4822 872 00983October 2000, Rev.2, 2/01© 2000 Fluke Corporation. All rights reserved. Printed in the Netherlands.All product names are trademarks of their respective companies.

Fluke 192/196/199ScopeMeter

Users Manual

Page 2: 0MASTER - RS Components
Page 3: 0MASTER - RS Components

LIMITED WARRANTY & LIMITATION OF LIABILITY

Each Fluke product is warranted to be free from defects in material and workmanship under normal use and service. The warranty period is three yearsand begins on the date of shipment. Parts, product repairs and services are warranted for 90 days. This warranty extends only to the original buyer or end-usercustomer of a Fluke authorized reseller, and does not apply to fuses, disposable batteries or to any product which, in Fluke’s opinion, has been misused, altered,neglected or damaged by accident or abnormal conditions of operation or handling. Fluke warrants that software will operate substantially in accordance with itsfunctional specifications for 90 days and that it has been properly recorded on non-defective media. Fluke does not warrant that software will be error free oroperate without interruption.

Fluke authorized resellers shall extend this warranty on new and unused products to end-user customers only but have no authority to extend a greater ordifferent warranty on behalf of Fluke. Warranty support is available if product is purchased through a Fluke authorized sales outlet or Buyer has paid theapplicable international price. Fluke reserves the right to invoice Buyer for importation costs of repair/replacement parts when product purchased in one countryis submitted for repair in another country.

Fluke’s warranty obligation is limited, at Fluke’s option, to refund of the purchase price, free of charge repair, or replacement of a defective product which isreturned to a Fluke authorized service center within the warranty period.

To obtain warranty service, contact your nearest Fluke authorized service center or send the product, with a description of the difficulty, postage and insuranceprepaid (FOB Destination), to the nearest Fluke authorized service center. Fluke assumes no risk for damage in transit. Following warranty repair, the productwill be returned to Buyer, transportation prepaid (FOB Destination). If Fluke determines that the failure was caused by misuse, alteration, accident or abnormalcondition of operation or handling, Fluke will provide an estimate of repair costs and obtain authorization before commencing the work. Following repair, theproduct will be returned to the Buyer transportation prepaid and the Buyer will be billed for the repair and return transportation charges (FOB Shipping Point).

THIS WARRANTY IS BUYER’S SOLE AND EXCLUSIVE REMEDY AND IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUTNOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. FLUKE SHALL NOT BE LIABLE FORANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, INCLUDING LOSS OF DATA, WHETHER ARISING FROM BREACHOF WARRANTY OR BASED ON CONTRACT, TORT, RELIANCE OR ANY OTHER THEORY.

Since some countries or states do not allow limitation of the term of an implied warranty, or exclusion or limitation of incidental or consequential damages, thelimitations and exclusions of this warranty may not apply to every buyer. If any provision of this Warranty is held invalid or unenforceable by a court of competentjurisdiction, such holding will not affect the validity or enforceability of any other provision.

Fluke Corporation, P.O. Box 9090, Everett, WA 98206-9090 USA, or

Fluke Industrial B.V., P.O. Box 90, 7600 AB, Almelo, The Netherlands

Page 4: 0MASTER - RS Components

SERVICE CENTERS

To locate an authorized service center, visit us on the World Wide Web:

http://www.fluke.com

or call Fluke using any of the phone numbers listed below:

+1-888-993-5853 in U.S.A. and Canada

+31-40-2675200 in Europe

+1-425-356-5500 from other countries

Page 5: 0MASTER - RS Components

i

Table of Contents

Chapter Title PageUnpacking the Test Tool Kit ...................................................................................... 2Safety Information: Read First .................................................................................. 4

1 Using The Scope .................................................................................................... 7

Powering the Test Tool............................................................................................. 7Resetting the Test Tool ............................................................................................ 8Navigating a Menu ................................................................................................... 9Hiding Key Labels and Menus .................................................................................. 10Input Connections .................................................................................................... 10Making Scope Connections ...................................................................................... 11Displaying an Unknown Signal with Connect-and-View™ .......................................... 12Making Automatic Scope Measurements .................................................................. 13Freezing the Screen ................................................................................................. 14Using Average, Persistence and Glitch Capture........................................................ 15Acquiring Waveforms ............................................................................................... 18Analyzing Waveforms............................................................................................... 20

Page 6: 0MASTER - RS Components

Fluke 192/196/199Users Manual

ii

2 Using The Multimeter.............................................................................................. 21

Making Meter Connections........................................................................................ 21Making Multimeter Measurements............................................................................. 22Freezing the Readings .............................................................................................. 25Selecting Auto/Manual Ranges.................................................................................. 25Making Relative Measurements................................................................................. 26

3 Using The Recorder Functions............................................................................... 27

Opening the Recorder Main Menu............................................................................. 27Plotting Measurements Over Time (TrendPlot™)....................................................... 28Recording Scope Waveforms In Deep Memory (Scope Record™)............................. 31Analyzing a TrendPlot or Scope Record .................................................................... 34

4 Using Replay, Zoom and Cursors .......................................................................... 35

Replaying the 100 Most Recent Scope Screens ........................................................ 35Zooming in on a Waveform ....................................................................................... 38Making Cursor Measurements................................................................................... 40

5 Triggering on Waveforms....................................................................................... 43

Setting Trigger Level and Slope................................................................................. 44Using Trigger Delay or Pre-trigger ............................................................................. 45Automatic Trigger Options......................................................................................... 46Triggering on Edges.................................................................................................. 47Triggering on External Waveforms ............................................................................ 50Triggering on Video Signals....................................................................................... 51Triggering on Pulses ................................................................................................. 53

Page 7: 0MASTER - RS Components

Contents (continued)

iii

6 Using Memory, PC and Printer .............................................................................. 57

Saving and Recalling ................................................................................................ 57Documenting Screens .............................................................................................. 61

7 Tips ......................................................................................................................... 65

Using the Standard Accessories ............................................................................... 65Using the Independently Floating Isolated Inputs ...................................................... 68Using the Tilt Stand .................................................................................................. 70Resetting the Test Tool ............................................................................................ 70Suppressing Key Labels and Menu’s ........................................................................ 70Changing the Information Language ......................................................................... 71Adjusting the Contrast and Brightness ...................................................................... 71Changing Date and Time.......................................................................................... 72Saving Battery Life ................................................................................................... 73Changing the Auto Set Options................................................................................. 74

8 Maintaining the Test Tool ...................................................................................... 75

Cleaning the Test Tool.............................................................................................. 75Storing the Test Tool ................................................................................................ 75Charging the Batteries.............................................................................................. 76Extending Battery Operation Time ............................................................................ 77Replacing the NiMH Battery Pack BP190.................................................................. 78Calibrating the Voltage Probes ................................................................................. 78Displaying Calibration Information............................................................................. 80Parts and Accessories.............................................................................................. 80Troubleshooting........................................................................................................ 85

Page 8: 0MASTER - RS Components

Fluke 192/196/199Users Manual

iv

9 Specifications ......................................................................................................... 87

Introduction............................................................................................................... 87Dual Input Oscilloscope............................................................................................. 88Automatic Scope Measurements............................................................................... 90Meter ........................................................................................................................ 94DMM Measurements on Meter Inputs........................................................................ 94Recorder................................................................................................................... 96Zoom, Replay and Cursors........................................................................................ 97Miscellaneous ........................................................................................................... 97Environmental........................................................................................................... 99Safety ....................................................................................................................... 10010:1 Probe ................................................................................................................ 101Electromagnetic Immunity ......................................................................................... 103

Index

Page 9: 0MASTER - RS Components

Unpacking the Test Tool Kit

1

Declaration of Conformity

for

Fluke 192/196/199

ScopeMeter® test tools

Manufacturer

Fluke Industrial B.V.Lelyweg 1

7602 EA AlmeloThe Netherlands

Statement of Conformity

Based on test results using appropriate standards,the product is in conformity with

Electromagnetic Compatibility Directive 89/336/EECLow Voltage Directive 73/23/EEC

Sample tests

Standards used:

EN 61010.1 (1993)Safety Requirements for Electrical Equipment for

Measurement, Control, and Laboratory Use

EN-IEC61326-1 (1997)Electrical equipment for

measurements and laboratoryuse -EMC requirements-

The tests have been performed in atypical configuration.

This Conformity is indicated by the symbol ,i.e. “Conformité Européenne”.

Page 10: 0MASTER - RS Components

Fluke 192/196/199Users Manual

2

Unpacking the Test Tool KitThe following items are included in your test tool kit:

Note

When new, the rechargeable NiMH battery is notfully charged. See Chapter 8.

FL UKE S COPEMETER PR ODU CT

R EGIS T RATION CARD

4

3

7

1 2

911

12

10

13

5

6

(2x)

(2x)(2x)

(2x)

(2x)

(2x)

(2x)

8

(2x)(2x)

e

e

(1x)f

(2x)b

b

a

a

c

c

(2x)d

d

Figure 1. ScopeMeter Test Tool Kit

Page 11: 0MASTER - RS Components

Unpacking the Test Tool Kit

3

# Description

1 ScopeMeter Test Tool

2 Battery Charger (country dependent)

3 10:1 Voltage Probe Set (red)a) 10:1 Voltage Probe (red)b) Hook Clip for Probe Tip (red)c) Ground Lead with Hook Clip (red)d) Ground Lead with Mini Alligator Clip (black)e) 4-mm Test Probe for Probe Tip (red)f) Ground Spring for Probe Tip (black)

4 10:1 Voltage Probe Set (gray)a) 10:1 Voltage Probe (gray)b) Hook Clip for Probe Tip (gray)c) Ground Lead with Hook Clip (gray)d) Ground Lead with Mini Alligator Clip (black)e) 4-mm Test Probe for Probe Tip (gray)

5 Test Leads (red and black)

6 Accessory Set (only for Fluke 196 and 199)a) 2-mm Test Probe for Probe Tip (red)b) Industrial Alligator for Probe Tip (red)c) 2-mm Test Probe for Banana Jack (red)d) Industrial Alligator for Banana Jack (red)e) Ground Lead with 4-mm Banana Jack

(black)

# Description

7 Accessory Set (only for Fluke 196 and 199)a) 2-mm Test Probe for Probe Tip (gray)b) Industrial Alligator for Probe Tip (gray)c) 2-mm Test Probe for Banana Jack (gray)d) Industrial Alligator for Banana Jack (gray)e) Ground Lead with 4-mm Banana Jack

(black)

8 Users Manual (this book)

9 Product Registration Card with Envelope

10 Shipment box (basic version only)

Fluke 192-S, 196-S and 199-S versions include also thefollowing items:

# Description

11 Optically Isolated RS-232 Adapter/Cable

12 FlukeView® ScopeMeter® Software forWindows®

13 Hard Case

Page 12: 0MASTER - RS Components

Fluke 192/196/199Users Manual

4

Safety Information: Read FirstCarefully read the following safety information beforeusing the test tool.

Specific warning and caution statements, where theyapply, appear throughout the manual.

A “Warning” identifies conditions andactions that pose hazard(s) to the user.

A “Caution” identifies conditions and actionsthat may damage the test tool.

The following international symbols are used on the testtool and in this manual:

See explanation inmanual

Double Insulation(Protection Class)

Disposal information Earth

Recyclinginformation

ConformitéEuropéenne

Safety Approval Safety Approval

WarningTo avoid electrical shock or fire, use only theFluke power supply, Model BC190 (BatteryCharger / Power Adapter).

Page 13: 0MASTER - RS Components

Safety Information: Read First

5

WarningTo avoid electrical shock or fire if a test toolinput is connected to more than 42 V peak (30Vrms) or on circuits of more than 4800 VA:

• Use only insulated voltage probes and testleads (and adapters) supplied with the testtool, or equivalents as specified in theaccessory list, see Chapter 8.

• Before use, inspect voltage probes, testleads and accessories for mechanicaldamage and replace when damaged.

• Remove all probes, test leads andaccessories that are not in use.

• Always connect the battery charger first tothe ac outlet before connecting it to the testtool.

• Do not connect the ground spring tovoltages higher than 42 V peak (30 Vrms)from earth ground.

• Do not apply voltages that differ more than600 V from earth ground to any input whenmeasuring in a CAT III environment.Do not apply voltages that differ more than1000 V from earth ground to any input whenmeasuring in a CAT II environment.

• Do not apply voltages that differ more than600 V from each other to the isolated inputswhen measuring in a CAT III environment.Do not apply voltages that differ more than1000 V from each other to the isolated inputswhen measuring in a CAT II environment.

• Do not apply input voltages above the ratingof the instrument. Use caution when using1:1 test leads because the probe tip voltagewill be directly transmitted to the test tool.

• Do not use exposed metal BNC or bananaplug connectors.

• Do not insert metal objects into connectors.

• Always use the test tool only in the mannerspecified.

Voltage ratings that are mentioned in the warnings, aregiven as limits for “working voltage”. They representV ac rms (50-60 Hz) for ac sinewave applications and asV dc for dc applications.

Overvoltage Category III refers to distribution level andfixed installation circuits inside a building.Overvoltage Category II refers to local level, which isapplicable for appliances and portable equipment.

Page 14: 0MASTER - RS Components

Fluke 192/196/199Users Manual

6

The terms ‘Isolated’ or ‘Electrically floating’ are used inthis manual to indicate a measurement in which the testtool input BNC or banana jack is connected to a voltagedifferent from earth ground.

The isolated input connectors have no exposed metal andare fully insulated to protect against electrical shock.

