Dispatch Date : 29-Jan-2018 Page 1 Syphilis serology Distribution : 4217 Laboratory : Operated by Public Health England NIS - Specialist Microbiology Services 133-155 Waterloo Road Wellington House London SE1 8UG © Copyright. The data in UK NEQAS reports are confidential. Participants must consult the scheme organiser before quoting data from the scheme. UK NEQAS for Microbiology PO Box 63003 London NW9 1GH Published at 8:26 on Thursday 12 July 2018 Your Score Intended Result Your Report Specimen 4189 Reagin positive Reagin positive 2 TP Agglutination positive TP Agglutination positive 2 FTA-Abs positive FTA-Abs positive 2 Immunoassay IgG/total Ab positive Immunoassay IgG/total Ab positive 2 Immunoassay IgM positive Immunoassay IgM positive 2 Immunoblot positive Immunoblot positive 2 Specimen 4190 Reagin negative Reagin negative Not scored TP Agglutination positive TP Agglutination positive 2 FTA-Abs positive FTA-Abs positive Not scored Immunoassay IgG/total Ab positive Immunoassay IgG/total Ab positive 2 Immunoassay IgM negative Immunoassay IgM negative 2 Immunoblot positive Immunoblot positive Not scored Specimen 4191 Reagin negative Reagin negative 2 TP Agglutination negative TP Agglutination negative 2 FTA-Abs negative FTA-Abs negative 2 Immunoassay IgG/total Ab negative Immunoassay IgG/total Ab negative 2 Immunoassay IgM negative Immunoassay IgM negative 2 Immunoblot negative Immunoblot negative 2 Specimen 4192 Reagin negative Reagin negative 2 TP Agglutination negative TP Agglutination negative 2 FTA-Abs negative FTA-Abs negative 2 Immunoassay IgG/total Ab negative Immunoassay IgG/total Ab negative 2 Immunoassay IgM negative Immunoassay IgM negative 2 Immunoblot negative Immunoblot negative 2 Specimen 4193 Reagin negative Reagin negative Not scored TP Agglutination positive TP Agglutination positive 2 FTA-Abs positive FTA-Abs positive 2 Immunoassay IgG/total Ab positive Immunoassay IgG/total Ab positive 2 Immunoassay IgM negative Immunoassay IgM negative 2 Immunoblot positive Immunoblot positive 2 Specimen 4194 Reagin negative Reagin negative Not scored TP Agglutination positive TP Agglutination positive 2 FTA-Abs positive FTA-Abs positive 2 Immunoassay IgG/total Ab positive Immunoassay IgG/total Ab positive 2 Immunoassay IgM negative Immunoassay IgM negative 2 Immunoblot positive Immunoblot positive 2 Please note that from this distribution onwards the report will have an additional graph displaying confirmatory test results(Syphilis line blot). Comments: Please check last page (Page 19) for additional comments. Results for each marker by methods used by participants are presented on subsequent pages. Images of the Fujirebio INNO-LIA IgG line assay of the positive specimens can be viewed on: www.ukneqasmicro.org.uk, by logging onto the Results and Reports section and selecting the 'Dist' button for the scheme. Turn around time: The time taken to report your results was 0-days. This information is provided for your own use and does not form part of your performance assessment. Acknowledgements: We thank colleagues at PHE- Public Health Laboratories in Birmingham, Bristol, Manchester and Newcastle; and the Department of Microbiology, Sheffield for assistance with pre-distribution tests. Enquiries: Any queries about this distribution please contact Beatrix Kele at the e-mail address below. Pre-distribution test results are available should you experience a technical failure and wish to discuss the results. For repeat specimens please order using the web form or e-mail [email protected] stating your laboratory identification number, the distribution number and specimen numbers. Report authorised by: Dr. Sanjiv Rughooputh, Scheme Organiser. of 15

Upload: others

Post on 24-Aug-2020

2 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Syphilis serology Laboratory : Distribution : 4217 Page 1 ... · Axis-Shield 18 (26.1) 26 (11.0) Becton Dickinson 1 (1.4) 27 (11.4) bioMerieux 2 (2.9) 36 (15.3) Number of Reports

Dispatch Date : 29-Jan-2018

Page 1

Syphilis serology

Distribution : 4217

Laboratory :

Operated by Public Health England NIS - Specialist Microbiology Services 133-155 Waterloo Road Wellington House London SE1 8UG

© Copyright. The data in UK NEQAS reports are confidential. Participants must consult the scheme organiser before quoting data from the scheme. UK NEQAS for Microbiology PO Box 63003 London NW9 1GH Published at 8:26 on Thursday 12 July 2018

Your Score Intended Result Your Report

Specimen 4189 Reagin positive Reagin positive 2 TP Agglutination positive TP Agglutination positive 2 FTA-Abs positive FTA-Abs positive 2 Immunoassay IgG/total Ab positive Immunoassay IgG/total Ab positive 2 Immunoassay IgM positive Immunoassay IgM positive 2 Immunoblot positive Immunoblot positive 2

Specimen 4190 Reagin negative Reagin negative Not scored TP Agglutination positive TP Agglutination positive 2 FTA-Abs positive FTA-Abs positive Not scored Immunoassay IgG/total Ab positive Immunoassay IgG/total Ab positive 2 Immunoassay IgM negative Immunoassay IgM negative 2 Immunoblot positive Immunoblot positive Not scored

Specimen 4191 Reagin negative Reagin negative 2 TP Agglutination negative TP Agglutination negative 2 FTA-Abs negative FTA-Abs negative 2 Immunoassay IgG/total Ab negative Immunoassay IgG/total Ab negative 2 Immunoassay IgM negative Immunoassay IgM negative 2 Immunoblot negative Immunoblot negative 2

Specimen 4192 Reagin negative Reagin negative 2 TP Agglutination negative TP Agglutination negative 2 FTA-Abs negative FTA-Abs negative 2 Immunoassay IgG/total Ab negative Immunoassay IgG/total Ab negative 2 Immunoassay IgM negative Immunoassay IgM negative 2 Immunoblot negative Immunoblot negative 2

Specimen 4193 Reagin negative Reagin negative Not scored TP Agglutination positive TP Agglutination positive 2 FTA-Abs positive FTA-Abs positive 2 Immunoassay IgG/total Ab positive Immunoassay IgG/total Ab positive 2 Immunoassay IgM negative Immunoassay IgM negative 2 Immunoblot positive Immunoblot positive 2

Specimen 4194 Reagin negative Reagin negative Not scored TP Agglutination positive TP Agglutination positive 2 FTA-Abs positive FTA-Abs positive 2 Immunoassay IgG/total Ab positive Immunoassay IgG/total Ab positive 2 Immunoassay IgM negative Immunoassay IgM negative 2 Immunoblot positive Immunoblot positive 2

Please note that from this distribution onwards the report will have an additional graph displaying confirmatory test results(Syphilis line blot). Comments: Please check last page (Page 19) for additional comments. Results for each marker by methods used by participants are presented on subsequent pages. Images of the Fujirebio INNO-LIA IgG line assay of the positive specimens can be viewed on: www.ukneqasmicro.org.uk, by logging onto the Results and Reports section and selecting the 'Dist' button for the scheme. Turn around time: The time taken to report your results was 0-days. This information is provided for your own use and does not form part of your performance assessment. Acknowledgements: We thank colleagues at PHE- Public Health Laboratories in Birmingham, Bristol, Manchester and Newcastle; and the Department of Microbiology, Sheffield for assistance with pre-distribution tests. Enquiries: Any queries about this distribution please contact Beatrix Kele at the e-mail address below. Pre-distribution test results are available should you experience a technical failure and wish to discuss the results. For repeat specimens please order using the web form or e-mail [email protected] stating your laboratory identification number, the distribution number and specimen numbers. Report authorised by: Dr. Sanjiv Rughooputh, Scheme Organiser.

