1 - clinical and vaccine immunology · 2011-07-13 · 115 69,107 and registered at clintrial.gov...
TRANSCRIPT
1
Performance of a Rapid and Simple HIV Testing Algorithm in a 1
Multicenter Phase III Microbicide Clinical Trial 2
3
4
5
Tania Crucitti 1 *, Doug Taylor 2, Greet Beelaert 1, Katrien Fransen 1, Lut 6
Van Damme 2, 3 7
8
9
1 Institute of Tropical Medicine, Antwerp, Belgium 10
2 FHI, NC/VA, USA 11
3 CONRAD, Arlington, VA, USA 12
13
* Corresponding author: 14
Tania Crucitti, HIV/STI Reference Laboratory, Institute of Tropical Medicine, 155 15
Nationalestraat, 2000 Antwerp, Belgium, e-mail: [email protected], tel + 32 3 247 65 16
52, fax + 32 3 247 63 33 17
18
Copyright © 2011, American Society for Microbiology and/or the Listed Authors/Institutions. All Rights Reserved.Clin. Vaccine Immunol. doi:10.1128/CVI.05069-11 CVI Accepts, published online ahead of print on 13 July 2011
on February 13, 2021 by guest
http://cvi.asm.org/
Dow
nloaded from
2
ABSTRACT 19
A multi-test sequential algorithm based on Rapid/Simple (R/S) assays was 20
applied for the diagnosis of HIV infection among participants in a phase 3 21
microbicide effectiveness trial. HIV testing was performed on finger-prick 22
blood samples from patients after their enrollment in the trial. The specimens 23
were tested in a serial procedure using three different rapid tests (Determine 24
HIV-1/2 (Abbott), SD Bioline HIV 1/2 3.0 (Standard Diagnostics), and Uni-Gold 25
HIV (Trinity Biotech)). In the event of discordant results between the 26
Determine HIV 1/2 and SD Bioline HIV 1/2 3.0 tests, a third assay (i.e., Uni-27
Gold HIV) determined the final outcome. When the final outcome was positive, 28
a second specimen was collected and tested with the same algorithm, only if 29
a positive result was obtained with this sample the participant was informed of 30
her positive serostatus. A total of 5734 post-enrollment specimens obtained 31
from 1398 women were tested. Forty-six women tested positive according to 32
the testing algorithm performed on the first collected specimen. Confirmatory 33
testing results obtained at the ITM confirmed that 42 women were truly 34
infected. Two out of four initial false positives tested negative upon testing of a 35
second blood specimen. The other two tested false positive twice using 36
specimens collected on the same day. A high percentage of specimens 37
reactive with the Determine HIV-1/2 assay only were observed in the study 38
site in Kampala. This result did not appear to be associated with pregnancy or 39
malaria infection. 40
HIV testing algorithms including only R/S assays are suitable for use in clinical 41
trials, provided that adequate quality assurance procedures are in place. 42
on February 13, 2021 by guest
http://cvi.asm.org/
Dow
nloaded from
3
INTRODUCTION 43
Phase 2b/3 trials assessing the effectiveness of products to prevent the 44
heterosexual acquisition of HIV are almost exclusively conducted in African 45
and Asian countries with HIV epidemics and high incidence of infection. Even 46
in such settings, effectiveness trials must be conducted in multiple centers to 47
observe the large number of HIV infection events required to achieve 48
adequate statistical power to detect a treatment effect. 49
50
Study clinics may be located in remote areas, small towns or villages, where 51
well-equipped laboratories are lacking. There is often access only to a small 52
on-site laboratory where rapid and simple (R/S) assays to detect HIV 53
antibodies are performed. Testing using enzyme-linked immunosorbent 54
assays is preferable when large numbers of specimens are to be tested; 55
however, equipment, storage of reagents in refrigerated conditions, and 56
extensive training of laboratory staff is required for these assays. In contrast, 57
R/S assays require little to no equipment and have no storage restrictions (if 58
ambient temperatures are below 27°C). In addition, the use of R/S assays 59
(compared to traditional multiwell immunoassays) has the advantage that 60
results can be provided the day samples are collected; greatly reducing the 61
likelihood that individuals will fail to receive their test results. As with multiwell 62
immunoassays, there exists a time interval between infection and detectable 63
antibody levels (i.e., a window period, often measured in weeks), and false 64
positive results can never be excluded (4). Nonetheless, once patients are out 65
on February 13, 2021 by guest
http://cvi.asm.org/
Dow
nloaded from
4
of the window period, the sensitivity and specificity of the R/S HIV assays are 66