The red and gray BNC jacks, and the red and black4-mm banana jacks can independently be connected to avoltage above earth ground for isolated (electricallyfloating) measurements and are rated up to 1000 VrmsCAT II and 600 Vrms CAT III above earth ground.

If Safety Features are Impaired

Use of the test tool in a manner not specified mayimpair the protection provided by the equipment.Before use, inspect the test leads for mechanical damageand replace damaged test leads!

Whenever it is likely that safety has been impaired, thetest tool must be turned off and disconnected from theline power. The matter should then be referred toqualified personnel. Safety is likely to be impaired if, forexample, the test tool fails to perform the intendedmeasurements or shows visible damage.

Page 15: 0MASTER - RS Components

7

Chapter 1Using The Scope

About this Chapter

This chapter provides a step-by-step introduction to thescope functions of the test tool. The introduction does notcover all of the capabilities of the scope functions butgives basic examples to show how to use the menus andperform basic operations.

Powering the Test ToolFollow the procedure (steps 1 through 3) in Figure 2 topower the test tool from a standard ac outlet.See Chapter 8 for instructions on using battery power.

Turn the test tool on with the on/off key.

The test tool powers up in its last setup configuration. Figure 2. Powering the Test Tool

Page 16: 0MASTER - RS Components

Fluke 192/196/199Users Manual

8

Resetting the Test ToolIf you want to reset the test tool to the factory settings, dothe following:

1 Turn the test tool off.

2 Press and hold.

3 Press and release.

The test tool turns on, and you should hear a doublebeep, indicating the reset was successful.

4 Release.

Now look at the display; you will see a screen that lookslike Figure 3.

Figure 3. The Screen After Reset

Page 17: 0MASTER - RS Components

Using The ScopeNavigating a Menu 1

9

Navigating a MenuThe following example shows how to use the test tool’smenus to select a function. Subsequently follow steps1 through 4 to open the scope menu and to choose anitem.

1 Press the SCOPE key to displaythe labels that define the presentuse for the four blue functionkeys at the bottom of the screen.

NoteTo hide the labels for full screen view, press theSCOPE key again. This toggling enables you tocheck the labels without affecting your settings.

2 Open the Waveform Optionsmenu. This menu is displayed atthe bottom of the screen.

Figure 4. Basic Navigation

3a

3b

Use the blue arrow keys tohighlight the item.Press the blue ENTER key toaccept the selection.

4 Press the ENTER key until you exitthe menu.

Note

Repeatedly pressing lets you to stepthrough a menu without changing the settings.

Page 18: 0MASTER - RS Components

Fluke 192/196/199Users Manual

10

Hiding Key Labels and MenusYou can hide a menu or key label at any time:

Hide any key label or menu.

To display menus or key labels, press one of the yellowmenu keys, e.g. the SCOPE key.

Input ConnectionsLook at the top of the test tool. The test tool has foursignal inputs: two safety BNC jack inputs (red input A andgray input B) and two safety 4-mm banana jack inputs(red and black). Use the two BNC jack inputs for scopemeasurements, and the two banana jack inputs for metermeasurements.

Isolated input architecture allows independent floatingmeasurements with each input.

Figure 5. Measurement Connections

Page 19: 0MASTER - RS Components

Using The ScopeMaking Scope Connections 1

11

Making Scope ConnectionsTo make dual input scope measurements, connect thered voltage probe to input A, and the gray voltage probeto input B. Connect the short ground leads of eachvoltage probe to its own reference potential. (See Figure6.)

Note

To maximally benefit from having independentlyisolated floating inputs and to avoid problemscaused by improper use, read Chapter 7: “Tips”.

Figure 6. Scope Connections

Page 20: 0MASTER - RS Components

Fluke 192/196/199Users Manual

12

Displaying an Unknown Signal withConnect-and-View™The Connect-and-View feature lets the test tool displaycomplex, unknown signals automatically. This functionoptimizes the position, range, time base, and triggeringand assures a stable display of virtually any waveform. Ifthe signal changes, the setup is automatically adjusted tomaintain the best display result. This feature is especiallyuseful for quickly checking several signals.

To enable the Connect-and-View feature, do thefollowing:

1 Perform an Auto Set. AUTO appears atthe top right of the screen.

The bottom line shows the range, the time base, and thetrigger information.

The waveform identifier (A) is visible on the bottom rightside of the screen, as shown in Figure 7. The input A zeroicon (-) at the left side of the screen identifies the groundlevel of the waveform.

2 Press a second time to select themanual range again. MANUAL appearsat the top right of the screen.

Figure 7. The Screen After an Auto Set

Use the light-gray RANGE, TIME and MOVE keys at thebottom of the keypad to change the view of the waveformmanually.

Page 21: 0MASTER - RS Components

Using The ScopeMaking Automatic Scope Measurements 1

13

Making Automatic Scope MeasurementsThe test tool offers a wide range of automatic scopemeasurements. You can display two numeric readings:READING 1 and READING 2. These readings are selectableindependently, and the measurements can be done onthe input A or input B waveform

To choose a frequency measurement for input A, do thefollowing:

1 Display the SCOPE key labels.

2 Open the Reading 1 menu.

3 Select on A. Observe that thehighlight jumps to the presentmeasurement.

4 Select the Hz measurement.

Observe that the top left of the screen displays the Hzmeasurement. (See Figure 8.)

To choose also a Peak-Peak measurement for Input B assecond reading, do the following:

1 Display the SCOPE key labels.

2 Open the Reading 2 menu.

3 Select on B. The highlight jumpsto the measurements field.

4 Open the PEAK menu.

5 Select the Peak-Peakmeasurement.

Page 22: 0MASTER - RS Components

Fluke 192/196/199Users Manual

14

Figure 8 shows an example of the screen. Note that thePeak-Peak reading for input B appears next to the input Afrequency reading at the top of the screen.

Figure 8. Hz and V peak-peak as Scope Readings

Freezing the ScreenYou can freeze the screen (all readings and waveforms)at any time.

1 Freeze the screen. HOLD appearsat the right of the reading area.

2 Resume your measurement.

Page 23: 0MASTER - RS Components

Using The ScopeUsing Average, Persistence and Glitch Capture 1

15

Using Average, Persistence and GlitchCapture

Using Average for Smoothing Waveforms

To smooth the waveform, do the following:

1 Display the SCOPE key labels.

2 Open the Waveform Optionsmenu.

3 Jump to Waveform:

4 Select Average... to open theAverage menu

5 Select Normal, jump to Factor:and select Average 8x.Thisaverages the outcomes of 8acquisitions.

6 Exit the menu.

You can use the average functions to suppress randomor uncorrelated noise in the waveform without loss ofbandwidth. Waveform samples with and withoutsmoothing are shown in Figure 9.

Figure 9. Smoothing a Waveform

Page 24: 0MASTER - RS Components

Fluke 192/196/199Users Manual

16

Using Persistence to Display Waveforms

When persistence is selected, the test tool displays theupper and lower boundaries of dynamic waveforms.

1 Display the SCOPE key labels.

2 Open the Waveform Optionsmenu.

3 Jump to Waveform: and SelectPersistence , then select Infinite.

4 Start monitoring the waveform.

Figure 10. Using Persistence to Display WaveformVariations

Observe that the fluctuating portion of the waveformappears in gray on the screen. See Figure 10.

You can use persistence to observe variations in time oramplitude of input waveforms over time.

Page 25: 0MASTER - RS Components

Using The ScopeUsing Average, Persistence and Glitch Capture 1

17

Displaying Glitches

To capture glitches on a waveform, do the following:

1 Display the SCOPE key labels.

2 Open the Waveform Optionsmenu.

3 Select Display Glitches: Yes

4 Exit the menu.

You can use this function to display events (glitches orother asynchronous waveforms) of 50 ns (nanoseconds)or wider, or you can display HF modulated waveforms.

Suppressing High Frequency Noise

Switching Display Glitches to No will suppress the highfrequency noise on a waveform. Averaging will suppressthe noise even more.

1 Display the SCOPE key labels.

2 Open the Waveform Optionsmenu.

3 Select Display Glitches: No,then select Average: to open theAverage menu

4 Select Average: Normal, andFactor : 8x

TipGlitch capture and average do not affectbandwidth. Further noise suppression is possiblewith bandwidth limiting filters. See Chapter 1:“Working with Noisy Waveforms”.

Page 26: 0MASTER - RS Components

Fluke 192/196/199Users Manual

18

Acquiring Waveforms

Selecting AC-Coupling

After a reset, the test tool is dc-coupled so that ac and dcvoltages appear on the screen.

Use ac-coupling when you wish to observe a small acsignal that rides on a dc signal. To select ac-coupling, dothe following:

1 Display the INPUT A key labels.

2 Highlight AC.

Observe that the bottom left of the screen displays theac-coupling icon: .

Reversing the Polarity of the DisplayedWaveform

To invert the input A waveform, do the following:

1 Display the INPUT A key labels.

2 Open the Input A menu.

3 Select Inverted and acceptinverted waveform display.

4 Exit the menu.

For example, a negative-going waveform is displayed aspositive-going waveform which may provide a moremeaningful view. An inverted display is identified by aninversed trace identifier ( ) at the right of the waveform.

Page 27: 0MASTER - RS Components

Using The ScopeAcquiring Waveforms 1

19

Working with Noisy Waveforms

To suppress high frequency noise on waveforms, you canlimit the working bandwidth to 10 kHz or 20 MHz. Thisfunction smoothes the displayed waveform. For the samereason, it improves triggering on the waveform.

To choose HF reject, do the following:

1 Display the INPUT A key labels.

2 Open the Input A menu.

3 Jump to Bandwidth.

4 Select 10kHz (HF reject) toaccept the bandwidth limitation.

Tip

To suppress noise without loss of bandwidth,use the average function or turn off DisplayGlitches.

Using Waveform Mathematics Functions

When adding (A+B), subtracting (A-B), or multiplying(A*B) the input A and input B waveform, the test tool willdisplay the mathematical result waveform and the input Aand input B waveforms.

A versus B provides a plot with input A on the verticalaxis and input B on the horizontal axis.

The Mathematics functions perform a point-to-pointoperation on waveforms A and B.

To use a Mathematics function, do the following:

1 Display the SCOPE key labels.

2 Open the Waveform Optionsmenu.

Page 28: 0MASTER - RS Components

Fluke 192/196/199Users Manual

20

3 Jump to Waveform: and SelectMathematics... to open theMathematics menu.

4 Select Function: A+B, A-B, A*Bor A vs B.

5 Select a scale factor to fit themathematical result waveform ontothe display, and return.

The sensitivity range of the mathematical result is equalto the sensitivity range of the least sensitive input dividedby the scale factor.

Analyzing WaveformsYou can use the analysis functions CURSOR, ZOOM andREPLAY to perform detailed waveform analysis. Thesefunctions are described in Chapter 4: “Using Cursors,Zoom and Replay”.

Page 29: 0MASTER - RS Components

21

Chapter 2Using The Multimeter

About this Chapter

This chapter provides a step-by-step introduction to themultimeter functions of the test tool (hereafter called“meter”). The introduction gives basic examples to showhow to use the menus and perform basic operations.

Making Meter ConnectionsUse the two 4-mm safety red ( ) and black (COM)banana jack inputs for the Meter functions.(See Figure 11.)

Note

Typical use of the Meter test leads andaccessories is shown in Chapter 7.

Figure 11. Meter Connections

Page 30: 0MASTER - RS Components

Fluke 192/196/199Users Manual

22

Making Multimeter MeasurementsThe screen displays the numeric readings of themeasurements on the meter input.

Measuring Resistance Values

To measure a resistance, do the following:

1 Connect the red and black test leads from the4-mm banana jack inputs to the resistor.

2

Display the METER key labels.

3 Open the Measurement menu.

4

Highlight Ohms.

5 Select Ohms measurement.

The resistor value is displayed in ohms. Observe alsothat the bargraph is displayed. (See Figure 12.)

Figure 12. Resistor Value Readings

Page 31: 0MASTER - RS Components

Using The MultimeterMaking Multimeter Measurements 2

23

Making a Current Measurement

You can measure current in both Scope mode and Metermode. Scope mode has the advantage of two waveformsbeing displayed while you perform measurements.Meter mode has the advantage of high measurementresolution.

The next example explains a typical currentmeasurement in Meter mode.

WarningCarefully read the instructions about thecurrent probe you are using.

To set up the test tool, do the following:

1 Connect a current probe (e.g. i400, optional)from the 4-mm banana jack outputs to theconductor to be measured.

Ensure that the red and black probe connectorscorrespond to the red and black banana jackinputs. (See Figure 13.)

2 Display the METER key labels.

Figure 13. Measurement Setup

3 Open the Measurement menu.

4 Highlight A ac....

5 Open the Current Probesubmenu.

Page 32: 0MASTER - RS Components

Fluke 192/196/199Users Manual

24

6 Observe the sensitivity of thecurrent probe. Highlight thecorresponding sensitivity in themenu, e.g. 10 mV/A.

7 Accept the current measurement.

Now, you will see a screen like in Figure 14.

Figure 14. Ampere Measurement Readings

Page 33: 0MASTER - RS Components

Using The MultimeterFreezing the Readings 2

25

Freezing the ReadingsYou can freeze the displayed readings at any time.

1 Freeze the screen. HOLD appearsat the top right of the readingarea.

2 Resume your measurement.

You can use this function to hold accurate readings forlater examination.

Note

For saving screens into memory, see Chapter 6.