of 15

Page 2: Syphilis serology Laboratory : Distribution : 4217 Page 1 ... · Axis-Shield 18 (26.1) 26 (11.0) Becton Dickinson 1 (1.4) 27 (11.4) bioMerieux 2 (2.9) 36 (15.3) Number of Reports

Dispatch Date : 29-Jan-2018

Page 2

Syphilis serology

Distribution : 4217

Laboratory :

Operated by Public Health England NIS - Specialist Microbiology Services 133-155 Waterloo Road Wellington House London SE1 8UG

© Copyright. The data in UK NEQAS reports are confidential. Participants must consult the scheme organiser before quoting data from the scheme. UK NEQAS for Microbiology PO Box 63003 London NW9 1GH Published at 8:26 on Thursday 12 July 2018

Cumulative score information Total number of specimens sent to you for UK NEQAS for Syphilis serology over the last 2 distributions is 12 For these distributions specimen numbers 3927 3928 3929 3930 3931 3932 4189 4190 4191 4192 4193 4194 have been sent. Number of specimens reported too late for analysis (not scored) 0 Your cumulative score for the specimen/test combinations that you reported was 122 out of a possible total of 122. The mean score calculated from the reports returned by UK laboratories testing the specimen/test combinations you examined was 119.44 (with a standard error of 2.28). Performance rating Your performance rating for UK NEQAS for Syphilis serology (i.e. the number of standard errors by which your cumulative score lies above or below the mean) for UK laboratories is 1.12. A performance rating of more than 1.96 standard errors below the mean indicates possible poor performance. For multiple analyte schemes no penalty is incurred for non return of reports. However non reporting may be used as a measure of performance. Your performance rating may change if other participants' results are amended.

Per

form

ance

Rat

ing

Distribution

Your performance rating over the past 12 distributions Your current performance rating is 1.12

MICPERF2

3059 3164 3266 3370 3508 3588 3720 3804 3944 4028 4131 4217-4-3-2-101234

All

UK

Tot

al s

core

Distribution

Total score you achieved for each of the last 12 distributionsYour total score for this distribution is 62

MICPERF2X

3059 3164 3266 3370 3508 3588 3720 3804 3944 4028 4131 421720

30

40

50

60

70

80

of 15

Page 3: Syphilis serology Laboratory : Distribution : 4217 Page 1 ... · Axis-Shield 18 (26.1) 26 (11.0) Becton Dickinson 1 (1.4) 27 (11.4) bioMerieux 2 (2.9) 36 (15.3) Number of Reports

Dispatch Date : 29-Jan-2018

Page 3

Syphilis serology

Distribution : 4217

Laboratory :

Operated by Public Health England NIS - Specialist Microbiology Services 133-155 Waterloo Road Wellington House London SE1 8UG

© Copyright. The data in UK NEQAS reports are confidential. Participants must consult the scheme organiser before quoting data from the scheme. UK NEQAS for Microbiology PO Box 63003 London NW9 1GH Published at 8:26 on Thursday 12 July 2018

Six sera (plasma converted by thrombinisation) were issued with requests to report on antibody reactivity indicating the presence of Treponema pallidum (TP). Specimens 4193 and 4194 were pool of two donations. Specimens 4189 and 4190 were single donations of syphilis positive material and specimen 4191 and 4192 were single donations of negative specimens.

Specimen : 4189 Syphilis serology positive Reagin positive

UK (%) All (%)

35 (50.7) 56 (23.7)Biokit RPR Reditest3 (4.3) 4 (1.7)Newmarket0 (0.0) 5 (2.1)Sclavo3 (4.3) 7 (3.0)BioRad1 (1.4) 8 (3.4)Unspecified1 (1.4) 14 (5.9)Plasmatec1 (1.4) 18 (7.6)Omega

18 (26.1) 26 (11.0)Axis-Shield1 (1.4) 27 (11.4)Becton Dickinson2 (2.9) 36 (15.3)bioMerieux

Number of Reports0 12 24 36 48 60

Positive Equivocal Negative

Your Report : Reagin positive

Reagin equivocal

69

Reagin positive

2

Overall Results

236

-1

% Correct

00

99.2

------------------------------------------------------------UK

2340

Total

1Reagin negative

69

All

100

------------------------------------------------------------

2

Score

Your Score : 2

Specimen : 4189 Syphilis serology positive TP Agglutination positive

UK (%) All (%)

48 (80.0) 124 (61.1)Fujirebio : Serodia0 (0.0) 3 (1.5)Far Diagnostics0 (0.0) 3 (1.5)Sclavo1 (1.7) 4 (2.0)Randox0 (0.0) 4 (2.0)Unspecified0 (0.0) 6 (3.0)Plasmatec4 (6.7) 7 (3.4)Newmarket0 (0.0) 8 (3.9)Axis-Shield0 (0.0) 8 (3.9)Omega6 (10.0) 12 (5.9)Biokit: Syphagen

Number of Reports0 30 60 90 120 150

Positive Equivocal Negative

Your Report : TP Agglutination positive

TP Ag. equivocal

60

TP Ag. positive

1

Overall Results

203

-1

% Correct

00

99.5

------------------------------------------------------------UK

2020

Total

1TP Ag. negative

60

All

100

------------------------------------------------------------

2

Score

Your Score : 2

Specimen : 4189 Syphilis serology positive FTA-Abs IgG positive

UK (%) All (%)

1 (100) 3 (14.3)Zeus0 (0.0) 1 (4.8)VIRO-IMMUN 0 (0.0) 2 (9.5)Scimedx0 (0.0) 4 (19.0)Alphadia0 (0.0) 4 (19.0)Unspecified0 (0.0) 7 (33.3)Euroimmun

Number of Reports0 2 4 6 8 10

Positive Equivocal Negative

Your Report : FTA-Abs positive

FTA equivocal

1

FTA positive

0

Overall Results

21

-1

% Correct

00

100

------------------------------------------------------------UK

210

Total

1FTA negative

1

All

100

------------------------------------------------------------

2

Score

Your Score : 2

Operated by Public Health England NIS - Specialist Microbiology Services 133-155 Waterloo Road Wellington House London SE1 8UG

© Copyright. The data in UK NEQAS reports are confidential. Participants must consult the scheme organiser before quoting data from the scheme. UK NEQAS for Microbiology PO Box 63003 London NW9 1GH Published at 8:26 on Thursday 12 July 2018

of 15

Page 4: Syphilis serology Laboratory : Distribution : 4217 Page 1 ... · Axis-Shield 18 (26.1) 26 (11.0) Becton Dickinson 1 (1.4) 27 (11.4) bioMerieux 2 (2.9) 36 (15.3) Number of Reports

Dispatch Date : 29-Jan-2018

Page 4

Syphilis serology

Distribution : 4217

Laboratory :

Six sera (plasma converted by thrombinisation) were issued with requests to report on antibody reactivity indicating the presence of Treponema pallidum (TP). Specimens 4193 and 4194 were pool of two donations. Specimens 4189 and 4190 were single donations of syphilis positive material and specimen 4191 and 4192 were single donations of negative specimens.

