0 pl-950 safeaire ii fh manualsash and remove blocking, being careful not to damage sill or baffles....

24
Fisher Hamilton ® SafeAire ® II Fume Hoods Operation, Maintenance and Installation Instructions

Upload: others

Post on 22-Oct-2020

2 views

Category:

Documents


0 download

TRANSCRIPT

  • Fisher Hamilton®

    SafeAire®II Fume Hoods

    Operation, Maintenance and Installation Instructions

  • 2

    SafeAire® II Laboratory Fume Hoods

    Operation, Maintenance and Installation Instructions

    General Information

    Fisher Hamilton is pleased to have had the privilege offurnishing your laboratory equipment. This is a majorinvestment, and your equipment is built to withstand extremesof temperature, stress and corrosion provided you give itreasonable care and use.

    This manual gives you full information for the operation andcare of fume hoods and other Fisher Hamilton equipmentitems.

    Utility and fine appearance can be assured for many years by following simple procedures. A regular schedule ofmaintenance will be most effective.

    If you have a special problem or would like extra copies of thismanual, write to the Technical Service Department, FisherHamilton, Two Rivers, Wisconsin 54241.

    Table of ContentsGeneral Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2–3

    General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2

    Perchloric Acid Hoods . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3

    Hood Identification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3

    Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4–10

    Fume Hood Superstructures . . . . . . . . . . . . . . . . . . . . . . . . . . . 4

    HOPEC IV Superstructures . . . . . . . . . . . . . . . . . . . . . . . . . . 5–6

    Baffles . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7–8

    Blower Enclosures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

    Fume Hood Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10

    Exhaust Filter Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10

    Minihelic Gauge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10

    Warning and Operating Instructions . . . . . . . . . . . . . . . . . . . . 11

    Maintenance and Adjustments . . . . . . . . . . . . . . . . . . . . . . 12–18

    General Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12

    Fume Hood Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12

    Cleaning Fume Hood Interiors . . . . . . . . . . . . . . . . . . . . . . . . . 12

    How to Replace Fluorescent Light Tube . . . . . . . . . . . . . . . . . 13

    Servicing Fume Hood Fixtures . . . . . . . . . . . . . . . . . . . . . . . . . 13

    Replacing Sash Glass and Cables . . . . . . . . . . . . . . . . . . 14–16

    Exhaust Filter Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . 17

    Blower RPM Adjustments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17

    Manometer, Monitoring Exhaust Filters . . . . . . . . . . . . . . . . . . 18

    Field Testing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19–21

    Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22

    General

    Fume hoods are exposed to extremes of temperature,reagent fumes and working surface abuse. Regular care willprolong service life and insure safe working conditions.

    The exhaust system and blower of a fume hood must function properly for safety. Maintenance personnel shouldservice the fan and motor assembly regularly, lubricate asrequired, and make sure that the exhaust system is freefrom obstructions. Semiannually, accumulated depositsshould be removed from the impeller blade and housing.

    A simple test with lighted match or smoke will show if the airis being drawn into the hood. More accurate checks of airvelocity can be made with a thermal anemometer. SeeInspection and Field Evaluation procedure.

    Always place equipment and apparatus as far back into thefume hood as possible since this provides greater assur-ance of proper fume collection and removal.

    Large, bulky apparatus or equipment should be placed inthe fume hood to permit air flow around it, and never placedso as to interfere with the operation of the baffle system.Raise large items an inch or two above work surface. Spilledliquids, acids, or corrosive materials should be immediatelywiped up and the surface neutralized with water or the proper neutralizing agent so as to prevent damage to thework surface and the hood interior or to apparatus andequipment installed in the hood.

    Remember that special fume hoods are required for the handling of Perchloric Acid. See next page.

  • SafeAire® II Laboratory Fume Hoods

    Operation, Maintenance and Installation Instructions

    General Information

    3

    Perchloric Acid Hoods

    The properties of perchloric acid require that a speciallydesigned fume hood be set aside for exclusive use with thismaterial. The hood is equipped with a cold water spraydevice for washdown of the interior surfaces. A trough isplaced across the back of the hood for collection and disposal of washdown waters. Operating personnel shouldbe well trained in the proper handling techniques and befamiliar with the characteristics of this material.

    Frequency of washdown, both hood interior and exterior system, is determined by the usage and concentration ofreagents. This can range from a weekly procedure to onethat occurs after every use. Washdown should always be followed by an inspection to verify that all areas are cleanand that the wash system is functioning properly.

    Some of the hazards of perchloric acid which justify the useof a special hood are:

    1. Perchloric acid is a very strong acid, capable of produc-ing severe burns when in contact with skin, eyes or respiratory tract.

    2. As an aqueous solution, it can cause violent explosions ifimproperly handled.

    3. It reacts with other substances to form unstable materialswhich are susceptible to exploding either by impact, friction, or spontaneous combustion.

    Persons using perchloric acid should be thoroughly familiarwith its hazards. Many reported laboratory accidents haveinvolved less than one gram of reactant. Listed below aresome common safety practices that should be followed:

    • Spilled perchloric acid should be thoroughly washedaway with large amounts of water.

    • The use of organic chemicals or materials in the hoodshould be avoided.

    • Goggles or other effective eye protection should be usedwhenever possible, as well as utilization of the fumehood sash for additional safety.

    • Gas flames or oil baths should not be used within thehood.

    • Organic chemicals should not be kept in storage areasset aside for perchloric acid storage.

    • A schedule should be made for regular washdown andinspection of hood interior, ductwork and blower toguard against a build-up of dangerous perchloric materials.

    • Only a fluorocarbon grease should be used as a blower lubricant, since any other type is to be considered potentially hazardous.

    • Washdown procedure should be performed after completion of usage with all apparatus removed fromhood.

