Pressure Intensity

• The pressure intensity at a point can be calculated using the solution of wave equation:– 1D case– Circular aperture

• The pressure distribution is proportional with the fourier transform of the aperture.

Page 3: Pressure Intensity

Side Lobes

Side lobes are small beams of greatly reduced intensity that are emitted at angles tothe primary beam and they often cause image artifacts.

• the origin of these lobes are due from radial vibrations from the edges of the transducer

Page 4: Pressure Intensity
Page 5: Pressure Intensity

Focused Transducers

High frequency beams have two advantages over low-frequency beams: (1) axial resolution is superior; and (2) the Fresnel zone is longer

It would seem logical to use high frequencies for all imaging. High frequencies however, have a major drawback related to penetration. Tissue absorption increaseswith increasing frequency, so a relatively low frequency beam is required to penetrate thick parts.

It would then seem logical to use low frequency transducers and to increase the size of the transducer to keep the beam coherent for sufficient depth to reach the point of interest (longer Fresnel zone). Although larger transducers improve coherence they deteriorate lateral resolution. The dilemma is at least partially resolved with the use focused transducers.

NOTE• focused transducers reduce beam width which improves lateral resolution• they also concentrate beam intensity thereby increasing penetration and echo intensity thus improving image quality

Page 6: Pressure Intensity

• the focal zone is the region over which the beam is focused• the focal length is the distance from the transducer to the centre of the focal zone• the depth of focus is the distance over which the beam is in a reasonable focus• a small diameter transducer has a shorter focal zone and spreads more rapidly in the far zone• most diagnostic transducers are focused, which is achieved using a either a curved piezoelectric crystal, an acoustic lens or electronics (phased arrays)

Page 7: Pressure Intensity

Ultrasound Pulse Production and Reception

A transducer is a device that can convert one form of energy into another. Ultrasoundtransducers are used to convert an electrical signal into ultrasonic energy that can betransmitted into tissue, and to convert ultrasonic energy reflected back from the tissueinto an electrical signal.

The general composition of an ultrasound transducer is shown below:• the most important component is a thin piezoelectric (crystal) element located near the face of the transducer• the front and back face of the element is coated with a thin conducting film to ensure good contact with the two electrodes• the outside electrode is grounded to protect the patient from electrical shock• an insulated cover is used to make the device watertight• an acoustic insulator made of cork or rubber is used to prevent the passing of sound into the housing (i.e.: reduces transducer vibrations)

• the inside electrode is against a thick backing block that absorbs sound waves transmitted back into the transducer

Page 8: Pressure Intensity

Transducers, (apertures)

• Bulk Acoustic wave transducers• Surface acoustic wave transducers

quartz crystal

microbalance

Page 9: Pressure Intensity

• In the bulk of an ideally infinite unbounded solid, two types of bulk acoustic waves (BAW) can propagate. They are the longitudinal waves, also called compressional/extensional waves, and the transverse waves, also called shear waves, which respectively identify vibrations where particle motion is parallel and perpendicular to the direction of wave propagation.

• When a single plane boundary interface is present forming a semi- infinite solid, surface acoustic waves (SAW) can propagate along the boundary.

Page 10: Pressure Intensity

• Probably the most common type of SAWs are the Rayleigh waves, which are actually two-dimensional waves given by the combina- tion of longitudinal and transverse waves and are confined at the surface down to a penetration depth of the order of the wavelength.

• Shear horizontal (SH) particle displacement has only a very low pene- tration depth into a liquid, hence a device with pure or pre- dominant SH modes can operate in liquids without significant radiation losses in the device.

• Love waves (LW), where the acoustic wave is guided in a foreign layer

• Plate waves, also called Lamb waves, require two parallel boundary planes.

• flexural plate wave (FPW)

Page 11: Pressure Intensity

• Bulk Wave:– Conventional piezo– CMUT– Thin film

• SAW– Microfabrication– CMUT

Page 12: Pressure Intensity

Frequency- Material width Relation

• df/dl=?

Page 13: Pressure Intensity
Page 14: Pressure Intensity

Why should the transducer thickness be equal to 1/2 of the desired wavelength?

When the piezoelectric element is driven by a alternating voltage the crystalvibrates (i.e.: the width of the crystal moves back and forth). The front face of the crystal emits sound both in the forward and backward directions as does the back surface.

