tunable laser

4
TDLS200 Tunable Diode Laser Gas Analyzer and its Application to Industrial Process Yokogawa Technical Report English Edition Vol.53 No.2 (2010) TDLS200 Tunable Diode Laser Gas Analyzer and its Application to Industrial Process Kazuto Tamura *1 Tomoaki Nanko *2 Yukihiko Takamatsu *1 Junichi Matsuo *3 Yokogawa’s TDLS200 laser gas analyzer, which is based on tunable diode laser spectroscopy, features high selectivity and long-term stability, and offers fast in-situ analysis of even high-temperature or corrosive gases. This report introduces the measurement principle of the TDLS200 and an example of real-time residual ammonia measurement in a flue to control ammonia injection volume during the denitration and dust collection process. INTRODUCTION A laser gas analyzer employing the tunable diode laser absorption spectroscopy (TDLAS) method has a feature capable of measuring the concentration of the sample gas with high selectivity for gas components and without contact only by irradiating the gas sample with light from a tunable diode laser. Thus fast and accurate in-situ measurements can be performed in process lines under various measurement conditions. For example, measurements can be performed at normal temperature and up to temperatures as high as 1500 °C, and in environments with fluctuating pressure. It can also measure corrosive or highly hazardous gases. Use of the accurate and highly responsive measurement signals produced by the analyzer for process control systems helps improve yields, energy efficiency, and safety in various industrial processes. A simple configuration without moving parts and components with limited lifespan ensures virtually maintenance-free operation and management. Yokogawa’s TDLS200 laser gas analyzer is a new type of laser gas analyzer evolved for process measurement. Applying a measurement method based on the peak area integration method, it eliminates errors caused by the presence of other gases and pressure changes, which lead to measurement errors. It also enables accurate measurement of the concentration of gas components, even under simultaneous fluctuations in temperature and other factors. This paper provides an overview of the TDLS200 laser gas analyzer and describes its features and measurement principle, and introduces examples of its applications. INSTALLATION AND CONFIGURATION OF TDLS200 LASER GAS ANALYZER Figure 1 shows an example of installation and configuration of the TDLS200 laser gas analyzer. The TDLS200 has a launch unit and a detection unit that are usually placed in a facing orientation on either side of the duct (cross duct) through which process gas flows. The cross-duct installation shown in Figure 1 can accommodate ducts up to approximately 20 m wide. Figure 1 Installation and Configuration Example of the TDLS200 Laser Gas Analyzer Optical windows separate the interior of the analyzer from the process gas. The light from the semiconductor laser in the launch unit goes through the optical window, the process gas and the optical window of the detection unit and reaches the photodetector. The photodetector converts the power of the received laser beam into an electrical signal. The arithmetic circuit in the launch unit determines the absorption spectrum of the measured component, calculates the peak area of the spectrum, converts it into component concentration, and outputs it as a 4 to 20 mA analog signal. The optical axis alignment part has a bellows structure that allows easy angle adjustment while maintaining airtightness essential for the industrial process. The connection of the launch unit and the detection unit via the optical axis alignment part enables easy optical axis adjustment not only for regular installation configuration where the Process gas flow Measuring laser beam Semiconductor laser Photodetector (Photo diode) Mounting flange (on both sides) Optical axis alignment (on both sides) Purge gas for preventing contamination of the optical window (on both sides) Optical path length Launch unit (Light emitter) Detection unit (Light receiver) *1 Analytical Products Business Center, Industrial Automation Business Headquarters *2 Quality Assurance Dept., Industrial Automation Business Headquarters *3 Laser Analysis Division, Yokogawa Corporation of America 51 113

Upload: pushkar-sharma

Post on 21-Apr-2015

82 views

Category:

Documents


2 download

TRANSCRIPT

Page 1: Tunable Laser

TDLS200 Tunable Diode Laser Gas Analyzer and its Application to Industrial Process

Yokogawa Technical Report English Edition Vol.53 No.2 (2010)

TDLS200 Tunable Diode Laser Gas Analyzer and its Application to Industrial ProcessKazuto Tamura *1 Tomoaki Nanko *2

Yukihiko Takamatsu *1 Junichi Matsuo *3

Yokogawa’s TDLS200 laser gas analyzer, which is based on tunable diode laser spectroscopy, features high selectivity and long-term stability, and offers fast in-situ analysis of even high-temperature or corrosive gases. This report introduces the measurement principle of the TDLS200 and an example of real-time residual ammonia measurement in a flue to control ammonia injection volume during the denitration and dust collection process.

