SPL Short Cryomodule Design. Mock-up status, pressure relief devices and vacuum vessel mechanical calculations. P. Azevedo (CERN-TE/MSC) SPL Internal Meeting, 07/12/2012. Summary. Overview of cavity supporting system Status of supporting system mock-up - PowerPoint PPT Presentation

TRANSCRIPT

Page 1: SPL Short  Cryomodule  Design

SPL Short Cryomodule Design

P. Azevedo (CERN-TE/MSC)

SPL Internal Meeting, 07/12/2012

Mock-up status, pressure relief devices and vacuum vessel mechanical calculations

Page 2: SPL Short  Cryomodule  Design

Summary

1. Overview of cavity supporting system

2. Status of supporting system mock-up

3. Pressure relief devices: introduction

4. Bursting discs for LHe 2K volume

5. Vacuum vessel relief plate

6. Lower protection levels (for process and vacuum volumes)

7. Vacuum vessel mechanical calculations

SPL Internal Meeting, 07/12/2012 2 / 34

Page 3: SPL Short  Cryomodule  Design

1. Overview of cavity supporting system

3 / 34

• The power coupler double tube acts as vertical support and longitudinal positioner

• The design is simplified

• Better thermal performance - less heat conduction paths from room temperature

Supporting concept: Power coupler double tube as support

SPL Internal Meeting, 07/12/2012

Page 4: SPL Short  Cryomodule  Design

1. Overview of cavity supporting system

4 / 34

TOLERANCES BUDGET FOR SPL CRYOMODULE

Step Description Position

tolerance of every cavity

(3σ)

Cumulative tolerances

1_0 Cavities delivery to CERN ±0.4 mm ±0.4 mm

6_1 Assembling tuner outside clean room. ±0.1mm (?) ±1.6mm (?)

1 Alignment string of He vessels under assembly girder ± 0.1 mm ± 0.1 mm

2 Load transfer to vac.vessel ± 0.3 mm ± 0.4 mm

3 Re-alignment via vessel flange screws adjustment ± 0.2 mm ± 0.2 mm

4 Mechanical mounting of top lid ± 0.3 mm ± 0.5 mm

5 Pump down ± 0.1 mm ± 0.6 mm

6 Transport & Handling ± 0.1 mm (?) ± 0.7 mm (?)

7 Cryostat @ cold (nominal operating T) ± 0.1 mm (?) ± 0.8 mm (?)

8 RF power on < ± 0.1 mm < ± 0.9 mm (?)

9 CD/WU cycles ± 0.1 mm (?) < ± 1.0 mm (?)

Sum of tolerances (mm)

Arithmetic Quadratic

Cavity / He tank 1.6 0.55

Cryomodule 1 0.42

Total 2.6 0.69

1.2 mm is an acceptable value for the cavities misalignment (3σ)*

*Summary of the 4th SPL SCM Working Group meeting held on 26/06/2012; R.Bonomi

Alignment tolerances

Cryomodule tolerances, V. ParmaSPL Internal Meeting, 07/12/2012

Page 5: SPL Short  Cryomodule  Design

5 / 34

1. Overview of cavity supporting system

Inter-cavity support

Helium tankInterface with vacuum vessel

Double tube of the power couplerVacuum vessel

SPL Internal Meeting, 07/12/2012

Page 6: SPL Short  Cryomodule  Design

6 / 34

1. Overview of cavity supporting system

Spherical joint

Sliding cylinder

He vessels flanges

241 mm

Deformable “triangle”

For more information on the design of these components and calculation notes concerning the supporting system, check the SPL workspace (references)..

Vacuum vessel / double tube interface and inter-cavity support Courtesy of P.Duthil / S.Rousselot (CNRS/IPNO)

SPL Internal Meeting, 07/12/2012

Page 7: SPL Short  Cryomodule  Design

7 / 34

2. Status of supporting system mock-up

• Validate the supporting and alignment concept

• Test critical components of unknown behaviour, the interface with vacuum vessel and the inter-cavity support, during assembly and cool-down

• Validate thermal calculations – namely the thermal model of actively cooled double

• Learn about alignment survey methods and other measurements relevant for the SPL short cryomodule

Flow of GN2

Filled with LN2

Mock-up developed by J-B. Deschamps, A. Vande Craen, R. Bonomi and P. Azevedo , in collaboration with different CERN groups. For more information, check the SLHiPP2 meeting presentation Mock-ups of the SPL cavity supporting system