Specimen : 4189 Syphilis serology positive Immunoassay IgG/total Ab positive

UK (%) All (%)

15 (9.7) 17 (4.5)Biokit : Bioelisa36 (23.2) 57 (15.2)Roche

2 (1.3) 4 (1.1)Murex : ICE0 (0.0) 5 (1.3)Vircell5 (3.2) 8 (2.1)Trinity1 (0.6) 8 (2.1)BioRad2 (1.3) 8 (2.1)Unspecified

14 (9.0) 27 (7.2)Siemens:ADVIA Centaur11 (7.1) 45 (12.0)DiaSorin : Liaison52 (33.5) 155 (41.2)Abbott : Architect

Number of Reports0 40 80 120 160 200

Positive Equivocal Negative

Your Report : Immunoassay IgG/total Ab positive

Immunoassay equivocal

139

Immunoassay positive

2

Overall Results

342

-1

% Correct

10

99.4

------------------------------------------------------------UK

3400

Total

1Immunoassay negative

138

All

99.3

------------------------------------------------------------

2

Score

Your Score : 2

Specimen : 4189 Syphilis serology positive Immunoassay IgM positive

UK (%) All (%)

10 (35.7) 13 (17.3)Mikrogen recomWell0 (0.0) 2 (2.7)BioRad2 (7.1) 3 (4.0)Captia2 (7.1) 4 (5.3)Lab 210 (0.0) 4 (5.3)Virotech1 (3.6) 5 (6.7)Unspecified0 (0.0) 7 (9.3)Mikrogen recomLine1 (3.6) 7 (9.3)Euroimmun2 (7.1) 9 (12.0)Diesse

10 (35.7) 14 (18.7)Trinity

Number of Reports0 4 8 12 16 20

Positive Equivocal Negative

Your Report : Immunoassay IgM positive

IgM equivocal

28

IgM positive

1

Overall Results

75

-1

% Correct

10

98.7

------------------------------------------------------------UK

740

Total

1IgM negative

27

All

96.4

------------------------------------------------------------

2

Score

Your Score : 2

Specimen : 4189 Syphilis serology positive Immunoblot positive

UK (%) All (%)

9 (90.0) 27 (55.1)Fujirebio INNO-LIA1 (10.0) 1 (2.0)DiaSorin : Liaison0 (0.0) 2 (4.1)Virotech0 (0.0) 3 (6.1)Euroimmun0 (0.0) 5 (10.2)Viramed0 (0.0) 5 (10.2)Mikrogen recomLine0 (0.0) 6 (12.2)Unspecified

Number of Reports0 6 12 18 24 30

Positive Equivocal Negative

Your Report : Immunoblot positive

Immunoblot equivocal

10

Immunoblot positive

0

Overall Results

49

-1

% Correct

00

100

------------------------------------------------------------UK

490

Total

1Immunoblot negative

10

All

100

------------------------------------------------------------

2

Score

Your Score : 2

Operated by Public Health England NIS - Specialist Microbiology Services 133-155 Waterloo Road Wellington House London SE1 8UG

© Copyright. The data in UK NEQAS reports are confidential. Participants must consult the scheme organiser before quoting data from the scheme. UK NEQAS for Microbiology PO Box 63003 London NW9 1GH Published at 8:26 on Thursday 12 July 2018

of 15

Page 5: Syphilis serology Laboratory : Distribution : 4217 Page 1 ... · Axis-Shield 18 (26.1) 26 (11.0) Becton Dickinson 1 (1.4) 27 (11.4) bioMerieux 2 (2.9) 36 (15.3) Number of Reports

Dispatch Date : 29-Jan-2018

Page 5

Syphilis serology

Distribution : 4217

Laboratory :

Six sera (plasma converted by thrombinisation) were issued with requests to report on antibody reactivity indicating the presence of Treponema pallidum (TP). Specimens 4193 and 4194 were pool of two donations. Specimens 4189 and 4190 were single donations of syphilis positive material and specimen 4191 and 4192 were single donations of negative specimens.

Specimen : 4190 Syphilis serology positive Reagin negative

UK (%) All (%)

36 (52.2) 57 (24.1)Biokit RPR Reditest0 (0.0) 5 (2.1)Human Diagnostics0 (0.0) 5 (2.1)Sclavo2 (2.9) 6 (2.5)BioRad1 (1.4) 8 (3.4)Unspecified1 (1.4) 14 (5.9)Plasmatec1 (1.4) 18 (7.6)Omega

18 (26.1) 26 (11.0)Axis-Shield1 (1.4) 27 (11.4)Becton Dickinson2 (2.9) 36 (15.2)bioMerieux

Number of Reports0 12 24 36 48 60

Positive Equivocal Negative

Your Report : Reagin negative

Reagin equivocal

69

Reagin positive

220

Overall Results

237

NS

% Correct

670

92.8

------------------------------------------------------------UK

170

Total

NSReagin negative

2

All

97.1

------------------------------------------------------------

NS

Score

Your Score : Not scored

Specimen : 4190 Syphilis serology positive TP Agglutination positive

UK (%) All (%)

48 (81.4) 124 (61.7)Fujirebio : Serodia0 (0.0) 3 (1.5)Far Diagnostics0 (0.0) 3 (1.5)Sclavo0 (0.0) 3 (1.5)Unspecified1 (1.7) 4 (2.0)Randox0 (0.0) 6 (3.0)Plasmatec4 (6.8) 7 (3.5)Newmarket0 (0.0) 8 (4.0)Axis-Shield0 (0.0) 8 (4.0)Omega6 (10.2) 12 (6.0)Biokit: Syphagen

Number of Reports0 30 60 90 120 150

Positive Equivocal Negative

Your Report : TP Agglutination positive

TP Ag. equivocal

59

TP Ag. positive

17

Overall Results

201

-1

% Correct

03

90.0

------------------------------------------------------------UK

1810

Total

1TP Ag. negative

59

All

100

------------------------------------------------------------

2

Score

Your Score : 2

Specimen : 4190 Syphilis serology positive FTA-Abs IgG positive

UK (%) All (%)

1 (100) 3 (13.6)Zeus0 (0.0) 1 (4.5)VIRO-IMMUN 0 (0.0) 2 (9.1)Scimedx0 (0.0) 4 (18.2)Alphadia0 (0.0) 5 (22.7)Unspecified0 (0.0) 7 (31.8)Euroimmun

Number of Reports0 2 4 6 8 10

Positive Equivocal Negative

Your Report : FTA-Abs positive

FTA equivocal

2

FTA positive

4

Overall Results

22

NS

% Correct

12

72.7

------------------------------------------------------------UK

160

Total

NSFTA negative

1

All

50.0

------------------------------------------------------------

NS

Score

Your Score : Not scored

Operated by Public Health England NIS - Specialist Microbiology Services 133-155 Waterloo Road Wellington House London SE1 8UG

© Copyright. The data in UK NEQAS reports are confidential. Participants must consult the scheme organiser before quoting data from the scheme. UK NEQAS for Microbiology PO Box 63003 London NW9 1GH Published at 8:26 on Thursday 12 July 2018

of 15

Page 6: Syphilis serology Laboratory : Distribution : 4217 Page 1 ... · Axis-Shield 18 (26.1) 26 (11.0) Becton Dickinson 1 (1.4) 27 (11.4) bioMerieux 2 (2.9) 36 (15.3) Number of Reports

Dispatch Date : 29-Jan-2018

Page 6

Syphilis serology

Distribution : 4217

Laboratory :

Six sera (plasma converted by thrombinisation) were issued with requests to report on antibody reactivity indicating the presence of Treponema pallidum (TP). Specimens 4193 and 4194 were pool of two donations. Specimens 4189 and 4190 were single donations of syphilis positive material and specimen 4191 and 4192 were single donations of negative specimens.