INTRODUCTION

A laser gas analyzer employing the tunable diode laser absorption spectroscopy (TDLAS) method has a feature

capable of measuring the concentration of the sample gas with high selectivity for gas components and without contact only by irradiating the gas sample with light from a tunable diode laser. Thus fast and accurate in-situ measurements can be performed in process lines under various measurement conditions. For example, measurements can be performed at normal temperature and up to temperatures as high as 1500 °C, and in environments with f luctuating pressure. It can also measure corrosive or highly hazardous gases. Use of the accurate and highly responsive measurement signals produced by the analyzer for process control systems helps improve yields, energy efficiency, and safety in various industrial processes. A simple configuration without moving parts and components with limited lifespan ensures virtually maintenance-free operation and management.

Yokogawa’s TDLS200 laser gas analyzer is a new type of laser gas analyzer evolved for process measurement. Applying a measurement method based on the peak area integration method, it eliminates errors caused by the presence of other gases and pressure changes, which lead to measurement errors. It also enables accurate measurement of the concentration of gas components, even under simultaneous f luctuations in temperature and other factors. This paper provides an overview of the TDLS200 laser gas analyzer and describes its features and measurement principle, and introduces examples of its applications.

INSTALLATION AND CONFIGURATION OF TDLS200 LASER GAS ANALYZER

Figure 1 shows an example of i ns t a l la t ion and conf iguration of the TDLS200 laser gas analyzer. The TDLS200 has a launch unit and a detection unit that are usually placed in a facing orientation on either side of the duct (cross duct) through which process gas flows. The cross-duct installation shown in Figure 1 can accommodate ducts up to approximately 20 m wide.

Figure 1 Installation and Configuration Example of the TDLS200 Laser Gas Analyzer

Optical windows separate the interior of the analyzer from the process gas. The light from the semiconductor laser in the launch unit goes through the optical window, the process gas and the optical window of the detection unit and reaches the photodetector. The photodetector converts the power of the received laser beam into an electrical signal. The arithmetic circuit in the launch unit determines the absorption spectrum of the measured component, calculates the peak area of the spectrum, converts it into component concentration, and outputs it as a 4 to 20 mA analog signal.

The optical axis alignment part has a bellows structure that allows easy angle adjustment while maintaining airtightness essential for the industrial process. The connection of the launch unit and the detection unit via the optical axis alignment part enables easy optical axis adjustment not only for regular installation configuration where the

Process gas flow

Measuringlaser beam

Semiconductor laser Photodetector(Photo diode)

Mounting flange(on both sides)

Optical axis alignment(on both sides)

Purge gas for preventingcontamination of the optical

window (on both sides)

Optical path length

Launch unit(Light emitter)

Detection unit(Light receiver)

*1 Analytical Products Business Center, Industrial Automation Business Headquarters

*2 Quality Assurance Dept., Industrial Automation Business Headquarters

*3 Laser Analysis Division, Yokogawa Corporation of America

51 113

Page 2: Tunable Laser

TDLS200 Tunable Diode Laser Gas Analyzer and its Application to Industrial Process

Yokogawa Technical Report English Edition Vol.53 No.2 (2010)

two units are placed on both sides of the pipe as shown in Figure 1, but also for various types of installation like those shown in Figure 2. This makes it possible to determine the installation configuration that best suits the components to be measured and the application to ensure optimal conditions for measurements.

