3 he polarization tests at uiuc danielle chandler david howell uiuc
TRANSCRIPT
3He Polarization Tests at UIUC
Danielle ChandlerDavid Howell
UIUC
Overview
• Confirmation of successful 3He polarization using 1083-nm optical signal
• Use of optical signal to calibrate perturbing field (B1)
• NMR setup and initial tests
Tests using optical signalcell
light detector
cell
Confirmation of 3He Polarization
Baseline drops on the 1083-nm light signal when B0 is turned on
1083-nm light produced in cell decreases when 3He is polarized because polarized atoms do not interact with the circularly-polarized laser light
Response of optical signal to continuous oscillatory B1 field
20
2
10
)γ/ωB(B
BA)ωA(
1
Scanned through the B1 resonance frequency and recorded the optical signal
At resonance, the perturbing field rotates the 3He spins away from the B0 axis, which makes the optical pumping more efficient… therefore there is a peak in the optical signal at resonance.
Results of B1 frequency scans
3 1Peak width 2 3 γ B2
02
1
)γ/ωB(B
BheightPeak
1
1
0
1
peak height is independent of B
on resonance, B ω/ 0, or for
sufficiently large B
Non-zero width as B0 → 0 reflectsnonuniformity in B0/B1 fields
B1 dependence suggests B0/B1 nonuniformity
Response of optical signal as a function of time
Amplitude of oscillation damps out as cell becomes depolarized
Signal decays as laser light repolarizes the 3He cell
end of pulse
Optical signal peaks when the pulse time gives a π pulse, since this should orient the spins antiparallel to the B0 axis, making optical pumping maximally efficient
Using optical signal to calibrate π/2 burst pulse
4
T timepulse
2
π
First maximum corresponds to π pulse
T
B1 program voltage = 2.5 Vpp
For a given B1:
Can use this optical method to calibrate B1:
γT
2π
(T/4)2γ
πB1
NMR Measurement Setup
Adiabatic fast passage method
Free induction decay method
NMR Measurements
With current setup, we observe the sensitivity to detect NMR signal at a level of 0.2 μV
Need to improve the design of the NMR for a better signal/noise ratio
Calculate that NMR signal should be of the same order of magnitude
Response of NMR coil to B1 pulse burst
Conclusion
Optical signal tests were successful and provide a method of calibrating B1
Preparing to make improvements on NMR sensitivity
(for example)increase number of turns on pickup coilsoptimize position and geometry of coilsoptimize Q-factorintroduce second pair of pickup coils