Introduction

SPL Internal Meeting, 07/12/2012

Page 8: SPL Short  Cryomodule  Design

8 / 34

2. Status of supporting system mock-up

Optical wire positioning monitor (stretched wire) will be installed in a second phase

Instrumentation Scheme:

SPL Internal Meeting, 07/12/2012

Page 9: SPL Short  Cryomodule  Design

9 / 34

2. Status of supporting system mock-up

• Design is finished – all components, cryogenic equipment and sensors have been defined / ordered

• Vacuum vessel has been manufactured

• Interfaces with vacuum vessel welded to double tubes (EBW)

• Cold mass (LN2 tanks) ready in a couple of weeks

• Assembly and instrumentation process defined

• Assembly and first alignment measurements: December 2012 / January 2013

• Cool down and first cold tests: first months of 2013

SPL Internal Meeting, 07/12/2012

Courtesy of J-B. Deschamps

Page 10: SPL Short  Cryomodule  Design

10 / 34

2. Status of supporting system mock-up

Courtesy of E. Rigutto

Courtesy of E. Rigutto

SPL Internal Meeting, 07/12/2012

Page 11: SPL Short  Cryomodule  Design

3. Pressure relief devices: IntroductionPressure / Temperature table:

Line Description

Pipe Size (ID min value)

Normal operating pressure

Normal operatin

g temperat

ure

Cool-down /

warm-up pressure

Cool-down /

warm-up temperat

ure

T rangeMaximum operating pressure

Maximum pressure in case of MCI

Design pressure

Test pressure Comment

[mm] [MPa] [T] [MPa] [K] [K] [MPa] [MPa] [MPa] [MPa]

Z cavity/beam vacuum N.A. I.P. 10-9

mbar (tbc) 2 N.A. N.A. 2-293 N.A. 0.2 @ 2K 0.15 @ 293K N.A.design pressure limited by cavity

plastic deformation

L Cavity-helium vessel enclosure

cavity OD + 10 0.0031 2

0.13 @ 293K

293-2 2-293

0.15 @ 293K;

0.2 @ 2K 0.15 @ 293K N.A.design pressure limited by cavity

plastic deformation0.2 @ 2K 0.2 @ 2K (tbc)

X Bi-phase pipe 100 0.0031 2

0.13 @ 293K

293-2 2-293

0.15 @ 293K

0.2 @ 2K 0.15 @ 293K N.A. "0.2 @ 2K 0.2 @ 2K

(tbc)

Y Cavity top connection 80 0.0031 2

0.13 @ 293K

293-2 2-293

0.15 @ 293K

0.2 @ 2K 0.15 @ 293K N.A. "0.2 @ 2K 0.2 @ 2K

(tbc)

XB Pumping line 80 0.0031 2

0.13 @ 293K

293-2 2-293

0.15 @ 293K

0.2 @ 2K 0.15 @ 293K N.A. "0.2 @ 2K 0.2 @ 2K

(tbc)

E Thermal shield supply 15 1.8 ~50 2 293-50 50-293 2 N.A. 2 2.5 Heat intercept

E’ Thermal shield return 15 1.8 ~50 2 293-50 50-293 2 N.A. 2 2.5 Return only

W Cryostat vacuum vessel TBD I.P. 10-

6mbar 293 vacuum 293 237-293 O.P. 0.1 I.P. 0.15 @237K

O.P. 0.1 @ 293K; I.P. 0.15 @237K

N.A.  

C/C1 Cavity filling 6 0.1 4.5 0.1 293-4.5 4.5-293 0.15 @4.5K N.A. 0.15 @ 293K N.A. Liquid supply

C2 Coupler cooling 6 0.1 4.5-293 0.1 293-4.5 4.5-293 0.15 @ 4.5K N.A. 0.15 @ 293 K N.A. Gaseous supply

C3 Cavity top supply 10 0.1 2 0.1 293-4.5 2-293 0.15 @ 4.5K N.A. 0.15 @ 293 K N.A. Liquid supply

V. Parma; SPL Pressure / Temperature Table 11 / 34 SPL Internal Meeting, 07/12/2012

Page 12: SPL Short  Cryomodule  Design

Cryogenic Scheme:

O. Pirotte; SM18 PID SPL Bunker

3. Pressure relief devices: Introduction

12 / 34 SPL Internal Meeting, 07/12/2012

Page 13: SPL Short  Cryomodule  Design

Risks overview:

Hazard Cause / Component Consequences Probability / Frequence

Control measures related to pressure

relief

Leak to vacuum vessel

Thermal cyclesMechanical loadsCorrosion

WeldsBellowsPipeFlanges

Pressure increase in vacuum vesselLoss of insulating vacuum (heat load)

LowVacuum vessel relief plate; Ps = 0.5 barg (design pressure)

Temporary (small) leak of air into vacuum volume

Vaporization of condensed air

HighVacuum vessel relief valve; Ps < 0.5 barg

Pressure increase

Loss of insulating vacuum

Pressure increase in 2K circuit

Low2K circuit bursting discs; Ps = 0.5 barg (design pressure)

Loss of beam vacuum

Overpressure in cryogenic supply

Return pipe blocked

Power failure Static heat loads not compensated

High2K circuit relief valve; Ps < 0.5 barg

3. Pressure relief devices: Introduction

13 / 34 SPL Internal Meeting, 07/12/2012

Work in progress

Page 14: SPL Short  Cryomodule  Design

4. Bursting discs for LHe 2K volume

14 / 34

Heat input due to loss of beam vacuum: different tests, different results• LHe bath cooled Nb deflector: 3 mm thick; 30 mm opening; w = 18 kW/m2• LEP cavity: 25 mm opening; 120 g/s; w = 10 kW/m2

80 mm opening; 1200 g/s, w = 40 kW/m2  • XFEL cryomodule: beam pipe opening; w = 23 kW/m2 (+/- 50 % uncertainty); w = 14.2 kW/m2

(+/- 10 % uncertainty) – different heat load estimation methods • Work in progress: these values are the result of tests carried out with different equipment and in

different conditions. A proper understanding of the geometrical and physical parameters is required before estimations can be made for the SPL cryomodule (cavity geometry, venting diameter, relief devices set pressure, peak pressure)

Experimental tests of fault conditions during the cryogenic operation of a XFEL prototype cryomodule; Boeckmann et al;

Safety aspects for LHe cryostats and LHe transport containers; Lehmann, Zahn; (1978)

Pressure protection against vacuum failures on the cryostats for LEP SC cavities; Cavallari et al; 1989

SPL Internal Meeting, 07/12/2012

Page 15: SPL Short  Cryomodule  Design

4. Bursting discs for LHe 2K volume

15 / 34

Heat input: loss of beam vacuum (no insulation) Estimate

Table by C. Parente (DGS-SEE-XP) based on multiple sources; from calculation sheet developed by A. Henriques (DGS-SEE-XP)

For the moment, the safety experts at CERN recommend 38 kW / m2

W. Lehmann

Heat input determines mass flow:

EN 13648-3

For an overpressure of 10% (Prelief = 1.55 bara):

W/S (kW/m2) W (kW) Qm (kg/s)

38 266 12

20 140 6.3 for 0.4*Pcrit < Prelief < Pcrit

SPL Internal Meeting, 07/12/2012

Page 16: SPL Short  Cryomodule  Design

4. Bursting discs for LHe 2K volume

16 / 34

Heat input: loss of insulating vacuum (with MLI) Estimate

For an overpressure of 10% (Prelief = 1.55 bara):

W/S (kW/m2) W (kW) Qm (kg/s)

6 48 2.2

For comparison purposes, an estimation of the heat input and relief mass flow in the event of loss of insulation vacuum (not the dimensioning scenario) was carried out:

SPL Internal Meeting, 07/12/2012

Page 17: SPL Short  Cryomodule  Design

4. Bursting discs for LHe 2K volume

17 / 34

Relieving temperature and bursting disc(s) discharge coefficient

EN ISO 4126-6

• The saturation temperature at the relieving pressure is 4.7 K. A value of 5 K was taken as relieving temperature, based on U. Wagner’s* initial estimates (conservative)

• A discharge coefficient (α) of 0.73 was taken – depending on final design, this value may be conservative

*U. Wagner; Cryogenic scheme, pipes and valves dimensions; SPL Conceptual Design Review; 04/11/2011

SPL Internal Meeting, 07/12/2012

Page 18: SPL Short  Cryomodule  Design

4. Bursting discs for LHe 2K volume

18 / 34

Results: Sizing of bursting discs

w/S (kw/m2)

No. Burst. Discs Dmin (mm)

201 93

2 65

381 127

2 90

• Formula units are not always consistent with the units presented

• Kb is a correction factor for subcritical flow (function of the isentropic expansion coefficient k and pressure ratio)