Specimen : 4190 Syphilis serology positive Immunoassay IgG/total Ab positive

UK (%) All (%)

15 (9.7) 17 (4.5)Biokit : Bioelisa36 (23.4) 57 (15.2)Roche

2 (1.3) 4 (1.1)Murex : ICE0 (0.0) 5 (1.3)Vircell0 (0.0) 7 (1.9)BioRad2 (1.3) 7 (1.9)Unspecified5 (3.2) 8 (2.1)Trinity

14 (9.1) 27 (7.2)Siemens:ADVIA Centaur11 (7.1) 45 (12.0)DiaSorin : Liaison52 (33.8) 155 (41.4)Abbott : Architect

Number of Reports0 40 80 120 160 200

Positive Equivocal Negative

Your Report : Immunoassay IgG/total Ab positive

Immunoassay equivocal

137

Immunoassay positive

2

Overall Results

340

-1

% Correct

02

98.8

------------------------------------------------------------UK

3360

Total

1Immunoassay negative

137

All

100

------------------------------------------------------------

2

Score

Your Score : 2

Specimen : 4190 Syphilis serology positive Immunoassay IgM negative

UK (%) All (%)

9 (32.1) 12 (16.0)Mikrogen recomWell0 (0.0) 2 (2.7)BioRad2 (7.1) 3 (4.0)Captia2 (7.1) 4 (5.3)Lab 210 (0.0) 4 (5.3)Virotech1 (3.6) 5 (6.7)Unspecified0 (0.0) 7 (9.3)Mikrogen recomLine1 (3.6) 7 (9.3)Euroimmun2 (7.1) 9 (12.0)Diesse

11 (39.3) 15 (20.0)Trinity

Number of Reports0 4 8 12 16 20

Positive Equivocal Negative

Your Report : Immunoassay IgM negative

IgM equivocal

28

IgM positive

73

Overall Results

75

2

% Correct

260

97.3

------------------------------------------------------------UK

20

Total

0IgM negative

2

All

92.9

------------------------------------------------------------

-1

Score

Your Score : 2

Specimen : 4190 Syphilis serology positive Immunoblot positive

UK (%) All (%)

9 (90.0) 27 (56.3)Fujirebio INNO-LIA1 (10.0) 1 (2.1)DiaSorin : Liaison0 (0.0) 2 (4.2)Virotech0 (0.0) 2 (4.2)Euroimmun0 (0.0) 4 (8.3)Viramed0 (0.0) 6 (12.5)Mikrogen recomLine0 (0.0) 6 (12.5)Unspecified

Number of Reports0 6 12 18 24 30

Positive Equivocal Negative

Your Report : Immunoblot positive

Immunoblot equivocal

10

Immunoblot positive

4

Overall Results

48

NS

% Correct

06

79.2

------------------------------------------------------------UK

380

Total

NSImmunoblot negative

10

All

100

------------------------------------------------------------

NS

Score

Your Score : Not scored

Operated by Public Health England NIS - Specialist Microbiology Services 133-155 Waterloo Road Wellington House London SE1 8UG

© Copyright. The data in UK NEQAS reports are confidential. Participants must consult the scheme organiser before quoting data from the scheme. UK NEQAS for Microbiology PO Box 63003 London NW9 1GH Published at 8:26 on Thursday 12 July 2018

of 15

Page 7: Syphilis serology Laboratory : Distribution : 4217 Page 1 ... · Axis-Shield 18 (26.1) 26 (11.0) Becton Dickinson 1 (1.4) 27 (11.4) bioMerieux 2 (2.9) 36 (15.3) Number of Reports

Dispatch Date : 29-Jan-2018

Page 7

Syphilis serology

Distribution : 4217

Laboratory :

Six sera (plasma converted by thrombinisation) were issued with requests to report on antibody reactivity indicating the presence of Treponema pallidum (TP). Specimens 4193 and 4194 were pool of two donations. Specimens 4189 and 4190 were single donations of syphilis positive material and specimen 4191 and 4192 were single donations of negative specimens.

Specimen : 4191 Syphilis serology negative Reagin negative

UK (%) All (%)

19 (48.7) 34 (19.3)Biokit RPR Reditest0 (0.0) 3 (1.7)Immucor/Immunotrep0 (0.0) 3 (1.7)BioRad0 (0.0) 5 (2.8)Sclavo1 (2.6) 7 (4.0)Unspecified1 (2.6) 14 (8.0)Plasmatec0 (0.0) 16 (9.1)Omega

13 (33.3) 20 (11.4)Axis-Shield1 (2.6) 22 (12.5)Becton Dickinson0 (0.0) 27 (15.3)bioMerieux

Number of Reports0 8 16 24 32 40

Positive Equivocal Negative

Your Report : Reagin negative

Reagin equivocal

39

Reagin positive

176

Overall Results

176

2

% Correct

390

100

------------------------------------------------------------UK

00

Total

1Reagin negative

0

All

100

------------------------------------------------------------

-1

Score

Your Score : 2

Specimen : 4191 Syphilis serology negative TP Agglutination negative

UK (%) All (%)

27 (77.1) 83 (55.3)Fujirebio : Serodia0 (0.0) 2 (1.3)Abbott : Architect0 (0.0) 3 (2.0)Fortress0 (0.0) 3 (2.0)Sclavo0 (0.0) 3 (2.0)Randox0 (0.0) 6 (4.0)Plasmatec3 (8.6) 6 (4.0)Newmarket0 (0.0) 8 (5.3)Axis-Shield0 (0.0) 8 (5.3)Omega5 (14.3) 11 (7.3)Biokit: Syphagen

Number of Reports0 20 40 60 80 100

Positive Equivocal Negative

Your Report : TP Agglutination negative

TP Ag. equivocal

35

TP Ag. positive

150

Overall Results

150

2

% Correct

350

100

------------------------------------------------------------UK

00

Total

1TP Ag. negative

0

All

100

------------------------------------------------------------

-1

Score

Your Score : 2

Specimen : 4191 Syphilis serology negative FTA-Abs IgG negative

UK (%) All (%)

1 (100) 3 (15.8)Zeus0 (0.0) 2 (10.5)Scimedx0 (0.0) 4 (21.1)Alphadia0 (0.0) 4 (21.1)Unspecified0 (0.0) 6 (31.6)Euroimmun

Number of Reports0 2 4 6 8 10

Positive Equivocal Negative

Your Report : FTA-Abs negative

FTA equivocal

1

FTA positive

19

Overall Results

19

2

% Correct

10

100

------------------------------------------------------------UK

00

Total

0FTA negative

0

All

100

------------------------------------------------------------

-1

Score

Your Score : 2

Operated by Public Health England NIS - Specialist Microbiology Services 133-155 Waterloo Road Wellington House London SE1 8UG

© Copyright. The data in UK NEQAS reports are confidential. Participants must consult the scheme organiser before quoting data from the scheme. UK NEQAS for Microbiology PO Box 63003 London NW9 1GH Published at 8:26 on Thursday 12 July 2018

of 15

Page 8: Syphilis serology Laboratory : Distribution : 4217 Page 1 ... · Axis-Shield 18 (26.1) 26 (11.0) Becton Dickinson 1 (1.4) 27 (11.4) bioMerieux 2 (2.9) 36 (15.3) Number of Reports

Dispatch Date : 29-Jan-2018

Page 8

Syphilis serology

Distribution : 4217

Laboratory :

Six sera (plasma converted by thrombinisation) were issued with requests to report on antibody reactivity indicating the presence of Treponema pallidum (TP). Specimens 4193 and 4194 were pool of two donations. Specimens 4189 and 4190 were single donations of syphilis positive material and specimen 4191 and 4192 were single donations of negative specimens.