high and comparable to multiwell immunoassays (2, 24, 25, 26). 67
The HIV testing algorithms described in the literature often include the 68
confirmation of reactive rapid or immunoassay tests by supplemental testing 69
such as western blot (WB) (3, 4, 19). Performing WB assays requires 70
additional training of the laboratory staff and specific equipment and is 71
expensive. The UNAIDS/WHO program on HIV/AIDS testing proposes three 72
testing strategies using R/S assays without the requirement for additional 73
confirmatory testing, making them affordable and suitable for application in 74
small laboratories (2, 7, 13, 14). To maximize specificity and positive 75
predictive value, multi-test sequential algorithms have been implemented. 76
However, the sensitivity of any such algorithm can only be as high as that of 77
the first test in the sequence (1). Here, we present the results obtained with a 78
sequential HIV multi-test algorithm based on R/S assays used for the 79
diagnosis of HIV infection among participants in a phase 3 microbicide 80
effectiveness trial. 81
82
MATERIALS AND METHODS 83
Study design 84
The study in which the R/S algorithm was employed has been described 85
elsewhere (23). Briefly, the study was a randomized double blind placebo-86
controlled trial of a candidate vaginal microbicide (cellulose sulfate gel) for HIV 87
prevention. Women were recruited at five sites: a community clinic and a clinic 88
on February 13, 2021 by guest
http://cvi.asm.org/
Dow
nloaded from
5
for sexually transmitted infections in Cotonou, Benin; the Y.R. Gaitonde 89
Center for AIDS Research and Education (Y.R.G. CARE) in Chennai, India; 90
the Medical Research Council in Durban, South Africa; the Mulago Hospital 91
(Makarere University) in Kampala, Uganda; and clinics in Mudhol and 92
Jhamkandi in Karnataka, India (in collaboration with the Karnataka Health 93
Promotion Trust, Bangalore). Eligibility criteria included negative HIV antibody 94
tests at screening and enrollment and agreement to visit the clinic for HIV 95
testing 1, 3, 6, 9, and 12 months after enrollment. 96
97
HIV testing was performed on serum at screening; on plasma at enrollment, 98
the final visit and at any visit at which product use was discontinued; and on 99
whole blood obtained by finger prick at the follow-up visits at 1, 3, 6, and 9 100
months. The sites applied their national or local HIV testing algorithm for the 101
determination of HIV infection at screening and enrollment, but the sequential 102
R/S study algorithm (Figure 1) was applied on all other visits. This paper 103
presents the results and performance of the HIV algorithm used during follow-104
up and at the final visit and does not include the evaluation of the national or 105
local testing algorithm used at screening and enrollment. 106
107
Ethical approval 108
The study was approved by the Institutional Review Board of the Eastern 109
Virginia Medical School and of the Institute of Tropical Medicine and by local 110
ethics committees at each site where women were recruited. All approvals 111
on February 13, 2021 by guest
http://cvi.asm.org/
Dow
nloaded from
6
were granted prior to study initiation. Participants gave written informed 112
consent before screening and enrollment. The trial was conducted under the 113
Food and Drug Administration's Investigational New Drug application number 114
69,107 and registered at clintrial.gov (study number NCT00153777). 115
116
Specimens 117
Whole blood was collected in an EDTA microtainer (Becton Dickinson, 118
Sparks, USA) by finger prick using the Glucolet 2 device (Bayer HealthCare 119
LLC, Mishawaka, USA) at 1, 3, 6, and 9 month follow-up visits. The collected 120
volumes ranged between 250 and 500 microliters. At the final visit, permanent 121
product withdrawal visit or in the event the first finger-prick blood sample was 122
reactive for HIV antibodies, 7 ml of blood was collected in an EDTA blood 123
collection tube and processed within four hours. After centrifugation at 800 x g 124
(± 2000-2200 rpm) at 20°C ± 5°C for 10 minutes, the plasma was transferred 125
to a 15 ml conical tube under sterile conditions. The tube was centrifuged for 126
30 minutes at 1200 x g (± 2200 - 2400 rpm) at 20°C ± 5°C. After HIV R/S 127
testing on site, plasma aliquots were stored at -70°C until shipment on dry ice 128
to the Institute of Tropical Medicine (ITM), Antwerp, Belgium, for additional 129
testing. 130
131
HIV Assays 132
on February 13, 2021 by guest
http://cvi.asm.org/
Dow
nloaded from
7