• C is a function of k

EN ISO 4126-6

Heat input value should be clarified before final design decisions

Also affecting the design of the cryomodule: “The liquid container shall be protected against overpressure by a minimum of two relief devices in parallel, preferably of different types”*

*Safety instruction IS 47: The use of cryogenics fluids ; CERN EDMS doc. 335812, by the Safety Comission

SPL Internal Meeting, 07/12/2012

Work in progress – preliminary results

Page 19: SPL Short  Cryomodule  Design

4. Bursting discs for LHe 2K volume

19 / 34

Pressure drop limits and cryomodule design

w/S (kw/m2)

No. Burst. Discs* ΔP (mbar)** ΔP/Ps (%)

201 16 3

2 3 1

381 59 12

2 10 2

* For 2 bursting discs, these are considered to be placed on opposite ends of the bi-phase pipe.** Pressure drop along the bi-phase pipe: no local pressure drops considered

• The pressure drop along the bi-phase pipe is significant - the total pressure drop should be limited to 3% of Ps (EN 13648-3) – 15 mbar

• Local pressure drops have to be determined: since the dynamic/ velocity pressure of the discharged mass flow is 119 mbar, for an heat input of 20 kW/m2, we are limited to very small local pressure drop coefficients

• An additional problem is the fact that only part of the bi-phase pipe will constitute a “free relief path” for a zero slope configuration (common LHe bath as opposed to the “roman fountain” solution for a positive slope)

Courtesy of P. Duthil, S. Rousselot et al, CNRS / IPNO

SPL Internal Meeting, 07/12/2012

Work in progress – preliminary results

Page 20: SPL Short  Cryomodule  Design

5. Vacuum vessel relief plate

20 / 34

Methodology – there is no “standard” method; 2 different methods were used:

A) An orifice in the 2K LHe circuit causes a discharge of LHe into the vacuum vessel (incompressible fluid). This mass flow, which depends on the orifice diameter, is the mass flow discharged by the relief plate, at subcritical flow and higher Trelief.

• Highly dependent upon orifice hole and Trelief • Turns a highly transient phenomenon (LHe

release into the vacuum) into a steady state process

B) Complete rupture of the 2K LHe enclosure: the vacuum vessel becomes a non-insulated cryostat. The heat load to the helium volume causes a discharge through the relief plate

• The He density in the vacuum vessel is lower than the saturated vapour density at the relief pressure – mass flow calculation is not trivial

For both cases, the process volume relief devices (same set pressure) are ignored – conservative assumption

SPL Internal Meeting, 07/12/2012

Page 21: SPL Short  Cryomodule  Design

0 5 10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90 95 1000.0

100.0

200.0

300.0

400.0

500.0

Tsat20 K70 K

Which relief temperature?

Complete rupture of C3 line bellows

5. Vacuum vessel relief plate

21 / 34

Method A

Complete rupture of bi-phase pipe bellows

1st Step - incompressible flow through an orifice, into vacuum:

A

• A is the area of the orifice• Kd is the orifice coefficient of discharge;

Kd=0.62 (HSE recommendation)• P is the pressure in the process volume; P=1.5 bar (design pressure)

• Kd,RP is the coefficient of discharge of the relief plate; Kd=0.73 was taken

• Y is the expansion factor for the He vapour• ARP is the area of the relief plate orifice• PV is the relief pressure (1.5 bara)• Pb is the back pressure (atmosphere)

2nd Step - compressible and subcritical flow through vacuum vessel relief plate:

DRP (mm) vs Dorifice (mm)

For the moment we can assume a 10 mm diameter as “reasonable”. This value corresponds to the complete rupture of the line C3, and to a “reasonable size for an hypothetical orifice in the bi-phase pipe bellows

SPL Internal Meeting, 07/12/2012

Work in progress – preliminary results

Page 22: SPL Short  Cryomodule  Design

5. Vacuum vessel relief plate

22 / 34

Method B

Due to large vacuum vessel volume, helium density is lower than density of saturated vapour at relief pressure. Two hypothesis:

1) Vaporization of LHe mass correspondent to cryomodule volume (identical to the 2K LHe relief flow presented before)

2) Transient heat conduction to the Ghe mass filling the vacuum vessel; mass flow correspondent to density decrease due to temperature increase

Which relief temperature?