Specimen : 4191 Syphilis serology negative Immunoassay IgG/total Ab negative

UK (%) All (%)

15 (9.9) 17 (4.6)Biokit : Bioelisa35 (23.0) 56 (15.1)Roche

2 (1.3) 4 (1.1)Murex : ICE0 (0.0) 5 (1.3)Vircell0 (0.0) 7 (1.9)BioRad5 (3.3) 8 (2.2)Trinity2 (1.3) 8 (2.2)Unspecified

14 (9.2) 26 (7.0)Siemens:ADVIA Centaur11 (7.2) 45 (12.1)DiaSorin : Liaison52 (34.2) 155 (41.7)Abbott : Architect

Number of Reports0 40 80 120 160 200

Positive Equivocal Negative

Your Report : Immunoassay IgG/total Ab negative

Immunoassay equivocal

138

Immunoassay positive

339

Overall Results

340

2

% Correct

1380

99.7

------------------------------------------------------------UK

10

Total

1Immunoassay negative

0

All

100

------------------------------------------------------------

-1

Score

Your Score : 2

Specimen : 4191 Syphilis serology negative Immunoassay IgM negative

UK (%) All (%)

5 (31.3) 6 (10.5)Mikrogen recomWell2 (12.5) 2 (3.5)Captia0 (0.0) 2 (3.5)BioRad1 (6.3) 3 (5.3)Lab 210 (0.0) 3 (5.3)Virotech1 (6.3) 4 (7.0)Unspecified0 (0.0) 6 (10.5)Mikrogen recomLine0 (0.0) 6 (10.5)Euroimmun0 (0.0) 7 (12.3)Diesse7 (43.8) 11 (19.3)Trinity

Number of Reports0 4 8 12 16 20

Positive Equivocal Negative

Your Report : Immunoassay IgM negative

IgM equivocal

16

IgM positive

57

Overall Results

57

2

% Correct

160

100

------------------------------------------------------------UK

00

Total

0IgM negative

0

All

100

------------------------------------------------------------

-1

Score

Your Score : 2

Specimen : 4191 Syphilis serology negative Immunoblot negative

UK (%) All (%)

4 (80.0) 10 (45.5)Fujirebio INNO-LIA0 (0.0) 1 (4.5)Viramed1 (20.0) 1 (4.5)DiaSorin : Liaison0 (0.0) 2 (9.1)Virotech0 (0.0) 2 (9.1)Euroimmun0 (0.0) 3 (13.6)Mikrogen recomLine0 (0.0) 3 (13.6)Unspecified

Number of Reports0 4 8 12 16 20

Positive Equivocal Negative

Your Report : Immunoblot negative

Immunoblot equivocal

5

Immunoblot positive

22

Overall Results

22

2

% Correct

50

100

------------------------------------------------------------UK

00

Total

1Immunoblot negative

0

All

100

------------------------------------------------------------

-1

Score

Your Score : 2

Operated by Public Health England NIS - Specialist Microbiology Services 133-155 Waterloo Road Wellington House London SE1 8UG

© Copyright. The data in UK NEQAS reports are confidential. Participants must consult the scheme organiser before quoting data from the scheme. UK NEQAS for Microbiology PO Box 63003 London NW9 1GH Published at 8:26 on Thursday 12 July 2018

of 15

Page 9: Syphilis serology Laboratory : Distribution : 4217 Page 1 ... · Axis-Shield 18 (26.1) 26 (11.0) Becton Dickinson 1 (1.4) 27 (11.4) bioMerieux 2 (2.9) 36 (15.3) Number of Reports

Dispatch Date : 29-Jan-2018

Page 9

Syphilis serology

Distribution : 4217

Laboratory :

Six sera (plasma converted by thrombinisation) were issued with requests to report on antibody reactivity indicating the presence of Treponema pallidum (TP). Specimens 4193 and 4194 were pool of two donations. Specimens 4189 and 4190 were single donations of syphilis positive material and specimen 4191 and 4192 were single donations of negative specimens.

Specimen : 4192 Syphilis serology negative Reagin negative

UK (%) All (%)

20 (48.8) 35 (19.7)Biokit RPR Reditest0 (0.0) 3 (1.7)Immucor/Immunotrep1 (2.4) 4 (2.2)BioRad0 (0.0) 5 (2.8)Sclavo1 (2.4) 7 (3.9)Unspecified1 (2.4) 14 (7.9)Plasmatec0 (0.0) 16 (9.0)Omega

13 (31.7) 20 (11.2)Axis-Shield1 (2.4) 22 (12.4)Becton Dickinson0 (0.0) 27 (15.2)bioMerieux

Number of Reports0 8 16 24 32 40

Positive Equivocal Negative

Your Report : Reagin negative

Reagin equivocal

41

Reagin positive

178

Overall Results

178

2

% Correct

410

100

------------------------------------------------------------UK

00

Total

1Reagin negative

0

All

100

------------------------------------------------------------

-1

Score

Your Score : 2

Specimen : 4192 Syphilis serology negative TP Agglutination negative

UK (%) All (%)

28 (75.7) 84 (55.3)Fujirebio : Serodia0 (0.0) 2 (1.3)Abbott : Architect0 (0.0) 3 (2.0)Fortress0 (0.0) 3 (2.0)Sclavo0 (0.0) 3 (2.0)Randox0 (0.0) 6 (3.9)Plasmatec3 (8.1) 6 (3.9)Newmarket0 (0.0) 8 (5.3)Axis-Shield0 (0.0) 8 (5.3)Omega5 (13.5) 11 (7.2)Biokit: Syphagen

Number of Reports0 20 40 60 80 100

Positive Equivocal Negative

Your Report : TP Agglutination negative

TP Ag. equivocal

37

TP Ag. positive

151

Overall Results

152

2

% Correct

360

99.3

------------------------------------------------------------UK

10

Total

1TP Ag. negative

1

All

97.3

------------------------------------------------------------

-1

Score

Your Score : 2

Specimen : 4192 Syphilis serology negative FTA-Abs IgG negative

UK (%) All (%)

1 (100) 3 (15.8)Zeus0 (0.0) 2 (10.5)Scimedx0 (0.0) 4 (21.1)Alphadia0 (0.0) 4 (21.1)Unspecified0 (0.0) 6 (31.6)Euroimmun

