hod

Sum

mar

y

Slide 21Chairperson: John Vis, Audio: +44 207 162 0125

Comparison of SIM/scan Implementations

Implementation (1) – ‘coarse granularity’

Agilent Implementation– ‘fine granularity’ for whole acquisition time

Implementation (2) – ‘fine granularity’ at specific times

11.20 11.40 11.60 11.80 12.00 12.20 12.40 12.60 12.80 13.00 13.20 13.40

Scan SIM

TIME

Page 22: 5975 inert MSD - Agilent

5975

Che

mSta

tion

eMet

hod

Sum

mar

y

Slide 22Chairperson: John Vis, Audio: +44 207 162 0125

• AUTOSIM is a standard feature of G1701DA Software

• AUTOSIM provides rapid and automatic conversion of a Scan method to SIM/Scan

• AUTOSIM creates SIM method based on Calibration table

• The operator has fine control over SIM parameter settings

• The SIM ions/groups/times are saved automatically to the method

Synchronous SIM/Scan –AutoSIM

Page 23: 5975 inert MSD - Agilent

5975

Che

mSta

tion

eMet

hod

Sum

mar

y

Slide 23Chairperson: John Vis, Audio: +44 207 162 0125

Inject Standard

Page 24: 5975 inert MSD - Agilent

5975

Che

mSta

tion

eMet

hod

Sum

mar

y

Slide 24Chairperson: John Vis, Audio: +44 207 162 0125

AutoSIM

Application note: 5988-4188EN

Page 25: 5975 inert MSD - Agilent

5975

Che

mSta

tion

eMet

hod

Sum

mar

y

Slide 25Chairperson: John Vis, Audio: +44 207 162 0125

Edit SIM Method

AutoSIM automatically creates SIM groups and populates ions in each group

Page 26: 5975 inert MSD - Agilent

5975

Che

mSta

tion

eMet

hod

Sum

mar

y

Slide 26Chairperson: John Vis, Audio: +44 207 162 0125

Synchronous SIM/Scan

Page 27: 5975 inert MSD - Agilent

5975

Che

mSta

tion

eMet

hod

Sum

mar

y

Slide 27Chairperson: John Vis, Audio: +44 207 162 0125

1.00 2.00 3.00 4.00 5.00 6.00 7.00 8.00 9.00 10.0011.0012.0013.00

10000

30000

50000

70000

90000

110000

130000TIC: 1 ppb 60 VOCs.D\DATA.MS

1.00 2.00 3.00 4.00 5.00 6.00 7.00 8.00 9.00 10.0011.0012.0013.00

10000

30000

50000

70000

90000

TIC: 1 ppb 60 VOCs.D\DATASIM.MS

Scan and SIM Chromatograms from SIM/Scan Analysis of 60 VOCs at 1 ppb Each

SIM

Scan

• Velocity XPT/6890N/5975 P&T/GC/MS system• 20-m X 0.18 mm X 1.0 µmDB-VRX column• Scan sampling rate = 2; Dwell =20 ms for SIM

Page 28: 5975 inert MSD - Agilent

5975

Che

mSta

tion

eMet

hod

Sum

mar

y

Slide 28Chairperson: John Vis, Audio: +44 207 162 0125

Synchronous SIM/Scan Analysis of VOC Target Gases at 1 ppb each

1.20 1.30 1.40 1.50 1.60 1.70 1.80 1.90 2.000

1000

2000

3000

4000

5000

6000

7000

8000

9000

10000

1.151.201.251.301.351.401.451.501.551.601.651.701.751.801.851.901.950

1000

2000

3000

4000

5000

6000

7000

8000

9000

10000

Scan

SIM

1

2 3

4 5

6

1

2 3

4 5

6

VOC Gases (1 ppb each)

1. Dichlorodifluoromethane

2. Chloromethane

3. Vinyl chloride

4. Bromomethane

5. Ethyl chloride

6. Trichlorofluoromethane

Page 29: 5975 inert MSD - Agilent

5975

Che

mSta

tion

eMet

hod

Sum

mar

y

Slide 29Chairperson: John Vis, Audio: +44 207 162 0125

SIM/Scan and Scan-Only Sensitivity Comparison: 5-mL Purge of 1 ppb Ethyl Chloride (a Low Responding Gas)

2075Scan(SIM/Scan)

1773Scan(Scan only)

254749SIM(SIM/Scan)

Peak/Peak S/N

RMS S/N

• SIM as part of SIM/Scan run gives 10X greater sensitivity

• Still get ≥≥≥≥ 8 scans across peak using same scan rate (n = 2)

