SelectraTelescopic CS Lead Delivery System

BIOTRONIK

Cardiac Rhythm Management

Cardiac Resynchronization Therapy

Page 2: Selectra Telescopic CS Lead Delivery System BIOTRONIK Cardiac Rhythm Management Cardiac Resynchronization Therapy

2

Coronary sinus lead delivery system

Selectra. The first choice.

Selectra Outer Catheters

A variety of advanced catheters incorporating superior and innovative features

Selectra Inner Catheters

Optimal vessel sub-selection for direct delivery of low profile coronary sinus leads

New Selectra Slitter Tool

Ergonomic design for a smooth slitting process with lead dislodgement prevention

Page 3: Selectra Telescopic CS Lead Delivery System BIOTRONIK Cardiac Rhythm Management Cardiac Resynchronization Therapy

3

Telescopic system with inner catheters

Advanced and atraumatic soft tip design

New and innovative ergonomic handle

Balanced and superior shaft design

A variety of configurations

Coronary sinus lead delivery system

Selectra. The first choice.

Page 4: Selectra Telescopic CS Lead Delivery System BIOTRONIK Cardiac Rhythm Management Cardiac Resynchronization Therapy

4

The ultimate handling experience

Innovative design

The innovative handle design allows for a smooth slitting process from every angle

An integrated hemostatic valve enables effortless introduction of inner catheters or coronary sinus leads

The build-in side-port enables simultaneous manipulation and contrast injection provide an extraordinary handling experience

Page 5: Selectra Telescopic CS Lead Delivery System BIOTRONIK Cardiac Rhythm Management Cardiac Resynchronization Therapy

5

Lubricious and balanced like no other

Unique handling

Superior kink resistance and sensitive torque control are achieved by smooth stiffness profiles resulting in a pushable proximal shaft and flexible distal section

Page 6: Selectra Telescopic CS Lead Delivery System BIOTRONIK Cardiac Rhythm Management Cardiac Resynchronization Therapy

6

The unique hydrophilic inner coating provides excellent lubricity allowing for effortless introduction of inner catheters, delivery of coronary sinus leads and system removal

Lubricious and balanced like no other

Unique handling

Page 7: Selectra Telescopic CS Lead Delivery System BIOTRONIK Cardiac Rhythm Management Cardiac Resynchronization Therapy

7

Atraumaticity matters

Superior quality

Round soft tip design provides high atraumaticity during CS cannulation minimizing the risk for dissections

During lead manipulation, the high tip stability reduces the risk for prolapse

Excellent visibility during system positioning due to highly radiopaque polymeric x-ray marker

Page 8: Selectra Telescopic CS Lead Delivery System BIOTRONIK Cardiac Rhythm Management Cardiac Resynchronization Therapy

8

Teamwork

Superior quality

The combination of Selectra outer catheters with inner sub-selection catheters enables direct delivery of low profile coronary sinus leads such as the new Sentus OTW

Optimized coronary sinus cannulation and in deep-seating outer catheters due to the perfect balance of stiffness and flexibility

Page 9: Selectra Telescopic CS Lead Delivery System BIOTRONIK Cardiac Rhythm Management Cardiac Resynchronization Therapy

9

Superior quality

Selectra inner catheters

Slittable in-line luer with integrated hemostatic valve for smooth slitting procedures and ergonomic one-handed manipulation and contrast injection

Hydrophilic inner coating for easy insertion of CS leads and removal of delivery system without the risk for lead dislocation

50° and 90° degree tip angulations for easy vessel sub-selection

Page 10: Selectra Telescopic CS Lead Delivery System BIOTRONIK Cardiac Rhythm Management Cardiac Resynchronization Therapy

10

A variety of configurations

Eight curve shapes adapt to different patient coronary sinus anatomies

Available in 45 cm and 55 cm lengths

Unique handling

Page 11: Selectra Telescopic CS Lead Delivery System BIOTRONIK Cardiac Rhythm Management Cardiac Resynchronization Therapy

11

Selectra “Straight”

For CS approach from above with EP catheter

Support

None

Shape similar to

BSX Acuity BA CS-ST

MDT Attain Command Straight

SJM CPS Direct SL Straight

Reach

None

Page 12: Selectra Telescopic CS Lead Delivery System BIOTRONIK Cardiac Rhythm Management Cardiac Resynchronization Therapy

12

Selectra “BIO2”

For CS approach from above

Support

Medium, from vena cava superior

Shape similar to

SJM CPS Direct SL MP

Reach

Less

Page 13: Selectra Telescopic CS Lead Delivery System BIOTRONIK Cardiac Rhythm Management Cardiac Resynchronization Therapy

13

Selectra “Multipurpose EP”