    Hood Identification

    Fisher Hamilton L.L.C. Two Rivers, WI 54241

    Item/Room No. . . . Item: 158C Room: 207Job/Unit No. . . . . TX 394381 012 H012M Unit: 0005093Product No. . . . . . FH3943810244 Entity:Description . . . . . Desc: X54L836PD – DWG. HD01Project Name . . . . Project Name: RICE UNIVERSITY – OLD CHEMISTRYShop Order No. . . . Shop Order #: 161914 P.O. No.: 22492/26108

    Clr:G1G1UNCRATED

    Bar Code No. . . . . 018021957

    Front Panel Side

    Panel

    2-Part Label(Located on backside of collarand bottom of the airfoil sill)

    Liner

    SashBaffles

    TX 3

    9438

    1

    FH

    3943

    8102

    44X5

    4L83

    6PD

    – D

    WG.

    HD0

    1Placed on backside of collar

    Placed on bottomof airfoil sill

    GasketedAccess

    Panel

    Airfoil Sill

  • 4

    Dimensions are nominal and illustrations and specifications are basedon the latest product information available at the time of publication.The right is reserved to make changes at any time without notice.

    SafeAire® II Laboratory Fume Hoods

    Operation, Maintenance and Installation Instructions

    Installation

    ShippingScrews

    WeightPan

    Work Surface

    Top of Fume Hood

    ShippingScrew

    FrontPanel

    No. 10 X 5/8"Screws

    2. Remove two (2) shipping screws that secure the counter- balance weight pan to the rear top rail.

    3. Remove screws from sash hold-down clips. Open thesash and remove blocking, being careful not to damagesill or baffles.

    Note: Baffles can be installed at this time, or after the hoodis in place on the work surface. See Page 7.

    8. Drill two (2) each 1/8" diameter pilot holes at each sidestructural frame into the work surface and secure samewith four (4) No. 10 X 5/8" screws saved from the shipping skid.

    9. Reinstall end panels.

    10. Caulk hood to work surface with silicone sealant.

    6. Remove two (2) shipping screws from the top of the frontpanel. These screws must be removed to allow futureaccess to the fluorescent lights.

    7. Check the following items:

    • That the counterweight operates free of obstructions.• That the cables align properly in the pulleys.• That there is proper horizontal sash alignment and

    counterweight balance.

    • That the sash does not bind in the sash guides.

    4. Remove shipping screws holding the fume hood frame tothe skid. Save four (4) of these screws, No. 10 X 5/8", tosecure hood to the work surface.

    5. Place the fume hood on the work surface taking care to protect the work surface.

    Catches

    Side Panel

    Lower Lip

    Fume Hood SuperstructureInstallation

    Side Enclosure Panels

    1. Lower the side panel into the side frame of the fumehood superstructure, engaging the frame’s lower lip.While pressing down on the two black catches, rotate theside panel and engage the frame’s upper lip. Gentlyapply additional pressure to the panel and release thecatches to secure the panel to the hood.

  • SafeAire® II Laboratory Fume Hoods

    Operation, Maintenance and Installation Instructions

    Installation

    5

    Dimensions are nominal and illustrations and specifications are basedon the latest product information available at the time of publication.The right is reserved to make changes at any time without notice.

    HOPEC IV Fume Hood Installation

    1. Install and level two fume hood base cabinets with rearfillers and kneespace panel as required.

    2. Depending on height requirements, install either the topfiller frame assembly alone or in combination with theoptional-height filler frame.

    HOPEC IV Hood

    BaseCabinet

    TroughAssembly

    Run siliconesealant bead on back flange oftrough. Place worksurface into frameand onto troughflange.

    If the top filler frame only isrequired, insert rough assem-bly into top filler frame andlocate front lower edge oftrough flush with cabinet oroptional height filler. Secure tothe base cabinet using eight(8) No. 10 X 5/8" PPHSMSthrough the side returns of theTop Filler Frame into the frontand rear returns of the basecabinet.

    If the optional height fillerframe is required, it isinstalled by aligning it’s bot-tom 3/4" return on the basecabinet and securing with (4)No. 10 X 5/8" PPHSMSscrews. Insert through assem-bly into Top Filler Frame andlocate front lower edge oftrough flush with cabinet oroptional height filler. Attach theTop Filler Frame to theOptional Filler Frame usingeight (8) No. 10 X 5/8"PPHSMS through the sidereturns of the Top Filler Frameinto the front and rear returnsof the base cabinet.

    36" Work Surface

    Height

    Top Filler Frame

    Optional Filler

    Fille

    r

    End view showing HOPEC IV hood,Top Filler Frame and Base Cabinet.

    OptionalHeight Filler

    Top FillerFrame

    WorkSurface

    BaseCabinet

    HOPEC IV Hood

    BaseCabinet

    36" Work Surface

    Height

    Top Filler Frame

    Fille

    r

    End view showing HOPEC IV hood,Top Filler Frame, Optional HeightFiller and Base Cabinet.

    TroughAssembly

    Run siliconesealant bead on back flange oftrough. Place worksurface into frameand onto troughflange.

    Lower frontedge of troughshould be flushwith front ofcabinet

    Top FillerFrame

    WorkSurface

    BaseCabinet

  • 6

    Dimensions are nominal and illustrations and specifications are basedon the latest product information available at the time of publication.The right is reserved to make changes at any time without notice.

    SafeAire® II Laboratory Fume Hoods

    Operation, Maintenance and Installation Instructions

    Installation

    HOPEC IV Fume Hood Installation (cont.)

    3. Set the fume hood trough onto the top fillerframe. The rear edge of the trough will set overthe front horizontal rail and the return edge onthe trough will secure it in place.

    Note If a drain tube is required, it should beinstalled at this time. The tube is located at theleft end of the trough and is fed through the topof the left hand base cabinet into a one galloncontainer.

    4. Install poly vent, if required, through pre-drilledholes in the backs of base cabinets. The tailpiece of the vent is supplied with a double compression assembly to lock it in place.

    5. Seal the top edge of the trough assembly with a RTV silicon or other applicable sealer andthen place the fume hood work surface into the Top FillerFrame. If base cabinet poly vents are used, align them tothe pre-cut holes in the work surface.

