Page 1: Product Manual for Acclaim Surfactant · 2008-11-12 · PRODUCT MANUAL for the ACCLAIM SURFACTANT GUARD COLUMN (4.3 x 10mm, P/N 063215) ... Any HPLC pump capable of delivering

Acclaim® Surfactant

Page 2: Product Manual for Acclaim Surfactant · 2008-11-12 · PRODUCT MANUAL for the ACCLAIM SURFACTANT GUARD COLUMN (4.3 x 10mm, P/N 063215) ... Any HPLC pump capable of delivering

PRODUCT MANUAL

for the

ACCLAIM SURFACTANT GUARD COLUMN

(4.3 x 10mm, P/N 063215) (2.1 x 10mm, P/N 068124)

ACCLAIM SURFACTANT ANALYTICAL COLUMN

(4.6 x 150mm, P/N 063201) (4.6 x 250mm, P/N 063203) (2.1 x 150mm, P/N 068123)

©2008 DIONEX Corporation Document No. 065087

Revision 02 June 2008

Page 3: Product Manual for Acclaim Surfactant · 2008-11-12 · PRODUCT MANUAL for the ACCLAIM SURFACTANT GUARD COLUMN (4.3 x 10mm, P/N 063215) ... Any HPLC pump capable of delivering

Acclaim Surfactant Product Manual Page 2 of 41

Table of Contents

SECTION 1 – INTRODUCTION TO THE ACCLAIM SURFACTANT COLUMN .........................................4 1.1 FEATURES OF THE ACCLAIM SURFACTANT COLUMN...................................................................................4 1.2 ACCLAIM SURFACTANT OPERATING LIMITS AND SPECIFICATIONS .............................................................6 1.3 FORMATS OF THE ACCLAIM SURFACTANT COLUMN ...................................................................................6

1.3.1 Acclaim Surfactant Products .........................................................................................................6 1.4 ACCLAIM SURFACTANT OPERATING CONDITIONS.......................................................................................7 1.5 KEY APPLICATIONS OF THE ACCLAIM SURFACTANT COLUMN ....................................................................7

SECTION 2 – CHROMATOGRAPHY SYSTEM ..................................................................................................8 2.1 DETECTORS.................................................................................................................................................8

2.1.1 Evaporative Light Scattering Detection (ELSD) ...........................................................................8 2.1.2 Suppressed Conductivity Detection (SCD) ....................................................................................8 2.1.3 UV Detection .................................................................................................................................8 2.1.4 Mass Spectrometric Detection (MS) ..............................................................................................8 2.1.5 Refractive Index Detection (RI) .....................................................................................................9

2.2 PUMPS .........................................................................................................................................................9 2.3 INJECTORS...................................................................................................................................................9

SECTION 3 - INSTALLATION .............................................................................................................................10 3.1 SYSTEM REQUIRMENTS .............................................................................................................................10 3.2 SYSTEM VOID VOLUME ............................................................................................................................10 3.3 THE INJECTION LOOP ................................................................................................................................10 3.4 COLUMN INSTALLATION...........................................................................................................................10 3.5 COLUMN GUARD.......................................................................................................................................10

SECTION 4 – OPERATION ...................................................................................................................................11 4.1 MOBILE PHASE SELECTION .......................................................................................................................11

4.1.1 Evaporative Light Scattering Detection (ELSD) .........................................................................11 4.1.2 Suppressed Conductivity Detection (SCD) ..................................................................................11 4.1.3 UV Detection ...............................................................................................................................11 4.1.4 Mass Spectrometric Detection (MS) ............................................................................................11 4.1.5 Refractive Index Detection (RI) ...................................................................................................11

4.2 VALIDATING COLUMN PERFORMANCE ......................................................................................................11 4.2.1 Procedure for Validating Column Performance: ........................................................................12

4.3 EQUILIBRATING THE COLUMN...................................................................................................................12 4.4 QUICKSTART FOR ACCLAIM SURFACTANT COLUMN................................................................................12

4.4.1 Using the chromatographic conditions for Column Quality Assurance Report:.........................12 4.4.2 Column Quickstart.......................................................................................................................12

4.5 CARING FOR THE COLUMN.........................................................................................................................13 4.6 CHEMICAL PURITY REQUIREMENTS .........................................................................................................13

4.6.1 Inorganic Chemicals....................................................................................................................13 4.6.2 De-ionized Water .........................................................................................................................13 4.6.3 Solvents ........................................................................................................................................14

4.7 MOBILE PHASE PREPARATION ...................................................................................................................14 4.7.1 0.1 M Ammonium Acetate Buffer (1 L) ........................................................................................14 4.7.2 20 mM Sodium Borate Buffer (1 L) .............................................................................................14 4.7.3 20 mM Acetic Acid Buffer (1 L) ...................................................................................................14 4.7.4 Mobile Phases Containing Solvents ............................................................................................14

Document 065087 Rev. 02 ©Dionex Corporation June 2008

Page 4: Product Manual for Acclaim Surfactant · 2008-11-12 · PRODUCT MANUAL for the ACCLAIM SURFACTANT GUARD COLUMN (4.3 x 10mm, P/N 063215) ... Any HPLC pump capable of delivering

Acclaim Surfactant Product Manual Page 3 of 41

Document 065087 Rev. 02 ©Dionex Corporation June 2008

SECTION 5 – METHODS DEVELOPMENT.......................................................................................................15 5.1 IONIC STRENGTH .......................................................................................................................................15 5.2 MOBILE PHASE PH ....................................................................................................................................16 5.3 ORGANIC MODIFIER...................................................................................................................................17 5.4 ION-PAIRING AGENT.................................................................................................................................17 5.5 TEMPERATURE ..........................................................................................................................................17

SECTION 6 – APPLICATIONS .............................................................................................................................18 6.1 ANIONIC SURFACTANTS.............................................................................................................................18

6.1.1 Linear Alkylbenzene Sulfonates ...................................................................................................18 6.1.2 Sodium Lauryl Sulfate.................................................................................................................. 19

6.2 CATIONIC SURFACTANTS...........................................................................................................................21 6.2.1 Cationic Surfactants with ELS Detection ....................................................................................21 6.2.2 Cationic Surfactants with Suppressed Conductivity Detection....................................................22 6.2.3 Benzalkonium Salts ......................................................................................................................23 6.2.4 Quaternary Imidazolium Compounds..........................................................................................24

6.3 NON-IONIC SURFACTANTS ........................................................................................................................25 6.3.1 Ethoxylated Quaternary Compounds...........................................................................................25 6.3.2 PEG Monoethyl Ether (MW-550) ................................................................................................26 6.3.3 Triton X-100.................................................................................................................................27 6.3.4 NEODOL 25-7 .............................................................................................................................28 6.3.5 ZONYL FSO Perfluorosurfactant ................................................................................................29 6.3.6 Polyethylene Glycols (PEG) ........................................................................................................30

6.4 SURFACTANTS IN COMMERCIAL PRODUCTS...........................................................................31 6.4.1 Laundry Detergent.......................................................................................................................31 6.4.2 Shampoo ......................................................................................................................................32 6.4.3 Fabric Softener ............................................................................................................................33

SECTION 7 – TROUBLESHOOTING GUIDE ....................................................................................................34 7.1 HIGH BACKPRESSURE ...............................................................................................................................34

7.1.1 Finding the Source of High System Pressure ..............................................................................34 7.1.2 Replacing Column Bed Support Assemblies ................................................................................34

7.2 HIGH BACKGROUND OR NOISE .........................................................................................................................35 7.2.1 Preparation of Mobile Phases......................................................................................................35 7.2.2 A Contaminated Guard or Analytical Column .............................................................................35 7.2.3 Contaminated Hardware ..............................................................................................................35

7.3 POOR PEAK RESOLUTION..................................................................................................................................35 7.3.1 Loss of Column Efficiency ...........................................................................................................35 7.3.2 Poor Resolution Due to Shortened Retention Times....................................................................36 7.3.3 Loss of Front End Resolution .......................................................................................................36

LOT VALIDATION – 5µM 120Å ...........................................................................................................................40 Specifications: Parameter Analyte Requirement Result...................................................................41

Page 5: Product Manual for Acclaim Surfactant · 2008-11-12 · PRODUCT MANUAL for the ACCLAIM SURFACTANT GUARD COLUMN (4.3 x 10mm, P/N 063215) ... Any HPLC pump capable of delivering

Acclaim Surfactant Product Manual Page 4 of 41

SECTION 1 – INTRODUCTION TO THE ACCLAIM SURFACTANT COLUMN

1.1 FEATURES OF THE ACCLAIM SURFACTANT COLUMN

The Acclaim Surfactant column is a specialty silica-based column that incorporates a proprietary bonding chemistry developed specifically for the separation of surfactants. The separation mechanism of the Acclaim Surfactant is a combination of reversed phase, anion-exchange, and dipole-dipole interactions. The Acclaim Surfactant column offers ideal selectivity and unprecedented capability for separating cationic, nonionic, and anionic surfactants in a single chromatographic run, using simple and volatile (MS compatible) mobile phases. Acclaim Surfactant columns consist of 5-μm high purity spherical silica particles with 120 Ǻ diameter pores bonded with proprietary functional groups, which provide unsurpassed resolution and peak symmetry for a variety of surfactants, as shown in Figure 1. The Acclaim Surfactant column provides several features.