Post-enrollment specimens were tested according to a sequential HIV R/S 133
algorithm based on the modified WHO/UNAIDS testing approach for strategy 134
III, which was also included as Point Of Contact algorithm 4 in the status 135
report on HIV testing algorithms published by the Association of Public Health 136
Laboratories (1, 21). Specimens were first tested with the Determine HIV-1/2 137
test (Abbott Laboratories, United Kingdom). If a sample was reactive, it was 138
tested with the SD Bioline HIV-1/2 3.0 test (Standard Diagnostics, South 139
Korea). In the event of a discordant result, the Uni-Gold HIV test (Trinity 140
Biotech, Ireland) was used. The final result of the algorithm scored a sample 141
as positive if two out of the three R/S HIV assays were reactive. In this case, a 142
second specimen was collected from the participant, preferably on the same 143
day, and tested with the same algorithm. The participant was informed of her 144
positive serostatus only if a positive result with the second specimen was 145
obtained. 146
147
The ITM tested additional specimens to exclude the possibility that 148
participants entered or exited the study during their window period. To this 149
end, all enrollment plasma specimens from participants with detectable HIV 150
antibodies within 3 months after enrollment and all plasma specimens 151
collected at the final visits of patients without detectable HIV antibodies were 152
tested using a qualitative HIV-1 RNA PCR assay (COBAS Ampliscreen HIV-1 153
test, Roche Molecular Systems, USA). Due to the large number of specimens, 154
pools of a maximum of 24 final visit plasma samples were tested according to 155
the manufacturer’s specifications. 156
on February 13, 2021 by guest
http://cvi.asm.org/
Dow
nloaded from
8
157
In the context of the quality control established for the study, aliquots of stored 158
plasma specimens from participants who tested HIV positive after enrollment 159
were also tested at the ITM to confirm these results using a different testing 160
algorithm, which is presented in Fig 2. 161
The details regarding guaranteeing the quality of the HIV results obtained at 162
the different study sites have been reported previously (5). In brief, this quality 163
control assessment consisted of the assessment of the study site laboratories, 164
provision of hands-on training, conduction of supervisory visits, distribution of 165
quality and batch control panels, and re-testing of pre-defined specimens. 166
Pregnancy and Malaria testing 167
Participants were tested for pregnancy at all visits using a urine HCG rapid 168
test. When malaria was suspected, a thick blood smear film was examined for 169
the microscopic visualization of Plasmodium parasites. 170
171
Statistical analysis 172
Overall, 38.8% of the potential participants were HIV positive at screening and 173
thus excluded from study participation (Table 1). Here, we summarize the 174
results of the sequential R/S HIV algorithm as applied to 5734 post-enrollment 175
specimens of 1398 enrolled women who had at least one follow-up HIV test 176
specimen as well as confirmatory specimens collected following an initial 177
positive result (specimens collected after a confirmed positive HIV result and 178
on February 13, 2021 by guest
http://cvi.asm.org/
Dow
nloaded from
9
specimens contributed by three women later found to have been in their 179
window period at enrollment are excluded). Descriptive tests of associations 180
between participant characteristics and discordant rapid test results were 181
conducted using logistic regression with generalized estimating equations to 182
account for repeated measures on each enrolled participant. 183
Final confirmation of positive results was performed at the ITM according to 184
the testing algorithm presented in Fig. 2 and/or using the HIV-1 RNA PCR 185
assay. Negative results were confirmed by testing all plasma samples 186
collected at final visits with the HIV-1 RNA PCR assay. 187
188
RESULTS 189
A total of 46 first collected specimens tested positive according to the 190
sequential R/S testing algorithm at the sites (Table 2), with all but one 191
specimen reactive with both Determine HIV-1/2 and SD Bioline HIV-1/2 3.0 192
(Figure 3). Confirmatory results were available for all 46 women, with 42 193
(91.3%) confirmed HIV-1 positive and 4 (8.7%) confirmed HIV negative (one 194
of the 42 confirmed positives became infected after the administrative 195
censoring date of the trial and was excluded from the primary effectiveness 196
