Process Gas Chromatography Siemens NeSSI Update Slide 1; 03/08 Advances in NeSSI TM communications An Implementation of the Generation 2 Bus Siemens Activities Silke Ebbing CPAC Rome, March 2008

Upload: others

Post on 22-Jan-2020

7 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Process Gas Chromatography - University of Washingtondepts.washington.edu/cpac/Activities/Meetings/Satellite/2008/Friday/... · Process Gas Chromatography Traditional Sample Conditioning

Process Gas Chromatography

Siemens NeSSI Update Slide 1; 03/08

Advances in NeSSITM communications

An Implementation of the Generation 2 BusSiemens Activities

Silke Ebbing

CPAC Rome, March 2008

Page 2: Process Gas Chromatography - University of Washingtondepts.washington.edu/cpac/Activities/Meetings/Satellite/2008/Friday/... · Process Gas Chromatography Traditional Sample Conditioning

Siemens NeSSI Update.ppt Slide 2; SE 03/08

Process Gas Chromatography

Continuous Gas Analytics Laser Spectrometry

System Integration

Portfolio Process Analytics

Gas Chromatography

Page 3: Process Gas Chromatography - University of Washingtondepts.washington.edu/cpac/Activities/Meetings/Satellite/2008/Friday/... · Process Gas Chromatography Traditional Sample Conditioning

Siemens NeSSI Update.ppt Slide 3; SE 03/08

Process Gas Chromatography

How to get a valid measurement???

- Reliable, robust analyzer

- Regular validation & calibration

- Representative sample

Page 4: Process Gas Chromatography - University of Washingtondepts.washington.edu/cpac/Activities/Meetings/Satellite/2008/Friday/... · Process Gas Chromatography Traditional Sample Conditioning

Siemens NeSSI Update.ppt Slide 4; SE 03/08

Process Gas Chromatography

Topics

Sample Handling Systems for Gas Chromatographs► Status quo

► NeSSI

Siemens Implementation of Generation 2► A major project

► The Maxum connection

► Cooperation with NeSSI component suppliers

► Summary of key design requirements

► Status

Page 5: Process Gas Chromatography - University of Washingtondepts.washington.edu/cpac/Activities/Meetings/Satellite/2008/Friday/... · Process Gas Chromatography Traditional Sample Conditioning

Siemens NeSSI Update.ppt Slide 5; SE 03/08

Process Gas Chromatography

Traditional Sample Conditioning SystemsCustom designed and built

Discrete components with tubing interconnection

Layout in cabinets or on panels

Electrical components are minimized; pneumatic devices sometimes used► Valves (stream or bottle selection, atmospheric reference, other)► Pressure controllers or sensors► Flow controllers or sensors► Temperature controllers and sensors

Interconnection to analyzers

Page 6: Process Gas Chromatography - University of Washingtondepts.washington.edu/cpac/Activities/Meetings/Satellite/2008/Friday/... · Process Gas Chromatography Traditional Sample Conditioning

Siemens NeSSI Update.ppt Slide 6; SE 03/08

Process Gas Chromatography

Sample system with mechanical components

inside cabinet

Maxum GC

Typical, Conventional “Not Smart” System

High Voltage electrical power in Explosion Proof housings for heat. No automated control or sensing.

Mechanical regulators, flow indicators and gauges with no electrical interface.

Note:

Some sampling systems mounted separately from GC.

Some systems mounted attached to GC.

Air tubing bundle to operate valves; electrical pilot valves are mounted inside GC

Page 7: Process Gas Chromatography - University of Washingtondepts.washington.edu/cpac/Activities/Meetings/Satellite/2008/Friday/... · Process Gas Chromatography Traditional Sample Conditioning

Siemens NeSSI Update.ppt Slide 7; SE 03/08

Process Gas Chromatography

Sample system with mechanical components

inside cabinet

Maxum GC

Conventional “Smart” System – Very Rarely Provided

Air tubing bundle to operate valves; electrical pilot valves are mounted inside GC

All mechanical indicators, sensors and controllers must be replaced with electronics. All electronics must conform to Explosion Proof or non-incendive design.(NOT SHOWN in this photograph!)

Wiring bundle between GC and Sampling System;Discrete I/O electronics inside GC;Wiring conforms to Explosion Proof, non-incendive methods.

NOTE:This is rarely done because of the high expense and

awkward construction techniques required.