For CS approach from above

Support

Medium, from vena cava superior

Shape similar to

BSX Acuity BA CS-MP

MDT Attain Command MB2

SJM CPS Direct SL 115°

Reach

Less

Page 14: Selectra Telescopic CS Lead Delivery System BIOTRONIK Cardiac Rhythm Management Cardiac Resynchronization Therapy

14

Selectra “Multipurpose Hook”

For CS approach from below

Support

Medium, from VCS right atrial wall

Shape similar to

BSX Acuity BA CS-H

MDT Attain Command MP

Reach

Medium

Page 15: Selectra Telescopic CS Lead Delivery System BIOTRONIK Cardiac Rhythm Management Cardiac Resynchronization Therapy

15

Selectra “Right”

For CS approach during right-sided implants

Support

Medium, from VCS

Shape similar to

BSX Acuity BA CS-EH ST R

MDT Attain Command Right

SJM CPS Direct SL Right

Reach

Medium

Page 16: Selectra Telescopic CS Lead Delivery System BIOTRONIK Cardiac Rhythm Management Cardiac Resynchronization Therapy

16

Selectra “Amplatz 6.0”

For CS approach when severe left heart dilatation is present

Support

Medium, from VCS

Shape similar to

BSX Acuity BA CS_A6

MDT Attain Command Amplatz

Reach

Medium

Page 17: Selectra Telescopic CS Lead Delivery System BIOTRONIK Cardiac Rhythm Management Cardiac Resynchronization Therapy

17

Selectra “Hook”

For CS approach from above

Support

Medium, from VCS

Shape similar to

SJM CPS Direct SL 135°

Reach

Medium

Page 18: Selectra Telescopic CS Lead Delivery System BIOTRONIK Cardiac Rhythm Management Cardiac Resynchronization Therapy

18

Selectra “Extended Hook”

For CS approach from below

Support

More, from VCS and right atrial wall

Shape similar to

BSX Acuity BA CS-EH

MDT Attain Command EH

SJM CPS Direct SL Wide

Reach

More

Page 19: Selectra Telescopic CS Lead Delivery System BIOTRONIK Cardiac Rhythm Management Cardiac Resynchronization Therapy

19

New Selectra slitter tool

Integrated lead clip fixates the lead body during the slitting process

Secure and easy slitting of the handle and the entire catheter due to optimized blade

Ergonomic design allows for comfortable handling during catheter removal or continuation of the slitting process in case of slipping

Unique handling

Page 20: Selectra Telescopic CS Lead Delivery System BIOTRONIK Cardiac Rhythm Management Cardiac Resynchronization Therapy

Cardiac Rhythm Management

Cardiac Resynchronization Therapy

Handling Tips & Tricks

Selectra Hemostatic Valve Design- INTERNAL EDUCATION ONLY -

Page 21: Selectra Telescopic CS Lead Delivery System BIOTRONIK Cardiac Rhythm Management Cardiac Resynchronization Therapy

21

INTERNALEDUCATION

ONLY

The valve is designed to ensure a low friction environment for optimal lead pushability and safety during sheath removal.

The valve consist of a proximal sealing ring and a distal back-pressure valve.

Selectra. The first choice.

Selectra hemostatic valve design

Back-pressure valve

Sealing ring

Valve closed Valve open

Insertion of tools open the back-pressure valve

Page 22: Selectra Telescopic CS Lead Delivery System BIOTRONIK Cardiac Rhythm Management Cardiac Resynchronization Therapy

22

INTERNALEDUCATION

ONLY

Selectra hemostatic valve design update

The hemostatic valve performance was continuously improved based on observations from the PME and initial market feedback.

Process improvements lead to avoidance of valve prolapse during lead or catheter removal ensuring the integrity and functionality of the back-pressure valve.

Short-term: The TVI with dilator is offered separately and can be used to introduce 0.14” guide wires. The dilator seals onto the guide wire and reduces back-bleeding.

Medium-term: A new TVI which seals onto a 0.14” guide wire will be developed and included into the accessory kit.

Recommendation: Close the proximal end of the catheter with your thumb during aspiration through the side port.

Selectra hemostatic valve design.

Page 23: Selectra Telescopic CS Lead Delivery System BIOTRONIK Cardiac Rhythm Management Cardiac Resynchronization Therapy

23

INTERNALEDUCATION

ONLY

Use case #1 - Nothing is inserted

Due to the fact that the venous blood pressure is higher than the atmospheric pressure the back-pressure valve is fully closed when nothing is inserted from proximal!

Selectra hemostatic valve design.

Higher pressure

Lower pressure

Nothing is inserted

Page 24: Selectra Telescopic CS Lead Delivery System BIOTRONIK Cardiac Rhythm Management Cardiac Resynchronization Therapy

24

INTERNALEDUCATION

ONLY

Use case #2 - Dilator insertion

Observation:

If only a dilator is inserted, back-bleeding can occur through inner lumen of the dilator because of an “open” system.

