217A Multiparameter Patient Simulator217A Multiparameter Patient SimulatorThe 217A Multiparameter Patient Simulator has been labeled solid and dependable by satisfied users worldwide. This “veteran" is quite popular with customersrequiring a comprehensive and portable multiparameter patient simulator.
This product has been discontinued. This product has been discontinued. Service and repair is guaranteed until October 17, 2011.Service and repair is guaranteed until October 17, 2011. We recommend the We recommend the in place of this product.in place of this product.
Standard Accessories
Miscellaneous AccessoriesMiscellaneous Accessories
Controller/Co Simulator
217A-21A, Controller/Co Simulator (part# 2246913)(for use with 217A)
9530-0019FG
9530-0019FG Overlay kit (part# 2392790)(for use with 21A Controller for 217A)
Cables (IBP)Cables (IBP)
3010-0048FG
3010-0048FG unterminated UU-1 BP cable (part# 2392173)
3010-0076
3010-0076 TK-1 (6M) BP Cable (part# 2198879)
3010-0110
HP-3 BP cable (12M 5\u00b5V) 120 in, Mini-DIN to 12 pin HP/PH (part# 2198902)
3010-0111
3010-0111 HP-4 BP cable (12M 40\u00b5V) (part# 2198916)
3010-0114
Siemens-SM-1, BP cable
3010-0115
3010-0115 Mennen Medical MM-1 (6M) (6-pin) (part# 2198933)
3010-0116
3010-0116 PC-1 BP cable (6M) (6-pin) (part# 2198940)
3010-0122
3010-0122 GE Marquette Medical MQ-2 (8M round) (7-pin) (part# 2198978)
3010-0208
Nihon Kohden-NK-1 BP cable (5M) (5-pin) (part# 2462263)
3010-0248
QM-1 (6M) Blood pressure cable (10-pin orange) (part# 2198925)
3010-0308
Unterminated BP cable 4100-01 (part# 2199393)(7-pin DIN, one end only)
3010-0357
GE Marquette Medical MQ-3 BP cable (rectangular -11 M) (part# 2199627)
3010-0366
3010-0366 SM-3 BP cable (15M) (part# 2199666)
3010-0584
3010-0584 DX-1 BP cable (10F) (10-pin) (part# 2200955)
CablesCables
11451
E-input meas. Cable 2M, black
3010-0117 (part number)
SpaceLabs (ALPHA9/703R) SL-1 (5M) BP Cable
3010-0120
LT-1 BP cable - (intrauterine pressure – 10M) (part# 2198969)
Provides interface with the following:
Airshields (Fetascan 1400)Advanced Medical Systems (FM 660, FM 670)
3010-0206 (part number)
Kontron Medical (Mini-, Super-, Color-Mon series) KT-1 (6M) BP Cable
3010-0284
3010-0284 COA-1 Cardiac output adapter (part# 2199240)
4100-04 HP cable (12M)
Hewlett Packard 5\u00b5V (78300, -500, -800, and Merlin/Viridia/Omnicare) 4100-04 (12 meter)
HP/Philips HP-9 BP cable
HP/Philips (8040A) HP-9 (intrauterine pressure only – 12M 5 \u00b5V) BP Cable
3010-0377
3010-0377 DS-A IABP Sync Datascope (part# 2199747)(for use with 217A/medSim 300B)
3010-0386FG
3010-0386FG CI-2B Interface Cable (part# 2392224)(for use with 21A Controller to 217A)
3010-0441
Interface cable, medTester to IDA 4 Plus (without wedge adapter) (RS-232; female DB9 to female DB25) (part# 2200102) (medTester toSigmaPace\™1000/PC/Index 2XL/IDA 4 Plus [IDA 4 to wedge requires cable# 2201042]; Impulse 4000 to PC)