Page 8: Process Gas Chromatography - University of Washingtondepts.washington.edu/cpac/Activities/Meetings/Satellite/2008/Friday/... · Process Gas Chromatography Traditional Sample Conditioning

Siemens NeSSI Update.ppt Slide 8; SE 03/08

Process Gas Chromatography

NeSSI Intended Benefits

Reduced size and cost of sampling systems► Standard, modular components

Improved on-line “system availability”► Fewer tubing fittings, couplings► Opportunity for electronic maintenance

monitoring

Improved on-line performance► Smaller components► Precise electronic control

Page 9: Process Gas Chromatography - University of Washingtondepts.washington.edu/cpac/Activities/Meetings/Satellite/2008/Friday/... · Process Gas Chromatography Traditional Sample Conditioning

Siemens NeSSI Update.ppt Slide 9; SE 03/08

Process Gas Chromatography

Status of NeSSIGeneration 1A mechanical specification is published (SP76.02)► Several companies now

produce NeSSI standard mechanical components

Generation 2An electrical interface concept is published► No one has previously

developed NeSSI interface electronics

Generation 3Concepts for integrated sensors and analyzers► Various models and

prototypes have been shown

Page 10: Process Gas Chromatography - University of Washingtondepts.washington.edu/cpac/Activities/Meetings/Satellite/2008/Friday/... · Process Gas Chromatography Traditional Sample Conditioning

Siemens NeSSI Update.ppt Slide 10; SE 03/08

Process Gas Chromatography

Technical OverviewNeSSI Modular Mechanical Concept

Small, modular components

Standardized “substrate” (or “platform”) that the components mount to► Substrate can be made in a variety of ways► Components can be made in a variety of ways► The interface between the substrate and the components is always the same

Conventional Sampling System

Large mechanical components.Tube and fitting construction.

Modular Sample System(Mechanical Concept)

Small modular components.Standard substrate interconnections.

(Actual systems typically in a cabinet.)

Illustration courtesy of CIRCOR Tech.

Page 11: Process Gas Chromatography - University of Washingtondepts.washington.edu/cpac/Activities/Meetings/Satellite/2008/Friday/... · Process Gas Chromatography Traditional Sample Conditioning

Siemens NeSSI Update.ppt Slide 11; SE 03/08

Process Gas Chromatography

Technical OverviewExample NeSSI Systems

Modular interconnection substrates.

Note:

In actual systems, the

components and substrates are

mounted inside a cabinet.

Modular components

Page 12: Process Gas Chromatography - University of Washingtondepts.washington.edu/cpac/Activities/Meetings/Satellite/2008/Friday/... · Process Gas Chromatography Traditional Sample Conditioning

Siemens NeSSI Update.ppt Slide 12; SE 03/08

Process Gas Chromatography

Pro-E rendering of Swagelok system; rendering by Siemens – component representation approximate.

Technical OverviewNeSSI Illustration; Mechanical; exploded view

Substrate blockFootprint: 3.8cm x 3.8cm

Air-operated Valve

Mechanical Flow Meter

Mechanical Guage

Page 13: Process Gas Chromatography - University of Washingtondepts.washington.edu/cpac/Activities/Meetings/Satellite/2008/Friday/... · Process Gas Chromatography Traditional Sample Conditioning

Siemens NeSSI Update.ppt Slide 13; SE 03/08

Process Gas Chromatography

Major project enforced Siemens Activities for Gen 2

Project► Large petrochemical manufacturer; major plant expansion► Several hundred analyzers with sample conditioning systems► User desires “smart systems” implemented with NeSSI

Generation 2 technology to maximize on-going system performance

► Project issues:• Risk mitigation; project requires proven techniques to

minimize risk

Problem► No NeSSI bus standard yet existed► Significant technical issues to be resolved with available

techniques

Page 14: Process Gas Chromatography - University of Washingtondepts.washington.edu/cpac/Activities/Meetings/Satellite/2008/Friday/... · Process Gas Chromatography Traditional Sample Conditioning

Siemens NeSSI Update.ppt Slide 14; SE 03/08

Process Gas Chromatography

The Siemens Solution

Implement bus using available technology► Buffers and barriers for rugged I.S. environment

Cooperate with major substrate and component suppliers to provide compatible components► Swagelok

► CIRCOR Tech

► Parker Hannifin

Build demonstration systems for evaluation

Complete commercial engineering and certification following go-ahead approval

Page 15: Process Gas Chromatography - University of Washingtondepts.washington.edu/cpac/Activities/Meetings/Satellite/2008/Friday/... · Process Gas Chromatography Traditional Sample Conditioning