Number of Reports0 2 4 6 8 10

Positive Equivocal Negative

Your Report : FTA-Abs negative

FTA equivocal

1

FTA positive

18

Overall Results

19

2

% Correct

10

94.7

------------------------------------------------------------UK

10

Total

0FTA negative

0

All

100

------------------------------------------------------------

-1

Score

Your Score : 2

Operated by Public Health England NIS - Specialist Microbiology Services 133-155 Waterloo Road Wellington House London SE1 8UG

© Copyright. The data in UK NEQAS reports are confidential. Participants must consult the scheme organiser before quoting data from the scheme. UK NEQAS for Microbiology PO Box 63003 London NW9 1GH Published at 8:26 on Thursday 12 July 2018

of 15

Page 10: Syphilis serology Laboratory : Distribution : 4217 Page 1 ... · Axis-Shield 18 (26.1) 26 (11.0) Becton Dickinson 1 (1.4) 27 (11.4) bioMerieux 2 (2.9) 36 (15.3) Number of Reports

Dispatch Date : 29-Jan-2018

Page 10

Syphilis serology

Distribution : 4217

Laboratory :

Six sera (plasma converted by thrombinisation) were issued with requests to report on antibody reactivity indicating the presence of Treponema pallidum (TP). Specimens 4193 and 4194 were pool of two donations. Specimens 4189 and 4190 were single donations of syphilis positive material and specimen 4191 and 4192 were single donations of negative specimens.

Specimen : 4192 Syphilis serology negative Immunoassay IgG/total Ab negative

UK (%) All (%)

15 (9.8) 17 (4.6)Biokit : Bioelisa35 (22.9) 56 (15.0)Roche

2 (1.3) 4 (1.1)Murex : ICE0 (0.0) 5 (1.3)Vircell5 (3.3) 8 (2.1)Trinity1 (0.7) 8 (2.1)BioRad2 (1.3) 8 (2.1)Unspecified

14 (9.2) 27 (7.2)Siemens:ADVIA Centaur11 (7.2) 44 (11.8)DiaSorin : Liaison52 (34.0) 155 (41.6)Abbott : Architect

Number of Reports0 40 80 120 160 200

Positive Equivocal Negative

Your Report : Immunoassay IgG/total Ab negative

Immunoassay equivocal

139

Immunoassay positive

339

Overall Results

342

2

% Correct

1380

99.1

------------------------------------------------------------UK

30

Total

1Immunoassay negative

1

All

99.3

------------------------------------------------------------

-1

Score

Your Score : 2

Specimen : 4192 Syphilis serology negative Immunoassay IgM negative

UK (%) All (%)

5 (31.3) 6 (10.5)Mikrogen recomWell2 (12.5) 2 (3.5)Captia0 (0.0) 2 (3.5)BioRad1 (6.3) 3 (5.3)Lab 210 (0.0) 3 (5.3)Virotech1 (6.3) 4 (7.0)Unspecified0 (0.0) 6 (10.5)Mikrogen recomLine0 (0.0) 6 (10.5)Euroimmun0 (0.0) 7 (12.3)Diesse7 (43.8) 11 (19.3)Trinity

Number of Reports0 4 8 12 16 20

Positive Equivocal Negative

Your Report : Immunoassay IgM negative

IgM equivocal

16

IgM positive

57

Overall Results

57

2

% Correct

160

100

------------------------------------------------------------UK

00

Total

0IgM negative

0

All

100

------------------------------------------------------------

-1

Score

Your Score : 2

Specimen : 4192 Syphilis serology negative Immunoblot negative

UK (%) All (%)

4 (80.0) 10 (50.0)Fujirebio INNO-LIA0 (0.0) 1 (5.0)Viramed0 (0.0) 1 (5.0)Euroimmun1 (20.0) 1 (5.0)DiaSorin : Liaison0 (0.0) 2 (10.0)Virotech0 (0.0) 2 (10.0)Unspecified0 (0.0) 3 (15.0)Mikrogen recomLine

Number of Reports0 4 8 12 16 20

Positive Equivocal Negative

Your Report : Immunoblot negative

Immunoblot equivocal

5

Immunoblot positive

20

Overall Results

20

2

% Correct

50

100

------------------------------------------------------------UK

00

Total

1Immunoblot negative

0

All

100

------------------------------------------------------------

-1

Score

Your Score : 2

Operated by Public Health England NIS - Specialist Microbiology Services 133-155 Waterloo Road Wellington House London SE1 8UG

© Copyright. The data in UK NEQAS reports are confidential. Participants must consult the scheme organiser before quoting data from the scheme. UK NEQAS for Microbiology PO Box 63003 London NW9 1GH Published at 8:26 on Thursday 12 July 2018

of 15

Page 11: Syphilis serology Laboratory : Distribution : 4217 Page 1 ... · Axis-Shield 18 (26.1) 26 (11.0) Becton Dickinson 1 (1.4) 27 (11.4) bioMerieux 2 (2.9) 36 (15.3) Number of Reports

Dispatch Date : 29-Jan-2018

Page 11

Syphilis serology

Distribution : 4217

Laboratory :

Six sera (plasma converted by thrombinisation) were issued with requests to report on antibody reactivity indicating the presence of Treponema pallidum (TP). Specimens 4193 and 4194 were pool of two donations. Specimens 4189 and 4190 were single donations of syphilis positive material and specimen 4191 and 4192 were single donations of negative specimens.

Specimen : 4193 Syphilis serology positive Reagin negative

UK (%) All (%)

35 (50.7) 56 (23.6)Biokit RPR Reditest0 (0.0) 5 (2.1)Human Diagnostics0 (0.0) 5 (2.1)Sclavo3 (4.3) 7 (3.0)BioRad1 (1.4) 8 (3.4)Unspecified1 (1.4) 14 (5.9)Plasmatec1 (1.4) 18 (7.6)Omega

18 (26.1) 26 (11.0)Axis-Shield1 (1.4) 27 (11.4)Becton Dickinson2 (2.9) 36 (15.2)bioMerieux

Number of Reports0 12 24 36 48 60

Positive Equivocal Negative

Your Report : Reagin negative

Reagin equivocal

69

Reagin positive

212

Overall Results

237

NS

% Correct

632

89.5

------------------------------------------------------------UK

230

Total

NSReagin negative

6

All

91.3

------------------------------------------------------------

NS

Score

Your Score : Not scored

Specimen : 4193 Syphilis serology positive TP Agglutination positive

UK (%) All (%)

47 (79.7) 122 (61.0)Fujirebio : Serodia0 (0.0) 3 (1.5)Far Diagnostics0 (0.0) 3 (1.5)Sclavo0 (0.0) 3 (1.5)Unspecified1 (1.7) 4 (2.0)Randox0 (0.0) 6 (3.0)Plasmatec4 (6.8) 7 (3.5)Newmarket0 (0.0) 8 (4.0)Axis-Shield0 (0.0) 8 (4.0)Omega6 (10.2) 12 (6.0)Biokit: Syphagen

Number of Reports0 30 60 90 120 150

Positive Equivocal Negative

Your Report : TP Agglutination positive

TP Ag. equivocal

59

TP Ag. positive

5

Overall Results

200

-1

% Correct

01

97.0

------------------------------------------------------------UK

1940

Total

1TP Ag. negative

59

All

100

------------------------------------------------------------

2

Score

Your Score : 2

Specimen : 4193 Syphilis serology positive FTA-Abs IgG positive

UK (%) All (%)

