The Future of NVIS Display MeasurementThe Future of NVIS Display Measurement
ALI Advanced Seminar, 2006ALI Advanced Seminar, 2006 Optronic LaboratoriesOptronic Laboratories
The Future of NVIS DisplayThe Future of NVIS DisplayThe Future of NVIS Display
MeasurementMeasurementMeasurement
Dr. Richard Young
Optronic Laboratories, Inc.
-
8/6/2019 1224787448p28 Future of Nvis Display Msmt
2/30
The Future of NVIS Display MeasurementThe Future of NVIS Display Measurement
ALI Advanced Seminar, 2006ALI Advanced Seminar, 2006 Optronic LaboratoriesOptronic Laboratories
IntroductionIntroductionIntroduction
Thanks to the excellent quality of modern
scanning instruments, NVIS compatibility
measurements have become almost routine.
What improvements could instrument
manufacturers make
-
8/6/2019 1224787448p28 Future of Nvis Display Msmt
3/30
The Future of NVIS Display MeasurementThe Future of NVIS Display Measurement
ALI Advanced Seminar, 2006ALI Advanced Seminar, 2006 Optronic LaboratoriesOptronic Laboratories
IntroductionIntroductionIntroduction
Better sensitivity, so low luminance small spot
sizes can be measured as easily as highluminance large spot sizes.
Less variation in results
Faster scans Greater portability and ruggedness
Lower cost
A wish-list for improvements
-
8/6/2019 1224787448p28 Future of Nvis Display Msmt
4/30
The Future of NVIS Display MeasurementThe Future of NVIS Display Measurement
ALI Advanced Seminar, 2006ALI Advanced Seminar, 2006 Optronic LaboratoriesOptronic Laboratories
Sensitivity and VariabilitySensitivity and VariabilitySensitivity and VariabilityAn Example of OL 750S-NVG performance vs MIL sensitivity specification
1.0E-15
1.0E-14
1.0E-13
1.0E-12
1.0E-11
1.0E-10
1.0E-09
350 400 450 500 550 600 650 700 750 800 850 900 950
Wavlength [nm]
Radiance@1
0:1S/N[W/(cmn
msr)]
0.18mm spot size
0.3mm spot size
0.6mm spot size
1.8mm spot size
3.0mm spot size
MIL specification
Much of theMuch of the
variability isvariability is
due to thedue to the
longerlongerwavelengthswavelengths
Scanning systems haveScanning systems have
plenty of sensitivityplenty of sensitivity
-
8/6/2019 1224787448p28 Future of Nvis Display Msmt
5/30
The Future of NVIS Display MeasurementThe Future of NVIS Display Measurement
ALI Advanced Seminar, 2006ALI Advanced Seminar, 2006 Optronic LaboratoriesOptronic Laboratories
Sensitivity and VariabilitySensitivity and VariabilitySensitivity and Variability
OL 750S-NVG and OL 770-NVS performance vs MIL sensitivity specification
1.0E-15
1.0E-14
1.0E-13
1.0E-12
1.0E-11
1.0E-10
1.0E-09
350 400 450 500 550 600 650 700 750 800 850 900 950
Wavlength [nm]
Ra
diance@1
0:1S/N[W/(cmnmsr)]
0.18mm spot size
0.3mm spot size
0.6mm spot size
1.8mm spot size
3.0mm spot size
OL 770-NV
@ 0.17mm spot size
MIL specification
There is noThere is no
dramaticdramatic
change inchange inlongerlonger
wavelengthwavelength
sensitivitysensitivity
The OL 770The OL 770--NV detectoris based onNV detectoris based on
silicon, andhence has a different shape,silicon, andhence has a different shape,
-
8/6/2019 1224787448p28 Future of Nvis Display Msmt
6/30
The Future of NVIS Display MeasurementThe Future of NVIS Display Measurement
ALI Advanced Seminar, 2006ALI Advanced Seminar, 2006 Optronic LaboratoriesOptronic Laboratories
Spectroradiometer designSpectroradiometer designSpectroradiometer design
De ector
Exit lit
Entrance lit
-
8/6/2019 1224787448p28 Future of Nvis Display Msmt
7/30
The Future of NVIS Display MeasurementThe Future of NVIS Display Measurement
ALI Advanced Seminar, 2006ALI Advanced Seminar, 2006 Optronic LaboratoriesOptronic Laboratories
Spectroradiometer designSpectroradiometer designSpectroradiometer design
Detector
Exit lit
Entrance lit
Rotation
-
8/6/2019 1224787448p28 Future of Nvis Display Msmt
8/30
The Future of NVIS Display MeasurementThe Future of NVIS Display Measurement
ALI Advanced Seminar, 2006ALI Advanced Seminar, 2006 Optronic LaboratoriesOptronic Laboratories
Spectroradiometer designSpectroradiometer designSpectroradiometer design
Multi- h l
ete tor
Entr n Slit
NORot tion
(Each l ntb haves likea
detector+ slit)
-
8/6/2019 1224787448p28 Future of Nvis Display Msmt
9/30
The Future of NVIS Display MeasurementThe Future of NVIS Display Measurement
ALI Advanced Seminar, 2006ALI Advanced Seminar, 2006 Optronic LaboratoriesOptronic Laboratories
The Multiplex AdvantageThe Multiplex AdvantageThe Multiplex Advantage
When all the light is seen all thetime:
Noise is decreased. Sensitivity is increased.