Siemens NeSSI Update.ppt Slide 15; SE 03/08

Process Gas Chromatography

Topics for bus developmentImplementation with an existing bus used in Siemens analyzers► Existing hardware, software and bus protocol for proven reliability in

similar applications► Enhanced with standard circuitry for robustness in Intrinsic Safety► Tested for conformance with CE and IEC specification for immunity and

avoidance of EMI and RFI

Siemens provides:► Bus technology including circuit design and Intrinsic Safety design

parameters► Intrinsic Safety implementation technology plus design assistance and

certification support► Project management and point of overall responsibility for

implementation► Specialized Human Interface to facilitate field maintenance

Component vendors provide:► Bus compatible components including component-specific electronics► Integration of component-specific electronics and Siemens bus

interface electronics► Point of responsibility for component I.S. certification

Page 16: Process Gas Chromatography - University of Washingtondepts.washington.edu/cpac/Activities/Meetings/Satellite/2008/Friday/... · Process Gas Chromatography Traditional Sample Conditioning

Siemens NeSSI Update.ppt Slide 16; SE 03/08

Process Gas Chromatography

Electrical Concept Implementation“Smart” Analyzers

Sampling System

Chromatograph

GC Control Electronics

Ethernet Communications Network

Note:

Sampling system components are never directly accessed by System Bus above. This data is always provided via the analyzer / SAM which controls the sampling system.

SAM Embedded in GCSpecialized

HMI Software

Intrinsically Safe NeSSI Gen. 2 Bus

Page 17: Process Gas Chromatography - University of Washingtondepts.washington.edu/cpac/Activities/Meetings/Satellite/2008/Friday/... · Process Gas Chromatography Traditional Sample Conditioning

Siemens NeSSI Update.ppt Slide 17; SE 03/08

Process Gas Chromatography

V TPA

Continuous Gas Analyzers with system control and network integration provided by Siemens Network Access Unit (NAU) with integrated NeSSI Generation 2 Bus Control capability.

Sampling systems built with substrates and smart, Generation 2- compatible components.

Intrinsically safe, Generation 2 Bus

Electrical Concept Implementation“Dumb” analyzers

SAM functionality is external to the analyzer

Ethernet Communications Network

Page 18: Process Gas Chromatography - University of Washingtondepts.washington.edu/cpac/Activities/Meetings/Satellite/2008/Friday/... · Process Gas Chromatography Traditional Sample Conditioning

Siemens NeSSI Update.ppt Slide 18; SE 03/08

Process Gas Chromatography

System Concept ImplementationEthernet Network

Sample to Analyzers

Intrinsically Safe NeSSI

Bus

Intrinsically Safe NeSSI

Bus

Specialized HMI

Software

SAM Embedded in Controllers

Page 19: Process Gas Chromatography - University of Washingtondepts.washington.edu/cpac/Activities/Meetings/Satellite/2008/Friday/... · Process Gas Chromatography Traditional Sample Conditioning

Siemens NeSSI Update.ppt Slide 19; SE 03/08

Process Gas Chromatography

Illustration courtesy of CIRCOR Tech.

Technical OverviewNeSSI Electrical Concept

Input:Manifold air for valves

Input:Electrical power for high wattage heaters

Intrinsically Safe Bus; provides;- Signal for sensors and valve control- Power for sensors and valve electrical pilots

Actual components mounted inside a cabinet creating a

Division 1 or Zone 1 environment.

Page 20: Process Gas Chromatography - University of Washingtondepts.washington.edu/cpac/Activities/Meetings/Satellite/2008/Friday/... · Process Gas Chromatography Traditional Sample Conditioning

Siemens NeSSI Update.ppt Slide 20; SE 03/08

Process Gas Chromatography

Bus Electrical ImplementationUsing Existing I/O Link Structure

Detector Signal and Control

Multiple channels of Oven Temperature Control

Multiple channels of Electronic Carrier Gas

Pressure Control

Multiple Solenoid Valve Controls

Maxum currently uses an internal serial I/O linking system to manage devices required for the Chromatograph.

Maxum Electronic Section

Page 21: Process Gas Chromatography - University of Washingtondepts.washington.edu/cpac/Activities/Meetings/Satellite/2008/Friday/... · Process Gas Chromatography Traditional Sample Conditioning

Siemens NeSSI Update.ppt Slide 21; SE 03/08

Process Gas Chromatography

Maxum GC Electronics

Bus Electrical Implementation

Existing I/O link connected to new module.