Page 30: 5975 inert MSD - Agilent

5975

Che

mSta

tion

eMet

hod

Sum

mar

y

Slide 30Chairperson: John Vis, Audio: +44 207 162 0125

Why Our SIM/Scan is So Good

22

Performance Electronics

Nearly no fall off with increasing scan speed

21100

50

Phenanthrene

20 ~ 6250 amu/sec

23 ~ 781 amu/sec

Nearly no fall off with increasing scan speed

Standard Electronics

Fall off with increasing scan speed

21

22

100

50

Phenanthrene

23 ~ 781 amu/sec

Fall off with increasing scan speed

20 ~ 6250 amu/sec

Moving from 23 to 22

doubles scans

Page 31: 5975 inert MSD - Agilent

5975

Che

mSta

tion

eMet

hod

Sum

mar

y

Slide 31Chairperson: John Vis, Audio: +44 207 162 0125

For a 0.25 mm column, a typical peak width (at base) is 6 seconds. Switch from scan at 2^2 mode to SIM/scan@2^1, there will still be the same 15 cycles over the peak, but gets the 4-ion SIM group data for FREE.

Scan ���� SIM/scan45 to 555 => 510 amu

Scan

45 to 555 amu, Scan only @ 2^2

2.87 scan/sec1563 amu/s

SIM

4 ions, 40 ms

1 sec 2 sec

2^06250 amu/s

3.76 cycle/s

2^13125 amu/s

2.88 cycle/s

2^21563 amu/s

1.94 cycle/s

Same width (SIM)

a cycle

Page 32: 5975 inert MSD - Agilent

5975

Che

mSta

tion

eMet

hod

Sum

mar

y

Slide 32Chairperson: John Vis, Audio: +44 207 162 0125

Synchronous SIM/Scan –Key Features

• Performance Electronics really make the difference

• SIM Improvements – 100 groups and 60 ions per group

• Minimum dwell time is reduced by half to 5 ms

• Even more data points per SIM peak

• Synchronous SIM/Scan available on 5973 MSDs with performance electronics (req. Smart Card 3+)

• Upgrades: G1088A and G1088B AND SW upgrade

Page 33: 5975 inert MSD - Agilent

5975

Che

mSta

tion

eMet

hod

Sum

mar

y

Slide 33Chairperson: John Vis, Audio: +44 207 162 0125

Synchronous SIM/ScanApplication examples

•Allergens analysis: Proposed 5975 method, 26 Targets

SIM/Scan mode

- SIM table built with AUTOSIM

- Quant with SIM target ion

- Confirmation by DRS with Scan data

•Pesticides analysis with microfluidics technology for flexible GC and GC/MS analysis

Page 34: 5975 inert MSD - Agilent

5975

Che

mSta

tion

eMet

hod

Sum

mar

y

Slide 34Chairperson: John Vis, Audio: +44 207 162 0125

Allergens analysis in SIM/Scan

After shave1% v/v 1ul Split 20 : 1SCAN

SIM

Page 35: 5975 inert MSD - Agilent

5975

Che

mSta

tion

eMet

hod

Sum

mar

y

Slide 35Chairperson: John Vis, Audio: +44 207 162 0125

Allergens Overlaid SIM/Scan

Page 36: 5975 inert MSD - Agilent

5975

Che

mSta

tion

eMet

hod

Sum

mar

y

Slide 36Chairperson: John Vis, Audio: +44 207 162 0125

Allergens SIM/Scan analysis and deconvolution with DRS

With courtesy of Chris Sandy and J-F Garnier

Quant based on SIM target ion

Confirmation based on deconvoluted full mass spectrum

Page 37: 5975 inert MSD - Agilent

5975

Che

mSta

tion

eMet

hod

Sum

mar

y

Slide 37Chairperson: John Vis, Audio: +44 207 162 0125

Synchronous SIM/Scan –Summary

• Great for target analysis and identifying unknowns

• SIM data for quantitation

• Scan data for confirmation of unknowns

• Our SIM/Scan delivers real performance based on superior electronics

Page 38: 5975 inert MSD - Agilent

Overview of MicrofluidicTechnologies

Page 39: 5975 inert MSD - Agilent

5975

Che

mSta

tion

eMet

hod

Sum

mar

y

Slide 39Chairperson: John Vis, Audio: +44 207 162 0125

Connections to Microfluidic Splitter

•Deactivated, very low internal volume to prevent peak tailing orband broadening

•A single, special design metal ferrule

• More inert that graphite/vespel, does not leak at high oven temperatures

To Detector 1To Detector 2

Plate

Metal

Ferrule

Nut

Column

Channel

Page 40: 5975 inert MSD - Agilent

5975

Che

mSta

tion

eMet

hod

Sum

mar

y

Slide 40Chairperson: John Vis, Audio: +44 207 162 0125

Microfluidic Splitter hardware installation

Deactivated microfluidicsplitter operates to 350 C.