Selecting Auto/Manual RangesTo activate manual ranging, do the following during anyMeter measurement:

1 Activate manual ranging.

2 Increase or decrease the range.

Observe how the bargraph sensitivity changes.

Use manual ranging to set a fixed bargraph sensitivityand decimal point.

3 Choose auto ranging again.

When in auto ranging, the bargraph sensitivity anddecimal point are automatically adjusted while checkingdifferent signals.

Page 34: 0MASTER - RS Components

Fluke 192/196/199Users Manual

26

Making Relative MeasurementsA relative measurement displays the presentmeasurement result relative to a defined reference value.

The following example shows how to perform a relativevoltage measurement. First obtain a reference value:

1 Display the METER key labels.

2 Measure a voltage to be used asreference value.

3 Set RELATIVE to ON. (ON ishighlighted.)

This stores the reference value as reference forsubsequent measurements. The stored reference value isdisplayed in small digits at the bottom right side of thescreen after the word REFERENCE.

4 Measure the voltage to becompared to the reference.

Observe that the main reading is displayed as variationsfrom the reference value. The actual reading with itsbargraph is displayed beneath these readings. (SeeFigure 15.)

Figure 15. Making a Relative Measurement

You can use this feature when, for example, you need tomonitor input activity (voltage, resistance, temperature) inrelation to a known good value.

Page 35: 0MASTER - RS Components

27

Chapter 3Using The Recorder Functions

About this Chapter

This chapter provides a step-by-step introduction to therecorder functions of the test tool. The introduction givesexamples to show how to use the menus and performbasic operations.

Opening the Recorder Main MenuFirst choose a measurement in scope or meter mode.Now you can choose the recorder functions from therecorder main menu. To open the main menu, do thefollowing:

1 Open the RECORDER main menu.(See Figure 16.)

Figure 16. Recorder Main Menu

Page 36: 0MASTER - RS Components

Fluke 192/196/199Users Manual

28

Plotting Measurements Over Time(TrendPlot™)Use the TrendPlot function to plot a graph of Scope orMeter measurements as function of time.

Note

Because the navigations for the dual inputTrendPlot (Scope) and the single input TrendPlot(Meter) are identical, only TrendPlot (Scope) isexplained in the next sections.

Starting a TrendPlot Function

To start plotting a graph of the reading over time, do thefollowing:

1 Apply a signal to the red BNC input A and turnon Reading 1 in scope mode

2 Open the RECORDER main menu.

3 Highlight Trend Plot (Scope).

4 Start the TrendPlot recording.

The test tool continuously records the digital readings ofthe input A measurements and displays these as a graph.The TrendPlot graph rolls from right to left like a paperchart recorder.

Observe that the recorded time from start appears at thebottom of the screen. The present reading appears ontop of the screen. (See Figure 17.)

Note

When simultaneously TrendPlotting tworeadings, the screen area is split into twosections of four divisions each.

Page 37: 0MASTER - RS Components

Using The Recorder FunctionsPlotting Measurements Over Time (TrendPlot™) 3

29

Figure 17. TrendPlot Reading

When the Scope is in automatic mode, automatic verticalscaling is used to fit the TrendPlot graph on the screen.

5 Set RECORDER to STOP to freezethe recorder function.

6 Set RECORDER to RUN to continue.

Displaying Recorded Data

When in normal view (NORMAL), only the nine mostrecently recorded divisions are displayed on screen. Allprevious recordings are stored in memory.

VIEW ALL shows all data in memory:

7 Display an overview of the fullwaveform.

Press repeatedly to toggle between normal view(NORMAL) and overview (VIEW ALL)

When the recorder memory is full, an automaticcompression algorithm is used to compress all samplesinto half of the memory without loss of transients. Theother half of the recorder memory is free again tocontinue recording.

Page 38: 0MASTER - RS Components

Fluke 192/196/199Users Manual

30

Changing the Recorder Options

At the right bottom of the display you can choose todisplay the time elapsed from start and the actual time ofthe day.

To change the time reference, proceed from step 6 asfollows:

7 Open the Recorder Optionsmenu.

8 Select Time of Day and jump tothe next field.

Now the recorded time and the current time appear at thebottom of the screen.

The options Reading 1 and Reading 2 allow you toselect the scope readings to be recorded. (Or one metermeasurement when in TrendPlot meter mode.)

To continue without making further changes:

9 Close.

Turning Off the TrendPlot Display

10 Exit the recorder function.

Page 39: 0MASTER - RS Components

Using The Recorder FunctionsRecording Scope Waveforms In Deep Memory (Scope Record™) 3

31

Recording Scope Waveforms In DeepMemory (Scope Record™)The SCOPE RECORD function is a roll mode that logs one ortwo long waveforms. This function can be used to monitorwaveforms like motion control signals or the power-onevent of an Uninterruptable Power Supply (UPS). Duringrecording, fast transients are captured. Because of thedeep memory, recording can be done for more than oneday. This function is similar to the roll mode in manyDSO’s but has deeper memory and better functionality.

Starting a Scope Record Function

1 Apply a signal to the red BNC input A.

2 From the Recorder main menu,highlight Scope Record.

3 Start the recording.

The waveform moves across the screen from right to leftlike a normal chart recorder. (See Figure 18.)

Figure 18. Recording Waveforms

Observe that the top of the screen displays the following:

• Time from start at the top of the screen.

• The status at the bottom of the screen whichincludes the time/div setting as well as the totaltimespan that fits the memory.

Note

For accurate recordings it is advised to let theinstrument first warm up for five minutes.

Page 40: 0MASTER - RS Components

Fluke 192/196/199Users Manual

32

Displaying Recorded Data

In Normal view, the samples that roll off the screen arestored in deep memory. When the memory is full,recording continues by shifting the data in memory anddeleting the first samples out of memory.

In View All mode, the complete memory contents aredisplayed on the screen.

4 Press to toggle between VIEW ALL

(overview of all recordedsamples) and NORMAL view.

You can analyze the recorded waveforms using theCursors and Zoom functions. See Chapter 4: “UsingReplay, Zoom and Cursors”.

Using ScopeRecord in Single Sweep Mode

Use the recorder Single Sweep function to automaticallystop recording when the deep memory is full.

Continue from step 3 of the previous section:

4 Open the Recorder optionsmenu.

5 (2x) Jump to the Mode field.

6 Select Single Sweep and acceptthe recorder options.

Page 41: 0MASTER - RS Components

Using The Recorder FunctionsRecording Scope Waveforms In Deep Memory (Scope Record™) 3

33

Using Scope Record in Triggered Single SweepMode

To record, for example, the power-on event of anUninterruptable Power Supply (UPS) it might be useful tostart recording on an external trigger signal. The nextexample explains a typical triggered Single Sweeprecording.

To set up the test tool, continue from step 3 of theprevious section:

4 Apply an Uninterruptable Power Supply signalto the red BNC input A. Apply a start signal tothe red and black external trigger bananainputs. (See Figure 19.)

5 Open the Recorder Optionsmenu.

6 Jump to Display Glitches.

7 Select Yes, then jump to Mode.

Figure 19. Scope Record Single Sweep

8 Select on EXT to open theStart Single Sweep on Ext.menu.

9 Select positive slope and jump toLevel.

10 Select 1.2 V and accept allrecorder options.

Page 42: 0MASTER - RS Components

Fluke 192/196/199Users Manual

34

The test tool starts recording after receiving the startsignal. All samples are continuously saved in deepmemory until the memory is full. The last nine recordeddivisions are displayed on the screen. Use View All todisplay the full memory contents.

Note

To learn more about the Single Shot triggerfunction, see Chapter 5 “Triggering onWaveforms”.

Figure 20. Triggered Single Sweep Recording

Analyzing a TrendPlot or Scope RecordFrom a Scope TrendPlot or Scope Record you can usethe analysis functions CURSORS and ZOOM to performdetailed waveform analysis. These functions aredescribed in Chapter 4: “Using Replay, Zoom andCursors”.

Page 43: 0MASTER - RS Components

35

Chapter 4Using Replay, Zoom and Cursors

About this Chapter

This chapter covers the capabilities of the analysisfunctions Cursor, Zoom, and Replay. These functionscan be used with one or more of the primary functionsScope, TrendPlot or Scope Record.

It is possible to combine two or three analysis functions.A typical application using these functions follows:

• First replay the last screens to find the screen ofspecial interest.

• Then zoom in on the signal event.

• Finally, make measurements using the cursors.

Replaying the 100 Most Recent ScopeScreensWhen you are in scope mode, the test tool automaticallystores the 100 most recent screens. When you press theHOLD key or the REPLAY key, the memory contents arefrozen. Use the functions in the REPLAY menu to “go backin time” by stepping through the stored screens to find thescreen of your interest. This feature lets you capture andview signals even if you did not press HOLD.

Page 44: 0MASTER - RS Components

Fluke 192/196/199Users Manual

36

Replaying Step-by-Step

To step through the last scope screens, do the following:

1 From scope mode, open theREPLAY menu.

Observe that the trace is frozenand that REPLAY appears at thetop of the screen (see Figure 21).

2 Step through the previousscreens.

3 Step through the next screens.

Observe that the bottom of the waveform area displaysthe replay bar with a screen number and related timestamp:

Figure 21. Replaying a Waveform

The replay bar represents all 100 stored screens inmemory. The icon represents the picture beingdisplayed on the screen (in this example: SCREEN -84). Ifthe bar is partly white, the memory is not completely filledwith 100 screens.

From this point you can use the zoom and cursorfunctions to study the signal in more detail.

Page 45: 0MASTER - RS Components

Using Replay, Zoom and CursorsReplaying the 100 Most Recent Scope Screens 4

37

Replaying Continuously

You can also replay the stored screens continuously, likeplaying a video tape.

To replay continuously, do the following:

1 From Scope mode, open theREPLAY menu.

Observe that the trace is frozenand REPLAY appears at the top ofthe screen.

2 Continuously replay the storedscreens in ascending order.

Wait until the screen with the signal event of interestappears.

3 Stop the continuous replay.

Turning Off the Replay Function

4 Turn off REPLAY.

Capturing 100 Intermittents Automatically

When you use the test tool in triggered mode, 100triggered screens are captured. This way you could usePulse Triggering to trigger and capture 100 intermittentglitches or you could use External Triggering to capture100 UPS startups.

By combining the trigger possibilities with the capability ofcapturing 100 screens for later replay, you can leave thetest tool unattended to capture intermittent signalanomalies.

For triggering, see Chapter 5: “Triggering on Waveforms”.

Page 46: 0MASTER - RS Components

Fluke 192/196/199Users Manual

38

Zooming in on a WaveformTo obtain a more detailed view of a waveform, you canzoom in on a waveform using the ZOOM function.

To zoom in on a waveform, do the following:

1 Display the ZOOM key labels.

Observe that the trace is frozen,HOLD appears at the top of thescreen, and the waveform ismagnified.

2 Enlarge (decrease the time/div)or shrink (increase the time/div)the waveform.

3 Scroll. A position bar displays theposition of the zoomed part inrelation to the total waveform.

Tip

Even when the key labels are not displayed atthe bottom of the screen, you can still use thearrow keys to zoom in and out.

Figure 22. Zooming in a Waveform

Observe that the bottom of the waveform area displaysthe zoom ratio, position bar, and time/div (see Figure 22).The zoom range depends on the amount of data samplesstored in memory.

From this point you can use the cursor function for furthermeasurements on the waveform.

Page 47: 0MASTER - RS Components

Using Replay, Zoom and CursorsZooming in on a Waveform 4

39

Displaying the Zoomed Waveform

The VIEW ALL feature is useful when you quickly need tosee the complete waveform and then return to thezoomed part.

4 Display the complete waveform.

Press repeatedly to toggle between the zoomedpart of the waveform and the complete waveform.

Turning Off the Zoom Function

5 Turn off the ZOOM function.

Page 48: 0MASTER - RS Components

Fluke 192/196/199Users Manual

40

Making Cursor MeasurementsCursors allow you to make precise digital measurementson waveforms. This can be done on live waveforms,recorded waveforms, and on saved waveforms.

Using Horizontal Cursors on a Waveform

To use the cursors for a voltage measurement, do thefollowing:

1 From scope mode, display thecursor key labels.

2 Press to highlight . Observethat two horizontal cursors aredisplayed.

3 Highlight the upper cursor.

4 Move the upper cursor to thedesired position on the screen.

5 Highlight the lower cursor.

6 Move the lower cursor to thedesired position on the screen.

Note

Even when the key labels are not displayed atthe bottom of the screen, you still can use thearrow keys. This allows full control of bothcursors while having full screen view.

Figure 23. Voltage Measurement with Cursors

The screen shows the voltage difference between the twocursors and the voltage at the cursors. (See Figure 23.)

Use horizontal cursors to measure the amplitude, high orlow value, or overshoot of a waveform.

Page 49: 0MASTER - RS Components

Using Replay, Zoom and CursorsMaking Cursor Measurements 4

41

Using Vertical Cursors on a Waveform

To use the cursors for a time measurement, do thefollowing:

1 From scope mode, display thecursor key labels.

2 Press to highlight . Observethat two vertical cursors aredisplayed. Markers (—) identifythe point where the cursors crossthe waveform.

3 If necessary, choose the trace:TRACE A ,B, or M (Mathematics).

4 Highlight the left cursor.

5 Move the left cursor to thedesired position on thewaveform.

6 Highlight the right cursor.

Figure 24. Time Measurement with Cursors

7 Move the right cursor to thedesired position on thewaveform.

The screen shows the time difference between thecursors and the voltage difference between the twomarkers. (See Figure 24.)