• Ideal selectivity for simultaneous separation of anionic, nonionic and cationic surfactants • Excellent peak shapes for cationic surfactants • Improved resolution for oligomers of ethoxylated surfactants • Compatibility of highly aqueous mobile phase conditions • Proven chromatographic conditions suitable for various detection methods

(ELSD, UV, conductivity detection, etc.) • Broad range of applications

Document 065087 Rev. 02 ©Dionex Corporation June 2008

Page 6: Product Manual for Acclaim Surfactant · 2008-11-12 · PRODUCT MANUAL for the ACCLAIM SURFACTANT GUARD COLUMN (4.3 x 10mm, P/N 063215) ... Any HPLC pump capable of delivering

Acclaim Surfactant Product Manual Page 5 of 41

Column: Acclaim Surfactant, 5 µm Dimensions: 4.6 x 150 mm Mobile phase: A –ACN, B- 0.1 M NH4OAc, pH 5.4 Gradient: 25% to 85% A in 30 min., then hold for 10 min Temperature: 30 °C Flow rate: 1 mL/min Injection vol.: 25 µL Detection: ELS Detector

mV

Minutes0 5 10 15 20 25 30 35 40

1

2

3

4

5

6

710

9

8

Analytes:1. Xylene sulfonate2. Lauryldimethylbenzyl

ammonium chloride3. Octylphenoxyethoxyethyl

dimethylbenzyl ammonium chloride

4. Triton X-1005. Decyl sulfate6. Dodecyl sulfate7-10. Dodecylbenzene sulfonate

Figure 1 Separation of Cationic, Nonionic, and Anionic Surfactants

Document 065087 Rev. 02 ©Dionex Corporation June 2008

Page 7: Product Manual for Acclaim Surfactant · 2008-11-12 · PRODUCT MANUAL for the ACCLAIM SURFACTANT GUARD COLUMN (4.3 x 10mm, P/N 063215) ... Any HPLC pump capable of delivering

Acclaim Surfactant Product Manual Page 6 of 41

1.2 ACCLAIM SURFACTANT OPERATING LIMITS AND SPECIFICATIONS

Flow Rate Range: 4.6mm ID <2 mL/min and 2.1mm ID <0.5 mL/min (not to exceed backpressure limit) Shipping Solution: Acetonitrile / 0.1% Acetic Acid v/v 80/20 Storage Solution: Acetonitrile / 0.1% Acetic Acid v/v 80/20 Buffer pH Range: pH 2.5–7.5 Solvents: Acetonitrile, Methanol Temperature Range: <60 °C Pressure Limit: 4000 psi (266 bar) Physical Bonding: Proprietary Endcapping: Yes % C: 12 Pore size: 120 Å Surface area: 300 m2/g Particle size: 5 µm 1.3 FORMATS OF THE ACCLAIM SURFACTANT COLUMN

Acclaim Surfactant columns are available in 4.6 x 150 mm (P/N 063201), 4.6 x 250 mm (P/N 046203) or 2.1 x 150 mm (P/N 068123) formats. The guards (P/N 063215) are a 4.3 x 10 mm cartridge and 2.1 x 10 mm (P/N068124) cartridge packed in a PEEK body with titanium frits. The guard cartridge is held in a Guard Holder (P/N 059526). Acclaim surfactant columns are available by special order in other formats. Inquire for custom ordering information. 1.3.1 Acclaim Surfactant Products

Part Number Product Name Dimensions 046203 Acclaim Surfactant, Analytical Column 4.6 x 250mm 063201 Acclaim Surfactant, Analytical Column 4.6 x 150mm 068123 Acclaim Surfactant, Analytical Column 2.1 x 150mm 063215 Acclaim Surfactant, Guard Cartridge 4.3 x 10mm 068124 Acclaim Surfactant, Guard Cartridge 2.1 x 10mm 059526 Guard Holder and Coupler Kit

Document 065087 Rev. 02 ©Dionex Corporation June 2008

Page 8: Product Manual for Acclaim Surfactant · 2008-11-12 · PRODUCT MANUAL for the ACCLAIM SURFACTANT GUARD COLUMN (4.3 x 10mm, P/N 063215) ... Any HPLC pump capable of delivering

Acclaim Surfactant Product Manual Page 7 of 41

Document 065087 Rev. 02 ©Dionex Corporation June 2008

1.4 ACCLAIM SURFACTANT OPERATING CONDITIONS

The recommended pH range for Acclaim Surfactant columns is between pH 2.5 – 7.5. They are compatible with mobile phases containing 0–100% HPLC solvents, such as methanol or acetonitrile. The Acclaim Surfactant can be operated at any flow rate, as long as the backpressure remains below 4,000psi. When setting up the analytical system, please be sure to check the special precautions listed in Section 3, “Installation”. 1.5 KEY APPLICATIONS OF THE ACCLAIM SURFACTANT COLUMN

The Acclaim Surfactant column is a valuable tool for the analysis of all types of sample matrix, including food and beverages, pharmaceuticals, chemicals, and environmental samples. Applications by Class of Compounds:

• Anionic surfactants • Cationic surfactants • Non-ionic surfactants • Amphoteric surfactants

Application by Industries:

• Food and beverages • Pharmaceuticals • Chemicals • Semiconductors • Environmental

Page 9: Product Manual for Acclaim Surfactant · 2008-11-12 · PRODUCT MANUAL for the ACCLAIM SURFACTANT GUARD COLUMN (4.3 x 10mm, P/N 063215) ... Any HPLC pump capable of delivering

Acclaim Surfactant Product Manual Page 8 of 41

SECTION 2 – CHROMATOGRAPHY SYSTEM

Typical Flow Rate: 1 mL/min for the 4.6mm ID and 0.2 mL/min for the 2.1mm ID Injection Volume: 5–50 µL for the 4.6mm ID and 1-10 µL for the 2.1mm ID Autosampler: AS50 or ASI-100 System Void Volume: Minimize the lengths of all connecting tubing and remove all unnecessary switching

valves and couplers. Pumps: Any HPLC pump capable of delivering gradients Detectors: UV, Conductivity, ELSD, MS, RI 2.1 DETECTORS

The Acclaim surfactant column is compatible with a variety of detectors and UV, ELSD and RI detectors. Example applications using each of these detectors are shown in Section 6 “Applications”. 2.1.1 Evaporative Light Scattering Detection (ELSD)

When different surfactants are present in the sample with concentrations of no less than 50 ppm for each individual surfactant, ELSD is a good option. ELSD is a universal detection method that is compatible with gradient methods and is far more sensitive than RI. In addition, methods developed with ELSD can be easily transferred to LC-ESI-MS applications with little or no modification, because both detectors share the same mobile phase requirements. A UV detector can serve as a complementary tool and, if chosen, should be placed between the analytical column and ELSD.

2.1.2 Suppressed Conductivity Detection (SCD)

Suppressed conductivity detection can also be used for surfactant analysis and provides certain advantages for analyzing trace levels ionic surfactants in complex matrices. In fact, when analyzing ionic surfactants, especially at low concentrations (below 10 ppm), conductivity detection using a Dionex PEEK system, equipped with a gradient pump, electrochemical detector, MMS suppressor, and conductivity cell, is the best approach. Eluents for the separation of surfactants using conductivity detection include borate buffer and acetic acid. If required, a UV detector can serve as a complementary tool that should be placed between the analytical column and the MMS suppressor.

2.1.3 UV Detection

UV absorbance is the most popular detection method in HPLC, due to its ease of use and sensitivity. The drawback with this approach is that the analyte must have a chromophore to be detected and many surfactants do not. The UV detector can be used alone or with other detectors. 2.1.4 Mass Spectrometric Detection (MS)

Mass spectrometry (MS) is an inherently sensitive and universal method and has become the widely accepted tool for characterization of organic compounds. The soft ionization techniques, such as electrospray ionization (ESI), have greatly increased the applicability of MS detection to surfactant analysis. MS detectors are the detector of choice when less than 1 ppm individual surfactant is present and/or identification is required. Mass spectrometry compatible methods using ammonium acetate and acetonitrile have been developed. They are shown in Section 6, “Applications” in this manual.

Document 065087 Rev. 02 ©Dionex Corporation June 2008

Page 10: Product Manual for Acclaim Surfactant · 2008-11-12 · PRODUCT MANUAL for the ACCLAIM SURFACTANT GUARD COLUMN (4.3 x 10mm, P/N 063215) ... Any HPLC pump capable of delivering

Acclaim Surfactant Product Manual Page 9 of 41

MS detection can be used in series with other detectors, such as UV. Place the UV detector between the column and the MS or ELSD. 2.1.5 Refractive Index Detection (RI)

Although refractive index (RI) detection is a universal detection method capable of detecting all analytes, it is incompatible with gradient methods, exhibits low sensitivity, and thus is typically only used when other detection methods are not applicable or available.

2.2 PUMPS

Any HPLC pump can be used for the separation of surfactants with the Acclaim surfactant column. However, if suppressed conductivity detection is chosen, a PEEK pump from Dionex is recommended. The GM-3 Gradient Mixer is also recommended for gradient analysis.