1) 2)W (kW) 184 59

Qm (kg/s) 8.2 0.2

DRP (mm)"Tsat" 103 21

20 K 169 26

70 K 232 36

Calculation by R. Bonomi

SPL Internal Meeting, 07/12/2012

Work in progress – preliminary results

Page 23: SPL Short  Cryomodule  Design

6. Lower protection level

23 / 34

LHe 2 K circuit

• Event: power failure

• Heat input: static heat loads – 70 W heat load to LHe bath*

• The mass flow and discharge area calculation follows the method used for the rupture discs (loss of beam vacuum)

• Diameter is highly dependent on the coefficient of discharge (depends on the valve, and is usually lower for low set pressures)

Ps (barg) 0.3

Prel (bara) 1.33

Kd 0.5

Kdr 0.45

w (W) 69.7

Qm (g/s) 3.0

Trel* (K) 5

Dmin (mm) 3.0

ΔP/Ps (%) 0

*R. Bonomi, SPL Short Cryomodule Heat loads; 3rd SPL SCM WG Meeting, 22/5/2012

EN ISO 4126-1

• Formula units are not always consistent with the units presented

• Kb is a correction factor for subcritical flow (function of the isentropic expansion coefficient k and pressure ratio)

• C is a function of k• Kd (Kdr=0.9 Kd) is the coefficient of discharge

*Tsat = 4.5 K

SPL Internal Meeting, 07/12/2012

Work in progress – preliminary results

Page 24: SPL Short  Cryomodule  Design

6. Lower protection levelVacuum vessel

• Event: temporary leak of air / air freezes immediately / leak is not detected / pressure rises during warm-up

• Heat input calculation – appropriate method?

• Relief plate (0.5 barg) behaviour at relief valve set pressure (0.3 barg, for instance)?

• Possibility of using a specific relief plate design which can deal with both higher and lower protection levels (different relief pressures and discharge flows) – is being studied

24 / 34SPL Internal Meeting, 07/12/2012

Page 25: SPL Short  Cryomodule  Design

7. Vacuum vessel mechanical calculationsIntroduction

• Two-part vacuum vessel design carried out by CNRS/IPNO

• Finished detailed design expected soon

• Parallel calculations carried out at CERN

25 / 34SPL Internal Meeting, 07/12/2012

3Re-alignment via vessel flange screws adjustment

± 0.2 mm ± 0.2 mm Possible until top lid is open (survey of single He vessel by laser tracker)

4 Mechanical mounting of top lid

± 0.3 mm ± 0.5 mmIncludes: adjustment of lid/vessel tolerances, sealing compression and tightening of bolts

5 Pump down ± 0.1 mm ± 0.6 mm Depends on shape imperfections, stress relieving, etc.

From Cryomodule tolerances; V. Parma

Page 26: SPL Short  Cryomodule  Design

7. Vacuum vessel mechanical calculationsString of cavities: deflection caused by induced displacements on interfaces with vacuum vessel

26 / 34SPL Internal Meeting, 07/12/2012

May 2012• A simplified model was used to evaluate the importance of the stability of the interfaces

between power couplers double tubes and vacuum vessel for the alignment of the cavities

• Linear and angular displacements were applied to the double tubes flanges and the effect on the alignment of the string of cavities was estimated

Page 27: SPL Short  Cryomodule  Design

7. Vacuum vessel mechanical calculationsString of cavities: deflection caused by induced displacements on interfaces with vacuum vessel

Results: maximum allowable interfaces displacements as a function of cavity alignment tolerance:

27 / 34SPL Internal Meeting, 07/12/2012

Tolerance (mm)

Maximum vertical displacement (mm)

Maximum angular displacement

(mrad)

Maximum angular displacement

(mrad)*

+/- 0.1 0.1 4.0 0.7+/- 0.2 0.2 8.0 1.5+/- 0.3 0.3 12.0 2.2+/- 0.4 0.4 16.0 3.0+/- 0.5 0.5 20.0 3.7+/- 0.6 0.6 24.0 4.4+/- 0.7 0.7 28.0 5.2+/- 0.8 0.8 32.0 5.9+/- 0.9 0.9 36.0 6.7+/- 1.0 1.0 40.0 7.4

* considering the results obtained from the induced displacements on the 5th interface as valid

Low angle propagation and high relative bending stiffness of inter-cavity support suggest that significant values of displacement (vertical and angular) induced on the interfaces to the vacuum vessel may compromise structural integrity of the helium vessels – although results were obtained considering a fixed connection between double tubes and vacuum vessel and an infinitely rigid vacuum vessel