8 Turn off the cursors.

Page 50: 0MASTER - RS Components

Fluke 192/196/199Users Manual

42

Using Cursors on a A+B, A-B or A*B Waveform

Cursor measurements on a A*B waveform give a readingin Watts if input A measures (milli)Volts and input Bmeasures (milli)Amperes.

For other cursor measurements on a A+B, A-B or A*Bwaveform no reading will be available if the input A andinput B measurement unit are different.

Making Rise Time Measurements

To measure rise time, do the following:

1 From scope mode, display thecursor key labels.

2 Press to highlight (rise time).Observe that two horizontalcursors are displayed.

3 If only one trace is displayed,select MANUAL or AUTO (thisautomatically does steps 4 to 6).For multiple traces select therequired trace A, B, or M (if amath function is active).

4 Move the upper cursor to 100%of the trace height. A marker isshown at 90%.

5 Highlight the other cursor.

6 Move the lower cursor to 0% ofthe trace height. A marker isshown at 10%.

The reading shows the risetime from 10%-90% of thetrace amplitude.

Figure 25. Risetime Measurement

Page 51: 0MASTER - RS Components

43

Chapter 5Triggering on Waveforms

About this Chapter

This chapter provides an introduction to the triggerfunctions of the test tool. Triggering tells the test toolwhen to begin displaying the waveform. You can usefully automatic triggering, take control of one or moremain trigger functions (semi-automatic triggering), or youcan use dedicated trigger functions to capture specialwaveforms.

Following are some typical trigger applications:

• Use the Connect-and-View™ function to have fullautomatic triggering and instant display of virtuallyany waveform.

• If the signal is unstable or has a very low frequency,you can control the trigger level, slope, and triggerdelay for a better view of the signal. (See nextsection.)

• For dedicated applications, use one of the fourmanual trigger functions:

• Edge triggering

• External triggering

• Video triggering

• Pulse Width triggering

Page 52: 0MASTER - RS Components

Fluke 192/196/199Users Manual

44

Setting Trigger Level and SlopeThe Connect-and-View™ function enables hands-offtriggering to display complex unknown signals.

When your test tool is in manual range, do the following:

Perform an auto set. AUTO appearsat the top right of the screen.

Automatic triggering assures a stable display of virtuallyany signal.

From this point, you can take over the basic triggercontrols such as level, slope and delay. To optimizetrigger level and slope manually, do the following:

1 Display the TRIGGER key labels.

2 Trigger on either positive slope ornegative slope of the chosenwaveform.

3 Enable the arrow keys for manualtrigger level adjustment.

Figure 26. Screen with all Trigger Information

4 Adjust the trigger level.

Observe the trigger icon that indicates the triggerposition, trigger level, and slope.

At the bottom of the screen the trigger parameters aredisplayed. (See Figure 26.) For example, means that input A is used as the trigger source with apositive slope.

When no trigger is found, the trigger parameters appearin gray.

Page 53: 0MASTER - RS Components

Triggering on WaveformsUsing Trigger Delay or Pre-trigger 5

45

Using Trigger Delay or Pre-triggerYou can begin to display the waveform some time beforeor after the trigger point has been detected. Initially, youhave 2 divisions of pre-trigger view (negative delay).

To set the trigger delay, do the following:

5 Hold down to adjust the triggerdelay.

Observe that the trigger icon on the screen moves toshow the new trigger position. When the trigger positionmoves left off of the screen, the trigger icon changes into

to indicate that you have selected a trigger delay.Moving the trigger icon to the right on the display givesyou a pre-trigger view.

In case of a trigger delay, the status at the bottom of thescreen will change. For example:

This means that input A is used as the trigger source witha positive slope. The 500.0 ms indicates the (positive)delay between trigger point and waveform display.

When no trigger is found, the trigger parameters appearin gray.

Figure 27. Trigger Delay or Pre-trigger View

Figure 27 shows an example of a trigger delay of 500 ms(top) and an example of pre-trigger view of 8 divisions(bottom).

Page 54: 0MASTER - RS Components

Fluke 192/196/199Users Manual

46

Automatic Trigger OptionsIn the trigger menu, settings for automatic triggering canbe changed as follows. (See also Chapter 1: “Displayingan Unknown Signal with Connect-and-View”)

1 Display the TRIGGER key labels.

Note

The TRIGGER key labels can differ depending onthe latest trigger function used.

2 Open the Trigger Options menu.

3 Open the Automatic Triggermenu.

If the frequency range of the automatic triggering is set to> 15 Hz, the Connect-and-View™ function respondsmore quickly. The response is quicker because the testtool is instructed not to analyze low frequency signalcomponents. However, when you measure frequencieslower than 15 Hz, the test tool must be instructed toanalyze low frequency components for automatictriggering:

4 Select > 1 HZ and return to themeasurement screen.

Page 55: 0MASTER - RS Components

Triggering on WaveformsTriggering on Edges 5

47

Triggering on EdgesIf the signal is instable or has a very low frequency, useedge triggering to obtain full manual trigger control.

To trigger on rising edges of the input A waveform, do thefollowing:

1 Display the TRIGGER key labels.

2 Open the Trigger Options menu.

3 Open the Trigger on Edge menu.

When Free Run is selected, the test tool updates thescreen even if there are no triggers. A trace alwaysappears on the screen.

When On Trigger is selected, the test tool needs atrigger to display a waveform. Use this mode if you wantto update the screen only when valid triggers occur.

When Single Shot is selected, the test tool waits for atrigger. After receiving a trigger, the waveform isdisplayed and the instruments is set to HOLD.

In most cases it is advised to use the Free Run mode:

4 Select Free Run, jump to Noisereject Filter.

5 Set Noise reject Filter to Off.

Observe that the key labels at the bottom of the screenhave adapted to allow further selection of specific edgetrigger settings:

Page 56: 0MASTER - RS Components

Fluke 192/196/199Users Manual

48

Triggering on Noisy Waveforms

To reduce jitter on the screen when triggering on noisywaveforms, you can use a noise rejection filter. Continuefrom step 3 of the previous example as follows:

4 Select On Trigger, jump to Noisereject Filter.

5 Set Noise reject Filter to On.

Observe that the trigger gap has increased. This isindicated by a taller trigger icon .

Making a Single Acquisition

To catch single events, you can perform a single shotacquisition (one-time screen update). To set up the testtool for a single shot of the input A waveform, continuefrom step 3 again:

4 Select Single Shot.

5 Accept the settings.

The word WAITING appears at the top of the screenindicating that the test tool is waiting for a trigger. Assoon as the test tool receives a trigger, the waveform isdisplayed and the instrument is set to hold. This isindicated by the word HOLD at top of the screen.

Page 57: 0MASTER - RS Components

Triggering on WaveformsTriggering on Edges 5

49

The test tool will now have a screen like Figure 28.

6 Arm the test tool for a new singleshot.

Tip

The test tool stores all single shots in the replaymemory. Use the Replay function to look at allthe stored single shots.

Figure 28. Making a Single Shot Measurement

Page 58: 0MASTER - RS Components

Fluke 192/196/199Users Manual

50

Triggering on External WaveformsUse external triggering when you want to displaywaveforms on inputs A and B while triggering on a thirdsignal. You can choose external triggering with automatictriggering or with edge triggering.

1 Supply a signal to the red and black 4-mmbanana jack inputs. See Figure 29.

In this example you continue from the Trigger on Edgesexample. To choose the external signal as trigger source,continue as follows:

2 Display the TRIGGER (On Edges)key labels.

3 Select Ext (external) edge trigger.

Observe that the key labels at the bottom of the screenhave been adapted to allow selection of two differentexternal trigger levels: 0.12 V and 1.2 V:

Figure 29. External Triggering

4 Select 1.2V under the Ext LEVELlabel.

From this point the trigger level is fixed and is compatiblewith logic signals.

Page 59: 0MASTER - RS Components

Triggering on WaveformsTriggering on Video Signals 5

51

Triggering on Video Signals

To trigger on a video signal, first select the standard ofthe video signal you are going to measure:

1 Apply a video signal to the red input A.

2 Display the TRIGGER key labels.

3 Open the Trigger Options menu.

4 Select Video on A to open theTrigger on Video menu.

5 Select positive signal polarity forvideo signals with negative goingsync pulses.

Figure 30. Measuring Interlaced Video Signals

6 Select the video standard andreturn.

Trigger level and slope are now fixed.

Observe that the key labels at the bottom of the screenhave been changed to allow further selection of specificvideo trigger settings:

Page 60: 0MASTER - RS Components

Fluke 192/196/199Users Manual

52

Triggering on Video Frames

Use FIELD 1 or FIELD 2 to trigger either on the first halfof the frame (odd) or on the second half of the frame(even).

To trigger on the second half of the frame, do thefollowing:

7 Choose FIELD 2.

The signal part of the even field is displayed on thescreen.

Triggering on Video Lines

Use ALL LINES to trigger on all line synchronization pulses(horizontal synchronization).

7 Choose ALL LINES.

The signal of one line is displayed on the screen. Thescreen is updated with the signal of the next lineimmediately after the test tool triggers on the horizontalsynchronization pulse.

To view a specific video line in more detail, you can selectthe line number. For example, to measure on video line123, continue from step 5 as follows:

7 Enable video line selection.

8 Select number 123.

The signal of line 123 is displayed on the screen.Observe that the status line now also shows the selectedline number. The screen is continuously updated with thesignal of line 123.

Page 61: 0MASTER - RS Components

Triggering on WaveformsTriggering on Pulses 5

53

Triggering on PulsesUse pulse width triggering to isolate and display specificpulses that you can qualify by time, such as glitches,missing pulses, bursts or signal dropouts.

Detecting Narrow Pulses

To set the test tool to trigger on narrow positive pulsesshorter than 5 ms, do the following:

1 Apply a video signal to the red input A.

2 Display the TRIGGER key labels.

3 Open the Trigger Options menu.

4 Select Pulse Width on A... toopen the Trigger on PulseWidth menu.

5 Select the positive pulse icon,then jump to Condition.

6 Select <t, then jump to Update.

7 Select On Trigger.

The test tool is now prepared to trigger on narrow pulsesonly. Observe that the trigger key labels at the bottom ofthe screen have been adapted to set the pulseconditions:

Page 62: 0MASTER - RS Components

Fluke 192/196/199Users Manual

54

To set the pulse width to 5 ms, do the following:

7 Enable the arrow keys to adjustthe pulse width.

8 Select 5 ms.

All narrow positive pulses shorter than 5 ms are nowdisplayed on the screen. (See Figure 31.)

Tip

The test tool stores all triggered screens in thereplay memory. For example, if you setup yourtriggering for glitches, you can capture 100glitches with time stamps. Use the REPLAY key tolook at all the stored glitches.

Figure 31. Triggering on Narrow Glitches

Page 63: 0MASTER - RS Components

Triggering on WaveformsTriggering on Pulses 5

55

Finding Missing Pulses

The next example covers finding missing pulses in a trainof positive pulses. In this example it is assumed that thepulses have a 100 ms distance between the rising edges.If the time accidently increases to 200 ms, a pulse ismissing. To set the test tool to trigger on such missingpulses, let it trigger on gaps bigger than about 150 ms.Do the following:

1 Display the TRIGGER key labels.

2 Open the Trigger Options menu.

3 Select Pulse Width on A... toopen the Trigger on PulseWidth menu.

4 Select the positive pulse icon totrigger on the gap between thepositive pulses, then jump toCondition.

5 Select >t, then jump to Update.

6 Select On Trigger.

The test tool is now prepared to trigger on pulse gaps.Observe that the trigger menu at the bottom of thescreen has been adapted to set the pulse condition:

Page 64: 0MASTER - RS Components

Fluke 192/196/199Users Manual

56

To set the pulse width to 150 ms, continue as follows:

7 Enable the arrow keys to adjustthe pulse width.

8 Select 150 ms.

Figure 32. Triggering on Missing Pulses

Page 65: 0MASTER - RS Components

57

Chapter 6Using Memory, PC and Printer

About this Chapter

This chapter provides a step-by-step introduction to thegeneral functions of the test tool that can be used in thethree main modes: Scope, Meter, or Recorder. You willfind information on printer and computer communicationat the end of this chapter.

Saving and RecallingYou can:

• Save screens and setups to memory, and recallthem again from memory. The test tool has 10screen and setup memories and 2 record and setupmemories.

• Recall screens and recordings to analyze or print thescreen image at a later date.

• Recall a setup to continue a measurement with therecalled operating configuration.

Page 66: 0MASTER - RS Components

Fluke 192/196/199Users Manual

58

Saving Screens with Associated Setups

To save a screen in memory location 10, do the following:

1 Display the SAVE/PRINT key labels.

From this point the screen is frozen until you hide theSAVE/PRINT key labels again.

2 Open the Save menu.

Observe that free memorylocations are indicated by an opensquare (p). Filled memorylocations are indicated by a solidsquare (n).

3 Highlight SCREEN+SETUP location 10.

4 Save the actual screen.

Note

The two record+setup memory locations storemore than what is just visible on the screen.In TrendPlot or scope record mode the fullrecording is saved. In scope mode you can saveall 100 replay screens in a single record+setupmemory location.

