real-time 3d mr respiratory motion acquisition for simultaneous pet-mr imaging

6
Real-time 3D MR Respiratory Motion Acquisition for Simultaneous PET / MR Imaging Liheng Guo 1 , Mark Ahlman 3 , Tao Feng 2 , Michael Guttman 4 , Elliot McVeigh 1 , David Bluemke 3 , Benjamin Tsui 2 , Daniel Herzka 1 1 Biomedical Engineering, Johns Hopkins University, Baltimore, MD 2 Department of Radiology, Johns Hopkins University, Baltimore, MD 3 Radiology and Imaging Sciences, National Institutes of Health, Bethesda, MD 4 Cardiology, School of Medicine, Johns Hopkins University, Baltimore, MD Podium Presentation. Annual Meeting of the International Society for Nuclear Medicine and Molecular Imaging (SNMMI). St. Louis, MO. June 11, 2014. Liheng Guo SNMMI 2014

Upload: liheng-guo

Post on 17-Feb-2017

219 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Real-time 3D MR Respiratory Motion Acquisition for Simultaneous PET-MR Imaging

Real-time 3D MR Respiratory Motion Acquisition for SimultaneousPET / MR Imaging

Liheng Guo1, Mark Ahlman3, Tao Feng2, Michael Guttman4, Elliot McVeigh1,David Bluemke3, Benjamin Tsui2, Daniel Herzka1

1Biomedical Engineering, Johns Hopkins University, Baltimore, MD2Department of Radiology, Johns Hopkins University, Baltimore, MD3Radiology and Imaging Sciences, National Institutes of Health, Bethesda, MD4Cardiology, School of Medicine, Johns Hopkins University, Baltimore, MD

Podium Presentation. Annual Meeting of the International Society for Nuclear Medicine and Molecular Imaging (SNMMI). St. Louis, MO. June 11, 2014.

Liheng GuoSNMMI 2014

Page 2: Real-time 3D MR Respiratory Motion Acquisition for Simultaneous PET-MR Imaging

Motivation• Breathing motion corrupts positron emission tomography (PET) images

Goerres, et al. %Respiration-induced Attenuation Artifact at PET/CT: Technical Considerations. Radiology. 2003; 226:906–910.

Liver

Lung

TimeHeart

http://www.siemens.com/press/en/presspicture/

Positron Emission Tomography (PET)

Motion-induced false signal void

Page 3: Real-time 3D MR Respiratory Motion Acquisition for Simultaneous PET-MR Imaging

Objective• To track breathing motion in 3D in real time, using simultaneous PET-MRI• To generate 3D motion vector field to correct PET images

http://www.siemens.com/press/en/events/healthcare/2010-11-Biograph-mMR.php

Page 4: Real-time 3D MR Respiratory Motion Acquisition for Simultaneous PET-MR Imaging

Method: Imaging Data Acquisition• Radial 3D MRI

Barger, et al. MRM 2002;48:297-305

• Golden Angle Sampling Pattern500 views

1000

2000

Early Late

Page 5: Real-time 3D MR Respiratory Motion Acquisition for Simultaneous PET-MR Imaging

Method: Image Reconstruction

Imaging Data Acquisition

. . .

• Sliding window frames• Iterative compressed-sensing reconstruction in 3D wavelet domain

Page 6: Real-time 3D MR Respiratory Motion Acquisition for Simultaneous PET-MR Imaging

Frame 1 Frame 2 Frame 3 Frame 4

Coronal View(slice 60 of 128)

Coronal Projection(all slices)

MotionVector Field

(slice 60 of 128)

3.1×3.1×3.1mm3 voxels 2s temporal res 4D registration

Results: Real-Time 3D Images• Sufficient Resolution for Motion Mapping