생체계측 강의록 medical instrucmentation#3
DESCRIPTION
생체계측 강의록 Medical instrucmentation#3. 2010103821 생체의공학과 최준민 제출일 2014.04.01( 화 ). Dynamic Characteristics. Dynamic Characteristics : 동적인 특성. H. y(t). x(t). Dynamic Characteristics. Laplace transform. Algebraic Equation. Differential Equation. Inverse Laplace transform. - PowerPoint PPT PresentationTRANSCRIPT
Dept. of Biomed. Eng. BME302: Medical Instrumentation Kyung Hee Univ.
1 최준민
생체계측 강의록Medical instrucmentation#3
2010103821 생체의공학과 최준민제출일 2014.04.01( 화 )
Dept. of Biomed. Eng. BME302: Medical Instrumentation Kyung Hee Univ.
2 최준민
Dynamic Characteristics• Dynamic Characteristics : 동적인 특성
Hx(t) y(t)
시간
시간시스템특성 dttxty )()(
xbdtxdb
dtxdbya
dtyda
dtyda m
n
mn
n
n 01
1
101
1
1
위해서미분방정식을이용하기
Dept. of Biomed. Eng. BME302: Medical Instrumentation Kyung Hee Univ.
3 최준민
Dynamic CharacteristicsDifferential Equation Algebraic Equation
Response in S-domainResponse in timeDomain
Laplacetransform
Inverse Laplacetransform
dtetxsXtxL
transformLaplacest)(
Dept. of Biomed. Eng. BME302: Medical Instrumentation Kyung Hee Univ.
4 최준민
Dynamic Characteristics• Laplace Transform’s example
Laplace Laplace Laplace
S
tuL 1)( aS
tueL at
1)( 1)( tL
t t t
Dept. of Biomed. Eng. BME302: Medical Instrumentation Kyung Hee Univ.
5 최준민
Dynamic Characteristics
formLaplaceBy
txLsXsXSdt
txdL mm
m
)( )()(
)()()()(
)()()(
: )( )()()(
)()( )()(
11
01
01
01
1
101
1
1
sXsHLsYLty
transformInversesXsHsY
nctionTransferFusHbsbSabsbSb
sXsYsH
formLaplaceBy
xbdtxdb
dtxdbya
dtyda
dtyda
txLsXsXSdt
txdL
nn
mm
m
n
mn
n
n
mm
m
합의각각의바꾼후부분함수로
Dept. of Biomed. Eng. BME302: Medical Instrumentation Kyung Hee Univ.
6 최준민
Dynamic Characteristics
V s
R
C V c (t )
RCSSVsV
SVsVsSVCR
formRaplaceBy
tuVtVdttdVCR
AC
ACC
Ai
11)( )()(
)()()(1
Dept. of Biomed. Eng. BME302: Medical Instrumentation Kyung Hee Univ.
7 최준민
Dynamic Characteristics
)( 1)()(
111)(
)()(
)()()(
1
1
tueVsVLtV
RCS
VSV
RCSSVsV
SVsVsSVCR
formRaplaceBy
tuVtVdttdVCR
RCt
ACC
AAAC
ACC
Ai
Dept. of Biomed. Eng. BME302: Medical Instrumentation Kyung Hee Univ.
8 최준민
Dynamic Characteristics①0 차 기기
E R (1-k)RkR
Vo
= x(t)
= y(t)
kEERkKR
kRVo
)1(
입력과 출력사이의 모양 차이 X
크기만 달라져 !
0 차기기 : 출력 = 입력의 상수배주파수와 관계 X입력 :
x(t)
출력 : y(t)
k
감쇄기입력 : x(t)
출력 : y(t)
Dept. of Biomed. Eng. BME302: Medical Instrumentation Kyung Hee Univ.