2.3 INJECTORS

Either an autosampler or a manual injector can be used for sample injection.

Document 065087 Rev. 02 ©Dionex Corporation June 2008

Page 11: Product Manual for Acclaim Surfactant · 2008-11-12 · PRODUCT MANUAL for the ACCLAIM SURFACTANT GUARD COLUMN (4.3 x 10mm, P/N 063215) ... Any HPLC pump capable of delivering

Acclaim Surfactant Product Manual Page 10 of 41

Document 065087 Rev. 02 ©Dionex Corporation June 2008

SECTION 3 - INSTALLATION

3.1 SYSTEM REQUIRMENTS

The Acclaim Surfactant column can be run on any HPLC system, including Dionex IC systems, and the UltiMate 3000 HPLC systems. Each of the possible configurations offers multiple sampling options. However, consistently reproducible quantification and an absence of disturbing artifacts are best achieved using the “full loop” mode and in conjunction with a 5 - 50 μL loop for the 4.6mm ID column and 1 - 10µL for the 2.1mm ID column. Optimal reproducibility of retention time results can be achieved by regulating the temperature of the column using a column heater. 3.2 SYSTEM VOID VOLUME

For best performance, minimize the lengths of all connecting tubing and remove all unnecessary switching valves and couplers. If you need assistance in properly configuring your system, contact the nearest Dionex Office (see, “Dionex Worldwide Offices” on the Dionex Reference Library CD-ROM). 3.3 THE INJECTION LOOP

DIONEX recommends using 5-50 µL injection loops for 4.6mm ID columns and 1-10 µL for 2.1mm ID columns. 3.4 COLUMN INSTALLATION

Connect the column with the direction arrow on the label pointing toward the detector. 3.5 COLUMN GUARD

When using a guard cartridge, install the cartridge in the guard holder. Connect the guard holder to the analytical column using the column connector fitting.

Page 12: Product Manual for Acclaim Surfactant · 2008-11-12 · PRODUCT MANUAL for the ACCLAIM SURFACTANT GUARD COLUMN (4.3 x 10mm, P/N 063215) ... Any HPLC pump capable of delivering

Acclaim Surfactant Product Manual Page 11 of 41

SECTION 4 – OPERATION

4.1 MOBILE PHASE SELECTION

4.1.1 Evaporative Light Scattering Detection (ELSD)

ELSD requires volatile buffers, such as ammonium acetate, ammonium formate, acetic acid, or formic acid. 20 to 100 mM buffer concentration is a good starting point. Acetonitrile is the preferred organic modifier because it generates a much lower backpressure on the column than does methanol.

4.1.2 Suppressed Conductivity Detection (SCD)

SCD requires a low conductivity or suppressible mobile phase to achieve high sensitivity. A good eluent for the analysis of cationic surfactants is 10 to 20 mM acetic acid with acetonitrile. For analysis of anionic surfactants, 10 to 30 mM ammonium borate buffer ~pH6 with acetonitrile is a good starting point. Note that when borate buffer is used, acetonitrile (not methanol) should be used as the organic modifier.

4.1.3 UV Detection

The mobile phase for the Acclaim Surfactant column with UV detection should have low UV background. 4.1.4 Mass Spectrometric Detection (MS)

Like ELSD, MS detection requires volatile buffers, such as ammonium acetate, ammonium formate, acetic acid, or formic acid. 20 to 100 mM buffer concentration is a good starting point. Acetonitrile is the preferred organic modifier because it generates a much lower backpressure on the column than does methanol.

4.1.5 Refractive Index Detection (RI)

The mobile phase for the Acclaim Surfactant column with RI detection should have low RI background.

4.2 VALIDATING COLUMN PERFORMANCE

Dionex recommends that you perform an efficiency test on your Acclaim Surfactant column before use. The purpose of column performance validation is to ensure no damage has occurred during shipping. Test the column using the conditions described on the Quality Assurance Report enclosed in the column box (also included in Appendix A of this manual.) Repeat the test periodically to track the column performance over time. Note that slight variations may be obtained on two different HPLC systems due to system electronic, plumbing, operating environment, reagent quality, column conditioning, and operator technique. Please see the example Quality Assurance Report in Appendix A.

Document 065087 Rev. 02 ©Dionex Corporation June 2008

Page 13: Product Manual for Acclaim Surfactant · 2008-11-12 · PRODUCT MANUAL for the ACCLAIM SURFACTANT GUARD COLUMN (4.3 x 10mm, P/N 063215) ... Any HPLC pump capable of delivering

Acclaim Surfactant Product Manual Page 12 of 41

4.2.1 Procedure for Validating Column Performance:

1) Connect the column to the LC system. 2) Purge the column with the mobile phase listed on the QA report for 20 to 40 column volumes. 3) Inject the test mix shown in the QA report and collect the data. 4) Compare your result with the QA report provided in the column box. 5) If the chromatogram is satisfactory, you can use the column for your application work.

4.3 EQUILIBRATING THE COLUMN

Equilibrate the column after installing it for the first time. Always re-equilibrate the column prior to use following periods of storage. Purge the column with the shipping or storage solvent until the baseline is stable. The shipping and recommended storage solvent is acetonitrile / 0.1% acetic acid v/v 80/20. Purge the column with acetonitrile / 0.1% acetic acid v/v 80/20. Then equilibrate the column with at least 15 column volumes of the mobile phase until a stable baseline is achieved. 4.4 QUICKSTART FOR ACCLAIM SURFACTANT COLUMN

4.4.1 Using the chromatographic conditions for Column Quality Assurance Report:

Column: 5 micron particle size Mobile phase: Acetonitrile/0.1M NH4OAc, pH5.4 v/v 50/50 Flow rate: 4.6 x 150 mm use 1 mL/min ; 2.1 x 150mm use 0.2 mL/min Temperature: 30 ˚C Injection vol.: 5 µL Detection: UV at 220 nm

4.4.2 Column Quickstart

1) Prepare the mobile phase (see Section 4.7). 2) Install the mobile phase, 0.1% acetic acid, DI water, and acetonitrile on the LC pump. 3) Turn on the UV detector and set it to 220nm. 4) Prime the pump. 5) Install the column. 6) Flush the column with at least 15 column volumes of acetonitrile / 0.1% acetic acid v/v 80/20, then 10

column volumes of acetonitrile / 0.1% acetic acid v/v 50/50. 7) Equilibrate the column under initial mobile phase conditions for at least 15 minutes. 8) Conduct tests. 9) At the end of the day, flush the column with acetonitrile / 0.1% acetic acid v/v 50/50 for approximately 10

minutes. 10) Flush the column with acetonitrile / 0.1% acetic acid v/v 80/20 for approximately 10 minutes. 11) Turn off the pump and detector. 12) Remove the column and plug the ends, if the column will not be used for more than 2–3 days.

*For short-term storage (such as overnight), you may store the column in mobile phase.

Document 065087 Rev. 02 ©Dionex Corporation June 2008

Page 14: Product Manual for Acclaim Surfactant · 2008-11-12 · PRODUCT MANUAL for the ACCLAIM SURFACTANT GUARD COLUMN (4.3 x 10mm, P/N 063215) ... Any HPLC pump capable of delivering

Acclaim Surfactant Product Manual Page 13 of 41

4.5 CARING FOR THE COLUMN

To ensure the high performance of the Acclaim Surfactant column, the following guidelines should be followed:

1) Protect the column from contamination using an Acclaim Surfactant guard cartridge, 2) Make sure that solvents are miscible when changing mobile phases. 3) Always degas and filter mobile phases through a 0.22-μm membrane filter. 4) When switching to a new mobile phase, the column should be equilibrated for at least 30 column

volumes before injecting the sample. 5) The recommended pH range is from pH 2.5 to 7.5. However, it is preferred that the column be used

between pH3 and pH6.5 to achieve longer lifetime. 6) The column can be stored in mobile phase for short time storage (e.g. overnight). However, it is

highly recommended that the column be stored in acetonitrile / 10-50 mM NH4OAc buffer (pH 4.0-5.0) v/v 80/20 to 50/50 for long-term storage. If the mobile phase contained a buffer, first flush the column with 10 column volumes of acetonitrile / 10 mM NH4OAc buffer (pH 5) v/v 50/50 before changing over to acetonitrile / 10-50 mM NH4OAc buffer (pH 4.5) v/v 80/20. Do not store the column under highly aqueous conditions for an extended period of time.

7) The recommended operating maximum temperature is below 60 ˚C. In most cases, temperature control between ambient and 30 ˚C gives good results.