Measurements can be faster.
The multiplex advantage means that an instrument simultaneously
measuring a signal over a range of frequencies obtains a t1/2 advantage
in the time t required to obtain a given signal-to-noise ratio compared
to that which would be necessary using dispersive methods if the
noise is detector-limited.
http://scienceworld.wolfram.com/physics/MultiplexAdvantage.html
-
8/6/2019 1224787448p28 Future of Nvis Display Msmt
10/30
The Future of NVIS Display MeasurementThe Future of NVIS Display Measurement
ALI Advanced Seminar, 2006ALI Advanced Seminar, 2006 Optronic LaboratoriesOptronic Laboratories
Spectroradiometer designSpectroradiometer designSpectroradiometer design
The OL 770-NVS is a special type ofspectroradiometer, many decades betterthanother multi-channel devices.
Far higher sensitivity Much lower noise
Orders of magnitude lower stray light
Vastly improved electronics
Dedicated software
State-of-the-arttechnology
-
8/6/2019 1224787448p28 Future of Nvis Display Msmt
11/30
The Future of NVIS Display MeasurementThe Future of NVIS Display Measurement
ALI Advanced Seminar, 2006ALI Advanced Seminar, 2006 Optronic LaboratoriesOptronic Laboratories
MIL Spec: OthersMIL Spec: OthersMIL Spec: Others
A.3.3. Wavelength accuracy & repeatability
Accuracytypical
-
8/6/2019 1224787448p28 Future of Nvis Display Msmt
12/30
The Future of NVIS Display MeasurementThe Future of NVIS Display Measurement
ALI Advanced Seminar, 2006ALI Advanced Seminar, 2006 Optronic LaboratoriesOptronic Laboratories
MIL Spec: OthersMIL Spec: OthersMIL Spec: Others
A.3.6. Linearity
In scale
-
8/6/2019 1224787448p28 Future of Nvis Display Msmt
13/30
The Future of NVIS Display MeasurementThe Future of NVIS Display Measurement
ALI Advanced Seminar, 2006ALI Advanced Seminar, 2006 Optronic LaboratoriesOptronic Laboratories
MIL Spec: OthersMIL Spec: OthersMIL Spec: Others
A.3.9. Spectroradiometer Optics
Full scale at
-
8/6/2019 1224787448p28 Future of Nvis Display Msmt
14/30
The Future of NVIS Display MeasurementThe Future of NVIS Display Measurement
ALI Advanced Seminar, 2006ALI Advanced Seminar, 2006 Optronic LaboratoriesOptronic Laboratories
Other AdvantagesOther AdvantagesOther Advantages
Sincethere are no moving parts
the construction is rugged and compact
results are highly repeatable
timetaken in moving the monochromator,
changing filters etc. is avoided
-
8/6/2019 1224787448p28 Future of Nvis Display Msmt
15/30
The Future of NVIS Display MeasurementThe Future of NVIS Display Measurement
ALI Advanced Seminar, 2006ALI Advanced Seminar, 2006 Optronic LaboratoriesOptronic Laboratories
Other AdvantagesOther AdvantagesOther Advantages
Sincethere is no photomultiplier
A veryexpensive component is avoided
The detector is not damaged by strong light
High voltages are not required
Since all the wavelengths are measured at
the sametime
It doesnt matter ifthe source is d.c., a.c.,
modulated or flashed
-
8/6/2019 1224787448p28 Future of Nvis Display Msmt
16/30
The Future of NVIS Display MeasurementThe Future of NVIS Display Measurement
ALI Advanced Seminar, 2006ALI Advanced Seminar, 2006 Optronic LaboratoriesOptronic Laboratories
An Intercomparison StudyAn Intercomparison StudyAn Intercomparison Study
-
8/6/2019 1224787448p28 Future of Nvis Display Msmt
17/30
The Future of NVIS Display MeasurementThe Future of NVIS Display Measurement