May 2012

Page 28: SPL Short  Cryomodule  Design

7. Vacuum vessel mechanical calculationsStability of interfaces with double tubes during pump down

28 / 34SPL Internal Meeting, 07/12/2012

Courtesy of P. Duchesne et al ; CNRS / IPNO

• The effect of the vacuum load on the stability of the interfaces with the double tubes was studied in parallel with CNRS/IPNO (P. Duschesne)

• Similar, but not identical, models, contact properties and boundary conditions were used by CERN and CNRS/IPNO

September 2012 (work presented in WG meeting 6)

Page 29: SPL Short  Cryomodule  Design

29 / 34

Results – vertical displacement (uz) of interfaces with double tubes for vacuum load:

0.5 1 1.5 2 2.5 3 3.5 4 4.5 5 5.5

-0.05

0

0.05

0.1

0.15

0.2

CNRS / Nonlinear contacts

CERN / Linear contacts

CERN / Linear contacts / Simple interfaces

7. Vacuum vessel mechanical calculationsStability of interfaces with double tubes during pump down

SPL Internal Meeting, 07/12/2012

September 2012 (work presented in WG meeting 6)

uz [mm]

Interface

Page 30: SPL Short  Cryomodule  Design

7. Vacuum vessel mechanical calculationsMounting of the lid: effect of flat flanges shape imperfections

30 / 34SPL Internal Meeting, 07/12/2012

• The effect of the shape imperfections of the two vacuum vessel connection flanges on the stability of the vacuum vessel during its closing was studied in parallel with CNRS/IPNO (P. Duchesne)

• Different assumptions and calculations by CERN and CNRS/IPNO

• In both cases, the closing of the vessel was not “simulated” – assumptions were made and the results are not expected to be representative

• A planarity tolerance of 0.6 mm for each flange was considered

Courtesy of P. Duchesne et al; CNRS / IPNO

Page 31: SPL Short  Cryomodule  Design

7. Vacuum vessel mechanical calculationsMounting of the lid: effect of flat flanges shape imperfections; CNRS/IPNO approach

31 / 34SPL Internal Meeting, 07/12/2012

From Test cryomodule: Modélisation d’un défaut géométrique; P. Duchesne et al

1.2mmExemple de défaut de planéité =1.2mm

Modèle: géométrie parfaite chargement du défaut

Configuration Bearing 1 Bearing 2 Bearing 3 Bearing 4 Bearing 5

Cas 1 0.0005 -0.0036 0.0021 -0.0001 0.0002

Cas 2 0.0026 -0.0167 0.0062 -0.0003 0.0011

Cas 3 -0.045 0.036 0.063 0.036 -0.045

Cas 4 0.024 -0.016 -0.029 -0.018 0.024

Déplacement (mm) suivant Z :

242mm

1.2mm rattrapé par 1 vis

242mm

1.2mm rattrapé par 2 vis contigües

7000mm

1.2mm gap rattrapé par 2 vis au centre (gauche et droit)

Liaisons boulonnées uniquement sur les extrémités de l’enceinte

1.2mm gap rattrapé par 4 vis à l’extrémité

3500mm

Liaisons boulonnées uniquement sur la moitié de l’enceinte

CAS 1

CAS 4

CAS 3

CAS 2

Elément poutre

Connecteur rigide

Many assumptions were made: results are not expected to be representative or definitive

Page 32: SPL Short  Cryomodule  Design

• Displacements applied to bottom flange taking into account the relative stiffness of top and bottom part of vessel

• Entire vessel is modelled and a perfect contact between flanges is used - this should be conservative because a more rigid connection should impose a larger flange deformation, and consequently a larger deformation of the whole bottom part of the vessel

7. Vacuum vessel mechanical calculationsMounting of the lid: effect of flat flanges shape imperfections; CERN approach

32 / 34SPL Internal Meeting, 07/12/2012

1.2 mm

1.2 mm

Load case 1

Load case 2

Page 33: SPL Short  Cryomodule  Design

7. Vacuum vessel mechanical calculationsMounting of the lid: effect of flat flanges shape imperfections; CERN approach

33 / 34SPL Internal Meeting, 07/12/2012

Load case 1

Load case 2

Deformed shape:

Load Case Sign

uz (mm) d=0.42mm (LC1)

uz (mm) d=0.45mm (LC2)