8) The recommended maximum back pressure is 4000 psi. 4.6 CHEMICAL PURITY REQUIREMENTS

Obtaining reliable, consistent and accurate results require mobile phases that are free of ionic and spectrophotometric impurities. Chemicals, solvents and de-ionized water used to prepare mobile phase should be of the highest purity available. Maintaining low trace impurities and low particle levels in mobile phases helps to protect your columns and system components. DIONEX cannot guarantee proper column performance when the quality of the chemicals, solvents and water used to prepare the mobile phase has been compromised. 4.6.1 Inorganic Chemicals

Reagent Grad inorganic chemicals should always be used to prepare ionic eluents. Whenever possible, inorganic chemicals that meet or surpass the latest American Chemical Society standard for purity should be used. These inorganic chemicals should be labeled, with the purity included on the lot analysis. Some recommended reagents are HPLC grade. 4.6.2 De-ionized Water

The de-ionized water used to prepare the mobile phase should be Type I Reagent Grade Water with a specific resistance of 18.2 megohm-cm. The de-ionized water should be free of ionized impurities, organics, microorganisms and particulate matter larger than 0.2 μm. Many commercial water purifiers are designed for HPLC applications and are suitable for these applications.

Document 065087 Rev. 02 ©Dionex Corporation June 2008

Page 15: Product Manual for Acclaim Surfactant · 2008-11-12 · PRODUCT MANUAL for the ACCLAIM SURFACTANT GUARD COLUMN (4.3 x 10mm, P/N 063215) ... Any HPLC pump capable of delivering

Acclaim Surfactant Product Manual Page 14 of 41

Document 065087 Rev. 02 ©Dionex Corporation June 2008

4.6.3 Solvents

The solvents used must be free from ionic and UV-absorbing impurities. However, since most manufacturers of solvents do not test for ionic impurities, it is important that the highest grade of solvents available be used. Currently, several manufacturers manufacture ultrahigh purity solvents that are compatible with HPLC and spectrophotometric applications. Use of these ultrahigh purity solvents will usually ensure that your chromatography is not affected by ionic impurities in the solvent. Currently at Dionex, we have obtained consistent results using High Purity Solvents manufactured by Burdick and Jackson and Optima® Solvents by Fischer Scientific. When using a solvent in an ionic mobile phase, column generated backpressure will depend on the solvent used, concentration of the solvent, the ionic strength of the buffer and the flow rate used. The column back pressure will vary as the composition or the water-methanol or water-acetonitrile mixture varies. The practical backpressure limit for the Acclaim Surfactant columns is 4,000 psi (27.57 MPa).

NOTE Since aqueous-methanol mixtures result in much higher back-pressures, aqueous-acetonitrile mobile phases

in any ratio are preferred and recommended.

4.7 MOBILE PHASE PREPARATION

4.7.1 0.1 M Ammonium Acetate Buffer (1 L)

Dissolve 7.8 g ammonium acetate (43,131-1, Aldrich) in 1000 g D.I. water. Adjust the pH to the final value using acetic acid or ammonium hydroxide. 4.7.2 20 mM Sodium Borate Buffer (1 L)

Combine 200 mL of 0.1N sodium hydroxide solution (31,948-1, Aldrich) and 800 mL D.I. water and mix thoroughly. Add 31 g boric acid (>99.5%, 15660, Fluka) and sonicate the mixture under vacuum for 5 minutes. 4.7.3 20 mM Acetic Acid Buffer (1 L)

Dissolve 1.2 g of acetic acid (9524-33, J.T. Baker) in 1000 g DI water followed by sonication under vacuum for 5 minutes. 4.7.4 Mobile Phases Containing Solvents

When mixing solvents with water, mix the solvent with water on a volume-to-volume basis. For example, if a procedure requires a mobile phase of 10% acetonitrile, prepare the mobile phase by adding 100 mL of acetonitrile to a mobile phase reservoir. Then add 900 mL of the aqueous portion (DI water or buffer) to the acetonitrile in the reservoir. Using this procedure to mix the solvent with the aqueous phase will ensure that a consistent true volume/volume mobile phase is obtained. Premixing water with solvent will minimize the possibility of out-gassing in the detector cell.

NOTE Degas the aqueous component of the mobile phase and then add to the solvent component. Avoid excessive purging or degassing of mobile phases containing solvents, if possible, since a volatile solvent can be ‘boiled’

off from the solution.

Page 16: Product Manual for Acclaim Surfactant · 2008-11-12 · PRODUCT MANUAL for the ACCLAIM SURFACTANT GUARD COLUMN (4.3 x 10mm, P/N 063215) ... Any HPLC pump capable of delivering

Acclaim Surfactant Product Manual Page 15 of 41

SECTION 5 – METHODS DEVELOPMENT

The Acclaim Surfactant column is compatible with a wide range of mobile phase systems that are compatible with UV, ELSD, MS, and conductivity detection. Not only can it be used as a QA column for the analysis of individual surfactants, but it is also valuable for the analysis of surfactants in complex sample matrices. In addition, by changing the composition of the mobile phase with respect to ionic strength or pH, the Acclaim Surfactant column can be used to identify the nature of species of interest in unknown sample, as shown in Figures 4 and 5.

NOTE The Acclaim Surfactant column features reversed-phase anion-exchange and hydrogen-bonding

interaction. Thus, it should not be used in the same way as conventional reversed-phase columns. 5.1 IONIC STRENGTH

The ionic strength is a powerful selectivity modifier for separating surfactants. Increasing the ionic strength of the mobile phase produces several results: a decrease in the retention time for anionic surfactants, an increase in the retention for cationic surfactants, but no change for nonionic surfactants. In this way, a clear distinction between the different classes of surfactants present in the sample can be made, as shown in Figure 3.

Column: Acclaim® Surfactant, 5 µmDimensions: 4.6 x 150 mmMobile Phase: (A) ACN, (B) 0.1 M NH4OAc,

pH 5.4Gradient: 25% to 85% A in 30 min Temperature: 30 °CFlow Rate: 1 mL/minInj. Volume: 25 µLDetection: ELSD

Peaks:1. Laurylpyridium chloride2. Lauryldimethylbenzyl ammonium chloride3. Triton X-1004. Decyl sulfate5. Dodecyl sulfate

Minutes

5 10 15 30

mV

2520

1

2

3 4

5

1

2

3 4

5

50 mM NH4OAc

100 mM NH4OAc

Minutes

5 10 15 30

mV

2520

1

2

3 4

5

1

2

3 4

5

50 mM NH4OAc

100 mM NH4OAc

Figure 3

Effect of Ionic Strength on Selectivity

Document 065087 Rev. 02 ©Dionex Corporation June 2008

Page 17: Product Manual for Acclaim Surfactant · 2008-11-12 · PRODUCT MANUAL for the ACCLAIM SURFACTANT GUARD COLUMN (4.3 x 10mm, P/N 063215) ... Any HPLC pump capable of delivering

Acclaim Surfactant Product Manual Page 16 of 41

5.2 MOBILE PHASE pH

Another powerful selectivity modifier is the mobile phase pH. Increasing the pH of the mobile phase produces several results: a decrease in the retention time of anionic surfactants, an increase in the retention time of cationic surfactants, but no change in the retention time of non-ionic surfactants. These effects are shown in Figures 4 - 5.

Column: Acclaim® Surfactant, 5 µmDimensions: 4.6 x 150 mmMobile Phase: CH3CN/0.1 M NH4OAc, pH 5.4

v/v, 70/30Temperature: 30 °CFlow Rate: 1 mL/minInj. Volume: 10 µLDetection: ELSD

Peaks:1. Triton X-1002. Decyl Sulafate3. Dodecyl Sulfate

1 2

pH5.4

pH4.5

pH3.6

0 5 10 15 20

Minutes

mV

3

12

3

1

2

3

Figure 4 Effect of pH on the Selectivity of Anionic Surfactants

Column: Acclaim® Surfactant, 5 µmDimensions: 4.6 x 150 mmMobile Phase: CH3CN/0.1 M NH4OAc, pH 5.4

v/v, 40/60Temperature: 30 °CFlow Rate: 1 mL/minInj. Volume: 10 µLDetection: ELSD

Peaks:1. Lauryldimethylbenzyl ammonium chloride2. Octylphenoxyethoxyethyl dimethylbenzyl

ammonium chloride

0 4 8 12

Minutes

mV

1 2

1 2

1 2

pH5.4

pH4.5

pH3.6

Figure 5 Effect of pH on the Selectivity of Cationic Surfactants

Document 065087 Rev. 02 ©Dionex Corporation June 2008

Page 18: Product Manual for Acclaim Surfactant · 2008-11-12 · PRODUCT MANUAL for the ACCLAIM SURFACTANT GUARD COLUMN (4.3 x 10mm, P/N 063215) ... Any HPLC pump capable of delivering

Acclaim Surfactant Product Manual Page 17 of 41

Document 065087 Rev. 02 ©Dionex Corporation June 2008

5.3 ORGANIC MODIFIER

Increasing the concentration of the organic modifier in the mobile phase generally results in a decrease in the retention time for all classes of surfactant. Acetonitrile or methanol is the recommended organic modifier. Acetonitrile is preferred over methanol because it generates less backpressure on the column, compared with methanol. 5.4 ION-PAIRING AGENT

The use of different ion-pairing agents affects the retention times of both cationic and anionic surfactants. For example, the use of tetramethyl ammonium acetate results in an increase in the retention time for anionic surfactants compared with the use of ammonium acetate. Similarly, the use of larger anions in the mobile phase gives rise to an increase in the retention time for cationic surfactants. 5.5 TEMPERATURE

Temperature has little effect on the selectivity of surfactants using the Acclaim Surfactant column; all the surfactants are eluted earlier. Since there is no advantage to higher temperatures, it is recommended that the Acclaim Surfactant column be used in a controlled environment, between 25 oC and 35 oC to ensure consistent retention times. Operation above 50 oC will result in an increased background signal using ELSD or MS detection.