    6. Install cupsinks if required. The fume hood base cabinetsare supplied with pre-punched holes in the back panelsabove the stone liner for cupsink drains. Cupsink assem-blies are supplied with the sink, an extension pipe, anelbow, and a pipe section that protrudes through the holein the base cabinet back. The cupsink assembly isdesigned to fit two conditions.

    A. When the fume hood assembly is supplied with theOptional Height Filler Frame, the cupsink assembly isfished through the sink cutout hole in the work surface and the drain line hole at the rear of the basecabinet.

    B. When the fume hood is supplied with only the TopFiller Frame, the extension pipe below the cupsink isremoved to maintain alignment with the base cabinetrear drain hole. The pipe extension should be savedin the event that the Optional Height Filler Frame berequired at a later date.

    7. Move the HOPEC IV fume hood to the area of installation.Perform Steps 1–10 on Page 4.

    Drain Tube

    Must Be Flush

    BaseCabinet

    Top FillerFrame

    Work SurfaceFront Horizontal Rail

    Trough

  • SafeAire® II Laboratory Fume Hoods

    Operation, Maintenance and Installation Instructions

    Installation

    7

    Dimensions are nominal and illustrations and specifications are basedon the latest product information available at the time of publication.The right is reserved to make changes at any time without notice.

    Baffle Installation and Adjustment

    Fisher Hamilton provides two choices of baffle design concepts:1. Fixed.2. Remote exterior control

    Baffle Installation

    Install baffles according to information on the next page.

    See label located on right front post for Remote ExteriorBaffle adjustment instructions.

    Baffle Adjustment

    Fisher Hamilton fume hoods have two baffle settings:A. Lighter-than-Air or high thermal loading.B. Heavier-Than-Air

    Baffle position A is for lighter-than-air gases, or high heat loads. The top slot is open to maximum. Fume hood air flowis concentrated in the upper portion of enclosure for efficientcapture and exhaust of lighter-than-air fumes and hot gases.

    The baffle at position B provides maximum air flow acrossthe work surface for the collection and removal of fumesgenerated at the work surface and heavier-than-air gases.

    A – High Heat Loads or Lighter-Than-Air Gases

    B – Heavier-Than-Air Gases and Surface Fumes

  • 8

    Dimensions are nominal and illustrations and specifications are basedon the latest product information available at the time of publication.The right is reserved to make changes at any time without notice.

    SafeAire® II Laboratory Fume Hoods

    Operation, Maintenance and Installation Instructions

    Installation

    Fixed Baffle Installation

    Installing Fixed Baffle

    1. Place upper baffle into position by engaging top edgeinto front exhaust collar extrusion. Lift baffle up and intoupper support brackets to lock into position.

    2. Place top edge of lower baffle into front slots on bottomedge of upper support brackets. Lift and lock baffle intothe slot of lower support bracket.

    Exhaust Collar Extrusion

    Upper Baffle

    Lower Baffle

    Upper Support Bracket

    Lower Support Bracket

    Upper Remote Baffle

    Lower Fixed Baffle

    Remote Control Baffle Installation

    Installing Remote Exterior Control Baffle

    1. Place upper baffle into position by aligning forked slot ofremote arm straight up and down. Slide upper baffle thrufingers of remote arm and drop into top slot of uppersupport bracket. (Baffle should be in front of exhaust collar extrusion).

    2. Place top edge of lower baffle into front slots on bottomedge of upper support brackets. Lift and lock baffle intothe slot of lower support bracket.

    Exhaust Collar Extrusion

    Upper Baffle

    Lower Baffle

    Upper Support Bracket

    Lower Support Bracket

    Installing Lower Baffle – Fixed and Remote

  • SafeAire® II Laboratory Fume Hoods

    Operation, Maintenance and Installation Instructions

    Installation

    9

    Dimensions are nominal and illustrations and specifications are basedon the latest product information available at the time of publication.The right is reserved to make changes at any time without notice.

    Installing Blower Enclosure to Fume Hood

    Hardware included:1 - Front panel1 - Front panel-removable2 - Side panels2 - Front support brackets2 - Rear support brackets1 - Bag assembly

    Tools required:#2 Phillips screwdriver.

    1. Fasten support brackets (1) in front top of frame assemblyin holes provided. Use 1/4-20 X 3/4" hex head thread cut-ting screws (HHTCS) (2) furnished in bag assembly tofasten brackets.

    2. Fasten rear support (3) to rear top of frame assembly. UseNo. 8 X 5/8" Phillips pan head sheet metal screws(PPHSMS) (4) furnished in bag assembly.

    3. Install right-hand (5) and left-hand (6) side panels to supports using No. 8 X 5/8" PPHSMS (4).

    4. Install front panel (7) to side panels with No. 8 X 5/8"PPHSMS.

    5. Subassemble thumb screws (8) and push retainers (9) toremovable panel (10).

    6. Insert hole fasteners (11) in larger hole in flange of side panels.

    7. Insert top flange of removable panel under and behindfront panel. Fasten thumb screws and tighten.

    4

    3

    4

    4

    3

    12

    4

    1

    2

    4

    4

    5

    117

    9

    8

    89

    10

    11

    4

    4

    6

  • 10

    Dimensions are nominal and illustrations and specifications are basedon the latest product information available at the time of publication.The right is reserved to make changes at any time without notice.

    SafeAire® II Laboratory Fume Hoods

    Operation, Maintenance and Installation Instructions

    Installation

    Exhaust Filter Installation

    Product Numbers 54L296, 54L297, 54L298 and54L299

    Installation of Minihelic Gauge

    Replacement filter sets consist of one rough and one HEPA filter.

    Product Number 54L302 – Filter set for 54L296 or 54L298.

    Product Number 54L300 – Filter set for 54L297 or 54L299.