1 + 0.13 -

1 - 0.12 -

2 + - 0.08

2 - - 0.08

Maximum interface displacement as a function of flange displacement magnitude:

Many assumptions were made: results are not expected to be representative or definitive

Page 34: SPL Short  Cryomodule  Design

34 / 34

References• SPL Workspace: https://espace.cern.ch/spl-cryomodule

• V. Parma; “Cryomodule tolerances”

• R.Bonomi; “Summary of the 4th SPL SCM Working Group meeting held on 26/06/2012”

• P. Azevedo et al; “Mock-ups of the SPL cavity supporting system”; SLHiPP2 meeting; 03/05/2012

• EN ISO 4126: Safety devices for protection against excessive pressure (parts 1,6, and 7)

• EN 13648: Cryogenic vessels – Safety devices for protection against excessive pressure (part 3)

• EN 13458: Cryogenic vessels – Static vacuum-insulated vessels (part 2)

• U. Wagner; “Cryogenic scheme, pipes and valves dimensions”; SPL Conceptual Design Review; 04/11/2011

• R. van Weelderen; “Open Cryogenic Action Items”; 4th SPL SC WG meeting; 26/06/2012, updated 04/07/2012

• R. Bonomi; “SPL Short Cryomodule Heat loads”; 3rd SPL SCM WG Meeting, 22/5/2012

• O. Pirotte; “SM18 PID SPL Bunker”

• V. Parma; “SPL Pressure / Temperature Table”

• A. Henriques; "Safety Accessory Calculation Tool for Cryogenic Vessels“

• Lehmann and Zahn; “Safety aspects for LHe cryostats and LHe transport containers” (1978)

• Cavallari et al; “Pressure protection against vacuum failures on the cryostats for LEP SC cavities” (1989)

• Boeckmann et al; “Experimental tests of fault conditions during the cryogenic operation of a XFEL prototype cryomodule”

• CERN Safety Comission; “Safety instruction IS 47: The use of cryogenics fluids”; CERN EDMS doc. 335812

• P. Azevedo; “String of cavities – deflection caused by induced displacements on interfaces with vacuum vessel”

• P. Azevedo; “FE calculations of the vacuum vessel of the SPL Short Cryomodule”; 6th SPL SC WG meeting; 18/09/2012

• P. Duchesne, P. Duthil and S. Rousselot; “ Test cryomodule: Mechanical studies“; 6th SPL SC WG meeting; 18/09/2012

• P. Duchesne, P. Duthil and S. Rousselot; “ Test cryomodule: Modélisation d’un défaut géométrique“

SPL Internal Meeting, 07/12/2012

Page 35: SPL Short  Cryomodule  Design

Thank you for your attention

Page 36: SPL Short  Cryomodule  Design

Spare slides

Page 37: SPL Short  Cryomodule  Design

Heat loads table - TOTALSubassembly Type Source Desti-

nation 2 K 4.5 K 50 K

Double-walled tube

cd radRF

DWT bath 13 (1) x 5= 65

0.1 (2) x 5= 0.5

0.5 (3) x 4+ 0.1 x 1

= 2.1

22 (4) x 4+ 13 x 1= 101

- -

cv DWT gas - - (1) 60 (2) x 5= 300

60 (3) x 5= 300 - (4) -

Cold-warm transition

cd WF TS - - 23.0 x 2= 46.0

cd TS CM 0.8 x 2= 1.6

0.8 x 2= 1.6

0.8 x 2= 1.6

0.8 x 2= 1.6 - -

rad WF + wall CM 1.0 x 2= 2.0

1.0 x 2= 2.0

1.0 x 2= 2.0

1.0 x 2= 2.0 - -

rad WF TS - - 0.2 x 2= 0.4

rad VV TS - - 45

rad TS CM 1.6 1.6 1.6 1.6 - -

Cavity RF cavity CM - (1) - (2)20.0 (3) x

4= 80.0

20.0 (4) x 4

= 80.0- -

TOT for SCM (W) 68.7 (1) 5.7 (2) 87.3 (3) 186.2 (4) - 300 (2) 300 (3) - 91.4

Static heat loads(1) RF off, cool off(2) RF off, cool onDynamic heat loads(3) RF on, cool on(4) RF on, cool off

(R.Bonomi)