analysis reported in (23)). HIV-2 infection was not detected in the post-197
enrollment specimens. According to the confirmed results of a total of 5734 198
specimens tested, the sequential R/S HIV testing algorithm showed an initial 199
sensitivity of 100% (95% CI: 89.6%-100%) and an initial specificity of 99.9% 200
(95% CI: 99.8%-100%). One of the four women with negative confirmatory 201
on February 13, 2021 by guest
http://cvi.asm.org/
Dow
nloaded from
10
results was subsequently lost to follow-up; two of the four had HIV negative 202
results at three or more subsequent visits; and one had a positive HIV result 203
on her first but a negative result on her second collected specimen, followed 204
by negative results at her last two visits. The initial positive results for these 205
four participants were considered false positive. 206
For two of the participants, including the participant with an unconfirmed 207
positive result at her next visit, the final result was false positive. We cannot 208
exclude that the false positive results were partially caused by the use of 209
deteriorated Determine HIV-1/2 test strips. Indeed, we observed that the 210
Determine HIV-1/2 test strips were not stored properly and were exposed to 211
humidity at this test site. In addition, the SD Bioline HIV-1/2 3.0 assay had a 212
false reaction leading to a final false positive result. 213
There were 124 specimens from 98 women with a reactive result in the 214
Determine HIV-1/2 test only (i.e., both SD Bioline HIV-1/2 3.0 and Uni-Gold 215
HIV were not reactive; Table 3). Five of the 98 women were not tested again, 216
2 were HIV antibody positive according to the R/S algorithm at their next study 217
visit and 91 were subsequently tested but never diagnosed with HIV. Of the 218
91, 22 (24.1%) were reactive only in the Determine HIV-1/2 test on multiple 219
occasions (Table 4). 220
221
The proportion of specimens that were only reactive in the Determine HIV-1/2 222
assay was significantly higher in Uganda (7.2%) than in the study as a whole 223
(2.2%; p-value<0.001). Given this observation, we sought to explore the 224
on February 13, 2021 by guest
http://cvi.asm.org/
Dow
nloaded from
11
possible influences of pregnancy and malaria on the performance of the 225
Determine HIV-1/2 assay. Pregnancy and malaria were both highly prevalent 226
in Uganda, and there is evidence in the literature to suggest that these 227
conditions may impact the specificity of the Determine HIV-1/2 assay (9, 10, 228
11, 15,18, 28). However, we found no significant difference in the proportion 229
of specimens in which only the Determine HIV-1/2 test was reactive between 230
pregnant and non-pregnant women (3.6% reactive among women testing 231
positive for pregnancy, 2.1% for women not testing positive for pregnancy; p-232
value=0.24). Malaria was only diagnosed in Benin and Uganda. Restricting 233
analysis to those two sites, we likewise found no significant difference in the 234
proportion of specimens in which only the Determine HIV-1/2 test was reactive 235
according to malaria status (4.7%, regardless of previous malaria diagnosis; 236
p-value=0.61). 237
238
Having a reactive result only in the Determine HIV-1/2 assay was significantly 239
and positively associated with the probability of obtaining a reactive result only 240
on the Determine HIV-1/2 test at a subsequent visit (p-value=0.014), but this 241
result is confounded by the effect of site (20 of the 22 women with repeat 242
samples only reactive in the Determine HIV-1/2 test were from Uganda). 243
Given these results, we are unable to explain the high prevalence of 244
specimens reactive only in the Determine HIV-1/2 test in Uganda. 245
246
on February 13, 2021 by guest
http://cvi.asm.org/
Dow
nloaded from
12
Using the qualitative HIV-1 RNA PCR assay on stored plasma specimens, 247
three participants were found to be HIV infected at enrollment. These three 248
participants were from Uganda, Benin, and South Africa; sites which used a 249
different testing algorithm at enrollment than at follow-up and hence do not 250
provide information on the utility of the R/S algorithm presented here. 251
Another participant was found to be recently infected at her final visit. 252
253
DISCUSSION 254
HIV testing plays a crucial role in HIV prevention clinical trials as it determines 255
the eligibility of potential participants and is used to identify the primary study 256
endpoints. 257
258
The use of R/S HIV tests provides the opportunity for small or field-based 259
laboratories to deliver HIV results on the same day as the study visit. 260