    Pre-Filter Cover

    Front ofFume Hood

    Securely attach the filter inlet collar to hood exhaust transition using same method as followed in the duct system. Filter outlet may be attached to duct using flexibleconnector or same as inlet connection.

    Sensor block and sensortube, two req’d each

    Connect to low side ofMinihelic Gauge

    Thick-wall vinyl laboratory tubing, cut to length

    Minihelic Gaugerange 0"-3"W.C., reads filterpressure drop

    Front Post

    Connect to high side ofMinihelicGauge

    The pre-filter coverfaces forward onRestricted Bypass andConstant VolumeBypass hoods.

    Fume Hood Monitor

    Proper fume hood operation is key to laboratory safety,comfort and energy management. OSHA requires thatlaboratories take measures to ensure proper and adequateoperation of fume hoods. Recommendations include the useof a continuous air monitoring device. The ANSI Z9.5 andNFPA 45 standards reinforce these requirements.

    Fume hood monitors have the ability to monitor true fumehood face velocity using thermal sensors located in theinstrument. The thermal sensors are exposed to cleanlaboratory air only. They can be surface-mounted in minuteseliminating the need for expensive panel cutouts.

    Each model is equipped with indicator lights that illuminatebased on a pre-determined set-point. An audible 85dBpiezoelectric alarm sounds and a red indicator lightilluminates to warn of potentially dangerous low air flowconditions.

    Fume hood monitors are shipped with operationmanuals

  • SafeAire® II Laboratory Fume Hoods

    Operation, Maintenance and Installation Instructions

    Warning and Operating Instructions

    11

    Dimensions are nominal and illustrations and specifications are basedon the latest product information available at the time of publication.The right is reserved to make changes at any time without notice.

    Warning and Operating Instructions

    Refer to OperationInstructions label on

    left side of sash glass. Data from the

    “As Manufactured” test should be loggedin this area along withany further Field Test

    results. (See Page 18 and 19)

    Warning

    This product is intended for use with certain chemicals that can cause serious injury or illness through inhalation orphysical contact. While this product is intended to minimizeexposure to certain hazardous chemicals when selected,installed and operated properly, its performance and thesafety of the user is affected by a number of factors. Theseinclude the HVAC system, the specific chemicals andprocesses being used, proper operation and the conditionof the room.

    Before using this fume hood, consult the owner’s industrialhygienist or safety representative to make sure:

    1) The specific fume hood alarms, controls and the HVACsystem have been properly selected and are operatingcorrectly;

    2) The hood has been tested after installation androutinely thereafter to ensure the fume hood isproviding the proper containment for the specificchemicals and processes being used;

    3) There has been appropriate training on the correct useof the fume hood and handling of the specificchemicals and the fume hood operating instructionshave been reviewed;

    4) Any personal protective devices that are required areproperly selected and provided;

    5) The fume hood is being operated at the appropriateface velocity. The fume hood should never be operatedwith the sash in the full open position.

    Operating Instructions

    Failure to follow these instructions could result inphysical injury or illness.

    Caution: Do not use hood for perchloric acidprocedures.

    1. Do not use this fume hood unless you have receivedproper training from the owner’s industrial hygienist orsafety representative.

    2. This fume hood is not intended to be used with allchemicals or all chemical processes. Consult the owner’sindustrial hygienist or safety representative to determinewhether the hood is appropriate for the chemicals andprocesses to be used.

    3. Verify that the fume hood exhaust system and controlsare operating properly and providing the necessary airflow. If in doubt, the owner’s industrial hygienist or safetyrepresentative should be consulted. It is recommendedthat the hood be equipped with an air flow monitoringdevice. Before using the fume hood, verify that themonitor is operating properly by testing the monitor.

    4. The hood should not be operated with the sash in the fullopen (set-up) position. When the hood is in use, theopening of the sash glass should be kept at a minimum.On a vertical rising sash, the sash glass should be nohigher than 18". The sash should remain closed when thehood is not in use.

    5. Place chemicals and other work materials at least six (6)inches inside the sash.

    6. Do not restrict air flow inside the hood. Do not put largeitems in front of the baffles. Large apparatus should be elevated on blocks. Remove all materials not needed forthe immediate work. The hood must not be used forstorage purposes.

    7. Never place your head inside the hood.

    8. External air movement can affect the performance of thehood. Do not operate near open doors, open windows orfans. Avoid rapid body movements. Do not open thehood if there are cross-drafts or turbulence in front of thehood. Do not open the sash rapidly.

    9. Wear gloves and other protective clothing if contact with contaminants is a hazard.

    10. Clean spills immediately.

    11. If fumes or odors are present, stop operating the hood,close the sash and contact the owner’s industrialhygienist or safety representative immediately.

    12. It is recommended that this fume hood be tested and certified annually by the owner according to applicableindustry and government standards.

  • 12

    Dimensions are nominal and illustrations and specifications are basedon the latest product information available at the time of publication.The right is reserved to make changes at any time without notice.

    SafeAire® II Laboratory Fume Hoods

    Operation, Maintenance and Installation Instructions

    Maintenance and Adjustments

    General Maintenance of FumeHoods

    Fume hood maintenance procedures consist primarily ofclean-up, adjustment, lubrication, and replacement of worn,damaged or non-functioning parts. Lubrication of sashguides, cables, pulley wheels, and other working partsshould be accomplished as required and replacement ofbroken, worn, or non-functioning parts as needed. Thefollowing items should be inspected and serviced at leastsemi-annually:

    • Liner and baffles for condition and cleanliness.• Low air flow detectors.• Service fixtures and lights.• Pulleys and belts.• Sash operation and counterbalance cables including a

    complete visual check of the entire system.

    • Make sure there is a 1/16" clearance between cable keepers and pulley sheaves.

    • Velocity and pressure sensing detectors.• Low or no flow alarms, both visible (lights) and audible

    (horns or bells).

    • Signal transmission for alarms designed to activate sig-nals at more than one location.

    • Instrument verification of fume hood face velocity anddetermination of usage by observation and interview.