• Pepperl-Fuchs Intrinsically Safe Power supply: 9.5v at 1 Amp.

• Powers I/O link and components.• Use additional power supplies as needed

for large numbers of components.

NeSSI Generation 2 Bus;Intrinsically safe;

carries signal and power for devices.

• Existing I/O link used for control of GC Inputs and Outputs.

• Existing protocol supports several hundred I/O channels.

New module with power drivers and Intrinsic Safety Barriers.

V TPA

Page 22: Process Gas Chromatography - University of Washingtondepts.washington.edu/cpac/Activities/Meetings/Satellite/2008/Friday/... · Process Gas Chromatography Traditional Sample Conditioning

Siemens NeSSI Update.ppt Slide 22; SE 03/08

Process Gas Chromatography

Component Electrical Implementation

NeSSI Substrates by component suppliers

Maxum GC

Proprietary mechanical NeSSI component by component

vendors

Link driver and I.S. barriers by Siemens

Software and I/O link master protocol

by Siemens

Proprietary component-specific electronics by component

vendor

Chip with I/O link slave software;Intrinsic Safety barriers; and

Standardized interface to component electronics by

Siemens

Certification of the complete component by component

vendor.

Page 23: Process Gas Chromatography - University of Washingtondepts.washington.edu/cpac/Activities/Meetings/Satellite/2008/Friday/... · Process Gas Chromatography Traditional Sample Conditioning

Siemens NeSSI Update.ppt Slide 23; SE 03/08

Process Gas Chromatography

Key Physical and Functional Requirements

SamplingSystem

Analyzer

SamplingSystem

Analyzer

SamplingSystem

Analyzer House (typical)

Analyzer

Bus to support very large number of participating

component suppliers (unlimited ~ >1000’s)

Number of bus “instances” to be limited to simplify analyzer

control and reduce cost; <= 2 instances preferred, excluding

power sources.

Analyzer may be up to 10m from sampling system. Total cable length <=20m

Page 24: Process Gas Chromatography - University of Washingtondepts.washington.edu/cpac/Activities/Meetings/Satellite/2008/Friday/... · Process Gas Chromatography Traditional Sample Conditioning

Siemens NeSSI Update.ppt Slide 24; SE 03/08

Process Gas Chromatography

Key Physical and Functional Requirements

Bus to provide for sensing and control of: valves, pressure, flow, temperature. Component ambient

temperature to be >=70°C

Design goal for flammability ignition temperature to be T4 (135°C) with minimum at least T3 (200°C).

Bus to be I.S.; components to mount in Div. 1 / Zone 1 hazardous area.

Mounting area may be hazardous or

not.

Interior of cabinet is considered hazardous due to leakage of flammable gases in a closed area.

Capability for the standard functions to be implemented in a variety of ways.

Must support power loading and control of power permitted by I.S. bus: Start-up, minimum, maximum, abnormal, other power-management issues

Any single physical device should be able to represent itself as using multiple device types.

Page 25: Process Gas Chromatography - University of Washingtondepts.washington.edu/cpac/Activities/Meetings/Satellite/2008/Friday/... · Process Gas Chromatography Traditional Sample Conditioning

Siemens NeSSI Update.ppt Slide 25; SE 03/08

Process Gas Chromatography

Key Physical and Functional RequirementsSampling system to be capable of supporting at least 30 separate streams

Bus to support replacement of components while live without interruption of bus operation

Bus to provide for control of high-power components that do not draw power from the bus itself (eg. heaters)

Systems are not ad hoc. Device function does not change while live. Device recognition during application is useful.

Interchangeability requires that the system be able to confirm that a replacement device is of comparable type to device being replaced

Page 26: Process Gas Chromatography - University of Washingtondepts.washington.edu/cpac/Activities/Meetings/Satellite/2008/Friday/... · Process Gas Chromatography Traditional Sample Conditioning

Siemens NeSSI Update.ppt Slide 26; SE 03/08

Process Gas Chromatography

Pro-E rendering of Swagelok system; rendering by Siemens – component representation approximate.

Key Physical and Functional RequirementsCabling must connect in

available “footprint” of 3.8cm x 3.8cm

Number of components on bus:Typical: 5-15

Common: 20-30Maxum: 200

(Note: not all are shown connected in this illustration.)

Bus must provide signal AND power.

Components not large enough to

permit 2 cable connections.

