May 3, 2013 1
10201 Wayzata Blvd., Suite 310 Minneapolis, MN 55305 Sales 877.273.1764 www.atirix.com
TJC Sentinel Event Alert 47: Centralized QC in Practice
Steve Backes, CEO
Copyright © 2013 Atirix Medical Systems, Inc. All rights reserved. Atirix® and QC-Track® are registered trademarks of Atirix Medical Systems, Inc. QC-Track is protected by issued and pending patents, including U.S. 8,428,969 “System
and method for tracking medical imaging quality” and continuations
Device Quality Control is needed for imaging systems… • Process to ensure imaging equipment is performing
per manufacturer specifications, regulatory requirements and physicist needs – Quantitative and qualitative tests
– Phantom
– Documentation
• Regulations include:
– Legislation, such as MQSA, California SB 1237 (CT dosing), MN …
– Accreditation, such as ACR
– Advisory, including The Joint Commission, AAPM (e.g. TG142 for therapy centers)
• A well designed QC program is necessary to ensure image quality
May 3, 2013 3
Jordan, S. The Real Thing. Canadian Bottle and Stoneware Collector 20. April, 1997
Oak Ridge Associated Universities (ORAU) Health Physics Historical Instrumentation
Museum Collection.
Action Item 14: “Implement a system for centralized quality and safety performance monitoring…” to include equipment located anywhere in the hospital or clinics, including “radiology department, cardiac catheterization suite, the OR, … physician offices,…”
Action Item 16: “Ensure that recommended quality control, testing (including daily functional tests) and preventive maintenance activities are performed in accordance with manufacturer’s guidelines.”
Sentinel Event Alert 47, “Radiation risks of diagnostic imaging”, 2011
… in response to regulations…
Medicare Improvements for Patients and Providers Act (MIPPA), 2008
Mammography Quality Standards Act, 1994 Sec. 900.12 Quality standards (d)(2) Quality assurance records. The lead interpreting physician, quality control technologist, and medical
physicist shall ensure that records concerning mammography technique and procedures, quality control (including monitoring data, problems detected by analysis of that data, corrective actions, and the effectiveness of the corrective actions), safety, protection and employee qualifications to meet assigned quality assurance tasks are properly maintained and updated.
‘‘‘SEC. 135. IMAGING PROVISIONS ‘‘(B) ADVANCED DIAGNOSTIC IMAGING SERVICES DEFINED.—…includes ‘‘(i) diagnostic magnetic resonance imaging, computed tomography, and nuclear medicine (including positron emission tomography); and ‘‘(ii) such other diagnostic imaging services… (excluding X-ray, ultrasound, and fluoroscopy),
(3) CRITERIA FOR ACCREDITATION (E) standards that require the establishment and maintenance of a quality assurance and quality control program by the supplier that is adequate and appropriate to ensure the reliability, clarity, and accuracy of the technical quality of diagnostic images produced by such supplier;.
4 May 3, 2013
… and to avoid headlines.
5
West Virginia Hospital Overradiated Brain Scan Patients, Records Show By WALT BOGDANICH
Published: March 5, 2011
THE RADIATION BOOM
After Stroke Scans, Patients Face Serious Health Risks By WALT BOGDANICH
Published: July 31, 2010
THE RADIATION BOOM
X-Rays and Unshielded Infants By WALT BOGDANICH and KRISTINA REBELO
Published: February 27, 2011
THE RADIATION BOOM
As Technology Surges, Radiation Safeguards Lag By WALT BOGDANICH
Published: January 26, 2010
May 3, 2013
Minnesota: Advanced Diagnostic Imaging Accreditation and Reporting Requirements Nov, 2012
“… advanced diagnostic imaging services [CT, MRI, Nuc Med] eligible for reimbursement shall only be reimbursed if the facility where the services has been provided:
1. Has obtained accreditation by August 1, 2013 and
2. Has demonstrated holding accreditation annually in reports to the Commissioner of Health.
May 3, 2013 6
Regardless of motivation, new regulations are being implemented around the country, including here in Minnesota…
Action Item 14: “Implement a system for centralized quality and safety performance monitoring…” to include equipment located anywhere in the hospital or clinics, including “radiology department, cardiac catheterization suite, the OR, … physician offices,…”
Action Item 16: “Ensure that recommended quality control, testing (including daily functional tests) and preventive maintenance activities are performed in accordance with manufacturer’s guidelines.”