Page 19: Product Manual for Acclaim Surfactant · 2008-11-12 · PRODUCT MANUAL for the ACCLAIM SURFACTANT GUARD COLUMN (4.3 x 10mm, P/N 063215) ... Any HPLC pump capable of delivering

Acclaim Surfactant Product Manual Page 18 of 41

SECTION 6 – APPLICATIONS

6.1 ANIONIC SURFACTANTS

6.1.1 Linear Alkylbenzene Sulfonates

Column: See chromatograms Dimensions: 4.6 x 150 mm Temperature: 30 °C Flow rate: 1 mL/min Injection vol.: 10 µL Detection: UV, 225 nm Mobile phases: For Acclaim Surfactant - ACN/100 mM NH4OAc, pH 5.4 v/v 70/30 For Acclaim PA - ACN/100 mM NH4OAc, pH 5.4 v/v 50/50

C (CH2)m

H

CH3

SO3Na

(CH2)n CH3

Sodium Dodecylbenzene Sulfonatem + n = 10

0 5 10 15 20 24

Acclaim® Surfactant

Acclaim® PA

Minutes

AU

C10

C11

C12

C13

C10

C11

C12

C13

Figure 6 Separation of Linear Alkylbenzene Sulfonates

Document 065087 Rev. 02 ©Dionex Corporation June 2008

Page 20: Product Manual for Acclaim Surfactant · 2008-11-12 · PRODUCT MANUAL for the ACCLAIM SURFACTANT GUARD COLUMN (4.3 x 10mm, P/N 063215) ... Any HPLC pump capable of delivering

Acclaim Surfactant Product Manual Page 19 of 41

6.1.2 Sodium Lauryl Sulfate

Column: Acclaim Surfactant, 5 µm Dimension: 4.6 x 150 mm Mobile Phase: CH3CN/0.1 M NH4OAc v/v 75/25 Temperature: 25 °C Flow Rate: 1 mL/min (0.2 mL/min to MS) Inj. Volume: 25 µL Detection: MS (ESI negative) Sample: SLS (1 ppm)

0 5 10 15

Minutes

Abun

danc

e

O - S - O Na

O

O

Sodium Lauryl Sulfate

Figure 7

Analysis of Sodium Lauryl Sulfate (LC-ESI-MS)

Document 065087 Rev. 02 ©Dionex Corporation June 2008

Page 21: Product Manual for Acclaim Surfactant · 2008-11-12 · PRODUCT MANUAL for the ACCLAIM SURFACTANT GUARD COLUMN (4.3 x 10mm, P/N 063215) ... Any HPLC pump capable of delivering

Acclaim Surfactant Product Manual Page 20 of 41

6.1.3 POE(30) Lauryl Sulfate Column: Acclaim Surfactant, 5 µm Dimension: 4.6 x 150 mm Mobile Phase: CH3CN/0.5 M H3BO4 and 20 mM NaOH v/v 55/45 Temperature: 30 °C Flow Rate: 1 mL/min Inj. Volume: 10 µL Detection: Suppressed CD (AMMS® III) Sample: POE(30) Lauryl Sulfate (1000 ppm)

0 10 20 30

0

4

Minutes

µS

(OCH2CH2)n- O-S-O Na

O

O

Sodium POE (30) Lauryl Sulfate

n ~ 30

Analysis of POE(30) Lauryl Sulfate with Suppressed Conductivity Detection

Figure 8

Document 065087 Rev. 02 ©Dionex Corporation June 2008

Page 22: Product Manual for Acclaim Surfactant · 2008-11-12 · PRODUCT MANUAL for the ACCLAIM SURFACTANT GUARD COLUMN (4.3 x 10mm, P/N 063215) ... Any HPLC pump capable of delivering

Acclaim Surfactant Product Manual Page 21 of 41

6.2 CATIONIC SURFACTANTS

6.2.1 Cationic Surfactants with ELS Detection

Column: Acclaim Surfactant, 5 µm Dimensions: 4.6 x 150 mm Mobile Phase: (A) Acetonitrile (B) 0.1 M NH4OAc, pH 5.4 Gradient: 25% to 85% A in 30 min Temperature: 30 °C Flow Rate: 1 mL/min Inj. Volume: 25 µL Detection: ELSD

Peaks:1. Lauryl pyridinium chloride2. Lauryldimethylbenzyl ammonium chloride3. Octylphenoxyethoxyethyl dimethylbenzyl

ammonium chloride5. Cetyltrimethylammonium chloride6. Cetylpyridinium chloride

0 10 20 30

1

2

3

4

5

Minutes

mV

Figure 9

Analysis of Cationic Surfactants with Evaporative Light Scattering Detection

Document 065087 Rev. 02 ©Dionex Corporation June 2008

Page 23: Product Manual for Acclaim Surfactant · 2008-11-12 · PRODUCT MANUAL for the ACCLAIM SURFACTANT GUARD COLUMN (4.3 x 10mm, P/N 063215) ... Any HPLC pump capable of delivering

Acclaim Surfactant Product Manual Page 22 of 41

6.2.2 Cationic Surfactants with Suppressed Conductivity Detection

Column: Acclaim Surfactant, 5 µm Dimensions: 4.6 x 150 mm Mobile Phase: (A) Acetonitrile (B) 20 mM Acetic acid Gradient: 20-50% A in 15 min Temperature: 30 °C Flow Rate: 1 mL/min Inj. Volume: 10 µL Detection: Suppressed CD (CMMS® III)

0 5 10 15

12

3

4

5

Minutes

µS

Peaks: 1. Lauryl pyridinium chloride2. Lauryldimethylbenzyl ammonium chloride3. Octylphenoxyethoxyethyl

dimethylbenzyl ammonium chloride4. Cetyltrimethylammonium chloride5. Cetylpyridinium chloride

Figure 10 Separation of Cationic Surfactants with Suppressed Conductivity Detection

Document 065087 Rev. 02 ©Dionex Corporation June 2008

Page 24: Product Manual for Acclaim Surfactant · 2008-11-12 · PRODUCT MANUAL for the ACCLAIM SURFACTANT GUARD COLUMN (4.3 x 10mm, P/N 063215) ... Any HPLC pump capable of delivering

Acclaim Surfactant Product Manual Page 23 of 41

6.2.3 Benzalkonium Salts

Column: Acclaim Surfactant, 5 µm Dimensions: 4.6 x 150 mm Mobile Phase: Acetronitrile/ 0.1 M NH4OAc, pH 5.4 v/v 60/40 Temperature: 25 °C Flow Rate: 1 mL/min (0.2 mL/min to MS) Inj. Volume: 5 µL Detection: MS (ESI positive) Peak: Benzyl dimethyl hydrogenated tallow ammonium chloride (10 ppm)

C18

C16

C17C15

Abun

danc

e

0 5 10 15

Minutes

R

CH3

CH3

CH2N

R = C14 to C18

+

-Cl

Benzalkonium Salts

Figure 11 Analysis of Benzalkonium Salts with LC-ESI-MS Detection

Document 065087 Rev. 02 ©Dionex Corporation June 2008

Page 25: Product Manual for Acclaim Surfactant · 2008-11-12 · PRODUCT MANUAL for the ACCLAIM SURFACTANT GUARD COLUMN (4.3 x 10mm, P/N 063215) ... Any HPLC pump capable of delivering

Acclaim Surfactant Product Manual Page 24 of 41

6.2.4 Quaternary Imidazolium Compounds

Column: Acclaim Surfactant, 5 µm Dimensions: 4.6 x 150 mm Mobile Phase: (A) Acetonitrile (B) 0.1 M NH4OAc, pH 5.4 Gradient: 25–85% A in 30 min Temperature: 30 °C Flow Rate: 1 mL/min Inj. Volume: 10 µL Detection: ELSD

Diethyl heptadecyl imidazolinium ethylsulfate

2-Hydroxyethylbenzyl coco imidazolinium chloride

0 5 10 15 20 25 30

Minutes

mV

N

N

C

R1

R3 +

O

R3

X -

CH2CH2NHC

Quaternary Imidazolium Compounds

Figure 12 Analysis of Imidazolium Compounds with Evaporative Light Scattering Detection

Document 065087 Rev. 02 ©Dionex Corporation June 2008

Page 26: Product Manual for Acclaim Surfactant · 2008-11-12 · PRODUCT MANUAL for the ACCLAIM SURFACTANT GUARD COLUMN (4.3 x 10mm, P/N 063215) ... Any HPLC pump capable of delivering

Acclaim Surfactant Product Manual Page 25 of 41

6.3 NON-IONIC SURFACTANTS

6.3.1 Ethoxylated Quaternary Compounds

Column: Acclaim Surfactant, 5 µm Dimensions: 4.6 x 150 mm Mobile Phase: Acetonitrile/0.2 M NH4OAc, pH 5.4 v/v 40/60 Temperature: 30 °C Flow Rate: 1 mL/min Inj. Volume: 25 µL Detection: ELSD Analyte: Ethoquad 18/25 (0.2%)