Moreover, R/S tests do not require refrigeration as long as temperatures are 261
below 27°C, nor do they require special equipment, processing of specimens 262
in batches, or highly skilled laboratory staff. 263
264
In general, multiwell immunoassays are considered to be superior to R/S tests 265
for the detection of HIV antibodies (16). However, all but one infection taking 266
place during this study were detected using the sequential R/S HIV algorithm. 267
The single missed infection was detected at the patient’s final visit using a 268
on February 13, 2021 by guest
http://cvi.asm.org/
Dow
nloaded from
13
PCR RNA test and was most likely in the window period. This recent infection 269
would thus probably not have been detected using a third generation 270
immunoassay either. 271
272
The advantage of most R/S assays is that they can be performed on finger-273
prick blood, in contrast to multiwell immunoassays, for which venous blood 274
has to be collected. We chose to collect 250 - 500 µl of whole blood from 275
finger pricks in an EDTA microtainer because the procedure is less invasive 276
and often more acceptable to participants and provides sufficient volume to 277
perform the 3 R/S assays included in the testing algorithm. 278
279
Testing of a second specimen following an initial positive HIV algorithm result 280
was done to exclude transcription errors and to control for possible specimen 281
mix-ups (20). This procedure is routine practice in Belgium but not in other 282
countries. The first specimen tested was blood obtained by finger prick, which 283
in some instances may influence the reliability of the visual reading of the R/S 284
test (12). Among the four participants with HIV false positive results, one 285
participant had substances in her blood which cross-reacted at all testing 286
occasions with the SD Bioline HIV-1/2 3.0 assay. For this participant only, the 287
sequence of the rapid assays was changed to the Determine HIV-1/2 test first 288
and in the event of a reactive result, the Unigold HIV test was performed 289
instead of the SD Bioline HIV-1/2. 290
291
on February 13, 2021 by guest
http://cvi.asm.org/
Dow
nloaded from
14
We found 124/5734 (2.2%) specimens that were only reactive in the 292
Determine HIV-1/2 test, with a much higher rate (7.2%) in Uganda than in any 293
other site. The lack of specificity of the Determine HIV-1/2 assay with 294
specimens from Uganda has been reported previously (6, 12, 22). We 295
hypothesize that in Kampala, Uganda, in particular, participants had 296
undiagnosed endemic non-HIV infections or circulating substances in their 297
blood that may have cross-reacted with the Determine HIV-1/2 test (12). 298
Cross-reactivity of HIV assays is not uncommon; a recent publication reported 299
cross-reactivity of the Murex HIV Ag/Ab combination EIA with S. haematobium 300
IgG and a study from our group showed cross-reactivity of the Determine HIV-301
1/2 test with Trypanosoma brucei gambiense (8, 17). The 124 results came 302
from 98 women, and only two were ultimately diagnosed with HIV during the 303
trial. This counters the impression expressed by laboratory staff in the field 304
that the Determine HIV-1/2 test is more sensitive at detecting recent 305
infections. 306
It should be noted that the Determine HIV 1/2 rapid test was initially 307
developed by Abbott; however, in May 2005 Abbott entered an agreement to 308
sell the rapid test to Inverness Medical Innovation. Currently, the rapid test is 309
manufactured by Alere Medical in Japan. Per the new manufacturer, the 310
performance characteristics of the assay has not changed with the change in 311
manufacturer. 312
However, we note that some of the false reactive results obtained with the 313
Determine HIV-1/2 assay could be attributed to the deterioration of the test 314
strips that was observed at the start of the study and the over-interpretation of 315
on February 13, 2021 by guest
http://cvi.asm.org/
Dow
nloaded from
15
reddish ‘shadows’ or ‘ghost lines’ especially when finger-prick blood was 316
tested (15). Re-training on the storage and handling of the strips and test 317
procedure, including the interpretation of ‘reactivity,’ was provided (5). It 318
should be noted that according to the guidelines of the World Health 319
Organization, the discordance of test results between first and second or 320
tiebreaker tests should not exceed 5% (27). Although the number of false 321
reactive Determine HIV-1/2 assays obtained in Kampala decreased following 322