Power Requirements:Sensors: 10-40ma

Controllers: 20-70maValve pilots: 80-350ma

Distance to analyzer <=20m

Components must be scanned every 10 to

30 seconds.

Valves must operate within 0.5 sec.

At least 4 valves may be active

simultaneously

Bus must meet Zone 1 and EMI/RFI

requirements.

Page 27: Process Gas Chromatography - University of Washingtondepts.washington.edu/cpac/Activities/Meetings/Satellite/2008/Friday/... · Process Gas Chromatography Traditional Sample Conditioning

Siemens NeSSI Update.ppt Slide 27; SE 03/08

Process Gas Chromatography

Status of Siemens ImplementationAnalyzer and software► Communications Software / Protocol: existing proprietary, complete

External bus► Buffer / drivers and I.S. barriers: testing complete; production design in

progress► I.S. certification: preliminary review complete► EMI / RFI testing: bus validation tests complete► Robustness, bit error rate testing: bus validation tests complete

Components► Bus interface chip and circuitry: Version 1 operating in test► Component interface and software: in test► I.S. certification: pending► Component functions: in test

Compatibility with NeSSI Vision► Direct implementation of NeSSI concept; meets all key requirements► Architecture consistent with other Bus technologies currently being

considered (CAN, Fieldbus, Profibus)

Page 28: Process Gas Chromatography - University of Washingtondepts.washington.edu/cpac/Activities/Meetings/Satellite/2008/Friday/... · Process Gas Chromatography Traditional Sample Conditioning

Siemens NeSSI Update.ppt Slide 28; SE 03/08

Process Gas Chromatography

Siemens Process Analytics

Questions?

Page 29: Process Gas Chromatography - University of Washingtondepts.washington.edu/cpac/Activities/Meetings/Satellite/2008/Friday/... · Process Gas Chromatography Traditional Sample Conditioning

Siemens NeSSI Update.ppt Slide 29; SE 03/08

Process Gas Chromatography

Backup

Page 30: Process Gas Chromatography - University of Washingtondepts.washington.edu/cpac/Activities/Meetings/Satellite/2008/Friday/... · Process Gas Chromatography Traditional Sample Conditioning

Siemens NeSSI Update.ppt Slide 30; SE 03/08

Process Gas Chromatography

What Is NeSSI?New Sampling System Initiative

User group – committee► Hosted by CPAC

Center for Process Analytical Chemistry;University of Washington; Seattle

► End users and vendors► Meeting for about 8 year

Rationale of NeSSI► Process analyzers require sample conditioning systems► Sampling systems account for a high portion of total system

maintenance► There has been no development or significant improvement in process

sampling systems in many years► A group of end users began meeting 8 years ago to address this topic

Goals► Publish related specifications► Encourage and enable standardization and enhanced designs

Page 31: Process Gas Chromatography - University of Washingtondepts.washington.edu/cpac/Activities/Meetings/Satellite/2008/Friday/... · Process Gas Chromatography Traditional Sample Conditioning

Siemens NeSSI Update.ppt Slide 31; SE 03/08

Process Gas Chromatography

NeSSI Committee(New Sampling System Initiative)

Major customer participants► Dow► ExxonMobil► ConocoPhillips► Shell► BP► Others

Major analyzer supplier participants► Siemens► ABB► Emerson-Rosemount

Major component supplier participants► Swagelok► CIRCOR Tech► Parker / Hannifin

Page 32: Process Gas Chromatography - University of Washingtondepts.washington.edu/cpac/Activities/Meetings/Satellite/2008/Friday/... · Process Gas Chromatography Traditional Sample Conditioning

Siemens NeSSI Update.ppt Slide 32; SE 03/08

Process Gas Chromatography

The Traditional Analyzer System

HMI

DumbAnalyzer

Air

Power SmartAnalyzer

HMIAir

Power

System Communication Network

ProcessProcess

Discrete Electrical and / or pneumatic connections.

Sampling System Controller

Page 33: Process Gas Chromatography - University of Washingtondepts.washington.edu/cpac/Activities/Meetings/Satellite/2008/Friday/... · Process Gas Chromatography Traditional Sample Conditioning

Siemens NeSSI Update.ppt Slide 33; SE 03/08

Process Gas Chromatography

The NeSSI VisionNeSSI Electrical Concept

SAM HMI

System Communication Network

SmartAnalyzer

HMIAir

PowerDumb

Analyzer

Air

Power

ProcessProcess

Serial electrical bus.