0 10 20 30 40

Minutes

mV

Figure 13

Analysis of Ethoxylated Quats

Document 065087 Rev. 02 ©Dionex Corporation June 2008

Page 27: Product Manual for Acclaim Surfactant · 2008-11-12 · PRODUCT MANUAL for the ACCLAIM SURFACTANT GUARD COLUMN (4.3 x 10mm, P/N 063215) ... Any HPLC pump capable of delivering

Acclaim Surfactant Product Manual Page 26 of 41

6.3.2 PEG Monoethyl Ether (MW-550)

Column: Acclaim Surfactant, 5 µm Dimensions: 4.6 x 150 mm Mobile Phase: (A) Acetonitrile (B) H2O Gradient: 5–20% A in 25 min Temperature: 30 °C Flow Rate: 1 mL/min Inj.Volume: 10 µL Detection: ELSD

PEG Monoethyl Ether (MW-550)

Et (OCH2CH2)nOH

n ~ 12

Minutes

mV

0 10 20155 25

Figure 14 Analysis of PEG Monoethyl Ether (MW-550)

Document 065087 Rev. 02 ©Dionex Corporation June 2008

Page 28: Product Manual for Acclaim Surfactant · 2008-11-12 · PRODUCT MANUAL for the ACCLAIM SURFACTANT GUARD COLUMN (4.3 x 10mm, P/N 063215) ... Any HPLC pump capable of delivering

Acclaim Surfactant Product Manual Page 27 of 41

6.3.3 Triton X-100

Column: Acclaim Surfactant, 5 µm Dimensions: 4.6 x 150 mm Mobile Phase: CH3CN/50mM NH4OAc, pH5.4 v/v 45/55 Temperature: 25 °C Flow Rate: 1 mL/min (0.2 mL/min to MS) Inj. Volume: 25 µL Detection: MS (ESI positive) Sample: Triton X-100 (100 ppm)

Triton X-100

n ~ 10

C8H17 (OCH2CH2)nOH

0 10 20155 2

Minutes

Abun

danc

e

n = 10

n = 9

n = 8

n = 7

n = 6

n = 5

n = 4

n = 14

5

n = 17

Figure 15

Analysis of Triton X-100 (LC-ESI-MS)

Document 065087 Rev. 02 ©Dionex Corporation June 2008

Page 29: Product Manual for Acclaim Surfactant · 2008-11-12 · PRODUCT MANUAL for the ACCLAIM SURFACTANT GUARD COLUMN (4.3 x 10mm, P/N 063215) ... Any HPLC pump capable of delivering

Acclaim Surfactant Product Manual Page 28 of 41

6.3.4 NEODOL 25-7

Column: Acclaim Surfactant, 5 µm Dimensions: 4.6 x 150 mm Mobile Phase: (A) Acetonitrile (B) H2O Gradient: 40–65% A in 30 min; then hold at 65%, A for 5 min Temperature: 30 °C Flow Rate: 1 mL/min Inj. Volume: 10 µL Detection: ELSD

NEODOL 25–7

R (OCH2CH2)nOH

R = C12 to C15

n ~ 7

0 10 20 30 35Minutes

mV

Figure 16

Analysis of Ethoxylated Alcohol (NEODOL 25-7)

Document 065087 Rev. 02 ©Dionex Corporation June 2008

Page 30: Product Manual for Acclaim Surfactant · 2008-11-12 · PRODUCT MANUAL for the ACCLAIM SURFACTANT GUARD COLUMN (4.3 x 10mm, P/N 063215) ... Any HPLC pump capable of delivering

Acclaim Surfactant Product Manual Page 29 of 41

6.3.5 ZONYL FSO Perfluorosurfactant

Column: Acclaim Surfactant, 5 µm Dimensions: 4.6 x 150 mm Mobile Phase: (A) Acetonitrile (B) H2O Gradient: 40–70% A in 30 min; then hold at 70% A for 5 min Flow Rate: 1 mL/min Inj. Volume: 10 µL Detection: ELSD

Zonyl FSO fluorosurfacant

RfCH2CH2O(CH2CH2O)xH

Rf = F(CF2CF2)yx = 0 to about 15y = 1 to about 7

0 10 20 30 35

Minutes

mV

Figure 17 Analysis of ZONYL FSO Perfluorosurfactant

Document 065087 Rev. 02 ©Dionex Corporation June 2008

Page 31: Product Manual for Acclaim Surfactant · 2008-11-12 · PRODUCT MANUAL for the ACCLAIM SURFACTANT GUARD COLUMN (4.3 x 10mm, P/N 063215) ... Any HPLC pump capable of delivering

Acclaim Surfactant Product Manual Page 30 of 41

6.3.6 Polyethylene Glycols (PEG)

Column: Acclaim Surfactant, 5 µm Dimensions: 4.6 x 150 mm Mobile phase: ACN/0.1M NH4OAc, pH 5.4 v/v 60/40 Temperature: 30 °C Flow rate: 1 mL/min Injection vol.: 10 µL Detection: ELSD

mV

Minutes0 10 20 30

Figure 18 POE (30) Ammonium Lauryl Sulfate

Document 065087 Rev. 02 ©Dionex Corporation June 2008

Page 32: Product Manual for Acclaim Surfactant · 2008-11-12 · PRODUCT MANUAL for the ACCLAIM SURFACTANT GUARD COLUMN (4.3 x 10mm, P/N 063215) ... Any HPLC pump capable of delivering

Acclaim Surfactant Product Manual Page 31 of 41

6.4 SURFACTANTS IN COMMERCIAL PRODUCTS

6.4.1 Laundry Detergent

Column: Acclaim Surfactant, 5 µm Dimensions: 4.6 x 150 mm Mobile Phase: (A) Acetonitrile (B) 0.1 M NH4OAc, pH 5.4 Gradient: 25–80% A in 30 min., then hold at 80% A for 15 min. Temperature: 30 °C Flow Rate: 1 mL/min Inj. Volume: 5 µL Detection: ELSD with UV at 225 nm Sample Prep: Dilute 10x with 70% acetonitrile, then filtered through 0.2-µm membrane

0 5 10 15 20 25 30 35 40 45Minutes

UV, 225 nmDodecylbenzene Sulfonate

ELSDDodecylbenzene Sulfonate and Laureth Sulfate

Figure 19

Analysis of a Laundry Detergent

Document 065087 Rev. 02 ©Dionex Corporation June 2008

Page 33: Product Manual for Acclaim Surfactant · 2008-11-12 · PRODUCT MANUAL for the ACCLAIM SURFACTANT GUARD COLUMN (4.3 x 10mm, P/N 063215) ... Any HPLC pump capable of delivering

Acclaim Surfactant Product Manual Page 32 of 41

6.4.2 Shampoo

Column: Acclaim Surfactant, 5 µm Dimensions: 4.6 x 150 mm Mobile Phase: (A) Acetonitrile (B) 0.1 M NH4OAc, pH 5.4 Gradient: 25–80% A in 30 min; then hold at 80% A for 15 min. Temperature: 30 °C Flow Rate: 1 mL/min Inj. Volume: 10 µL Detection: ELSD and UV

UV, 225 nm

ELSDLaureth Sulfate

0 5 10 15 20 25 30 35 40 45

Minutes

Figure 20 Analysis of Surfactants in a Shampoo

Document 065087 Rev. 02 ©Dionex Corporation June 2008

Page 34: Product Manual for Acclaim Surfactant · 2008-11-12 · PRODUCT MANUAL for the ACCLAIM SURFACTANT GUARD COLUMN (4.3 x 10mm, P/N 063215) ... Any HPLC pump capable of delivering

Acclaim Surfactant Product Manual Page 33 of 41

6.4.3 Fabric Softener

Column: Acclaim Surfactant, 5 µm Dimensions: 4.6 x 150 mm Mobile Phase: (A) Acetonitrile (B) THF (C) 0.1 M NH4OAc, pH5.4 Gradient: 40% A/10% B/50% C to 75% A/10% B/75% C in 20 min Temperature: 30 °C Flow Rate: 1 mL/min Inj. Volume: 20 µL Detection: ELSD Peaks: N-methyl bis[ethyl(tallowate)]-2-hydroxyethyl ammonium methyl sulfate

0 10 20 30

Minutes

mV

OCH2CH2

O

O

+

CH2CH2OH

OCH2CH2NCH3

CH3OSO3-

Figure 21

Analysis of Surfactants in a Fabric Softener

Document 065087 Rev. 02 ©Dionex Corporation June 2008

Page 35: Product Manual for Acclaim Surfactant · 2008-11-12 · PRODUCT MANUAL for the ACCLAIM SURFACTANT GUARD COLUMN (4.3 x 10mm, P/N 063215) ... Any HPLC pump capable of delivering

Acclaim Surfactant Product Manual Page 34 of 41

Document 065087 Rev. 02 ©Dionex Corporation June 2008

SECTION 7 – TROUBLESHOOTING GUIDE

The following instructions should help you to locate and eliminate problems traceable to hardware and chemistry issues. It also provides a selection of cleanup and reconditioning procedures that have been found effective by many users. Please keep in mind that some problems may be due to other reasons, such as sample contamination, poor water quality, etc. If you cannot solve your problem with the help of this manual, please contact the nearest DIONEX office (see, “DIONEX Worldwide Offices” on the Dionex Reference Library CD-ROM).