Uses metal ferrules to eliminate leaks and

retightening

To MSD

To Det 1

Column in

Makeup from Aux EPC

Page 41: 5975 inert MSD - Agilent

5975

Che

mSta

tion

eMet

hod

Sum

mar

y

Slide 41Chairperson: John Vis, Audio: +44 207 162 0125

AUX EPC

MSD

Column in

Blocked

Microfluidic Splitter operating modes * (1) Single Column

* Aux EPC Flow enables column change-over without venting MSD

* Aux EPC Flow enables inlet / septum maintenance without ingressof air into column.

Page 42: 5975 inert MSD - Agilent

5975

Che

mSta

tion

eMet

hod

Sum

mar

y

Slide 42Chairperson: John Vis, Audio: +44 207 162 0125

Microfluidic Splitter operating modes (2) Splitter

AUX EPCColumn in

MSD

Detector 1

* Aux EPC Flow enables column change-over without venting MSD

* Aux EPC Flow enables inlet / septum maintenance without ingressof air into column.

Page 43: 5975 inert MSD - Agilent

5975

Che

mSta

tion

eMet

hod

Sum

mar

y

Slide 43Chairperson: John Vis, Audio: +44 207 162 0125

Splitting to 2 Detectors

Effluent Splitter

with Makeup

6890N GC

Column

AUX EPC

3.8 psig

0.814 m X 0.18 mm id

1.1 m X 0.18 mm id

15 m X 0.25 mm id X 0.25 um HP-5MS

Configuration for 2:1 Split to MSD

mECD

Auto-sampler

5975 InertMSD

Page 44: 5975 inert MSD - Agilent

5975

Che

mSta

tion

eMet

hod

Sum

mar

y

Slide 44Chairperson: John Vis, Audio: +44 207 162 0125

Microfluidic Splitter – ‘Chromatographically Transparent’

MSD

ECD

Page 45: 5975 inert MSD - Agilent

5975

Che

mSta

tion

eMet

hod

Sum

mar

y

Slide 45Chairperson: John Vis, Audio: +44 207 162 0125

100 ppb VOAs, 100 ppm Gasoline in Water, Ambient Headspace

0 2 4 6 8 10 12

0 2 4 6 8 10 12

0 2 4 6 8 10 12

Scan

ECD

SIM

Page 46: 5975 inert MSD - Agilent

5975

Che

mSta

tion

eMet

hod

Sum

mar

y

Slide 46Chairperson: John Vis, Audio: +44 207 162 0125

AUX EPCColumn 1 / Inlet 1

MSD

Column 2 / Inlet 2

Microfluidic Splitter operating modes :(3) Dual Column Inlet

* Aux EPC Flow enables column change-over without venting MSD

* Aux EPC Flow enables inlet / septum maintenance without ingressof air into column.

Page 47: 5975 inert MSD - Agilent

5975

Che

mSta

tion

eMet

hod

Sum

mar

y

Slide 47Chairperson: John Vis, Audio: +44 207 162 0125

Dual Column Configuration*

Back : 30m x 0.25mm ID x 0.25um HP5-MS

Front : 12.4m x 0.25mm ID x 0.25um HP5-MS

25 mins

6 mins

*Requires performance turbo pump

Page 48: 5975 inert MSD - Agilent

5975

Che

mSta

tion

eMet

hod

Sum

mar

y

Slide 48Chairperson: John Vis, Audio: +44 207 162 0125

AUX EPCColumn 1 / Inlet 1

MSD

Column 2 / Inlet 2

Microfluidic Splitter operating modes :(4) Dual Column Inlet + Olfactory Port + MSD

ODO II* Aux EPC Flow enables column change-over without venting MSD

* Aux EPC Flow enables inlet / septum maintenance without ingressof air into column.