training and thorough supervision, it remained approximately 5%. This 323
observation reinforces our hypothesis that interfering blood substances played 324
a role in the observed false reactive Determine HIV-1/2 results. In addition, 325
this finding emphasizes the need for and importance of assay and algorithm 326
validation in the populations and geographical areas in which assays will be 327
used (7, 12). 328
Although rapid assays are simple, training in the test procedure and reading of 329
the test devices is still necessary (5, 12). Likewise, quality control and 330
assurance procedures should be put into place for any clinical trial. First, the 331
quality of the test devices should be assessed when received by the 332
laboratory. Once the test devices pass batch-entry control, internal controls 333
should be run at regular intervals. The staff administering the tests should also 334
be assessed at a pre-defined frequency by testing quality control panels (20). 335
Although the assays in this study do not require refrigeration, they cannot be 336
stored at temperatures higher than 27°C or 30°C, and the tests should not be 337
performed in conditions exceeding those temperatures. In addition, the test 338
devices are susceptible to humidity and need to be stored in sealed pouches 339
containing a desiccant. 340
on February 13, 2021 by guest
http://cvi.asm.org/
Dow
nloaded from
16
341
With the use of a qualitative HIV RNA PCR assay, we were able to more 342
accurately determine the time of infection and exclude three participants from 343
the analysis who were found to be HIV infected but still in their window period 344
at enrollment. We were also able to detect one early/recent infection in a 345
participant who did not develop HIV antibodies at her last study visit and who 346
was likely in her window period. Not only could this endpoint be included in 347
the data analysis and improve the precision of the estimated treatment effect, 348
the use of PCR on the final visit sample allowed the participant to be referred 349
for further HIV infection management. Excluding this recent infection detected 350
at a final visit, no additional infections other than those determined by the HIV 351
testing algorithm were detected using HIV RNA PCR. This observation 352
confirms the sensitivity of the HIV testing algorithm and its suitability in clinical 353
trials. 354
355
In conclusion, our results demonstrate that HIV algorithms using only rapid 356
and simple HIV assays with results confirmed on a second specimen provide 357
reliable HIV results and can easily be implemented in HIV prevention trials 358
conducted in resource-limited areas. If more accurate time point 359
determinations of HIV infection are required for outcome analysis, we 360
recommend using HIV RNA PCR in a look-back procedure on stored 361
specimens. 362
363
on February 13, 2021 by guest
http://cvi.asm.org/
Dow
nloaded from
17
Acknowledgements 364
The authors thank all study participants and study staff, especially the 365
laboratory staff, at the study sites for their dedication. 366
We thank Marjan Mangelschots, Mirriam Van Rooy, and Sergio Garcia for the 367
HIV testing and Wendy Thys for her data entry work. 368
This study was supported by grants from the United States Agency for 369
International Development and the Bill & Melinda Gates Foundation. 370
No potential conflict of interest relevant to this article was reported. 371
372
REFERENCES 373
1. Association of Public Health Laboratories and the Centres for Disease 374
Control & Prevention. HIV Testing Algorithms: A Status Report 2009. 375
Available at 376
http://www.aphl.org/aphlprograms/infectious/hiv/Pages/HIVStatusRepor377
t.aspx 378
2. Branson, B.M. 2000. Rapid tests for HIV antibody. Aids rev. 2:76-83. 379
3. Centers for Disease Control and Prevention. 2004. Notice to 380
Readers: Protocols for Confirmation of Reactive Rapid HIV tests 381
MMWR 53:221-222. 382
Available http://www.cdc.gov/mmwr/preview/mmwrhtml/mm5310a7.htm 383
on February 13, 2021 by guest
http://cvi.asm.org/
Dow
nloaded from
18
4. Chen, M.Y., and C.S. Estcourt. 2009. Time to roll out rapid testing for 384
HIV ? Yes, but with appropriate safeguards. Sexual Health 6:1-3. 385
5. Crucitti, T., K. Fransen, R. Maharaj, T. Tenywa, M. Massinga 386
Loembe, K.G. Murugavel, K. Mendoca, S. Abdellati, G. Beelaert, 387
and L. Vandamme. 2010.Obtaining valid laboratory data in clinical 388
trials conducted in resource diverse settings: Lessons learned from a 389
microbicide phase III clinical trial. PlosOne 5 (10) e13592. 390