Page 34: Process Gas Chromatography - University of Washingtondepts.washington.edu/cpac/Activities/Meetings/Satellite/2008/Friday/... · Process Gas Chromatography Traditional Sample Conditioning

Siemens NeSSI Update.ppt Slide 34; SE 03/08

Process Gas Chromatography

Key Bus Design RequirementsThe Old Way

Stream Select

Bypass Flow

Analyzer Flow

Flow / Pressure Adjust

Filters

Page 35: Process Gas Chromatography - University of Washingtondepts.washington.edu/cpac/Activities/Meetings/Satellite/2008/Friday/... · Process Gas Chromatography Traditional Sample Conditioning

Siemens NeSSI Update.ppt Slide 35; SE 03/08

Process Gas Chromatography

Representative Test System DesignAn Introduction to NeSSI Bus Design Requirements

Process Stream Inlet

Two Standard Bottle Inlets

Analyzer Valve Loop

Sample Shut Off and Atmospheric Reference

Speed Loop Return

Active Bus Components:• 3 selector valves• Sample Shut off valve• Atmospheric reference valve• 2 pressure transmitters• 2 flow transmitters (optioned)

Active Bus

Page 36: Process Gas Chromatography - University of Washingtondepts.washington.edu/cpac/Activities/Meetings/Satellite/2008/Friday/... · Process Gas Chromatography Traditional Sample Conditioning

Siemens NeSSI Update.ppt Slide 36; SE 03/08

Process Gas Chromatography

FISCO Power Delivery

Zündgrenzkurven nach W39

0

50

100

150

200

250

300

350

400

450

500

10 12 14 16 18 20 22 24 26 28 30

U0 (V)

I K (m

A)

FISCO=-Bereich

IIB

IIC

Power Delivery Curves from W39

Page 37: Process Gas Chromatography - University of Washingtondepts.washington.edu/cpac/Activities/Meetings/Satellite/2008/Friday/... · Process Gas Chromatography Traditional Sample Conditioning

Siemens NeSSI Update.ppt Slide 37; SE 03/08

Process Gas Chromatography

Cabling Issues

Cable size, bend radius, height

Connection methods and system design

Jacket materials

Cost

Page 38: Process Gas Chromatography - University of Washingtondepts.washington.edu/cpac/Activities/Meetings/Satellite/2008/Friday/... · Process Gas Chromatography Traditional Sample Conditioning

Siemens NeSSI Update.ppt Slide 38; SE 03/08

Process Gas Chromatography

Representative 4-wire Bus Design

AxDriver Barrier

IS PS Components

Signal

Power

Page 39: Process Gas Chromatography - University of Washingtondepts.washington.edu/cpac/Activities/Meetings/Satellite/2008/Friday/... · Process Gas Chromatography Traditional Sample Conditioning

Siemens NeSSI Update.ppt Slide 39; SE 03/08

Process Gas Chromatography

Design Topics and Issues

Bus Software

Component Physical Issues

EMI / RFI and Robust Design

I.S. Design Issues

Cabling

Page 40: Process Gas Chromatography - University of Washingtondepts.washington.edu/cpac/Activities/Meetings/Satellite/2008/Friday/... · Process Gas Chromatography Traditional Sample Conditioning

Siemens NeSSI Update.ppt Slide 40; SE 03/08

Process Gas Chromatography

Technical OverviewMaxum GC Electronics Interior

Location of electrical boards required to provide discrete analog and digital I/O to interface sensors.

Pilot valves to control air-operated valves in GC and Sampling System.

(Locations for additional pilot valves if required.)

To sampling system

Page 41: Process Gas Chromatography - University of Washingtondepts.washington.edu/cpac/Activities/Meetings/Satellite/2008/Friday/... · Process Gas Chromatography Traditional Sample Conditioning

Siemens NeSSI Update.ppt Slide 41; SE 03/08

Process Gas Chromatography

Maxum Detail

Bus Driver Module

ProE Illustration of Module on location in

Maxum GC.

Page 42: Process Gas Chromatography - University of Washingtondepts.washington.edu/cpac/Activities/Meetings/Satellite/2008/Friday/... · Process Gas Chromatography Traditional Sample Conditioning

Siemens NeSSI Update.ppt Slide 42; SE 03/08

Process Gas Chromatography

Maxum and NAU Detail

P&F I.S. Power Supply Bus Interface Module (prototype) in Maxum Network Access Unit

Example component electronics - bus interface