spectradyne’s nanoparticle analyzer technology · spectradyne’s nanoparticle analyzer...

2
www.nanoparticleanalyzer.com © Spectradyne LLC [email protected] R1.0 Particle Analysis ® 0 100 Time ( s) Particle size Passage through NC generates signal Signal scales with particle size NC Nanoscale constriction (NC) Electrical detection Individual particles Spectradyne’s Nanoparticle Analyzer Technology A Modern Implementation of Proven Technology Spectradyne’s nCS1instrument (Fig.1) characterizes nanoparticles using Microfluidic Resistive Pulse Sensing (MRPS, a.k.a. the Coulter Principle), a proven technique that is considered the gold standard for whole blood measurements. The RPS technique has been updated using Spectradyne’s patented nanoparticle analyzer (NPA) technology. The heart of the technology is the microfluidic cartridge (Fig.1), which allows the electrical detection of nanoparticles as they pass one by one through a nanoconstriction. The nCS1 Difference: No dependence on particle material type • High-resolution size distribution Sizing range: 50nm to 2µm diameter • Arbitrary polydispersity Total sample analysis in minutes. Disposable microfluidic cartridges Only 3µL sample required Truly orthogonal to light-based techniques Spectradyne nCS1 and Microfluidic Cartidges Spectradyne nCS1 Applications Nanomedicines Biologics: protein aggregates, viruses, etc. Drug ingredients: excipients, carriers, lipids, etc. General nanoparticles: gold/silver, silicon- derived, metal-oxide nanoparticles, etc. How it Works: How Microfluidic Resistive Pulse Sensing (MRPS) works: Particles in fluid pass through a nanoscale constriction (NC) as shown on left side. A voltage drop is applied continuously across the two sides of the NC. As particles pass through the NC, the output signal changes in proportion to the volume of the particle. Particles are measured individually, with no dependence on particle material. Figure 1: The Spectradyne nCS1 occupies a small bench top footprint, approximately 1.5 sq ft (left). Only 3 μL of analyte is required for analysis using a disposable microfluidic cartridge (right), which prevents contamination between measurements and eliminates cleaning requirements.

Upload: others

Post on 15-Jul-2020

7 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Spectradyne’s Nanoparticle Analyzer Technology · Spectradyne’s Nanoparticle Analyzer Technology A Modern Implementation of Proven Technology Spectradyne’s nCS1™ instrument

www.nanoparticleanalyzer.com © Spectradyne LLC [email protected]

Particle Analysis

®

0 100Time ( s)

Particle size

Passage through NC generates signal Signal scales with particle size

NCNanoscale constriction (NC) Electrical detection Individual particles

Spectradyne’s Nanoparticle Analyzer Technology

A Modern Implementation of Proven Technology

Spectradyne’s nCS1™ instrument (Fig.1) characterizes nanoparticles using Microfluidic Resistive Pulse Sensing (MRPS, a.k.a. the Coulter Principle), a proven technique that is considered the gold standard for whole blood measurements. The RPS technique has been updated using Spectradyne’s patented nanoparticle analyzer (NPA) technology. The heart of the technology is the microfluidic cartridge (Fig.1), which allows the electrical detection of nanoparticles as they pass one by one through a nanoconstriction.The nCS1 Difference:

• No dependence on particle material type• High-resolution size distribution• Sizing range: 50nm to 2µm diameter• Arbitrary polydispersity• Total sample analysis in minutes.• Disposable microfluidic cartridges• Only 3µL sample required• Truly orthogonal to light-based techniques

Spectradyne nCS1 and Microfluidic Cartidges

Spectradyne nCS1 Applications• Nanomedicines• Biologics: protein aggregates, viruses, etc.• Drug ingredients: excipients, carriers, lipids,

etc.• General nanoparticles: gold/silver, silicon-

derived, metal-oxide nanoparticles, etc.

How it Works:

How Microfluidic Resistive Pulse Sensing (MRPS) works: Particles in fluid pass through a nanoscale constriction (NC) as shown on left side. A voltage drop is applied continuously across the two sides of the NC. As particles pass through the NC, the output signal changes in proportion to the volume of the particle. Particles are measured individually, with no dependence on particle material.

Figure 1: The Spectradyne nCS1 occupies a small bench top footprint, approximately 1.5 sq ft (left). Only 3 μL of analyte is required for analysis using a disposable microfluidic cartridge (right), which prevents contamination between measurements and eliminates cleaning requirements.

Page 2: Spectradyne’s Nanoparticle Analyzer Technology · Spectradyne’s Nanoparticle Analyzer Technology A Modern Implementation of Proven Technology Spectradyne’s nCS1™ instrument

See what you’ve been missing!Contact Spectradyne today and ask for a complimentary sample analysis:

Spectradyne LLC, 23875 Madison St., Suite A, Torrance, CA 90505(424) 271-9262 www.nanoparticleanalyzer.com

www.nanoparticleanalyzer.com © Spectradyne LLC [email protected]

Spectradyne nCS1NTADLS

nCS1

TM -

Conc

entr

atio

n Sp

ectr

al D

ensi

tyTM

(108 P

artic

les.

mL-1

. nm

-1)

NTA

- C

once

ntra

tion

(108 P

artic

les.

mL-1

)

DLS

- N

orm

aliz

ed In

tens

ity (a

.u.)

Polydispersity Measurement

0

4

0

4

0

1

0 50 100Particle Diameter (nm)

150 200 nCS1

TM -

Conc

entr

atio

n Sp

ectr

al D

ensi

tyTM

(107 P

artic

les.

mL-1

. nm

-1)

Aggregation Stress Comparison

More Stress: More Aggregates

0

1

150 200 300 400Particle Diameter (nm)

500

Example nCS1 Applications

Measurement of polydisperse samples:

In the graph above, results are shown for an equal parts mixture of polystyrene beads having NIST-certified mean diameters of 52, 94,122 and 150nm measured with the nCS1, and also by an independent laboratory using DLS and nanotracking analysis (NTA).

Only the nCS1 has high enough resolution to clearly resolve the four components of the mixture, and properly return concentrations for each within the manufacturers specification.

Spectradyne nCS1 SpecificationsTechnology Microfluidic Resistive Pulse Sensing (MRPS)Acceptable particle types All materials (e.g. transparent/opaque, conducting/insulating, etc.)Particle Size Range 40nm to 2,000nmSizing/Concentration Precision Sizing < ±3%, Concentration < ±10% (particles/ml)Measurable Concentration Range 10⁵ to 10¹² particles/ml (sample dependent)Sample Size Required 3 μLMaximum Particle Detection Rate ≈ 10,000 particles/sec (sample dependent)Instrument Control Interface USB to Windows computerData Analysis Software Proprietary signal extraction method, real-time signal display,

real-time concentration display, multiple filtering methods, multi-dimensional data display, PDF report export

Physical Characteristics 13” W x 15” L (33 cm W x 38 cm L)Electrical Standard 120/220V, 50/60 Hz AC

Measurement of submicron protein aggregates:

In the example shown above, protein aggregates in a proprietary protein formulation were accurately quantified by the nCS1. Five samples were provided: one unstressed sample and four samples subjected to increasing levels of stress (10, 20, 30 and 60 minutes of stress). The sample was run “as-is”: no dilution or additives were required.

The nCS1 was able to clearly show increased aggregation with increased stress. Protein aggregates are difficult to measure by NTA or DLS, since they rely on optical contrast, which is very low in this type of sample.