Figure 2 TDLS200 Installation Configurations

MEASUREMENT PRINCIPLE OF THE TDLS200 LASER GAS ANALYZER

The TDLS200 uses the TDLAS method to measure optical absorption spectra peculiar to molecules ranging from infrared to near-infrared due to vibration and rotation energy transition of molecules in the measured component through a semiconductor laser with an extremely narrow spectral linewidth. The optical absorption spectra peculiar to major molecules such as O2, NH3, H2O, CO and CO2 are in the region from infrared to near-infrared, and measurements of the amount of light absorbed (absorbance) at a specific wavelength make it possible to calculate the concentration of the measured component.

Infrared and Near-infrared Absorption Spectra Due to Molecule Vibration and Rotation

Figure 3 shows an example of molecule vibration and rotation mode (1) while Figure 4 shows a typical CO2 absorption spectrum in the infrared to the near-infrared range. Molecules consisting of two or more atoms absorb light at frequencies specific to the vibration and rotation modes of stretching, scissoring, etc. as shown in Figure 3. The width of absorption band for CO2 shown in Figure 4 is about 10 nm and the spectral linewidth of each absorption peak is approximately 0.05 nm.

Figure 3 Examples of Molecule Vibration and Rotation Modes

Figure 4 CO2 Absorption Spectrum

Component Selectivity of Tunable Diode Laser SpectroscopyFigure 5 shows the absorption spectra for CO2 and

H2O. A wavelength resolution of conventional spectrometers for the infrared to near-infrared region is around 0.3 to 3 nm. This value is equivalent to or larger than the distance between absorption peaks of different gases (0.5 nm or less), making it impossible to resolve gas absorption peaks. This example explains the case measuring concentration of CO2 in the presence of H2O. A conventional spectrometer will be affected by H2O spectrum interference. To reduce the effect of interference, measures such as removing interfering gases must be taken prior to measurement.

Figure 5 CO2 and H2O Absorption Spectra

In contrast, unlike low resolution spectrometers, the TDLS200 uses a laser beam with an extremely narrow spectral linewidth emitted by the tunable diode laser, and its wavelength can be changed by adjusting laser temperature and drive current. This makes it possible to measure a single absorption peak among those in the absorption spectrum like that shown in Figure 5. Thus, as shown in Figure 6, one absorption peak that is not affected by interfering gases can be selected for measurement.

Since wavelength selectivity is high and measurements are not affected by other interfering components, there is no need to remove interfering gases prior to measurement, allowing in-situ measurement of process gas.

36-inch pipe

24-inch pipe

Restriction

36 to 14 inch concentric

reducer

24 to 14 inch concentric

reducer

Minimum opticalpath length36 inches

Minimum opticalpath length36 inches

Minimum opticalpath length36 inches

Minimum opticalpath length36 inches

Symmetricstretching

Antisymmetricstretching

In-planescissoring

In-planerocking

Out-of-planewagging

Out-of-planetwisting

0

0.0005

0.001

0.0015

0.002

0.0025

1.5381.5361.5341.5321.531.528

Abs

orba

nce

[A.U

.]

Wavelength [µm]

1.533 1.5335 1.534 1.5345 1.535 1.5355Wavelength [µm]

Abs

orba

nce

[A.U

.] CO2

0.00050.0010.00150.0020.00250.0030.00350.0040.00450.005

CO2

H2O

The distance between CO2 and H2O peaks is 0.5 nm or less and cannot be resolved by conventional spectrometers

52114

Page 3: Tunable Laser

TDLS200 Tunable Diode Laser Gas Analyzer and its Application to Industrial Process

Yokogawa Technical Report English Edition Vol.53 No.2 (2010)

Figure 6 Absorption Spectrum Measured Using TDLS200

Measurements of Component Concentration Using Peak Area Integration Method

The TDLS200 is capable of measuring an accurate spectrum without being affected by interfering components by sweeping the oscillation wavelength of the semiconductor laser around a single absorption peak of the component to be measured.