Page 49: 5975 inert MSD - Agilent

5975

Che

mSta

tion

eMet

hod

Sum

mar

y

Slide 49Chairperson: John Vis, Audio: +44 207 162 0125

AUX EPC3.8 psig Column inlet

23.73 psig

MSD

Detector 1

Splitter mode – acquiring data

AUX EPC25.0 psig Column inlet

1.0 psig

MSD

Backflush mode

Detector 1

Microfluidic Splitter operating modes *

(5) Splitter mode with Backflush

* Aux EPC Flow enables column

change-over without venting MSD* Aux EPC Flow enables inlet / septum

maintenance without ingress of air into column

Page 50: 5975 inert MSD - Agilent

5975

Che

mSta

tion

eMet

hod

Sum

mar

y

Slide 50Chairperson: John Vis, Audio: +44 207 162 0125

6890N/5975 inert Pesticide Analysiswith 3-Way Splitter

3-way Effluent Splitter with Makeup

Auto-sampler

6890N GC

Column

Phosphorus FPD

30 m X 0.25 mm id X 0.25 um HP-5MS

1X method with 10:10:1 split FPD:MSD:ECD

5975InertMSD

uECDAUX EPC

3.8 psig

Page 51: 5975 inert MSD - Agilent

5975

Che

mSta

tion

eMet

hod

Sum

mar

y

Slide 51Chairperson: John Vis, Audio: +44 207 162 0125

1-injection-4-signal System

• 5975, RTLocked, 30m x 0.25mm x 0.25um HP-5MS

• Samples: pesticide mix at 10, 1, and 0.1 ppm

• Injection: 10:1 split

• At the end of the column, eluent flow is split 3 ways: MSD:FPD:µECD = 10:10:1

10:1 split

SIM/Scan

µECD

FPD

~480 pg

~480 pg

~48 pg

Injection 10 ppm (10000 pg)

1000 pg

Page 52: 5975 inert MSD - Agilent

5975

Che

mSta

tion

eMet

hod

Sum

mar

y

Slide 52Chairperson: John Vis, Audio: +44 207 162 0125

5975 SIM/scan + 3-way splitterUltimate Pesticide System

Scan: ~ 480 pg

SIM: ~ 480 pg

FPD: ~ 480 pg7 peaks

5.00 10.00 15.00 20.00 25.00 30.00 35.00 40.00 45.00

µECD: ~ 48 pg

Page 53: 5975 inert MSD - Agilent

5975

Che

mSta

tion

eMet

hod

Sum

mar

y

Slide 53Chairperson: John Vis, Audio: +44 207 162 0125

QuickSwap MSD interfaceAvailable October 1st, 2005

• Replacement for NoVent

• Remove column w/o venting

• Backflush mode

• Inlet maintenance reverse flow

• Removes heavies from column

• Maintain constant flow to MSD

MS

Transferline

Column

EPC inlet

Microfluidic technology

Requires performance turbo for flow rates > 2 cc/min

Page 54: 5975 inert MSD - Agilent

Introduction to eMethods and the new MSD Chemstation

Page 55: 5975 inert MSD - Agilent

5975

Che

mSta

tion

eMet

hod

Sum

mar

y

Slide 55Chairperson: John Vis, Audio: +44 207 162 0125

What are eMethods?

• eMethods is short for Electronic Methods.

• Can now package existingGC/MSD methods

• Export them to other 5973 MSDs

• Export them to 5975 inert MSDs

Page 56: 5975 inert MSD - Agilent

5975

Che

mSta

tion

eMet

hod

Sum

mar

y

Slide 56Chairperson: John Vis, Audio: +44 207 162 0125

eMethod Benefits

• Fast setup of new 5975 inertMSD

• Method sharing between5973 and 5975 MSDs

• Download apps from websiteand import entire method intoMSD ChemStation

Page 57: 5975 inert MSD - Agilent

5975

Che

mSta

tion

eMet

hod

Sum

mar

y

Slide 57Chairperson: John Vis, Audio: +44 207 162 0125

New 5975 inert eMethods

• PBDE Flame Retardants

• SIM/Scan – Pesticides in Water and Food

• VOCs via Headspace – European method

• Residual Solvents via Headspace

• NIDA Drugs

Page 58: 5975 inert MSD - Agilent

5975

Che

mSta

tion

eMet

hod

Sum

mar

y

Slide 58Chairperson: John Vis, Audio: +44 207 162 0125

MSD ChemStation Software

• Two GC/MS perChemStation

• GC and GC/MS control

• Retention time locking

• Custom reports

• eMethods

• Four configurations

• Enhanced mode

• EnviroQuant

• DrugQuant

• Reformulated Gas

Most popular GC/MS software in the world

Page 59: 5975 inert MSD - Agilent

5975

Che

mSta

tion

eMet

hod

Sum

mar

y

Slide 59Chairperson: John Vis, Audio: +44 207 162 0125

Instrument Control Enhancements

• Combined Top/Instrument Control Screen

• New icons for quick access to tune, EMF and print

• Configurable sequence table

• More right mouse click functions in editing sequences

• Printing selected sections of method and sequence

• Import of GC ChemStation method parameters(excluding 5890 GC)