    • Ductwork and blower.

    Clean-up should be accomplished by, or under thesupervision of, a knowledgeable technician and shouldinclude removal of all baffles for clean-up of all interiorsurfaces.

    Flush all spills immediately using neutralizing compounds asrequired and clean thoroughly. Use good housekeeping in laboratory fume hoods at all times.

    Fume Hood Inspection Procedures

    Safety considerations require that a schedule of inspectionand documentation be set up for every laboratory fumehood at least annually.

    An inspection record should be maintained. This record maybe in the form of a label attached to the fume hood, or a logheld by the laboratory director or health safety director.

    Inspection procedures should include instrument verificationof fume hood face velocity and a determination of usage byobservation and interview. These procedures should alsoconsist of a physical examination of liner condition andcleanliness, baffle and sash operation and condition,counterbalance cables, light operation and condition, andservice fixture function.

    Inspection results should be recorded and reported to the proper authority for any required action.

    Note: Special purpose fume hoods such as those usedwith radioactive materials or perchloric acid requireadditional inspection procedures to cover specialequipment and requirements.

    Options, such as low air flow detectors, when installed,should be inspected at least annually. Where extremehazardous or corrosive conditions exist or when filters arepresent in the system, the inspection frequency should beincreased appropriately. Velocity and pressure sensingdetectors should be tested at each inspection. Low-flow ofno-flow alarms of the visible (lights) or audible (horns orbells) type should be tested for correct operation at least ateach inspection. Signal transmission for alarms designed toactivate signals at more than one location should be verifiedat each location during each inspection. Frayed or brokenbelts should be replaced promptly.

    Cleaning Fume Hood Interiors

    Fume hood liners are maintained by an occasionalwashdown with detergent and warm water. Stains and saltdeposits can be removed with a weak acid solution (5%) oran appropriate solvent – DO NOT USE ACETONE. Removebaffles for access to all surfaces. See Page 8 for removaland installation procedures.

    The use of organic chemicals or materials in a specialized perchloric acid fume hood with a stainless steel interiorshould be avoided.

    Warning

    Use only fluorocarbon grease on blower since anyother type is to be considered potentially dangerous.

    Warning

    Frayed or broken cables should be replaced to avoid personal injury or damage to fume hood.

    Not all cable manufacturer’s cables are the samequality and cycle life. We can only warranty

    cables furnished by Fisher hamilton.

  • SafeAire® II Laboratory Fume Hoods

    Operation, Maintenance and Installation Instructions

    Maintenance and Adjustments

    13

    Dimensions are nominal and illustrations and specifications are basedon the latest product information available at the time of publication.The right is reserved to make changes at any time without notice.

    Installation of Gasked Access panel

    Twist the corners of gasket towardscutout before insertion. Replace thepanel and work the entire periphery ofthe gasket to be sure that the gasket iscompletely snapped into position.Gasket should be smooth and tightwhen properly seated.

    The fixtures used within fume hoods are needle valve type,and if they wear, stainless steel cone and seat replacementkits can be ordered from Fisher Hamilton. It is necessary toremove the handle from the valve and then remove the valvemechanism. This can be done through the access panels(shown above) if fixtures are mounted in the superstructure,or from inside the cupboard if the fixtures are deckmounted.

    Access to the valves by removal of the exterior end panels(Step 1 on Page 4) permits seat replacement without theneed to remove the valve. This approach is recommendedwhen ends are exposed and accessible.

    Access to service fixture valves on fume hoods withoutaccess panels is obtained by removal of the exterior endpanels (Step 1 on Page 4) when hoods are free-standing.

    Access Through Front Posts

    When ends are not accessible, access is gained through the front posts. Remove front panel, unscrew index button, control knob, and retainer flange from fixture handle rod. Remove two (2) screws from post as shown above, lift up and outward to remove post. Electrical fixturesare connected to post with flexible conduit and can remainattached.

    Fluorescent Light TubeReplacement – Bench Top and Hi-Line Fume Hoods

    1. Remove sash enclosure if applicable.

    2. Remove two (2) screws securing bottom of front panel tofront corner posts. Pull front panel up to disengage fromstuds and remove panel.

    3. Squeeze bottom edge of lamp housing to disengagefrom galvanized channel. Rotate lamp housing up toexpose bulbs. Replace bulbs with same type as in unit.turn on light switch to verify connections.

    4. Reverse Steps 1–3 to return hood to usable condition.

    Remove two screws

    2

    1

    Fume Hood Service Fixtures

    Removal of Gasketed Access Panel

    Inser t screwdriver andwedge out panel and

    gasket assembly

    Lift up and pull out Screws

    Front Corner Post

    Index Button, Screw, Control Knob,

    and Retainer Flange

  • 14

    Dimensions are nominal and illustrations and specifications are basedon the latest product information available at the time of publication.The right is reserved to make changes at any time without notice.

    SafeAire® II Laboratory Fume Hoods

    Operation, Maintenance and Installation Instructions

    Maintenance and Adjustments

    Frame Glass and CableReplacement

    Note: Special parts, options, and accessories should be maintained as required.

    Sashes occasionally require service. The glass may fog due to the condensation of chemical vapors, and such materialsshould be removed promptly by washing with water and detergent to prevent etching of the glass.

    WARNING

    If cable is frayed or damaged, it MUST be replaced to avoid personal injury or damage to the fume hood

    1. Hold the sash in the closed/down position by clampingthe sash counterbalance cable to the top structural framemember. In some situations, access to the top of thefume hood may be gained by:• Removing end panel if end of hood is clear - See Page

    4.• Removing front louver panel - See Page 13.• Removing blower enclosure panels- See Page 9.

    2. Remove the outer portion of the sash moulding strip onall four sides of the frame. Use a flat blade screwdriver to

    pry away from inner portion of moulding. Dispose ofbroken and both portions of the sash moulding strip.