6. Eller, L.A., M.A. Eller, B.J. Ouma, P. Kataaha, B.S. Bagaya, R.L. 391
Olemukan, S. Erima, L. Kawala, M.S. de Souza, H. Kibuuka, F. 392
Wabwire-Mangen, S.A. Peel, R.J. O’Connell, M.L. Robb, and N.L. 393
Michael. 2007. Large-scale human immunodeficiency virus rapid test 394
evaluation in a low-prevalence Ugandan blood bank population. J. Clin. 395
Microbiol. 45:3281-3285. 396
7. Everett, D.B., K. Baisley, J. Changalucha, A. Vallely, D. Watson-397
Jones, C. Cook, L. Knight, D.A. Ross, K. Mugeye, S. McCormack, 398
C.J. Lacey, U. Jentsch, and R.J. Hayes. 2009. Suitability of simple 399
Human Immunodeficiency Virus rapid tests in clinical trials in 400
community-based clinic settings. J. Clin. Microbiol. 47:1058-1062. 401
8. Everett , D.B., K. Baisley, R. McNerney, T. Chirwa, D.A. Ross, J. 402
Changalucha, D. Watson-Jones, H. Helmby, D.W. Dunne, D. 403
Mabey, and R.J. Hayes.2010. Association of schistosomiasis with 404
false positive HIV test results in an African adolescent population. J. 405
Clin. Microbiol. 48:1570-1577. 406
9. Fonseca, M.O, L. Pang, L. de Avilla Sdo, V.G. Arruk, T.R. Tozetto-407
Mendoza, A.W. Ferreira, A. Saes-Alquezar, and M. Boulos. 2000. 408
on February 13, 2021 by guest
http://cvi.asm.org/
Dow
nloaded from
19
Cross-reactivity of anti-Plasmodium falciparum antibodies and HIV 409
tests. Trans. R. Soc. Trop. Med. Hyg. 94:171-172. 410
10. Gasasira, A.F., G. Dorsey, M. Kamya, D. Havlir, M. Kiggundu, P.J. 411
Rosenthal, and E.D. Charlebois. 2006. False-positive results of 412
Enzyme Immunoassays for Human Immunodeficiency virus in patients 413
with uncomplicated malaria. J. Clin. Microbiol. 44:3021-3024. 414
11. Ghosh, K., K.N. Javeri, D. Mohanty, B.D. Parmar, R.R. Surati, and 415
S.H. Joshi. 2001. False-positive serological tests in acute malaria. Br. 416
J. Biomed. Sci. 58:20-23. 417
12. Gray, R., F. Makumbi, D. Serwadda, T. Lutalo, F. Sewankambo, and 418
M.J. Wawer. 2007. Limitations of rapid HIV-1 tests during screening for 419
trials in Uganda: diagnostic test accuracy study. BMJ 335:188-190. 420
13. Joint United Nations Programme on HIV/AIDS (UNAIDS). 1997. 421
Revised recommendations for the selection and use of HIV antibody 422
tests. Wkly Epidemiol. Rec. 72:81-87. 423
14. Joint United Nations Programme on HIV/AIDS (UNAIDS). 1998. The 424
importance of simple/rapid assays in HIV testing. Wkly Epidemiol. Rec. 425
73:321-328. 426
15. Klarkowski, D.B., J.M. Wazome, K.M. Lokuge, L. Shanks, C.F. 427
Mills, and D.P. O’Brien. 2009. The evaluation of a rapid In Situ HIV 428
confirmation test in a programme with a high failure rate of the WHO 429
HIV two-test diagnostic algorithm. PlosOne.4: 430
e4351.doic:10.1371/journal.pone.0004351 431
16. Laforgerie, E., B. Boucher, T. Dong Ly, L. Maisoneuve, J. Izopep, 432
C. Delaugerre, and F. Simon. 2010. Sensitivity of 8 CE (European 433
on February 13, 2021 by guest
http://cvi.asm.org/
Dow
nloaded from
20
Community) –approved rapid disposable tests for anti-HIV antibody 434
detection during and after seroconversion. J. Virol. Methods. 165:105-435
107. 436
17. Lejon, V., D.M. Ngoyi, M. Ibunga, G. Beelaert, I. Maes, P. Büsher, 437
and K. Fransen. 2010. Low specificities of HIV diagnostic tests 438
caused by Trypanosoma brucei gambiense sleeping sickness. J. Clin. 439
Microbiol. 48:2836-2839. 440
18. Magee, L.A., K.E. Murphy, and P. van Dadelszen. 1999. False-441
positive results in antenatal HIV screening. CMAJ 160:1285. 442
19. Padian, N.S., A. van der Straten, G. Ramjee, T. Chipato, G. de 443
Bruyn, S. Shiboki, E.T. Montgomery, H. Francher, H. Cheng, M. 444
Rosenblum, M. van der Laan, N. Jewell, J. McIntyre, and MIRA 445
team. 2007. Diaphragm and lubricant gel for prevention of HIV 446
acquisition in southern African women: a randomised controlled trial. 447
The Lancet 370:251-261. 448
20. Plebani M. 2009. Does POCT reduce the risk of error in laboratory 449
testing? Clin. Chim. Acta 404:59-64. 450
21. Respess, R.A., M.A. Rayfield, and T.J. Dondero. 2001. Laboratory 451
testing and rapid HIV assays: applications for HIV surveillance in hard-452
to-reach populations. AIDS. 15:S49-S59. 453
22. Singer D.E., N. Kiwanuka, D. Serwadda, F. Nalugoda, L. Hird, J. 454
Bulken-Hoover, G. Kigozi, J.A. Malia, E.K. Calero, W. Sateren, M.L. 455
Robb, F. Wabwire-Magen, M. Wawer, R.H. Gray, N. Sewankambo, 456
D.L. Birx, and N.L. Michael. 2005. Use of stored serum from Uganda 457
for development and evaluation of a human immunodeficiency virus 458
on February 13, 2021 by guest
http://cvi.asm.org/
Dow
nloaded from
21
type 1 testing algorithm involving multiple rapid immunoassays. J. Clin. 459