Page 67: 0MASTER - RS Components

Using Memory, PC and PrinterSaving and Recalling 6

59

Deleting Screens with Associated Setups

To delete all screens and associated setups, continuefrom step 2 of the previous example as follows:

3 Delete all saved screens andsetups.

To delete only one screen and setup, continue from step2 of the previous example as follows:

3 Highlight SCREEN+SETUP location 5.

4 Delete the saved screen+setup ofmemory location 5.

Recalling Screens with Associated Setups

To recall screen+setup 2, do the following:

1 Display the SAVE/PRINT key labels.

2 Open the Recall menu.

3 Highlight SCREEN+SETUP location 2.

4 Use RECALL to recall the savedscreen.

Observe that the recalled waveform is displayed and thatHOLD appears on the screen. From this point you can usecursors and zoom for analysis or you can print therecalled screen.

Page 68: 0MASTER - RS Components

Fluke 192/196/199Users Manual

60

Recalling a Setup Configuration

To recall the setup configuration from memory 2, do thefollowing:

1 Display the SAVE/PRINT key labels.

2 Open the Recall menu.

3 Highlight SCREEN+SETUP location 2.

4 Use RECALL SETUP to recall thesaved setup.

Observe that RUN appears at the top right of the screen.From this point you continue in the new operatingconfiguration.

Viewing Stored Screens

To scroll through the memories while looking at thestored screens, do the following:

1 Display the SAVE/PRINT key labels.

2 Open the viewer.

3 Scroll through all stored screens.

Use this function to quickly find a stored screen.

Page 69: 0MASTER - RS Components

Using Memory, PC and PrinterDocumenting Screens 6

61

Documenting ScreensWith the FlukeView® software you can upload waveformdata and screen bitmaps to your PC or notebookcomputer for further processing. Printing can also bedone by connecting the test tool directly to a printer.

Connecting to a Computer

To connect the test tool to a PC or notebook computerand use the FlukeView software for Windows® (SW90W),do the following:

l Use the Optically Isolated RS-232 Adapter/Cable(PM9080) to connect a computer to the OPTICALPORT of the test tool. (See Figure 33.)

Figure 33. Connecting a Computer

Note

For information about installing and using theFlukeView ScopeMeter software, see theSW90W Users Manual.

A Software & Cable Carrying Case Kit is optionallyavailable as model number SCC190.

Page 70: 0MASTER - RS Components

Fluke 192/196/199Users Manual

62

Connecting to a Printer

To print a screen directly to a printer, use one of thefollowing adapters:

l The Optically Isolated RS-232 Adapter/Cable(PM9080) to connect a serial printer to the OPTICALPORT of the test tool. (See Figure 34.)

l The Print Adapter Cable (PAC91, optional) toconnect a parallel printer to the OPTICAL PORT ofthe test tool. (See Figure 35.)

Before printing, you must setup the test tool for a specificprinter.

Figure 34. Connecting a Serial Printer

Figure 35. Connecting a Parallel Printer

Page 71: 0MASTER - RS Components

Using Memory, PC and PrinterDocumenting Screens 6

63

Setting up the Printing Configuration

This example demonstrates how to set up the test tool forprinting on a postscript printer with a 9600 baud rate:

1 Display the USER OPTIONS keylabels.

2 Open the User Options menu.

3 Open the Printer Setup submenu.

4 Select Postscript and jump toBaud Rate.

5 Select a baud rate of 9600 andreturn to normal mode.

Whenever possible, choose the option Postscript whenprinting screens. This option gives the best printingresults. Consult the manual that came with your printer tofind out whether it has Postscript printing possibilities.

Printing a Screen

To print the currently displayed screen, do the following:

1 Clear the menu if you do not wantto print it.

2 Display the SAVE/PRINT key labels.

3 Start printing.

A message appears at the bottom of the screenindicating that the test tool is busy printing.

Page 72: 0MASTER - RS Components

Fluke 192/196/199Users Manual

64

Page 73: 0MASTER - RS Components

65

Chapter 7Tips

About this Chapter

This chapter gives you information and tips on how youcan make the best use of the test tool.

Using the Standard AccessoriesThe following illustrations show the use of the standardaccessories such as voltage probes, test leads, and thevarious clips.

Page 74: 0MASTER - RS Components

Fluke 192/196/199Users Manual

66

Figure 36. Probing Using 2-mm Heavy DutyTest Probes

WarningTo avoid electrical shock or fire, do notconnect the ground spring to voltages higherthan 30 Vrms from earth ground.

Figure 37. HF Voltage Probe Connection UsingGround Spring

Figure 38. Heavy Duty Fixed Connections for ScopeMeasurements Using Industrial Alligator Clips

Figure 39. Electronic Connections for ScopeMeasurements Using Hook Clips and Hook Clip

Grounding

Page 75: 0MASTER - RS Components

TipsUsing the Standard Accessories 7

67

Figure 40. Electronic Connections for ScopeMeasurements Using Hook Clips and Alligator Clip

Grounding

Figure 41. Fixed Electronic Connections for ScopeMeasurements with Fixed External Triggering

Figure 42. Manual Probing for Meter Measurementsusing 2-mm Test Probes

Figure 43. Heavy Duty Fixed Connections for MeterMeasurements Using Industrial Alligator Clips

Page 76: 0MASTER - RS Components

Fluke 192/196/199Users Manual

68

Using the Independently FloatingIsolated InputsYou can use the independently floating isolated inputs tomeasure signals that are independently floating fromeach other.

Independently floating isolated inputs offer additionalsafety and measurement capabilities compared to inputswith common references or grounds.

Measuring Using Independently FloatingIsolated Inputs

The test tool has independently floating isolated inputs.Each input section (A, B, External Trigger / DMM) has itsown signal input and its own reference input. Thereference input of each input section is electricallyisolated from the reference inputs of the other inputsections. The isolated input architecture makes the testtool about as versatile as having three independentinstruments. The advantages of having independentlyfloating isolated inputs are:

• It allows simultaneous measurement ofindependently floating signals.

• Additional safety. Since the commons are not directlyconnected, the chance of causing short circuit whenmeasuring multiple signals is greatly reduced.

• Additional safety. When measuring in systems withmultiple grounds, the ground currents induced arekept to a minimum.

Because the references are not connected togetherinside the test tool, each reference of the used inputsmust be connected to a reference voltage.

Independently floating isolated inputs are still coupled byparasitic capacitance. This can occur between the inputreferences and the environment, and between the inputreferences mutually (see Figure 44). For this reason, youshould connect the references to a system ground oranother stable voltage. If the reference of an input isconnected to a high speed and / or high voltage signal,you should be aware of parasitic capacitance.(See Figures 45, 46, 47 and 48.)

Page 77: 0MASTER - RS Components

TipsUsing the Independently Floating Isolated Inputs 7

69

Figure 44. Parasitic capacitance between probes,instrument and environment

M

MOTION/MOTOR CONTROLLERDC BUS

DC BUS

DIGITALGROUND

+

-

Figure 45. Parasitic capacitance between analog anddigital reference

ANALOG INPUT DIGITAL CONTROLLER

ANALOGGROUND

DIGITALGROUND

Figure 46. Correct connection of reference leadsANALOG INPUT DIGITAL CONTROLLER

ANALOGGROUND

DIGITALGROUND

Figure 47. Wrong connection of reference leads

Noise that is picked up by reference lead B can betransmitted by parasitic capacitance to the analog inputamplifier.

Page 78: 0MASTER - RS Components

Fluke 192/196/199Users Manual

70

Using the Tilt StandThe test tool is equipped with a tilt stand, allowing viewingfrom an angle while placed on a table. From this positionyou can access the OPTICAL PORT at the side of thetest tool. The typical position is shown in Figure 48.

Figure 48. Using the Tilt Stand

Resetting the Test ToolIf you want to reset the test tool to the factory settings, dothe following:

1 Turn the test tool off.

2 Press and hold.

3 Press and release.

The test tool turns on, and you should hear a doublebeep, indicating the reset was successful.

4 Release.

Suppressing Key Labels and Menu’sYou can hide a menu or key label at any time:

Hide any key label or menu.

To display menus or key labels, press one of the yellowmenu keys, e.g. the SCOPE key.

Page 79: 0MASTER - RS Components

TipsChanging the Information Language 7

71

Changing the Information LanguageDuring operation of the test tool, messages may appearat the bottom of the screen. You can select the languagein which these messages are displayed. The combinationof selectable languages (one or more) depends on theversion ordered.

In this example you can select English or French . Tochange the language from English to French, do thefollowing:

1 Display the USER key labels.

2 Open the Language Selectmenu.

3 Highlight FRENCH.

4 Accept French as the language.

Adjusting the Contrast and BrightnessAfter power-up, the screen has a bright display. To adjustthe contrast and backlight brightness, do the following:

1 Display the USER key labels.

2 Enable the arrow keys for manualcontrast and backlightadjustment.

3 Adjust the contrast of the screen.

4 Change the backlight.

NoteThe new contrast and brightness are stored untila new adjustment is made.

To save battery power, the test tool is in economicbrightness mode when operated on the battery. The highbrightness intensity increases when you connect thepower adapter.

NoteUsing dimmed light lengthens maximum batterypower operation by about one hour.

Page 80: 0MASTER - RS Components

Fluke 192/196/199Users Manual

72

Changing Date and TimeThe test tool has a date and time clock. For example, tochange the date to 19 April, 1999, do the following:

1 Display the USER key labels.

2 Open the User Options menu.

3 Open Date Adjust menu.

4 Choose 1999, jump to Month.

5 Choose 04, jump to Day.

6 Choose 19, jump to Format.

7 Choose DD/MM/YY, accept thenew date.

You can change the time in a similar way by opening theTime Adjust menu (steps 2 and 3.)

Page 81: 0MASTER - RS Components

TipsSaving Battery Life 7

73

Saving Battery LifeWhen operated on the battery (no battery chargerconnected), the test tool conserves power by shuttingitself down. If you have not pressed a key for at least 30minutes, the test tool turns itself off automatically.

Note

If the power adapter is connected, there is noautomatic power shutdown.

Automatic power shutdown will not occur if TrendPlot orScope Record is on, but the backlight will dim. Recordingwill continue even if the battery is low, and retention ofmemories is not jeopardized.

Setting the Power Down Timer

Initially the power shutdown time is 30 minutes. You canset the power shutdown time to 5 minutes as following:

1 Display the USER key labels.

2 Open the User Options menu.

3 Open the Battery Save Optionsmenu.

4 Select 5 Minutes.

Page 82: 0MASTER - RS Components

Fluke 192/196/199Users Manual

74

Changing the Auto Set OptionsWith the next procedure you can choose how auto setbehaves when you press the AUTO (auto set) key.

1 Display the USER key labels.

2 Open the User Options menu.

3 Open the Auto Set Adjustmenu.

If the frequency range is set to > 15 Hz, the Connect-and-View function responds more quickly. The response isquicker because the test tool is instructed not to analyzelow frequency signal components. However, when youmeasure frequencies lower than 15 Hz, the test tool mustbe instructed to analyze low frequency components forautomatic triggering:

4 Select Signal > 1 Hz, then jumpto Coupling.

With the coupling option you can choose how auto setbehaves. When you press the AUTO (auto set) key, thecoupling can either be set to dc or left unchanged:

5 Select Unchanged.

Note

The auto set option for the signal frequency issimilar to the automatic trigger option for thesignal frequency. (See Chapter 5: “AutomaticTrigger Options”). However, the auto set optiondetermines the behavior of the auto set functionand shows only effect when you press the autoset key.

Page 83: 0MASTER - RS Components

75

Chapter 8Maintaining the Test Tool

About this Chapter

This chapter covers basic maintenance procedures thatcan be performed by the user. For complete service,disassembly, repair, and calibration information, see theService Manual. You will find the part number of theService Manual in the section “Parts and Accessories” inthis chapter.

Cleaning the Test ToolClean the test tool with a damp cloth and a mild soap. Donot use abrasives, solvents, or alcohol. These maydamage the text on the test tool.

Storing the Test ToolIf you are storing the test tool for an extended period oftime, charge the NiMH (Nickel-Metal Hydride) batteriesbefore storing.

Page 84: 0MASTER - RS Components

Fluke 192/196/199Users Manual

76

Charging the BatteriesAt delivery, the NiMH batteries may be empty and mustbe charged for 4 hours (with the test tool turned off) toreach full charge. When fully charged, the batteriesprovide 4 hours of use.

When battery power is used, the battery indicator at thetop of the screen informs you about the condition of thebatteries. The battery symbols are: . Thesymbol indicates that there are typically five minutesof operating time left.

To charge the batteries and power the instrument,connect the battery charger as shown in Figure 49. Tocharge the batteries more quickly, turn off the test tool.

Caution

To avoid overheating of the batteries duringcharging, do not exceed the allowableambient temperature given in thespecifications.

Note

No damage will occur if the charger is connectedfor long periods, e.g., during the weekend. Theinstrument then automatically switches to tricklecharging.

Figure 49. Charging the Batteries

Page 85: 0MASTER - RS Components

Maintaining the Test ToolExtending Battery Operation Time 8

77

Extending Battery Operation TimeTypically, NiMH batteries always meet the specifiedoperating time. However, if the batteries have beenextremely discharged (for example, when empty batterieswere stored for a long period) it is possible that thebattery condition has deteriorated.

To keep the batteries in optimal condition, observe thefollowing guidelines:

• Operate the test tool on batteries until the symbol appears at the bottom of the screen. Thisindicates that the battery level is low and that theNiMH batteries need to be recharged.

• To obtain optimal battery condition again, you canrefresh the batteries. During a battery refresh, thebatteries will be fully discharged and charged again.A complete refresh cycle takes about 12 hours andshould be done about four times a year. You cancheck the latest battery refresh date. See section“Displaying Calibration Information”.