Carrying CasesCarrying Cases
9530-0062
9530-0062, Multipurpose Hard-Sided Watertight Carrying Case 1400 (part# 2248546)(for use with 217A)
9530-0064
Multipurpose Hard-Sided Watertight Carrying Case (part# 2248568)(contains “pick & pluck” foam) (17” L x 11 \u00be” W x 6” H)
Patient Monitor SimulatorsPatient Monitor Simulators
PT MONITOR W/ECG/RESP/BP/TEMP(ZOLL)
SIM TEST SET USA,TEST SET
217A-multiparameter-patient-simulator217A-multiparameter-patient-simulator
ECG GeneralECG General
Lead Configuration:Lead Configuration:
12-lead ECG with nine independent outputs for each signal leadreferenced to RL
Output Update Rate:Output Update Rate:
2 ms
Output Impedance:Output Impedance:
1000 \u03a9 to RL
1 V Output:1 V Output:
Lead II at 1 V peak
Amplitude Accuracy:Amplitude Accuracy:
5 % 2 Hz square wave at 1 mV p-p (Lead II)
Normal Sinus RhythmNormal Sinus Rhythm
Rates:Rates:
30, 60, 80, 120, 160, 200, 240, and 300 BPM
Amplitudes:Amplitudes:
0.5, 1.0, 1.5, 2.0 mV (Lead II)
ST Segment:ST Segment:
0.05, 0.10, 0.15, 0.20, 0.50, and 0.80 mV (positive and negativepolarity)
Axis Deviation:Axis Deviation:
Normal, horizontal, and vertical
Pediatric ECG:Pediatric ECG:
40 ms R-Wave width
ECG PerformanceECG Performance
Square Wave:Square Wave:
2 Hz at 1 mV bipolar
Pulse:Pulse:
4 s at 1 mV
Sine Waves:Sine Waves:
10, 40, 50, 60, and 100 Hz at 1 mV
Triangle Wave:Triangle Wave:
2 Hz and 3 mV
ArrhythmiasArrhythmias
Selections:Selections:
45 total ventricular, supraventricular, premature, conduction, andpacemaker
Base Rhythm:Base Rhythm:
80 BPM, 1.0 mV/Lead II
Pacemaker Spike Duration:Pacemaker Spike Duration:
0.15 ms
Pacemaker Spike Amplitude:Pacemaker Spike Amplitude:
4.8 mV on lead II, 2.4 mV on leads I and III and V1-V6, 5 %
Output ConnectorsOutput Connectors
High Level:High Level:
0.25 in standard phone jack
Low Level:Low Level:
Ten AHA (American Heart Association) or European (IEC) color-coded binding posts (as ordered)
Binding posts compatible with disposable snaps and 3.2 or 4.0 mmelectrodes
TemperatureTemperature
TemperatureTemperature
Channel One:Channel One:
Fixed at 37 \u00b0C/98.6 \u00b0F
Channel Two:Channel Two:
30 \u00b0C / 86 \u00b0F; 37 \u00b0C / 98.6 \u00b0 F; 40\u00b0C / 104 \u00b0F (selectable)
Temp Probe Compatibility:Temp Probe Compatibility:
400/700 YSI series
Accuracy:Accuracy:
0.25 \u00b0C
ArtifactArtifact
ECG:ECG:
50 Hz, 60 Hz, muscle and baseline wander. Any single artifact can beadded to the ECG selection at three different levels to all or to anysingle-signal lead.
BP/Respiration:BP/Respiration:
Two pressure values in phase with respiration can be injected intoall BP channels. Arterial and LV waveforms are modulated at theselected respiratory rate. All right-heart waveforms have therespiration offset signal added to the output.