1 (100) 3 (14.3)Zeus0 (0.0) 1 (4.8)VIRO-IMMUN 0 (0.0) 2 (9.5)Scimedx0 (0.0) 4 (19.0)Alphadia0 (0.0) 4 (19.0)Unspecified0 (0.0) 7 (33.3)Euroimmun

Number of Reports0 2 4 6 8 10

Positive Equivocal Negative

Your Report : FTA-Abs positive

FTA equivocal

1

FTA positive

1

Overall Results

21

-1

% Correct

00

95.2

------------------------------------------------------------UK

200

Total

1FTA negative

1

All

100

------------------------------------------------------------

2

Score

Your Score : 2

Operated by Public Health England NIS - Specialist Microbiology Services 133-155 Waterloo Road Wellington House London SE1 8UG

© Copyright. The data in UK NEQAS reports are confidential. Participants must consult the scheme organiser before quoting data from the scheme. UK NEQAS for Microbiology PO Box 63003 London NW9 1GH Published at 8:26 on Thursday 12 July 2018

of 15

Page 12: Syphilis serology Laboratory : Distribution : 4217 Page 1 ... · Axis-Shield 18 (26.1) 26 (11.0) Becton Dickinson 1 (1.4) 27 (11.4) bioMerieux 2 (2.9) 36 (15.3) Number of Reports

Dispatch Date : 29-Jan-2018

Page 12

Syphilis serology

Distribution : 4217

Laboratory :

Six sera (plasma converted by thrombinisation) were issued with requests to report on antibody reactivity indicating the presence of Treponema pallidum (TP). Specimens 4193 and 4194 were pool of two donations. Specimens 4189 and 4190 were single donations of syphilis positive material and specimen 4191 and 4192 were single donations of negative specimens.

Specimen : 4193 Syphilis serology positive Immunoassay IgG/total Ab positive

UK (%) All (%)

15 (9.7) 17 (4.5)Biokit : Bioelisa36 (23.2) 57 (15.1)Roche

2 (1.3) 4 (1.1)Murex : ICE0 (0.0) 5 (1.3)Vircell5 (3.2) 8 (2.1)Trinity1 (0.6) 8 (2.1)BioRad2 (1.3) 9 (2.4)Unspecified

14 (9.0) 27 (7.2)Siemens:ADVIA Centaur11 (7.1) 45 (11.9)DiaSorin : Liaison52 (33.5) 155 (41.1)Abbott : Architect

Number of Reports0 40 80 120 160 200

Positive Equivocal Negative

Your Report : Immunoassay IgG/total Ab positive

Immunoassay equivocal

139

Immunoassay positive

3

Overall Results

342

-1

% Correct

10

99.1

------------------------------------------------------------UK

3390

Total

1Immunoassay negative

138

All

99.3

------------------------------------------------------------

2

Score

Your Score : 2

Specimen : 4193 Syphilis serology positive Immunoassay IgM negative

UK (%) All (%)

9 (32.1) 12 (16.0)Mikrogen recomWell0 (0.0) 2 (2.7)BioRad2 (7.1) 3 (4.0)Captia2 (7.1) 4 (5.3)Lab 210 (0.0) 4 (5.3)Virotech1 (3.6) 6 (8.0)Euroimmun1 (3.6) 6 (8.0)Unspecified0 (0.0) 7 (9.3)Mikrogen recomLine2 (7.1) 9 (12.0)Diesse

11 (39.3) 15 (20.0)Trinity

Number of Reports0 4 8 12 16 20

Positive Equivocal Negative

Your Report : Immunoassay IgM negative

IgM equivocal

28

IgM positive

69

Overall Results

75

2

% Correct

262

92.0

------------------------------------------------------------UK

41

Total

0IgM negative

1

All

92.9

------------------------------------------------------------

-1

Score

Your Score : 2

Specimen : 4193 Syphilis serology positive Immunoblot positive

UK (%) All (%)

9 (90.0) 26 (53.1)Fujirebio INNO-LIA1 (10.0) 1 (2.0)DiaSorin : Liaison0 (0.0) 2 (4.1)Virotech0 (0.0) 3 (6.1)Euroimmun0 (0.0) 5 (10.2)Viramed0 (0.0) 6 (12.2)Mikrogen recomLine0 (0.0) 6 (12.2)Unspecified

Number of Reports0 6 12 18 24 30

Positive Equivocal Negative

Your Report : Immunoblot positive

Immunoblot equivocal

10

Immunoblot positive

3

Overall Results

49

-1

% Correct

03

87.8

------------------------------------------------------------UK

430

Total

1Immunoblot negative

10

All

100

------------------------------------------------------------

2

Score

Your Score : 2

Operated by Public Health England NIS - Specialist Microbiology Services 133-155 Waterloo Road Wellington House London SE1 8UG

© Copyright. The data in UK NEQAS reports are confidential. Participants must consult the scheme organiser before quoting data from the scheme. UK NEQAS for Microbiology PO Box 63003 London NW9 1GH Published at 8:26 on Thursday 12 July 2018

of 15

Page 13: Syphilis serology Laboratory : Distribution : 4217 Page 1 ... · Axis-Shield 18 (26.1) 26 (11.0) Becton Dickinson 1 (1.4) 27 (11.4) bioMerieux 2 (2.9) 36 (15.3) Number of Reports

Dispatch Date : 29-Jan-2018

Page 13

Syphilis serology

Distribution : 4217

Laboratory :

Six sera (plasma converted by thrombinisation) were issued with requests to report on antibody reactivity indicating the presence of Treponema pallidum (TP). Specimens 4193 and 4194 were pool of two donations. Specimens 4189 and 4190 were single donations of syphilis positive material and specimen 4191 and 4192 were single donations of negative specimens.

Specimen : 4194 Syphilis serology positive Reagin negative

UK (%) All (%)

36 (51.4) 57 (23.9)Biokit RPR Reditest0 (0.0) 5 (2.1)Human Diagnostics0 (0.0) 5 (2.1)Sclavo3 (4.3) 7 (2.9)BioRad1 (1.4) 8 (3.4)Unspecified1 (1.4) 14 (5.9)Plasmatec1 (1.4) 18 (7.6)Omega

18 (25.7) 26 (10.9)Axis-Shield1 (1.4) 27 (11.3)Becton Dickinson2 (2.9) 36 (15.1)bioMerieux

Number of Reports0 12 24 36 48 60

Positive Equivocal Negative

Your Report : Reagin negative

Reagin equivocal

70

Reagin positive

213

Overall Results

238

NS

% Correct

661

89.5

------------------------------------------------------------UK

240

Total

NSReagin negative

4

All

94.3

------------------------------------------------------------

NS

Score

Your Score : Not scored

Specimen : 4194 Syphilis serology positive TP Agglutination positive

UK (%) All (%)

48 (80.0) 123 (61.2)Fujirebio : Serodia0 (0.0) 3 (1.5)Far Diagnostics0 (0.0) 3 (1.5)Sclavo0 (0.0) 3 (1.5)Unspecified1 (1.7) 4 (2.0)Randox0 (0.0) 6 (3.0)Plasmatec4 (6.7) 7 (3.5)Newmarket0 (0.0) 8 (4.0)Axis-Shield0 (0.0) 8 (4.0)Omega6 (10.0) 12 (6.0)Biokit: Syphagen