Microbiol. 43:5312-5315. 460
23. Van Damme, L., R. Govinden, F.M. Mirembe, F. Guedou , S. 461
Solomon, M.L. Becker, B.S. Pradeep, A.K. Krishnan, M. Alary, B. 462
Pande, G. Ramjee, J. Deese, T. Crucitti, D. Taylor, for the CS 463
group. 2008. Lack of effectiveness of cellulose sulfate gel for 464
prevention of vaginal HIV transmission. N. Engl. J. Med. 359:463-472. 465
24. World Health Organization. 2002. HIV simple/rapid assays: 466
operational characteristics, Report 12: whole blood specimens. World 467
Health Organization, Geneva, Switzerland. 468
http://whqlibdoc.who.int/publications/2002/9241592354.pdf 469
25. World Health Organization. 2004. HIV simple/rapid assays: 470
operational characteristics, Report 14. World Health Organization, 471
Geneva, Switzerland. 472
http://whqlibdoc.who.int/publications/2004/9241592168.pdf 473
26. World Health Organization. 2009. HIV antigen/antibody assays: 474
operational characteristics, Report 16: rapid assays. World Health 475
Organization, Geneva, Switzerland. 476
http://www.who.int/diagnostics_laboratory/publications/Report16_final.p477
df. 478
27. World Health Organization. 2004. Rapid HIV tests: guidelines for use 479
in HIV testing and counselling services in resource-constrained 480
settings. World Health Organization, Geneva, Switzerland. 481
28. Zacharias, N.M., I.D. Athanassaki, H. Sangi-Haghpeykar, and M.O. 482
Gardner. 2004. High false-positive rate of human immunodeficiency 483
on February 13, 2021 by guest
http://cvi.asm.org/
Dow
nloaded from
22
virus rapid serum screening in a predominantly Hispanic prenatal 484
population. J. Parasitology 24:743-747. 485
486
on February 13, 2021 by guest
http://cvi.asm.org/
Dow
nloaded from
23
Figures 487
488
Figure 1: HIV testing algorithm used in the trial during follow-up visits and at 489
final visits. 490
491
Figure 2: HIV testing algorithm used for re-testing and final confirmation of 492
HIV seroconversion at the Institute of Tropical Medicine, Belgium. 493
494
Legend to Figure 2 495
EIA 1: Enzygnost® Anti-HIV 1/2 Plus, Dade Behring GmbH, Duitsland; EIA 2: 496
Vironostika® HIV Uni-Form II plus O, Organon Teknika, France; LIA: INNO-497
LIATM HIV I/II Score, Innogenetics, Belgium; Ag: INNOTEST® HIV Antigen 498
mAb, Innogenetics, Belgium; IND: indeterminate ; +:positive; -: negative, 499
500
Figure 3: HIV test results on the first specimens collected 501
502
on February 13, 2021 by guest
http://cvi.asm.org/
Dow
nloaded from
24
TABLE 1 HIV prevalence rates in women who presented for screening at the 503
5 study sites 504
Site
Number of women
tested HIV positive (%)
Benin, Cotonou 464 129 (27.8)
South Africa, Durban 1432 723 (50.5)
Uganda, Kampala 571 185 (32.4)
India, Bangalore 72 36 (50.0)
India, Chennai 377 60 (15.9)
Pooled 2916 1133 (38.8)
505
506
507
508
509
510
on February 13, 2021 by guest
http://cvi.asm.org/
Dow
nloaded from
25
TABLE 2 Positive HIV rapid test results of first specimens collected at follow-511
up 512
Site
Number of first
specimens tested HIV Positive (%)
Benin, Cotonou 909 11 (1.2)
South Africa, Durban 2366 28 (1.2)
Uganda, Kampala 1506 7 (0.5)
India, Bangalore 49 0 (0.0)
India, Chennai 904 0 (0.0)
Pooled 5734 46 (0.8) a,b
513
a All but one of the 46 specimens were Determine reactive/SD Bioline reactive 514
(HIV-1); one specimen was Determine reactive/Unigold reactive 515
b 42 of the 46 were confirmed positive at the ITM 516
517
518
on February 13, 2021 by guest
http://cvi.asm.org/
Dow
nloaded from
26
TABLE 3 Specimens reactive only in the Determine HIV-1/2 assay 519
Site
Number of first
collected specimens Reactive (%)
Benin, Cotonou 909 5 (0.6)
South Africa, Durban 2366 9 (0.4)
Uganda, Kampala 1506 108 (7.2)
India, Bangalore 49 0 (0.0)
India, Chennai 904 2 (0.2)
Pooled 5734 124 (2.2) a
520
a contributed by 98 women 521
522
523
on February 13, 2021 by guest
http://cvi.asm.org/
Dow
nloaded from
27
TABLE 4 Women with reactive results only in the Determine HIV-1/2 assay 524
Women (% enrolled) with
reactive results only in
the Determine assay
Number of study visits
with reactive results
Site 1 2 3 4
Benin, Cotonou 5 (2.2) 5 0 0 0
South Africa, Durban 8 (1.5) 7 1 0 0
Uganda, Kampala 84 (27.7) 64 17 2 1
India, Bangalore 0 (0.0) 0 0 0 0
India, Chennai 1 (0.4) 0 1 0 0
Pooled 98 a (7.0) 76 19 2 1
525
a Five were not tested again; 2 tested positive for HIV according to the R/S 526
algorithm at their next study visit; and 91 had subsequent test visits but were 527
never diagnosed with HIV 528
529
530
531
on February 13, 2021 by guest
http://cvi.asm.org/
Dow
nloaded from