Sentinel Event Alert 47 “Radiation risks of diagnostic imaging”
Aug 2011
7 May 3, 2013
… and The Joint Commission has weighed in, too.
How is TJC SEA 47 different? It defines how to implement QC…
May 3, 2013 8
MQSA MIPPA SEA 47
“a QC process is needed for MQSA
accreditation”
“ACR accreditation needed.” “An adequate
QC process”
“Implement a centralized QC system”
“What to do – mammography”
“What to do – CT, MRI, Nuc Med”
“How to implement QC? – Centralized”
… but the reality in most facilities: Binders
9
• QC processes at remote facilities are
difficult to manage
• RSO’s and medical physicists need to travel
to remote facilities to review QC data
• Binders get lost or damaged
• Manual calculations and graphing are time
consuming and error-prone
• Many hours are spent organizing data prior
to an audit
May 3, 2013
ATIRIX SOLUTION: QC-TRACK®
May 3, 2013 10
Device Quality Control
Credentials Management
Web-based device QC and credential tracking system
• Device QC Tracking, Graphing and Trend Reporting
• Automated Phantom Image QC
• Device Failure Tracking
• Apron Tracking
• Credential Tracking
• Device Inventory Listing
• QC Data Repository
• Tablet support
• Designed for intranet operations, with built-in security, audit logging, provisioning and LDAP integration
11
Personal Apron Tracking
May 3, 2013
Atirix QC-Track system
QC-Track has the workflow needed in centralized QC systems...
QC Dashboard
QC Checklist
Worksheet/Templates MQSA
Checklist Device Worksheet
Sign Off
Test Failure Graph
QC-Connector™
12
QC-Track.Aprons
Reports
Trend Analysis
Apron Tracking
Batch Reporting
Custom Reporting
May 3, 2013
Credential Dashboard
Facility/Room Credential
Worksheet/Template
Professional Credential
Worksheet/Templat
e Professional License/
Certificates
ACR Accreditation/ Physicist Survey
… and the credential workflow for a complete QC tracking program
13 May 3, 2013
Credential reports by Modality
The QC Dashboard reveals QC status at-a-glance…
14
… providing links to worksheets, graphs and reports May 3, 2013
Built-in alerts if a test fails
QC-Track Worksheets are based on the vendor’s QC Manual, ACR guidelines, and medical physicist needs
!
Technologist records the QC results on a worksheet
Each test has its own worksheet • Calculations with thresholds
• Pass/Fail
• Customizable
QC-Track performs the calculations
QC-Track verifies results and notifies if test fails with yield icon
Facility can modify worksheets using Worksheet Editor for special needs of inspectors or medical physicists
15 May 3, 2013
And Atirix provides many Worksheets “off the shelf”
16
CT ACR Phantom QC Daily & Weekly Worksheets
MAMMOGRAPHY Fuji CR GE Senographe DS/Essential Hologic M-IV, Selenia Dimensions 2D/Dimensions 3D (Tomo), Selenia/ Selenia Tungsten Siemens Mammomat Novation DR
MRI ACR Phantom QC Daily & Weekly Worksheets
CR/DR/FLUORO SYSTEMS DISC Plus, IBA Primus, PTW Normi Phantom QC
Daily & Weekly Worksheets
STEREOTACTIC BIOPSY SYSTEMS ACR Stereotactic QC GE Upright Hologic MultiCare Platinum
LASER PRINTERS AGFA Drystar AXYS/ 4500M/5500/5300/5503 Fuji DryPix 4000/5000/7000 Kodak 5850/6800/6850/8900 Konica DryPro 793/873 Sony UP-DF750
DIAGNOSTIC REVIEW WORKSTATIONS AGFA Impax Barco Coronis MFGD5421/MFGD 5621 HC Eizo Nanao G5521 Hologic SecurView Kodak Carestream Mammography Planar Dome C5i E5 Siemens Syngo MammoReport Tokoku ME25
DIGITIZERS Hologic – R2 DX/LS Vidar R2 Image Checker - Vidar
DEPARTMENTAL - Mammography Operations Reject Analysis (Multiple imaging unit)
May 3, 2013
… clean, professional graphs and reports,…
17
QC-Track meets accreditation reporting needs • Generate device QC reports
• Modalities • Accessories, including printers, review
workstations, digitizers, etc.