9 최준민
Dynamic Characteristics• Step Response : transient ( 과도현상 ) : 전원을 넣자마자의 상태• Frequency Response : Steady state → transient 사라짐 전원을 넣고 시간이 오래 지났을 때의 상태②1 차기기
xbyadtdya 001
R
C
0
0 차기기 일때에는 입력이 들어가면 바로 출력1 차기기 일 때는 출력이 expotential 하게 입력을 따라가는 현상이 나타남→Transient 현상
Dept. of Biomed. Eng. BME302: Medical Instrumentation Kyung Hee Univ.
10 최준민
Dynamic Characteristics• 1 차의 경우 : 출력이 입력의 주파수에 따라서 크기가 달라진다 .
→ 주파수 함수를 frequency Response 로 표현 → 주파수에 따라서 전달함수를 표현
) :)(( 1)(
)()(
)()()(
functiontransferfrequencyjHjk
jXjYjH
jssXsYsH
Dept. of Biomed. Eng. BME302: Medical Instrumentation Kyung Hee Univ.
11 최준민
Dynamic Characteristics• 예를 들어 100Hz 가 들어오면 다음 수 만큼 출력이 달라진다 .
Response : )( Response : )(
)2100(1
1)()()(
PhasejHmaginutudejH
jk
jk
jXjYjH
log w
log w
Hlog H
21
0.707 -45°
-90°
주파수↑ → 출력↓ , 주파수↓
Dept. of Biomed. Eng. BME302: Medical Instrumentation Kyung Hee Univ.
12 최준민
Distortionless Measurement• 일반적으로 입력이 step 함수 → 출력 step 함수를 원한다 . 즉 , 왜곡이 없는 함수 → 이상적인 0 차기기
)()()()()()()()(
020121
21
ttAxttAxtytytytxtxtx
Delay time 같은 경우 → A 배만큼 커지고 모양이 같다
Hx(t) y(t)
Dept. of Biomed. Eng. BME302: Medical Instrumentation Kyung Hee Univ.
13 최준민
Distortionless Measurement
t t t
)(1 tx )(2 tx )(tx
t t t
)(1 ty )(2 ty )(ty
t0 t0 t0
1 1
2
A A
2A
Dept. of Biomed. Eng. BME302: Medical Instrumentation Kyung Hee Univ.
14 최준민
Distortionless Measurement예 ) 진폭 특성에 의한 왜곡 )(2)()( 0201 ttAxttAxty
t t t
)(1 tx )(2 tx )(tx
t t t
)(1 ty )(2 ty )(ty
t0 t0 t0
1 1
2
A A
2A
3A
-A
입력과 모양 다른 왜곡 생김
Dept. of Biomed. Eng. BME302: Medical Instrumentation Kyung Hee Univ.
15 최준민
Distortionless Measurement• 예 ) 위상 특성에 의한 왜곡 )2(2)()( 0201 ttxttAxty
t t t
)(1 tx )(2 tx )(tx
t t t
)(1 ty )(2 ty )(ty
t0 2t0 t0
1 1
2
A A
2A
Dept. of Biomed. Eng. BME302: Medical Instrumentation Kyung Hee Univ.
16 최준민
Distortionless Measurement
D
D
D
tj
st
st
D
kejwH
kesXsYsH
sXkesY
ttkxty
)(
)()()(
)(
)()(
H
모든 주파수에 대해 선형적으로 일정
H
DtH
기울기 = -tD 로 일정
Dept. of Biomed. Eng. BME302: Medical Instrumentation Kyung Hee Univ.
17 최준민
Distortionless Measurement예 1)
예 2) 신호원의 대역폭 10Hz ~ 100Hz, 무왜곡 측정을 위한 주파수 전달함수 그려라
주파수위상
,)sinsin
)(sin)(sin)(sinsin)(
2211
222111
2211
1211
DD
DD
DD
ttttAttAkttAttAkty
tAtAtx
H
10Hz 100Hz
10Hz~ 100Hz 사이에서 일정 → Flat 한 Magnitude re-sponse
10Hz 100Hz
w
w
H
Magnitude response 는 flat 하며통과시간이 같아야 무 왜곡 !