To refresh the battery, make sure that the test tool is linepowered and proceed as follows:

1 Display the USER key labels.

2 Open the User Options menu.

A message appears asking whether you want to start therefresh cycle now.

3 Start the refresh cycle.

Do not disconnect the battery charger during therefresh cycle. This will interrupt the refresh cycle.

Note

After starting the refresh cycle, the screen will beblack.

Page 86: 0MASTER - RS Components

Fluke 192/196/199Users Manual

78

Replacing the NiMH Battery Pack BP190Usually it should not be necessary to replace the batterypack. However, if replacement is needed, this should bedone by qualified personnel only. Contact your nearestFluke center for more information.

Calibrating the Voltage ProbesTo meet full user specifications, you need to adjust thered and gray voltage probes for optimal response. Thecalibration consists of a high frequency adjustment and adc calibration for 10:1 probes. The dc calibration is notpossible for 100:1 probes.

This example shows how to calibrate the 10:1 voltageprobes:

1 Display the input A key labels.

2 Open the Probe on A menu.

3 Select Voltage, then jump toAttenuation.

Figure 50. Adjusting Voltage Probes

If the 10:1 option is already selected, proceed with step 5.

4 Select 10:1, then return.

Page 87: 0MASTER - RS Components

Maintaining the Test ToolCalibrating the Voltage Probes 8

79

Repeat steps 2 and 3 and proceed as follows:

5 Select Probe Cal with the arrowkeys, then accept.

A message appears asking you whether to start the 10:1probe calibration.

6 Start the probe calibration.

A message appears telling you how to connect the probe.Connect the red 10:1 voltage probe from the red input Ajack to the red banana jack. Connect the reference leadto the black banana jack. (See Figure 50.)

7 Adjust the trimmer screw in theprobe housing until a pure squarewave is displayed.

8 Continue with DC calibration.Automatic DC calibration is onlypossible for 10:1 voltage probes.

The test tool automatically calibrates itself to the probe.During calibration you should not touch the probe. Amessage indicates when the DC calibration hascompleted successfully.

9 Return.

Repeat the procedure for the gray 10:1 voltage probe.Connect the grey 10:1 voltage probe from the grey inputB jack to the red banana jack. Connect the reference leadto the black banana jack.

Note

When using 100:1 voltage probes, choose 100:1attenuation to perform a HF adjustment.Automatic dc calibration is not possible with thisprobe type.

Page 88: 0MASTER - RS Components

Fluke 192/196/199Users Manual

80

Displaying Calibration InformationYou can display version number and calibration date:

1 Display the USER key labels.

2 Open the Version & Calibrationmenu.

The screen gives you information about the modelnumber with software version, the calibration number withlatest calibration date, and the latest battery refresh date.

3 Return.

Recalibration must be carried out by qualified personnel.Contact your local Fluke representative for recalibration.

Parts and Accessories

Standard Accessories

The following tables list the user-replaceable parts for thevarious test tool models. For additional optionalaccessories, see the ScopeMeter Accessories booklet.

To order replacement parts or additional accessories,contact your nearest service center.

Page 89: 0MASTER - RS Components

Maintaining the Test ToolParts and Accessories 8

81

Table 1. Standard Accessories

Item Ordering Code

Battery Charger, available models:Universal Europe 230 V, 50 and 60 HzNorth America 120 V, 50 and 60 HzUnited Kingdom 240 V, 50 and 60 HzJapan 100 V, 50 and 60 HzAustralia 240 V, 50 and 60 HzUniversal 115 V/230 V, 50 and 60 Hz ** The 230V rating of the BC190/808 is not for use in North America.

A line plug adapter complying with the applicable NationalRequirements may be provided to alter the blade configurations fora specific country. The universal adapter is standard equipped witha North American line cord.

BC190/801BC190/803BC190/804BC190/806BC190/807BC190/808

Voltage Probe Set (Red), designed for use with the Fluke ScopeMeter190 series test tool.

The set includes the following items (not available separately):• 10:1 Voltage Probe (red)• 4-mm Test Probe for Probe Tip (red)• Hook Clip for Probe Tip (red)• Ground Lead with Hook Clip (red)• Ground Lead with Mini Alligator Clip (black)• Ground Spring for Probe Tip (black)

VPS200-R

Page 90: 0MASTER - RS Components

Fluke 192/196/199Users Manual

82

Item Ordering Code

Voltage Probe Set (Gray), designed for use with the Fluke ScopeMeter190 series test tool.

The set includes the following items (not available separately):• 10:1 Voltage Probe (gray)• 4-mm Test Probe for Probe Tip (gray)• Hook Clip for Probe Tip (gray)• Ground Lead with Hook Clip (gray)• Ground Lead with Mini Alligator Clip (black)

VPS200-G

Flexible Test Leads (red and black) TL24 (General Purpose Leads)

Accessory Set (Red)

The set includes the following items (not available separately):• Industrial Alligator for Probe Tip (red)• 2-mm Test Probe for Probe Tip (red)• Industrial Alligator for Banana Jack (red)• 2-mm Test Probe for Banana Jack (red)• Ground Lead with 4-mm Banana Jack (black)

AS200-R

Accessory Set (Gray)

The set includes the following items (not available separately):• Industrial Alligator for Probe Tip (gray)• 2-mm Test Probe for Probe Tip (gray)• Industrial Alligator for Banana Jack (gray)• 2-mm Test Probe for Banana Jack (gray)• Ground Lead with 4-mm Banana Jack (black)

AS200-G

Page 91: 0MASTER - RS Components

Maintaining the Test ToolParts and Accessories 8

83

Item Ordering Code

Replacement Set for Voltage Probe

The set includes the following items (not available separately):• 2x 4-mm Test Probe for Probe Tip (red and gray)• 3x Hook Clip for Probe Tip (2 red, 1 gray)• 2x Ground Lead with Hook Clip (red and gray)• 2x Ground Lead with Mini Alligator Clip (black)• 5x Ground Spring for Probe Tip (black)

RS200

Table 2. Users Manuals

Item Ordering Code

Users Manual (English)

Users Manual (German)

Users Manual (French)

Users Manual (Spanish)

Users Manual (Portuguese)

Users Manual (Italian)

Users Manual (Chinese)

Users Manual (Japanese)

Users Manual (Korean)

4822 872 00983

4822 872 00984

4822 872 00985

4822 872 00986

4822 872 00987

4822 872 00988

4822 872 00989

4822 872 00991

4822 872 00992

Page 92: 0MASTER - RS Components

Fluke 192/196/199Users Manual

84

Optional Accessories

Item Ordering Code

Software & Cable Carrying Case Kit

Set contains the following parts:

Optically Isolated RS-232 Adapter/Cable

Hard Carrying Case

FlukeView® ScopeMeter® Software for Windows 95® , 98® , Me® , 2000® and NT4®

SCC190

PM9080

C190

SW90W

Optically Isolated RS-232 Adapter/Cable PM9080

Hard Case C190

Soft Case C195

Current Shunt 4-20 mA CS20MA

Print Adapter Cable for Parallel Printers PAC91

Optional Service Manual

Item Ordering Code

Service Manual (English) 4822 872 05376

Page 93: 0MASTER - RS Components

Maintaining the Test ToolTroubleshooting 8

85

Troubleshooting

The Test Tool Does Not Start Up

• The batteries may be completely empty. In this casethe test tool will not start up, even if it is powered bythe battery charger. Charge the batteries first: powerthe test tool with the battery charger without turning iton. Wait about 15 minutes and try turning on the testtool again.

The Test Tool Shuts Down After A Few Seconds

• The batteries may be empty. Check the batterysymbol at the top right of the screen. A symbolindicates that the batteries are empty and must becharged.

The Screen Remains Black

• Make sure that the test tool is on.

• You might have a problem with the screen contrast.Press , then press . Now you can use thearrow keys to adjust the contrast.

The Operation Time Of Fully Charged BatteriesIs Too Short

• The batteries may be in poor condition. Refresh thebatteries to optimize the condition of the batteriesagain. It is advised to refresh the batteries about fourtimes a year.

The Printer Does Not Print

• Make sure that the interface cable is properlyconnected between the test tool and the printer.

• Make sure that you have selected the correct printertype. (See Chapter 6.)

• Make sure that the baud rate matches with theprinter. If not, select another baud rate.(See Chapter 6.)

• If you are using the PAC91 (Print Adapter Cable),make sure that it is turned on.

Page 94: 0MASTER - RS Components

Fluke 192/196/199Users Manual

86

FlukeView Does Not Recognize The Test Tool

• Make sure that the test tool is turned on.

• Make sure that the interface cable is properlyconnected between the test tool and the PC.

• Make sure that the correct COM port has beenselected in FlukeView. If not, change the COM portsetting or connect the interface cable to anotherCOM port.

Battery Operated Fluke Accessories Do NotFunction

• When using battery operated Fluke accessories,always first check the battery condition of theaccessory with a Fluke multimeter.

Page 95: 0MASTER - RS Components

87

Chapter 9Specifications

IntroductionPerformance Characteristics

FLUKE guarantees the properties expressed in numericalvalues with the stated tolerance. Specified non-tolerancenumerical values indicate those that could be nominallyexpected from the mean of a range of identicalScopeMeter test tools.

Environmental Data

The environmental data mentioned in this manual arebased on the results of the manufacturer’s verificationprocedures.

Safety Characteristics

The test tool has been designed and tested inaccordance with Standards ANSI/ISA S82.01-1994, EN61010.1 (1993) (IEC 1010-1), CAN/CSA-C22.2No.1010.1-92 (including approval), UL3111-1 (includingapproval) Safety Requirements for Electrical Equipmentfor Measurement, Control, and Laboratory Use.

This manual contains information and warnings that mustbe followed by the user to ensure safe operation and tokeep the instrument in a safe condition. Use of thisequipment in a manner not specified by the manufacturermay impair protection provided by the equipment.

Page 96: 0MASTER - RS Components

Fluke 192/196/199Users Manual

88

Dual Input Oscilloscope

Isolated Inputs A and B (Vertical)

Bandwidth, DC CoupledFLUKE 199 ....................................... 200 MHz (-3 dB)FLUKE 196 ....................................... 100 MHz (-3 dB)FLUKE 192 ......................................... 60 MHz (-3 dB)

Lower Frequency Limit, AC Coupledwith 10:1 probe ...................................... <2 Hz (-3 dB)direct (1:1) ............................................. <5 Hz (-3 dB)

Rise TimeFLUKE 199 ....................................................... 1.7 nsFLUKE 196 ....................................................... 3.5 nsFLUKE 192 ....................................................... 5.8 ns

Analog Bandwidth Limiters.............. 20 MHz and 10 kHz

Input Coupling ................................................... AC, DC

Polarity ................................................ Normal, Inverted

Sensitivity Rangeswith 10:1 probe ............................50 mV to 1000 V/divdirect (1:1) ...................................... 5 mV to 100 V/div

Trace Positioning Range ..............................±4 divisions

Input Impedance on BNCDC Coupled ..................... 1 MΩ (±1 %)//15 pF (±2 pF)

Max. Input Voltagewith 10:1 probe ......................................600 V CAT III

1000 V CAT IIdirect (1:1) .............................................300 V CAT III(For detailed specifications, see “Safety”)

Vertical Accuracy ....................±(1.5 % + 0.04 range/div)

Digitizer Resolution ................... 8 bits, separate digitizerfor each input

Horizontal

Maximum Time Base Speed:FLUKE 199 .....................................................5 ns/divFLUKE 196 .....................................................5 ns/divFLUKE 192 ...................................................10 ns/div

Minimum Time Base Speed (Scope Record) ....2 min/div

Real Time Sampling Rate (for both inputs simultaneously)FLUKE199:

5 ns to 2 µs /div ................................. up to 2.5 GS/s5 µs to 120 s/div .......................................... 20 MS/s

FLUKE 196:5 ns to 2 µs /div .................................... up to 1 GS/s5 µs to 120 s/div .......................................... 20 MS/s

FLUKE 19210 ns to 2 µs /div .............................. up to 500 MS/s5 µs to 120 s/div .......................................... 20 MS/s

Page 97: 0MASTER - RS Components

SpecificationsDual Input Oscilloscope 9

89

Record LengthScope Record Mode.......... 27500 points on each inputScope Normal Mode............ 1000 points on each inputScope Glitch Capture Mode...500 points on each input

Glitch Detection5 µs to 120 s/div....... displays glitches as fast as 50 ns

Waveform Display...............A, B, A+B, A-B, A*B, A vs BNormal, Average (2,4,8,64 x), Persistence

Time Base Accuracy ...................................... ±100 ppm

Trigger and Delay

Trigger Modes..................................... Automatic, Edge,External, Video, Pulse Width

Trigger Delay ............................... up to +1000 divisions

Pre Trigger View ...........................one full screen length

Max. Delay...................................................10 seconds

Automatic Connect-and-View Trigger

Source ........................................................... A, B, EXT

Slope ................................................. Positive, Negative

Edge Trigger

Screen Update.......... Free Run, On Trigger, Single Shot

Source ........................................................... A, B, EXT

Slope ................................................. Positive, Negative

Trigger Level Control Range.........................±4 divisions

Trigger Sensitivity A and BDC to 5 MHz at >5 mV/div ........................0.5 divisionsDC to 5 MHz at 5 mV/div .............................. 1 division200 MHz (FLUKE 199) ................................. 1 division250 MHz (FLUKE 199) ............................... 2 divisions100 MHz (FLUKE 196) ................................. 1 division150 MHz (FLUKE 196) ............................... 2 divisions60 MHz (FLUKE 192) ................................... 1 division100 MHz (FLUKE 192) ............................... 2 divisions

Isolated External Trigger

Bandwidth ...........................................................10 kHz

Modes.................................................. Automatic, Edge

Trigger Levels (DC to 10 kHz) ................. 120 mV, 1.2 V

Video Trigger

Standards ........................... PAL, PAL+, NTSC, SECAM

Modes.....................Lines, Line Select, Field 1 or Field 2

Source ........................................................................ A

Polarity...............................................Positive, Negative

Sensitivity..................................... 0.7 division sync level

Page 98: 0MASTER - RS Components

Fluke 192/196/199Users Manual

90

Pulse Width Trigger

Screen Update...........................On Trigger, Single Shot

Trigger Conditions.......... <T, >T, ≈T (±10 %), ≠T(±10 %)

Source.........................................................................A

Polarity ...................................Positive or negative pulse

Pulse Time Adjustment Range .......1/100 div. to 250 div.with a maximum resolution of 50 ns.