7.1 HIGH BACKPRESSURE

7.1.1 Finding the Source of High System Pressure

If the system pressure is excessively high, determine the cause of the high pressure. The system should be used with a high-pressure in-line filter for mobile phases. The filter should be positioned between the gradient pump pressure transducer and the injection valve. Make sure you have a high-pressure in-line filter in place and that it is not contaminated.

A. Make sure that the pump is set to the correct flow rate. Higher than recommended mobile phase flow rates will cause higher pressure. Measure the pump flow rate if necessary with a stop watch and graduated cylinder.

B. Determine which part of the system is causing the high pressure. It could be a piece of tubing that has plugged, collapsed tubing walls from over tightening, an injection valve with a plugged port, a column with particulates plugging the bed support, a plugged High-pressure In-line filter, or the detector cell. To identify which part of the chromatographic system is causing the problem, disconnect the pump fluid line from the injection valve and turn the pump on. Watch the pressure; it should not exceed 50 psi (0.34 MPa). Continue adding the system components back into the fluid path, one by one, while watching the system pressure.

7.1.2 Replacing Column Bed Support Assemblies

If the column inlet bed support is determined to be the cause of the high back pressure, it should be replaced. To change the inlet bed support assembly, refer to the following instructions, using one of the two spare inlet bed support assemblies included in the column box.

A. Disconnect the column from the system. B. Carefully unscrew the inlet (top) column fitting. Use two open-end wrenches. If the column

packing appears soupy, carefully remove excess moisture by touching a paper towel to the edge of the packing bed. Allow water (not bonded phase) to soak into the paper towel.

C. Remove the bed support assembly. Turn the end fitting over and tap it against a bench top or

other hard, flat surface to remove the bed support and seal assembly. If the bed support must be pried out of the end fitting, use a sharp pointed object such as a pair of tweezers, but be careful that you DO NOT SCRATCH THE WALLS OF THE ENDFITTING. Discard the old bed support assembly.

D. Place a new bed support assembly into the end fitting. Make sure that the end of the column tube

is clean and free of any particulate matter so that it will properly seal against the bed support assembly. Use the end of the column to carefully start the bed support assembly into the end fitting.

Page 36: Product Manual for Acclaim Surfactant · 2008-11-12 · PRODUCT MANUAL for the ACCLAIM SURFACTANT GUARD COLUMN (4.3 x 10mm, P/N 063215) ... Any HPLC pump capable of delivering

Acclaim Surfactant Product Manual Page 35 of 41

Document 065087 Rev. 02 ©Dionex Corporation June 2008

CAUTION

If the column tube end is not clean when inserted into the end fitting, particulate matter may obstruct a proper seal between the end of the column tube and the bed support assembly. If this is the case, additional tightening may not seal the column but instead damage the column tube or the end fitting. Carefully wipe the sealing surfaces clean before assembling.

E. Screw the end fitting back onto the column. Tighten the end fitting finger-tight, then an

additional 1/4 turn (25 in. lb). Tighten further only if leaks are observed. F. Reconnect the column to the system and resume operation.

NOTE

DO NOT attempt to remove the column outlet end fitting because the packing will extrude out of the column and ruin it.

7.2 HIGH BACKGROUND OR NOISE

7.2.1 Preparation of Mobile Phases

A. Ensure the mobile phase is made correctly.

B. Ensure the eluents are made from chemicals with the recommended purity. C. Ensure the deionized water used to prepare the reagents has a specific resistance of 18.2 megohm-

cm.

7.2.2 A Contaminated Guard or Analytical Column

Remove the Acclaim Surfactant Analytical Column from the system. Connect the fluid lines to a piece of back pressure tubing. If the background decreases, then the column is the cause of the high background. See Section 4.4.2, “Column Quickstart” Steps 6 through 10. 7.2.3 Contaminated Hardware

To eliminate the hardware as the source of the high background, pump deionized water with a specific resistance of 18.2 megohmcm through the system. If it is not, check the detector cell by injecting deionized water directly into it. 7.3 POOR PEAK RESOLUTION

Poor peak resolution can be due to any or all of the following factors. 7.3.1 Loss of Column Efficiency

A. Check to see if headspace has developed in the analytical column. This may be due to improper use of the column such as submitting it to high pressures. Remove the column’s top end fitting (see Section 7.1.2, “Replacing Column Bed Support Assemblies”). If the resin does not fill the column body all the way to the top, the resin bed has collapsed, creating a headspace. Headspace of 1–2 mm is the maximum allowable before the column demonstrates significant losses of efficiency. If more than 2 mm of headspace is observed, the column must be replaced.

Page 37: Product Manual for Acclaim Surfactant · 2008-11-12 · PRODUCT MANUAL for the ACCLAIM SURFACTANT GUARD COLUMN (4.3 x 10mm, P/N 063215) ... Any HPLC pump capable of delivering

Acclaim Surfactant Product Manual Page 36 of 41

Document 065087 Rev. 02 ©Dionex Corporation June 2008

B. Extra-column system effects can result in sample band dispersion, decreasing peak efficiencies. Make sure you are using tubing with an i.d. of no greater than 0.010" to make all liquid line connections between the injection valve and the detector cell inlet on standard bore (4-mm) systems. Check for leaks.

7.3.2 Poor Resolution Due to Shortened Retention Times

Even with adequate system and column efficiency, resolution of peaks will be compromised if analytes elute too fast.

A. Check the eluent flow rate. Determine if the actual flow rate is different than the flow rate specified by the analytical protocol. Measure the eluent flow rate after the column using a stopwatch and graduated cylinder. Allow the system to equilibrate for at least 5 minutes before making the measurement to allow time for the pump pressure feedback to engage.

B. Check to see if the mobile phase compositions and concentrations are correct. For isocratic analysis, a mobile phase buffer component that is too strong will cause the peaks to elute later. Prepare fresh buffer solution. If you are using a gradient pump to proportion the final mobile phase from concentrated solutions in two or three different reservoirs, the composition of the final mobile phase may not be accurate enough for the application. Use one reservoir containing the correct final mobile phases composition to see if proportioning accuracy is the problem. This may be a problem when one of the concentrated solutions is proportioned at less than 5%.

C. Column contamination can lead to a loss of column efficiency. Refer to Section 4.5, “Caring for the Column”, for recommended column cleanup procedures. Possible sources of column contamination are impurities in chemicals, in the deionized water, or from the sample matrix being used. Be especially careful to make sure that chemicals with recommended purity are used. The deionized water should have a specific resistance of at least 18.2 megohm-cm.

If you need assistance in solving resolution problems, contact the Dionex Regional Office nearest you (see, “Dionex Regional Offices” on the Dionex Reference Library CD-ROM). 7.3.3 Loss of Front End Resolution

If poor resolutions and efficiencies are observed for the very early eluting peaks near the system void volume compared to the later eluting peaks, check the following:

A. Improper mobile phase concentration. Remake the mobile phase as required for your application. Ensure that the water and chemicals used are of the required purity.

B. Column overloading. Reduce the amount of sample components injected onto the analytical column by either diluting the sample or injecting a smaller sample volume onto the column. Ensure that late eluting components, present in high concentrations, are not overloading the column.

C. Sluggish operation of the injection valve. For pneumatically driven injector valves, check the air pressure and make sure there are no gas leaks. Check for partially plugged port faces. Refer to the injector valve manual for instructions.

D. Improperly swept volumes anywhere in the system prior to the guard and analytical columns. Swap components and refit tubing connections, one at a time, in the system prior to the analytical column and test for front-end resolution after every system change.

E. Sample-related problems. Too much organic solvent or extremes of pH in the sample matrix can be a problem. Dilute the sample or adjust the pH to be similar to the mobile phase.

F. Guard cartridge. The guard cartridge should be replaced on at regular intervals. Keep a trending log of retention times, peak efficiencies, and resolution of key analytes. Replace the guard when these parameters indicate deterioration of the columns. A compacted or contaminated guard cartridge should be replaced.