    3. Press the four (4) sections of the inner sash mouldingstrip over the sash frame edge.

    4. Place the new sash glass onto the inner sash mouldingand hold in place.

    5. Press the outer moulding strip into the groove of theinner strip.

    6. Remove clamps holding the sash cable in place and testfor balance.

    7. Sash counterbalance cable is replaced by removing frontcorner post (See Page 13) for access. Field repairs canbe made on existing cable if the problem is an endfastening device or disconnection. If cable is frayed orbroken, order replacement cable – describe hood type,size and cable length.

    Top of Fume Hood

    Cable

    Clamp

    LaminatedSafety Glass

    OuterMoulding

    Inner SashMoulding

    LaminatedSafety Glass

    Outer Moulding

    Inner SashMoulding

    Sash Frame

    Inner SashMoulding

    Sash Frame

    LaminatedSafety Glass

    Inner SashMoulding

  • SafeAire® II Laboratory Fume Hoods

    Operation, Maintenance and Installation Instructions

    Maintenance and Adjustments

    15

    Dimensions are nominal and illustrations and specifications are basedon the latest product information available at the time of publication.The right is reserved to make changes at any time without notice.

    Stainless SteelEdge Protector

    Right Sash Guide

    Sash Glass

    Right Corner Post

    Left Corner Post

    Bottom Sash Frame

    Left Sash Guide

    Fume HoodFrame

    Assembly

    Sash Glass Replacement – Fume Hoods That Have Been Installed

    1. Remove both front corner posts, See Page 13.

    2. After removal of one of the corner posts, clamp the sashcable to the top of the side frame assembly. See page14.

    3. Remove screws that hold the sash guide to the frontframe upright and slide the guide up and away from thesash. Care should be taken when the sash guide isremoved so that the glass does not fall from theremaining guide.

    Warning

    If cable is frayed or damaged, it MUST be replaced toavoid personal injury or damage to the fume hood.

    Front Panel

  • 16

    Dimensions are nominal and illustrations and specifications are basedon the latest product information available at the time of publication.The right is reserved to make changes at any time without notice.

    SafeAire® II Laboratory Fume Hoods

    Operation, Maintenance and Installation Instructions

    Maintenance and Adjustments

    4. While holding the sash glass, first remove the upperstainless steel edge protector. Then force the glass fromthe bottom sash pull by pulling upward while holding thesash pull in place. At this point the glass and gasketmaterial should come loose, the bottom sash frameremains attach to the sash cables and one sash guide.

    5. Place gasket material on the bottom of the new sashglass. Align with the bottom frame member and pressinto place. Replace the stainless steel edge protector onthe top horizontal edge of the glass. Replace the sashguide and remove clamps holding the sash cables. Movethe sash up and down to test for proper alignment in thesash guides. If at this point you notice the glass is notcompletely seated into the bottom frame member, tapgently on the bottom with a rubber mallet to seat theglass.

    6. Replace the front corner posts and the front panel.

    Top Sash Edge Protector

    Access Holes to Remove Screws from Sash Guide

    Sash Assembly

    BottomSash Pull

    Frame Front Upright

    Sash Glass Replacement – FumeHoods That Have Been Installed(Cont.)

  • SafeAire® II Laboratory Fume Hoods

    Operation, Maintenance and Installation Instructions

    Maintenance and Adjustments

    17

    Dimensions are nominal and illustrations and specifications are basedon the latest product information available at the time of publication.The right is reserved to make changes at any time without notice.

    Blower RPM Adjustments

    1. Remove housing over motor blower assembly.

    2. Loosen the four (4) bolts [A] which hold the motormounting plate stationary so that the plate has a verticalmovement, as shown in illustration below. This should bedone so that a later adjustment for correcting belt tensioncan be made.

    3. Make all adjustments ONLY with the outside half sheave[B] on the driving shaft.

    4. To increase the RPM of the blower, increase the diameterof the driving sheave by loosening the Allen screw [C]and turning the outside half-sheave toward the motor.Tightening the Allen screw to the flat portion of thethreaded shaft then fixes the diameter of the sheave.

    5. To decrease the RPM of the blower, decrease thediameter of the driving sheave by loosening the Allenscrew [C] and turning the outside half-sheave away fromthe motor. Tightening the Allen screw to the flat portion ofthe threaded shaft then fixes the diameter of the sheave.

    6. Correct belt tension (side play 1/2" to 3/4") can now be set by adjusting the loosened motor mounting plate andtightening the four (4) bolts.

    Terminology:Sheave = Pulley

    Driven = Attached to blower shaftDriving = Attached to motor shaft

    Motor

    To Blower

    Driving Shaft

    Driven Shaft

    Allen Screw [C]

    Outside Half-Sheave [B]

    Motor Side Half-Sheave

    [A]

    Exhaust Filter Replacement

    To remove pre-filter – remove six (6) screws that secure thepre-filter cover. Slide the filter out. Slide new filter into the housing and secure cover in place.

    To remove final filter – loosen thenuts above the springs to unclampthe filter. Continue turning the nutsuntil they contact the upperflange of the spring bracketand lifts the plenum away fromthe filter approximately 1/8 inch.

    Slide the old filter out (therods will support the upperplenum and duct work).Insert the new final filter.Turn the four (4) nuts sownto compress the springs.Inspect seal afterinstallation and periodicallythereafter.

    Cover

    Final Filter

    Cover

    Pre-FilterScrews

    Filter Housing

    Loosen Nutto ReleaseFinal Filter

    WARNING

    Wear adequate protective devices and use techniques to prevent contamination of the laboratoryenvironment. Filter exchange should be accomplishedby knowledgeable personnel who perform sterilization

    procedures or are trained for the safe handling ofradioactive materials. Some users prefer to wet the

    filter media with a steam or aerosol generator toincrease adhesion of particles.

  • 18

    Dimensions are nominal and illustrations and specifications are basedon the latest product information available at the time of publication.The right is reserved to make changes at any time without notice.