위상지연위상지연
Phase response 는 – tD 로 일정하며 위상지연은 주파수가 커짐에 따라 비례하여 커지기 때문에 다음과 같은 그래프를 보인다 .
Dept. of Biomed. Eng. BME302: Medical Instrumentation Kyung Hee Univ.
18 최준민
Bio Potential ( 생체전위 )
1) 신경 감각신경 : Dendrite 쪽에 Sensor 가 달려있다 . 적절한 센서를 통해 외부의 자극들이 신경으로 들어와 뇌로 전달 -> 감각을 느낌2) 신경 운동신경 : 운동중추 끝에 근육이 존재해 그 자극으로 움직인다 .
수십 μm
두께 : 수 μm
최대 1m
Dept. of Biomed. Eng. BME302: Medical Instrumentation Kyung Hee Univ.
19 최준민
Bio Potential
Cl-
Na+
K+Na+
Cl-
intracellular fluid
Extracellular fluid
K+
Na+
펌프세포 내부의 농도차이→ Diffustion 발생( 골고루 퍼지려고 함 )
* 확산이 일어나는데도세포 내 외 농도차가 생겨 ?
→ 세포의 Pump 로 인해서
전해질로 채워진세포 내외부
Na+ 3 개당 : K+ 1 개의 비율로 펌프
Dept. of Biomed. Eng. BME302: Medical Instrumentation Kyung Hee Univ.
20 최준민
Bio Potential
Cell Neuron
O PAMP
+
-
O U T
0여기서 ground 는 extra cellular fluid 에 연결
Vo
①
②
t1
-60mVResting Membrane Potential
① 은 extra celluer fluid 영역이므로 전위차가 생기지 않는다 .
(ground = extra cellular fluid)
세포막 안쪽 : -60mV
Dept. of Biomed. Eng. BME302: Medical Instrumentation Kyung Hee Univ.
21 최준민
Bio PotentialV
Na+
Cl- Cl-Na+
0.1% NaCl 1% NaClV
Na+
Cl- Cl-Na+
0.1% NaCl 1% NaCl
① 막이 Insulating Membrane 일 때
Insulating Membrane
이온들이 이동하지 못한다 .
→ i = 0, V = 0
② 막이 Permeable Membrane 일 때Permeable Membrane: to both Na+, Cl-
Na+ 이온과 Cl- 이온 모두 움직 일 수 있기 때문에→ i = 0, V = 0
Diffusion 은 농도차에 비례하는데 Na+ 가 이동하면Diffusion 으로 인해 Cl- 도 이동하므로 → Net Charge X
Dept. of Biomed. Eng. BME302: Medical Instrumentation Kyung Hee Univ.
22 최준민
Bio Potential③ 막이 Semi Premeable Membrane 일 때
V
Na+
Cl- Cl-Na+
0.1% NaCl 1% NaCl
Semi Permeable Membrane: to Na+
++
++
+ +
+ +
-
-
-- --- -
-
-
-
-
+++ +
농도차에 의한 확산으로 + 차지가 왼쪽으로 이동 (Diffusion)
Diffusion
Electric FieldNa+ 의 이동으로 인해 왼쪽이 + 오른쪽이 –가 되면서 Electric field 생긴다 .
Diffusion+ Vrmp -
Dept. of Biomed. Eng. BME302: Medical Instrumentation Kyung Hee Univ.
23 최준민
Bio Potential1) 막이 Semi Permeable membrane to Na+ 일 때 현상은 다음과 같다 .Diffusion 으로 인해 Na+ 가 오른쪽으로 이동한다 .
2)Diffusion 으로 인해 Electric field 가 생겨 왼쪽에서 오른쪽으로 가려는 Coulomb Force 가 생긴다 .
3)Diffusion 의 힘이 초기에 Coulomb force 보다 강하기 때문에 위 과정을 두개의 힘이 같아 질때 까지 반복한다 → 전위차 ↑그로 인해 Vrpm 이 생성된다 .