Although an absorption spectrum obtained by the TDLS200 isolates it from interfering components, the shape of the spectrum varies depending on the broadening effect caused by the following factors: gas temperature, gas pressure, coexisting gas components. Therefore measurements subject to such environmental changes require compensation. Figure 7 shows the changes of a 0.76 µm-band single absorption spectrum of O2 with a 10% O2 concentration caused by temperature changes, pressure changes and inclusion of CO2, N2 or He as background gases respectively.

The TDLS200 sweeps a wavelength using a semiconductor laser and measure absorption spectra, and calculate component concentration from the spectrum area employing the peak area integration method. Table 1 shows a comparison of the influence of each factor on the peak area integration, peak height and 2f methods (2).

Conventional methods used so far include the peak height method that calculates the concentration of measured components from the peak height of absorption spectra, and the 2f method that calculates the concentration of the measured component from the P-P (Peak to Peak) value of the waveform generated by the frequency doubling modulation from a

wavelength scanning signal. However, all of these methods were greatly affected by changes in temperature, pressure and coexisting gases. In contrast, the peak area integration method is theoretically not affected by changes in coexisting gas components and spectrum area remains constant. And the spectrum area changes linearly with pressure changes in principle.

As shown in Table 1, the effect of temperature, pressure and coexisting gas components on the peak height method and the 2f method are nonlinear, which makes compensation difficult when more than one of these factors are involved. Since the peak area integration method permits linear compensation for gas pressure changes and non-linear compensation for gas temperature changes, it ensures accurate compensation.

Figure 7 Changes in Absorption Spectra Caused by Temperature, Pressure, and Coexisting Gasses

EXAMPLES OF INDUSTRIAL APPLICATIONS … MEASUREMENTS OF RESIDUAL NH3 CONCENTRATION IN FLUE GAS

The TDLS200, thanks to its capability of fast in-situ measurement in actual processes, can be applied to high-speed control where signals indicating component concentration measurements essential for management are directly delivered to process control systems, and can be applied to real-time management of process states as well. The TDLS200 can thus

1.533 1.5335 1.534 1.5345 1.535 1.5355Wavelength [µm]

Abs

orba

nce

[A.U

.] CO2

0.00050.0010.00150.0020.00250.0030.00350.0040.00450.005

CO2

H2O

00.00020.00040.00060.00080.0010.00120.00140.00160.00180.002

150140130120110100908070605040302010

Data no.

Abs

orba

nce

[A.U

.]CO2 absorption peaks not subject to interference with other gases

Select only single CO2 peak not subject to interferencewith other gases

Table 1 Factors Affecting Spectra and Comparison of their Effects

Factors causing changes

Peak area integration method

Peak height method / 2f method

Gas temperature Nonlinear NonlinearGas pressure Linear Nonlinear

Coexisting gas components No effect Nonlinear

0

0.002

0.004

0.006

0.008

0.01

0.012

0.014

30025020015010050Data no.

Abs

orba

nce

[A.U

.] 296 K396 K496 K

0

0.002

0.004

0.006

0.008

0.01

0.012

0.014

30025020015010050Data no.

Abso

rban

ce [A

.U.] 0.5 atm

1 atm1.5 atm

0

0.002

0.004

0.006

0.008

0.01

0.012

0.014

1481411341271201131069992857871645750433629221581

Data no.

Abs

orba

nce

[A.U

.]

10% O2 in CO2

10% O2 in N2

10% O2 in He

(a) Absorption spectrum when temperature changes

(b) Absorption spectrum when pressure changes

(c) Absorption spectrum when other gas is mixed

53 115

Page 4: Tunable Laser

TDLS200 Tunable Diode Laser Gas Analyzer and its Application to Industrial Process

Yokogawa Technical Report English Edition Vol.53 No.2 (2010)

help improve various industrial processes. In this section, we will introduce measurements of residual NH3 concentration in flue gas. Note that optimum combustion control applications have been reported in another Yokogawa technical report (3). Please refer to that report for details.