Page 38: SPL Short  Cryomodule  Design

Expansion factor

Page 39: SPL Short  Cryomodule  Design

Critical / subcritical flow

Page 40: SPL Short  Cryomodule  Design

EN 16648-3 – Pressure drop

Page 41: SPL Short  Cryomodule  Design

Volume of 2K circuit  (l)

4 cavities 64 l  

1 phase separator 5 l  

1 x line 47.1 l l=6000; d=100

4 Y lines 2.5 l l=100; d=100

  total 318.3 l  

Helium volume inside SPL cryomodule

Page 42: SPL Short  Cryomodule  Design

From A. Henriques; "Safety Accessory Calculation Tool for Cryogenic Vessels"

Heat input table by C. Parente

Page 43: SPL Short  Cryomodule  Design

Component or FittingMinor Loss Coefficient

- ξ -

90o bend, sharp 1.3

90o bend, with vanes 0.7

90o bend, roundedradius/diameter duct <1

0.5

90o bend, roundedradius/diameter duct >1 0.25

45o bend, sharp 0.5

45o bend, roundedradius/diameter duct <1

0.2

45o bend, roundedradius/diameter duct >1 0.05

T, flow to branch(applied to velocity in branch) 0.3

Examples of Minor loss coefficients for different components common in air duct distribution systems:(http://www.engineeringtoolbox.com/minor-loss-air-ducts-fittings-d_208.html)

Local pressure drop coefficients

Page 44: SPL Short  Cryomodule  Design

String of cavities: deflection caused by induced displacements on interfaces with vacuum vessel; results

Load case

Cavity 1 Cavity 2 Cavity 3 Cavity 4angle (mrad)

disp. (mm)

angle (mrad)

disp. (mm)

angle (mrad)

disp. (mm)

angle (mrad)

disp. (mm)

1z 0.37 0.08 0.00 0.00 0.00 0.00 0.00 0.001zb 0.37 -0.08 0.00 0.00 0.00 0.00 0.00 0.001α 0.02 -0.03 0.01 0.00 0.00 0.00 0.00 0.001αb 0.02 0.03 0.01 0.00 0.00 0.00 0.00 0.002z 0.41 0.47 0.37 0.08 0.00 0.00 0.00 0.002zb 0.41 -0.47 0.37 -0.08 0.00 0.00 0.00 0.002α 0.03 0.04 0.02 -0.03 0.01 0.00 0.00 0.002αb 0.03 -0.04 0.02 0.03 0.01 0.00 0.00 0.003z 0.04 -0.05 0.41 0.47 0.37 0.08 0.00 0.003zb 0.04 0.05 0.41 -0.47 0.37 -0.08 0.00 0.003α 0.00 0.00 0.03 0.04 0.02 -0.03 0.01 0.003αb 0.00 0.00 0.03 -0.04 0.02 0.03 0.01 0.004z 0.00 0.01 0.04 -0.05 0.41 0.47 0.37 0.084zb 0.00 -0.01 0.04 0.05 0.41 -0.47 0.37 -0.084α 0.00 0.00 0.00 0.00 0.04 0.04 0.04 -0.054αb 0.00 0.00 0.00 0.00 0.04 -0.04 0.04 0.055z 0.00 0.00 0.00 0.00 0.04 -0.04 0.37 0.425zb 0.00 0.00 0.00 0.00 0.04 0.04 0.37 -0.425α 0.00 0.00 0.00 0.00 0.03 -0.03 0.24 0.275αb 0.00 0.00 0.00 0.00 0.03 0.03 0.24 -0.27

Page 45: SPL Short  Cryomodule  Design

Example (load case 2α ):

Von Mises stress (MPa); model and mesh not detailed enough – this is not a stress analysis

String of cavities: deflection caused by induced displacements on interfaces with vacuum vessel; stress results

Page 46: SPL Short  Cryomodule  Design

Opening of 5 mm between flanges (warped vessel)

Mounting of the lid: effect of flat flanges shape imperfections; different calculation

June 2012

For this first step load (0.6 MPa), the displacements of the interfaces are quite small. For comparison, the pressure value correspondent to 60 kN/m (reference value for 30% of O-ring deformation) is around 0.8 MPa.

Displacements of interfaces with double tube (maximum absolute values)

P (MPa) ux,max uy,max uz,max αmax

0.6 0.03 0.03 0.04 0.50

1.2 0.08 0.03 0.02 0.54

1.8 0.13 0.04 0.01 0.78

2.4 0.18 0.04 0.02 1.23

3 0.23 0.04 0.04 1.86