Continuous Auto Set

Autoranging attenuators and time base, automaticConnect-and-View™ triggering with automatic sourceselection.

ModesNormal..................................15 Hz to max. bandwidthLow Frequency .......................1 Hz to max. bandwidth

Minimum Amplitude A and BDC to 1 MHz .....................................................10 mV1 MHz to max. bandwidth..................................20 mV

Automatic Capturing Scope Screens

Capacity .......................... 100 dual input scope Screens

For viewing screens, see Replay function.

Automatic Scope MeasurementsThe accuracy of all readings is within ± (% of reading +number of counts) from 18 °C to 28 °C. Add 0.1x (specificaccuracy) for each °C below 18 °C or above 28 °C. Forvoltage measurements with 10:1 probe, add probeaccuracy unless the probe has been calibrated on thetest tool. At least 1.5 waveform period must be visible onthe screen.

GeneralInputs ..................................................................A and B

DC Common Mode Rejection (CMRR) ................ >100 dBAC Common Mode Rejection at 50, 60, or 400 Hz ......>60 dB

DC Voltage (VDC)

Maximum Voltagewith 10:1 probe .................................................. 1000 Vdirect (1:1) ........................................................... 300 V

Maximum Resolutionwith 10:1 probe .....................................................1 mVdirect (1:1) ......................................................... 100 µV

Full Scale Reading ........................................ 1100 counts

Accuracy at 5 s to 5 µs/div ..................±(1.5 % +5 counts)

Normal Mode AC Rejection at 50 or 60 Hz ........... >60 dB

Page 99: 0MASTER - RS Components

SpecificationsAutomatic Scope Measurements 9

91

AC Voltage (VAC)

Maximum Voltagewith 10:1 probe................................................ 1000 Vdirect (1:1)......................................................... 300 V

Maximum Resolutionwith 10:1 probe................................................... 1 mVdirect (1:1)....................................................... 100 µV

Full Scale Reading ..................................... 1100 counts

AccuracyDC coupled:

DC to 60 Hz..............................±(1.5 % +10 counts)

AC coupled, low frequencies:50 Hz direct (1:1) .....................±(2.1 % + 10 counts)60 Hz direct (1:1) .....................±(1.9 % + 10 counts)With the 10:1 probe the low frequency roll off point willbe lowered to 2 Hz, which improves the AC accuracyfor low frequencies. When possible use DC couplingfor maximum accuracy.

AC or DC coupled, high frequencies:60 Hz to 20 kHz.......................±(2.5 % + 15 counts)20 kHz to 1 MHz.........................±(5 % + 20 counts)1 MHz to 25 MHz......................±(10 % + 20 counts)For higher frequencies the instrument’s frequency rolloff starts affecting accuracy.Normal Mode DC Rejection........................... >50 dB

All accuracies are valid if:• The waveform amplitude is larger than one division• At least 1.5 waveform period is on the screen

AC+DC Voltage (True RMS)

Maximum Voltagewith 10:1 probe................................................ 1000 Vdirect (1:1)......................................................... 300 V

Maximum Resolutionwith 10:1 probe...................................................1 mVdirect (1:1)....................................................... 100 µV

Full Scale Reading ..................................... 1100 counts

AccuracyDC to 60 Hz ...............................±(1.5 % + 10 counts)60 Hz to 20 kHz..........................±(2.5 % + 15 counts)20 kHz to 1 MHz............................±(5 % + 20 counts)1 MHz to 25 MHz ........................±(10 % + 20 counts)For higher frequencies the instrument’s frequency rolloff starts affecting accuracy.

Page 100: 0MASTER - RS Components

Fluke 192/196/199Users Manual

92

Amperes (AMP)

With Optional Current Probe or Current Shunt

Ranges........................... same as VDC, VAC, VAC+DC

Probe Sensitivity ................100 µV/A, 1 mV/A, 10 mV/A,100 mV/A, 1 V/A, 10 V/A, and 100 V/A

Accuracy ........................ same as VDC, VAC, VAC+DC(add current probe or current shunt accuracy)

Peak

Modes .........................Max peak, Min peak, or pk-to-pk

Maximum Voltagewith 10:1 probe ............................................... 1000 Vdirect (1:1) ........................................................ 300 V

Maximum Resolutionwith 10:1 probe ................................................ 10 mVdirect (1:1) ......................................................... 1 mV

Full Scale Reading....................................... 800 counts

AccuracyMax peak or Min peak............................. ±0.2 divisionPeak-to-peak .......................................... ±0.4 division

Frequency (Hz)Range....................................1.000 Hz to full bandwidth

Full Scale Reading .................................... 9 999 countswith at least 10 waveform periods on screen.

Accuracy1 Hz to full bandwidth .................... ±(0.5 % +2 counts)

Duty Cycle (DUTY)

Range...................................................4.0 % to 98.0 %

Pulse Width (PULSE)

Resolution (with GLITCH off)....................... 1/100 division

Full Scale Reading ....................................... 999 counts

Accuracy1 Hz to full bandwidth .................... ±(0.5 % +2 counts)

Page 101: 0MASTER - RS Components

SpecificationsAutomatic Scope Measurements 9

93

Power

Power Factor ..................... ratio between Watts and VARange ......................................................0.00 to 1.00

Watt................................. RMS reading of multiplicationcorresponding samples of input A (volts)

and Input B (amperes)Full Scale Reading .................................... 999 counts

VA ............................................................ Vrms x ArmsFull Scale Reading .................................... 999 counts

VA Reactive ................................................ √((VA)2-W2)Full Scale Reading .................................... 999 counts

Phase

Range.......................................... -180 to +180 degrees

Resolution....................................................... 1 degree

Accuracy0.1 Hz to 1 MHz ........................................±1 degrees1 MHz to 10 MHz ......................................±3 degrees

Temperature (TEMP)

With Optional Temperature Probe

Ranges (°C or °F) ............................... -40.0 to +100.0 °-100 to +250 °-100 to +500 °

-100 to +1000 °-100 to + 2500 °

Probe Sensitivity ..........................1 mV/°C and 1 mV/°F

Decibel (dB)

dBV..............................................dB relative to one volt

dBm................. dB relative to one mW in 50 Ω or 600 Ω

dB on ........................................VDC, VAC, or VAC+DC

Accuracy.........................same as VDC, VAC, VAC+DC

Page 102: 0MASTER - RS Components

Fluke 192/196/199Users Manual

94

Meter

Meter Input

Input Coupling ..........................................................DC

Frequency Response.....................DC to 10 kHz (-3 dB)

Input Impedance..............1 MΩ (±1 %)//10 pF (±1.5 pF)

Max. Input Voltage ............................ 1000 V CAT II600 V CAT III

(For detailed specifications, see “Safety”)

Meter Functions

Ranging.................................................... Auto, Manual

Modes ................................................. Normal, Relative

DMM Measurements on Meter InputsThe accuracy of all measurements is within ± (% ofreading + number of counts) from 18 °C to 28 °C.Add 0.1x (specific accuracy) for each °C below 18 °C orabove 28 °C.

General

DC Common Mode Rejection (CMRR) ................ >100 dBAC Common Mode Rejection at 50, 60, or 400 Hz ......>60 dB

Ohms (Ω)

Ranges.............................. 500.0 Ω, 5.000 kΩ, 50.00 kΩ,500.0 kΩ, 5.000 MΩ, 30.00 MΩ

Full Scale Reading500 Ω to 5 MΩ ........................................... 5000 counts30 MΩ........................................................ 3000 counts

Accuracy.............................................±(0.6 % +5 counts)

Measurement Current ................. 0.5 mA to 50 nA, ±20 %decreases with increasing ranges

Open Circuit Voltage.................................................<4 V

Continuity (CONT)

Beep.......................................................<50 Ω (±30 Ω)

Measurement Current ............................ 0.5 mA, ±20 %

Detection of shorts of ...........................................≥1 ms

Page 103: 0MASTER - RS Components

SpecificationsDMM Measurements on Meter Inputs 9

95

Diode

Maximum Voltage Reading.................................... 2.8 V

Open Circuit Voltage ..............................................<4 V

Accuracy.............................................±(2 % +5 counts)

Measurement Current ............................ 0.5 mA, ±20 %

Temperature (TEMP)

With Optional Temperature Probe

Ranges (°C or °F) ............................... -40.0 to +100.0 °-100.0 to +250.0 °-100.0 to +500.0 °

-100 to +1000 °-100 to + 2500 °

Probe Sensitivity ..........................1 mV/°C and 1 mV/°F

DC Voltage (VDC)

Ranges ....500.0 mV, 5.000 V, 50.00 V, 500.0 V, 1100 V

Full Scale Reading ..................................... 5000 counts

Accuracy..........................................±(0.5 % +5 counts)

Normal Mode AC Rejection at 50 or 60 Hz ±1 % >60 dB

AC Voltage (VAC)Ranges ....500.0 mV, 5.000 V, 50.00 V, 500.0 V, 1100 V

Full Scale Reading ..................................... 5000 counts

Accuracy15 Hz to 60 Hz ...............................±(1 % +10 counts)60 Hz to 1 kHz.............................±(2.5 % +15 counts)For higher frequencies the frequency roll off of theMeter input starts affecting accuracy.

Normal Mode DC Rejection ................................ >50 dB

AC+DC Voltage (True RMS)Ranges ....500.0 mV, 5.000 V, 50.00 V, 500.0 V, 1100 V

Full Scale Reading ..................................... 5000 counts

AccuracyDC to 60 Hz ...................................±(1 % +10 counts)60 Hz to 1 kHz.............................±(2.5 % +15 counts)For higher frequencies the frequency roll off of theMeter input starts affecting accuracy.

All accuracies are valid if the waveform amplitude islarger than 5 % of full scale.

Page 104: 0MASTER - RS Components

Fluke 192/196/199Users Manual

96

Amperes (AMP)

With Optional Current Probe or Current Shunt

Ranges........................... same as VDC, VAC, VAC+DC

Probe Sensitivity ................100 µV/A, 1 mV/A, 10 mV/A,100 mV/A, 1 V/A, 10 V/A, and 100 V/A

Accuracy ........................ same as VDC, VAC, VAC+DC(add current probe or current shunt accuracy)

Recorder

TrendPlot (Meter or Scope)

Chart recorder that plots a graph of min and max valuesof Meter or Scope measurements over time.

Measurement Speed...................> 2.5 measurements/s

Time/Div .......................................10 s/div to 20 min/div

Record Size................................ 13500 points per input

Recorded Time Span ...........................90 min to 8 days

Time Reference ................... time from start, time of day

Scope Record

Records scope waveforms in deep memory whiledisplaying the waveform in Roll mode.

Source...................................................Input A, Input B

Max. Sample Speed (10 ms/div to 1 min/div)..... 20 MS/s

Glitch capture (10 ms/div to 1 min/div) .................. 50 ns

Time/Div in normal mode ............. 10 ms/div to 2 min/div

Record Size .................................27500 points per input

Recorded Time Span ............................11 s to 30 hours

Acquisition Modes.....................................Single SweepContinuous Roll

External Triggering

Time Reference ................... time from start, time of day

Page 105: 0MASTER - RS Components

SpecificationsZoom, Replay and Cursors 9

97

Zoom, Replay and Cursors

Zoom

Horizontal MagnificationScope Record ........................................... up to 100xTrendPlot .....................................................up to 50xScope ............................................................up to 8x

Replay

Displays a maximum of 100 captured dual input Scopescreens.

Replay modes..........Step by Step, Replay as Animation

Cursor Measurements

Cursor Modes ................................single vertical cursordual vertical cursors

dual horizontal cursors (Scope mode)

Markers .................... automatic markers at cross points

Measurements ..................................... value at cursor 1value at cursor 2

difference between values at cursor 1 and 2time between cursors

Time of Day (Recorder modes)Time from Start (Recorder modes)

Rise Time

Miscellaneous

Display

View Area ......................................132 mm (5.2 inches)

Backlight ................... Cold Cathode Fluorescent (CCFL)Temperature compensated

Brightness............................. Power Adapter: 60 cd / m2

Batteries: 35 cd / m2

PowerRechargeable NiMH Batteries:

Operating Time ...............................................4 hoursCharging Time.................................................4 hours

Allowable ambienttemperature during charging:.0 to 40 °C (32 to 104 °F)

Auto power downtime (battery saving):............5 min, 30 min or disabled

Battery Charger / Power Adapter BC190:• BC190/801 European line plug 230 V ±10 %• BC190/803 North American line plug 120 V ±10 %• BC190/804 United Kingdom line plug 230 V ±10 %• BC190/806 Japanese line plug 100 V ±10 %• BC190/807 Australian line plug 230 V ±10 %• BC190/808 Universal switchable adapter 115 V ±10 %

or 230 V ±10 %, with plug EN60320-2.2G

Line Frequency ........................................ 50 and 60 Hz

Page 106: 0MASTER - RS Components

Fluke 192/196/199Users Manual

98

Probe Calibration

Manual pulse adjustment and automatic DC adjustmentwith probe check.