Page 38: Product Manual for Acclaim Surfactant · 2008-11-12 · PRODUCT MANUAL for the ACCLAIM SURFACTANT GUARD COLUMN (4.3 x 10mm, P/N 063215) ... Any HPLC pump capable of delivering

Acclaim Surfactant Product Manual Page 37 of 41

Acclaim® Surfactant5µm 120Å (4.6 x 150 mm)

Product No. 063201

Date: 10-Mar-05 10:47Serial No. :

Lot No. : 013-05-002

Mobile Phase: 50:50 v/v Acetonitrile:0.10 M NH4OAc, pH 5.4Flow Rate: 1.00 mL/min Temperature: 30 °CDetection: UV, 220 nm Injection Volume: 5.0 µL

Storage Solution: 50:50 v/v Acetonitrile:0.10 M NH4OAc, pH 5.4

No. Peak Name Ret.Time Asymmetry Efficiency (1) Concentration(min) (EP) (EP) (µg/mL)

1 Uracil 1.68 1.22 14254 100

2 Quat (2) 2.58 1.15 7773 100

3 p-Toluenesulfonate 3.90 1.16 13977 200

4 Propylbenzene 7.52 1.08 14237 400

(1) Efficiency (Plates/Column)(2) Quat = 2-(4-octylphenoxy)ethoxyethyldimethylbenzyl ammoniumchloride

QA Results:Analyte Parameter Specification ResultsPropylbenzene Efficiency >=12,150 PassedPropylbenzene Asymmetry 0.95-1.27 PassedPropylbenzene Retention Time 7.45-8.66 Passed

Pressure <=1100 715

ProductionReference:Datasource: QARDirectory: Acclaim\SurfactantSequence: 063201Sample No: 1 6.80 SP3 Build 2345 (128616) (Demo-Installation)

Chromeleon®Dionex Corp. 1994-2008067911-01 (QAR)

0.0 1.3 2.5 3.8 5.0 6.3 7.5 8.8 10.0-0.050

0.200

0.400

0.600 Serial No.:AU

min

1

2

3

4

Document 065087 Rev. 02 ©Dionex Corporation June 2008

Page 39: Product Manual for Acclaim Surfactant · 2008-11-12 · PRODUCT MANUAL for the ACCLAIM SURFACTANT GUARD COLUMN (4.3 x 10mm, P/N 063215) ... Any HPLC pump capable of delivering

Acclaim Surfactant Product Manual Page 38 of 41

Acclaim® Surfactant5µm 120Å (4.6 x 250 mm)

Product No. 063203

Date: 10-Mar-05 17:57Serial No. :

Lot No. : 013-05-004

Mobile Phase: 50/50 v/v Acetonitrile/0.1M NH4OAc, pH 5.4Flow Rate: 1.00 mL/min Temperature: 30 °CDetection: UV, 220 nm Injection Volume: 5.0 µL

Storage Solution: 50/50 v/v Acetonitrile/0.1M NH4OAc, pH 5.4

No. Peak Name Ret.Time Asymmetry Efficiency (1) Concentration(min) (EP) (EP) (µg/mL)

1 Uracil 2.80 1.22 25258 125

2 Quat (2) 4.47 1.11 13052 125

3 p-Toluenesulfonate 6.62 1.09 23591 250

4 Propylbenzene 12.91 1.05 24749 500

(1) Efficiency (Plates/Column)(2) Quat = 2-(4-octylphenoxy)ethoxyethyldimethylbenzyl ammoniumchloride

QA Results:Analyte Parameter Specification ResultsPropylbenzene Efficiency >=19,800 PassedPropylbenzene Asymmetry 0.95-1.27 PassedPropylbenzene Retention Time 12.42-14.29 Passed

Pressure <=1760 1110

Production Reference:Datasource: QARDirectory: Acclaim\SurfactantSequence: 063203Sample No: 1 6.80 SP3 Build 2345 (128616) (Demo-Installation)

Chromeleon®Dionex Corp. 1994-2008067913-01 (QAR)

0.0 2.0 4.0 6.0 8.0 10.0 12.0 14.0 17.0-0.050

0.100

0.200

0.300 Serial No.:AU

min

1

2

3

4

Document 065087 Rev. 02 ©Dionex Corporation June 2008

Page 40: Product Manual for Acclaim Surfactant · 2008-11-12 · PRODUCT MANUAL for the ACCLAIM SURFACTANT GUARD COLUMN (4.3 x 10mm, P/N 063215) ... Any HPLC pump capable of delivering

Acclaim Surfactant Product Manual Page 39 of 41

Acclaim® Surfactant5µm 120Å (2.1 x 150 mm)

Product No. 068123

Date: 28-Apr-08 11:09Serial No. : 000010

Lot No. : 013-05-016

Mobile Phase: 50:50 v/v Acetonitrile:0.10 M NH4OAc, pH 5.4Flow Rate: 0.21 mL/min Temperature: 30 °CDetection: UV, 220 nm Injection Volume: 1.0 µL

Storage Solution: Acetonitrile / 0.1% Acetic acid v/v 80/20

No. Peak Name Ret.Time Asymmetry Efficiency (1) Concentration(min) (EP) (EP) (µg/mL)

1 Uracil 1.77 0.76 7562 100

2 Quat (*) 3.25 1.21 6185 100

3 p-Toluenesulfonate 3.53 1.16 9843 200

4 Propylbenzene 7.86 1.10 11145 400

(1) Efficiency (Plates/Column)(2) Quat = 2-(4-octylphenoxy)ethoxyethyldimethylbenzyl ammonium chloride

QA Results:Analyte Parameter Specification ResultsPropylbenzene Efficiency >=8,100 PassedPropylbenzene Asymmetry 0.95-1.32 PassedPropylbenzene Retention Time 7.44-8.87 Passed

Pressure <=1100 613

Production Reference:Datasource: QARDirectory: Acclaim\SurfactantSequence: 068123Sample No.: 1 6.80 SP3 Build 2345 (128616) (Demo-Installation)

Chromeleon® Dionex Corp. 1994-2008

0.0 1.3 2.5 3.8 5.0 6.3 7.5 8.8 10.0-50

200

500 Serial No.: 10mAU

min

1

2

3

4

WVL:220 nm

Document 065087 Rev. 02 ©Dionex Corporation June 2008

Page 41: Product Manual for Acclaim Surfactant · 2008-11-12 · PRODUCT MANUAL for the ACCLAIM SURFACTANT GUARD COLUMN (4.3 x 10mm, P/N 063215) ... Any HPLC pump capable of delivering

Acclaim Surfactant Product Manual Page 40 of 41

Document 065087 Rev. 02 ©Dionex Corporation June 2008

ACCLAIM® SURFACTANT

LOT VALIDATION –5µM 120Å

Lot #: 013-05-002 Date: 28-Mar-2005 Bonded Silica Tests

0.0 2.0 4.0 6.0 8.0 10.0-0.050

0.700 2005-03-10-063201-val idation #2 #1 SFC_P01AU

min

1

2

3

4

0.0 10.0 20.0 30.0-100

500

900 LOT VALIDATION #6 #1, SFC_P01 (isocraticV

min

1 2

3 4

5

Isocratic Selectivity Gradient Surfactant Resolution

Silica Qualification Test Parameter Specification Results

Raw Particle size (µm) 4.2-4.5 4.3 Pore Volume (mL/g) 0.9-1.0 0.98 Pore diameter (Å) 120-140 127 Particle distribution(40/90) 1.2-1.3 1.24 Surface area (m2/g) 290-320 PASSED Metal (ppm) Na, Fe, Al <3 PASSED Bonded Total carbon (%) 11.8 – 13.2 12.8

Page 42: Product Manual for Acclaim Surfactant · 2008-11-12 · PRODUCT MANUAL for the ACCLAIM SURFACTANT GUARD COLUMN (4.3 x 10mm, P/N 063215) ... Any HPLC pump capable of delivering

Acclaim Surfactant Product Manual Page 41 of 41

Document 065087 Rev. 02 ©Dionex Corporation June 2008

Test Conditions and Results (Tested using a 4.6 x 150 mm Analytical column)

Isocratic Selectivity This parameter is used to monitor batch-to-batch reproducibility as measured by the relative retention of an anionic surfactant to a neutral marker. Mobile Phase: 50:50 v/v Acetonitrile: 0.10 M Ammonium acetate, pH 5.4 Flow Rate: 1.0 mL/min Temperature: 30 oC Injection Volume: 5 µL Detection: UV 220 nm, 6mm cell Peaks: 1) Uracil, 0.10 mg/mL; 2) Octylphenoxyethoxyethyldimethylbenzyl ammonium chloride, 0.10 mg/mL; 3) Sodium p-Toluenesulfonate, 0.20 mg/mL; 4) Propylbenzene, 0.40 mg/mL

Specifications:

Parameter Analyte Requirement Result

Retention time (4) 7.45—8.66 7.52 Asym. (4) 0.95—1.27 1.08 Efficiency (4) ≥ 12,150 N/column 14237 Retention time (3) 3.57—4.34 3.90 Asym. (3) < 1.80 max 1.16 Retention ratio (3)/(4) 0.47—0.61 0.52 Retention time (2) 2.45—2.95 2.58 Asym. (2) < 1.80 max 1.15

Gradient Surfactant Resolution An application-based test for the resolution of anionic, nonionic and cationic surfactants. Mobile Phase A: Acetonitrile Mobile Phase B: 0.10 M Ammonium acetate, pH 5.4 Time %A %B 0.0 25 75 25.0 85 15 30.0 85 15 Flow Rate: 1.0 mL/min Temperature: 30 oC Injection Volume: 20 µL Detection: Evaporative Light Scattering Analytes: 1) Laurylpyridinium chloride, 0.10 mg/mL; 2) Lauryldimethylbenzyl ammonium chloride, 0.10 mg/ mL ; 3) Triton X-100, 0.40 mg/mL; 4) Sodium decylsulfate, 0.20 mg/mL; 5) Sodium dodecylsulfate, 0.10 mg/mL

Specifications:

Parameter Analyte Requirement Result

Retention ratio (2)/(3) 0.66—0.86 0.76 Retention ratio (3)/(5) 0.45—0.65 0.51