• Display of IP address(s) of GC and MSD instruments

Page 60: 5975 inert MSD - Agilent

5975

Che

mSta

tion

eMet

hod

Sum

mar

y

Slide 60Chairperson: John Vis, Audio: +44 207 162 0125

Data Analysis Enhancements

• Explorer-style navigation tool

• Interactive editing of Integration Parameters

• Switchable right mouse click functions

• Spreadsheet-like SIM Parameters Table

• Icons for custom tools, DA options, variable watch

• Toolbar on/off capabiity

• Selectable DA color themes

Page 61: 5975 inert MSD - Agilent

5975

Che

mSta

tion

eMet

hod

Sum

mar

y

Slide 61Chairperson: John Vis, Audio: +44 207 162 0125

Data Analysis Enhancements

•QEdit functionality is improved:

• Separate window for Baseline Display

• Generating reports directly from QEdit

• Custom configuring X-axis

Page 62: 5975 inert MSD - Agilent

5975

Che

mSta

tion

eMet

hod

Sum

mar

y

Slide 62Chairperson: John Vis, Audio: +44 207 162 0125

Explorer Navigation

Explorer navigation

Page 63: 5975 inert MSD - Agilent

5975

Che

mSta

tion

eMet

hod

Sum

mar

y

Slide 63Chairperson: John Vis, Audio: +44 207 162 0125

Interactive Edit Integration

Page 64: 5975 inert MSD - Agilent

5975

Che

mSta

tion

eMet

hod

Sum

mar

y

Slide 64Chairperson: John Vis, Audio: +44 207 162 0125

Right Click Features

Page 65: 5975 inert MSD - Agilent

5975

Che

mSta

tion

eMet

hod

Sum

mar

y

Slide 65Chairperson: John Vis, Audio: +44 207 162 0125

SIM Parameter Table

Page 66: 5975 inert MSD - Agilent

5975

Che

mSta

tion

eMet

hod

Sum

mar

y

Slide 66Chairperson: John Vis, Audio: +44 207 162 0125

QEdit

Reports

Baseline Display

Page 67: 5975 inert MSD - Agilent

5975

Che

mSta

tion

eMet

hod

Sum

mar

y

Slide 67Chairperson: John Vis, Audio: +44 207 162 0125

Library Options

• NIST05

• Wiley 7th ed

• Stan pesticide

• Pfleger/Maurer/Weber MS Drug Library, 3rd ed

• NIST Chemical Structures Library

NEW

Page 68: 5975 inert MSD - Agilent

5975

Che

mSta

tion

eMet

hod

Sum

mar

y

Slide 68Chairperson: John Vis, Audio: +44 207 162 0125

Other Software Options

• Deconvolution Reporting Software

- New Revison A.02 released Aug 1, 2005

• Security ChemStation

• Headspace software

• RTL Pesticide Database : FAMES, Flavors, Volatiles Organics, Organotin derivatives, PCB congeners, Forensic toxicology

Page 69: 5975 inert MSD - Agilent

5975

Che

mSta

tion

eMet

hod

Sum

mar

y

Slide 69Chairperson: John Vis, Audio: +44 207 162 0125

Expanding the portfolio of RTL / DRS Databases

• Updated RTL Pesticides Database, Nov 2005

Adding ~ 200 new compounds

• TIC Toxic Industrial Chemicals Database, Nov 2005

~ 800 compounds (VOC, SVOC, Dioxins, PCBs andPesticides)

Page 70: 5975 inert MSD - Agilent

5975

Che

mSta

tion

eMet

hod

Sum

mar

y

Slide 70Chairperson: John Vis, Audio: +44 207 162 0125

GC/MS Configurations

• 5973N EI with diffusion pump

• 5975 inert EI with standard turbo pump

• 5975 inert EI with performance turbo pump

• 5975 XL EI/PCI/NCI with performance turbo pump

All systems available with choice of either 6850 or 6890 GCs

Page 71: 5975 inert MSD - Agilent

5975

Che

mSta

tion

eMet

hod

Sum

mar

y

Slide 71Chairperson: John Vis, Audio: +44 207 162 0125

Summary

Agilent 5975Inert MSD

• Better reliability and quality

• Enhanced performance

• eMethods

• AutoSIM and SIM/scan

• Higher mass range

• AutoCI

• Sensitivity

• Software improvements

Page 72: 5975 inert MSD - Agilent

5975

Che

mSta

tion

eMet

hod

Sum

mar

y

Slide 72Chairperson: John Vis, Audio: +44 207 162 0125

Thank you for attending today’s Agilent e-Seminar.

Our e-Seminar schedule is expanding regularly.

Please check our web site frequently at:

www.agilent.com/chem

If you would like to receive regular updates please