• Generate weekly, monthly, quarterly, annual reports and graphs on demand
• Enables access from report summaries down to detailed worksheets via built-in drill-down capability
Batch Reports
• Full audit packages can be generated at any time
MQSA Reports
Checklist Device Worksheet
Sign Off
Test Failure Graph
May 3, 2013
… as well as Custom Trend Reporting, for analysis not possible in a paper-based “binder” system.
18
• Compare data elements and identify trends • Multiple values within a device • Multiple devices within a facility • Devices across facilities
• Customize the report
• Name • Visual indicators • Time frame
• Save the report • Re-run the same report
• PDF or print
May 3, 2013
QC-Track also has built-in QC Review and Sign-off, for Interested Parties
19
• Interested Parties - lead interpreting radiologist, medical physicist, lead QC technologist and facility management - are able to review and sign off electronically
• See FDA website article posted Feb. 2010
• QC-Track includes a report that shows the sign-off status for each of the devices in a facility
• Interested Party utilizes QC-Track to review QC status to their satisfaction
• QC-Track reminds Interested Party that sign-off is due
May 3, 2013
Professional and Facility Credentials
20
• Dashboards for clinical professionals and facility tracking include credentials and their status
• Track and manage • Credits/CEUs
• Exams
• Related documents
• Identify at-a-glance pending or past-due requirements or renewals
• Emails sent when credits coming due
• Links to detailed views of uploaded physicist reports, scanned certificates, and documents
• Generate reports per ACR requirements for each modality
May 3, 2013
…so, with all that, which regulations to follow?
Analog and FFDM quality control
MQSA MIPPA / ACR Accreditation
Joint Commission SEA47
Mammo professional and facility credentials Medical audit/patient letters
QC sign-offs
CT quality control MRI quality control
Breast MRI quality control
Professional credentials
Centralized QC system (Action Item 14) Follow vendor’s QC process (Action Item 16)
Procedure documentation
Annual 18 – 36 month On-site Audit schedule Random
State/AAPM/ Medical Physicist
As needed
As needed
As needed
21
Fluoro/CR/DR/other
varies
Device Inventory listing
May 3, 2013
SEA 47 IN PRACTICE: UNIVERSITY OF VIRGINIA
May 3, 2013 22
University of Virginia Charlottesville, VA
• 578 bed academic hospital
– 12 affiliated locations
• 150+ imaging units (CT, MRI, FFDM, US…)
• 3 full time medical physicists
May 3, 2013 23
UVA Site 2
QC Data
UVA Site 1
UVA Situation in 2010 • Seeking automated solution to support vision for
QC as a key component of overall image quality
• Limited QC automation: RIT Radia for CT, MRI
– Limited integration where QC automated…
• “System silos”, “Data silos”
• Complex MQSA QC
– 7 FFDMs across 4 locations
– Multiple vendors and generations of FFDM systems
• Explored cost of building in-house QC system…
May 3, 2013 24
QC Data
UVA installs QC-Track
May 3, 2013 25
1. FFDM
2. CT, with Radia & QC-Connector
(XML)
3. Apron Pilot
4. Aprons
5. CT with Radia.Net
(embedded)
6. MRI with Radia.Net
(embedded)
2012 2013
RSNA 2011 “Can you add apron tracking to the QC-Track system?”
Allen Goode, Lead Medical Physicist
26
QC workflow with QC-Track
1. Dashboard indicates QC tests are due in a department
2. QC data gathering. FFDM console reviewed, phantom is scanned, or aprons are gathered and worksheets are completed
3. Results are matched to devices and tests, comments added, and data is entered into QC-Track. Failures are recorded and tracked.