Front surfaceBack surface

Thickness (t)

A B C D

• wave front (A) will get absorbed by the transducer’s backing material• wave front (D) will enter into the patient • the wave front (C) is reflected at the back face of the disk, and by the time it joins wave front (D), it has traveled an extra distance 2t. If this distance equals a wavelength the wave fronts (D) and (C) reinforce for they are in phase, and constructive interference or resonance occurs.• if wave fronts (D) and (C) are not in phase, then there will be some destructive interference• same reasoning applies to wave front (B)

PatientBacking Block

Page 15: Pressure Intensity

Creating a sound wave from an electrical pulse

When a positive voltage (A) is applied across the surface of the crystal, it creates an electric field across the crystal surface which cause the molecules (dipoles) in the crystal to realign and thus changing the shape (width) of the crystal.

When the voltage polarity is changed from positive to negative, there is a point in timewhen the electric field across the crystal is zero (at voltage equal to zero) and the crystalrelaxes (B). When the voltage polarity is reversed (i.e.: negative) the crystal realigns once again and changes its width once again (C).

A B C

Positive

Negative

Voltage Pulse

Time

Page 16: Pressure Intensity

The net effect the alternating voltage pulse has on the crystal is to make it oscillate backand forth about its width. This change in shape of the crystal increases and decreasesthe pressure in front of the transducer, thus producing ultrasound waves.

Ultrasound wave direction

Ultrasound wave directionCompression region created when crystalsurface is expanding (more pressure on surface)

Rarefaction region created when crystalsurface is contracting (less pressure on surface)

wavefront diagram

Page 17: Pressure Intensity

Creating sound wave using CMUT

Page 18: Pressure Intensity

CMUT vs Piezoelectric

• The frequency of operation depends on the cell size (cavity of membrane), and on the stiffness of the material used as a membrane.

• As it is built on silicon, the integration of electronics would be easier for the CMUTs compared to other transducer technologies.

• Large Bandwith/High frequency

• Smaller dimensions

Page 19: Pressure Intensity

Resonant Frequency

The frequency at which the transducer is the most efficient as a transmitter of soundis also the frequency at which it is most sensitive as a receiver of sound. Thisfrequency is called the natural or resonant frequency of the transducer.

• the thickness and the material (i.e.: speed of sound in the crystal) of the piezoelectric crystal determines the resonant frequency of the transducer• transducers crystals are normally manufactured so that their thickness (t) is equal to one-half of the wavelength () of the ultrasound produced by the transducer

Bandwidth

The range of frequencies in the emitted ultrasound wave is called the bandwidth and is defined to be the full width of the frequency distribution at half maximum (FWHM).

Resonant Frequency

bandwidth SPL Spatial Pulse Length (SPL)

Page 20: Pressure Intensity

Continuous voltage waveform Pulsed voltage waveform

Continuous waveformcan be represented bya single sine wave (onefrequency), thus frequencydistribution is very narrow

Pulsed waveformcan be represented bythe sum of many sinewaves each of differentfrequency, thus frequencydistribution is wide

Frequency distribution of emitted ultrasound wave

Page 21: Pressure Intensity

Q-factor

The Q-factor of a transducer system describes the shape of the frequency distribution(response curve) and is defined as

where f0 is the resonance frequency, f1 is the frequency below resonance at whichintensity is reduced by half and f2 is the frequency above resonance at whichintensity is reduced by half

Q-factor = f0

(f2 - f1)Bandwidth = (f2 - f1)

• high Q transducers produce relatively pure frequency spectrums and low Q transducers produce a wider range of frequencies• short pulses correspond to reduce Q values and vice versa

bandwidth Q-factor

Page 22: Pressure Intensity

Pulse Ultrasound Mode

Because a transducer can be a transmitter and a receiver of ultrasonic energy, it clearly stands to reason that a continuous voltage waveform can not be used. If such a waveform was used, the transducer would always function as a transmitter. Since the internally generated sound waves are stronger than the returning echoes, the returning signal is lost in the noise of the system. To over come this problem, most transducers are used in a pulse mode where the voltage waveform consists of many pulses each separated by a fixed distance and time. The transducer functions as a transmitter during pulse excitation and as a receiver during the time interval between pulses.

Voltage waveform Ultrasound pulses produced by transducerNOTE• most transducers are designed to have short pulses (improved resolution) with low Q values (broad bandwidth - desirable in order to receive echoes of many different frequencies)

Page 23: Pressure Intensity

Pulse Repetition Frequency (PRF)

Pulse Repetition Period (PRP)

• PRF is the number of pulses occurring in 1 second

• PRP is the time from the beginning of one pulse to the beginning of the next pulse

Page 24: Pressure Intensity

Spatial Pulse Length (SPL)

SPL is the length of space over which a single pulse occurs, and is defined as

SPL = n •

where n is the number of cycles in the pulse and is the wavelength.

NOTEAn important parameter when considering axial resolution