Generator Output....................................3 Vpp / 500 Hzsquare wave

Memory

Number of Scope Memories....................................... 10Each memory can contain two waveforms pluscorresponding setups

Number of Recorder Memories ....................................2Each memory can contain:• a dual input TrendPlot

(2 x 13500 points per input)• a dual input Scope Record

(2 x 27500 points per input)• 100 dual input Scope screens

Mechanical

Size ..................... 64 x 169 x 254 mm (2.5 x 6.6 x 10 in)

Weight ..................................................1.95 kg (4.3 lbs)including battery

Optical InterfacePort

Type .......................................RS-232, optically isolated

To Printer........................... supports Epson FX, LQ, andHP Deskjet®, Laserjet®, and Postscript

• Serial via PM9080 (optically isolated RS-232 Adapter/Cable, optional).

• Parallel via PAC91 (optically isolated Print AdapterCable, optional).

To PC/Notebook• Serial via PM9080 (optically isolated RS-232 Adapter/

Cable, optional), using SW90W (FlukeView® softwarefor Windows 95®, 98®, Me®, 2000® and NT4®).

Page 107: 0MASTER - RS Components

SpecificationsEnvironmental 9

99

EnvironmentalEnvironmental....................... MIL-PRF-28800F, Class 2

TemperatureOperating:

battery only.........................0 to 50 °C (32 to 122 °F)power adapter ....................0 to 40 °C (32 to 104 °F)

Storage .......................... -20 to +60 °C (-4 to +140 °F)

HumidityOperating:

0 to 10 °C (32 to 50 °F)..................... noncondensing10 to 30 °C (50 to 86 °F).................................. 95 %30 to 40 °C (86 to 104 °F) .................................75 %40 to 50 °C (104 to 122 °F) ...............................45 %

Storage:-20 to +60 °C (-4 to +140 °F) ............ noncondensing

AltitudeOperating ....................................... 3 km (10 000 feet)Storage ........................................ 12 km (40 000 feet)

Vibration (Sinusoidal) ....................................... max. 3 g

Shock ............................................................ max. 30 g

Electromagnetic Compatibility (EMC)Emission and immunitiy ........... EN-IEC61326-1 (1997)

Enclosure Protection ........................... IP51, ref: IEC529

Page 108: 0MASTER - RS Components

Fluke 192/196/199Users Manual

100

SafetyDesigned for measurements on 1000 V Category IIInstallations, 600 V Category III Installations, PollutionDegree 2, per:• ANSI/ISA S82.01-1994• EN61010-1 (1993) (IEC1010-1)• CAN/CSA-C22.2 No.1010.1-92• UL3111-1

Max. Input Voltages

Input A and B directly ............................. 300 V CAT III

Input A and B via 10:1 probe ................. 1000 V CAT II600 V CAT III

METER/EXT TRIG inputs...................... 1000 V CAT II600 V CAT III

Max. Floating Voltage

From any terminal to ground ................. 1000 V CAT II600 V CAT III

Between any terminal............................ 1000 V CAT II600 V CAT III

Voltage ratings are given as “working voltage”. Theyshould be read as Vac-rms (50-60 Hz) for ACsinewave applications and as Vdc for DCapplications.

Figure 51. Max. Input Voltage v.s. Frequency

Note

Overvoltage Category III refers to distributionlevel and fixed installation circuits inside abuilding. Overvoltage Category II refers to locallevel, which is applicable for appliances andportable equipment.

Page 109: 0MASTER - RS Components

Specifications10:1 Probe 9

101

Figure 52. Safe Handling: Max. Input Voltage BetweenScope References, and Between

Scope References and Meter Reference

10:1 Probe

Safety

Max. Input Voltage ..........................1000 V CAT II600 V CAT III

Max. Floating Voltagefrom any terminal to ground...................1000 V CAT II

600 V CAT IIIup to 400 Hz

Electrical specifications

Input Impedance at probe tip10 MΩ (±2 %)//14 pF (±2pF)

Capacity Adjustment Range .........................10 to 22 pF

Attenuation at DC (1 MΩ input) ....................10 x (±2 %)

Bandwidth (with FLUKE 199)...... DC to 200 MHz (-3 dB)

EnvironmentalTemperature

Operating ...........................0 to 50 °C (32 to 122 °F)Storage........................ -20 to +60 °C (-4 to +140 °F)

AltitudeOperating .................................... 3 km (10 000 feet)Storage...................................... 12 km (40 000 feet)

HumidityOperating at 10 to 30 °C (50 to 86 °F).............. 95 %

Page 110: 0MASTER - RS Components

Fluke 192/196/199Users Manual

102

0.011

2

5

10

20

50

100

200

500

1000

0.02 0.05 0.1 0.2 0.5 1 2 5 10 20 50 100 200

FRE QUE NCY (MHz)

MAX. INPUTVOLT AGE (Vrms)

CAT III

CAT II

Figure 53. Max. Voltage From Probe Tip to Groundand From Probe Tip to Probe Reference

0.011

2

5

10

20

50

100

200

500

1000

0.02 0.05 0.1 0.2 0.5 1 2 5 10 20 50 100 200

FRE QUE NCY (MHz)

MAX. VOLT AGEF ROM PROBE RE FE RE NCET O GROUND

CAT II I

CAT II

Figure 54. Safe Handling: Max. Voltage From ProbeReference to Ground

Page 111: 0MASTER - RS Components

SpecificationsElectromagnetic Immunity 9

103

Electromagnetic ImmunityThe Fluke 190 series, including standard accessories, conforms with the EEC directive 89/336 for EMC immunity, asdefined by EN-61326-1, with the addition of the following tables.

Scope Mode (10 ms/div): Trace disturbance with VPS200 voltage probe shorted

Table 1

No visible disturbance E = 3V/mFrequency range 10 kHz to 20 MHz 5 mV/div to 100 V/divFrequency range 20 MHz to 100 MHz 100 mV/div to 100 V/divFrequency range 100 MHz to 1 GHz 500 mV/div to 100 V/div *)

(*) With the 20 MHz Bandwidth Filter switched on: no visible disturbance.With the 20 MHz Bandwidth Filter switched off: disturbance is max 2 div.

Table 2

Disturbance less than 10% of full scale E = 3V/mFrequency range 20 MHz to 100 MHz 10 mV/div to 50 mV/div

Test Tool ranges not specified in tables 1 and 2 may have a disturbance of more than 10% of full scale.

Meter Mode (Vdc, Vac, Vac+dc, Ohm and Continuity): Reading disturbance with test leads shorted

Table 3

Disturbance less than 1% of full scale E = 3V/mFrequency range 10 kHz to 1 GHz 500mV to 1000V , 500Ohm to 30 MOhm ranges

Page 112: 0MASTER - RS Components

Fluke 192/196/199Users Manual

104

Page 113: 0MASTER - RS Components

105

Index

—Numerics—10-1 Voltage Probe, 812-mm Test Probes, 3, 824-mm Test Probes, 3, 81

—A—Accessories, 65, 80AC-Coupling, 18Acquiring the Waveform, 18Altitude, 99, 101Ampere Measurement, 23Amperes, 92, 96Analysis Functions, 35, 97AS200 Accessory Set, 3, 82Auto Ranges, 25Auto Set, 90

Automatic Connect-and-ViewTrigger, 89

Automatic Power Shutdown, 73Automatic Scope Measurements, 13Automatic Triggering, 46Average, 15A+B, 19A-B, 19A*B, 19A versus B, 19

—B—Banana Jack Inputs, 10, 21, 28Bandwidth, 88, 94Bargraph, 22Battery

Charger, 3, 81

Charging, 2, 76Indicator, 76Life, 73Refresh, 77Refresh Date, 80Replacing, 78

BC190 Battery Charger, 3, 81

—C—C190 Hard Case, 3, 84C195 Soft Case, 84Calibrating the Test Tool, 80Calibrating Voltage Probes, 78, 98Calibration, 98Capturing 100 Screens, 37, 90Case, 84Charge Time, 97

Page 114: 0MASTER - RS Components

Fluke 192/196/199Users Manual

106

Charger, 81Charging, 76Cleaning, 75Clear Menu, 10, 70Common Leads, 3Connect-and-View, 43, 90Connecting a Computer, 61Connecting A Printer, 62Connections, 10, 21Continuity, 94Contrast, 71CS20MA Current Shunt, 84Current Measurement, 23Current Probe, 23Current Shunt, 84Cursors Measurements, 40

—D—Date, 72DC Voltage (VDC), 90, 95Decibel (dB), 93Delay, Trigger, 89Deleting Screens, 59Diode, 95Display, 97Displaying Recorded Data, 29, 32DMM Measurements, 22Documenting Screens, 61

Duty Cycle, 92

—E—Earth Ground, 6Edge Trigger, 47, 89Electrical Shock, 5Electrically Floating, 6Electromagnetic Compatibility, 1, 99Electronic Scope Connections, 66Emission, 99Environmental, 99Environmental Data, 87External Trigger, 89External Triggering, 50

—F—FlukeView, 61, 84Freezing the Readings, 25Freezing the Screen, 14Frequency (Hz), 92Frequency Response, 88, 94

—G—Glitch Capture, 17, 32, 33Ground Leads, 3, 81Ground Spring, 3, 81

—H—Hard Case, 3, 84HF Voltage Probe Connection, 66Hook Clips, 3, 81Horizontal Cursors, 40Humidity, 99Hz, 92

—I—Immunity, 99Information Language, 71Input A Measurement, 13Input B Measurement, 13Input Coupling, 94Input Impedance, 88, 94, 101Interface, 98Inverted Display, 18Isolated, 6

—L—Language, 71

—M—Maintenance, 75Manual, 83Manual Ranges, 25Mathematics Functions, 19

Page 115: 0MASTER - RS Components

Index (continued)

107

Max. Floating Voltage, 100, 101Max. Input Voltage, 100Measurement Connections, 10, 21Measurements, 13, 22Measurements on Meter Inputs, 94Mechanical, 98Memory, 98Meter Connections, 67Meter Measurements, 22Multi Meter Measurements, 22

—N—Navigating a Menu, 9NiMH Battery, 75, 76Noisy Waveforms, 19, 48

—O—Ohm (Ω), 94Operating Time, 97Optical Interface, 61, 62, 98Oscilloscope, 88OVERVIEW, 39

—P—PAC91, 62, 84Parallel Print Cable, 84Parallel Printer, 62

Parts, 80Peak, 92Performance Characteristics, 87Persistence, 16Phase, 93PM9080, 61, 62, 84Polarity, 18Power, 97Power Adapter, 73, 81Power Down Timer, 73Powering the Test Tool, 7Pre Trigger, 45Print Cable, 84Probe, 78Probe Calibration, 78, 98Product Registration Card, 3Pulse Trigger, 53Pulse Width, 92Pulse Width Trigger, 90

—R—Readings, 13Recalibrating, 80Recalling Screens, 59Recalling Setups, 60Record Length, 89Record+Setup Memory, 58Recorder, 96

Recorder Options, 30Recording Waveforms, 31Refreshing Batteries, 80Relative Measurements, 26Replaceable Parts, 80Replacement Set, 83Replacing Batteries, 78Replay, 35, 58, 97Resetting the Test Tool, 8, 70Resistance Measurement, 22Reversing the Polarity, 18Rise Time,42, 88RMS Voltage, 91Roll Mode Function, 96RS200 Replacement Set, 83RS-232 Adapter/Cable, 3, 61, 62, 84

—S—Safety, 100Safety Characteristics, 87Safety Requirements, 1Sampling Rate, 88Saving, 58SCC 190, 61, 84Scope, 88Scope Connections, 11, 66Scope Cursor Measurements, 97Scope Measurements, 13

Page 116: 0MASTER - RS Components

Fluke 192/196/199Users Manual

108

Scope Record, 31, 96Screen Contrast, 71Screen Without Menus, 10, 70Serial Printer, 62Service Manual, 84Shock, 99Single Shot, 48Single Sweep Mode, 32Slope, 44, 89Slow Variations, 28Smooth, 15Soft Case, 84Software, 3, 84Software Version, 80

Specifications, 87Spike Capture, 17Stand, 70Storing, 75SW90W Software, 3, 61, 84

—T—Temperature, 93, 95, 99, 101Test Leads, 3Test Probes, 3, 81Tilt Stand, 70Time, 72Time Base Accuracy, 89TL24 Test Leads, 82TrendPlot™ Function, 28TrendPlot (Meter), 96Trigger

Delay, 45, 89Level, 44Modes, 89Pre-trigger, 45

Trigger Sensitivity, 89Triggering

Automatic, 46, 89External, 50On Edges, 47On Pulses, 53On Video, 51

On Waveforms, 43Troubleshooting, 85TV Triggering, 51

—U—Unpacking, 2Users Manual, 3, 83

—V—Vertical Accuracy, 88Vertical Cursors, 41Vibration, 99Video Frames, 52Video Lines, 52Video Trigger, 51, 89Viewing Stored Screens, 60Voltage Probes, 3, 78, 81VPS200 Voltage Probe Set, 3

—Z—Zoom, 38, 97