4. QC data available for tracking, reporting
May 3, 2013
Situation today…
27
• Full digital QC for 2 modalities • 7 FFDMs in 4 locations • 5 CTs in 2 locations
• Apron tracking
• 2,500+ aprons in 35 locations
• 85+ users trained in 35+ locations
Result: Interested Parties now have centralized system to track status across locations
May 3, 2013
35+ UVA Sites
QC Data
UVA Intranet
Metrics – as of April 25, 2013 • 11,974 worksheets recorded • 1,192 weekly CT Phantom “uploads” • 997 reports have been run, including 12
batch reports
May 3, 2013 28
Test Count Test Count
Aprons Protective Shielding 1,851 FFDM Viewbox and Viewing Condition 596
Flat Field Test 453
Workstations Phantom Image - DRW 593 Weekly Checklist 226
Quality Control Check 391 SNR/CNR 200
Phantom Image Quality and CNR - AWS 387 Detector Flat-Field Calibration 199
Artifact Evaluation - AWS 198 Compression Thickness Indicator 191
DRW - Phantom Image - 2D 49 Visual Checklist 160
Phantom Image 139
Monitors Monitor Cleaning 1,575 AOP and SNR Check 88
LCD Calibration 64 Phantom Image Quality - AWS 71
Phantom Images 69
Biopsy systems Localization Accuracy - Lateral Approach 332 CNR and MTF Measurement 68
Localization Accuracy - Vertical Approach 331 MTF Measurement 67
Needle Path Block Test 331 Empty Reject Bin 60
Repeat Analysis (<=20%) - Digital 6 Phantom Image - 2D 60
Compression 3 Phantom Image - Tomo 58
Repeat Analysis Check 33
Printers Laser Imager Quality Control 700 Compression Force 27
Artifact Evaluation - Printer 278 Repeat Reject Analysis 24
DICOM Printer Quality Control 278 Repeat Analysis 23
Geometry Calibration 4
Failures: “Better visibility than ever before”
• 288 logged in 2 years
• All failures logged in Failure Log
• Types of failures – Actual “failures”
» Printers » Visual checklist
– “Technical failures but acceptable”
» Known CT phantom ROI positioning issue vs. spiral CT
– Training issues…
May 3, 2013 29
So, what have we learned? • Gaining the commitment needed for an enterprise QC system is a complex
advocacy process for the physics team
– Radiology director – funding
– Department managers – implementation, on-going use
• Aprons – 35 departments….
• Early visible success in FFDM helped build acceptance
• Lots of aprons de-commissioned when faced with tracking “seldom used” pieces
• Flexibility of QC-Track model helpful in meeting federal and state-specific regulations
May 3, 2013 30
… and an obvious lesson: “Enterprise” = “Lots of parties”
May 3, 2013 31
Centralized QC System
Imaging Vendors
IT
Regulators Inspectors
Medical Physicists
Systems (Atirix QC-Track®)
Technologists RSO
Radiology Management
… a few other observations…
• Sponsorship, particularly with recommended span of SEA 47
– E.g., at UVA the physics team manages only the ionizing radiation systems
• Atirix needs QC-Track automated modules for more modalities, esp. Nuc Med
• SEA 47 -- “guideline” versus “enforced”… versus experience of MQSA: MQSA resulted in improved image quality – Spelic. FDA Report 2003.
– Pisano et al. AJR: 174, April 2000.
– Suleiman,OH et al. Radiology 1999; 210: 345-351.
May 3, 2013 32
…and the good news: a well designed digital QC program can also have a positive financial impact.
• Improved Productivity – Reduced travel, documentation expenses
– Reduced QC time, increasing imaging unit availability
– Process consistency
• Increased information availability – Inspections, prep and on-site response time
– Failures and other service events
• Improved Reimbursement – Government payers
• MIPPA requires accreditation (ACR, The Joint Commission, IACTL) for Medicare reimbursement
– Private payers
• United Health, other insurers require ACR accredited imaging units
May 3, 2013 33
- Questions - Upcoming conferences: AHRA - Thanks to University of Virginia Radiological Imaging Technologies (RIT)
- For more information
May 3, 2013 34
Steven J. Backes President/CEO
Moving Medical Decisions
Atirix Medical Systems, Inc. 10201 Wayzata Boulevard Minneapolis, MN 55305 952.674.4726 office 952.546.2023 fax 612.723.5268 cell
www.atirix.com
Sales 877.273.1764 www.atirix.com
Copyright © 2013 Atirix Medical Systems, Inc. All rights reserved. Atirix® and QC-Track® are registered trademarks of Atirix Medical Systems, Inc. QC-Track is protected by issued and pending patents, including U.S. 8,428,969 “System
and method for tracking medical imaging quality” and continuations