Ammonia (NH3) gas is injected into flue gas to remove NOx (denitration), increase the efficiency of a dust collector and prevent corrosion. Excess NH3 injection increases the running costs and the amount of residual NH3, resulting in generation of foul odors. Therefore, the amount of NH3 in the exhaust gas must be measured, controlled and monitored. For example, the exhaust gas denitration equipment for a large combustion furnace uses the DeNOx SCR (selective catalytic reduction) process, where NOx is reduced to N2 and H2O by NH3 injection and selective reduction catalysis, and residual NH3 concentration (ppm level) in the flue gasses is measured online.

Conventional NH3 meters that use indirect NOx measuring methods such as the chemiluminescence detection and ion electrode method have slow response and require sample line equipment involving heated pipes to prevent NH3 adhesion and large maintenance costs depending on such a complex measurement system. On the other hand, as shown in Figure 8, the TDLS200 is installed directly into process lines and measures NH3, which substantially improves responsiveness and maintainability. In addition, the highly responsive NH3 concentration signals can be employed in NH3 injection control to optimize NH3 injection.

Figure 8 Use of TDLS200 in Denitration Equipment for Measurement and Injection Control of NH3

Figure 9 shows an installation location examples of the TDLS200 in an exhaust gas denitration process of a boiler.

Figure 9 Installation location examples of TDLS200 in exhaust gas denitration process of a boiler

Figure 10 shows the changes in NH3 injection volume and measurement values of TDSL200 in the case where denitration equipment for improving the efficiency of the dust collector and preventing corrosion is not equipped with; TDSL200 is installed at point A in Figure 9.

Figure 10 NH3 Injection Volume and Changes in Measured Values of TDLS200

CONCLUSION

This paper has presented an overview of the TDLS200 laser gas analyzer and described its features, measurement principle, and applications. The high component selectivity, fast response and ease of maintenance achieved by this technology and product assure its application in a broad range of industrial processes. These include not only use in the NH3 measurement applications introduced in this paper, but also in measurements of CO and O2 in optimum combustion control, measurements of minute amounts of water in electrolytic plants, etc. We expect the TDSL200 to greatly contribute to the preservation of the environment and reduction of running costs through its applications to process control not limited to monitoring.

REFERENCES

(1) The Chemical Society of Japan compiled by, The Fifth Series of Experimental Chemistry Vol. 9 Spectroscopy and Diffraction I, Maruzen Co. Ltd., 2005 in Japanese

(2) Masayuki Satou, “New type of laser gas analyzer technology and its application to industrial processes,” Instrumentation Control Engineering, Vol. 52, No. 2, 2009, pp. 73-76 in Japanese

(3) Yoshitaka Yuki, Akihiro Murata, “Optimum Combustion Control by TDLS200 Tunable Diode Laser Gas Analyzer,” Yokogawa Technical Report English Edition, Vol. 53, No. 1, 2010, pp. 19-22

* TDLS is a trademark of Yokogawa Electric Corporation.

NOx

High-speed measurement of residual NH3

Control of NH3 injection volumeFIC

Denitrationequipment

DeNOxSCR

NH3 injection

TDLS200

Dust collector

Air heater

Denitration equipment

Economizer

(When denitration equipment is not provided)

Point of NH3 injection

Point of NH3 injectionA

BC

Examples of TDLS200installation location

0

2

4

6

8

10

12

15:0014:3014:0013:3013:0012:3012:0011:3011:00

Time

Vol

ume

of in

ject

ed a

mm

onia

[kg/

h]

Volume of injected ammonia

4

6

8

10

12

14

16

TDLS

indi

cate

d va

lue

[ppm

]

TDLS indicated value

-2

0

2

4

6

8

10

16:0015:3015:0014:3014:0013:3013:0012:3012:00

Time

Vol

ume

of in

ject

ed a

mm

onia

[kg/

h]

Volume of injected ammonia

0

2

4

6

8

10

12

TDLS

indi

cate

d va

lue

[ppm

]

TDLS indicated value

14:32 Boiler restart13:14 Boiler stoppage

54116