Selectra hemostatic valve design.

Dilator

Back-bleeding

Only dilator inserted

Higher pressure

Page 25: Selectra Telescopic CS Lead Delivery System BIOTRONIK Cardiac Rhythm Management Cardiac Resynchronization Therapy

25

INTERNALEDUCATION

ONLY

Use case #2 - Dilator insertion (cont.)

Solution:

Use dilator only in combination with the 0.35” guide wire in order to “close” the inner lumen of the dilator!

0.35” Guide Wire

Dilator

Dilator and guide wire inserted

Selectra hemostatic valve design.

Page 26: Selectra Telescopic CS Lead Delivery System BIOTRONIK Cardiac Rhythm Management Cardiac Resynchronization Therapy

26

INTERNALEDUCATION

ONLY

Use case #3 – Valve prolapse

Observation:

Prolapse may occur when a dilator or lead is quickly removed from the sheath pulling the flap back through valve

Selectra hemostatic valve design.

Higher pressure

Page 27: Selectra Telescopic CS Lead Delivery System BIOTRONIK Cardiac Rhythm Management Cardiac Resynchronization Therapy

27

INTERNALEDUCATION

ONLY

Use case #3 - Valve prolapse (cont.)

Solution:

Leakage will stop once a lead or catheter is advanced through the valve, pushing the flap back from the valve

Selectra hemostatic valve design.

Dilator, LV Lead or EP Catheter

Insert dilator or lead

Page 28: Selectra Telescopic CS Lead Delivery System BIOTRONIK Cardiac Rhythm Management Cardiac Resynchronization Therapy

28

INTERNALEDUCATION

ONLY

Use case #3 - Valve prolapse (cont.)

Solution:

Process improvements aims to avoid the valve prolapse during lead or catheter removal ensuring the integrity and functionality of the back-pressure valve.

Please note that the improvement is being introduced seamlessly into running production. Therefore you may receive products before or after the process change depending on inventory. Care should be taken to observe a potential valve prolapse and if required, to take the action recommended on the previous slide.

Selectra hemostatic valve design.

Page 29: Selectra Telescopic CS Lead Delivery System BIOTRONIK Cardiac Rhythm Management Cardiac Resynchronization Therapy

29

INTERNALEDUCATION

ONLY

Use case #4 - Lead or EP catheter insertion

The sealing ring fully seals onto a lead or EP catheter and no back-bleeding occurs!

Selectra hemostatic valve design.

LV Lead or EP Catheter

Lead is inserted

Page 30: Selectra Telescopic CS Lead Delivery System BIOTRONIK Cardiac Rhythm Management Cardiac Resynchronization Therapy

30

INTERNALEDUCATION

ONLY

Use case #5 - Aspiration with syringe

Observation:

Aspiration can lead to lower pressure distally of the valve causing it to open slightly, which might cause air bubbles being aspirated.

Selectra hemostatic valve design.

Higher pressure

Lower pressure

Air bubbles might enter the catheter and will be directly aspirated into the syringe

Aspiration through side port

Page 31: Selectra Telescopic CS Lead Delivery System BIOTRONIK Cardiac Rhythm Management Cardiac Resynchronization Therapy

31

INTERNALEDUCATION

ONLY

Use case #5 - Aspiration with syringe (cont.)

Solution:

Close the proximal end of the catheter with your thumb or sealing cap to avoid valve opening during aspiration!

After aspiration the valve closes again due to the pressure change and there is no risk of air embolism for the patient!

Selectra hemostatic valve design.

The back-pressure valve closes due to the higher pressure inside the system!Higher pressure

Lower pressure

Aspiration through side port

Page 32: Selectra Telescopic CS Lead Delivery System BIOTRONIK Cardiac Rhythm Management Cardiac Resynchronization Therapy

32

INTERNALEDUCATION

ONLY

Use case #6 - Guide wire only insertion

Observation:

The sealing ring does not fully seal onto 0.14” and 0.35” guide wires if they are introduced alone, which might cause slight back-bleeding.

Selectra hemostatic valve design.

Guide Wire

Only guide wire inserted

Back-bleeding

Page 33: Selectra Telescopic CS Lead Delivery System BIOTRONIK Cardiac Rhythm Management Cardiac Resynchronization Therapy

33

INTERNALEDUCATION

ONLY

Use case #6 - Guide wire only insertion (cont.)

Solution:

If only a 0.14” guide wire is inserted use the TVI with dilator. The TVI and dilator seals onto the guide wire and centers it avoiding excessive back-bleeding.

Selectra hemostatic valve design.

TVI Dilator

TVI

Guide Wire

Guide Wire and TVI with dilator inserted