Number of Reports0 30 60 90 120 150

Positive Equivocal Negative

Your Report : TP Agglutination positive

TP Ag. equivocal

60

TP Ag. positive

2

Overall Results

201

-1

% Correct

00

99.0

------------------------------------------------------------UK

1990

Total

1TP Ag. negative

60

All

100

------------------------------------------------------------

2

Score

Your Score : 2

Specimen : 4194 Syphilis serology positive FTA-Abs IgG positive

UK (%) All (%)

1 (100) 3 (14.3)Zeus0 (0.0) 1 (4.8)VIRO-IMMUN 0 (0.0) 2 (9.5)Scimedx0 (0.0) 4 (19.0)Alphadia0 (0.0) 4 (19.0)Unspecified0 (0.0) 7 (33.3)Euroimmun

Number of Reports0 2 4 6 8 10

Positive Equivocal Negative

Your Report : FTA-Abs positive

FTA equivocal

1

FTA positive

1

Overall Results

21

-1

% Correct

00

95.2

------------------------------------------------------------UK

200

Total

1FTA negative

1

All

100

------------------------------------------------------------

2

Score

Your Score : 2

Operated by Public Health England NIS - Specialist Microbiology Services 133-155 Waterloo Road Wellington House London SE1 8UG

© Copyright. The data in UK NEQAS reports are confidential. Participants must consult the scheme organiser before quoting data from the scheme. UK NEQAS for Microbiology PO Box 63003 London NW9 1GH Published at 8:26 on Thursday 12 July 2018

of 15

Page 14: Syphilis serology Laboratory : Distribution : 4217 Page 1 ... · Axis-Shield 18 (26.1) 26 (11.0) Becton Dickinson 1 (1.4) 27 (11.4) bioMerieux 2 (2.9) 36 (15.3) Number of Reports

Dispatch Date : 29-Jan-2018

Page 14

Syphilis serology

Distribution : 4217

Laboratory :

Six sera (plasma converted by thrombinisation) were issued with requests to report on antibody reactivity indicating the presence of Treponema pallidum (TP). Specimens 4193 and 4194 were pool of two donations. Specimens 4189 and 4190 were single donations of syphilis positive material and specimen 4191 and 4192 were single donations of negative specimens.

Specimen : 4194 Syphilis serology positive Immunoassay IgG/total Ab positive

UK (%) All (%)

15 (9.7) 17 (4.5)Biokit : Bioelisa36 (23.2) 57 (15.2)Roche

2 (1.3) 4 (1.1)Murex : ICE0 (0.0) 5 (1.3)Vircell5 (3.2) 8 (2.1)Trinity1 (0.6) 8 (2.1)BioRad2 (1.3) 8 (2.1)Unspecified

14 (9.0) 27 (7.2)Siemens:ADVIA Centaur11 (7.1) 44 (11.7)DiaSorin : Liaison52 (33.5) 155 (41.3)Abbott : Architect

Number of Reports0 40 80 120 160 200

Positive Equivocal Negative

Your Report : Immunoassay IgG/total Ab positive

Immunoassay equivocal

139

Immunoassay positive

1

Overall Results

342

-1

% Correct

00

99.7

------------------------------------------------------------UK

3410

Total

1Immunoassay negative

139

All

100

------------------------------------------------------------

2

Score

Your Score : 2

Specimen : 4194 Syphilis serology positive Immunoassay IgM negative

UK (%) All (%)

9 (32.1) 12 (16.0)Mikrogen recomWell0 (0.0) 2 (2.7)BioRad2 (7.1) 3 (4.0)Captia2 (7.1) 4 (5.3)Lab 210 (0.0) 4 (5.3)Virotech1 (3.6) 5 (6.7)Unspecified0 (0.0) 7 (9.3)Mikrogen recomLine1 (3.6) 7 (9.3)Euroimmun2 (7.1) 9 (12.0)Diesse

11 (39.3) 15 (20.0)Trinity

Number of Reports0 4 8 12 16 20

Positive Equivocal Negative

Your Report : Immunoassay IgM negative

IgM equivocal

28

IgM positive

67

Overall Results

75

2

% Correct

273

89.3

------------------------------------------------------------UK

50

Total

0IgM negative

1

All

96.4

------------------------------------------------------------

-1

Score

Your Score : 2

Specimen : 4194 Syphilis serology positive Immunoblot positive

UK (%) All (%)

9 (90.0) 26 (54.2)Fujirebio INNO-LIA1 (10.0) 1 (2.1)DiaSorin : Liaison0 (0.0) 2 (4.2)Virotech0 (0.0) 3 (6.3)Euroimmun0 (0.0) 5 (10.4)Viramed0 (0.0) 5 (10.4)Mikrogen recomLine0 (0.0) 6 (12.5)Unspecified

Number of Reports0 6 12 18 24 30

Positive Equivocal Negative

Your Report : Immunoblot positive

Immunoblot equivocal

10

Immunoblot positive

3

Overall Results

48

-1

% Correct

00

93.8

------------------------------------------------------------UK

450

Total

1Immunoblot negative

10

All

100

------------------------------------------------------------

2

Score

Your Score : 2

Operated by Public Health England NIS - Specialist Microbiology Services 133-155 Waterloo Road Wellington House London SE1 8UG

© Copyright. The data in UK NEQAS reports are confidential. Participants must consult the scheme organiser before quoting data from the scheme. UK NEQAS for Microbiology PO Box 63003 London NW9 1GH Published at 8:26 on Thursday 12 July 2018

of 15

Page 15: Syphilis serology Laboratory : Distribution : 4217 Page 1 ... · Axis-Shield 18 (26.1) 26 (11.0) Becton Dickinson 1 (1.4) 27 (11.4) bioMerieux 2 (2.9) 36 (15.3) Number of Reports

Comments on distribution 4217

Comments

Overall performance was very good with correct reports for scored specimens of 99.7% for reagin, 97.5% for TP Agglutination, 97% for FTA, 99.3% for Immunoassay IgG/total, 96.3% for Immunoblot and 96.2% for Immunoassay IgM.

Specimens 4190, 4193 and 4194 Reagin, not scored Experience has shown that treponemal antibody positive specimens with reagin titres of less

than 1:4 may be reported as negative by a significant number of participants, due to variation in test sensitivity, local reporting practice, and to the normal statistical variation that is likely to occur. Only those specimens having reagin titres of 1:4 or more during pre-distribution

testing will be included in scoring (Guidance from the National Quality Assurance Advisory Panel).

TP agglutination titres:

TP agglutination titres were reported as follows: Specimen 4189: median titre 1:2560 (range 1:5 to 1:20480), 167 labs

Specimen 4190: median titre 1:160 (range 1:1 to 1:1280), 155 labs Specimen 4193: median titre 1:160 (range 1:1 to 1:2560), 162 labs Specimen 4194: median titre 1:640 (range 1:2 to 1:10240), 165 labs

Specimens 4191 and 4192 were negative for treponemal antibodies.

Page 15

Syphilis serology

Distribution : 4217

Laboratory :

Operated by Public Health England NIS - Specialist Microbiology Services 133-155 Waterloo Road Wellington House London SE1 8UG

© Copyright. The data in UK NEQAS reports are confidential. Participants must consult the scheme organiser before quoting data from the scheme. UK NEQAS for Microbiology PO Box 63003 London NW9 1GH Published at 8:26 on Thursday 12 July 2018

of 15