RespirationRespiration
Baseline Impedances:Baseline Impedances:
500 \u03a9 or 1000 \u03a9
Impedance Variations:Impedance Variations:
0.0, 0.1, 0.2, 0.5, 1.0, and 3.0 \u03a9
Rates:Rates:
15, 20, 30, 40, 60, and 120 BrPM
Apnea:Apnea:
Continuous (off), momentary, 12 s. and 32 s
Ventilator Simulation:Ventilator Simulation:
Rate 40 BPM, delta impedance 3 \u03a9
Output Configuration:Output Configuration:
Lead I, lead II / RL-LL
Blood Pressure (Three Channels)Blood Pressure (Three Channels)
Input /Output Impedance:Input /Output Impedance:
300 \u03a9
Exciter Input Voltage Range:Exciter Input Voltage Range:
2 to 10 V
Exciter Input Frequency Range:Exciter Input Frequency Range:
DC to 4000 Hz
Transducer Sensitivity:Transducer Sensitivity:
5 to 40 uV/V/mmHg
Level Accuracy:Level Accuracy:
1 % of range + 1 % of setting + 1 count
Calibrated Rate:Calibrated Rate:
80 BPM
Waveform Update Rate:Waveform Update Rate:
10 ms
BP Selections:BP Selections:
Dynamic (mmHg):Dynamic (mmHg):
BP1BP1ATM(0)Arterial (120/80)Left ventricle (120/0)Right ventricle (25/0)Pulmonary artery (25/10)PA wedge (10/2)
BP2BP2ATM (0)Central venous pressure (15//10)Right ventricle (25/0)Pulmonary artery (25/10)PA wedge (10/2)
BP3BP3
Simulates a Swan-Ganz catheterization procedure, including right-heart navigation, and balloon inflation/deflation and removal
Static (mmHg):Static (mmHg):
BP1BP1
ATM (0), 10, 20, 40, 80, 100, 200, 250, and 300
BP2BP2
5, 10, 20, and 30
BP3BP3
(No static level selections)
Arrhythmia SelectionsArrhythmia Selections
SupraventricularSupraventricularAtrial fibrillation 1Atrial fibrillation 2Atrial flutter
Sinus arrhythmia
Sinus arrhythmiaMissed beat 80 BPMMissed beat 120 BPMParoxysmal atrial tachycardia (PAT)NodalSupraventricular tachycardia
Premature beatsPremature beatsAtrial (PAC)Nodal (PNC)PVC 1 StandardPVC 1 EarlyPVC 1 R on TPVC 2 StandardPVC 2 EarlyPVC 2 R on TMultifocal PVCs
Conduction defectsConduction defectsFirst degreeSecond-degree type 1Second-degree type 2Third degreeRight-bundle-branch blockLeft-bundle-branch block
Ventricular rhythmsVentricular rhythmsPVCs 6/minutePVCs 12/minutePVCs 24/minuteFrequent multifocal PVCs 20/minBigeminyTrigeminyPair PVCsRun of 5 PVCsRun of 11 PVCsVentricular tachycardiaV-Fibrillation (fine)V-Fibrillation (coarse)Dying heart complex (EMD)Asystole
Pacemaker (transvenous)Pacemaker (transvenous)AsynchronousDemand with frequent sinus beatDemand with occasional sinus beatAV sequentialNoncaptureNonfunction
Defibrillator TrainingDefibrillator Training
Emergency 1:Emergency 1:
Normal Sinus Rhythm (NSR) with PVC activity terminating into V-FIB
Emergency 2:Emergency 2:
NSR with a run of five PVCs terminating into V-FIB
Cardioversion:Cardioversion:
Atrial fibrillation is output until defibrillator is discharged, at whichpoint conversion to NSR will occur if properly synchronized with theR-Wave. If the discharge is not synchronized, the ECG output willdegrade to V-FIB. Emergency defibrillation must then be performedto achieve conversion to NSR.
Intra-Aortic Balloon Assist Pump (IABP) SimulationIntra-Aortic Balloon Assist Pump (IABP) Simulation
Demonstrates the interactive simulation of the patient and the IABP.Synchronization signal from the IABP augments the simulated arterialpressure waveform.
Available ECG Simulations:Available ECG Simulations:
Normal sinus rhythm(30, 60, 80, and 120 BPM)Atrial fibrillationPremature ventricular contraction (PVCs 6/min)Asynchronous pacemaker (75 BPM)Run of five PVCsRun of eleven PVCs
CaseCase
High-impact polystyrene
PowerPower
9 V alkaline battery or battery eliminator (115 V or 230 V available)
DimensionsDimensions
7.5 in L x 5.4 in W x 1.8 in H (19.0 cm L x 13.8 cm W x 4.4 cm H)
WeightWeight
2 lb (0.9 kg)