    SafeAire® II Laboratory Fume Hoods

    Operation, Maintenance and Installation Instructions

    Maintenance and Adjustments

    Monitoring Exhaust Filters Using aManometer

    When a filter device is installed on a fume hood exhaust system, it is important that the filter performance andcondition be monitored to ensure proper performance of thefilter and of the fume hood to which it is connected.

    By measuring the pressure drop across the filters, themanometer assembly will provide information on filterfunction and condition.

    Manometer reading with clean filters should be recordedand marked with grease pencil on the face of the unit. Whenthe reading changes by one inch, replace filter(s). Amaintenance schedule should be set up for periodic readingof the manometer. Frequency can be determined by usage.

    Any change in manometer reading should be investigated. It may indicate filter damage, over-pressure, or an unsafe operating condition. If it is subjected to an over-pressure, disassemble and examine for fluid in loops and tubes. Drain and re-install per instructions.

    The manometer assembly requires a periodic cleaning ofthe exterior with water or naphtha and inspection andadjustment of the oil level. Adjust micrometer knob for zeroreading as required. Add .826 sp. gr. red gauge oil whenneeded, to maintain zero reading. Oil can be obtained fromFisher Hamilton.

    Proper use of the manometer assembly provides acontinuous indication of filter and operation conditions.

    Note – Be sure to remove manometer when changing filters.

    � �� �

    ���

  • SafeAire® II Laboratory Fume Hoods

    Operation, Maintenance and Installation Instructions

    Field Testing

    19

    Dimensions are nominal and illustrations and specifications are basedon the latest product information available at the time of publication.The right is reserved to make changes at any time without notice.

    three sections of equal area across the top half of the fumehood face and three sections of equal area across the bot-tom half of the fume hood face.

    Note: If not in accordance with specified face velocity, refer to Troubleshooting section on Page 22, for aid in determin-ing the cause of variation in air flow. If face velocity can notbe corrected to that which is specified, reclassify fume hoodto conform to actual face velocity. Shut off auxiliary air when testing an Auxiliary Air fume hood.

    Sash Operation

    Check operation of the sash by moving it through its fulltravel. Sash operation shall be smooth and easy. Vertical rising sashes shall hold at any height without creeping up or down.

    Air Flow

    Fume Hoods

    Turn fume hood exhaust blower on. With sash in the open position, check air flow into the fume hood using a cottonswab dipped in titanium tetrachloride or other smokesource. A complete traverse of the fume hood face shouldverify that air flow is into the fume hood over the entire facearea. A reverse flow of air indicates unsafe fume hood oper-ation. Consult the Troubleshooting section on Page 22, forpossible causes and take corrective action. Move a lightedsmoke bomb throughout the fume hood work area directingsmoke across the work surface and baffle.Smoke should becontained within the fume hood and be rapidly exhausted.

    Low Air Flow Monitor

    On fume hoods with low flow warning devices, verify that monitor functions properly and indicates unsafe conditions.

    Fume Hood Evaluation in the Field

    It is recommended that the user make provisions to have the following tests performed on all laboratory fume hoods.These tests should be performed by qualified personnel toverify proper operation of the fume hoods before they areput to use. The tests of the fume hoods should be performed after the installation is complete, the building ventilation system has been balanced, and all connectionsmade. Any unsafe conditions disclosed by these testsshould be corrected before using the hood.

    Test Procedures

    Test Conditions

    Verify that building make-up air system is in operation, thedoors and windows are in normal operating position, andthat all other hoods and exhaust devices are operating atdesigned conditions.

    Room Conditions

    Check room condition in front of the fume hood using athermal anemometer and a smoke source to verify that thevelocity of cross drafts does not exceed 20% of the specifiedaverage fume hood face velocity. Any cross drafts thatexceed these values shall be eliminated before proceedingwith the fume hood test.

    Equipment List

    (a) A properly calibrated hot-wire thermal anemometer simi-lar or equal to Alnor Model No. 8500.

    (b) A supply of 1/2 minute smoke bombs.

    (c) A bottle of titanium tetrachloride and a supply of cottonswabs or other recognized device for producing smoke.

    Note: It must be recognized that no fume hood can operateproperly if excessive cross drafts are present.

    Face Velocity

    Determine specified average face velocity for the fume hoodbeing tested. Perform the following tests to determine iffume hood face velocities conform to specifications. Withthe sash in normal operating position, turn ON the exhaustblower. The face velocity shall be determined by averagingthe velocity of six readings taken at the fume hood face.Readings shall be taken at the centers of a grid made up of

    CAUTION

    Titanium tetrachloride fumes are toxic and corrosive. Use sparingly, avoid inhalation and exposure to body,

    clothing and equipment.

  • 20

    Dimensions are nominal and illustrations and specifications are basedon the latest product information available at the time of publication.The right is reserved to make changes at any time without notice.

    SafeAire® II Laboratory Fume Hoods

    Operation, Maintenance and Installation Instructions

    Fume Hood Testing

    The test sheet attached to the hood reflects hoodperformance parameters. This sheet represents “AM”testing.

    Fisher Hamilton strongly recommends that the ASHRAE110-1995 test procedure be subjected to this hood under“AU” (as used) conditions.

    Refer to the ASHRAE Standard 110-1995 or contact FisherHamilton for further information.

    If, for some reason, the above test cannot be performed at the job site, Fisher Hamilton strongly suggests use of the SEFA-1.2 1996 test procedure as minimal proof of properhood performance.

    This test consists of a face velocity grid test and a smoketest procedure.

    Information and copies of this procedure are available fromFisher Hamilton.

    Fisher Hamilton also recommends at least semi-annual verification that this above criteria is subjected to and met by all hoods at your particular facility.