ALI Advanced Seminar, 2006ALI Advanced Seminar, 2006 Optronic LaboratoriesOptronic Laboratories
An Intercomparison StudyAn Intercomparison StudyAn Intercomparison Study
Green BGreen B NVIS RedNVIS RedNVIS YellowNVIS Yellow
-
8/6/2019 1224787448p28 Future of Nvis Display Msmt
18/30
The Future of NVIS Display MeasurementThe Future of NVIS Display Measurement
ALI Advanced Seminar, 2006ALI Advanced Seminar, 2006 Optronic LaboratoriesOptronic Laboratories
An Intercomparison StudyAn Intercomparison StudyAn Intercomparison StudySpectral Radiance of Samples
1.0E-12
1.0E-11
1.0E-10
1.0E-09
1.0E-08
1.0E-07
1.0E-06
350 400 450 500 550 600 650 700 750 800 850 900 950
Wavelength [nm]
SpectralRadiance[Wcm-
nm
-1]
NVI
Green B
-
8/6/2019 1224787448p28 Future of Nvis Display Msmt
19/30
The Future of NVIS Display MeasurementThe Future of NVIS Display Measurement
ALI Advanced Seminar, 2006ALI Advanced Seminar, 2006 Optronic LaboratoriesOptronic Laboratories
An Intercomparison StudyAn Intercomparison StudyAn Intercomparison StudySpectral Radiance of Samples
1.0E-11
1.0E-10
1.0E-09
1.0E-08
1.0E-07
1.0E-06
350 400 450 500 550 600 650 700 750 800 850 900 950
Wavelength [nm]
SpectralRadiance[Wcm-
nm
-1]
NVI
Yellow
-
8/6/2019 1224787448p28 Future of Nvis Display Msmt
20/30
The Future of NVIS Display MeasurementThe Future of NVIS Display Measurement
ALI Advanced Seminar, 2006ALI Advanced Seminar, 2006 Optronic LaboratoriesOptronic Laboratories
An Intercomparison StudyAn Intercomparison StudyAn Intercomparison StudySpectral Radiance of Samples
1.0E-11
1.0E-10
1.0E-09
1.0E-08
1.0E-07
1.0E-06
350 400 450 500 550 600 650 700 750 800 850 900 950
Wavelength [nm]
SpectralRadiance[Wcm-
nm
-1]
NVI
Red
-
8/6/2019 1224787448p28 Future of Nvis Display Msmt
21/30
The Future of NVIS Display MeasurementThe Future of NVIS Display Measurement
ALI Advanced Seminar, 2006ALI Advanced Seminar, 2006 Optronic LaboratoriesOptronic Laboratories
An Intercomparison StudyAn Intercomparison StudyAn Intercomparison Study
cd/m2
fL N a N b u' v' int time
51. 14. 1.46E- 6 1.11E- . .5398 0.45
51.31 14.98 1.46E-06 1.09E-07 0.3999 0.5399 0.45
51.31 14.98 1.46E-06 1.10E-07 0.3999 0.5399 0.45
51.35 14.99 1.46E-06 1.08E-07 0.3999 0.5398 0.4551.32 14.98 1.46E-06 1.07E-07 0.3999 0.5398 0.45
17.00 4.96 1.44E-06 1.08E-07 0.3994 0.5399 1.32
17.00 4.96 1.44E-06 1.10E-07 0.3993 0.5399 1.32
17.01 4.96 1.44E-06 1.09E-07 0.3993 0.5399 1.32
17.01 4.96 1.44E-06 1.05E-07 0.3994 0.5400 1.32
17.01 4.96 1.44E-06 1.08E-07 0.3993 0.5399 1.32
0.33 0.10 1.44E-06 1.07E-07 0.3992 0.5399 68.9
0.33 0.10 1.44E-06 1.10E-07 0.3992 0.5399 68.9
0.33 0.10 1.44E-06 1.11E-07 0.3993 0.5399 68.9
0.33 0.10 1.44E-06 1.09E-07 0.3992 0.5399 68.9
0.33 0.10 1.44E-06 1.06E-07 0.3992 0.5399 68.9
NVIS e Display
-
8/6/2019 1224787448p28 Future of Nvis Display Msmt
22/30
The Future of NVIS Display MeasurementThe Future of NVIS Display Measurement
ALI Advanced Seminar, 2006ALI Advanced Seminar, 2006 Optronic LaboratoriesOptronic Laboratories
NVIS adiance for reen B display (scaled to 0.1 fL)
0.0E+00
1.0E-11
2.0E-11
3.0E-11
4.0E-11
5.0E-11
6.0E-11
7.0E-11
8.0E-11
9.0E-11
1.0E-10
1.1E-10
1.2E-10
1.3E-10
1.4E-10
1.5E-10
1.6E-10
1.7E-10
0 2 4 6 8 10 12 14 16
Luminance of Display [fL]
NVIS
adiance[W/(cm
sr)]
System 1
System 2System 3
= NRa = NRb
open = day 1 solid = day 2
Spec. Max.