    Fume Hood Testing

    ANSI/ASHRAE 110-1995

    The performance of a laboratory fume hood in providing protection for the worker at the face of the hood is stronglyinfluenced by the laboratory room ventilation, and by other features of the laboratory in which it is installed. Therefore,there arises a need for a performance test which can beused to establish an “as manufactured” and an “as used”performance rating, including the influences of thelaboratory arrangement and it’s ventilating system.The test presumes a conditioned environment. No test can be devised which would, conducted once or infrequently,(viz., annually), reflect the results which would be obtainedin a non-conditioned laboratory with various conditions of windows, wind velocity, etc.

    This procedure is a performance test method.

    It remains for the user, the hygienist, or the applications engineer to specify what level of hood performance isdesired or required. It should be noted that the performancetest does not give a direct correlation between testing with atracer gas and operator exposures. Many factors, such asthe physical properties of the material, the rate and mode ofevolution, the amount of time the worker spends at the faceof the hood, and several other factors must be integrated,by a trained observer, into a complete evaluation of workerexposure. The performance test does, however, give arelative and quantitative determination of the efficiency ofhood capture under a set of strict, although arbitrary,conditions. The same test can be used to evaluate hoods inthe manufacturer’s facilities under (presumable) idealconditions, or under some specified condition of room airsupply.

    The test may be used as part of a specification once the appropriate release rate and required control level aredetermines. If so used, an “AM” (as manufactured)specification places a responsibility on the hoodmanufacturer, and an “AU” (as used) specification placesresponsibilities on others, viz., the designer of the room airsupply, the designer of the room layout, etc.

  • SafeAire® II Laboratory Fume Hoods

    Operation, Maintenance and Installation Instructions

    Fume Hood Testing

    21

    Dimensions are nominal and illustrations and specifications are basedon the latest product information available at the time of publication.The right is reserved to make changes at any time without notice.

    Field Evaluation of Laboratory Fume Hoods

    Face Velocity Test

    Square footage of hood opening ____________________

    . . . and Bypass . . . if any ____________________

    TOTAL ____________________

    1A ______________________________________ F.P.M.

    1B ______________________________________ F.P.M.

    2A ______________________________________ F.P.M.

    2B ______________________________________ F.P.M.

    3A ______________________________________ F.P.M.

    3B ______________________________________ F.P.M.

    4A ______________________________________ F.P.M.

    4B ______________________________________ F.P.M.

    TOTAL ____________ = ____________8 avg.

    TOTAL C.F.M. = (Avg. X Sq. Ft. of open sash & any bypass)

    1 2 3 4

    EQUAL EQUAL EQUAL EQUAL

    "

    "

    A

    B

    Project Name ____________________________________

    Location ________________________________________

    Order Number ____________________________________

    Room _______________ Item ______________________

    Fume Hood Identification ____________________________

    ________________________________________________

    Sash Operation ____________________________________

    Light Operation ____________________________________

    Baffle Operation __________________________________

    Services: A G V W NIT. STEAM

    OTHER ______________________________

    Conclusion & Comments ____________________________

    ________________________________________________

    ________________________________________________

    ________________________________________________

    ________________________________________________

    ________________________________________________

    ________________________________________________

    ________________________________________________

    ALARM CONDITION: FUNCTIONAL ________________

    NON-FUNCTIONAL ________________

    SMOKE TEST: POSITIVE ________________

    NEGATIVE ________________

    I certify that the above results were obtained on ____ ____ ____ by __________________________________________________

    Evaluation procedures conducted by ________________________________________ __________________________________Name Title

  • 22

    Dimensions are nominal and illustrations and specifications are basedon the latest product information available at the time of publication.The right is reserved to make changes at any time without notice.

    SafeAire® II Laboratory Fume Hoods

    Operation, Maintenance and Installation Instructions

    Troubleshooting

    Check List

    • Who stated that the unit did not operate properly?Position? Title? Employed by?

    • What tests were performed? Instruments used? When calibrated? Results?

    • Fume hood type and model number? Size?• Is hood location correct/acceptable? Cross currents

    present? Traffic past hood?

    • Is adequate free or make-up air available? Always? Whatis supply source? Can it be altered or cut off?

    • Did hood ever function properly? Have authorized modifications been made? Is baffle properly installed?Adjusted?

    • Have hoods ever been set? Calibrated? Tested?Balanced? By whom?

    • Have recent changes been made in heating/cooling system? Describe.

    Troubleshooting

    When fume hood test procedures detect an improperfunction, the cause is normally due to insufficient quantity oFair flowing through the hood, or due to room cross draftsblowing into or across the face of the fume hood, or acombination of both. The following suggestions are offeredto help pinpoint and correct the problem.

    Room Cross Drafts

    Air moving through an open door located adjacent to thefume hood can cause cross drafts.

    An open window or a room air supply located to one side oracross from the fume hood can cause disturbing crossdrafts.

    High-velocity air from ceiling-mounted diffusers can cause aflow of air down and into the top half of the fume hood facethat can cause reverse flows of air out of the bottom half ofthe face.

    Insufficient Air Flow

    Insufficient air flow through the fume hood can be causedby one or more of the following conditions. Each conditionshould be checked, and eliminated if possible, to determinewhich one or combination of conditions may exist.

    One possible explanation for low face velocity readings is inaccurate face velocity readings. Check air flow velocitymeter type. Was the recommended model used? When wasit calibrated last?

    If the recommended model was not used, check to makesure the instrument is recommended for low air velocities inthe 50 to 100 feet-per-minute range.

    If possible, verify readings with another air velocity meter orby checking air volume using a pitot tube traverse ofexhaust duct.

  • 23

  • Fisher Hamilton L.L.C.1316 18th StreetTwo Rivers, WITel: 793-1121Fax: 793-3085www.fisherhamilton.com

    DO NOT DISCARDIMPORTANT TEST AND

    CALIBRATION DATAENCLOSED!

    TO BE REMOVED ONLY BY HOOD USER

    PL-950 October 2005 Part No. 70051 © Copyright 2005, Fisher Hamilton L.L.C.

    Remove Shipping Screws From Topof Front Panel After Placing Hood in

    Position on Work Surface