NVIS Green B ResultsNVIS Green B ResultsNVIS Green B Results
For
-
8/6/2019 1224787448p28 Future of Nvis Display Msmt
23/30
The Future of NVIS Display MeasurementThe Future of NVIS Display Measurement
ALI Advanced Seminar, 2006ALI Advanced Seminar, 2006 Optronic LaboratoriesOptronic Laboratories
NVIS adiance for reen B display (scaled to 0.1 fL)
0.0E+00
1.0E-11
2.0E-11
3.0E-11
4.0E-11
5.0E-11
6.0E-11
7.0E-11
8.0E-11
9.0E-11
1.0E-10
1.1E-10
1.2E-10
1.3E-10
1.4E-10
1.5E-10
1.6E-10
1.7E-10
0 2 4 6 8 10 12 14 16
Luminance of Display [fL]
NVIS
adiance[W/(cm
sr)]
System 1
System 2System 3
= NRa = NRb
open = day 1 solid = day 2
Spec. Max.
Scanning monochromator
performance - 0.007" spotsize
NVIS Green B ResultsNVIS Green B ResultsNVIS Green B Results
For 2For 2
minuteminute
scanscan
For 1For 1--22
minuteminute
scanscan
For
-
8/6/2019 1224787448p28 Future of Nvis Display Msmt
24/30
The Future of NVIS Display MeasurementThe Future of NVIS Display Measurement
ALI Advanced Seminar, 2006ALI Advanced Seminar, 2006 Optronic LaboratoriesOptronic Laboratories
NVIS Green B ResultsNVIS Green B ResultsNVIS Green B ResultsCIE 1976 u v Chromaticity Diagram
0
0.1
0.
0.3
0.4
0.5
0.6
0.7
0 0.1 0.
0.3 0.4 0.5 0.6 0.7
u'
v'
Monochromatic loc
us
Green
Yellow
Red
System1
System
System 3
0.5774
0.5775
0.5776
0.5777
0.5778
0.5779
0.1305 0.1306 0.1307 0.1308
-
8/6/2019 1224787448p28 Future of Nvis Display Msmt
25/30
-
8/6/2019 1224787448p28 Future of Nvis Display Msmt
26/30
-
8/6/2019 1224787448p28 Future of Nvis Display Msmt
27/30
The Future of NVIS Display MeasurementThe Future of NVIS Display Measurement
ALI Advanced Seminar, 2006ALI Advanced Seminar, 2006 Optronic LaboratoriesOptronic Laboratories
NVIS Yellow ResultsNVIS Yellow ResultsNVIS Yellow ResultsNVIS Radiance for Yellow display (scaled to 15 fL)
0.0E+00
2.0E-08
4.0E-08
6.0E-08
8.0E-08
1.0E-07
1.2E-07
1.4E-07
0 2 4 6 8 10 12 14 16 18
Luminanceof isplay [fL]
NRbRadiance[W/(cmsr)]
System1
System2
System3
Spec. Max.
Spec. Min.
-
8/6/2019 1224787448p28 Future of Nvis Display Msmt
28/30
The Future of NVIS Display MeasurementThe Future of NVIS Display Measurement
ALI Advanced Seminar, 2006ALI Advanced Seminar, 2006 Optronic LaboratoriesOptronic Laboratories
NVIS Yellow ResultsNVIS Yellow ResultsNVIS Yellow ResultsCIE 1976 u'v' Chromaticity Diagram
0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0 0.1 0.2 0.3 0.4 0.5 0.6 0.7
u'
v'
Monochromatic loc
us
Green
Yellow
Red
System1
System
System 3
0.5600
0.5601
0.5602
0.5603
0.2618 0.2620 0.2622 0.2624 0.2626
-
8/6/2019 1224787448p28 Future of Nvis Display Msmt
29/30
The Future of NVIS Display MeasurementThe Future of NVIS Display Measurement
ALI Advanced Seminar, 2006ALI Advanced Seminar, 2006 Optronic LaboratoriesOptronic Laboratories
Study ConclusionsStudy ConclusionsStudy Conclusions
All systems gaveextremely good repeatability in
NRa,NRb and chromaticity.
All systems gavethe same results with very low
uncertainties. All systems showed thatNRa and NRb were
essentially constant across a wide luminance range
foreach display.
All systems showed chromaticity was essentially
constant with luminance foreach display.
-